SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

Similar documents
JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

Thermcraft Tube Furnace General Use

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Standard Operating Manual

Birck Nanotechnology Center

RAPID THERMAL PROCESSOR (Annealsys AS-ONE 150) Lab Manual

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

STS DRIE User SOP Location: NRF Cleanroon, Dry Etch Bay

Xactix XeF2 OPERATION MANUAL

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

OXFORD PLASMALAB 80PLUS (CLOEY)

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

University of MN, Minnesota Nano Center Standard Operating Procedure

Savannah S100 ALD at SCIF, UC Merced Standard operating Procedure

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

Arizona State University Center for Solid State Electronic Research. Table of Contents. Issue: C Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8

Armfield Distillation Column Operation Guidelines

Nordiko Metal Sputtering System Standard Operating Procedure

5890II GC Standard Operating Procedure 9/2/2005

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Angstrom Dielectric Sputterer Operation Manual

Usage Policies Notebook for Trion RIE / ICP Dry Etch

Standard Operating Procedure. For. PVD E-Beam

Notes-PECVD: Chamber 1

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

University of Minnesota, MN Nano Center Standard Operating Procedure

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Revised: June 7, 2017

The MRL Furnaces USED FOR THIS MANUAL COVERS

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Xactix Xenon Difluoride Etcher

Standard Operating Manual

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Birck Nanotechnology Center

Unaxis PECVD. SiH4 (5% in He)

Standard Operating Manual

CLEANLINESS STANDARD NCPRE tool contamination classification

Standard Operating Manual

University of Minnesota Nano Fabrication Center Standard Operating Procedure

Touch Screen Guide. OG-1500 and OG Part # T011

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

Usage Policies Notebook for STS DRIE System

SPLIT HOPKINSON PRESSURE BAR OPERATIONS MANUAL 5/9/2014

AX5000 Operational Manual

Standard Operational Procedure for Capillary Electrophoresis

Operating Procedures for Metal Evaporator I

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

STS PECVD Instructions

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Operating Procedures for the. SAMCO ICP RIE System

Oerlikon Sputtering Evaporator SOP

Standard Operating Manual

Standard Operating Manual

Using the Akta Prime plus October 22, 2012

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Standard Operating Procedure

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

Varian 3300 Gas Chromatograph Abbreviated Operating Instructions

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

Karl Suss MA6 Mask Aligner SOP

Equipment Operating Procedure Glove Box

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cambridge NanoTech: Savannah S100. Table of Contents

WATER CONTROL SYSTEM QUICK START

High Pressure Chem-SCAN Operating Manual

STS ICP-RIE. Scott Munro (2-4826,

BUBBLER CONTROL SYSTEM

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

Arizona State University NanoFab PLASMATHERM 790 RIE. Version A

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

University of MN, Minnesota Nano Center Standard Operating Procedure

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

Warnings: Notes: Revised: October 5, 2015

Operation of the Perkin Elmer TGA-GC/MS

MP15 Jockey Pump Controller

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

Boyle s Law: Pressure-Volume Relationship in Gases

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

Arizona State University NanoFab XACTIX ETCHER. Rev A

KEM Scientific, Inc. Instruments for Science from Scientists

Race Screen: Figure 2: Race Screen. Figure 3: Race Screen with Top Bulb Lock

TR Electronic Pressure Regulator. User s Manual

Agilent ICP-MS Interactive Troubleshooting tool for Plasma ignition problem

University of Minnesota Nano Center Standard Operating Procedure

User Manual for the Mars Calibration Bench

New Cortex Software User Guide

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Standard Operating Procedure Porosity Analyzer - The Micromeritics Tri-Star II (surface area) with Micrometrics VacPrep 061 (Sample Degas System)

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Transcription:

