GILMONT ACCUCAL FLOWMETERS

Similar documents
Laboratory Flowmeters

Installation, Operation, and Maintenance Manual

Operation Manual Piston Sensed Gas Pressure Regulators

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

INSTRUCTIONS FOR FM4460N SERIES HIGH PRESSURE, METAL TUBE FLOWMETERS

Installation, Operation, and Maintenance Manual

BP48-1 AND SSP48-1 IN-GROUND POST INSTALLATION INSTRUCTIONS

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

PTFE-SINGLE GLASS FLOW METERS

Compatible with a Wide Range of Gases or Liquids for Laboratory or Plant Applications

OPERATING MANUAL EASY-LOAD PUMP HEADS

INSTRUCTIONS FOR SERIES 50 & 50K FLOWMETERS

KOBOLD KSM FLOWMETER. User Instructions

shutoff valves 900 Series psig Manual Shutoff Valves

Pressure Dump Valve Service Kit for Series 2300 Units

CSA Sample Draw Aspirator Adapter Operator s Manual

PP-1, PP-1W, PP-2, & PP-3

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

INSTALLATION AND OPERATING INSTRUCTIONS

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

INSTRUCTIONS FOR SERIES 150 TWO TUBE GAS BLENDER

2 Sentry MVS Installation, Operation & Maintenance

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI

LRS(H)4 Pressure-Reducing Regulator User Manual

Tank Blanketing Pressure Regulators RHPS Series

RK-IR Sample Draw Aspirator Adapter Operator s Manual

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER CONSTANT PRESSURE CYLINDER

Installation, Operation, and Maintenance Manual

PURA. Pure Gas Dewpoint Transmitter. Users Guide

Multiple Tube - Teflon PTFE-Glass Flowmeters

BULLETIN FOR ACCURATE INDICATION AND CONTROL OF GAS AND LIQUID FLOW IN LOW RANGES.

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

BPB26-1 AND SSPB26-1 PATIO BASE INSTALLATION INSTRUCTIONS

Model VR6 System. Installation, Operation & Maintenance

Instructions Manual. Variable area flowmeter 60 M-1. Series. instrumentation for fluids. Technical Data. C-MI-60M1 Rev.: 0 English Version

MUELLER. A Wall Type. Indicator Post. Reliable Connections. General Information 2. Technical Data/ Dimensions 3. Installation 4-5.

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

INSTRUCTIONS FOR MODEL SSC LOW DELIVERY PRESSURE GAS REGULATORS

Precision and Instrument Regulators 1/4" port size

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

Discontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves.

Multichannel Pump Head pages 1-8. Tête de pompe multicanal pages F1-F8. Mehrkanal- Pumpenkopf Seiten D1-D8

Instruction Manual - Diaframless TM Ejector Chlorine, Sulfur Dioxide and Ammonia

Anti-flood device Model B-1

INSTRUCTIONS FOR MODELS UPH & UPHS HIGH FLOW, HIGH PURITY GAS REGULATORS

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Single Tube Flowmeters

Your safety and the safety of others are very important.

Operation Manual. Diaphragm Sensed Gas Pressure Regulators

Binks SG-2 Plus TM Rotary 2 QT. PRESSURE CUP

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

Constant Pressure Crude Oil Container Model CPCCP

Related Products: Auto Drain Valve. Model/Specifications. Model

Binks SG-2 Plus TM 2 QT. PRESSURE CUPS Model No (Standard) Model No (with OscillatingAgitator)

TECHNICAL DATA ANTI-FLOOD DEVICE MODEL B-1 1. DESCRIPTION

Gas Proportioners - Back Pressure Compensated

Installation, Operation & Maintenance Manual

SPECIFICATIONS. Approximate Weight: 10 oz. Surface Finish: Ra micro inch or less MATERIALS OF CONSTRUCTION

MUELLER A A Non-Adjustable. Vertical Indicator Posts. Reliable Connections. General Information 2. Technical Data 3.

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER FILTER DRYER

DPC-30 DPC-100. Reference Manual

Variable Area Flow Meters

Installation/Care/Use Manual

RD(H)20/25 Pressure-Reducing Regulator User Manual

1200B2 Series Service Regulators. Instruction Manual

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

Section 3- Repair and Calibration Procedures

INSTALLATION, OPERATION, AND MAINTENANCE MANUAL WELKER RELIEF VALVE

Brooks Flow Controllers for Gas and Liquid Service Model FC 8744, Series FC 8800 and FC 8900

INSTRUCTIONS FOR SINGLE STAGE LECTURE BOTTLE REGULATORS

TECHNICAL DATA ACCELERATOR MODEL E-1

INSTRUCTIONS FOR MODELS APC & APG HIGH PURITY, CORROSIVE GAS REGULATORS

TECHNICAL DATA CAUTION

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

Multiple Tube Flowmeters

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

TECHNICAL BULLETIN. Sigma-Aldrich Pure-Pac II System. Technical Bulletin AL-236

HP-104 & HP-106 Tapping Machine OPERATIONS MANUAL

Pressure Dump Valve Service Kit for Series 3000 Units

Varea-Meter Flowmeters. For Metering and Controlling Flows of Process Fluids

DTG - LCD Digital Temperature Gauge USER MANUAL

! WARNING. Model PFC-1-G (direct acting) PFC-1-GR (reverse acting) Modulating Pneumatic Liquid Level Controls INSTRUCTION MANUAL MM-110B

