Clinical Implementation of Volumetric Modulated Arc Therapy

Similar documents
CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

SnapShot IMRT with compensators and FFF beams

Planning Assignment (3 field rectum)

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

IMRT in clinical practice at the UMC-Utrecht. Utrecht

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Verification for Small Fields

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Constancy checks of well-type ionization chambers with external-beam radiation units

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Math Review. Overview

Zoubir Ouhib Lynn Cancer Institute

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Dosimetric Calculations. Lonny Trestrail

Relative Dosimetry. Photons

Scientific Journal of Nuclear Medicine & Radiation Therapy

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

A beam-matching concept for medical linear accelerators

THE development of more advanced techniques in radiotherapy,

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator

Review of fundamental photon dosimetry quantities

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

Introduction. Materials and Methods

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Commissioning and periodic tests of the Esteya electronic brachytherapy system

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

Dynamic wedges dosimetry and quality control

kvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more The Best of Both Worlds

Abstract: Introduction:

Disclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

1. Question Answer cgy / MU cgy / MU 2. Question Answer

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Iranian Journal of Medical Physics

Surface buildup dose dependence on photon field delivery technique for IMRT

Proton Therapy QA Tools

Radiotherapy Treatment Planning and Delivery

Quality control for Intensity-modulated radiation therapy

3D Treatment Planning and verification with hand calculations

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

TG-51: Experience from 150 institutions, common errors, and helpful hints

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Commissioning of a new total body irradiation protocol

Evaluation of six TPS algorithms in computing entrance and exit doses

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

High Speed Direct SAD Radiosurgery Beam Scanner

Radiotherapy physics & Equipments

Statistical determinations of the gating windows in the. respiratory gated radiotherapy using the visible

The dose distribution of medium energy electron boosts to the laryngectomy stoma

Total Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Low Dose Rate (LDR) LDR = 10 mrads(si)/s

How to Design Medical Accelerator Systems for Reliability: IBA PT System

Development of Graphical User Interfaces (GUI) software and database for radiation therapy applications

TEST REPORT BEAMSCAN WITH HALCYON

ADDENDUM IMPORTANT DOCUMENT INVITATION TO BID ADDENDUM. OPENING DATE & TIME: January 3, 2018 at 3:00PM

Varian HDR surface applicators commissioning and clinical implementation

Automatic treatment planning an MCO perspective

USER MANUAL PROGRAM VERSION 1.0.1

A gated deep inspiration breath-hold radiation therapy technique using a linear position transducer

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

Total Ionization Dose (TID) Test Results of the RH1028MW Ultralow Noise Precision High Speed Operational Low Dose Rate (LDR)

Total Ionization Dose (TID) Test Results of the RH1021BMH-10 Precision 10V Low Dose Rate (LDR) LDR = 10 mrads(si)/s

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Exposure System Selection

Physics and Quality Assurance

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

THE FUNDAMENTALS OF THE AIR SAMPLER CALIBRATION-VERIFICATION PROCESS

Breast Treatment System REFERENCE MANUAL

Hand-held pressure calibrator with integrated pump Model CPH6600

Risk Assessment Guidance for QM in Brachytherapy. Conflicts. Learning Objectives. Example Procedure

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD

Quality Management Determined from Risk Assessment

Transcription:

Clinical Implementation of Volumetric Modulated Arc Therapy UT M.D. Anderson Cancer Center Ramaswamy Sadagopan, Rebecca M. Howell, Weiliang Du and Peter Balter

Definition 2 Intensity Modulated Arc therapy where the following three parameters are modulated simultaneously Gantry rotation Dose rate Leaf speed.

Overview Advantages of Modulated Arc Therapy 3 Configuring the Accelerator / R & V system Treatment Planning System commissioning Patient specific QA (film and ion-chamber) RPC phantom and partial treatments verifications Routine QA (output @ dose rates, DMLC QA, linearity check at several dose rates, constancy check)

Advantages of modulated arc treatment -Varian Demo Plan 4 IMRT Treatment plan RapidArc Plan 1147MU & 5.5 min 3D dose max = 108.8% 804 MU & 1.5 min 3D dose max = 114.7% http://www.varian.com/us/oncology/treatments/treatment_techniques/rapidarc/compari son.html

Advantages of modulated arc treatment 5 Speed of delivery (MLC travel optimized, ref) Less MU, less secondary dose to normal tissue and hence low probability for secondary cancer (Aperture based optimization, ref) Potential for improved optimization due to increased number of beam directions Less wear on MLC

Configuring the Accelerator 6 Prerequisites: On-Board Imager kv Imaging System (OBI) with OBI Advanced Imaging Software - Millennium 120 MLC or High Definition MLC (HDMLC) with dynamic MLC option - ARIA Oncology Information System v8.5 or later, or RapidArccompatible, 4DITC-compatible 3rd party information system - Eclipse Treatment Planning with RapidArc treatment planning software license. -4D-ITC

Configuring the Accelerator 7 Need clutch less system for gantry rotation, dual read outs for gantry angle (CBCT include these) and a stronger chain for the drive. 4DTC software version 8 (Enable the rapid arc treatment on the console under Treatment Administration software) Clinac software v7.8 Enable the Dynalog files if needed.

