MASS-STREAM TM Instruction Manual. D-6200 Analog Mass Flow Meters / Controllers

Similar documents
Operating Instructions for Direct Flow Mass Flow Meter for gases. Model: DMW

Analogue and Digital Mass Flow Meters and Controllers for Gases MASS-STREAM

Digital Mass Flow Meters and Controllers for Gases

Instruction manual. CEM controlled evaporator and mixer. Doc. no.: H Date:

Instruction manual. MASS-VIEW series. Doc. nr.: F Date:

Quick Installation Guide Doc nr.: C Date:

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

Model 601CV. Mass Flow Controller. Product Features:

Mass Flow Meter (MFM) for gases

Brooks TMF-Zone 1 Models 5816-Ex/5864-Ex

Certified according to DIN EN ISO 9001 Technical Datasheet TRICOR Series Mass Flow Meters

DTG - LCD Digital Temperature Gauge USER MANUAL

Design Features. General Description

Operating Instructions DB41

Dual Solenoid Gas Valve Installation

C-Flow Coriolis Mass Flow Meters

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

Model PDT Dewpoint Transmitter

Instruction Manual. Doc. no.: A Date:

DPC-30 DPC-100. Reference Manual

GILMONT ACCUCAL FLOWMETERS

KTE6000 / KTU6000 Series OEM pressure transmitters for industrial media

Mass Flow Controller (MFC) for Gases

QBS Electronic Pressure Regulator

Mass Flow Controller (MFC) for Gases

SITRANS P measuring instruments for pressure

PM105 PRESSURE METER USER S MANUAL

Instruction Manual Differential Pressure Transmitter

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Mass Flow Meter (MFM) for Gases

Operating Instructions Temperature Measuring System MF420-5T-100

Instruction and Maintenance Manual

BGA158 Series CE Approved Class A Shutoff Gas Valve

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

GP1 & GP2. Electropneumatic Regulators FOR PRESSURE CONTROL TO 1,000 PSI

Pressure Automated Calibration Equipment

Instruction Manual Contact Pressure Vacuum Gauge

Maintenance Manual. Liquid Level Transmitter CT801-SI, CT801-TA

GAS FUEL VALVE FORM AGV5 OM 8-03

Instruction Manual Differential Pressure Transmitter

QPL... Pressure Switches. Siemens Building Technologies HVAC Products

OPTIBAR P 1010 C Technical Datasheet

Measurement accessories METPOINT OCV for the measurement in systems up to 40 bar

T EK-SUB 4800C 19 mm Submersible Level Transmitter

BGA158 Series CE Approved Class B Shutoff Gas Valve

DK48 - DK800. Quick Start. Variable area flowmeter

Instruction Manual BA90 Precision Barometer

* * Data Sheet and Operating Manual. Digital manometer ME01 ## # 87 # HL S#### Dust explosion protection Zone 22.

Target disk flowmeter Series DP. Instructions manual DP65 DP500. R-MI-DP Rev.: 4 English version

FTC130 Transmitter. Operating Manual

Instruction and Maintenance Manual

Model FS5001B VA.1. MEMS Mass Flow Sensors ISO 9001 ISO OHSAS 18001

BGA158, BGA171 and BGA110 Series Shutoff Gas Valve

Orifice Plate Flow Meter

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

Electro-Pneumatic Converter YT-940 SERIES

ATM/GR. Electronic pressure switches CUSTOMER BENEFITS

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

CORI-FLOW PRECISION MASS FLOW METER / CONTROLLER FOR LIQUIDS AND GASES

INSTALLATION & MAINTENANCE INSTRUCTIONS

EPR High Precision 3000 psi Electronic Pressure Controller Accurate to 0.25% of full scale

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

PLAQUE MFM0680 UNDERWATER LIGHT. Operating Instructions

VOLUME BOOSTER RELAYS YT-300 / 305 SERIES

P900. P900, Datenblatt Seite 1 DESCRIPTION

Features. Description

Constant Pressure Inlet (CCN) Operator Manual

Types and approvals. Page 1/6. Data Sheet

Operating Manual * * Pressure Transmitter. Table of Content. 1 Safety Guidelines. 1.3 Risks due to Non-Observance of Safety Instructions

