TSE Blood Pressure Monitor Invasive series

Similar documents
Research Grade Blood Pressure Transducer User's Manual

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Columbus Instruments

Selecting the right pressure sensor for your application

MANUAL KPS Pressure Control Valve

C-Flow Coriolis Mass Flow Meters

HI 2314 HI 2315 HI 23151

Hand-held pressure calibrator with integrated pump Model CPH6600

Pneumatic high-pressure controller Model CPC7000

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

DPC-30 DPC-100. Reference Manual

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Installation and Operation Manual

MCQ Gas Blender 100 Series 3 Channels Gas Mixer

Instruction Manual Differential Pressure Transmitter

ISQB1 INTRINSICALLY SAFE ELECTRO-PNEUMATIC CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS HAZARDOUS AREA CLASSIFICATION

Additel 761 Automated Pressure Calibrators Selection Guide

User s Guide. Vacuum Controller for liquid delivery systems

Pneumatic high-pressure controller Model CPC7000

HAEMODYNAMIC MONITORING SETUP IN THE CATHETERISATION LABORATORY MAHAT SAPION

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

Additel 761 Automated Pressure Calibrators Selection Guide

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

Additel 761 Automated Pressure Calibrators

[Instruments for vacuum measurement, checking and adjustment] 3

WIKA INSTRUMENT CORPORATION

QBS Electronic Pressure Regulator

SHORT COMMUNICATION A SIMPLE, SENSITIVE AND VERSATILE SOLID-STATE PRESSURE TRANSDUCER

HUMOR 20. High Accuracy Humidity Calibrator. Water T, p1 HUMOR 20 V1.0

Digital Melting Point Apparatus

Additel 761 Automated Pressure Calibrators

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

High-speed pressure sensor Model CPT6140

Transmitter CS 21 Operation Manual

BAPI Pressure Line of Products - FAQs

Sensoric 4-20 ma Transmitter Board Operation Manual

INTRODUCTION TO NETWORK WIND 3 MOUNTING THE UNIT 14 SELECTING THE DISPLAY MODE 5 ABBREVIATIONS AND DEFINITIONS 17

HUMOR 20 HUMOR 20. High-precision Humidity Calibrator. Water T, p1

Mouse and Rat Tail Cuff Blood Pressure Systems from 1-24 Animals. Mouse rat tail cuff blood pressure system components

Transmitter CS 21 Operation Manual

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

Pneumatic High-Speed Pressure Controller Model CPC3000

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Flamco Flamco-Fill STA Flamco-Fill STM

6 Common Hydrodynamic Testing Mistakes

Technical Data Sheet MF010-O-LC

INSTALLATION & MAINTENANCE INSTRUCTIONS

PPC4E Pressure Controller/Calibrator Superior rangeability and reliability

BUBBLER CONTROL SYSTEM

Proportional Pressure-Relief Cartridge Valve, Size 2...4

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

Technical Information Conducal CLY421

Proportional Pressure-Relief Cartridge Valve, Size 2...4

Precision pressure sensor Basic version Model CPT6020

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Additel 761 Automated Pressure Calibrators

Inverse Prop. Pressure-Relief Cart., Size 2 4

Operating Instructions for Intrinsically Safe Pressure Transmitters Series DMG/******** for Hazardous Application in Coal Mining Industry

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI

USING THE CS400/CS405 (KPSI SERIES 169/173) SUBMERSIBLE PRESSURE TRANSDUCER WITH CAMPBELL SCIENTIFIC DATALOGGERS INSTRUCTION MANUAL

Mounting instructions. Strain transducer SLB-700A. B 26.SLB700A.10 e

RM-80 respiration monitor

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

BS12N Piezoresistive OEM Pressure Sensor

BUYER S GUIDE AQUAlogger 520

BS12 Piezoresistive OEM Pressure Sensor

2/ Port G1/4

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

PC10 Piezoresistive Silicon Pressure Sensor

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.

FLAMMABLE GAS SENSOR/TRANSMITTER

Pressure Transmitters 36 F-27839

ACQUIDATA FAMILY CALIBRATION PROCEDURE CALIBRATION CHECKING

Operating Instructions

SPD310/SPD360 Air Differential

Digital electro-pneumatic regulators Series ER100

陕西创威科技有限公司. Shaanxi Creat Wit Technology Co., Ltd. Precision pressure calibrator-cwy300 Operating Manual

Research Grade Blood Pressure Transducer

Heavy Duty Dissolved Oxygen Meter

Inverse Proportional Pressure-Relief Cart., Size 2 4

2/2-Way Proportional Valve (motor-driven)

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

Digital Pressure Gauge Main features: Data Sheet: BIT02B.417.R6.EN COVER on request

ADT 761 Automated Pressure Calibrator

Instruction Manual. PLT Series. 1. General Specification. 2. Installation. 3. Wiring. 4. Function. 5. Setting & Calibration

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

COSMO INSTRUMENTS CO., LTD.


