Limb Perfusion Device

Similar documents
Inflatable Standing Aid Device

1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

Clinical Engineering. Equipment (Heart-lung machines)

Ergonomic Handle for a 2DoF Robotic Hand Rehabilitation Device

A complete family of Pediatric Perfusion Systems

Discussion and guidance on the definition and qualification of porous loads

Sampling Equipment Catalog. MicroPurge Low-Flow Pump Control

Treated w/ mildew and UV inhibitors to keep its beauty through all seasons

Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System

Procedures for operation of the TA Instruments DSC

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.

Project Number: P17453

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT

The next generation of Blood and Fluid Warming

Development of a High Pressure, Oil Free, Rolling Piston Compressor

BIOSTAT CultiBag RM Culturing Convenience

MECH690 MECHANICAL DESIGN PROJECT, SUMMER 2010 MID-SEMESTER REPORT INTEGRATED CLIP CARRYING CASE-M240B PROJECT MANAGER: SIGNATURE: DATE:

SERIES 20 CHAMBERS THE RC-25 AND RC-25F CHAMBERS ASSEMBLY

The next generation of Blood and Fluid Warming

COVER PHOTO: JONAS DAHM / DEEP SEA PRODUCTIONS VERSATILE REBREATHERS HANDCRAFTED IN DENMARK

Design of an Automated Perfusion Pump James J Higham

KEY CONCEPTS AND PROCESS SKILLS. 1. Creating models is one way to understand and communicate scientific information.

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

INSPIRE Min.I. Redefining minimally invasive approach to perfusion. Expand your choices with INSPIRE Min.I.

User Guide. Pall Laboratory Manifold. For laboratory use. Not for use in a manner other than indicated. Introduction. Regulatory References

Solutions for transcutaneous monitoring

PHOTO: JONAS DAHM / DEEP SEA PRODUCTIONS VERSATILE REBREATHERS FOR SKILLED DIVERS

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Solar Boat Hydrofoils

Tourniquet System. Ideal for single or bi-lateral orthopaedic surgery and pain management

Flight Medical presents the F60

Recommendations on Checking Anaesthesia Delivery Systems

NEW: CO2 incubator with sterilizable sensor

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

Test design for Tyco pressure relief valve

INSTALLATION AND OPERATION INSTRUCTIONS

Pressure measurement Applying remote seal, isolation diaphragms between process fluids and transmitters

Model # HC-MIS INSTRUCTION MANUAL

WARNING: EXPLOSION HAZARD

Head Impact Analysis Validation for Aluminium Bonnet

Barrier Development and Evaluation Methodology. D.S. Musgrave 1 1 Thermal Visions, Inc., Granville, USA

fabian Therapy evolution Our 2-in-1 device

Physiological Considerations for Compression Bandaging

PICU Resident Self-Study Tutorial The Basic Physics of Oxygen Transport. I was told that there would be no math!

UNITED STATES MARINE CORPS FIELD MEDICAL TRAINING BATTALION Camp Lejeune, NC

Traysealer T 100. Tray packaging in the smallest space

Next Generation Life Support (NGLS): Variable Oxygen Regulator Element

Unit II Problem 4 Physiology: Diffusion of Gases and Pulmonary Circulation

ACCU-PULSE Installation and Operation Instructions

Ranger Walking Initiation Stephanie Schneider 5/15/2012 Final Report for Cornell Ranger Research

Specifications...Easy CLIA Classification: Sample Type: Blood gas analyzer operation has never been simpler Sample Size: Measured Parameters PCO2

Name of product: Gripper for vertical façade FTV panels Types: PVF (60, 80, 100, 120, 133, 150, 172, 200 and 240 mm) INSTRUCTIONS FOR USE

Medical Instruments in the Developing World

Vehicle-mounted meters, pump supplied

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

InFlow Modular perfusion bioreactor

MSD Project Risk Assessment Template

In Vessel Composter Installation & Commissioning Manual

HONG KONG COLLEGE OF ANAESTHESIOLOGISTS TECHNICAL GUIDINES RECOMMENDATIONS ON CHECKING ANAESTHESIA DELIVERY SYSTEMS

CHAPTER 18 UNIVERSITY OF WISCONSIN- MILWAUKEE

Competency Record for Sonic Cleaning

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

Vibration isolation system 1VIS10W. User manual

Optimal Designs. A System to Quantify Heart Valve Leaflet Strain

Redesign - ATMOS S 201 Thorax. Digital thoracic drainage systems. ATMOS Thorax MOBILITY SAFETY COMFORT FLEXIBILITY

Pneumatic spool valve islands

Transforming Preservation

Degassing ERC KK Saitama, Japan

SERIES 20 CHAMBERS THE RC-20 AND RC-20H CHAMBER ASSEMBLY

Operator s Manual. with Maintenance Schedule. Fourth Edition Second Printing Part No

erc-o 2 -tech Ò innovative gas systems Markwerbener Str Weißenfels Germany

FLOW CALIBRATION RIG Mobile Calibration of In-line Flow Meters

Immerse yourself in nature

973-SF 6 Analyzer. Precise and Stable SF 6 Gas Analyzer REFLECTING YOUR STANDARDS

B. Assemble the Method-25 sample train with a heated probe/filter.

INSTRUCTIONS FOR CONTINUED AIRWORTHINESS AND EQUIPMENT MAINTENANCE MANUAL FOR OXYGEN BOTTLE MOUNT

This test shall be carried out on all vehicles equipped with open type traction batteries.

