FRG-700 Inverted Magnetron Pirani Gauge

Similar documents
Inverted Magnetron Pirani Gauge

Cold Cathode Pirani Gauge

Cold Cathode Pirani Gauge

Operating Instructions

Contents Safety Technical Data Installation Operation Deinstallation Maintenance, Repair Returning the Product Disposal Further Information

Ambient Capacitance Gauge

Operating Manual VSA100A. tina29e1 ( )

Pirani Capacitance Diaphragm Gauge

OPERATING INSTRUCTIONS

Capacitance Diaphragm Gauge

Operating Instructions

Capacitance Diaphragm Gauge

Product Identification 7 Validity 7 Intended Use 9 Functional Principle 9 Patents 9 Scope of Delivery 9

Capacitance Diaphragm Gauge

InstruTech. Cold Cathode Pirani Combination Vacuum Gauge WGM701 Module The Wasp. User Manual. InstruTech 1475 S. Fordham St. Longmont, CO USA

InstruTech. CDM900 Module The Micro Bee. User Manual. InstruTech 1475 S. Fordham St. Longmont, CO USA

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Instruction Manual Contact Pressure Vacuum Gauge

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

MANUAL KPS Pressure Control Valve

Instruction Manual BA90 Precision Barometer

DPC-30 DPC-100. Reference Manual

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

Model FS5001B VA.1. MEMS Mass Flow Sensors ISO 9001 ISO OHSAS 18001

DMK 458. Pressure Transmitter for Marine and Offshore. Ceramic Sensor. accuracy according to IEC 60770: standard: 0.25 % FSO option: 0.

BGA244 Binary Gas Analyzer

INSTALLATION & MAINTENANCE INSTRUCTIONS

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

INSTRUCTIONS. VP-100 Ion Pump. Model / R001

MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

CSA Sample Draw Aspirator Adapter Operator s Manual

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Norrsken Family Booklet

SDX Submersible Depth Transmitter User Manual

MAGICFX CONFETTIPISTOL USER AND INSTALLATION MANUAL

Instruction Manual EMA 84 Digital Pressure Gauge

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

Operating Instructions Temperature Measuring System MF420-5T-100

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

AA100 Acoustic Actuator

SDX Submersible Depth Transmitter User Manual

KTE6000 / KTU6000 Series OEM pressure transmitters for industrial media

I/P Transducer Positioner Module

92831 TEL: (714) FAX:

Digital Vacuum Regulator

Installation and Operation Manual

Type 3709 Pneumatic Lock-up Valve. Translation of original instructions. Mounting and Operating Instructions EB 8391 EN

RK-IR Sample Draw Aspirator Adapter Operator s Manual

x act i Precision- Pressure Transmitter for Food Industry, Pharmacy and Biotechnology Stainless Steel Sensor

Digital Melting Point Apparatus

High-performance submersible pressure transmitter For level measurement Model LH-10

Nitrogen Supply System

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

Gas density monitor With integrated transmitter Model GDM-100-TI

AMS 6916 Board mount pressure sensor with ratiometric analog output

LK-SX CO 2 +VOC. Application. Security Advice Caution. Notes on Disposal

Instruction Manual Differential Pressure Transmitter

! Warning, refer to accompanying documents.

HAYWARD FLOW CONTROL TFS SERIES FLOW METER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

High-performance submersible pressure transmitter For level measurement Model LH-10

Target disk flowmeter Series DP. Instructions manual DP65 DP500. R-MI-DP Rev.: 4 English version

Instruction Manual for Intrinsically Safe - Reed Switches

RGC-IR Remote Gas Calibrator for IR400

ASDX Series Silicon Pressure Sensors

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

Transmitter CS 21 Operation Manual

* * Data Sheet and Operating Manual. Digital manometer ME01 ## # 87 # HL S#### Dust explosion protection Zone 22.

