COMELEC C-30-S Parylene Coating System

Similar documents
Usage Policies Notebook for Parylene Coating System

March CS-1701F Reactive Ion Etcher

Photolithography. Operating Instructions

Standard Operating Manual

THERMAL EVAPORATION UNIT (for Al evaporation)

Standard Operating Manual

STS ICP-RIE. Scott Munro (2-4826,

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

Revised: June 7, 2017

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Standard Operating Manual

The SPI Sputter Coater Handbook

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

Nordiko Metal Sputtering System Standard Operating Procedure

University of MN, Minnesota Nano Center Standard Operating Procedure

STS PECVD Instructions

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

Xactix Xenon Difluoride Etcher

Plasma II (AXIC) Standard Operating Procedure. Revision: 1.0 Last Updated: Feb.6/2013, Revised by Grace Li

CHEMICAL ENGINEERING SENIOR LABORATORY CHEG Initiated Chemical Vapor Deposition

EE 432 Lab 3 PMOS source/drain lithography and diffusion

VACUUM CHAMBER PM OBJECTIVE: Vacuum Vacuum STS-RIE / ETCH CAVITY. Chamber. Vacuum CHAMBER WALL. 5 to 6 days. Procedure: particle issues.

Xactix XeF2 OPERATION MANUAL

Oerlikon Sputtering Evaporator SOP

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Plasma Cleaner. Yamato Scientific America. Contents. Innovating Science for Over 125 Years. Gas Plasma Dry Cleaner PDC200/210/510 PDC610G.

QUORUM TECH 150RES THE FIRST AND THIRD WEEK WILL BE SET UP FOR CARBON COATING THE SECOND AND LAST WEEK WILL BE SET UP FOR GOLD COATING

Standard Operating Procedure. For. PVD E-Beam

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Standard Operating Manual

COBILT CA-800 Mask Aligner Equipment Operation

MANUAL FOR SPTS APS (DIELECTRICS ETCHER)

Savannah S100 ALD at SCIF, UC Merced Standard operating Procedure

KOH Etchant Standard Operating Procedure

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

NORDSON MARCH PX-1000 PLASMA ASHER STANDARD OPERATING PROCEDURE Version: 1.0 July 2016

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Warnings: Notes: Revised: January 8,

Unaxis PECVD. SiH4 (5% in He)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

[needs to be discussed with faculty if any new material to be done]

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Standard Operating Manual

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility

Arizona State University NanoFab PLASMATHERM 790 RIE. Version A

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

MMRC PVD Evaporator Instructions

University of MN, Minnesota Nano Center Standard Operating Procedure

Standard Operating Manual

Vacuum Systems and Cryogenics for Integrated Circuit Fabrication Technology 01

Cambridge NanoTech: Savannah S100. Table of Contents

NRF Suss Delta 80 Spinner SOP Revision /14/2016 Page 1 of 11. Suss Delta 80 Spinner SOP

Operating Procedures for Metal Evaporator I

VACUUM CHAMBER PM TECHNIQUE LAM 9400 POLY ETCH WET SCRUB

The MRL Furnaces USED FOR THIS MANUAL COVERS

OXFORD PLASMALAB 80PLUS (CLOEY)

Cryo-Evaporator Operation

EXPERIMENT XI. Careful!! Improper handling of the vacuum line may result in the release of SO 2 which is an irritating and suffocating gas.

Arizona State University Center for Solid State Electronic Research. Table of Contents. Issue: C Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

Angstrom Dielectric Sputterer Operation Manual

NNCI ETCH WORKSHOP SI DRIE IN PLASMATHERM DEEP SILICON ETCHER. Usha Raghuram Stanford Nanofabrication Facility Stanford, CA May 25, 2016

Approved by Principal Investigator Date: Approved by Super User: Date:

OPERATION OF THE DIMPLER

Introduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

Fig. 1: Mechanical vacuum gauge (left) and capacitive vacuum gauge (right)

REPRODUCIBILITY IN OPTICAL THIN FILM PROCESSING PART 1, THE VACUUM AND PUMPING

Basic & Controlled Atmosphere Glove Boxes

Operating Instructions

Oxygen Contamination

STS DRIE User SOP Location: NRF Cleanroon, Dry Etch Bay

Eco Snow CO2 Clean System Operation

Plasma Asher: March PX-500 User guide (May-30, 2017)

