Neotecha SAPRO - Aseptic sampling valve SV Installation, operation and maintenance instructions

Similar documents
Neotecha SAPRO - Tank Valve Type ST for aseptic sampling with bottles Installation, Operation and Maintenance Instruction

Neotecha Tank Sampling Sapro

Neotecha Tank Sampling Sapro

Ball Float Steam Trap UNA 45 MAX, UNA 46 MAX, UNA 46A MAX PN 40/Class 300 DN 40, 50, 65

Declaration of Conformity as per Directive 97/23/EC

SCA Series Inverted Bucket Steam Traps

Ball Float Steam Trap UNA 43 PN 16/CL 125/JIS 10K UNA 46 PN 40/CL 150/CL 300/JIS 10K/JIS 20K DN 80, 100, 150, 3", 4", 6"

Pressure Reducing Valve DMV 750 set range: bar

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10

CAST IRON SAFETY VALVE TYPE 6301

Declaration of Conformity as per Directive 97/23/EC and Manufacturer s Declaration as per Directive 98/37/EC

G type DUCTED EXHAUST SAFETY VALVE 2871 AND 288X SERIES. Model/Ref:

Declaration of Conformity as per Directive 97/23/EC

Pressure Reducing Valve DMV 755

VALFONTA INSTRUCTIONS: OPERATION AND INSTALLATION PRESSURE REDUCING VALVE MODEL PRV44

Pressure relief valve DHV 712 DN 65-80: 0,5-10 bar, DN : 0,3-4 bar, DN 100: 0,5-6 bar

Pressure Reducing Valve DMV 755 set range: 1,0-9,0 bar

Contents. 1. General information BROEN BALLOMAX Steel Ball Valves Approvals Quality Management...2

Declaration of Conformity as per Directive 97/23/EC

FT16 Ball Float Steam Trap Installation and Maintenance Instructions

Steam Trap BK 15 (DN40, DN 50) Original Installation Instructions English

Installation & Operation Manual Proven Quality since 1892

Pressure relief valve DHV 716 set range: 0,5-10,0 bar

Pressure Relief Valve DHV 718

Series 44 Self-operated Pressure Regulators. Type 44-0 B Pressure Reducing Valve. ANSI version

M45 ISO Ball Valve DN25 to 150 Installation and Maintenance Instructions

CA10S, CA14 and CA14S Air and Gas Traps

CVS10 Sanitary Check Valves

Mounting and Operating Instructions EB 2172 EN. Series 43 Temperature Regulators. Type 43-5 and Type Type Type 43-6

TD45 Thermodynamic Steam Trap Installation and Maintenance Instructions

6301 TYPE CAST IRON SAFETY VALVES

AE50S Automatic Air and Gas Vent for Liquid Systems Installation and Maintenance Instructions

Series 44 Self-operated Pressure Regulators. Type 44-1 B Pressure Reducing Valve. Type 44-6 B Excess Pressure Valve. ANSI version

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

Operating instructions - Translation of the original -

HM and HM34 Inverted Bucket Steam Traps Installation and Maintenance Instructions

Type 3709 Pneumatic Lock-up Valve. Translation of original instructions. Mounting and Operating Instructions EB 8391 EN

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Pressure relief valve DHV 725 set range: 0,2-10,0 bar

Tank Blanketing Pressure Regulators RHPS Series

ITT RICHTER CHEMIE-TECHNIK

T 2523 EN Type 2406 Excess Pressure Valve Self-operated Pressure Regulators ANSI version

Operating instructions - Translation of the original -

Welcome to the LESER Seminar, Taipei 28. June Design_of_safety_relief_valves_250804_Cal

Serie 06-M6. Swing wafer check valve. made in. Application fields. Check valves E U R O P E WATER CONDITIONING INDUSTRY

Self-operated Pressure Regulators Type 2405 Pressure Reducing Valve

BSA6T and BSA64T Stainless Steel Bellows Sealed Stop Valves Installation and Maintenance Instructions

