REPRODUCIBILITY IN OPTICAL THIN FILM PROCESSING PART 1, THE VACUUM AND PUMPING

Similar documents
Introduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.

EXERCISE 2: THE VACUUM SYSTEM

Vacuum Systems and Cryogenics for Integrated Circuit Fabrication Technology 01

BASICS of MAINTENANCE and TROUBLESHOOTING of VACUUM DIFFUSION PUMPS

The Principles of Vacuum Technology

Vacuum Science Techniques and Applications Dan Dessau Adv. Lab 2007

E SC 412 Nanotechnology: Materials, Infrastructure, and Safety Wook Jun Nam The Pennsylvania State University

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.

Revision 2013 Vacuum Technology 1-3 day Good Vacuum Practice 1 Day Course Outline

A NOVEL SENSOR USING REMOTE PLASMA EMISSION SPECTROSCOPY FOR MONITORING AND CONTROL OF VACUUM WEB COATING PROCESSES

Fig. 1: Mechanical vacuum gauge (left) and capacitive vacuum gauge (right)

Introductory Lab: Vacuum Methods

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

THERMAL EVAPORATION UNIT (for Al evaporation)

Matching Vacuum Pump to Process

Session 4B 11:00 11:30 AM. BASICS of MAINTENANCE and TROUBLESHOOTING of VACUUM DIFFUSION PUMPS

School of Chemistry SOP For Operation Of Glove Boxes

Leak Checking Large Vacuum Chambers

Standard Operating Manual

Operating Procedures for Metal Evaporator I

Another convenient term is gauge pressure, which is a pressure measured above barometric pressure.

Mass Spec will not pump down

Cryo-Evaporator Operation

Standard Operating Manual

Chapter 8: Cryo-sorption pumps

How To Use Getters and Getter Pumps

VACUUM TECHNIQUES. Why? How else might you make the measurement?

ASPECTS OF VACUUM TECHNOLOGY. (a) To see something of the operation of vacuum pumps; Then, we will use the vacuum system for one of two experiments:

The SPI Sputter Coater Handbook

Oxygen Contamination

Leaks, contamination and closed loop control how RGAs make coating processes more profitable

Plasma Sources and Feedback Control in Pretreatment Web Coating Applications

Determination of the Vacuum System for R2R by using a New Pumping Simulation Program

How To Choose a Roughing/Backing Pump for the Turbo and Drag Family

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

Air Eliminators and Combination Air Eliminators Strainers

Short Introduction to Cryo-Pumps and Refrigerators. Dr Graham Rogers Leybold UK Ltd.

INTRODUCTION: EXPERIMENTAL NOTES:

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

Remote Plasma Cleaning from a TEM Sample Holder with an Evactron De-Contaminator ABSTRACT

VS Series Leak Detectors for Vacuum Furnaces LEAK DETECTION FOR VACUUM FURNACES

Gases and Pressure SECTION 11.1

Both physical and chemical effects come into play either separately or in combination.

Approved by Principal Investigator Date: Approved by Super User: Date:

ELEVATION CORRECTION FACTORS FOR RADON MONITORS*

States of Matter Review

Copyright, 2005 GPM Hydraulic Consulting, Inc.

End of Chapter Exercises

Nordiko Metal Sputtering System Standard Operating Procedure

Characterizing a Cu/Mn Alloy for Extracting Oxygen from Inert Gas. Streams

Oxford Instruments Austin, Inc.

ANNEX AMENDMENTS TO THE INTERNATIONAL CODE FOR FIRE SAFETY SYSTEMS (FSS CODE) CHAPTER 15 INERT GAS SYSTEMS

The Discussion of this exercise covers the following points: Pumps Basic operation of a liquid pump Types of liquid pumps The centrifugal pump.

Inert Air (N2) Systems Manual

A/C SYSTEM GENERAL DIAGNOSTIC PROCEDURES

Lesson 6: Flow Control Valves

Instruction Manual. Digital Thermocouple Gauge Control Units Models DTC & DTC Models DTC-06M-115 & DTC-06M-230

Oxford Instruments Austin, Inc.