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

Gas density monitor With integrated transmitter Model GDM-100-TI

ACV-10 Automatic Control Valve

ASDX Series Silicon Pressure Sensors

Air Sensor. FCS Ex. Manual. AQ Elteknik AB

PORTER. USER S MANUAL for Models LM, HM, LC and HC. FM-653 Rev. B 8/02

PSG 901. Pressure Switch for Gas. elektrogas.com EE141-07/12

VS18/26 with PROFINET and EtherNet/IP Interface. ATEX Installation Instructions

Brooks Flow Controllers for Gas and Liquid Service Model FC 8744, Series FC 8800 and 8900

453 Series Steam Heated Vaporizing Regulator

Solenoid gas valve. The solenoid gas valve is also applicable as a main gas valve Pilot gas valve for forced draft burners

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

INSTALLATION MANUAL Matheson Tri-Gas Cabinet Enclosures

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

X-Flow Mass Flow Controller 授权代理 : 北京品超思瑞科技学院公司

Mass Flow Controller (MFC) for Gases

DeltaSpan Pressure Transmitters LD10 Series Owner s Manual

EC214 EC215 - EC215R Bench Conductivity Meters

GEMS SENSORS & CONTROLS

Safety. Operating instructions Solenoid valve for gas VG 6 VG 15/10 DANGER. Contents WARNING CAUTION. Changes to edition 09.14

MANUAL KPS Pressure Control Valve

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Product information. Capacitive. Level detection in liquid VEGACAP 62 VEGACAP 63 VEGACAP 64 VEGACAP 66 VEGACAP 69. Document ID: 29983

780S Ultra High Purity Series

Pneumatic pressure switch 20DD

IN-FLOW CTA Industrial Mass Flow Meters and Controllers for Gases

Best Practice Guide, Servomex 2700

INSTRUCTIONS FOR MODEL SSC LOW DELIVERY PRESSURE GAS REGULATORS

OXY 3690 MP. Measuring transducer for air oxygen

Transcription:

MASS-STREAM TM Instruction Manual D-6200 Analog Mass Flow Meters / Controllers Doc. No.: 9.17.112A Date: 17-03-2017 ATTENTION Before installing and operating the instrument it is strongly recommended that this instruction manual should be read carefully. Not following the guidelines could result in personal injury and/or damage to the equipment.

SCOPE OF THIS INSTRUCTION MANUAL This instruction manual of M+W Instruments GmbH covers the installation of the analog Mass-Stream TM instrument model series D-6200 for mass flow measurement and control of gasses and gas mixes The information in this manual has been reviewed and is believed to be wholly reliable. No responsibility, however, is assumed for inaccuracies. The material in this manual is for information purposes only, and is subject to change without notice. Warranty The M+W Instruments GmbH grants a warranty on its products according to the General Terms and Conditions. The warranty is limited to the mass flow meters and controllers and its components. The correct use of the device under the observation of all specified operation conditions is the precondition for the warranty. ATTENTION! The warranty is solely limited to the mass flow meters and controllers and its components. No warranty claims are accepted in case of a faulty operation of the instrument, like a wrong electrical hook-up for example. Any liability is explicitly excluded for any secondary damage, caused by the instrument s failure or the malfunction of any component. All wear parts, like seals for example, are also excluded from the warranty. Page 2 off 12

The products of M+W Instruments GmbH are warranted against defects in material and workmanship for a period of three years from the date of shipment, provided they are used in accordance with the ordering specifications and they are not subjected to abuse or physical damage. The products of other manufacturers are warranted for a period of one year. Products that do not operate properly during this period may be repaired or replaced at no charge. Repairs are normally warranted for one year or the balance of the original warranty, whichever is the longer. See also paragraph A-V and B-II of the Terms and Conditions of sales of the M+W Instruments GmbH. The warranty includes all initial and latent defects, random failures, and indeterminable internal causes. It excludes failures and damage caused by the customer, such as contamination, improper electrical hook-up, physical shock etc. Re-conditioning of products primarily returned for warranty service that is partly or wholly judged non-warranty may be charged for. M+W Instruments GmbH prepays outgoing freight charges when any party of the service is performed under warranty, unless otherwise agreed upon beforehand. However, if the product has been returned collect to M+W Instruments GmbH, these costs are added to the repair invoice. Import and/or export charges, foreign shipping methods/carriers are paid for by the customer. Page 3 off 12

Table of Contents 1 ANALOG INSTRUMENTS... 5 1.1 General... 5 1.2 Installation... 7 1.3 Operation and Maintenance... 8 1.4 Specifications of analog MASS-STREAM TM Instruments D-6200.... 9 1.5 Hook-up with 6DIN Connector, 6-pin round... 10 1.6 Hook-up with 15-pin sub-d connector... 10 2 CTA MEASUREMENT PRINCIPLE...11 The dimensional drawings and hook-up diagram of the Mass-Stream TM instruments D-6200 are available for download on our website www.mw-instruments.com. Page 4 off 12