APCO ARV CLEAN WATER AIR RELEASE VALVES. Model 50A

Installation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Welker Magnetic Indicator (With Gravity Mixer) Model CP2GM-HP

ATD LB PRESSURE BLASTER INSTRUCTION MANUAL

INSTALLATION / OPERATION / MAINTENANCE INSTRUCTION

Air Operated Hydraulic Pumping Systems to 50,000 psi

!!!! SERVICE MANUAL PRESSURE POT 2 GALLON. Service Manual: LT Washington St 931 Progress Ave., #7

Model PSI Compressor with 3-Gallon Air Tank 12VDC

Model Series 62 Constant Differential Relay

GETZ EQUIPMENT INNOVATORS RECOVERY SYSTEM CLEAN AGENT FE-36 / HALOTRON 1 PART NO. 4G59751 OPERATIONS MANUAL

AIR/OVER HYDRAULIC JACK 20 TON

Choosing the right miniature instrument and control valve is a snap

INSTALLATION & MAINTENANCE INSTRUCTIONS

MUELLER. A A Adjustable. Vertical Indicator Posts. Reliable Connections. General Information 2. Technical Data 3.

1800C and 1800C-HC Series Service Regulators

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May

Transcription:

OPERATING MANUAL GILMONT ACCUCAL FLOWMETERS 28W092 Commercial Ave. Barrington, IL U.S.A. 60010-2392 (847) 381-4888 (847) 381-7053 (Fax) 800-962-7142 www.barnant.com e-mail: barnant@barnant.com A-1299-0766 Edition 01

TEFLON- Reg TM E.I. DuPont de Nemours, Inc. Trademarks bearing the symbol in this publication are registered in the U.S. and in other countries.

DESCRIPTION Variable area Flowmeters (rotameters) are used to measure the flow rate of liquids and gases in a variety of laboratory and industrial applications. These meters consist of a spherical float moving vertically in a glass flowtube with a tapered inside diameter. As the flow through the flowtube increases, the float rises in the flowtube. A scale marked on the flowtube (along with calibration charts and correction factors for fluid density, temperature and pressure) is used to obtain accurate measurements of flow rate. These Flowmeters contain the following features: Correlated and direct reading all in one easy to read meter. Accuracy as good as ± 2% of reading. Also available with NISTtraceable calibration ±1% of reading accuracy at calibrated points. Rugged construction ideal for industrial and laboratory use. Stainless steel, PCTFE or PTFE and glass wetted parts. Interchangeable dual, direct reading scales. Available with or without 14 turn precision valve. Two scale cards are supplied for direct reading of air and water under standard operating conditions (70 F, 1 ATM) and a blank scale is supplied for use with the correlated scale of the meter. These scales are installed over the process connections at the back of the Flowmeter and held in place by the two palnuts (see Figure 2). An additional float made of stainless steel is supplied with each Flowmeter. This float can be used in place of the black glass float installed in the Flowmeter to extend the range of the Flowmeter by approximately 1.75 times for air and 2.12 times for water. UNPACKING Carefully unpack and check contents. If any components are missing or damaged, save the packing list and material and advise the carrier of the discrepancy. ACCESSORIES Flowmeter Stand Kit. Includes one base, one rod and two mounting clamps Flowmeter Base Rod, 18 in L Mounting Clamp, includes wood screws GF-4004 GF-4001 GF-4002 GF-4003 1

INSTALLATION 1. Unpack the Flowmeter and retain all packing material until proper product operation has been verified. NOTE: If used for vacuum applications or when back pressure compensation is required (as in gas blending), the metering valve must be positioned at the outlet (top of the Flowmeter) by inverting the frame assembly in relation to the flowtube assembly. NOTE: Meters are supplied with the 0-100 scale divisions on the flowtube set up for use with the printed tables. If direct reading scales are to be used, see the instructions at the back of the manual. 2. Before installing, open valve of the Flowmeter (if equipped) and check that the float moves freely in the flowtube. In some cases, floats may appear sluggish or do not move due to condensation resulting from temperature changes during shipping. It is advisable in such cases to blow a dry clean gas through the meter until the float is freed. 3. Leave the Flowmeter valve partly open during installation (it is shipped in the partly open position). 4. It is important that all lines to be connected to the Flowmeter are purged of any dust or other residual contamination prior to installing the meter. NOTE: In some applications a particulate filter should be installed at the inlet of the Flowmeter. Select a suitable filter for the particulate matter to be removed. 5. Isolation or shut-off valves should be installed in the system to permit removal of the meter for cleaning or servicing. 6. Mounting: a. The meter must be mounted in a vertical position. A deviation of 5 degrees or more from vertical will affect the accuracy of the meter. b. Connectors on the meter are supplied with wrench flats which must be held firmly when threading mating connectors, to prevent loosening of the connectors. c. Paste sealants are generally preferred. However, applications which are sensitive to contamination and require inert sealants, TEFLON PTFE tape is recommended. CAUTION: Care must be taken to avoid shredding of TEFLON PTFE tape which can foul meter operation. 2