Air Plus User Manual

Start-up of the Three-shaft Closed Loop

Keep the balance in sight

Steam penetration in steam sterilization processes

Improving distillation tower operation

KTM OM-2 SPLIT BODY FLOATING BALL VALVES INSTALLATION AND MAINTENANCE INSTRUCTIONS

SERIES 20 CHAMBERS THE RC-26, RC-26P AND RC-26G CHAMBERS ASSEMBLY

Mass Spec will not Autotune

Replace Old Obsolete Systems!

Precision Liquid Settlement Array Manual

AUTOMATIC HOSE TEST UNIT, TYPE SPU

TIGHTNESS. Glass sealing Thanks to our glass-sealing technology, ODU products can meet the most demanding tightness requirements.

McMurdo S20 & Kannad Marine R10 AIS MOB Devices

IEC C LASSIFICATIONS:

LifeChoice Activox Portable oxygen concentrator. Portable oxygen therapy for patients with active lifestyles. ResMed.com

Alfred ECMO Circuit Priming and Storage Guide

3. Real-time operation and review of complex circuits, allowing the weighing of alternative design actions.

2015 Products Catalog Product Catalog. Products designed for the treatment and prevention of wounds

Mass Flowmeters. Mass Flowmeters for Gases

Instruction Sheet. Important Note: Information that requires special emphasis. Note: Information that supplements points in the main text.

Application of vacuum grease

AC : ACCURATE CRYOCHAMBER FOR A SMALL LABORATORY WITH SMALL BUDGET

May 2016 Med Tips. Helpful Hints and FAQs. Replacing AirSep VisionAire Mixing Tank

Transcription:

1 Limb Perfusion Device Design Team Peter Colby, Eric Greenberg, Nicholas Soo Hoo, Keith Travis Design Advisor Prof. Jeffrey Ruberti Abstract Every year thousands of people lose part or all of their upper extremities after having them severed in a traumatic incident. The problem arises as time elapses and the limb becomes ischemic as distal tissues begin to die. Currently the standard procedure for preserving the limb is to put it on ice and transport it to the nearest facility. This practice is not only inefficient, but limits tissue life to four hours. One way to keep the tissue alive is to constantly supply the tissue with oxygen, nutrients and medication via a perfusate. The goal of this project is to design a portable perfusion device that will keep a severed limb viable for up to 12 hours to allow for transplantation and replantation. Oxygenation of the tissue can be accomplished using an oxygenator which removes CO 2 from and adds O 2 to the perfusion solution. The oxygen is then transferred to the limb in a perfusion loop where O 2 is exchanged for CO 2 before exiting the arm. Calculations were performed to find the perfusate and oxygen flow rates required to sustain all metabolic functions. A peristaltic pump will be used to move the fluid throughout the system and a system of sensors will monitor the system for temperature, pressure, flow rate and gas concentrations. The temperature of the system will be kept between 4-10 o C to decrease metabolic requirements. The system will also include a bubble trap, filters, oxygen tank and cooling loop to ensure maximum performance and safety. A container and tray were designed to hold all of the components and the arm. The packaging and insulation were optimized to fit everything into the smallest possible space. Once these designs have been finalized and the parts assembled testing will be necessary to assure the device works.

2 The Need for Project A device that will extend the life The standard practice for keeping a severed limb alive is to of a severed limb to allow for a place it in a cooler of ice en route to the hospital. This technique only higher likelihood of reattachment preserves the limb for four hours by lowering its metabolic needs, but onto a patient. does not oxygenate or prevent ischemia. This short time frame lowers the chances for successful replantation or transplant. Creating a device that can prolong the life of the limb would improve the quality of life for many upper extremity amputee victims. Making the device portable and lightweight would allow it to be used in the field by medics, emergency staff and military personnel. The Design Project Objectives and Requirements The device must be able to keep Design Objectives a severed limb viable for up to The goal is to extend the life of the limb to 12 hours to allow 12 hours. It must also maintain for viable reattachment. The device needs to be portable and oxygenate portability standards and operate a severed limb by means of machine perfusion. The portability will be on battery power. covered by military standard MIL-HDBK-759C which states the size cannot exceed 485mmX 460mm X255mm and must weigh less than 11.2 kg. The size constraints created by the size of an arm make this a difficult objective to meet. The device will closely mimic the biologic functions of the human body by operating at rates calculated to meet the metabolic requirements of the arm. Design Requirements There are several requirements that must be met to ensure the limb is oxygenated and protected for the duration of the trip. Temperature of the fluid and arm will range from 4 to 37 o C but will optimally run at lower temperatures in order to decrease metabolic requirements. The flow rate of the perfusate must be between 300-500 ml/min and required oxygen consumption rates are 2-21 ml/min dependent on temperature. These specifications ensure the arm will experience a maximum pressure of 200mmHg which is the limit before damage will occur to the tissue. Design Concepts Considered Several concepts were The overall schematic and order of components of the considered with different perfusate system were determined by physiological requirements of the packaging and interface designs. arm. With this in mind it was crucial to find a way to optimize the packaging and minimize the overall dimensions of the container. The