A soft sealed atmospheric safety valve with large flow capacity and soft seat tightness for steam condenser and turbine applications

Operating instructions - Translation of the original -

MST21 Stainless Steel Balanced Pressure Thermostatic Steam Trap

Instruction Manual Contact Pressure Vacuum Gauge

Mounting and Operating Instructions EB EN. Type Supply Pressure Regulator. with increased air capacity

AVS32 Stainless Steel Air Vent for Steam Systems Installation and Maintenance Instructions

Ball valve HKSF-W100. Ball valve HKSF-W100. RMA Kehl GmbH & Co. KG Oststrasse 17 D Kehl / Germany

Steam Trap BK BK 212-ASME. Original Installation Instructions English

OPERATING MANUAL CHEMTRAC 840M MANUAL RETRACTABLE PROCESS HOLDER

Mounting and Operating Instructions EB 8546 EN. Type 4708 Supply Pressure Regulators. Translation of original instructions. Type Type

Pressure Relief Valve DHV 718

KBV21i and KBV40i Air Actuated Boiler Blowdown Valves

Size : Ends : Min Temperature : Max Temperature : Materials : Cast iron body

Manual Actuated Boiler Blowdown Valves

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

Pressure Reducing Valve

Atmospheric relief valve type 1100 Installation and maintenance instructions

Type 4708 Supply Pressure Regulator

SAPAG. Safety valves, type 5700 Storage, Use, Operation and Maintenance Instructions. IMPORTANT NOTICE

Installation / Operation / Maintenance Manual KLINGER Reflex Level Gauge Type R 25

Materials : Forged carbon steel A105N

Pressure relief valve

Mounting and Operating Instructions EB 3017 EN. Self-operated Regulators

Siphon vessel / Priming aid

Operating instruction

Combined Pressure / Vacuum Relief Valve KITO VD/o

Control Valve with ZK Radial Stage Nozzle ZK 210. Original Installation Instructions English

OFFICINE OROBICHE S.p.A. 1/9

2-Port Seat Valves with Flange, PN 16

Materials : Stainless steel ASTM A351 CF8M

KBV21i and KBV40i Key Operated Boiler Blowdown Valves Installation and Maintenance Instructions

Regulating valves Type REG 6-65 Type REG-SS REFRIGERATION AND AIR CONDITIONING. Technical leaflet

TP1 and TP2 Temporary Cone Shaped Strainers

Telefon (+45) Telefax (+45)

Instruction presentation Special valves

Type BBS-03, BBS-05, BBS-06, BBS-25

PENBERTHY SUBMERSIBLE AUTOMATIC SUMP DRAINER INSTALLATION, OPERATION AND MAINTENANCE INSTRUCTIONS

Solenoid valves, two-step on/off Type PMLX REFRIGERATION AND AIR CONDITIONING. Technical leaflet

CA16 and CA16S-14 Air and Gas trap Installation and Maintenance Instructions

CRP INSTALLATION, OPERATING AND MAINTENANCE INFORMATION FOR INLINE SAMPLING VALVES

A3S Bellows Sealed Stop Valve Installation and Maintenance Instructions

MAINTENANCE AND OPERATING MANUAL

Operating Manual FSV, FSÖV, SSVF

Instruction Manual. Alfa Laval SB Membrane Sample Valve ESE02963-EN Original manual

Diaphragm relief valve

Series 42 Self-operated Regulators Differential Pressure Regulators with Type 2424/Type 2428 Actuator (closing) Type Type 42-28

SV5 Safety Valve Installation and Maintenance Instructions

Installation, Operation and Maintenance Manual for Back Pressure Regulator

2-Port Seat Valves with Flange, PN 16

Type 2080 FALTENBALGVENTIL BELLOWS VALVE VANNE A SOUFFLET. Operating Instructions. Bedienungsanleitung Manuel d utilisation