HYDROVEX TTT Membrane Flow Regulator

RK LEL Sample Draw Aspirator Adapter Operator s Manual

QBS Electronic Pressure Regulator

URC Voltage Sensor SEN-VOLT for use with MRX units containing sensor ports

INSTRUCTIONS FOR FM4460N SERIES HIGH PRESSURE, METAL TUBE FLOWMETERS

Installation, Operation and Maintenance Manual. TSV Temperature Shut-off Valve SPD Rev. 6 10/07

English. Introduction. Safety Instructions. All Products. Inspection and Maintenance Schedules. Parts Ordering. Specifications WARNING WARNING

Type Operating Instructions. 2/2-way solenoid valve 2/2-Wege Magnetventil Electrovanne 2/2 voies.

A17W49_MASEM_F_EN MANUAL

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

LK-SX VOC. Application. Security Advice Caution. Notes on Disposal. Combined sensor mixed gas. Data sheet

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

ETL Water Level Meter

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0

The Broadest Gauging Line in the Industry

Advanced Test Equipment Rentals ATEC (2832)

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

Operating Instructions. Ball valve fitting according to ZB For pressure transmitter VEGABAR 82. Document ID: 50027

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

FLOWMETER MODELS: 1MFA, 4MFA, 6MFA and 8MFA Series SAVE THESE INSTRUCTIONS

TD-106-5D TD-106-5D-A Wind Direction Sensor User s Manual

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

Ambient Weather WS-03 Thermo-Hygrometer

AMS 2710 PCB pressure sensor module with V output

275 series Convection Vacuum Gauge

InstruTech. Cold Cathode Ionization Vacuum Gauge CCM500 Module The Hornet. User Manual. InstruTech 1475 S. Fordham St. Longmont, CO USA

Transmitter CS 21 Operation Manual

CF-SERIES HIGH FLOW REGULATORS

Transcription:

vacuum technologies FRG-700 Inverted Magnetron Pirani Gauge SHORT OPERATING INSTRUCTIONS Manual No. tqma48e1 Revision -- November 2008

FRG-700 Inverted Magnetron Pirani Gauge Copyright 2008 Vacuum Technologies 2

Warranty Products manufactured by Seller are warranted against defects in materials and workmanship for twelve (12) months from date of shipment thereof to Customer, and Seller s liability under valid warranty claims is limited, at the option of Seller, to repair, to replace, or refund of an equitable portion of the purchase price of the Product. Items expendable in normal use are not covered by this warranty. All warranty replacement or repair of parts shall be limited to equipment malfunctions which, in the sole opinion of Seller, are due or traceable to defects in original materials or workmanship. All obligations of Seller under this warranty shall cease in the event of abuse, accident, alteration, misuse, or neglect of the equipment. In-warranty repaired or replaced parts are warranted only for the remaining unexpired portion of the original warranty period applicable to the repaired or replaced parts. After expiration of the applicable warranty period, Customer shall be charged at the then current prices for parts, labor, and transportation. Reasonable care must be used to avoid hazards. Seller expressly disclaims responsibility for loss or damage caused by use of its Products other than in accordance with proper operating procedures. Except as stated herein, Seller makes no warranty, express or implied (either in fact or by operation of law), statutory or otherwise; and, except as stated herein, Seller shall have no liability under any warranty, express or implied (either in fact or by operation of law), statutory or otherwise. Statements made by any person, including representatives of Seller, which are inconsistent or in conflict with the terms of this warranty shall not be binding upon Seller unless reduced to writing and approved by an officer of Seller. 3

Warranty Replacement and Adjustment All claims under warranty must be made promptly after occurrence of circumstances giving rise thereto, and must be received within the applicable warranty period by Seller or its authorized representative. Such claims should include the Product serial number, the date of shipment, and a full description of the circumstances giving rise to the claim. Before any Products are returned for repair and/or adjustment, written authorization from Seller or its authorized representative for the return and instructions as to how and where these Products should be returned must be obtained. Any Product returned to Seller for examination shall be prepaid via the means of transportation indicated as acceptable by Seller. Seller reserves the right to reject any warranty claim not promptly reported and any warranty claim on any item that has been altered or has been returned by non-acceptable means of transportation. When any Product is returned for examination and inspection, or for any other reason, Customer shall be responsible for all damage resulting from improper packing or handling, and for loss in transit, not withstanding any defect or non-conformity in the Product. In all cases, Seller has the sole responsibility for determining the cause and nature of failure, and Seller s determination with regard thereto shall be final. If it is found that Seller s Product has been returned without cause and is still serviceable, Customer will be notified and the Product returned at Customer s expense; in addition, a charge for testing and examination may be made on Products so returned. 3/1/00 4