Page 1 of 9 SSI Solaris 150 RTA The Solaris 150 RTA is a rapid thermal annealing system capable of handling sample sizes up to 100mm (4 diameter) or smaller. The system can anneal in N 2 and Forming gas (N 2 with 4%H 2 ) up to a maximum temperature of 1100 C. Process time is limited to 5 minutes. The system is reserved for conventional semiconductors, dielectrics and metals. See the acceptable materials list at the end of the SOP. Prerequisites for operating the Solaris RTA system: a) Obtain a NRF ID (if you do not already have one) by completing the NRF Lab Use Request Form and safety training. b) Receive one on one training and certification from NRF Staff. Discuss your process with a staff member. Safety DANGER! Do not remove the covers of the instrument. Do not modify the instrument. HOT Components - The User must observe caution when loading/unloading samples from the sample stage or graphite susceptor. Samples are very hot when immediately removed from the process chamber. Sufficient cooling time must be allowed prior to sample handling. Chilled Water Leaks Due to chamber designs, we have had instances where chilled water has leaked out of the system. In the event of a leak, press EMO button and turn chiller off, toggle switch on the chillers front panel. Immediately contact NRF staff. 1.0 Pre-Operation 1.1 Tool Reservations may be made via the NRF Reservation Page. http://nimet.ufl.edu/servicecenter/resources/default.asp 1.2 Change gloves. WARNING No solvents or liquids are allowed near the machine, change your gloves before operation!! 1.3 Log into the tool by using the TUMI computer in this room (NRF 122).

Page 2 of 9 2.0 Hardware overview The Solaris RTA consists of the following components: computer, main oven unit, chiller, gas sources, and oxygen sensor. See the image below for system layout. O 2 sensor Oven unit Computer Chiller Figure 1. System layout. 3.0 User Interface The Solaris RTA operation is controlled through the software interface. The appropriate screens and functions will be addressed through the Operation portion of this SOP.

Page 3 of 9 4.0 Operation 4.1 Log into the Solaris RTA through the TUMI computer located in room 122. The TUMI computer is located in the back right corner of the room. This will enable the computer monitor and the chiller for the RTA. 4.2 Power on the main power on the oven unit. Figure 2. Main oven power switch and EMO button. Do not use the EMO as a power on and power off button. The EMO is the Emergency Manual Off button and should only be used to turn off the equipment in an emergency. 4.3 To load the sample, slowly lower the door handle with both hands and gently pull the sample door open. The quartz sample tray will withdraw with the door. The control thermocouple (TC) is located under the small Si disc on the sample tray. The sample tray is loosely attached to the door. NOTE: The sample or 100mm Si carrier must be placed over the TC for accurate processing. For small samples load small samples onto a Si carrier wafer (100mm). A Si carrier wafer with a second Si wafer as a lid can be used with transparent samples or samples requiring a group V overpressure for annealing. For large samples place the wafer sample on the 3 support pins and TC. (Refer to figure 4)

Page 4 of 9 Gently slide the door closed. Pay close attention to the side to side clearance of the sample tray and the RTA chamber. Raise the door handle to lock the door. THE SAMPLE TRAY IS VERY FRAGILE, USE WHEN OPENING AND CLOSING THE TRAY DOOR. Figure 4. Image of quartz sample tray with support pins and TC exposed (red circles). 4.4 Turn on the process gas. The N 2 is already on and available. Forming gas is in the gas cylinder located to the left of the system. Open the main tank valve, the regulator should already be set to 10psi. Please confirm the process gas pressure (high and low side of regulator) before continuing. If you have any questions, please consult with staff. Oxygen can be available by selecting Oxygen on the pneumatic switch over the gas bottles on the left wall. Open the main tank valve and secondary valve. The output pressure should be set at 10psi. On this setting, the Honey Well oxygen sensor will not read the O2 content in the chamber. The gas selecting manifold is designed so Oxygen cannot be run simultaneously with forming gas.

