VAPOMATIC ANESTHETIC VAPORIZER

Similar documents
M5400 Operators Manual

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Installation & Operation Manual 2011 Midmark Corp. 60 Vista Drive Versailles, OH USA

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

36E DSI, HSI & Proven Pilot Two-Stage Combination Gas Valve INSTALLATION INSTRUCTIONS

NARKOVET ANAESTHETIC UNIT

DFB FLOW METERS (DESIGNED FOR BLENDERS, DUAL TAPER)

Flow meter. bellow vaporizer. APL valve. Scavenging system

ANESTHESIA VAPORIZER SERVICE RECOMMENDATIONS

ACRYLIC FLOW METERS 0-30, 0-70 & LPM

Yoke Block Instruction Manual

Operator: Save these instructions for future use!

Interface Devices, Inc. Hydraulic Mini Mule

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

Yoke Block Instruction Manual

GILMONT ACCUCAL FLOWMETERS

Installation Instructions

User Instruction Manual

Needle valve. Contents. User s Manual. (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2

36E03 and 36E38 DSI and HSI Step Opening Combination Gas Valve INSTALLATION INSTRUCTIONS

Parker Hannifin Corporation Porter Instrument Division 245 Township Line Road Hatfield, PA USA Tel: Ref Rev.

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

2 CYLINDER MOBILE CART USER S MANUAL / INSTRUCTIONS

Sigma Delta Vaporizer User Instruction Manual

Installation and Use Manual

Installation Instructions

Installation/Care/Use Manual

with O 2 Controller Instruction Manual

INSTALLATION INSTRUCTIONS

MANUAL BE SERIES Test Benches

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

36E DSI and HSI Two-Stage Combination Gas Valve INSTALLATION INSTRUCTIONS

Datum 'L' Vaporizer. Blease. L Series. Anaesthesia Vaporizer. User Manual. Blease. User Manual

Budget Range Operators Handbook

Table Top and Convertible Anesthesia Machine Models: M3000 and M3100 (SHOWN WITH OPTIONAL EQUIPMENT)

INSTALLATION, CARE & USE MANUAL. EMASMB & LMASMB Surface Mount Bottle Filling Stations

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.

APC Pressure Reducer with Sensor

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

3/8" Dr. Air Butterfly Impact Wrench

Sample Concentrator SBH CONC/1 Instructions for use

453 Series Steam Heated Vaporizing Regulator

Instruction Manual Updated 7/26/2011 Ver. 2.2

FILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

OWNER S MANUAL. Page: 1 of 8

CSA Sample Draw Aspirator Adapter Operator s Manual

Instrument Operating Manual

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

Aquor House Hydrant V2

TABLE OF CONTENTS SAFETY FIRST!...3 BASIC OPERATION...4 ADJUSTMENTS...5. Cocking Pressure...5. Three-Way Valve Adjustment...6

AND OPERATION INSTRUCTIONS

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

RG1200 Service and Repair Manual

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Model PIV UL/FM Post Indicator Valve. Maintenance and Operation Manual UL/FM Post Indicator Valve Configuration AWWA C515

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

APCO ARV CLEAN WATER AIR RELEASE VALVES. Model 50A

Wafer Check Valve. Contents. User s Manual. (1) Be sure to read the following description of our product warranty 1

OXYWAY. Service and Repair Instructions

TITAN FLOW CONTROL, INC.