Configuring Accelerator contd 8 Clinac software v7.8

9 Configuring the Accelerator

Configuring R & V system 10 Mosaiq v1.6001 or up

TPS Commissioning 11 Eclipse V8.6 Commissioning (ready for clinical use) Model photon beam (AAA, PBC, DVO, PRO) Validation of Photon Models (5 depths 1.5 to 20 cm, FS 3 to 20 cm 2, agreement 1% in general) HU vs Electron density calibration table Dicom export/import setup Machine and DMLC Setup QA plan setup IMRT QA verification

TPS Commissioning 12 RPC phantom verification Dynalog files evaluation Dynamic MLC QA (necessary for sliding window IMRT) Constancy verification of RapidArc plan using monthly QA Setup Treatment planning strategies Pinnacle in Progress

13 TPS Commissioning

14 TPS Commissioning

Documents from Philips and Varian

Pinnacle Machine Commissioning This is a test machine!!!

Patient Specific QA Film and Ion Chamber QA 17 Film and Ion Chamber QA analogous to MDA IMRT QA procedure: Create verification plan in Eclipse for OmniPro phantom. Deliver dose to phantom. Compare measured and calculated ion chamber dose and film dose. Film at Z=2.0 cm isocenter a. b. Z=2.0 cm dose plane for export Bars away from beams

Patient Specific QA Ion Chamber/Film Results 18 Absolute Dosimetry Results - Ion Chamber Data Field Energy Ion Chamber Couch Coll Gantry MU (MV) Readings (nc) Dose (cgy) 1 6 0 45 179.9-180 792 3.796 Total Measured Calculated Dose % diff 330.8 330.8 328.0 0.9% Total Measured The ratio of the measured to calculated dose is between 0.95 and = 1.009 Total Calculated 1.05, monitor units do not need adjustment. Absolute dose agreed within 1%. Relative Dosimetry Results - Film Data a. Eclipse Dose Plane b. Measured Film Dose > 99% of film pixels had gamma 1.0 (3% and 3mm). c. Film Gamma Total number of pixels: 78400 Deviation: 0.15 Pixels in Ranges: 0.00 to 0.75 : 78253 (= 99.81 %) 0.75 to 0.87 : 147 (= 0.19 %) d. Isodose Overlay

Treatment Interlocks 19 If an interlock is asserted during treatment, follow normal procedures to clear interlock. If able to clear interlock, treatment will resume from point where interlock asserted. Interlock will not affect the accuracy of dose delivered to the patient. Tested by inserting interlock during treatment of QA phantom and then resuming treatment. Film and ion chamber data were analyzed. Absolute measured dose agreed with plan dose within 1%. The film analysis: >99% of the pixels had a gamma 1

Treatment Interlocks contd 20 Dose delivered to the isocenter is still accurate, even if beam is interrupted multiple times as in gated delivery. Tested by turning beam off repeatedly15 times during delivery to the QA phantom and comparing ion chamber reading obtained from uninterrupted delivery. Results are within 0.2%.

Partial Treatments 21 If interlock can not be cleared and have true partial treatment, the partially treated plan can be reconstructed in Eclipse to assess the dosimetric impact. Total plan MU is entire To assess the dose delivered treatment delivered. from a partial delivery - Adjust this value to the MU delivered. In this example, the treatment was stopped after delivery of 400 MU. Gantry end angle is 0.0. The Gantry end angle can be was compared to stop angle during treatment delivery. QA was performed on the partially delivered plan. Results were comparable to full treatment delivery. Eclipse calculation for one fraction partial arc 0 to 400 MU.

22 Monthly QA - Constancy Verification of RapidArc Delivery 1. Created verification plan of prostate plan using CT data set of the monthly QA phantom. 2. Determined dose to ion chamber (from Eclipse DVH). 3. Delivered RA plan to phantom, measure dose, and compare measured and calculated doses. This will be measured on monthly basis as routine QA.

Patient Specific QA Dynalog files 23 DynaLog file - a record of actual MLC positions for a dynamic treatment These values are compared with planned values. As part of commissioning, we delivered treatment 3X, and reviewed dynalog files. Maximum leaf position deviation for all 3 runs was in the range of 1mm to < 1.5 mm. The maximum deviation occurred for less than 3% of the leaf positions for all 3 deliveries. Negligible dosimetric effect. Recommend dynalog file analysis 3X during a patient s course of treatment.

