Navigator 600 Silica analyzers

Similar documents
ABB MEASUREMENT & ANALYTICS DATA SHEET series Dissolved oxygen systems

ATS430 turbidity sensor Retractable insertion assembly

9408 Series Dissolved Oxygen Systems. Dissolved oxygen systems designed for reliable operation with minimal operating costs

9408 series Dissolved oxygen systems

Service Calibration are you doing it properly?

AP100 series ph/redox (ORP) cartridge sensors with water-wash option

ABB MEASUREMENT & ANALYTICS DATA SHEET. AP100 ph/redox (ORP) cartridge sensors with water-wash option

UKAS calibration laboratory 8809 Temperature, pressure and electrical signals calibration guide

Service ISO17025 explained

UsER manual for Watersens ph -REDOX

C1960. Multi-recipe profile recorder/controller. Measurement made easy

Endura AZ40 Transmitter Main board spares kit

The benefits of the extended diagnostics feature. Compact, well-proven, and flexible

Endura AZ40 Sensor assembly High temperature filter assembly kit

PositionMaster EDP300 Extended Diagnostics. Compact, well-proven, and flexible

Observations Possible Causes Recommended Actions. Turquoise color

Operator Quick Guide ORBISPHERE 3654

FPD175 Compact orifice carrier assembly

Dissolved Oxygen Cartridge Electrode Systems

Calcium Reactor. CalcFeeder PRO. User Manual 1.1a. Model. CalcFeeder AC1 PRO CalcFeeder AC2 PRO CalcFeeder AC3 PRO CalcFeeder AC4 PRO

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Endura AZ40 Sensor assembly Smart electronics PCB

2600T Series Pressure Transmitters Calibration Frequency Determination

Touch Screen Guide. OG-1500 and OG Part # T011

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

6900 Maintenance Instruction System Flush

FX134, FX1234, FX3030 Problem and Solutions

Calcimeter Instruction Manual

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Operating Instructions

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

Quick Installation Guide Doc nr.: C Date:

ACV-10 Automatic Control Valve

CoriolisMaster mass flowmeter Diagnostics, verification and proof test

Pressure measurement Applying remote seal, isolation diaphragms between process fluids and transmitters

Dissolved Oxygen Systems

2600T Series Pressure Transmitters Plugged Impulse Line Detection Diagnostic. Pressure Measurement Engineered solutions for all applications

SEQUENTIAL SHOWER VALVE INSTRUCTION MANUAL PLEASE LEAVE THIS MANUAL WITH THE END USER

Instructions for Use

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Latvin Luxury Shower Panel. Telephone Product Specification. ~ Minimum Working Pressure 1.0 bar ~ Maximum Working Pressure 3.

Section 3- Repair and Calibration Procedures

HS Bubbler Technology. HyQuest Solutions New Zealand User Group August 2017

FAULT CODE TROUBLESHOOTING INDEX

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem

T004.BioTector TOC Analyzer Pressure and Flow Test failure analysis

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

AUTOMATIC GAS MANIFOLDS

NIV EAU MATIC ST User Manual Made in Canada

PRODUCT CONFORMITY CERTIFICATE

MODEL 1329 Tank Gauge

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Agilent Auxiliary Gas Module AGM 2. User s Guide

6800 Maintenance Instruction System Flush Procedure

Working correctly with gas

MoveRoll Conveyor Operating and Maintenance Manual

ABB measurement & analytics data sheet. JAA, JAF and JAG Pneumatic indicating controllers

SEMEMI SQA Unit Code H2B7 04 Assisting in the installation of compressed air equipment

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Agilent 1290 Infinity Pump Head Maintenance

AREA VALVE SERVICE UNITS

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Self-help Temperature Adjustment

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

SECTION 2 SMART PAYOUT MANUAL SET FIELD SERVICE MANUAL

Overview. SEMEMI216 SQA Unit Code HF4G 04 Carrying out fault location on fluid power equipment and circuits

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

BD Cytopeia Fluidic Kit User s Guide

High Pressure Chem-SCAN Operating Manual

ZIRCONIA OXYGEN DETECTOR SERVICE MANUAL

Considerations for Sampling Wet, High Pressure, and Supercritical Natural Gas. Presented by: Shannon M. Bromley

PANACEA P200 STAINLESS STEEL USER MANUAL

641/01 AVK SWING CHECK VALVE OPERATION & MAINTENANCE MANUAL. AVK UK Limited OM270710

PURA. Pure Gas Dewpoint Transmitter. Users Guide

5 Channel Calibrator

RAM Operation Manual

SPECIFICATION // DATA SHEET WATER DRAIN VALVE AM SENSORS LTD AMS-S-084/DS/ (ISSUE 1)

OPERATING INSTRUCTIONS Sherlock Chamber Testing System

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

SEMPEO SQA Unit Code FP2P 04 Maintaining fluid power equipment

1. Study the performance of a binary distillation column operated in batch mode.

ABB MEASUREMENT & ANALYTICS DATA SHEET. AK100 Series ATEX compliant gas analyzer system for hydrogen-cooled alternators