University of Minnesota Nano Center Standard Operating Procedure

End of Chapter Exercises

Application Note. ASTRON Remote Plasma Source Ignition Best Practices PROBLEM

Exercise 4-2. Centrifugal Pumps EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Pumps

Air entrainment in Dip coating under vacuum

High Performance Mass Spectrometers

Savannah S100 ALD at SCIF, UC Merced Standard operating Procedure

MS.RAJA ELGADY/PRESSURE PAPER 3

SERIES 30. Spring Operated Tank Blanketing Valve PROTECTOSEAL.

The Broadest Gauging Line in the Industry

SERIES 10. Pilot Operated Tank Blanketing Valve, 2" PROTECTOSEAL.

VACUUM CHAMBER MANUAL

ACCELERATOR VACUUM PROBLEMS: IN-LINE OIL TRAPPING BETWEgN THE STANFORD LINEAR ACCELERATOR AND THE BEAM SWITCHYARD

The Helium Leak Detector

MMRC PVD Evaporator Instructions

COMELEC C-30-S Parylene Coating System

WHP Operations and Methods { July William M. Smethie, Jr. Lamont Doherty Geological Observatory Palisades, New York U.S.A.

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

A/C SYSTEM DIAGNOSIS

Mass Spec will not Autotune

VACUUM REQUIREMENTS FOR LINACS P. Grand Yale University

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

Trillium US Inc. Cryo-Plex 10 Cryogenic High Vacuum Pump User s Manual Rev A / November 2015

Radar, Ultrasonic and RF Level Transmitters

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

Plasma Etching: Safety Operating Procedures

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

This educational seminar discusses creating, measuring, and troubleshooting Ultra-High Vacuum ( Torr) systems.

13.1!"#$#%"&'%()$*+%,+-.$+/*$#

Standard Operating Procedure. For. PVD E-Beam

Vacuum System at IUAC

OOC RGA. Ion Pump 400 l/s. 8" ports with baked viton o-ring seals. Welsh rotary roughing pump 650 l/s manual isolation valve.

Operating Procedures for the. SAMCO ICP RIE System

ONSITE PROVING OF GAS METERS. Daniel J. Rudroff WFMS Inc West Bellfort Sugar Land, Texas. Introduction

D R A F T. Operating Procedures for the NPDGamma Liquid Hydrogen Target at the BL 13. Version 1.00

Gases and Pressure. Main Ideas

EFFECTS OF BAFFLE SIZE ON PRESSURE DISTRIBUTION IN VACUUM CHAMBER DURING DYNAMIC GAS FLOW

Barrier Development and Evaluation Methodology. D.S. Musgrave 1 1 Thermal Visions, Inc., Granville, USA

XDF BURNERS DUAL FUEL EXCESS AIR BURNER FEATURES DESCRIPTION EXCESS AIR OPERATION

OLED vacuum deposition cluster system Sunicel Plus 400 / Plus 400L (Sunic Systems)

Transcription:

ABSTRACT REPRODUCIBILITY IN OPTICAL THIN FILM PROCESSING PART 1, THE VACUUM AND PUMPING Ronald R. Willey Willey Optical, Consultants, 13039 Cedar Street, Charlevoix, MI 49720, USA Ph 231-237-9392, ron@willeyoptical.com Variations in vacuum and pumping practices used in optical thin film coating processes can cause a lack of reproducibility and adhesion problems with the product if not properly handled. The quality of vacuum needed, mean free path effects, cross-over pressures, backstreaming, and checks for the health of the vacuum system are discussed and practical recommendations are made. INTRODUCTION In this series of short articles, we discuss ways to promote the reproducibility of processes for optical thin film coatings. One of the first things needed is an adequate vacuum for the process at hand. Variations in the quality of the vacuum can cause a lack of reproducibility in optical thin film results. Film density and index of refraction can vary if the pressure for the process is not reproducible. This could cause the spectral results and uniformity to vary from run to run. Adhesion failures may be due to oil backstreaming from both the mechanical pumps (MP) and diffusion pumps (DP). DPs will be addressed here for high vacuum pumps because they are in wide use around the world and are the most vulnerable to backstreaming. The pressure in a chamber is a measure of what vapor is still in it and what is competing at the substrate with the materials being deposited as a coating. The mean free path (MFP) is an important factor illustrating what is happening in the chamber on an atomic scale. If there were a small amount of residual gas in the chamber with which to collide, the evaporated molecules which collided with the gas would go in any and all directions and would probably also lose some of their energy (heat). If there were large amounts of gas, the molecules might actually lose so much energy by multiple collisions as to condense to solid particles before reaching any surface. Such high pressure depositions can result in powdery films or even dust which falls to the bottom of the chamber. We would like to assume that the vacuum in the chamber is such that the collisions are few and do not significantly affect the result as compared to a "perfect" vacuum. However, this is not always the case. HOW GOOD A VACUUM IS NEEDED? The MFP is the statistical average distance that a molecule will travel before colliding with another, at a given pressure and temperature. Equation 1 can be used to calculate the MFP at room temperature. 3 5.0 x 10 MFP = cm P Torr (1)