Inductive Linear Displacement Transducers Model IW 260 Measuring strokes: 80 mm, 170 mm, 240 mm, 360 mm

Inverse Proportional Pressure-Relief Cart., Size 2 4

P900. P900, Datenblatt Seite 1 DESCRIPTION

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

User s Guide. Pressure Controller for liquid delivery systems

BUYER S GUIDE AQUAlogger 530WTD

RTX 1000H Series. Versatile Transmitters for a World of Pressure. GE Measurement & Control. Features

Model PDT Dewpoint Transmitter

HUMOR 20. High-precision Humidity Calibrator HUMOR 20. Operation

For Air Pressure. AP-30 Series. Digital Display & Digital Control. EASY-TO-SEE 2-color display. Two-color digital display pressure sensor

Transcription:

TSE Blood Pressure Monitor Invasive 209100series For Invasive Measurements in Laboratory Animals

2 Specifications subject to change without notice USA / Canada / Mexico Phone 1-989-698-3067 Fax 1-989-698-3068 Toll-Free (USA / CND): Phone 1-866-466-8873 Fax 1-866-467-8873

3 TSE Blood Pressure Monitor Invasive - 209100series General information The TSE Blood Pressure Monitor is an easy-touse compact system designed for invasive pressure measurements on all laboratory animals. Other features are: easily portable due to its lightweight construction, stackable housing for multi-channel use. TSE Universal Transducer Amplifier With its integrated power supply this singlechannel amplifier provides power to the connected transducer and performs amplification and filtering of the measured signal. The pressure values are continuously displayed on the LCD display in mmhg. The unit comprises an output connector for convenient use with a strip chart recorder. A chart recorder with an input of +10V is required. Alternatively the amplifier can be connected to a data acquisition system. The desired measuring range of the amplifier is chosen via the front panel range knob: Range 1 0...20 mmhg venous pressure Range 2 0...200 mmhg arterial pressure (pulmonary) Range 3 0...300 mmhg arterial pressure The ranges are set ex works and can not be altered by the user. A filter can be chosen according to the quality of the input signal. The filter knob on the front panel allows a filter setting of 1 Hz, 10 Hz or 100 Hz frequency. Front panel: Front Panel 1 ZERO = Zero adjustment knob 2 3 ½ digit LCD display (pressure in mmhg) 3 Range knob 1: 0-20 mmhg 2: 0-200 mmhg 3: 0-300 mmhg 4 Filter frequency knob (1 Hz, 10 Hz, 100 Hz) Rear panel: 1 Power cable 2 ON/OFF switch 3 AMP OUT = Signal output connector (e.g. to chart recorder) 4 AMP IN = Transducer input connector 5 CAL = Calibration adjusting screws Worldwide Phone +49-(0)6172-789-0 Fax +49-(0)6172-789-500 info@tse-systems.com www.tse-systems.com

4 pressure, central venous pressure, pulmonary artery pressure, perfusion pressure etc. It has excellent stability and high accuracy over the whole pressure range. Specifications Rear Panel Control Elements Signal Input Differential signal from TSE blood pressure transducer 9100-series Sensitivity FS Range1 0...20 mmhg (venous) Range2 0...200 mmhg (pulmonary, arterial) Range 3 0...300 mmhg (arterial) Amplifier Output 1...+8 V DC FS Output Impedance 100 Ω maximum Non-Linearity ± 0.02 % FS Accuracy ± 1 % FS Filter 1 Hz, 10 Hz, 100 Hz Display 3 1/2 digit LCD display Environ. Temp. 10-50 C Power 230 V/50 Hz Fuses 2 fuses (50 ma, slow-blow) The pressure is transmitted via a catheter or a cannula to the sensor. This pressure is then translated into an electrical signal, which is proportional to the pressure level. The output is fed into the amplifier via a special connecting cable. The transducer is powered by the amplifier. Due to a specific surface treatment the pressure transducer can be easily ventilated. Other features are: integral 3-way stopcock, integrated fast flush device, over-pressure safety valve, crystal clear for easy detection of air bubbles, reusable. TSE Pressure Transducer 9100-series The TSE Pressure Transducer 9100-series is a universal physiological pressure sensor for pressure measurements up to +300mmHg. Specifications Pressure Range -50 to +300 mmhg Sensitivity 50µV/V/cmHg Max. Over-Pressure 10.000 mmhg Volume Displacement 0.04 mm³/100 mmhg pressure Zero Point Drift ±0.3 mmhg/ C Sensitivity Drift ±0.1 %/ C Zero Offset ± 25 mmhg Linearity/Hysteresis ± 2% or 1mm Hg (whatever is larger) Electrical Leakage 5µA at 220V/50Hz Maximum Excitation >10V DC or AC Voltage Insulation withstands 10.000 V DC Output Impedance 350 Ω, nominal Weight approx. 15g w/o cable The reliable and robust transducer is suitable for a variety of measurements including arterial USA / Canada / Mexico Phone 1-989-698-3067 Fax 1-989-698-3068 Toll-Free (USA / CND): Phone 1-866-466-8873 Fax 1-866-467-8873