3 container is a purchased part where all of the individual parts will be held and connected to the limb. Backpack Concept (see Figure left) The backpack concept uses a bag style container to hold all of the individual components of the perfusate, oxygen and cooling loops. They are affixed to the bag with straps and stitching. The bag can be opened and closed using a zipper and can also be insulated to reduce thermal leak. An additional case is needed for the arm as well as a way to connect the two compartments. Upright Box Concept For this concept all of the components are placed inside of an aluminum box with side handles and a hinged lid. The particular design allows all of the parts to be placed on the bottom of the box with room for insulation and a tray where the arm rests. Because the arm can fit into the box it reduces the need for an additional container. The components are secured to the box with standoffs and are packaged so that all of the parts are easily accessible. Insulated Concept (see Figure left) The insulated concept is an insulated container similar to a cooler that would house all of the system components with the exception of the arm. Similarly to the backpack concept, the limb would have to be placed in a separate container with an interface to attach the arm to the device. The particular container found for this concept is a box that is taller than it is wide and deep. Because of this the components have to be stacked one on top of the other in order to fit everything. Recommended Design Concept The recommended design is an Three different concepts were explored for packaging and aluminum box that can hold all of limb interface designs. All of these designs were feasible, but one the device components and a tray solution was thought to be the most practical and intuitive for use in a to hold the arm in place. medical emergency setting. Design Descriptions The overall dimensions of the aluminum container are 21 x 21 x 15. The box has two side handles and a hinged lid that can be latched closed. There is sufficient interior space to hold all of the components and arrange them on the bottom of the box. Various

4 standoffs and fasteners are used to secure the parts to ensure no hardware is damaged. Because the container is not insulated, closed cell foam will be used to thermally isolate the battery and pump from the arm and remaining parts. The arm is secured on an autoclavable tray, shown to the left, which is fitted to the top of the box with brackets. It can be removed to access the other components of the system and the lid can still be closed while the device is running. The tray has a built-in drain where the returning venous fluids are collected and returned to the perfusate loop with the aid of gravity. Having the limb and device components in one container makes the arm easier to attach to the device. Analytical Investigation The average metabolic rates for flow and oxygen consumption were found to be 350 ml/min and 21 ml/min respectively. A temperature curve was created for maximum flow rate and required oxygen delivery at lower temperatures. Using this curve and Henry s Law, a correlation was determined for partial pressure of oxygen in Perfadex at low temperatures. At 10 o C the required p0 2 in Perfadex delivered to the arm is approximately 10 kpa. In order to optimize the device and choose the most logical components, the individual systems and parts were researched. Trade studies were performed for pumps, oxygenators, filters and sensors to help decide which parts to order. Before purchasing any of the components it was important to verify that all of the parts would work correctly together and not fail when subjected to a load. The first step was to create SolidWorks models of the container and all of the components to show the parts could fit inside of the box. Once this was confirmed finite element analysis was applied to the selected standoffs and attaching hardware to make sure none of the components would fail during operation. Key Advantages of Recommended Concept The upright box container concept has several advantages over the other design considerations. Unlike the backpack and insulated container, the upright box allows all of the components to be placed on one layer of the box. This provides easy access to all of the parts if one needs to be replaced or repaired. It is also beneficial

5 because the arm is able to fit on a tray at the top of the box. This eliminates the need for multiple containers and greatly reduces the design work required for the arm interface. Financial Issues The device is built for medical testing and not mass production, so keeping costs down was not a primary objective. This device is currently being built for the Plastic Surgery team at Brigham & Women s Hospital located in Boston, MA. Because this will be used for medical testing, it is not necessary to design for massproduction. The hospital is responsible for the cost of all purchased parts. There are few parts that require machining and most of the parts can be purchased from various medical vendors. Many of the components come in contact with the perfusate and those that cannot be autoclaved or sterilized must be thrown away. This increases the overall cost of the device because multiple quantities must be ordered. The overall cost of one assembled device is approximately $5,100. Recommended Improvements With more time and resources Because a lot of time was spent on research and further development would allow physiological design, some of the components were over engineered for a better optimized system. or not optimized. For instance, some of the packaging hardware and fasteners were chosen based on their availability and ease of installment. With more time, these parts would be designed and machined to provide the greatest performance in the smallest space. Ultimately, if there was more time and resources available, the device would be made more custom. In terms of the container, packaging and sensors, everything would be a lot more streamlined and optimized. Another improvement that could be made is to incorporate the sensors and digital readouts into part of the container packaging. Currently, the information received by the sensors is displayed on a computer using LabView software. By creating a separate interface for these sensors it would eliminate the need for a computer to run the device.