M10HTi ISO Tobacco Ball Valve Screwed, SW, BW and Flanged versions

Regulating valves REG-SA and REG-SB

Series 44 Self-operated Pressure Regulators. Type 44-1 B Pressure Reducing Valve. Type 44-6 B Excess Pressure Valve

Transcription:

Before installation these instructions must be fully read and understood 2 Safety Please read these instructions carefully. 2.1 General danger potential Not paying attention to this instruction Incautious use or misuse Unqualified personnel or staff Use Valves for aseptic sampling of liquid mediums in tubes. 1 General information about this instruction These instructions include all directs to safely install and operate the SAPRO-aseptic sampling valve. If any problems arise, which cannot be solved with these instructions, please call for more information at the manufacturer. These instructions are corresponding to the valid EN-safety standards. Using this armature requires the operator, design engineer or the fitter of the plant to observe all valid national technical standards. Any technical change or improvement can only be lead trough the manufacturer. Using these instructions (and therefore also using the sampling valve) requires qualified personnel, installation and operating staff needs to be trained according to these instructions. 1.1 Validity of these installation, maintenance and repair instructions These installation, maintenance and repair instructions are valid for all SAPRO-aseptic sampling valves, built at Neotecha AG after March, 07. 2002. The period of validity ends with the release of a new revision of the instructions. 2.2 Use as directed 2.2.1 Area of use SAPRO-aseptic sampling valves are armatures, used for taking exact aseptic samples from mediums in tubes. In combination with the special aseptic bottle adapter, it is possible to take samples from any high-corrosive or hot medium without any contact of the medium with (maybe polluted) air. It is even possible to fill the sampling bottle with nitrogen before taking a sample, and also clean the inner parts of the bayonet sleeve and bottle adapter with cleaning liquid or to sterilize them with steam. When not taking samples, the SAPRO-aseptic sampling valve can be closed with a dummy adapter, preventing the bayonet sleeve from pollution. The SAPRO-aseptic sampling valve can only be installed in horizontal tubes, there is no possibility to install the valve in a vertical tube. All parts of the sampling valve in contact with medium are in stainless steel 1.4435 (316L), Hastelloy (different types), PTFE, PEEK or Perfluore. 2.2.2 Detailed function of the aseptic sampling valve The valve stem seals against the bottle adapter in the valve seat. By lifting up the stem with the dead man lever, the sampling medium flows through the bottle adapter into the bottle. The air (or nitrogen) in the bottle is blown out through a small drilling on the side of the bottle adapter, closed with a cross-cut silicone septum to prevent pollution reaching into the bottle. All parts in contact with the medium can be cleaned with cleaning liquid or steam before taking a sample, the bottle can be attached to the bottle adapter in the laboratory, the bottle may even be filled with nitrogen and the adapter keeps entirely closed until it s connected to the bayonet of the sampling valve. When not taking a sample, the bayonet sleeve of the valve can be closed with a dummy adapter as a protection against pollution. www.valves.emerson.com 2017 Emerson. All rights reserved. VCIOM-01983-EN 15/09