Page 5 of 9 4.5 Start Solaris 3.8 software. The main oven unit must be powered on before the Solaris software is launched. Failure to do so in the prescribed order will lead to a communication error. Once the Solaris software is running select Start RTP Process. The next screen (see below) will become visible. The menus that should be used by the user are Recipe Management and Run Process. All other menus are for NRF Staff and should not be accessed by users. Recipe management in this screen, the user may create or edit recipes for their processing. Many of the basic recipes are already available, these may be edited and saved under different names. The recipes are created as steps following a process flow. Each step can be edited for ramp rate, gas flow, hold time, etc. Steps can be added or inserted, depending on the type of processing required. The image of the Recipe Management screen is shown in Figure 5. NOTE: the default purge gas is N 2 and will be flowing at 3.0 SLPM when the Run Process menu is opened. It is highly recommended that the user edit an existing basic recipe and refer to NRF staff regarding the limits of the thermal process. Process time is limited to 5 min. Exceeding this time will cause the heaters to trip. Recipes are selected through a drop down list. When prompt for a password, click [OK]. Password is not required.

Page 6 of 9 Some basic recipes are: 300C_N2_60sec 400C_N2_60sec 500C_N2_60sec 600C_N2_60sec 700C_N2_60sec 800C_N2_60sec 900C_N2_60sec 1000C_N2_60sec Figure 5. Screen for Recipe Management. Once the recipe has been edited, it can be saved with a different file name. The save command will return the user to the RTP Process menus. 4.6 Next select the Run Process menu. In this menu, the user can select the process recipe and track the process flow with the graphic tool that follows the RTA process. Figure 6. a. Select the process recipe from the drop down list. Leave the calibration file set to factory calibration. The recipe cannot be edited from this page. b. Set the Overtemp Setpoint to 20 C above the recipe maximum temperature. This will prevent a possible temperature run away to protect the RTA system and the user s sample. Power Limit = 100%, Power Limit Time = 3.0, Purge MFC = 3.0. c. Leave Learn Mode off. This will prevent the software to from automatically adjust the recipe ramp and hold parameters. d. Leave the TC type set to Tray TC. This will read the temperature of the susceptor or the sample wafer as the process temperature.

Page 7 of 9 e. When ready, click the Start button on the menu. The user will be prompt to enter a sample name and user name, simply click the START PROCESS button and skip these fields. This will begin the RTA process. Recipe progress can be monitored on the graphical display as well as the table under the display. Figure 6. Screen for Run Process menu. Monitor the oxygen sensor readout which monitors the output gas of the RTA system. The lowest recorded oxygen concentration is 0.8%. Typically a 60 second pre purge is sufficient to achieve this limit. If the oxygen has not reached 0.8% before the temperature ramp step, stop the process with the STOP button. The purge time may need to be increased or there may be an issue with the system. f. When the process is complete, select Next Wafer. The Run Process Menu page will remain until the user clicks the Return to Main Menu button. It is generally a good practice to remain on the Run Process Menu page to monitor the sample temperature. 4.7 Unloading sample When the sample is cool enough, it can be removed from the system. Lower the door handle and gently pull the door open. Carefully remove the sample or susceptor from the right side of the tray. This will prevent possible damage to the TC or TC wires. Close the tray carefully when sample removal is complete.

Page 8 of 9 4.8 It is highly recommended to run an oxygen chamber clean anneal using the supplied Si wafer carrier before your samples are run. This will eliminate any potential contamination inflicted by previous users. The recipe runs oxygen at 1000C for 30s. Contact NRF staff for recommended recipe name. THE SAMPLE AND WAFER HOLDER MAY BE VERY HOT (OVER 100 C). USE TO NOT TOUCH THE SAMPLE TRAY OR SUSCEPTOR. 5.0 Shut Down Exit the Solaris software and return back to the main Windows desktop. Turn the main oven power off. (figure 2) Close any process gas cylinders by closing the main tank valve. The compressed air and N 2 should remain on. Log off the Solaris RTA system via the TUMI, record any comments. The Chiller will continue to run for an additional 10 minutes after logging off. This is to ensure proper cooling of the RTA oven.

Page 9 of 9 Acceptable Materials List. Semiconductors: Dielectrics: Si Ge Ga(Al,In)As Ga(Al,In)P Ga(Al,In)N Ga(Al,In)Sb Zn(Mg)O SiC Metals: SiO 2 Si 3 N 4 Sc 2 O 3 Mg(Ca)O Al 2 O 3 HfO 2 TiN TiO 2 Ti Al Pt Au Ni Pd Ir Boride containing alloys Nitride containing alloys