KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

Similar documents
KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

SOP for Karl Suss MJB3 #1 Mask Aligner

MJB-3 Mask Aligner Property of TAU MNCF August 2012

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Karl Suss MJB4 Mask Aligner

COBILT CA-800 Mask Aligner Equipment Operation

Standard Operating Manual

Karl Suss MA6 Mask Aligner SOP

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Warnings: Notes: Revised: January 8,

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Standard Operating Manual

Warnings: Notes: Revised: October 5, 2015

University of Minnesota Nano Fabrication Center

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility

University of Minnesota Nano Fabrication Center

Quintel Mask Aligner

Approved by Principal Investigator Date: Approved by Super User: Date:

OAI Model 200 Tabletop Mask Aligner Portland State University

Operation of the mask aligner MJB-55

Approved by Principal Investigator Date: Approved by Super User: Date:

Operation of the contact mask aligner Canon PPC 210 Projection Print Camera

Karl Suss Contact Aligner Operation

Photolithography. Operating Instructions

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

icreasepro Creaser Operators Manual

7130 Lancer Rear Drive Magnetic Commercial Indoor Cycling Bike

Usage Policies Notebook for Parylene Coating System

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

OPERATION OF THE DIMPLER

Standard Operating Manual

NORDSON MARCH PX-1000 PLASMA ASHER STANDARD OPERATING PROCEDURE Version: 1.0 July 2016

Cryo-Evaporator Operation

SPUTTER STATION STANDARD OPERATING PROCEDURE

Standard Operating Manual

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

Standard Operating Procedure. For. PVD E-Beam

Model 130M Pneumatic Controller

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

THERMAL EVAPORATION UNIT (for Al evaporation)

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

Revised: June 7, 2017

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

Unaxis PECVD. SiH4 (5% in He)

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

Standard Operating Procedure #COE-SOP-0001 Chemical Fume Hood Operation

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

The SPI Sputter Coater Handbook

Standard Operating Manual

Instruction Manual Otoflash G171 UV-flash-device for Light Curing

Squeegee Unit for MRS Multi-Recovery System

GV Standard X-Vent. Setup, Commissioning & Installation Guide

Approved by Principal Investigator Date: Approved by Super User: Date:

March CS-1701F Reactive Ion Etcher

NRF Suss Delta 80 Spinner SOP Revision /14/2016 Page 1 of 11. Suss Delta 80 Spinner SOP

MoveRoll Conveyor Operating and Maintenance Manual

MAGNETIC CYCLING TRAINER SF-B0419 USER MANUAL

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

[needs to be discussed with faculty if any new material to be done]

WARNING NOTES TABLE OF CONTENTS

EE 432 Lab 3 PMOS source/drain lithography and diffusion

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

COMELEC C-30-S Parylene Coating System

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

Usage Policies Notebook for Quintel 2001 Mask Aligner

VAPOMATIC ANESTHETIC VAPORIZER

Safety Tips for Using Germicidal Lamps

Product Description TUV PL-L

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

602 STRINGING MACHINE OWNER'S MANUAL

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Operating Instructions and Parts Manual 24529

Version 1.0. icrease pro + Serial Number. Date

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

O P E R A T O R S M A N U A O P E R A T O R S M A N U A L L

1. Features CONTENTS. English

LSX Low Smoke Convertor Operations Manual

Spectrum AquaLight (SAL ) Color Underwater Light Owner s Manual

User Manual GRI- 1500Li

RIPPLE TANK - with rippler & kit

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Side-of-Pole Mount for 1 Module (SPM1) For Module Types A & B

Standard Operating Manual

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

1)! DO NOT PROCEED BEYOND THIS MARK

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

M3-LED. Operator s Manual. Operator s Manual MADE IN THE USA USA

600 / 600FC OWNER'S MANUAL

INSTRUCTION MANUAL. January 23, 2003, Revision 0

Gas Check. Short testing time, with automatic start-up if no leakage was detected from previous operation.

O P E R ATING INSTRUCTIONS FOR MODEL SPR-45 Automatic Screen and Stencil Printer

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

WELCOME TO THE REVOLUTION

INSTRUCTION MANUAL TIPPING BUCKET RAIN GAUGE. MODEL TB3/0.1mm

QUARTZ HOUSING COATING PROCEDURE

OEM Manual. MODEL ½ Digit DRUM SCALE

DIP STATION WITH SAFETY CONNECTOR SF-BH6507 USER MANUAL

Transcription:

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure Version: 1.0 February 2014 UNIVERSITY OF TEXAS AT ARLINGTON Nanotechnology Research Center (NRC) 1

TABLE OF CONTENTS 1 Introduction 3 1.1 Scope of Work...3 1.2 Description..3 1.3 Safety Precautions... 4 2 Requirements.......5 2.1 Training... 5 2.2 Hardware,...5 2.3 Restrictions....... 5 3 Operating Procedure.........5 3.1 Lamp Powering Up Procedure.......5 3.2 Mask and Wafer Loading..8 3.3 Alignment and Setting Exposure Time.. 11 3.4 Shutdown Procedure...12 2

