Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Similar documents
13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Commissioning and periodic tests of the Esteya electronic brachytherapy system

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

Relative Dosimetry. Photons

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

Constancy checks of well-type ionization chambers with external-beam radiation units

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Zoubir Ouhib Lynn Cancer Institute

TG-51: Experience from 150 institutions, common errors, and helpful hints

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Victoreen B. Operators Manual. Image Intensifier Ion Chamber

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy

Monitor Unit Verification for Small Fields

Abstract: Introduction:

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC #

Proton Therapy QA Tools

Review of fundamental photon dosimetry quantities

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

Maintenance and general care of secondary standard and field instruments

THE development of more advanced techniques in radiotherapy,

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

ANSI N323A and ANSI N323B

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

IMRT in clinical practice at the UMC-Utrecht. Utrecht

SnapShot IMRT with compensators and FFF beams

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

General Accreditation Guidance. User checks and maintenance of laboratory balances

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Dosimetric Calculations. Lonny Trestrail

Exposure System Selection

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

A beam-matching concept for medical linear accelerators

Operational experiences with the EuroLoop Liquid Hydrocarbon Flow Facility

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Math Review. Overview

Investigating the effects of interchanging components used to perform ripple assessments on calibrated vector network analysers

A3 / A4 / A5 / A6 / A8 Spherical Ionization Chambers REF / / / / User Manual

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Victoreen and

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

BLOCKAGE LOCATION THE PULSE METHOD

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

How to Design Medical Accelerator Systems for Reliability: IBA PT System

ON THE RELATIVE ADEQUACY OF SOME AMERICAN CHAMBERS FOR MEASUREMENTS IN THE GRENZ RAY RANGE*

Improve Process Reliability

Thimble Ionization Chambers

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material

Increased Project Bankability : Thailand's First Ground-Based LiDAR Wind Measurement Campaign

MET-101 Introduction to Metrology

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber

Technical Procedure for General Laboratory Equipment

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP

ATION TITLE. Survey QC, Decision Making, and a Modest Proposal for Error Models. Marc Willerth, MagVAR

MIT Analytical X-Ray Safety Policies and Procedures

Cover Page for Lab Report Group Portion. Pump Performance

Temperature, salinity, density, and the oceanic pressure field

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

Impact of imperfect sealing on the flow measurement of natural gas by orifice plates

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. PRECISION MEASURING CORP Groesbeck Hwy. Fraser, MI Robert Meyer Phone:

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

Testing of Active Flow Control actuators at harsh environment

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY

LAKE WASHINGTON SOCKEYE SALMON STUDIES. Richard E. Thorne and James J. Dawson

Application Note AN-107

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Laboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.

Technical Report. 5th Round Robin Test for Multi-Capillary Ventilation Calibration Standards (2016/2017)

SUPPLEMENTARY INFORMATION

Gun Vacuum Performance Measurements February 2003

Transcription:

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin NCCAAPM Meeting Oct 11,2013

Larry DeWerd has a partial interest in Standard Imaging

Talks from people who have been in the field for awhile tend to be historical or philosophical My attempt at philosophical and hopefully informational material.

I don t have enough time Why worry it will all work out. I don t have to think about it it is late. Didn t the manufacturer tell me it was okay.

Some Manufacturers harp on the time saved, etc but their instrument is not as precise and can result in errors. For example, some digital sampling electrometers (depending on their sampling algorithm) may miss some signal from pulses of the accelerator. The physicist should ask.

Technicians use numbers, Medical physicists use understanding. Do it right good enough is not good enough

Transfer of Knowledge is Important. ADCLs transfer latest knowledge Â\y \ {täx áxxç yâüà{xü à{tç Éà{xÜá? à á uç áàtçw Çz âñéç à{x á{éâäwxüá Éy z tçàáê f Ü \áttv axãàéç? DIJI f Ü \áttv

Just doing a measurement and applying ADCL calibration does not mean that it is right You need to understand the equipment and see through the measurement. Numbers no matter how precise cannot by themselves imply anything. Precision and accuracy of dose measurements and reporting of the measurement details should be sufficient to allow the work to be interpreted and repeated and to allow valid comparisons to be made, both in the same laboratory and by other laboratories.

