Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

Similar documents
Angstrom Dielectric Sputterer Operation Manual

Standard Operating Procedure. For. PVD E-Beam

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

University of MN, Minnesota Nano Center Standard Operating Procedure

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

Oerlikon Sputtering Evaporator SOP

Cryo-Evaporator Operation

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Operating Procedures for Metal Evaporator I

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

Standard Operating Manual

Standard Operating Manual

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Approved by Principal Investigator Date: Approved by Super User: Date:

Standard Operating Manual

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

SPUTTER STATION STANDARD OPERATING PROCEDURE

Scanning Electron Microscope JEOL JSM F

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

Nordiko Metal Sputtering System Standard Operating Procedure

Using the Smart IRT diamond ATR.

Karl Suss MJB4 Mask Aligner

THERMAL EVAPORATION UNIT (for Al evaporation)

INSTALLATION INSTRUCTIONS

Revised: June 7, 2017

1)! DO NOT PROCEED BEYOND THIS MARK

STS ICP-RIE. Scott Munro (2-4826,

INSTALLATION INSTRUCTIONS. Parts List. Tools Required. Before You Begin. Installation. Customer Information BICYCLE ATTACHMENT JUL.

INSTALLATION INSTRUCTIONS

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

OPERATION OF THE DIMPLER

Smooth Vent Valve. Series XVD XL XL Q D- XVD XGT CYV

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Read Before Operating!

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

MMRC PVD Evaporator Instructions

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

Cover Page for Lab Report Group Portion. Boundary Layer Measurements

Linkam Scientific Instruments USER GUIDE DSC600. Temperature Controlled Stage

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer

TYPICAL OPERATION PROCEDURES

Smooth Vent Valve XVD. Series

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

SEM LEO 1550 MANUAL. I. Introduction. Generality

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Thank you for purchasing your new Empire Reloader B Sound-Activated 3-Speed Paintball Hopper!

Smooth Vent Valve. XVD Series

Operator s Manual. All-Terrain Wheelchair

INSTALLATION INSTRUCTIONS

Approved by BNC Managing Director Paul Lum Date: Approved by PI (s) /Super User (s): _Peter Hosemann/Andy Minor_Date:

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

UNITY 2. Uni-DAAMS Manual. Version 1.3 (Changes to Section 1.1) March 2010

Maintenance Manual. Note: This document is broken up into the following 6 sections of prescribed maintenance.

IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES. Tank valves must be installed or removed by qualified personnel.

Temescal BJD-1800 E-Beam Evaporator 1 System Overview - Cryo-pumped for typical base pressures in the low 10e-7 Torr range. - Four pockets in the

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

ATOC Meteorological Tower (6 meter-4 level) Guide

IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES. Tank valves must be installed or removed by qualified personnel.

Standard Operating Manual

OXFORD PLASMALAB 80PLUS (CLOEY)

9A5N Solid State CW Paddle

BAUSCH & LOMB ENGINEERING ROCHESTER. Title: Membrane Press Procedure Page 1 of 14

WARNING OLED & BASIC MENU QUICK START

MAINTENANCE PROCEDURE FOR X 650

STS PECVD Instructions

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

Linkam Scientific Instruments USER GUIDE TST350. Temperature Controlled Tensile Stress Testing Stage

[needs to be discussed with faculty if any new material to be done]

Standard Operating Manual

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement

Basic ICP Operating Procedures

SCUBAPRO Repair Guide. S600-S550 Second Stages

Read Before Operating!

QUORUM TECH 150RES THE FIRST AND THIRD WEEK WILL BE SET UP FOR CARBON COATING THE SECOND AND LAST WEEK WILL BE SET UP FOR GOLD COATING

Savannah S100 ALD at SCIF, UC Merced Standard operating Procedure

Experiment 11: The Ideal Gas Law

Warnings: Notes: Revised: January 8,

OLED vacuum deposition cluster system Sunicel Plus 400 / Plus 400L (Sunic Systems)

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

MJB-3 Mask Aligner Property of TAU MNCF August 2012

Purpose and Scope: Responsibilities: Warnings and precautions: Background Beers law

Unaxis PECVD. SiH4 (5% in He)

Konza District Pinewood Derby Track Parts And Assembly Instructions

3. Align ratchet drive with female insert on assembly, and raise ratchet handle until the ratchet drive is firmly seated into the assembly.

[USING THE VITROBOT April 9, Using the Vitrobot

High Altitude Deployment Article By Jim Jarvis Rockets Magazine, June 2011 Edited May 2013

Pressure Dump Valve Service Kit for Series 2300 Units

Mate ULTRA IMT Multi Tool Assembly Tooling Installation Instructions

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Parts: Included in the parts box: Inner Rear Tire Tray. Inner Front Tire Tray. Trail Doc Clamp. Pivot Assembly. Trail Doc Post.