Product Identification In all communications with VARIAN, please specify the information given on the product nameplate. For convenient reference copy that information into the space provided below. VARIAN Lexington MA 02421 USA Model: PN: SN: V W Validity This document applies to products with the following part numbers: FRG700KF25 (DN 25 ISO-KF) FRG700CF35 (DN 40 CF-F) The part number (PN) can be taken from the product nameplate. If not indicated otherwise in the legends, the illustrations in this document correspond to the gauge with vacuum connection DN 25 ISO-KF. They apply to the gauge with other vacuum connection by analogy. We reserve the right to make technical changes without prior notice. All dimensions in mm. 5

Intended Use The Inverted Magnetron Pirani Gauge FRG-700 has been designed for vacuum measurement in the pressure range of 5 10-9 1000 mbar. The Inverted Magnetron Pirani Gauge must not be used for measuring flammable or combustible gases which react in air. The gauge can be operated in connection with a VARIAN FRG Control Unit, a VARIAN Turbo AG Rack Controller, or with another controller. Functional Principle The gauge consists of two separate measurement systems (Pirani and cold cathode system) the signals of which are combined in such a way that one measurement signal is output. The Pirani measurement circuit is always on. 6

Contents Warranty 3 Warranty Replacement and Adjustment 4 Product Identification 5 Validity 5 Intended Use 6 Functional Principle 6 1 Safety 8 1.1 Symbols Used 8 1.2 Personnel Qualifications 8 1.3 General Safety Instructions 9 2 Technical Data 10 3 Installation 17 3.1 Vacuum Connection 17 3.2 Electrical Connection 21 4 Operation 22 4.1 Gas Type Dependence 22 4.2 Ignition Delay 22 4.3 Adjusting the Gauge 23 5 Deinstallation 26 6 Maintenance, Troubleshooting 27 7 Returning the Product 28 8 Disposal 28 9 Literature 29 EC Declaration of Conformity 30 For cross-references within this document, the symbol ( XY) is used, for cross-references to further documents, listed under literature, the symbol ( [Z]). 7

1 Safety 1.1 Symbols Used DANGER Information on preventing any kind of physical injury. WARNING Information on preventing extensive equipment and environmental damage. Caution Information on correct handling or use. Disregard can lead to malfunctions or minor equipment damage. Notice 1.2 Personnel Qualifications Skilled personnel All work described in this document may only be carried out by persons who have suitable technical training and the necessary experience or who have been instructed by the end-user of the product. 8

1.3 General Safety Instructions Adhere to the applicable regulations and take the necessary precautions for the process media used. Consider possible reactions with the product materials. Consider possible reactions (e.g. explosion) of the process media due to the heat generated by the product. Adhere to the applicable regulations and take the necessary precautions for all work you are going to do and consider the safety instructions in this document. Before beginning to work, find out whether any vacuum components are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts. DANGER DANGER: magnetic fields Strong magnetic fields can disturb electronic devices like heart pacemakers or impair their function. Maintain a safety distance of 10 cm between the magnet and the heart pacemaker or prevent the influence of strong magnetic fields by antimagnetic shielding. Communicate the safety instructions to all other users. 9

2 Technical Data Measuring range (air, N 2 ) Accuracy Repeatability 5 10-9 1000 mbar ±30% (in the range 1 10-8 100 mbar) ±5% (in the range 1 10-8 100 mbar) Output signal (measuring signal) Voltage range 0 +10.5 V Measurement range 1.82 8.6 V Voltage vs. pressure logarithmic, 0.6 V/decade Error signal <0.5 V no supply >9.5 V Pirani sensor defective (filament rupture) Output impedance Minimum loaded impedance Response time p > 10-6 mbar p = 10-8 mbar Identification gauge 2 10 Ω 10 kω, short-circuit proof (pressure dependent) <10 ms 1000 ms 85 kω, referenced to supply common Status pin 6 p > 10-2 mbar Low = 0 V Pirani-only mode p < 10-2 mbar Cold cathode not ignited Low = 0 V Pirani-only mode p < 10-2 mbar Cold cathode ignited High = 15 30 VDC Combined Pirani / cold cathode mode LED High voltage on (LED on) 10

