Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Similar documents
Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Review of fundamental photon dosimetry quantities

Dosimetric Calculations. Lonny Trestrail

1. Question Answer cgy / MU cgy / MU 2. Question Answer

Math Review. Overview

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Relative Dosimetry. Photons

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Monitor Unit Verification for Small Fields

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

Clinical Implementation of Volumetric Modulated Arc Therapy

3D Treatment Planning and verification with hand calculations

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

High Speed Direct SAD Radiosurgery Beam Scanner

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Constancy checks of well-type ionization chambers with external-beam radiation units

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Planning Assignment (3 field rectum)

Scientific Journal of Nuclear Medicine & Radiation Therapy

Zoubir Ouhib Lynn Cancer Institute

CHAPTER-2. Wedge filters are the most important beam modifiers used in

SnapShot IMRT with compensators and FFF beams

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

OUTPUT at dmax in tissue 8x8 20x20 At 100cm SSD At 100cm SAD

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

IMRT in clinical practice at the UMC-Utrecht. Utrecht

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Radiotherapy physics & Equipments

Introduction. Materials and Methods

Iranian Journal of Medical Physics

Commissioning and periodic tests of the Esteya electronic brachytherapy system

A beam-matching concept for medical linear accelerators

THE development of more advanced techniques in radiotherapy,

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

Investigation of Thermal Effects of CO 2 on Earth-Atmosphere System

TG-51: Experience from 150 institutions, common errors, and helpful hints

Abstract: Introduction:

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

Ultrasound Dose Calculations

How to Design Medical Accelerator Systems for Reliability: IBA PT System

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation

3 1 PRESSURE. This is illustrated in Fig. 3 3.

The use of the analytical balance, and the buret.

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

MATH GRADE 6 UNIT 6 RATE ANSWERS FOR EXERCISES

INTRODUCTION Porosity, permeability, and pore size distribution are three closely related concepts important to filter design and filter performance.

Underwater measurement of photosynthetically active radiation

Exercise 2-3. Flow Rate and Velocity EXERCISE OBJECTIVE C C C

Commissioning of a new total body irradiation protocol

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

General Accreditation Guidance. User checks and maintenance of laboratory balances

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

The dose distribution of medium energy electron boosts to the laryngectomy stoma

INSTRUMENT INSTRUMENTAL ERROR (of full scale) INSTRUMENTAL RESOLUTION. Tutorial simulation. Tutorial simulation

1. BEAM MODIFIERS IN RADIOTHERAPY

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Outlet Structures T-12

WMO LABORATORY INTERCOMPARISON OF RAIN INTENSITY GAUGES

Characterizers for control loops

Ratio & Rate Reasoning PRESENTED BY MR. LAWS 6 TH GRADE MATH

Development of Graphical User Interfaces (GUI) software and database for radiation therapy applications

THE BEHAVIOR OF GASES

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Surface buildup dose dependence on photon field delivery technique for IMRT

Preparation for Salinity Control ME 121

Exposure System Selection

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

DEVIL PHYSICS THE BADDEST CLASS ON CAMPUS AP PHYSICS

APPROVED FACILITY SCHOOLS CURRICULUM DOCUMENT SUBJECT: Mathematics GRADE: 6. TIMELINE: Quarter 1. Student Friendly Learning Objective

A Review of the Bed Roughness Variable in MIKE 21 FLOW MODEL FM, Hydrodynamic (HD) and Sediment Transport (ST) modules

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy

B. Radium Isotope Mass Balance Model Used to Derive Submarine Groundwater Discharge

Chem 110 General Principles of Chemistry

Unit 2 Kinetic Theory, Heat, and Thermodynamics: 2.A.1 Problems Temperature and Heat Sections of your book.

Student Manual. Lesson 4- Nuclear Density Testing. Version 1.0-3/09 4-1

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC

Performance of the Liquid Hydrogen Target

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

PX4 P L A S T I C PX4 PERFORMANCE WIL T-02

Ozobot Bit Classroom Application: Boyle s Law Simulation

Transcription:

Monitor Unit Calculations Part 2 George Starkschall, Ph.D. Department of Radiation Physics U.T. M.D. Anderson Cancer Center Calculation of machine setting reference dose machine setting =, reference dose output reference dose output = calibration dose output correction for collimator setting correction for distance from source. correction for beam modifiers correction for attenuation and scatter Collimator setting Multiply calibration dose output by output factor (field size factor) Dose at d max with collimator set to specified value divided by dose at d max for calibration value If calibration done at another depth d, use ratio of doses at d 1

Collimator setting Factors affecting output If f calibration dose output measured in air only factor causing change in output factor is collimator scatter (S c ) If calibration dose output measured in phantom additional factor due to phantom scatter (S p ) Phantom scatter dependent on field size and not collimator setting Collimator setting Note difference between collimator setting and field size Collimator setting field size of unblocked field at source to isocenter distance Field size is actual (or effective) size of treatment field at specified distance from source Field size may be larger than or smaller than collimator setting Collimator setting If f field size significantly difference from collimator setting, phantom scatter can be accounted for by multiplying by ratio of peakscatter factors PSF PSF field coll. 2