Ion chambers require calibration for the most precise measurements ADCLs willing to discuss measurements and methodology ADCL discuss the operation of instrumentation.

1 0.9 primary standard measurement frequency of result 0.8 0.7 0.6 0.5 0.4 0.3 clinical measurement conventional true value 0.2 0.1 0-6 -4-2 0 2 4 6 a.u.

NBS petitioned AAPM to create Regional Calibration Laboratories in 1975- In 1983 called ADCLs. NBS acknowledges traceability to primary standards (Proficiency tests) Agreement for Proficiency tests and round robins < 0.5% UWADCL founded 1981 by LAD

International & National Standards Primary Calibration Secondary Standards ADCL Hospital or Clinical Standards Secondary Calibration Tertiary Calibration Operational Standards

AAPM in conjunction with NIST set up secondary laboratories for medical ionizing radiation calibrations - For therapy applications the precision is within 0.5%. Traceability of radiation dose is dependent on standards that are the same throughout the world at primary labs.

All calibrations (especially for medical applications) should have a traceability train to a standard This provides confidence in the dose given because it was measured and can be traced back to a national standard at NIST However, the SSDL network is not the equivalent to the ADCL network

The ADCLs have proven track records of providing precise calibrations of equipment for Therapy and Diagnostic applications Safety and treatment delivery or imaging is improved because of ADCL calibration and research.

0.994 0.995 0.996 0.997 0.998 0.999 1.001 1.002 1 1.003 1.004 1.005 N u m b e r 16 14 12 10 8 6 4 2 0 Mean of ADCLs/ADCL

These tests have been in place over 20 years NIST and ADCLs agree within 0.5% for Cobalt-60 beams NIST and ADCLs agree less than 2.0 % for x-ray beams between 20 kvp and 250 kvp

Maintenance of accuracy and precision is very important for Medical Applications Knowledge of characteristics of chambers is very important Medical Applications rely heavily upon precise traceable calibrations.

Calibration of chambers should be done across the energy range of use. Bracket your point of interest. Generally there is a reference energy in use that is the minimum energy point used.

Because of variation in response with energy for chambers a range of calibration is necessary. The calibration should be done under correct criteria e.g. TG 61 There is no protocol using an absorbed dose to water calibration for x-ray beams. An SSDL offers absorbed dose in water but no protocol.

Within + 1% for Mammography range: 24 kvp to 35 kvp, 0.15 to 1.0 mm Al Within + 2 % or 2.5% for entire diagnostic range, up to 150 kvp For superficial or orthovoltage (radiobiology) 1 % to 2 % is desirable

There can be electron contamination and full buildup is needed for calibration of low energy (< 100kV) clinical beams This requires an addition to the window to be put over a plane-parallel chamber The window of the chamber (for example 2.5 mg / cm 2) should be subtracted from the values in the following table for the thickness of the foil for full buildup.

Tube potential (kv) Total wall thickness (mg cm -2 ) 40 3.0 50 4.0 60 5.5 70 7.3 80 9.1 90 11.2 100 13.4

Physicist should understand uncertainties as delivered by the ADCL For example, for cobalt calibrations

The expanded uncertainty (at k=2) is expected to include the conventional true value. The chance of the population mean (the true value) being within the measured value has a 95% certainty. Or we can be sure that our measurement indicates the correct value within 95%

ADCL Standard Chamber Parameter Type A Type B Charge 0.02 0.0 Timing 0.04 0.05 Air Density 0 0.1 Ionic Recombination 0 0 Distance from the source 0.1 0.07 Beam Uniformity 0 0.1 Air attenuation 0.1 0 Quadratic Sum 0.110 0.166 Type A and B Quadratic Sum 0.198 NIST Source Calibration 0.600 u NIST-ADCL Chamber Calibration 0.632 or at k=2 level, 1.26 %

Clinic Chamber Parameter Type A Type B Charge 0.02 0.1 Timing 0.04 0.05 Air Density 0 0.1 Ionic Recombination 0 0.1 Distance from the source 0.1 0.07 Beam Uniformity 0 0.1 Quadratic Sum 0.110 0.218 Type A and B Quadratic Sum 0.244 ADCL Electrometer Readout 0.057 ADCL Transfer Calibration 0.632 u ADCL-Clinic Chamber Calibration 0.680 or at k=2 level, 1.36 %

Standards changing Absorbed dose to water done with precision and uniformly. Calibration of chambers and sources essential for radiation therapy: Majority traceable through Cobalt and Cesium. Unfortunately, some manufacturers improvise dosimetry to market their product

Manufacturers need to pass some hoops to market product However, 510k from FDA does not mean that they have a calibration regimen for their machine. Also some manufacturers change without letting physicists know and assume everything still applies.