Supply DANGER The gauge may only be connected to power supplies, instruments or control devices that conform to the requirements of a grounded protective extralow voltage (SELV-E according to EN 61010). The connection to the gauge has to be fused 1). Voltage at the gauge Power consumption Fuse 1) Voltage at the supply unit with maximum cable length Adjustment Potentiometer <HV> Potentiometer <ATM> 15 30 VDC (ripple 1 V pp ) 2 W 1 AT 16 30 VDC (ripple 1 V pp ) 2) adjustment under 10-4 mbar adjustment at atmospheric pressure Electrical connection FCC68 socket, 8 poles Sensor cable 8 poles, shielded Line length 50 m (8 0.14 mm²) Operating voltage 3.3 kv Operating current 500 µa Grounding concept Vacuum connection measuring common Supply common signal common "Electrical Connection" connected via 10 kω (max. voltage differential with respect to safety ±50 V accuracy ±10 V) conducted separately 1) VARIAN controllers fulfill this requirement. 2) The minimum voltage of the power supply unit must be increased proportionally to the length of the sensor cable 11

Materials on the vacuum side Vacuum connection Measurement chamber Feedthrough isolation Internal seal Anode Ignition aid Pirani measurement tube Pirani filament Internal volume Pressure stainless steel stainless steel ceramic FPM 75 Mo stainless steel Ni, Au W 20 cm³ 10 bar (absolute), limited to inert gases Temperatures Operation Bakeout Pirani filament Storage Relative humidity Mounting orientation Use Type of protection IP 40 +5 +55 C 150 C (without electronics and magnetic shielding) 120 C -40 +65 C 80% at temperatures +31 C decreasing to 50% at +40 C any indoors only, altitude up to 2000 m 12

Dimensions [mm] 92 28 55 60 18 75 ø 63.5 DN 25 ISO-KF 24 DN 40 CF-F Weight FRG700KF25 FRG700CF35 700 g 980 g 13

Measuring Signal vs. Pressure Pressure p 1E+04 1E+02 p = 10 1.667U-d Pa mbar torr 1E+00 1E 02 1E 04 sensor error underrange overrange sensor error 1E 06 1E 08 0.00.5 1.5 2.5 3.5 4.5 5.5 6.5 7.5 8.5 9.5 10.5 U [V] Measurement signal U [V] p = 10 1.667 U-d U = c + 0.6log 10 p mbar Pa Torr d 11.33 9.33 11.46 c 6.8 5.6 6.875 valid in the range 5 10-9 mbar <p< 1000 mbar 3.8 10-9 Torr <p< 750 Torr 5 10-7 Pa <p< 1 10 5 Pa 14

Gas Type Dependence p (mbar) 10 2 8 6 4 2 10 1 8 6 4 2 10 0 8 6 4 2 Indication range >10-2 mbar) H 2 He Ne Air O2 CO N2 CO 2 Ar Freon 12 Kr Xe 10 1 8 6 4 2 10 2 Water vapor 10 3 2 4 6 10 2 2 4 6 10 1 2 4 6 10 0 2 4 6 10 1 2 4 6 10 2 p eff (mbar) 15

p (mbar) 10 1 8 6 4 2 10 2 8 6 4 Indication range 10-6... 10-1 mbar 2 10 3 8 6 4 2 10 4 8 6 4 2 10 5 8 6 4 He Ar Air N2 O2 CO 2 10 6 10-6 2 4 6 10-6 2 4 6 10 4 2 4 6 10-3 2 4 6 10-2 2 4 6 10-1 p eff (mbar) In the range below 10-5 mbar, the pressure indication is linear. For gases other than air, the pressure can be determined by means of a simple conversion formula: p eff = K pressure reading Where Gas type K (mean values) Air (O 2, CO, N 2 ) Xe Kr Ar H 2 Ne He 1.0 0.4 0.5 0.8 2.4 4.1 5.9 16

