Procedures for operation of the TA Instruments DSC

Similar documents
FOURIER TRANSFORM INFRARED SPECTROSCOPY

Purpose and Scope: Responsibilities: Warnings and precautions: Background Beers law

Operation of the Perkin Elmer TGA-GC/MS

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

The Gas Laws: Boyle's Law and Charles Law

Vapor Sorption Analyzer

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

Xactix XeF2 OPERATION MANUAL

RM-80 respiration monitor

Technology. In the My Files [My Files] submenu you can store all the programs that you have made on the NXT or downloaded from your computer.

FOOD INTAKE MONITOR. Hardware and Software User s Manual DOC-047 Rev 3.0. Copyright 2009 All Rights Reserved

[MYLAPS INTEGRATION]

Ideal gas law. Introduction

Boyle s Law: Pressure-Volume Relationship in Gases

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

SOP for use of the SX19 Stopped-Flow Instrument

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

P3000 Deadweight Tester Setup, Part 1: Limited Partial Correction Method

Physics Experiment 17 Ideal Gas Law Qualitative Study

5890II GC Standard Operating Procedure 9/2/2005

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

OPERATION MANUAL FOR VP 710E-3HM SERIES ALLIGATOR TUMBLE STIRRERS *US Patent # 6,176,609

Boyle s Law VC 09. Experiment 9: Gas Laws. Abstract

Standard Operating Manual

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

DSC 204 HP Phoenix. High-Pressure Differential Scanning Calorimetry up to 15 MPa Method, Technique and Applications

Usage Policies Notebook for Trion RIE / ICP Dry Etch

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Liquid Nitrogen Cooling Accessory CE

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

GCMSD-Headspace Analysis SOP

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Technology. Using Bluetooth

University of Minnesota, MN Nano Center Standard Operating Procedure

JABATAN KIMIA Kategori Arahan Kerja

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

1001ICT Introduction To Programming Lecture Notes

Armfield Distillation Column Operation Guidelines

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

LOW PRESSURE EFFUSION OF GASES revised by Igor Bolotin 03/05/12

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Discovery TGA 55, TGA 550, TGA Site Preparation Guide

LABORATORY INVESTIGATION

Instruction Manual. BZ7002 Calibration Software BE

Cover Page for Lab Report Group Portion. Head Losses in Pipes

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Basic Autoclave #1 Ideal Gas Law, Inert Gas

The MRL Furnaces USED FOR THIS MANUAL COVERS

LOW PRESSURE EFFUSION OF GASES adapted by Luke Hanley and Mike Trenary

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Exp. 5 Ideal gas law. Introduction

DDR Dive Data Recorder Manual - Rel /12

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

STS PECVD Instructions

Application Note Temperature dependence of viscosity of liquid crystal

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

Figure 1: Plug in ion inlet of mass spec

Fencing Time Version 4.3

Pneumatic high-pressure controller Model CPC7000

Ranger Walking Initiation Stephanie Schneider 5/15/2012 Final Report for Cornell Ranger Research

NEULOG OXYGEN LOGGER SENSOR GUIDE

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Standard Operating Manual

BASIC LABORATORY TECHNIQUES (Revised )

Manual for continuous distillation

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

CHM Introductory Laboratory Experiment (r17sd) 1/13

PGA3510 Portable 3-Gas IR Analyzer. Operations Manual

Heating of Land and Water

Basic ICP Operating Procedures

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Discovery HP-TGA 75/750. Site Preparation Guide

Chamber Test Testing Theory

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

RAM 4021-DPX Operation Manual

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

CHEMICAL ENGINEERING SENIOR LABORATORY CHEG Initiated Chemical Vapor Deposition

RAM 4021 Operation Manual

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark

AP Lab 11.3 Archimedes Principle

RAM Operation Manual

FLUORESCENCE DETERMINATION OF OXYGEN

NOVALYNX CORPORATION MODEL RAIN GAUGE CALIBRATOR INSTRUCTION MANUAL

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

University of MN, Minnesota Nano Center Standard Operating Procedure

RH800 & 2000 BASIC OPERATIONS

THE PMI AUTOMATED GAS/HELIUM PYCNOMETER PYC-G100A

Standard Operating Manual

The Race Director. IPICO Integration Direct Connect [IPICO INTEGRATION]

PitchTracker User Guide. Model: DKPT01 User Guide

6200 Isoperibol Calorimeter Model Number: 6200

Application Note AN-107

Transcription:

Procedures for operation of the TA Instruments DSC Purpose and Scope: This document describes the procedures and policies for using the MSE TA Instruments DSC. The scope of this document is to establish user procedures. Instrument maintenance and repair are outside the scope of this document. Responsibilities: This document is maintained by the department Lab manager. The Lab Manager is responsible for general maintenance and for arranging repair when necessary. If you feel that the instrument is in need of repair or is not operating correctly please notify the Lab Manager immediately. The Lab Manger will operate the instruments according to the procedures set down in this document and will provide instruction and training to users within the department. Users are responsible for using the instrument described according to these procedures. These procedures assume that the user has had at least one training session. Warnings and precautions: DO NOT HEAT YOUR SAMPLE TO DECOMPOSITION TEMPERATURES. TGA MUST ALWAYS BE PERFORMED FIRST! Please do not eat or drink while operating this instrument. NEVER touch the robotic arm. Never attempt to operate the instrument without nitrogen flow. Do not place a usb drive in any of the usb ports of the computer. Do not change any of the user preferences or instrument preferences. Notify the Lab Manger of any issues, do not attempt to fix a problem. The analysis software for the DSC and TGA are available for free on the TA Instruments web page at: Installation of Universal Analysis from the Advantage Software Package You will only need to download the Universal Analysis portion. There are also some training power-point videos available to further instruction on analysis. You are responsible for providing and caring for your own consumables for this instrument. T-Zero aluminum pans and lids are available from TA Instruments. Reference pans are provided. Before starting please verify that there is plenty of N2 in the cylinder and that the flow rate is at 50 ml/min. Background Differential Scanning Calorimetry is used to measure heat flow in or out of a material. From this information a number of thermal properties can be determined; Tg, melting temperature, crystallization temperature, oxidative induction time, % cure, etc. This method uses an empty sample pan as a reference that cancels the heat flow to and from the pan itself. Fig. 1 Typical transitions: TA Instruments training 02/2014