7. Base Mounting: Secure the Flowmeter to the base by means of a 1/4 in-20 screw threaded in the hole in the bottom block. (See ACCES- SORIES for Flowmeter Base model number GF-4001.) Panel Mounting: Locate mounting holes per Figure 1. Secure the Flowmeter to the panel by means of nuts threaded over the inlet and outlet fittings. Hold the fittings across the flats with an open end wrench when removing the nuts so that the O-ring seal remains tight. 8. Be sure that the piping is adequately supported to prevent undue strain on the Flowmeter. WARNING: A leak test must be performed when hazardous fluids are involved. Perform the leak test using a non-hazardous fluid. Start with a low pressure and gradually increase. FLOWMETER FIGURE 1 MOUNTING DIMENSIONS DIMENSIONS-INCHES (mm) A B C D E F G H 65 mm scale 5.65 in 0.60 in 4.50 in 1/8-27 0.78 in 1.00 in 1.75 in 1.25 in (All Flowtubes) (143.5 mm) (15.2 mm) (114.3 mm) NPT (19.8 mm) (25.4 mm) (44.5 mm) (31.8 mm) 150 mm scale 10.02 in 0.60 in 8.81 in 1/8-27 0.78 in 1.00 in 1.75 in 1.25 in (200-235 Flowtubes)(254.5 mm)(15.2 mm) (223.8 mm) NPT (19.8 mm) (25.4 mm) (44.5 mm) (31.8 mm) 150 mm scale 12.64 in 0.79 in 11.00 in 1/4-18 0.95 in 1.50 in 2.13 in 1.75 in (240-250 Flowtubes)(321.1 mm)(20.1 mm) (279.4 mm) NPT (24.1 mm) (38.1 mm) (54.1 mm) (44.5 mm) 3

WARNING: Close inlet valve prior to start-up to avoid sudden pressure surges and/or shock. OPERATION 1. Gradually open the metering valve and adjust the flow to the desired rate using the valve of the Flowmeter. The impact of the float at the top of the flowtube can damage the meter if exposed directly to full line pressure. Avoid sudden pressure surges and shock by closing inlet valve prior to start-up. WARNING: Do not exceed the maximum operating pressure 200 PSIG (1379 kpag) or maximum temperature of 250 F (121 C). 2. Flow rate for air and water at standard operating conditions 1 ATM (98 kpag), 70 F (21.1 C) can be determined from reading the scale at the center of the float and appropriate direct reading scale or supplied flow tables. INSTALLATION INSTRUCTIONS FOR DIRECT READING SCALES Proceed as follows: 1. Select desired direct reading scale, i.e., desired range, float and media. 2. Position scale such that the desired scale faces the front of the Flowmeter. 3. Align zero mark on scale with zero mark on flowtube and tighten the palnuts to retain in position. See Figure 2. 4. Remove the four screws holding the Front Shield in place. 5. Remove the Front Shield. 6. Using a 3/16 in hex wrench, loosen the Cylinder Screw at the top of the Flowmeter, approximately 1 to 2 turns. 7. Rotate the flowtube to position the scale on the flowtube to the side. 8. Using a 3/16 in hex wrench, tighten the Cylinder Screw to secure the flowtube. 9. Reinstall the Front Shield and the four screws. 4

FIGURE 2 SPECIFICATIONS Accuracy: Greater of ±2% of reading or ±1 scale division using correlated flow tables. Greater of ±5% of reading or ±3 mm of scale using direct reading scale. Maximum input pressure: 200 PSIG (1379 kpag) Maximum fluid temperature: 250 F (121 C) Maximum external temperature: 200 F (93.3 C) Process Connections: 65 mm scale: All flowtubes 1/8-27 NPT (Female) 150 mm scale: 200-235 flowtubes 1/8-27 NPT (Female) 240-250 flowtubes 1/4-18 NPT (Female) Reproducibility: 5 Greater of ±1% of reading or ±1/2 scale division.

WARRANTY Gilmont Instruments warrants its product to be free of material or manufacturing defects for the period noted on the warranty card from the date received. If any such defect is noticed, please return the instrument directly to our factory for repair, in which case, it will be repaired at no charge. PRODUCT RETURN To limit charges and delays, contact the seller or Manufacturer for authorization and shipping instructions before returning the product, either within or outside the warranty period. When returning the product, please state the reason for return. For your protection, pack the product carefully and insure against possible damage or loss. Any damages resulting from improper packaging are your responsibility. TECHNICAL ASSISTANCE If you have any questions about the use of this product, contact the Manufacturer or authorized seller. Printed in U.S.A. 010499 6