3 Installation 3.1 Vacuum Connection DANGER DANGER: overpressure in the vacuum system >1 bar Injury caused by released parts and harm caused by escaping process gases can result if clamps are opened while the vacuum system is pressurized. Do not open any clamps while the vacuum system is pressurized. Use the type clamps which are suited to overpressure. DANGER DANGER: overpressure in the vacuum system >2.5 bar KF flange connections with elastomer seals (e.g. O-rings) cannot withstand such pressures. Process media can thus leak and possibly damage your health. Use O-rings provided with an outer centering ring. 17

DANGER DANGER: protective ground Products that are not correctly connected to ground can be extremely hazardous in the event of a fault. Electrically connect the gauge to the grounded vacuum chamber. This connection must conform to the requirements of a protective connection according to EN 61010: CF flanges fulfill this requirement. For gauges with a KF flange, use a conductive metallic clamping ring. Caution Caution: vacuum component Dirt and damages impair the function of the vacuum component. When handling vacuum components, take appropriate measures to ensure cleanliness and prevent damages. Caution Caution: dirt sensitive area Touching the product or parts thereof with bare hands increases the desorption rate. Always wear clean, lint-free gloves and use clean tools when working in this area. 18

WARNING WARNING: electric arcing Helium may cause electric arcing with detrimental effects on the electronics of the product. Before performing any tightness tests put the product out of operation and remove the electronics unit. The gauge may be mounted in any orientation. To keep condensates and particles from getting into the measuring chamber preferably choose a horizontal to upright position and possibly use a seal with a centering ring and filter. If adjustment should be possible after the gauge has been installed, be sure to install it so that potentiometers <HV> and <ATM> can be accessed with a screwdriver ( "Adjusting the Gauge"). When making a CF flange connection, it can be advantageous to temporarily remove the electronics and the magnet unit ( Instruction Manual tqna48e1). 19

Remove the protective lid and connect the product to the vacuum system. Seal with centering ring or Protective lid Clamp Seal with centering ring and filter Keep the protective lid. 20

3.2 Electrical Connection Make sure the vacuum connection is properly made ( "Vacuum Connection"). If no sensor cable is available, make one according to the following diagram. 10 k Signal 10 10 Ident 6 3 5 4 1 2 + + 1 Electrical connection Pin 1 Supply (15 30 VDC) Pin 2 Supply common Pin 3 Signal output (measuring signal) Pin 4 Identification Pin 5 Signal common Pin 6 Status Pin 7, 8 n.c. 8 8 poles FCC-68 Connect the gauge to the controller using the sensor cable. 21

4 Operation When the supply voltage is applied, the measuring signal is available between pins 3 and 5. Over the whole measurement range, the measuring signal is output as a logarithm of the pressure (measuring signal vs. pressure "Technical Data"). Allow for a stabilizing time of 10 minutes. Once the gauge has been switched on, permanently leave it on irrespective of the pressure. The Pirani measurement circuit is always on. The cold cathode measurement circuit is controlled by the Pirani circuit and is activated only at pressures <1 10-2 mbar. 4.1 Gas Type Dependence The measurement value depends on the type of gas being measured. The value displayed is accurate for dry air, O 2, CO and N 2. It can be mathematically converted for other gases ( "Technical Data"). If the gauge is operated in connection with an VARIAN vacuum gauge controller, a calibration factor can be entered for correction of the reading. 4.2 Ignition Delay When cold cathode measurement systems are activated upon switching the gauge on, an ignition delay occurs, which is typically: 10-5 mbar 1 second 10-7 mbar 20 seconds 5 10-9 mbar 2 minutes As long as the cold cathode measurement circuit has not yet ignited, the measurement value of the Pirani is output as measuring signal ("Pirani underrange" is displayed for pressures <5 10-4 mbar). 22

4.3 Adjusting the Gauge The gauge is factory-calibrated. If used under different climatic conditions, through extreme temperatures, aging or contamination, and after exchanging the sensor, the characteristic curve can be offset and readjustment may become necessary. The cold cathode measurement circuit, which is dominant for low pressures (<1 10-3 mbar), is factory-calibrated. By way of contrast, the Pirani measurement circuit can be adjusted. Any adjustment has a negligible effect on the pressure range between approx. 10-2 mbar and 10 2 mbar. If you are using a seal with centering ring and filter, check that they are clean or replace them if necessary ( "Deinstallation"). Activate the gauge. Evacuate it to p << 10-4 mbar and wait at least 10 minutes. Turn the nameplate counter-clockwise until the mechanical stop is reached. 23