Collimator setting For high-energy (>10 MV or so) linac beams, peakscatter factor essentially independent of field size, so value of 1.0 is often used without introducing serious error Collimator setting Khan approach Khan prefers to separate collimator scatter explicitly from phantom scatter and write the collimator setting correction as S c S p S c : collimator scatter factor S p : phantom scatter factor Summary MDACC approach Khan approach PSF ( ) ( fs) OF cs Sc ( cs) S p ( fs) PSF( cs) 3

Where to find data Data typically available in machine data book (or files) Data normally presented for square fields For rectangular fields, need to determine equivalent square The square field that has the same dosimetric properties as the rectangular field Purely an empirical quantity Equivalent square Table look up Good rule of thumb s = 4 a/p = 2 lw/(l+w) The two are equivalent for all but extremely elongated fields Calculating an output factor Determine equivalent square Look up value in table If not explicitly expressed, estimate interpolation, e.g., output factor for 9 9 should be half the difference between 8 8 and 10 10 Express values to 3 decimal places, e.g., 1.035 4

Distance from source Corrects for differences between distance of calibration point from source and distance of reference point from source Calibration point may be at SAD+d max, while reference point may be at SAD Distance from source Example: Calibrated dose output at SAD + d max from source and reference point at isocenter Distance correction factor is SAD + d SAD max 2. Distance from source Example: Calibrated dose output at isocenter and reference point at isocenter Distance correction factor is 1.000 5

Beam modifiers Accounts for effects of trays, wedges, compensators, anything else that may be placed in beam between radiation source and patient Beam modifiers Beam modifier factors frequently taken to be constant In some cases, field size and/or depth dependent Beam modifiers Care when incorporating wedge factors, especially if extracting information from treatment planning system For some systems, reference dose is dose to reference point for unwedged field reference dose output must be for unwedged field, and no wedge factor included 6

Beam modifier data Beam modifier data typically found in machine data book (or files) Data generally acquired as part of beam commissioning process Effect of patient Must account for attenuation of beam in patient Most common factor is tissue-maximum ratio (TMR) We will present several other factors that may be easier to understand than TMR first Effect of patient If calibration c dose output specified in air -- all calculations up to this point give dose rate to reference point in air To get dose rate in tissue, multiply by tissue- air ratio (TAR) 7

Effect of patient If calibration c dose output specified in tissue at depth all calculations up to this point give dose rate to reference point at depth To get dose rate in tissue, multiply by either tissue-phantom ratio (TPR), or, if specified depth is d max, tissue-maximum ratio (TMR) Let Tissue-air ratio (TAR) D n dose delivered to small mass of tissue at some depth in water phantom D air dose delivered to same mass of tissue in air at same distance from the source, then D TAR = D n air Tissue-air ratio (TAR) D TAR = D n air 8

Tissue-air ratio (TAR) Example: Calculate treatment time required to deliver 150 cgy to a point at 6 cm depth using 8 cm 8 cm 60 Co beam with in-air dose rate of 135 cgy/min at 80 cm SAD. Tissue-air ratio (TAR) Dose rate in air 135 cgy/min TAR for 8 cm 8 cm 60 Co beam at depth of 6 cm found from TAR table to be 0.847 Dose rate in tissue: 135 cgy/min 0.847 = 114.3 cgy/min Tissue-air ratio (TAR) To o deliver dose of 150 cgy requires treatment time of 150 cgy = 1.31min. 114.3 cgy/min 9

Tissue-air ratio (TAR) TAR relates dose to a point to dose in air at the same point Used when patient set up isocentrically Used when calibration dose output expressed in air Typically only for low-energy photon beams TAR properties Even though TAR is no longer frequently used, we will examine some properties of the TAR, as analogous behavior is exhibited by the more commonly used TMR TAR properties Effect of depth: TAR decreases with increasing depth beam attenuation TAR evaluated at depth of central axis maximum (d max ) backscatter factor (BSF) For depths <d max, TAR increases with depth 10

Effect of beam quality: At low energies BSF increases with increasing energy increased penetration of scattered photons TAR properties Effect of beam quality: At high energies BSF decreases with increasing energy greater amount of forward scatter TAR properties Effect of beam quality: BSF reaches maximum value around HVL 0.4-0.8 mm Cu where it may be as high as 1.5 TAR properties 11

TAR properties Effect of field size: TAR increases with increasing field size increased amount of scatter reaching calculation point TAR properties Effect of source distance: TAR independent of source distance both D n and D air measured at same distance from source TAR data Find TAR tables: For 60 Co: BJR, supplement 25 For specific linacs: in physics data notebook Important to use correct table for appropriate machine 12

Tissue-air ratio (TAR) In n order to calculate dose rate at distance other than the SAD use inverse square relationship to go from one source distance to another Important to note that inverse square relationship only holds in air Tissue-air ratio (TAR) To relate dose rate in tissue at one distance to that at another: Calculate dose rate in air by dividing by appropriate TAR Use inverse square law to go from one distance to the other Calculate dose rate in tissue at new distance by multiplying in-air dose rate by appropriate TAR 13