2.2.3 Technical data Pressure range: 20Pa vacuum to 16 bar Temperature range: See diagram Sizes DIN PN 16: DN 40-100 ANSI cl. 150: NPS 1½ - 4 Testing pressure: 1.5 x PN = 24 bar 2.2.4 Restricted use The valve can only be used with products when the materials used (SS, PTFE, PEEK or Perfluore) are compatible. Information about the chemical resistance of these materials is found in technical literature or may be asked at the manufacturer. 2.2.5 Illegal technical changes It is illegal to do mechanical changes of the design of the SAPRO-aseptic sampling valve, or to use parts of other manufacturers for repairs, otherwise Neotecha can not guarantee for the safety of the valve. Repairs can only be done by Neotecha or personnel specially trained by Neotecha. 2.2.6 of misuse The Sapro aseptic sampling valve with all connection pipes and actuators can never be used to climb up or to stand on it. 2.2.7 Liability to obtain all directs about operation, maintenance and repair These instructions are a part of the product delivery and has to be available for all operators of the valve, stored well protected from pollution. 2.3 Danger sources 2.3.1 Chemical external The sleeve on the actuator should not be cleaned with solvents, use alcohol or a mild detergent. 2.4 Requests for the operators Operators are people who are familiar with the installation, operation, maintenance and repair of the SAPRO-aseptic sampling valve. They also have to be qualified enough to safely handle the valve and they have to: Keep all technical, internal and national regulations and instructions. Be trained according national safety standards, for example the use of their personal safety clothes or heat gloves, which have to be appropriate to production conditions. Have read and understood these instructions. 3. Transportation and storage The SAPRO-aseptic sampling valve is delivered with protection caps. Please leave these caps on the valve until it is ready to be installed. The aseptic sampling valve has to be stored in a dry, dust-free place until it s installed. 3.1. Transportation Transportation temperature range -20 C to 65 C. Protect against mechanical stress (shock, punch, vibration). 3.2. Storage Storage temperature range -20 C to 65 C, dry and dust-free. Please use driers or heating against condensation water in humid storerooms. 3.3. Handling before installation Remove the protection caps just before installation. Protect the sampling valve against wetness (otherwise dry before installation). Handling as directed protects from damage. Pressure/temperature diagram bar 20Pa vacuum Fig. 1 2.3.2 Electrically To prevent static charges, the entire aseptic sampling valve is made of conductive metal. 2.3.3 Thermically With a sample medium temperature range between -20 C to 200 C, the surface of the sampling valve may assume the same temperature. It is therefore necessary to wear gloves to avoid injuries when operating the valve. 2.3.4 Protection against opening the valve by mistake (see fig. 1) To prevent opening the valve by mistake, the locking pin (3) can be used. A padlock (10, not included in delivery) can additionally be used against deliberate opening. 2

4. Characteristics 4.1 General characteristics Connection versions: I-Clamp Tri-Clamp Other connection versions are available, ask the manufacturer. Body styles: Flanged DIN-length Flanged ANSI-length Screws: Screw quality A2 for all pressure loaded screws Adapters: Aseptic bottle adapter for glass bottles, thread ISO GL45 Dummy adapter to protect the valve from pollution Cleaning/sterilizing connections: 3 connections threads ¼" NPT for steam, rinse (with cleaning liquid) and drain Blocking medium connections: 2 connection threads ¼" NPT for blocking medium (normally alcohol) Operating: Dead man lever for manual operation Pneumatic actuator for automatic operation Weights: Weight table is only approximately, because of different connectionand adapter versions 4.1.3 Description of the connections The 3 connection threads ¼ NPT on the bayonet sleeve are: 1. Steam inlet connection to sterilize the sample passage in the bayonet sleeve and the bottle adapter in pos. 1 2. Rinse connection to clean the sample passage in the bayonet sleeve and the bottle adapter in pos. 1 3. Drain connection to let the steam or cleaning liquid flow out before taking a sample The 2 connection threads ¼ NPT on the upper part of the valve body are: 4 + 5: Connections for blocking media (normally alcohol), one of them is used to fill in the alcohol, the other can be used to take a sample of the alcohol. By taking an alcohol sample at this point the state of the blocking media and the stem seal can be monitored 4.1.4 Description of the bayonet positions (see fig. 5) Weight table Weight flanged version DN NPS kg 25 1 6.5 40 1½ 8.7 50 2 10.6 65 2½ 16.0 80 3 18.0 100 4 20.0 Fig. 2 4.1.1 Aseptic sampling valve (see fig. 2) 1. Sapro-aseptic sampling valve body flanged version, I-Clamp or Tri-Clamp (DIN/ANSI) 2. Dead man lever 3. Locking pin (can be completed with a padlock, not included), fixed with a SS chain 4. Set screw for medium flow regulation 5. Hexagon nut to fix the regulation set screw 6. Bayonet sleeve for the aseptic bottle adapter 7. Connections 3 x ¼ NPT for steam/rinse/ drain 8. Connections 2 x ¼ NPT for blocking medium Fig. 3 4.1.2 Aseptic bottle adapter and dummy adapter (see fig. 3 and 4) 1. Sapro-aseptic bottle adapter with thread ISO GL45 2. Sapro-aseptic dummy adapter (used when not taking a sample) 3. Sampling bottle 100 ml to 2000 ml ISO GL45 (standard version is 250 ml) 4. Vent, closed with a cross-cut silicone septum 5. Grips aluminium black anodized Fig. 4 Pos. 2 When bottle adapter is connected to pos. 2 the valve on the adapter is open. Fig. 5 Pos. 1 In this position steam or cleaning liquid can be put trough the product passage and exit trough the drain connection on the bayonet sleeve. 3

