TYPE APPROVAL CERTIFICATE Certificate No: TAA0000123 This is to certify: That the Pressure Transmitter with type designation(s) SITRANS P series DS III, P310, P410 Issued to Siemens AG - Process Industries and Drives Karlsruhe, Germany is found to comply with DNV GL rules for classification Ships, offshore units, and high speed and light craft Application : Product(s) approved by this certificate is/are accepted for installation on all vessels classed by DNV GL. Location classes: Temperature D Humidity B Vibration B EMC B Enclosure D (IP 66/68) Issued at Hamburg on 2017-05-18 This Certificate is valid until 2022-05-17. DNV GL local station: Augsburg Approval Engineer: Dariusz Lesniewski for DNV GL Joannis Papanuskas Head of Section This Certificate is subject to terms and conditions overleaf. Any significant change in design or construction may render this Certificate invalid. The validity date relates to the Type Approval Certificate and not to the approval of equipment/systems installed. Form code: TA 251 Revision: 2016-12 www.dnvgl.com Page 1 of 7 DNV GL 2014. DNV GL and the Horizon Graphic are trademarks of DNV GL AS.
SITRANS P transmitter for pressure Two-wire technology, series P310, P410, DS III Data position à (type: pressure) 2 4 Series P310 DS III / P410 4 5 Type Pressure relative 0 Pressure absolute 2 7 Communication 4-20 ma HART 3 Profibus PA 4 Foundation Fieldbus FF 5 8 Measuring cell filling Silicone oil 1 Inert liquid 3 Neobee oil 4 9 Measuring span Data position 5 = 0 Data position 5 = 2 to 0,25 bar A to 1 bar B to 4 bar C to 16 bar 0,25 bar D to 63 bar E to 160 bar 1,3 bar F to 400 bar 5 bar G to 30 bar H to 700 bar J to 160 bar L to 400 bar M to 700 bar N 10 Material of wetted parts seal diaphragm process connection Stainless steel Stainless steel A Hastelloy Stainless steel B Hastelloy Hastelloy C 11 Process connection Connection shank G½ A 0 Female thread ½ -14 NPT 1 Oval flange 7/16-20 UNF 2 Oval flange M10 3 Oval flange M12 4 M20x1,5 A 5 ½-14 NPT A 6 12 Material of non-wetted parts (Enclosure) Housing diecast aluminium 0 Housing stainless steel precision casting 3 Form code: TA 251 Revision: 2016-12 www.dnvgl.com Page 2 of 7
Data position à (type: pressure) 13 Version (Inscription and documentation) Standard version (German plate inscription, unit: bar) 1 International version (English plate inscriptions, doc. in 5 Languages, unit: bar) 2 Chines version (English plate inscriptions, unit: Pascal) 3 14 Explosion protection without explosion protection A ATEX Intrinsic safety (Ex ia) B ATEX Explosion-proof (Ex d) D ATEX Zone 2 (Ex na/ic) E FM + CSA intrinsic safe (is) F FM + CSA IS + XP, AEx ia/ex ia N ATEX Ex ia + Ex d P ATEX Ex ia + Ex d + Ex na/ic + Dust 1D/2D R FM + CSA (is + ep) + ATEX (Ex ia + Ex d + Zone 1D/2D) S 15 Elecrical Connection Female thread PG 13.5 A Female thread M 20 x 1.5 B Female thread M ½ 14 NPT C Connector M 12 (M20x1.5) F Connector M 12 (½ -14 NPT) G 16 Display without indicator 0 without window (digital display covered, setting: ma 1 without window (digital display covered, settings according order code Y21/Y22 2 with window (digital display, visible, setting: ma 6 with window (digital display, visible, setting acc. to spec, short spec. Y21/Y22 necessary 7 Other versions Add -Z to Article No. and add order code Desription Mounting bracket 316L Variants cable entry (M12, cable glands) Supplied with oval flange set Inscription and documentation in different languages Certifications Order code for P410 (increased measuring accuracy) Options for Oxygen applications Customized special configuration (labeling, documentation, presettings software) Manufacturer's declaration acc. to NACE (MR 0103-2012 and MR 0175-2009) Degree of protection IP66/IP68 Process flange screws made of Monel Approvals explosion proof Two coats of lacquer on casing and cover Marine Type Approval DNV-GL Valve manifold Customized special configurations (presettings software, labeling) Code A03 (only in connection with option 0 of the pos. 12) A50; D32; D38; D43; D52 D37 B00 B99 C11...C23; C61 C41 D01...02; E10 D05; D10; D35; D48; D50; D60; D61 D07 D12 D34 E01 E99 G10 S10 T02; T03; T06 Y01 Y98 Form code: TA 251 Revision: 2016-12 www.dnvgl.com Page 3 of 7
SITRANS P transmitter for differential pressure Two-wire technology, series P310, P410, DS III Data position à (type:differential pressure) 2 4 Series P310 DS III / P410 4 5 Type Differential pressure (absolute) 3 Differential pressure PN160 4 Differential pressure PN420 5 7 Communication 4-20 ma HART 3 Profibus PA 4 Foundation Fieldbus FF 5 8 Measuring cell filling Silicone oil 1 Inert liquid 3 Neobee oil 4 9 Measuring span Data position 5 = 3 Data position 5 = 4, 5, 6 to 0,02 bar B to 0,06 bar C to 0,25 bar 0,25 bar D to 0,60 bar E to 1,30 bar 1,60 bar F to 5 bar 5 bar G to 30 bar 30 bar H to 100 bar K 10 Material of wetted parts Seal diaphragm Measuring cell Process connection Stainless steel Stainless steel Stainless steel A Hastelloy Stainless steel Stainless steel B Hastelloy Hastelloy Stainless steel C Tantalum Tantalum Stainless steel E Monel Monel Stainless steel H Gold Gold Stainless steel L Order code "K01 or K02" changes material of process connection * to Hastelloy or Monel 11 Process connection (Mounting Thread) M10 0 M20 1 7/16-20 UNF (PN160) 2 7/16-20 UNF (PN420) 3 M10 vented 4 M12 vented 5 7/16-20 UNF vented (PN160) 6 7/16-20 UNF vented (PN420) 7 12 Material of non-wetted parts (Enclosure) Housing diecast aluminium 2 Housing stainless steel precision casting 3 Form code: TA 251 Revision: 2016-12 www.dnvgl.com Page 4 of 7