Pressure Dump Valve Service Kit for Series 2300 Units

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

User Manual for the Mars Calibration Bench

Safety Manual. Searchpoint Optima Plus

1 Square = Foot/Feet Revision G 1/18/2013 Titan Controls

COUNTER CURRENT UNIT. Instruction Manual

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

Modular valve block. Operating manual Series MVB 100/200. Version BA EN Print-No TR MA DE Rev001

Sedan Range. Evaporative Loss System Check Clarified Procedure

Model NDA Differential pressure transmitter. Deltapi N Series A complete range of pneumatic pressure transmitters

INSTALLATION, OPERATING & MAINTENANCE MANUAL WATER DRAIN VALVE

RHPS Series RD(H)F40 User Manual. Read the complete manual before installing and using the regulator.

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

O2 Portable. User Manual

Pressure Dump Valve Service Kit for Series 3000 Units

Transcription:

ABB MEASUREMENT & ANALYTICS INFORMATION INF09/020 REV. B Navigator 600 Silica analyzers Troubleshooting Measurement made easy Navigator 600 silica analyzers 1 Introduction This publication details troubleshooting procedures for Navigator 600 series silica analyzers. 2 For more information Further information is available from: www.abb.com/analytical or by scanning these codes: Sales Service

Note. If the procedures in this section do not resolve the problem, gather the information in IM/NAV6S, Section 10 and contact ABB Service. 3.1 Analyzer Malfunction In the majority of cases any problems experienced are found generally to be associated with the chemistry (check reagents/solutions first) and the liquid handling section. The most common problems are associated with the reagent or standard solutions. Any unpredictable problems may be due to the standard or reagent solutions or their flow through the analyzer. If any doubts exist regarding the integrity of these solutions, replace with freshly prepared solutions in the early stages of the fault finding investigations. If the analyzer fails to produce expected results, the most likely cause is the standards; either contaminated when handled or made up with poor quality water, possibly containing high levels of silica. Incorrectly prepared reagents may give a poor calibration factor. If the solutions are from a proprietary chemical supplier, take care when storing the bottles; date stamp, use in strict rotation and do not use after their expiry date. When measuring trace levels, contamination can be avoided only by preparing and handling solutions with great care. Laboratory glassware is not to be used for solutions; use plastic instead. Check mechanical components that are involved with the liquid handling systematically. For example, pumps, valves, tubing and tubing connections should be checked for correct operation and for leaks or blockages. Check that there have been no unauthorized modifications, for example, incorrect tubing fitted. A calibration fail could be caused by the liquid handling section of the analyzer, including the solutions. Noise may be due to air bubbles sticking in the pipework and cuvette. Degassing of the sample is normal due to pre-heating on entry into the liquid handling section. However, the analyzer is designed so this does not normally affect its performance. If the problem is excessive, a manual chemical clean should be performed see IM/NAV6S. Note. The analyzer is not suitable for samples with very high levels of entrained gas. Additional external sample de-gassing may be required in these applications. 3.2 Single-Stream Mode for Maintenance To carry out any maintenance it is necessary to switch to a single stream provided that a stream is available. If not, a solution can be introduced via the secondary calibration valve. This stops the multi-stream sequencing and enables the display and the current output to respond to immediate color changes in the cuvette. This mode is used to check the basic performance of the analyzer, such as response or drift, without waiting for the normal stream update. Note. In a multi-stream analyzer, ensure that only 1 stream is enabled for maintenance see IM/NAV6S. Selecting more than one stream puts the analyzer into multi-stream operation. 3.3 Effects of Loss of Power to the Analyzer The automatic action taken by the analyzer following an unforeseen loss of power is dependent upon the length of time the power was off and what was happening at the time of the power loss. The following table shows the automatic functions that are performed: Analyzer Status Between calibration or clean sequences During clean sequences Period of Loss of Power < 24 hours > 24 hours purge 2. Recovery purge 2. Recovery 3. Restart/Clean 4. Recovery purge 2. Clean sequence 3. Recovery Purge 2. Clean sequence 3. Recovery Table. 3.1 Automatic Functions by Status/Power Loss 2 INF09/020 Rev. B

3.4 Simple Checks Note. Before carrying out tests on the multi-stream version, it is essential that only one stream is selected. Caution. When carrying out the following procedures, take care not to overtighten the connectors as this can cause the tubes to distort and impede or block flow. Tighten the tube connectors finger-tight and then add a further 1 /4 turn using a6mmspanner. 3.4.1 Unstable or Erratic Readings, Calibration or Zero Factors Visual check for air bubbles Action Check for air bubbles in the reaction block and in the tube feeding the optical unit. If bubbles are found, check if they are present in the sample line or originate from one or more of the reagents. 1. From the sample line: a. Check tube connections from the constant-head unit to the pump assembly tighten if necessary. b. Check the solenoid valves and ensure the screws on the reverse side are tight. 2. From the reagent line(s): a. Check the pump tube connections. b. Check the reagent bottle caps. 3. In the reaction block and tubing: a. If bubbles in the reaction block and tubing are flowing smoothly, the problem is likely to be a poor seal on one or more of the tube connectors. b. Tighten tube connections as above. If the problem persists, replace the affected tubing assembly. 4. If bubbles are pulsing: a. Check if there is a blockage or partial blockage. Blockages can be caused by overtight tube connectors. b. Check the tightness of all connectors in the sample or reagent line affected. Degassing in cuvette To check for degassing in the cuvette: 1. Check the optical measurement mv values see IM/NAV6S. 2. If the mv values are unstable, carry out a Manual System Clean see IM/NAV6S. INF09/020 Rev. B 3