Fig. 2 and 3: Schematics of sample pan and chamber configuration: TA instruments Training: 02/2014 DSC operates under the following assumptions: (modified from TA Instruments Theory and Applications Training 02/2014) Fig. 4 Schematic and equation for basic DSC measurement: TA Instruments training 02/2014 The heat flow rate of an empty sample pan is zero Thermal resistance between the sample sensor and the furnace is the same as the resistance between the reference sensor and the furnace. The pan and sensor heat capacities are ignored The measured temperature equals the sample temperature No heat is exchanged with the surroundings. The sample material is making good contact with the bottom of the sample pan. Many parameters must be considered when preparing your sample and when interpreting your data. Sample prep is very important. The sample must make good contact with the surface of the sample pan. The bottom of the pans are very precisely machined to be flat (manufacturing tolerance of 0.00127mm). So sample prep should be done in as clean an environment as possible. There is already a prepared empty reference pan in the auto changer. Please do not remove this pan! Contact the Lab Manger if you think there is a problem.

The following figures are presented as examples taken from the TA instruments training manual. Please keep in mind that these are idealized representations and may not be typical results. Fig. 5 Endothermic vs Exothermic heat flow: TA Instruments training 02/2014 Fig. 6 Epoxy cure studies: TA Instruments training 02/2014 Fig. 7 Epoxy cure studies: TA Instruments training 02/2014

The following parameters should be considered to optimize your measurement: (modified from TA Instruments Theory and Applications Training 02/2014) Determine which pan type is best. The standard pan for our instrument is the T-Zero pan. If your sample loses about 0.5% or more due to volatile materials you should consider the hermitically sealed pans. What is the maximum temperature. Run TGA on your sample AT THE SAME RATE you plan to run your DSC measurements. Decomposition temperature is lower at lower rates. o o Decomposition will contaminate (and ultimately ruin) the sample chamber. DSC results are of little value once the sample has lost 5% of its weight due to decomposition (not desolvation) To determine volatile content you can heat about 5-10 mg of simple at 10 C/min. Again, only AFTER TGA. This is a kinetic measurement so consider rate. If you intend to do comparative studies, all conditions must remain the same. A sample measured at a rate of 20 C/min will not return the same results as a sample run at 10 C/min. Try to keep the sample thin. The sample should cover as much of the bottom of the pan as possible. If you are comparing a number of samples keep the sample masses as much the same as possible. Using your own tools and materials will reduce contamination. If you are measuring a liquid, you will only need a very small amount. Liquids can spill out of the sample pan either during pressing or during the scan. If this happens while in the chamber the chamber can become contaminated and create odd behavior both in the robotic operation and in the results. Please take care not to let this happen. Robotic Arm Sample Chamber Auto Changer Fig. 8 MSE DSC Instrument Procedures Place an empty T-Zero pan on the scale and tare Place your sample in the pan. It is very important that the sample make good contact with the surface of the pan. Good thermal contact is crucial to a good measurement. Note the weight of the sample (1 10 mg) Put the top die in place and put the pan in the bottom die. Place the lid on the pan and put the assembly in the press. Bring the press lever down carefully. No pressure is necessary. Please be gentle.

Move the lever up and back to its original position. Remove the pan and place in a position in the auto chamber. Take great care not to touch the robotic arm. It is very easy to bump it out of alignment. Repeat for each sample. (This will be demonstrated during training). If the software isn t already open, open instrument explorer and select the Q20 DSC. The last user s information will be displayed. In the summary tab select <Standard Mode> <Custom Test> and if you already have a completed procedures file you can open it here are well by clicking on the file icon. This is usually done during your training. Type in the sample name Select the correct sample pan position and reference pan position Select the icon to the right of Data file name and name the unique file. o Note: If you are running more than one sample you can name the first one and TA will add.00x to each subsequent run. Or you can provide a unique file name for each one. If necessary, select <Apply> If you have not created a procedures file, select the second tab, <Procedure> Under method, select <Editor> To the right of the screen will be a list of steps you can select. The most common is simply the rate and temperature. Standard would be 10 C/min to a temperature just past melting temperature. If you do not know what that temperature is you MUST RUN TGA to determine decomposition temperature. Do not run a sample to or past a materials decomposition temperature as this can damage the instrument. Select <Advanced> This is where you can define what temperature to load the next sample at. Enter a 10 degree range with the center range being the temperature you want to start your run at. Select OK then open the <End> conditions button. Enter a 10 degree range of temperatures to unload the completed sample. You can also choose to have the sample pan discarded. Check the <Notes> tab and enter your name in the appropriate field. Make sure that the flow rate is 50 ml / min. DO NOT CHANGE THIS FLOW RATE. To run more than one sample: After you have completed the first run parameters right click on the <Run> (left side of screen) and select <Copy and Append> Do this for every run. Go back and in each summary page type in the sample name, the sample weight, the pan number and if desired, provide a unique file name. Select <Apply> Double click the top run. Select <Run> at the top of the page.

Fig. 9 Programming and operations software When all of your measurements have been completed you can collect your data from the WORKGROUP computer. No other action is necessary. Please leave the computer on.