Alfred ECMO Circuit Priming and Storage Guide

Similar documents
1.060 Set-Up and Priming of the Bypass Circuit. Perfusion Technology Department. Perfusionists

Heart/Lung Perfusion Packs INSTRUCTIONS FOR USE

Skills & Scenarios 81

Skills & Scenarios 81

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

Contacts. Quick Start Guide

C O L O G Y. New. g u O N. l o t a C a. The high security connection

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

Patients Guide for Self Administering Intravenous Antibiotics General points

MyoSure Setup Simplified

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

Haemodialysis. Bloodlines, Tubing Systems, Accessories Product Range. Product name / Headline Subline

Digital Display large color coded LED display is easy to read precise measuring shows actual cuff pressure

New. The high security connection

WARNING: All medications should be closely monitored to detect any alterations in the effective concentration due to the ultrafiltration process.

AQUARIUS Continuous Renal Replacement Therapy Competency

Parts identification:

Bloodlines, Tubing Systems, Accessories

AR251/AR251-B. Venepuncture and Infusion Arm. Instruction Manual

Standard. Bag & Stand Advanced Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00442

OPERATOR S MANUAL HOTLINE 3 Blood and Fluid Warmer

CENTRAL IOWA HEALTHCARE Marshalltown, Iowa

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

InFlow Modular perfusion bioreactor

Chapter 40. Fluid, Electrolyte, and Acid Base Balance. Procedures Checklist INTRAVENOUS THERAPY. Procedure 40.1: Starting an Intravenous Infusion

(Delivery System) Pocket Guide. Series 3 software

Chest Drains. All Covenant Health Intermediate Care Nursery staff. Needle Aspiration

Welcome to the Specialized Medical Services Respiratory training webinar series!

Standard. Abdominal Aortic Aneurysm (AAA) Repair Trainer. Part No: 60610

IVIG Educational Presentation

Standard. Standard. Venepuncture Arm. Light. Part No: Brown. Part No: Black. Part No: 00332

Abdominal Aortic Aneurysm (AAA) Repair Trainer User Guide

EMERGENCY EQUIPMENT TROLLEY DAILY SIGNATURE SHEET. Please refer to the Emergency Equipment stock poster for detailed information

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

Introduction. Medley TM Medication Safety System. Alaris Medical Systems

patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

HAEMOFILTRATION GUIDELINES

EASYPUMP II THE FLEXIBLE SOLUTION FOR SHORT AND LONG TERM INFUSION THERAPIES

ila ACTIVVE SYSTEM PLATFORM

OPERATING MANUAL LO-CROSS-FLOW CF M

Troubleshooting Guide Aquarius

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Armfield Distillation Column Operation Guidelines

Ventam 85 Installation & Commissioning Instructions

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

Information for patients and nurses. Rocket Ambulatory Chest Drain

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

Tourniquet Systems Pressure Infusion

Installing ASI Umbilical Delivery Systems

AQUADEX FLEXFLOW SYSTEM QUICK REFERENCE GUIDE

Alaris Pump module FAQs

INFUSION SYSTEM (110V) (220V) USER S MANUAL

ARTERIAL MONITORING. Final Approval: VP Operations, Chief Nursing Officer Distribution: ICU, PACU POLICY:

AQUARIUS Continuous Renal Replacement Therapy with Regional Citrate Anticoagulation (RCA) Competency

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Lipodrop 2.0 device coating

Clinician Guide. Baxter Elastomeric Devices. Promoting patient-centric care

Ultrasound Compatible Lumbar Puncture/ Epidural Simulator

Oxygen Dialflow Meter. Instructions for Use

System Pressure Manager Standard & System Pressure Manager Plus

ECMO Levitronix CentriMag Protocols Neonatal & Infant

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

ETT Olfactometer: Getting started guide 2013a. ETT Olfactometer quick start

Varithena Delivery System Instructions for Use

HEALTH SERVICES CODE A.6 CATEGORY: RN - SNP

Instructions For Use. Elastomeric Infusion Device

Manual Solvent Purification System. Innovative Technology, Inc.