3.4.2 Low/High Calibration Factors Follow the guidance in the Operation General Information section see IM/NAV6S. If this does not rectify the situation, perform the following checks: Secondary calibration valve Secondary calibration solution flow Action To check the valve: 1. Manually energize the secondary calibration valve see IM/NAV6S ('Manual Test Settings'). 2. Check that the LED on the main board is lit and that there is an audible click from the solenoid valve. To check the secondary calibration solution: 1. Unscrew the lid on the secondary calibration solution bottle and withdraw the level sensor assembly to introduce some air so that it is possible to see if the solution is flowing. 2. De-energize the secondary calibration valve and replace the bottle lid. If solution is not flowing or it is pulsing: a. Check the sample connections on the lid and the solenoid valve; ensure they are not overtightened. b. Check the pump tubing and capstan, replace if necessary. 3.4.3 Analyzer Stability and Response Test If the analyzer exhibits short term drift or instability, see IM/NAV6S, Appendix 3.4.1 for an overview of possible causes. If this action does not help to resolve the problem, run the analyzer on a fixed concentration silica solution (see below) to establish if the analyzer is unstable or if the drift or changes in the reading are real changes in the sample. Short term drift or instability Procedure and Action Run the analyzer on a fixed concentration silica solution. To perform this procedure: 1. Manually energize the secondary calibration valve and exit 'Manual Test Settings' in 'Test' mode. 2. Change the operator view to show the chart screen see IM/NAV6S. 3. Run the analyzer in this way for as long as required to check for instability or drift. 4. De-energize the secondary valve when finished. 5. Follow the checks and procedures detailed in 'Unstable or Erratic Readings' see IM/NAV6S. 6. If the above checks do not resolve the issue, carry out a 'Manual System Clean' see IM/NAV6S. 4 INF09/020 Rev. B

3.4.4 Checking the Optical Unit If the analyzer fails to perform calibrations and normal measurement routines correctly, check the Optical Unit against outputs as a diagnostic/remedial procedure. Excess mv values Optical photocells Action To check excess mv values: 1. Press and select 'Diagnostic Information', 'Measurement Status'. The 'Measurement Status' window is displayed and a mv signal value is displayed for the cuvette and the cuvette reference sensors. 2. If the mv values are unstable and are moving by more than ± 10 mv over a short period of time, there could be bubbles developing within the cuvette. 3. If bubbles are confirmed in the cuvette, check the tube connections see IM/NAV6S and if the problem is not resolved by this procedure, perform a manual system clean see IM/NAV6S. If there is no response or calibrations fail, check the output of the optical photocells as follows: 1. Empty the cuvette of reacted chemicals: a. Press the key. b. Select 'Calibration & Maintenance'. c. Select 'Manual Test Settings'. d. Select 'Test Pumps'. e. Switch the reagent pump off and increase the speed of the sample pump to 5 rpm. f. Run the sample pump at the increased speed for 5 minutes. g. After 5 minutes have elapsed, set the sample pump speed back to its normal (previous) speed. 2. View the mv values in the 'Measurement Status' window. 3. If the optical unit is functioning correctly, the mv values are within the following ranges: Cuvette: 1000 to 2300 mv Cuvette Ref: 1000 to 2300 mv 4. If the Cuvette Ref signal is low, replace the optical unit part number AW600 091. 5. If the Cuvette signal is low, perform a manual system clean see IM/NAV6S. If this does not resolve the problem replace the optical unit part Number AW600 091. INF09/020 Rev. B 5

ABB Limited Measurement & Analytics Oldends Lane Stonehouse Gloucestershire GL10 3TA UK Tel: +44 (0)1453 826661 Fax: +44 (0)1453 829671 Email: instrumentation@gb.abb.com ABB Inc. Measurement & Analytics 125 E. County Line Road Warminster PA 18974 USA Tel: +1 215 674 6000 Fax: +1 215 674 7183 abb.com/measurement We reserve the right to make technical changes or modify the contents of this document without prior notice. With regard to purchase orders, the agreed particulars shall prevail. ABB does not accept any responsibility whatsoever for potential errors or possible lack of information in this document. We reserve all rights in this document and in the subject matter and illustrations contained therein. Any reproduction, disclosure to third parties or utilization of its contents in whole or in parts is forbidden without prior written consent of ABB. ABB 2018 INF09/020 Rev. B 08.2018