1 ANALOG INSTRUMENTS All necessary settings for this instrument are already performed at M+W Instruments GmbH. Following the next steps carefully is the quickest way to get the instruments operational in your own system. 1.1 General The analog instruments of M+W Instruments GmbH are measurement devices for thermal mass flow and control of gasses and gas mixes. The flow signal is measured directly in the gas flow from 0.5...10 l n */min up to 300...6,000 l n */min and controlled when a control valve is applied. [* n = normal conditions: At normal conditions the volume is regarded at a temperature of 0 C and a pressure of 1013.25 mbar (a)] As an option ex work the device can be equipped with an integrated 3½ digit LCD flow rate indication or with an integrated 8 digit counter. Output signal / setpoint: 0 5V dc OR 4 20 ma Power Supply: +15 Vdc OR +24 Vdc (please check on instrument label) ATTENTION! The input and output signals are NOT galvanically isolated! Scope of Delivery Immediately upon receipt of the incoming goods the package needs to be checked for any possible damages and the goods have to comply with the scope of delivery listed on the packing note. In case of any damages the supplier and the freight forwarder have to be contacted without delay. Page 5 off 12

In the event of any discrepancies please get in touch with your local distributor (see www.bronkhorst.com) or contact us directly: M+W Instruments GmbH Dorfstr. 1 D-85391 Leonhardsbuch Tel.: +49 8166 9921 0 Fax: +49 8166 9921 22 Email: sales@mw-instruments.com When the instrument needs to be returned please include a description of the damage in a closed envelope. ATTENTION! All instruments for return must have a fully filled-in Declaration of Decontamination. This declaration has to be attached to the shipping documents, allowing free access on the outside of the package. The declaration of decontamination is available on our accompanying CD with each instrument shipment or it can be downloaded from our website www.bronkhorst.com. Please mention the instrument s model code and the serial number on your order, as well as your VAT number when needed. Page 6 off 12

1.2 Installation To avoid personal injury and/or damage to the equipment only trained and qualified personnel shall perform the installation of the instruments: Read the instrument s name plate before installation and check the electrical connection, flow range, media to be measured, inlet and outlet pressure as well as input and output signals. The instruments contain electronic components which are sensitive to electronic discharges (ESD). Contact with electronically charged persons or objects could possibly endanger these components or even result in their failure. While installing the instrument take care of the arrow marking on the instrument s flow body, determining the flow direction. Ensure that all piping is absolutely clean and without any damage. It is recommended to install an inlet filter in front of the instrument. Avoid piping with small diameter at high flow rates. When possible no abrupt angles or other disturbances shall be mounted directly in front of the instrument. All devices have Withworth pipe threads type RP according to ISO228-1 (cylindrical G-threads in inches according to BSPP). The mounting of NPT-fittings will destroy the inlet threads in the instrument s body. Using an inlet pipe with sufficient piping diameter is recommended. The length of the inlet pipe should be at least 10 times the piping diameter. Before the first start-up check that the system has absolutely no leakage. If toxic, explosive and/or corrosive gases are to be used, ensure to purge the system long enough with a dry, inert gas. This is also required to remove these kinds of gases from the system. The instruments preferred mounting position is horizontally. Especially when installing bigger mass flow controllers (D-6273 and bigger) in different ways you should get in contact with your distributor or M+W Instruments GmbH beforehand. Page 7 off 12

All instruments described in this manual carry the CE-mark. Therefore they have to comply with the EMC requirements valid for these instruments. However compliance with the EMC requirements is not possible without the use of proper cables and connector/gland assemblies. In case the instrument is connected to other devices (e.g. power supply) ensure that the operating voltage complies with the supply voltage given on the name plate and that the function of the shielding is not affected. M+W Instruments GmbH recommends using standard cables following the CE-requirements and could supply them when needed. 1.3 Operation and Maintenance When the power is switched on a warm-up period of at least 30 sec is required allowing the instrument s thermal stabilisation. Afterwards the instrument will work with an accuracy of approximately 4% FS. It takes around 30 minutes until the optimal stability is reached and the instrument will work with an accuracy of < 3% FS. This is supposed to be performed with or without gas flow. To avoid creating gradients it is recommended that gas and ambient temperature are kept at the same level (max ΔT approx. ± 2 3 C). Otherwise the accuracy of this thermal measurement could be affected in an unpredictable manner. Be sure that the specified pressures have been applied. Pressure surges, as may occur during the system pressurisation, must be avoided. The control valve cannot be used as a shut-off valve. When starting up, always bring the instrument gradually up to the level of operating conditions. There are no regular maintenance checks required for the instrument s operation. It has to be ensured that a possible mounted inline filter is checked and cleaned on a regular basis. We recommend having the calibration checked every 24 months by the supplier or an authorized service partner. Page 8 off 12