Data position à (type:differential pressure) 13 Version (Inscription and documentation) Standard version (German plate inscription, unit: bar) 1 International version (English plate inscriptions, doc. in 5 Languages, unit: bar) 2 Chines version (English plate inscriptions, unit: Pascal) 3 14 Explosion protection without explosion protection A ATEX Intrinsic safety (Ex ia) B ATEX Explosion-proof (Ex d) D ATEX Zone 2 (Ex na/ic) E FM + CSA intrinsic safe (is) F FM + CSA IS + XP, AEx ia/ex ia N ATEX Ex ia + Ex d P ATEX Ex ia + Ex d + Ex na/ic + Dust 1D/2D R FM + CSA (is + ep) + ATEX (Ex ia + Ex d + Zone 1D/2D) S 15 Electrical Connection Female thread PG 13.5 A Female thread M 20 x 1.5 B Female thread M ½ 14 NPT C Connector M 12 (M20x1.5) F Connector M 12 (½ -14 NPT) G 16 Display without indicator 0 without window (digital display covered, setting: ma 1 without window (digital display covered, settings according order code Y21/Y22 2 with window (digital display, visible, setting: ma 6 with window (digital display, visible, setting acc. to spec, short spec. Y21/Y22 necessary 7 Other versions Add -Z to Article No. and add order code Desription Gasket variants Variants cable entry (M12, cable glands) Variants of fittings and mounting for process flanges Inscription and documentation in different languages Certifications Order code for P410 Options for Oxygen applications Customized special configuration Manufacturer's declaration acc. to NACE (MR 0103 2012 and MR 0175 2009) Degree of protection IP66/IP68 Process flange screws made of Monel Approvals explosion proof Two coats of lacquer on casing and cover Special options process flanges Material process flange Marine Type Approval DNV GL Valve manifold Customized special configurations Code A20 A29, A70 A79, J02 J05 A50, D32, D38, D43, D52 A40, J08 J09, L09 L10, D37 B00 B99 C11 C23, C61 C41 D01...02, E10 D05, D10, D35, D48, D50, D60, D61 D07 D12 D34 E01 E99 G10 H00 H99 K01, K02 S10 U02, U04 Y01 Y98 Form code: TA 251 Revision: 2016-12 www.dnvgl.com Page 5 of 7
Places of manufacture Siemens Production Automation S.A.S I IA SC MF-H 1, chemin de la Sandlach B.P. 60189 67506 Haguenau Cedex, France Siemens Canada Limited 1954 Technology Drive Peterborough ON K9J 6X7, Canada Siemens Ltd. Kalwa Works Thane Belapur Road Thane - 400601, INDIA Siemens Sensors and Communications Ltd. (SSCL) No.117, GuangXian Road, QiXianLing, High-Tech. Industry Zone, Dalian 116023, P.R.CHINA Application/Limitation The Type Approval covers hardware listed under. When the hardware is used in applications to be classed by DNV GL, documentation for the actual application is to be submitted for approval by the manufacturer of the application system in each case. Reference is made to DNV GL rules for classification of ships Pt.4 Ch.9 Control and monitoring systems. Assembly notes in the manufacturer s operating instruction (compact) to be observed. Type Approval documentation Documents as listed in the document: A5E38459162A _Rev02 DNV GL Typa approval assessment report issued at Haguenau on 2017-01-24 DNV GL Typa approval assessment report issued at Dalian on 2017-05-18 DNV GL Typa approval assessment report issued at Thane on 2017-04-10 Tests carried out Applicable tests according to class guideline DNVGL-CG-0339, November 2016. Marking of product The products to be marked with: - manufacturer name - model name - serial number - power supply Periodical assessment The scope of the periodical assessment is to verify that the conditions stipulated for the type are complied with, and that no alterations are made to the product design or choice of systems, software versions, components and/or materials. The main elements of the assessment are: Ensure that type approved documentation is available Inspection of factory samples, selected at random from the production line (where practicable) Review of production and inspection routines, including test records from product sample tests and control routines Ensuring that systems, software versions, components and/or materials used comply with type approved documents and/or referenced system, software, component and material specifications Form code: TA 251 Revision: 2016-12 www.dnvgl.com Page 6 of 7
Review of possible changes in design of systems, software versions, components, materials and/or performance, and make sure that such changes do not affect the type approval given Ensuring traceability between manufacturer s product type marking and the type approval certificate Periodical assessment is to be performed at least every second year and at renewal of this certificate. END OF CERTIFICATE Form code: TA 251 Revision: 2016-12 www.dnvgl.com Page 7 of 7