Nanofabrication Facility: Sputter Evaporator SOP Rev. 04, June 2007

Warnings: Notes: Revised: October 5, 2015

BS ml Bottle Sampling Unit

Plasma 1 (Technics) Standard Operating Procedure. Revision: 1.0 Last Updated: Feb.5/2013, Revised by Grace Li

Standard Operating Manual

University of Minnesota Nano Fabrication Center

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

VACUUM & AIRLESS. Select from Kimble vacuum traps, manifolds, Schlenk tubes, adapters, and bubblers for your vacuum and airless glassware needs.

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Figure 1: Hydrostatic Pressure Forces Are Perpendicular to the Surface

Dilution Refrigerator manual

DISASSEMBLY AND ASSEMBLY INSTRUCTIONS FOR LIQUID RING COMPRESSORS

VACUUM CHAMBER PM TECHNIQUE LAM 4520 Oxide ETCH Chamber

Inert Atmosphere Guide

SCS Field Revision No. 1.0 June 2001 STANDARD OPERATING PROCEDURE (SOP) FOR FIELD OPERATION OF THE SEQUENTIAL CYCLONE SAMPLER (SCS)

Leaks, contamination and closed loop control how RGAs make coating processes more profitable

Standard Operating Procedure Inert Vacuum Line. Ryan Mulvenna, May 2012

ESAIL D45.1 Final Report of Remote Unit Gas Thruster

University of Minnesota Nano Center Standard Operating Procedure

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

Manufacturing Processes for WSi 2 -GPSOI Substrates and their Influence on Cross-talk Suppression and Inductance

THERMODYNAMICS OF A GAS PHASE REACTION: DISSOCIATION OF N 2 O 4

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

Transcription:

COMELEC C-30-S Parylene Coating System Comelec C-30-S Parylene deposition system Introduction Parylene is a polymere deposited at room temperature in a vacuum chamber (few µb). Parylene coating is perfectly conformal. Parylene is a pinhole free layer resistant to HF, KOH and solvents. Its thickness can vary from few tens of nm to tens of μm with a deposition rate of few μm/hour. Parylene is easy to dry etch (istropic or anisotropic). Parylene can be use as a: Protective or support layer for sensitive device, Sacrificial layer with Al or Si device, Device layer with PR, SiO2 or Si as a sacrificial layer. Basic of deposition Parylene raw material is a solid dimmer. The system works under a primary vacuum (few μb). The dimmer is heated at about 150 C and changes to a vapor phase. It diffuses to Pyrolysis chamber at 670 C and is dissociated in a monomer. The monomer diffuses to the room

temperature or slightly heated (80 C) process chamber, and condensates to fabricate Parylene layer. Before being pump, residual monomers are trapped in a cryogenic device. The layer thickness depends on the Parylene solid load placed in the vaporization boat. Total surface and arrangement in the processing chamber influence the thickness and uniformity. At 100mm Silicon wafer scale layer uniformity is less than 1%. Parylene coating principle (Courtesy of www.comelec.ch). Parylene front panel, 10 slots holder and processing chamber Parylene Load Estimation Parylene load (g) 0.20 7.00 11.00 Avg. thickness 62nm 2.09µm 3.15µm Table: Processes performed with 3 Si wafers (100mm) loaded in lower, medium and higher slots.