5 Marks on the valve (see fig. 6) The CE-mark on the valve is only needed, if the product is according to PED/DGRL 97/23/EG. 6 Installation 6.1 General 6.1.1 Preparation for the installation The Sapro-aseptic sampling valve is designed for installation between flanged tubes. The standard clamp is an I-clamp with a standard (DIN or ANSI) face-to-face length. Please check the correct flange version before installing the valve. The dimensions of the valve are in the catalog otherwise ask the manufacturer The installation flanges need: Cleaned and undamaged connection flanges. To have the same size and standard as the connection flanges on the sampling valve have. Please check this out before any installation. 6.1.2 Installation position of the Sapro-aseptic sampling valve (see fig. 7) The Sapro-aseptic sampling valve can only be installed in horizontal tubes, where the sampling bottle stands vertically. There is no possibility to install the valve in a vertical tube. 6.2 Installation between the flanged tubes The flow direction of the sampling medium is not relevant for this aseptic sampling valve, but please check out if any special flange seals (gaskets) are needed. A sampling valve should not be used to spread the connection tubes, this could damage the connection flange surfaces. 6.2.1 Installation step by step (see fig. 8) 1. Check the correct size and type of the connection flanges on the sampling valve. Also check if the length of the valve body is correct for the installation. 2. Spread the connection tubes with a special tool (do not use the sampling valve itself to do this operation). 3. If needed, attach the connection gaskets to the connection flanges. 4. Turn the valve to the correct position and tighten the clamps (I-clamp or Tri-clamp) on the connection flanges. The valve is now ready to install the additional connections. 5. On the bayonet sleeve (6), there are 3 connections (¼ NPT). Connect the steam connection to an external steamer (if possible) and the rinse connection to a cleaning medium connection. The drain connection needs normally no external connection. 6. On the upper side of the body, there are 2 connections (¼ NPT). Connect one of them to enter with the blocking media (normally alcohol), the other connection is to take samples of the alcohol (to check the state of the blocking media and the stem seal can also be monitored). Please do not apply any welding procedures when the sampling valve is already installed! The inner parts like the soft seat or the packing sleeve could be damaged! Always tighten the clamps well before testing the sampling valve, even in low-pressure lines, never apply pressure to the valve without having all clamps tightened! Manufacturer Serial number Type of valve according order Fig. 6 Land of origin SV050LN-162S2B Year of production CE-mark with 4 digit notified body number (only for DN 40-100, NPS 1½ - 4) Fig. 7 Aseptic valve body Connection flanges Connection tube (customer-side) Fig. 8 Bayonet sleeve 4