Arizona State University Center for Solid State Electronic Research. Table of Contents. Issue: C Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8

MAGNETIC CYCLING TRAINER SF-B0419 USER MANUAL

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Transcription:

Angstrom E-Beam Instructions Tool Manager: Joe Palmer (jpalmer@princeton.edu) Office: 8-4706; Cell:609-751-1353 Backup: David Barth (dbarth@princeton.edu) Office: 8-4626; Cell: 610-405-8227 PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test. The Angstrom E-beam is a dual-chamber system for thin-film evaporation. The upper chamber is a load lock for increased pumping speed. A maximum of six different materials can be evaporated in one pump-down cycle, from the six pocket electron gun. The thickness of the metal is monitored in-situ by a thickness monitor. Substrates up to 6 inch diameter can be attached to the substrate holder. The sample temperature usually does not exceed 50 C. Only the following materials are available: Au, Pd, Ge, Ni, Pt, and Ti. 1) Loading your sample a. Record your name, date, and process to be executed in the log book. b. Log on to this computer and click on the Control Software Icon. When this program opens, its tab will be labeled as Princeton (01100). SR in the tab bar. Step 1) b.

c. Click Auto Vent under Upper Chamber in Mode Selection which is located on the right side of the screen. A window will pop up. Click Start.

d. Once the Upper Chamber has vented to atmospheric pressure, open the Upper Chamber door. Step 1) d.

e. Remove Sample Holder (i.e. the round metal disk hanging upside down in the Upper Chamber) i. Place your right hand (these directions assume that you are right-handed; please reverse if you are left-handed) under the sample holder and gently push it up while rotating the sample holder release knob (located on the top of the machine) with your left hand. This should be done so that the black dots on the sample holder release knob move to the left. Take care not to drop the sample holder! Step 1) e.

ii. When you feel the sample holder unlock itself, slowly lower the sample holder with your right hand and remove it from the machine. f. Once removed, securely fasten your sample to the sample holder using the Angstrom E-Beam clips, screws, and washers. Use only the clips, screws and washers that are labeled for Angstrom E-Beam use. Do not borrow from other machines. If you require your own set or do not wish to use clips, please see the staff. g. Put screws in any open holes on the sample holder so that metal isn t deposited on the heater. h. Put the sample holder back in the machine. i. Use your right hand to put the piston of the sample holder back in the hole. ii. Slowly rotate the sample holder with your right hand until its notches are aligned (once aligned, the sample holder will go further in). iii. Rotate the black dots on the sample holder release knob back to the right this will lock the sample holder in place. Step 1) h.

iv. Slowly remove your right hand from the sample holder and ensure that it is indeed locked in place. 2) Pumping Down a. Close the chamber door and finger tighten the door clamp. Step 2) a.

b. Click Full Pump Down under Upper Chamber in Mode Selection. Then click Start Selected Cycle. Step 2) b. Step 2) b.

c. Un-tighten the door clamp once the pressure becomes less than atmospheric pressure. Step 2) c.

d. Wait until the upper chamber pressure reaches 8.0e-7 Torr. To check this, click Upper Chamber Interlocks under Req D Interlocks on the left side of the screen. The Chamber Base Pressure and Chamber Safety Pressure Buttons will turn green when pressure becomes low enough. Step 2) d.

3) Depositing a. Go to File, Process, and then select the desired deposition process. Step 3) a.

b. Ensure that you have selected the correct process by checking the name of the process at the top of the screen. If correct, click Start Process. Name of Process Selected to be Executed Step 3) b.

c. Allow the process to run to completion under your supervision. You are required to watch the machine any time the E-Gun is active. 4) Removing Your Sample a. Once your process is fully completed, click Auto Vent under Upper Chamber in Mode Selection which is located on the right side of the screen. A window will pop up. Click Start. Step 4) a.

b. Once the Upper Chamber has vented to atmospheric pressure, open the Upper Chamber door. Step 4) b.

c. Remove the sample holder i. Place your right hand under the sample holder and gently push it up while rotating the sample holder release knob (located on the top of the machine) with your left hand. This should be done so that the black dots on the sample holder release knob moves to the left. Take care not to drop the sample holder! Step 4) c.

ii. When you feel the sample holder unlock itself, slowly lower the sample holder with your right hand and remove it from the machine. d. Remove your sample from the sample holder. e. Put the sample holder back in the upper chamber. i. Use your right hand to put the piston of the sample holder back in the hole. ii. Slowly rotate the sample holder with your right hand until its notches are aligned (once aligned, the sample holder will go further in). iii. Rotate the black dots on the sample holder release knob back to the right this will lock the sample holder in place. Step 4) e. iv. Slowly remove your right hand from the sample holder and ensure that it is indeed locked in place. f. Record the time you finished using the Angstrom E-Beam in the log book.