N N 0 = exp ( x MFP ) (2) In Eqn. 2, N is the number of molecules from a starting number N 0 that have not collided after a distance x. Figure 1 shows, from Eqn. 2, the fraction of uncollided atoms after a distance x measured in MFP. About 50% have not collided by 0.7 MFP, 36.8% by 1 MFP, and only 13.5% by 2 MFP. This implies that collisions in a one meter size box coater are not significant at 1 10 6 torr, where 98% would not have collided, but could be a factor to consider at 1 10 5 torr, where only 82% are uncollided. However, when the pressure is 1.5 10 4 torr, as might be used in reactive oxygen depositions of SiO 2 and TiO 2, only 5% of the molecules reach the other side of the chamber without collisions. In the case of uniformity masking, small changes in pressure at these levels can have measureable effects on material distributions. Figure 1. The fraction of uncollided atoms after traveling a distance x measured in MFP. Typical magnetron sputtering occurs at tens of millitorr. The MFP for 1.0 10 2 torr would be 0.5 cm. Therefore, at one inch or 2.54 cm distance from the sputtering cathode, only 0.6% of the sputtered molecules/atoms would not have collided. Clearly these sputtering processes are not "line of sight" depositions; an atom would typically suffer many collisions before reaching a substrate more than a centimeter away. It can also be shown that, at 1 10 6 torr, it only takes about a second for a monolayer of whatever gas is in the chamber (usually water vapor) to deposit on a surface. If the material depositing were contamination and it stayed on the surface, we might have some serious adhesion problems. HOW CAN WE CREATE THE VACUUM NEEDED? Figure 2 shows a typical box coater vacuum pumping systems. The chamber has two pumping port valves: the roughing valve and the high-vacuum valve. When the chamber has been loaded and the door closed, it starts at atmospheric pressure. The types of high vacuum pumps which can achieve the low pressures cannot typically operate at atmospheric pressure. It is therefore necessary to lower the chamber pressure sufficiently before the DP is brought into action. A mechanical rotary vane or piston displacement pump is commonly used to transfer