TJF-Q180V Cleaning and Disinfection Checklist

Summary of Cardiac Anesthesia Technician Responsibilities

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

Procedure for Operating Columbia (Nevis) LHe Electron Bubble Chamber Cryostat. Version 5.1

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

My CoughAssist. A patient guide to CoughAssist T70. Please visit

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

Quiz for Module 2: Aquarius Training (Rev. 1) Lesson 1: Introduction to Aquarius

Varithena Delivery System Instructions for Use INSTRUCTIONS FOR USE Please read the Product Monograph before using the product.

MANITOBA RENAL PROGRAM

ECONORESS ELECTRONIC EPS & EPT - ENHANCED PRESSURISATION SET INSTALLATION OPERATION & MAINTENANCE DOCUMENTATION

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

Shared Haemodialysis Care Handbook. Name Hospital number: Shared HD Care Nurse: Date: Machine type: Dialysis Unit:

Dialflow Regulator. Instructions for Use

Instruction Manual Updated 7/26/2011 Ver. 2.2

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

R E D I C O N T R O L S

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

TITLE: Pulse Oximetry COMPETENCY #: Resp #1 NEW COMPETENCY REVISION DATE: 10/18/12 EMPLOYEE NAME. DATE INITIAL RE-EVALUATION ANNUAL (if required) PRN

Nita Newborn Model 1800 User Manual

Title/Description: Fuel Cell Fill Procedure

Module 2: Aquarius Training

Adult All Integrated Mini Bypass System

BIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 AUDIT TO BE COMPLETED BY (DATE):

Bunnell LifePulse HFV Quick Reference Guide # Bunnell Incorporated

Model VR6 System. Installation, Operation & Maintenance

Peripheral Venous Catheter Placement Simulator

Copyright, 2005 GPM Hydraulic Consulting, Inc.

Procedure 85 Attaching The Humidifier To The Oxygen Flow Meter Or Regulator. Procedure 86 Administering Oxygen Through A Nasal Cannula

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

Transcription:

Alfred ECMO Circuit Priming and Storage Guide 1 March 2012 Check List Page 2. Assembly, Priming and Storage Page 3. Prepare ECMO trolley and Circuit immediately before use Page 6. Connection to Patient and Initiation of Support Page 7. ECMO circuit assembly diagram.. Page 8

Check List for Commencement of ECMO March 2012 2 Assembly, Priming and Storage Assemble full ECMO circuit (remove priming ports on access line) Purge entire circuit with CO2 Spike Normal Saline bag and Transfer Circuit to ECMO Trolley Prime ECMO Circuit o Transfer 0.9% normal saline from IV bag to circuit reservoir o Remove yellow cap from oxygenator o Prime Circuit (Release both clamps between reservoir and circuit) o Run ECMO pump o Remove residual CO2 from connectors and oxygenator Prepare trolley for Storage o Recap oxygenator o Cable tie ( 3) o Clamp return and access lines close to Lovell connections Establish trolley in storage location o Plug in power o Cover Prepare circuit immediately before use: In Storage location Add 100mls of 20% Albumin to reservoir bag Establish flow signal and circulate albumin prime Activate circuit heater. Raise water temp to 36.5 deg C. Detach power Transfer to bedside Prepare circuit immediately before use: At the bedside Connect Power Start Gas Flow to the oxygenator Start pump and check heater Clamp circuit (post oxygenator) Connect ECMO circuit to cannulae Go on ECMO Restart water heater