1.4 Specifications of analog MASS-STREAM TM Instruments D-6200 Accuracy (based on calibration with Air) Repeatability Pressure sensitivity Temperature sensitivity Response time sensor (τ 63%) Leak integrity (outboard) Measurement System ± 3% FS including non-linearity (± 2% FS on request) < ± 0.5% FS ± 0.3% RD / bar typical (Air) ± 0.3% RD / C (Air) appr. 1.2 sec < 2 10-7 mbar l/sec He Mechanical Parts Sensor Stainless steel SS316 (AISI 316L, 1.4404) Aluminium EN AW-6082-T6 (AL 50ST/51ST) Instrument body anodised OR Stainless steel SS316 (AISI 316L, 1.4404) Sieves Stainless steel Distance parts Teflon Protection IP40 Operating Limits Measuring range 5 100% (1:20) Temperature 0 50 C 0 10 bar(g) for aluminium body 0 20 bar(g) for stainless steel body Pressure The maximum operating pressure of controllers also depends on the selected valve! 30 min for optimum accuracy Warm-up time within 30 sec for accuracy ± 4% FS Supply voltage Current peak values Output signal Instrument connector Electrical Properties +15 Vdc OR 24 Vdc ± 10% with BJ-valve: +24 Vdc Inrush current: max. 250 ma Meter: approx. 75 ma at 0% flow approx. 175 ma at 100% flow Controller: add 250 ma 0 5 Vdc active OR 4 20 ma active 6-pin round DIN (male) optional for controllers: 15-pin sub-d (male) Page 9 off 12

1.5 Hook-up with 6DIN Connector, 6-pin round See hook-up document no. 9.16.089 of the shipment s accompanying documentation. Mass flow meter D-62 0: 1 + power 4 + output signal 2 0 power 5 n/c 3 0 output signal 6 n/c Mass flow controller D-62 1 / D-62 3: 1 + power 4 + output signal 2 0 power 5 + setpoint signal 3 0 output signal 6 n/c 1.6 Hook-up with 15-pin sub-d Connector Typ M for mass flow controller D-62 1 / D-62 3: 1 n/c 9 n/c 2 + output signal 10 n/c 3 n/c 11 n/c 4 n/c 12 0 output signal 5 0 power 13 n/c 6 n/c 14 n/c 7 + power 15 shield 8 + setpoint signal Typ B for mass flow controller D-62 1 / D-62 3: 1 0 power 9 0 power 2 + output signal 10 0 power 3 0 power 11 n/c 4 n/c 12 n/c 5 + power 13 n/c 6 n/c 14 shield 7 n/c 15 n/c 8 + setpoint signal Please pay attention to the label with the hook-up fitted at the cable. Be sure that the integrity of the shielding is not affected when connecting the system to other devices (e.g. to PLC). Only shielded wire terminals can be used. Page 10 off 12

2 CTA MEASUREMENT PRINCIPLE The direct through-flow measurement with CTA (Constant Temperature Anemometry) is also named inline measurement. The working principle is based on King s law of the ratio between the heater energy and the heat degradation by means of gas flow or liquid flows published in 1914. Two elements, a heater and a temperature probe, are positioned in the gas flow. CTA is aiming to keep the difference in temperature T between both sensor pins at a constant level. The flow rate and the heater energy required to maintain this constant T are proportional and thus indicate the mass flow of the gas. The relation between the heater energy and the mass flow is described by the following formula: n P = P 0 + C Φ m P Heater power P 0 Heater power offset at zero flow C Constant (device-dependent) Φ m Mass flow n dimensionless figure (type 0,5) Page 11 off 12

The calibration is processed with a mass flow controller which functions as a reference. The controller and the gas inlet pressure forward a defined gas flow, thus a signal of the sensor can be allocated to each specific mass flow. Each instrument is calibrated according to the customer s operation conditions within the specified accuracy limits. These customer specific information can be found on the instrument label and the calibration certificate. Page 12 off 12