5 Preparing the transducer for a measurement 6 5 1 3 4 Amplifier Pressure Line 2 1 Pressure sensor 5 Port B 2 3-way-stopcock 6 Port C 3 Flush-device yellow arrow Cable connector 4 Port A red arrow Pressure line 1. Flushing the sensor chamber Use 3-way-stopcock (2) to make connection sensor chamber/port B (5-1). Remove cap from port B (5) and connect a syringe filled with flush solution. Remove cap from port A (4). Pull tab of flush-device (3). Using the syringe flush the sensor chamber. Remove all air bubbles. Close port A with a non-vented cap if included in the scope of supply or use another 3-waystopcock. Take care no air bubbles are trapped inside the stopcock or sensor chamber. Repeat procedure if air bubbles remain inside the sensor chamber! 2. Flushing the pressure line Mount the transducer to the pressure line via port C (6) Use 3-way-stopcock (2) to make connection pressure line/port B (6-5). Using the syringe on port B flush the pressure line. Remove all air bubbles. Use 3-way-stopcock (2) to make connection pressure line/sensor chamber (6-1). Remove syringe. Close port B with a non-vented cap if included in the scope of supply or use another 3-waystopcock. 3. Connect pressure line to catheter/cannula It is recommended that the sensor is positioned at the level of the measuring point during measurement! Note: Stopcocks must be positioned 90 for OFFposition. Do not position stopcocks at 45 to achieve OFF-position, since this position is unprecise and may result in bleed back or air embolism! Measurement Any chart recorder with an input range from 0 to +10 V is suited for connection to the amplifier unit. Choose the chart speed according to your waveform requirements. Connect the recorder to the amplifier s rear BNC socket AMP OUT. In order to read off absolute pressure values the recorder has to be calibrated with zero pressure and a second pressure value, which should correspond to the maximum pressure level to be expected, e.g. 200mmHg! Remove the blue cap from the transducer connector. Worldwide Phone +49-(0)6172-789-0 Fax +49-(0)6172-789-500 info@tse-systems.com www.tse-systems.com

6 Carefully connect the transducer connector (white) to the amplifier cable (grey connector). Align the connector arrows and push the connectors together. Connect the amplifier cable to the rear panel 5-pole input socket AMP IN. Switch the amplifier on. The amplifier is switched on with the POWER mains switch on the rear panel. The LCD display lights up immediately. Choose the desired filter frequency 1 Hz, 10 Hz or 100 Hz with the filter knob on the front panel. Maximum signal smoothing, i.e. maximum suppression of artifacts, is achieved by using the 1 Hz filter. For blood or perfusion pressure measurements usually a setting of 10 or 100 Hz is recommended. The pressure can be read off the display in mmhg. Important: Allow approx. 5 minutes for the amplifier to warm-up in order to guarantee correct measurements! Calibrating the amplifier Each transducer is calibrated at the factory for a sensitivity of 50 µv/v/cmhg and can not be recalibrated by the user. The amplifier is also delivered calibrated. In case a recalibration is necessary please proceed as follows: Setting the amplifier: Depending on the measurement you want to perform choose the suitable measuring range with the front panel range knob. Range 1 0...20 mmhg Range 2 0...200 mmhg Range 3 0...300 mmhg venous pressure arterial pressure (pulmonary) arterial pressure Connect the pressure transducer to the amplifier (AMP IN). Connect a pressure-gauge to the transducer. Switch the amplifier ON. Choose the desired measuring range 1, 2 or 3. Zero balance the system to atmospheric pressure by removing the pressure line from the animal (or by opening port B with 3-waystopcock). Set the display on the amplifier to zero with knob ZERO. Note: If you change the measuring range the zeroing has to be repeated. Switch the recorder ON. Set the pen to the desired baseline position. Connect pressure line to the animal (or close port B with 3-way-stopcock). With the transducer vented to atmospheric pressure set the display to zero with the ZERO knob. Using the pressure-gauge apply a defined pressure to the transducer. The pressure is dependant on the range to be calibrated (e.g. range 3: 300mmHg). USA / Canada / Mexico Phone 1-989-698-3067 Fax 1-989-698-3068 Toll-Free (USA / CND): Phone 1-866-466-8873 Fax 1-866-467-8873

7 If the display deviates from the pressure applied, use the corresponding adjusting screw on the rear panel to set it to the correct value (CAL). CAL1, 2, 3 are used for range 1, 2 and 3 respectively. The transducer has been designed for multiple use. It is delivered non-sterile. Not for clinical use! The procedure described above can also be used to calibrate the chart recorder (omit step 7 then). Worldwide Phone +49-(0)6172-789-0 Fax +49-(0)6172-789-500 info@tse-systems.com www.tse-systems.com

12 Contact Maze us Sophisticated Life Science Research Instrumentation Copyright Phone (USA/Canada/Mexico): 2008 TSE Systems International Group All rights reserved TSE_Blood Pressure Monitor Invasive_20080724 1-989-698-3067 Toll-Free Phone (USA/Canada): 1-866-466-8873