3 A. Assembly and Priming (Surgically Clean Phase) Prepare surgically clean surface. Put on mask, hair cover, sterile gown and sterile gloves. Assistant is not sterile 1. Assemble full ecmo circuit a. Place following sterile items on the sterile drape: 20 ml syringe, scissors, gas-line filter, line clamp, 5 x three way taps with extension tubing b. Open Lovell attachment and place on surface to assemble c. Open ECMO circuit and place on clean surface to assemble and discard paper envelop d. Using sterile scissors, cut out the two priming ports on distal end of access tubing e. Connect this end of ECMO tubing to centrifugal pump inlet (central) to complete the ECMO circuit f. Attach extension loops with 3-way taps securely to Luer lock ports at points E, F, G, H, I, (see diagram) and ensure all taps are capped and turned off g. Cut ECMO circuit midpoint between blue and red markers h. Connect Lovell attachment to ECMO circuit. (Y limb of Lovell attachment goes on access line (blue markers). Other limb attaches to return line. i. Open all clamps on the Lovell attachment. 2. Purge entire circuit with CO2. a. Place tubing clamp on green patch (M) between ports E & F. b. Open port E c. Fit sterile 0.2 um filter to port E using piece of clean plastic tubing (3/16 th inch) d. Assistant connects CO2 line to.2 um filter and run CO2 at 1.0-3L per min. e. Expel CO2 via Lovell attachment filling line first. Then apply clamp C on filling line and Reservoir bag should start to fill f. Expel CO2 via all the other 4 circuit ports (I, H, G, F) in order g. Primer compresses bag and ensures CO2 venting from port F h. Primer changes 0.2 um filter to port F. i. Primer removes clamp from ECMO line at M (indicated by green tape). j. Assistant briefly runs CO2 at 2 l/min for 1.5 min (reverse flow) until CO2 leaves port E k. Assistant detaches 0.2um filter from port F and turns off CO2 flow. l. Primer ensures excess CO2 expelled from reservoir via the filling line, m. Primer closes 3- way taps at ports E and F and applies sterile caps to Luer connections.. n. Assistant connects spike of the filling line to priming solution (1 L (Normal) 0.9% Saline).

4 B. Assembly and Priming: (Transfer to Trolley) 3. Transfer circuit to ECMO trolley a. Hang priming solution and reservoir bag from IV pole on trolley b. Assistant to hold access and infusion lines together to left of trolley c. Place Rotaflo centrifugal pump head in to external drive unit. (Do not apply ultrasound paste to flowmeter at this stage) d. Attach oxygenator to holder e. Optimize position and inclination of rotaflo remote drive. (Drive head must not be inverted; outflow line must be close to vertical to aid de-airing of pump Remote drive should not be more than 1 foot below oxygenator) f. Make sure access and return lines are not tangled or twisted. 4. Prime ECMO circuit a. Transfer approximately 1.5L Normal saline 0.9% from IV bag to circuit reservoir i. Ensure clamp A is applied and tubing clamp applied at point B of circuit ii. Open clamp C and fill reservoir bag with approximately 1.5 L priming solution (Leave reservoir bag on trolley drip stand and apply pressure to Normal Saline 0.9% bag.) iii. Once reservoir bag filled apply clamp C. b. Commence Cyrstalloid Prime i. Remove yellow cap at L to aid de-airing process during prime. ii. Check and remove all clamps from ECMO circuit. iii. Open clamp A then B on Lovell attachment. iv. Make sure access line and pump head is virtually free of bubbles by tapping (if required) v. Ensure AC power indicator lit (if not check power cord connection; surge protector at the bottom of trolley; and circuit breaker at rear of pump) vi. Turn on console and increase speed to 1000 rpm (flow signal is not required) vii. Primer opens taps sequentially at E, F, G, and H briefly to ensure they are primed (Use sterile gloves. No touch technique of the 3-way tap). Caps should be placed on a sterile surface while 3-way tap is handled to prevent contamination during de-airing of side-ports. Take care not to contaminate surface viii. Detach and tap oxygenator and shake to displace bubbles to apex of oxygenator ix. Visible bubbles (>2mm) on the outlet side of the oxygenator can be removed via the fluid phase port (Port I)

5 c. Removing small residual CO2.bubbles from the oxygenator i. Engage Oxygenator in holder ii. Increase pump speed to 2000-3000 rpm (reduce if bubbles created in the reservoir bag with turbulence) and run for 5 minutes iii. Partially occlude ECMO circuit with tubing clamp just distal to oxygenator to raise blood phase pressure in the oxygenator to help expel CO2 bubbles from the oxygenator (ensure port L open) iv. Check no visible bubbles remain by disengaging the oxygenator and inspecting while tapping the oxygenator with palm. 5. Prepare trolley for storage a. Replace yellow cap for apical air release port (L) b. Ensure caps on all taps are firm and taps are turned off to the circuit c. Turn pump off at front console d. Assistant applies cable ties to all ECMO tubing connections fashioned during the priming process (3 in total): i. Between access line and pump head inlet and ii. Lovell connection points (Line A and B). (All other connections should come with cable ties applied by the manufacturer. 6. Establish trolley in storage location a. Disconnect from wall power b. Transfer trolley to designated storage area in ICU c. Make sure trolley is connected to power at new location d. Cover whole rig with clean drape.