What happens when manufacturers improvise for Standards? Errors! NIST, ADCLs, AAPM should insist on a standard There is a need to have traceability delivered by the Accredited Dosimetry Calibration Laboratories Hospital Physicists generally have requests Users must insist on traceable standards from NIST through the ADCLs

Sr-90 applicators introduced in 1950. Plane or concave. 28 year half life Typical 8mm diameter with a beta shield Extremely high dose rate Calibrated by each mfr wrt Bragg-Gray theory Bizarre units like reps, beta-roentgens Hospitals asked UW ADCL accuracy of dose.

Wing-like growth in conjunctiva Requires surgery to remove Surgery fails 90% of time One dose with Sr-90 prevents recurrence in roughly 90% of patients treated

One Chicago hospital had an Amersham Model SIA-20 applicator calibrated at NBS: calibrations differed by 38%. Each lab stood by their calibration! NIST had to review their calibrations

Sr-90 sources had only been intercompared IN AIR at distances of 20 to 30cm Desired dose rate is IN CONTACT with Sr-90 on surface of silver matrix Chris Soares developed an extrapolation chamber for calibration.

NIST traceable determination of the absorbed dose to water rate in the central 4 mm of the applicator including color enhanced contour plots and two dimensional dose profiles. Routine ADCL calibrations from UW ADCL using radiochromic film began in 1996. NIST quotes uncertainty of 7 % Calibration disagreement (old versus new) average -27 %.

Example of a color enhanced contour plot of a uniform Sr-90 ophthalmic applicator. The dose weighted isocenter is equidistant from the hash marks. The outer circle represents the source physical diameter. 70 60 50 40 30 20 10 80 90

2-Dimensional dose profiles for a uniform source. Horizontal Profile Vertical Profile 0.50 0.50 Dose Rate (Gy/sec) 0.40 0.30 0.20 0.10 Dose Rate (Gy/sec) 0.40 0.30 0.20 0.10 0.00 0.00-6.0-4.0-2.0 0.0 2.0 4.0 6.0 Distance (mm) -6.0-4.0-2.0 0.0 2.0 4.0 6.0 Distance (mm)

Example of a color enhanced contour plot of a skewed-sr-90 ophthalmic applicator. Notice the offset or shift of the dose weighted isocenter from the physical source center, and the nonuniform dose distribution. 40 30 20 10 50 60 70 90 80 90

Horizontal Profile Vertical Profile 0.40 0.40 0.30 0.30 0.20 0.20 0.10 0.10 0.00-6.0-4.0-2.0 0.0 2.0 4.0 6.0 Distance (mm) 0.00-6.0-4.0-2.0 0.0 2.0 4.0 6.0 Distance (mm)

The average difference between the old calibration and the present calibration is -27%. The distribution of source activity can be very unique for the individual source.

NIST standard Pd-103 has very short half-life (17.0 d) so NIST traceable 109 Cd source (half-life of 463.2d) was used as a reference source for 12 years, then replaced in 1997 However, self-shielding of the source encapsulation was different between these two isotopes This resulted in a sudden 9% shift in calibration by letter to users in 1997

There can be significant errors If traceable to NIST, even if wrong, everyone is consistent if traceable standard- at least traceable through ADCL NIST, ADCLs and CIRMS need to insist upon standards. Manufacturers try hard but they need to rely on NIST and ADCLs for standards

Âg{Éáx ã{é wé ÇÉà ÜxÅxÅuxÜ à{x Ñtáà tüx véçwxåçxw àé ÜxÑxtà àaê George Santayana, Harvard Professor and poet

Thanks are due to Students and staff of the UW ADCL All those who send us calibration instruments that support the research program of the UW Medical Radiation Research Center.