most of the gas from the chamber out to the atmosphere. With the chamber isolated by the roughing and HiVac valves closed, the mechanical pump (MP) is turned on and the foreline valve is opened to the HiVac pump. The HiVac pump might be a fractionating DP as illustrated. Figure 2. A typical box coater vacuum pumping system. The DP needs to be warmed up to operating temperature while being mechanically pumped. When it is warm enough that you cannot hold your hand on the foreline elbow, and it is pumped down to about 10 2 torr (10 microns), it is ready to use. To shut down a DP, the heaters are turned off while the cooling water is still running. When you can again hold your hand on the elbow of the foreline, the water cooling and MP can then be shut down also. When the DP is ready to use, the foreline valve can be closed (for a short while) and the roughing valve opened to the chamber. The MP then reduces the chamber pressure to about 10 1 torr (100 microns) where the DP can be used. Then, the rough valve must be closed and the foreline valve opened to clear the DP of any gas which has built up while the chamber was being rough pumped. When the foreline pressure is back to 10 2 torr or less, the HiVac valve is slowly opened to the chamber. If the valve is opened too quickly, the gas load on the DP might be too great and undesirable effects such as oil backstreaming would happen from the DP. To avoid this breakdown of the top jet, the indicator is that the foreline pressure should stay below 100 microns at all times. This 100 micron for the MP to DP crossover, and no more than 100 micron foreline pressure is a good rule of thumb ( 100 and 100 microns ) to avoid oil backstreaming from the MP and the DP. Note that it undesirable to rough the typical chamber much below 10 1 torr (100 microns ). This is because at lower pressures the gases flowing down the roughing line to the pump cease to be in viscous flow where they drive the pump oil molecules back into the pump. At the lower pressures, the molecules of MP oil may be in molecular flow, and some can travel back the roughing line into the chamber. Viscous flow would be like a stream which flowed so rapidly that no fish were able to swim upstream against it. However, when the stream was flowing slowly, the fish could swim anywhere. Similarly, in molecular flow, MP oil molecules can flow anywhere, including back into the coating chamber. It will be seen in Fig. 3 that the cross-over pressure to change from MP to DP is a compromise between upper and lower pressure limits,

and the pressure at which the transition from viscous to molecular flow regime occurs depends on the mean dimension of the openings involved. The 10 1 torr (100 microns) value mentioned above would be an appropriate lower limit for a 2-inch (50 mm) diameter roughing line and perhaps a 4-inch (100 mm) line. The background for this is discussed in more detail by Willey[1]. Figure 3. A chart of the pressures and mean gas conduit dimensions where flow changes from viscous to molecular. Dashed lines are typical, see text for other lines. In past times when pumping cycles were not automated, and even now on some manually operated pumping systems, backstreaming was more of a problem than it typically is now. It was not uncommon for an operator to start a system roughing and to go for a coffee break or lunch. When they returned, the chamber might be pumped down to a few microns, which is big NO-NO as indicated in Fig. 3. Many operators were unaware of this problem, and many coating runs had adhesion failures due to backstreaming. Figure 4 shows a recommendation from Langley and Hincle[2] which was preceded by Hoffman[3]. A valve, which could be as simple as a needle valve, is installed upstream of the MP by about 30 cm. (just to the right of the Roots blower), and it is set to give a pressure in that line of about 100 microns. This insures that this pumping line is always in viscous flow; and this avoids any backstreaming from the MP either to the chamber or to the DP. This is particularly good if there is a Roots blower in the system, because the purge gas will not influence the roughing time of the chamber, as it might in the absence of a Roots blower. Figure 4. To avoid backstreaming, a gas purge valve (upper right in figure) is set to give a pressure of about 100 microns in the vacuum line leading to the mechanical pump.

It is well to check the health of a coating chamber frequently by watching the pump-down versus time and also doing a leak-up test after the chamber is down to a given pressure. In a modern system with a computer display screen, it is easy to have a pressure versus time template of what the chamber does when it is clean-dry-and-empty and also an upper bound curve for when it is dirty and needs to be cleaned. Figure 5 illustrates such a screen. The operator can quickly see if the pumping is within bounds or if there is a problem like a leak, etc. Figure 5. Simulated pump-down screen to observe that conditions are normal during pumping. Similarly, when the chamber has pumped down normally within bounds, the HiVac valve can be shut and the pressure versus time rise observed. References of the rise with time when the seals were tight and the system was clean-dry-and-empty can be used, and also an upper bound of a dirty chamber, but no leaks rise time. If the latter is violated, it may be time to find and fix a leak. CONCLUSION When the various practices mentioned above are used, the run-to-run variations in the coating results due to vacuum and pumping should be minimal and the processes have a better chance of being the same tomorrow as they were today and yesterday. REFERENCES 1. R. R. Willey, Practical Production of Optical Thin Films, Willey Optical (2012). 2. R. Langley and L. Hinkle, How to Handle Backstreaming in a Diffusion Pumped Process System, Vacuum Technology and Coating, 9, 22-26(2008). 3. D. M. Hoffman, Operation and maintenance of a diffusion-pumped vacuum system, J. Vac. Sci. Technol. 16, 71-74(1978-9).