Linearity Check and Calibration 24 Monitor chamber linearity need to be verified at all available dose rates.(annual) Output calibration need to be verified at all available dose rates. (monthly) MU Ratio normalized to 200 MU @ 400 MU/min 100 MU/min 400 MU/min 600 MU/min 5 1.011 1.012 1.014 10 1.008 1.007 1.015 25 1.005 1.008 1.004 50 1.004 1.001 1.000 100 1.003 0.999 0.999 200 1.003 1.000 0.999 400 1.004 1.000 0.999

DMLC QA picket fence test 25 Arc+fixed Arc only

26 DMLC QA gantry speed & dose rate

RapidArc Prostate Treatment Planning Tips

MDA IMRT Treatment Plan 8 Beam Technique 28

M.D. Anderson IMRT Treatment Plan 29 In addition to DVH constraints, plan should minimize high isodose lines that bisect the rectum. The 30 Gy line should be conformal and not cover the posterior wall of the rectum and at minimum at least 45Gy should always split the rectum.

MDA DVH Criteria for 76 Gy in 38 Fx 30 CTV V100 76 PTV V98% 76 Rectum V60% 40 Rectum V50% 45 Rectum V40% 60 Rectum V20% 70 Rectum V15% 76 Rectum V5% 78-80 Bladder V20% 70 Femoral Heads V50% 45 Femoral Heads V10% 50

Rapid arc planning tips - prostate 31 360º (45 CA) rotation was not good even with NT objectives and avoidance structures Switched to 210-150 (300 arc) Increased the number of avoidance structures in the posterior Increased to two arcs CW and CCW of 210-150 Reduce dose to top and bottom most PTV slices by 5%.

32 Developing a Comparable RA Treatment 4 th Attempt 300 o arc (collimator 45 o ), typical constraints for bladder, rectum, and femoral heads + NT objectives and avoidance structures. Want want tighter to isodose improve lines posteriorly the dose fall-off posteriorly. 60% 14Gy 40% 29Gy 20% 47Gy

Avoidance Structures 1 and 2 33 Avoidance Rings: very similar to rings used by GU service for IMRT treatment planning. Ring 1: 1 cm expansion of PTV (remove PTV + 1 mm). Ring 2: 3 cm expansion of PTV, (remove PTV +1 mm and ring 1).

Avoidance Structures 3 and 4 34 Avoidance Structures: Overlap of rectum and ring 1. Overlap of rectum and ring 2. Ring 1 + Rectum Ring 2 + Rectum

Avoidance Structures 5 and 6 35 Expansions of Ring1+Rectum: Posterior Avoidance 1 (small) 1.5 cm laterally 5 cm posteriorly 1.5cm 4.5cm Posterior Avoidance 2 (large) 4.5 cm laterally 5 cm posteriorly 5cm

Best Results 36 Recall our first attempt plan: And now our best result plan:

Planning PTVs 37 In general, RA plans tend to be hot on the superior and inferior slices. Create Planning PTVs to cool down superior/inferior slices and maintain dose to rest of PTV. Treatment Planning PTVs Planning PTV1 = Only the sup/inf slices of PTV Planning PTV2 = PTV Planning PTV1 Dose Criteria for Planning PTVs 5% less dose to planning PTV 1 Evaluation Criteria for Treatment Plan Evaluate DVH for original PTV

Best Results 38 60% 14Gy 40% 24Gy 20% 44Gy Fall-off distance between 70Gy and 40Gy isodose lines is 0.5cm.

RapidArc Implementation Team 39 RapidArc Team Acknowledgements Rebecca M. Howell, PhD Peter A. Balter, PhD* Ramaswamy Sadagopan, MS* Andrew Lee, MD Catherine Wang, PhD Oleg N. Vassiliev, PhD Milos Vivic, PhD Mary K. Martel, PhD Michael Gillin, PhD Radhe Mohan, PhD Andrew Lee, M.D., Reviewed numerous treatment plans and helped develop the planning strategy. Physics Assistants: Scott LaNeave, Jared Ohrt, and Luke Whittlesey, provided assistance with demonstration and training regarding the current IMRT QA protocol and OmniPro software. Jennifer Johnson, assisted with the Diamond calculation test.

Key references: 40 Commissioning and quality assurance of rapidarc radiotherapy delivery system. Ling CC, Zhang P, LoSasso T. et al: IJROBP 575-581, 2008. Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing IMRT. Lossasso etal: Med Phy p 1919-1927, 25(10) 1998. Volumetric modulated arc therapy: IMRT in a single gantry arc. Otto, Karl, Med Phys, 35(1) p310-317, 2008. Direct aperture optimization: a turnkey solution for step-and-shoot IMRT. Shepard DM etal Med Phys 29(6), p1007-1018, 2002 Inverse planning for intensity-modulated arc therapy using direct aperture optimization. Earl MA etal Phys Med Biol, 48(8), p1075-1089, 2003.

THANK YOU Questions?