6.2.2 Final testing (see fig. 9) Before you use the aseptic sampling valve for the first time, please rinse the system with cleaning liquid or water to prevent the soft seat of the valve of damage. Please also rinse and clean the bottle adapter and dummy before taking the first sample. The cleaning can be done by taking the padlock away (if there is one), taking the locking pin (3) out and operate the dead man lever (2) for a few seconds. The bottle adapter can be set in position 2 for cleaning. After cleaning, please put the locking pin in again. 7 Starting 7.1 General Before starting, all specifications like material resistance/pressure and temperature of the production process need to be checked again. Any pollution (for example from welding processes) could cause problems on the valve. 7.2 Flow regulation (see fig. 10) Attach the bottle adapter with the bottle to the bayonet sleeve of the valve. By operating the dead man lever, the bottle can be filled. To control the amount of medium flowing through the adapter, screw the set screw (4) in (less medium) or out (more medium). When having found the right position of the set screw, it can be locked with the hexagon nut (5). 8 Danger potential during installation, operation and maintenance Safe operation of the Sapro-aseptic sampling valve is only guaranteed, when installed, operated and repaired by qualified personnel (see Qualified personnel ). All safety instructions and standards need to be fulfilled during installation, operation, maintenance and repair. 9 Operation (see fig. 11) 1. To take an aseptic sample, the bottle should be attached to the bottle adapter in the laboratory, where it can also be filled up with nitrogen. 2. Take the dummy away from the bayonet sleeve and attach the bottle adapter (with the bottle connected) to the aseptic sampling valve, turn to position 1 (see point 4.1.4: Description of the bayonet positions). 3. The inner parts of the bottle adapter and the bayonet sleeve can now be sterilized with steam or cleaned with cleaning liquid trough the two connections on the bayonet sleeve. Let the liquid flow trough the drain connection. Hexagon wrench, size 6 Fig. 9 Fig. 10 Dummy adapter Sampling valve body Aseptic bottle adapter Fig. 11 Bayonet sleeve Sampling bottle 5

4. After cleaning, turn the bottle adapter to position 2, the bottle adapter now opens the sampling bottle. 5. Take the locking pin out and operate the dead man lever to the maximum opening position (see point 7.2: Flow regulation). Now, the stem is lifted up and the sampling medium flows trough the bottle adapter into the sampling bottle. The air (or nitrogen) in the bottle is blown out trough the silicone septum on the side of the bottle adapter. Control the filling of the bottle. Don t overfill the bottle! 6. As the bottle contains enough sampling medium, let the dead man lever go, it closes the valve automatically. 7. Turn the bottle adapter back to position 1, the bottle is now closed again, do the same cleaning process as in point 3. 8. Take the bottle adapter away from the bayonet sleeve and attach the dummy adapter again. 9. The aseptic sampling process is now finished, the sampled medium can be taken to the laboratory to be analyzed. 1. To clean the bayonet sleeve and/or dismantle the soft seat and valve seat, simply open the four screws on the bottom side which connect the bayonet sleeve to the sampling valve body. 2. All parts can be cleaned now, use normal cleaning liquids. If the soft seat looks really bad, it has to be exchanged, ask the manufacturer for spare parts. 3. Install all parts again, tighten the screws well! If any problems on the upper part of the valve arise (dead man lever/problems with the stem seal), please contact the manufacturer. 11 Disposal The valve needs to be cleaned before any disposal! Incorrect cleaned sampling valves can cause injuries on hands or other parts of the human body, depending on the sampled medium. The manufacturer would not guarantee the safety of the valve when it is sold to other plants or factories after having it in sampling process. 10 Maintenance and repair Though almost all parts of the Sapro-sampling valve are made of SS, only few maintenance or repair is needed. If any problems with crystallizing medium arise, it may be necessary to clean the soft seat and/or the bayonet sleeve of the sampling valve. This operation should only be done by qualified personnel and only when there is no medium or pressure in the tube anymore! Prior to any maintenance or dismantling operation, the sampling valve and the tube needs to be empty and without any pressure! Depending on the medium flowing trough the valve, the system has to be rinsed with water or cleaning liquid to protect the operator. 2017 Emerson. All rights reserved. 6