CPR MEDICAL DEVICES INC. OPERATING MANUAL

SILENTAIRE TECHNOLOGY

OWNER'S MANUAL. Copyright 2003 GAMMA - All Rights Reserved

INSTRUCTION MANUAL Pressure Relief Device LPT

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

Types 749B and R130 Changeover Manifolds

Model VR6 System. Installation, Operation & Maintenance

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

RK-IR Sample Draw Aspirator Adapter Operator s Manual

OWNER'S MANUAL. Copyright 1999 ATS - All Rights Reserved

Techcon Systems TS1254 Pressure Pot

Eaton Filtration, LLC

Instruction Manual. PE-25/32/37 Tapping Machines. POLYTAPP Valves

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

MODEL 1329 Tank Gauge

DS06D,G Dial Set Pressure Regulating Valve

SILENTAIRE TECHNOLOGY

SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL

BlenderBuddy 2. Operating Manual & Instructions For Use ENGLISH R233M01 REV. A

Constant Pressure Crude Oil Container Model CPCCP

J Air and Water Kit Instructions Part# 02584

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

Float Operated Level Controllers

DEH Instructions. Three-Phase, Platform-Mounted Distribution Transformers

Lab Cock. Contents. User s Manual (1) Be sure to read the following warranty clauses of our product 1. (2) General operating instructions 2

KOBOLD KSM FLOWMETER. User Instructions

SALCO PRODUCTS, INC. PRESSURE RELIEF VALVE STORAGE, INSTALLATION, OPERATING, MAINTENANCE/TESTING, AND INSPECTION INSTRUCTIONS

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

Steel Spring Driven Reel Instruction Manual

KTM OM-2 SPLIT BODY FLOATING BALL VALVES INSTALLATION AND MAINTENANCE INSTRUCTIONS

High Pressure Inlet Kits for CONCOA BlendMites and BlendMasters

Eaton Filtration, LLC

Transcription:

VAPOMATIC ANESTHETIC VAPORIZER INSTRUCTION MANUAL SOLD FOR VETERINARY USE ONLY BY: A.M. BICKFORD, INC. 12318 BIG TREE ROAD WALES CENTER, NY 14169 1-800-795-3062 Fax: 716-652-2046 www.ambickford.com sales@ambickford.com

INSTALLATION: VAPOMATIC ANESTHETIC VAPORIZER The Vapomatic is normally mounted on the back bar of an anesthetic machine or in any case it should be solidly mounted to a rigid structure in an upright position. Two tapped holes in the rear mounting face (1/4-20) are provided for mounting. Since these are blind holes of 5/16 depth, screws of a proper length must be used so that they may be tightened before bottoming. A suggested mounting arrangement which allows flexibility in coping with various structure thicknesses is shown in Fig. 1 In some cases it may be desirable to mount the Vapomatic on a table or horizontal surface. A ¼ -20 tapped mounting hole is provided for this purpose in the center bolt at the bottom of the vaporizer. CONNECTIONS: Standard connections are (facing the front of the Vapomatic) left side, inlet, cage mount/23mm Male standard taper, and right side, outlet, cage mount/23mm Female standard taper. Normally, in veterinary use, mating connectors having tubing adapters are used, so that rubber tubing can be connected to the gas supply, either the flow meter common outlet, or the outlet terminal of the gas machine. CAUTION: The vaporizer inlet should not be connected downstream of the flush valve, otherwise excessive back pressure may be reflected back into the flow meters and other parts of the circuit when the flush valve is activated, possibly causing bursting of tubing or other components. The vaporizer outlet must be connected to the patient via the outlet of the gas machine or directly to the breathing circuit which would be, to the CO2 absorber, or to the non-rebreathing patient circuit. 2