While depressing the tactile switch with a cylindrical pin (ø 3 mm), adjust the <HV> potentiometer by means of a 1.5 mm screwdriver to 4.20 V or to 5 10-4 mbar Vent the gauge with air or nitrogen to atmospheric pressure, and wait at least 10 minutes. Turn the nameplate clockwise until the mechanical stop is reached. 24

Using the 1.5 mm screwdriver, adjust the <ATM> potentiometer to 8.60 V or to 1 10 3 mbar Turn the nameplate back to its original position (it catches). 25

5 Deinstallation DANGER DANGER: contaminated parts Contaminated parts can be detrimental to health and environment. Before beginning to work, find out whether any parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts. Caution Caution: vacuum component Dirt and damages impair the function of the vacuum component. When handling vacuum components, take appropriate measures to ensure cleanliness and prevent damages. Caution Caution: dirt sensitive area Touching the product or parts thereof with bare hands increases the desorption rate. Always wear clean, lint-free gloves and use clean tools when working in this area. Vent the vacuum system. 26

Put the gauge out of operation and unplug the sensor cable. Remove the gauge from the vacuum system and place the protective lid. When deinstalling a CF flange connection, it can be advantageous to temporarily remove the electronics and the magnet unit ( Instruction Manual tqna48e1). 6 Maintenance, Troubleshooting Instruction Manual tqna48e1 If operated at high pressures or under dirty conditions, the gauge must be regularly cleaned. Gauge failures due to contamination, as well as expendable parts (filament), are not covered by the warranty. 27

7 Returning the Product WARNING WARNING: forwarding contaminated products Contaminated products (e.g. radioactive, toxic, caustic or microbiological hazard) can be detrimental to health and environment. Products returned to VARIAN should preferably be free of harmful substances. Adhere to the forwarding regulations of all involved countries and forwarding companies and enclose a duly completed declaration of contamination. Products that are not clearly declared as "free of harmful substances" are decontaminated at the expense of the customer. Products not accompanied by a duly completed declaration of contamination are returned to the sender at his own expense. 8 Disposal DANGER DANGER: contaminated parts Contaminated parts can be detrimental to health and environment. Before beginning to work, find out whether any parts are contaminated. Adhere to the relevant regulations and take the necessary precautions when handling contaminated parts. 28

WARNING N WARNING: substances detrimental to the environment Products or parts thereof (mechanical and electric components, operating fluids etc.) can be detrimental to the environment. Dispose of such substances in accordance with the relevant local regulations. Separating the components After disassembling the product, separate its components according to the following criteria: Contaminated components Contaminated components (radioactive, toxic, caustic or biological hazard etc.) must be decontaminated in accordance with the relevant national regulations, separated according to their materials, and disposed of. Other components Such components must be separated according to their materials and recycled. 9 Literature [1] www.varianinc.com Instruction Manual Inverted Magnetron Pirani Gauge FRG-700 tqna38e1 Varian Vacuum Technologies, MA 02421, USA 29

EC Declaration of Conformity We, VARIAN, hereby declare that the equipment mentioned below complies with the provisions of the Directive relating to electrical equipment designed for use within certain voltage limits 2006/95/EC and the Directive relating to electromagnetic compatibility 2004/108/EC. FRG-700 Inverted Magnetron Pirani Gauge Part numbers FRG700KF25 FRG700CF35 Standards Harmonized and international/national standards and specifications: EN 61000-6-2 (EMC: generic immunity standard) EN 61000-6-3 (EMC: generic emission standard) EN 61010-1 (Safety requirements for electrical equipment for measurement, control and laboratory use) Signatures Varian Vacuum Technologies, USA 9 December 2008 Frederick C. Campbell Operations Manager 30

Notes 31

Original: English Varian Vacuum Technologies 121 Hartwell Avenue Lexington, MA, 02421 USA Tel: (781) 861 7200 Fax: (781) 861 5437 custserv@varianinc.com t qma48e1 www.varianinc.com