6 C. Prepare ECMO trolley and Circuit immediately before use: Initial Procedures In the storage location. 1. Add albumin a. Connect 100 ml Bottle of 20% albumin to filling line (after removal of Normal Saline 0.9% bag) and insert vent needle. b. Before releasing clamp C on the filling line, ensure the pump is not running (to prevent the possibility of massive air entrainment) c. Transfer albumin to reservoir bag by opening clamp C then elevating the albumin bottle and lowering the reservoir bag d. Re-apply Clamp C on the Filling Line once the Albumin is transferred e. Ensure any gas bubbles that have collected during storage are not located in the pump head f. Apply contact paste to flow sensor (it must completely surround the outflow tube) g. Start Rotaflo pump and establish flow (Clamps must be removed from line A and B) h. Set Rotaflow pump speed to 1000 rpm i. Make sure flow sensor is working properly i.e digital readout of pump flow is present. (if not reapply contact gel and try again) j. Check for Oxygenator leaks (inspect for leakage of priming solution) i. heater connection ports on the oxygenator ii. Fresh gas exhaust port 2. Activate heater/cooler. a. Fill heater with fluid via cap on top surface and make sure water level just below maximum b. Connect hosing of water heater to water connections at J and K on oxygenator (see fig 1.) c. Increase speed of rotaflo to 2000 rpm and run for 5 min. (water temp should settle at to 36.5 deg. C.) 3. Detach Power a. Turn off both rotaflo pump and water heater. b. Detach power cord. 4. Transfer to cannulation room

7 D. Prepare ECMO trolley and Circuit immediately before use: Procedures at Patient Bedside 1. Prepare the ECMO Trolley for active use a. Plug ECMO trolley into new power supply b. Connect to room supply of oxygen and air (if blender alarms the fault must be rectified) c. Briefly turn on Rotaflo pump and water heater to ensure they are both working properly then turn both of them off d. Ensure (yellow) cap is on port L e. Start gas flow to oxygenator at 5 l/min 100% oxygen f. Apply clamps low down on both Reservoir Bag lines (A and B) g. Apply safety clamp immediately post oxygenator. E. Connect Circuit to Patient 1. Connect distal ends of ECMO circuit to cannulae a. Assistant (non-sterile) clamps the ECMO circuit just be below the connection points to the Lovell Reservoir, starting with the access side and extends the tubing toward the cannulator for cleaning with Betadine b. Cannulator applies a second (sterile) clamp on the access line approximately 30cm below the Lovell connection point and cleans this distal segment (between the two clamps) with Betadine c. Once thoroughly cleaned with Betadine, the Cannulator then cuts the cleaned segment (close to the nonsterile clamp) and transfers the (cleaned) tubing to the cannula (removing torsion and tangles) d. Process is repeated for the Return line e. Cannulator connects the cleaned segments of tubing to the appropriate cannulae using an underwater seal 2. ECMO Initiation a. Remove clamps remaining at cannulae connection points b. Ensure safety clamp (post oxygenator) remains c. Ensure pump running ( Free Mode ) at 1000 revs and flow signal established (should be reading 0 L/min) d. Ensure gas flow to oxygenator connected e. Remove safety clamp SLOWLY (especially for V-A ecmo where retrograde flow should develop) watching the saline-blood interface in the access line tubing. (if retrograde flow occurs as in V-A ecmo you should see blood entering the return line f. Increase pump speed to achieve gradual inflow of the circuit prime and continue to gradually increase speed until the desired flow is achieved g. Restart water heater.

8 1 litre bag of Normal Saline (2) Filling line C Reservoir Bag Lovell Attachment Oxygenator (Side-view) A B Access Line Return Line J K Pump Head Oxygenator (End-view) I ECMO Circuit J & K are Water Connections for circuit heater Yellow Cap L E M F G H