How to proceed to use the system 1. Idle mode - Chamber is pumped: vacuum chamber and pump pressure should be few µbar, - LN2 trap is clean, empty and under vacuum, - Substrate holder is empty and under vacuum, - Parylene container is empty and under vacuum, - Zone 1 (vaporization) and zone 2 (pyrolysis) are off, 2. Make a coating - Before starting: o Switch on zone 2. It takes 1 hour to get it ready. You can switch it on automatically with a timer. Please fill process sheet: Time, pressures o Switch on the balance, - Chamber venting: o Close the valve chamber: vanne chambre light ON becomes OFF. It takes 30s to get it closed, o Vent the chamber: vanne air light OFF becomes ON. It takes approximately 1min to vent, - Wafers/samples loading: o Put on a pair of gloves, o Open the chamber lid, o Switch of venting, press vanne air, o Get the holder out and put it on the table, o Place wafers on the holder shelves (10slots), o Put back the holder in the chamber. Holder base must be properly positioned to rotate it later, o Press rotate and check if it goes well, o Close the lid, - Parylene load (keep the gloves): o Open the load door to get the Parylene container out, o Fit a new aluminium protection, o Move everything on Mettler-Toledo balance, adjust the zero (press tare), and fill with the appropriate Parylene quantity (see our table), o Move the filled container back to the loading room, o Close the door, - Pump the chamber: o 3 checks before pumping: LN2 trap placed, Chamber lid closed, And Parylene loading door closed, o Press vanne chamber to pump the chamber, Please fill process sheet: Time o Press F4 to put the lot number: quantity of Parylene g-run number, for example 7g-004 (7g Parylene load and run number 4), press F11, o Once vacuum is below 20µbar in the chamber, it takes between 5 to 15min (function of the state of the machine), it is time to fill the LN2 trap, Please fill process sheet: Time, pressures

- LN2 trap filling: o Put on safety glasses and LN2 gloves, o Fill the trap using the LN2 dewaer by applying some pressure on top (open the LN2 line + apply pressure with the manometer), o Once LN2 start spilling outside the trap, drop the flow by screwing the valve until the trap is full (10cm below the top). This operation takes about 5min. o If necessary, complete the filling with the thermal can near the system, o Put the LN2 stopper - height measurement plug, o Check on the C-30-S display if height is fine (90%), - Start deposition: o Once the chamber pressure is 5-6µbar (pump pressure always lower) press zone 1 to start Parylene dimmer vaporisation, Please fill process sheet: Time, pressures o Press F1 to monitor pressure chamber, - End of deposition: o Once pressure is back to 5-6µbar deposition is finished. There is no Parylene dimmer left in the system. Please fill process sheet: Time, pressures o Switch off zone 1 and WAIT 10MIN BEFORE VENTING TO CLEAR EVERYTHING in the system, Please fill process sheet: Time, pressures - Chamber venting: o Close the valve chamber: vanne chambre light ON becomes OFF. It takes 30s to get it closed. o Vent the chamber: vanne air light OFF becomes ON. It takes approximately 1min to vent. - Wafers/samples unloading: o Put on a pair of gloves, o Open the chamber lid, o Switch of venting, press vanne air, switch off rotation, o Get the holder out and put it on the table, o Take out wafers from, - LN2 trap cleaning: o Put on safety glasses and LN2 gloves, o Make ready an open thermal bottle, o Pull out the trap and empty the LN2 in the thermal bottle, o Take off glasses and LN2 gloves, o Go to the washbasin and scrub the outside of the trap, o Dry it and put back after O ring checking, - Chamber lid and O ring cleaning: o Below the lid Parylene film is easy to delaminated, o Remove the O ring and clear parylene on the top surface. You can use a blade to cut it. o Clean sealing surfaces with some IPA, o Wipe the O ring, put some vacuum grease and fit it in the groove, - Chamber pumping: o Put back the holder in the chamber. o Close the lid. o 3 checks before pumping: LN2 trap placed, Chamber lid closed,

And Parylene loading door closed, o Press vanne chamber to pump the chamber, Please fill process sheet: Time 3. Leave the system in Idle mode Photos Gallery o If no more deposition are planed, switch off zone 2 o System in Idle mode: Chamber is pumped: vacuum chamber and pump pressure should be few µbar, LN2 trap is clean, empty and under vacuum, Substrate holder is empty and under vacuum, Parylene container is empty and under vacuum, Zone 1 (vaporization) and zone 2 (pyrolysis) are off Fig. 1: Si wafer coated with 11.5um Parylene (0.5% uniformity). Fig. 2: Parylene conformal coating in deep Si cavities. Parylene is delaminated after sample cleavage. Fig. 3: Parylene isotropic etching to release asi structures. Fig. 4: Anisotropic etching of Parylene with asi mask.