CAUTION: Always check that the connection to the vaporizer and to the gas supply and patient are secure when using the equipment. Remember, the only way your patient can get life giving oxygen is through these connections. After installation, turn on the oxygen supply and check for leaks at the vaporizer connections by partly blocking the outlet of the machine and painting the joints with soap solution. INSTRUCTIONS FOR USE: CONTROLS; The dial knob on the top of the vaporizer is marked in % of volume of the anesthetic for which the vaporizer is calibrated. In the off or 0 position, the dial locks to prevent accidental turn on. To turn on the vaporizer, press down the locking key while rotating the dial counter clockwise. A resistance to turning will be noted during the first few degrees of rotation as a cam operated by-pass valve is activated. This portion of the dial scale is marked and should not be used. When in the 0 position, gas entering the vaporizer is by-passed to the outlet and the vaporizing chamber is sealed off no anesthetic is entering the gas stream. As the dial is rotated counter clockwise, the output concentration gradually increased in accordance with the calibration marked. FILLING AND DRAINING; To fill the vaporizer, turn the dial to 0 if the vaporizer is in an active situation (it is not then necessary to turn off the gas supply), unscrew the filler cap and fill the vaporizer with the appropriate anesthetic. The liquid level indicator is uncalibrated, but it is only necessary that the liquid level be seen to be acceptable for operation. A dry vaporizer will accept approximately 200 ml of liquid and approximately 150 ml when the wicks are wet. The top of the filler is lower than the working parts of the vaporizer so that it is not possible to overfill the vaporizer when in a level position. Replace the filler cap snugly so that it is leak tight but still can be removed by hand. To drain the vaporizer, remove the filler cap exposing the head of the drain screw. Position a collecting bottle under the drain spout and then loosen the drain screw a turn or two. After draining, secure the drain screw and replace the cap. It is advisable to drain the vaporizer when it is to be moved, transported, or shipped, to prevent draining, should be done to reduce the concentration of Thymol, a preservative, from collecting in the liquid. 3

TYPICAL VAPORIZER APPLICATIONS (ON A BASIC ANESTHESIA MACHINE) 4

CAUTION: Do not connect the vaporizer within breathing circuit. The vaporizer is of the high resistance to flow type, and will present excessive back pressure to the inlet circuit and to the vaporizer sump. Note that calibration is given for the normal flow rate of 0-10 liters/min. Direct connection to a ventilator is not recommended since the vaporizer will offer excessive resistance and develop abnormal back pressure which may cause connections to blow off or damage the vaporizer. These same precautions are relevant to all vaporizers of this type. MAINTENANCE AND SERVICE: Vaporizers used with halothane are particularly liable to suffer from the effect of accumulation of non-volatile additives in the vaporizing chamber. Similar effects may also occur with other anesthetics although more slowly. Anytime the performance of a vaporizer is suspect, or as a periodic routine, it maybe shipped to the manufacturer, A.M. Bickford, Inc., for service and calibration. If the vaporizer is damaged by dropping or other abuse, it should be returned for service and repair before being used again. A yearly service and calibration is recommended. CALIBRATON; Vapomatic vaporizers are calibrated for the specific agent marked on the labeling. Only the specified anesthetic should be used. Each vaporizer is individually calibrated by means of a Riken interference refractometer at a flow rate of 3 liters/min. and a room temperature of 72 degrees F using oxygen as a carrier gas. Calibration tolerances are met over flow rates of 100 ml/min. through 10 liters/min. with a substantially linear output, in ambient temperature range of 60-90 degrees F. 5

RETURNING VAPORIZERS: For routine repair or service: Very carefully tightly pack the vaporizer to prevent damage in shipping. Preferred: Ship by United Parcel Service. Address: A.M. Bickford, Inc. Service Department 12318 Big Tree Rd. Wales Center, NY 14169 We will perform the necessary work to restore the vaporizer to specifications, bill you our flat service charge plus replacement parts, if any, and return it to you promptly by UPS. LIMITED WARRANTY: Vapomatic Anesthetic Vaporizers are warranted against defective parts and materials for 1 year beyond the shipping date to the original customer. Such claims should be communicated to A.M. Bickford, Inc., and return authorization obtained prior to return of vaporizer. Choice of action by A.M. Bickford, Inc., will be limited to repair or replacement and return of the vaporizer at no cost if A.M. Bickford, Inc. concurs that inspection of the unit indicates such defects. FOR WARRANTY CLAIMS: Call or write describing the condition, give the serial number and other pertinent details. If the situation appears to be warranty related, we will authorize you to return the vaporizer for our further evaluation. 6