1 INTRODUCTION 1.1 Scope These procedures apply to the KARL SUSS MJB3 UV400 mask aligner. All maintenance should follow the procedures set forth in the manufacturer s maintenance and operations manuals. This document is for reference only. Personnel should be trained by authorized staff before operating MJB3 UV400 mask aligner. 1.2 Description The SUSS MJB3 UV400 mask aligner is designed for high resolution photolithography in a laboratory environment. The MJB3 mask aligner offers exceptional flexibility in the processing of irregular shaped samples of different thickness such as sample wafer pieces and glass slides up to standard size 2 diameter wafers. The SUSS MJB3 UV400 is equipped with 350W mercury short-arc lamp providing primary exposure wavelengths of 350-450nm. The machines equipped with a high precision stage allowing alignment accuracies to 0.25 microns. The resolution of the MJB3 UV400 is 1-2 microns under optimum conditions. 1.3 Safety Precautions 1.3.1 This equipment uses UV LIGHT SOURCE. Be very careful and aware of hazards that stem from the UV light source. The light emanating from the light source is a hazard to vision. DO NOT LOOK INTO THE OUTPUT BEAM of the UV Lamp or reflection of the beam. Permanent damage to the retina of the eye and subsequent blindness can result. If you encounter any hazards, contact NanoFAB staff immediately. 1.3.2 Prolonged exposure to diffused reflection from the output beam, illuminated surfaces in the beam, exposures of even a few seconds to the direct output beam or to the lamp itself, can cause ultraviolet skin burns or burns to the outer layers of the eye. 1.3.3 High voltage is used to run the UV lamp; do not attempt to disassemble supply 3

1.3.4 If the main lamp housing needs to be opened for service such and a stuck shutter the system that the light source is mounted on and the UV power supply should be powered down and locked out. Then the lamp N 2 cooling should be left on for 30 minutes to properly cool down the inside of the lamp housing before opening it for service. Call Nanofab staff for any maintenance issues. 1.3.5 This equipment uses a UV light source that is subjected to 20 to 30 atmospheres of internal pressure so there is always the possibility of a lamp exploding due to internal strains or physical abuse.. The lamp contains Mercury. Any vapor released by an exploded lamp would be highly dangerous. In the event of a lamp explosion, clear the area of personnel and contact the NanoFAB staff immediately. http://www.cdc.gov/niosh/npg/npgd0384.html http://www.sciencelab.com/msds.php?msdsid=9927224 1.3.6 Before changing the lamp, working on the light source, or performing any system maintenance requiring access to AC electrical power, Staff maintenance needs to turn OFF the power at the FACILITY AC POWER circuit breaker. After the power has been turned off, verify that the output voltage is zero. 1.3.7 Do not operate the system without proper room ventilation as some Deep UV lamps produce significant levels of OZONE. http://www.cdc.gov/niosh/topics/ozone/ http://www.ozoneapplications.com/info/ozone%20solutions%20msds%20ozone.pdf 1.3.8 Read any posted NanoFAB Engineering Change Notices (ECN) for any hardware, process or safety changes before running the tool. 4

2 REQUIREMENTS 2.1 Training You must be a qualified user on the SUSS MJB3 UV400 ALIGNER. 2.2 Hardware 2.2.1 High resolution contact exposure mode printing down to 1-2 um by using primary wavelengths of 350 to 450nm from a 350W mercury short-arc lamp. 2.2.2 High precision X, Y, and θ alignment stage and microscope manipulator. 2.2.3 Wafer, glass slides and substrate handling up to dia.50 mm (wafers), 1 x1 (substrates) 2.2.4 Special substrate chucks for pieces, III-V materials, thick substrates, hybrids. 2.2.5 Economical tool for laboratories and small series production. 2.3 Restrictions 2.3.1 Glass slides, small sample pieces up to 2 diameter wafer sizes can be aligned and exposed. Small sample pieces need to be fixed to a 2 carrier wafer. 2.3.2 Only 4 x 4 x (0.065-0.1 ) and 3 x 3 x (0.065-0.1 ) masks can be used. 2.3.3 MJB3 Aligner is available 7 days a week / 24 hours a day. 3 OPERATING PROCEDURE 3.1 Lamp Powering Up Procedure/ Mask and Wafer Loading/ Alignment and Exposure 3.1.1 The Hg-Arc lamp is peaked and calibrated to 3.0 mw/cm 2 with the UV probe. 3.1.2 Make sure the power to the aligner is OFF by checking the red Power button light is OFF. Turning ON the Illumination Controller when power to the aligner is ON, will damage the aligner. 5

3.1.3 Turn ON the lamp N 2 cooling switch to the up position. The N2 pressure should be 1.5 barr. 3.1.4 Turn on the green POWER switch on the Illumination Controller under the table by moving the lever to the up position then press the green Lamp button ( for nor more than 1 sec). The lamp should ignite. Press green channel button to toggle between sensor 1 and 2. Sensor 1 is 365nm sensor. Wait at least 10 minutes for the lamp to warm up, the display should read 275 W ( standby idle power on the lamp). 3.1.5 Turn the Aligner air pressure switch to the up position. The air pressure should be 2.0 barr. Check the mask vacuum switch is in the UP position (ON), if not move this switch to the up position. 3.1.6 Turn on the power to the aligner by pressing the red power button. 6

3.1.7 If you hear an alarm ( buzzer) sounding press the Nitrogen Loss button to silence the alarm (it was previously pressed by mistake) 3.1.8 Move the Contact Lever (1 st ) then the Separation Lever (2 nd ) to the separation position to before loading the mask or substrate. Contact Lever (1 st ) Separation Lever (2 nd ) 3.1.9 Loosen the Mask Holder clamping knobs by rotating both a few turns CCW, then slide the Mask holder OUT and place upside down on the table. 7

3.1.10 Carefully place your clean mask with the chrome side facing up (glass side in contact with the mask holder) on the mask holder. Make sure the mask is centered and its sides are perpendicular with the mask holder edges. Check to see if the masks die orientation is the correct orientation you need. Then press the Mask Vacuum button to turn ON mask vacuum. Check to see if the mask is holding with your wafer tweezers by gently tapping the mask edge. If you find a mask vacuum problem call staff to check ( to avoid a broken mask) 3.1.11 Carefully slide the mask + Mask Holder back into the mask guide rail and tightened the clamping knobs by turning them CW until they are finger tight. 3.1.12 Gently slide OUT the wafer holder tray and place you coated substrate on the chuck. Slowly slide the wafer holder tray and wafer all the way IN. 8

3.1.13 Move Separation Lever (1 st ) then the Contact Lever (2 nd ) to the Contact positions to remove any gap or wedge between the wafer and mask. The green contact light will go ON Separation Lever (1 st ) Contact Lever (2 nd t ) Green Contact light is ON 3.1.14 To aligne the substrate to the mask move the separation lever to the Separation position creating a gap. The green contact light will go OFF. 3.1.15 Turn on the Halogen lamp power supply to the Microscope illuminator to about 4 volts (middle of the scale). ON: Rotate dial CW direction to about 4 volts 3.1.16 Adjust the Microscope eyepiece for binocular vision, adjust the microscope manipulator for your field of view and use the focusing knobs to focus on the mask and die alignment markers. 9

3.1.17 Align your wafers alignment marks to the mask alignment marks by using X, Y, θ micrometers. 3.1.18 After the alignment is done move the Separation Lever back to the Contact position The green contact light should go ON. Contact Lever Green Contact light is ON 3.1.19 The substrate is now ready for exposure. Set the exposure time on the timer located at the right of the front panel. Your exposure time depends on the exposure Dose (mj /cm2) for your resist type and thickness and Lamp intensity (Mw/cm2). The exposure time is Dose Intensity. 3.1.20 To set the exposure time 1 st select the time units multiple (seconds, minutes, hours) by rotating the inner black knob to align with the black arrow to the time units multiple you need. Align with the black arrow to the time units you need ( ie. X 10s ) 10

3.1.21 Next rotate the Outer plastic dial the time base ( 0 to 3 ) you need. ( i.e. 2 X 10s) = 20 second exposure 3.1.22 The exposure time in now set. Press the blue Exposure button to expose the wafer. Turn away from the UV light, do not look at the UV light or you will damage your retinas permanently. 3.1.23 After the exposure is done move the Separation Lever (1 st ) then the Contact Lever (2 nd ) to the Seperation positions. The green contact light will go OFF. 3.1.24 Gently slide the wafer transport tray out and remove your wafer for further processing. 11

3.1.25 Loosen the Mask Holder clamping knobs by rotating both a few turns CCW, then slide the Mask holder OUT and place upside down on the table. Press the mask vacuum button to turn OFF mask vacuum and remove your mask. 3.1.26 After your mask is removed replace the mask holder and tightened the clamping knobs finger tight. 3.1.27 Turn OFF the microscope halogen illuminator power supply. 3.1.28 Turn OFF the Hg-Arc lamp power supply. 3.1.29 Turn OFF the MJB3 system power on the front panel. 12

3.1.30 Turn OFF the MJB3 air pressure by moving the switch to the down position. 3.1.31 Leave the mercury lamp N 2 cooling ON for 30 minutes to properly cool the lamp 3.1.32 After 30 minutes turn OFF the N2 cooling by moving the switch to the down position. 3.1.33 Enter the required information in the logbook. 13

14