Introduction. Materials and Methods

Similar documents
TG-51: Experience from 150 institutions, common errors, and helpful hints

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

Abstract: Introduction:

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

Relative Dosimetry. Photons

Review of fundamental photon dosimetry quantities

Clinical Implementation of Volumetric Modulated Arc Therapy

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

Math Review. Overview

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Monitor Unit Verification for Small Fields

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

1. Question Answer cgy / MU cgy / MU 2. Question Answer

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Certification of Class II Prescribed Equipment

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

Certification of Class II Prescribed Equipment

Exposure System Selection

Dynamic wedges dosimetry and quality control

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

THE development of more advanced techniques in radiotherapy,

A beam-matching concept for medical linear accelerators

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

Constancy checks of well-type ionization chambers with external-beam radiation units

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

Proton Therapy QA Tools

SnapShot IMRT with compensators and FFF beams

COMPARISON BETWEEN PHYSICAL WEDGES AND VIRTUAL WEDGES IN LINAC

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

CHAPTER-2. Wedge filters are the most important beam modifiers used in

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

OBJECTIVE 6: FIELD RADIOLOGICAL MONITORING - AMBIENT RADIATION MONITORING

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at

Scientific Journal of Nuclear Medicine & Radiation Therapy

Radiotherapy physics & Equipments

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

Variable Depth Bragg Peak Test Method

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Planning Assignment (3 field rectum)

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

Dosimetric Calculations. Lonny Trestrail

Zoubir Ouhib Lynn Cancer Institute

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Surface buildup dose dependence on photon field delivery technique for IMRT

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

Commissioning and periodic tests of the Esteya electronic brachytherapy system

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

User Manual. RW3 Slab Phantom T29672 and T

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

Varian HDR surface applicators commissioning and clinical implementation

Let it Roll: The Soup Can Experiment

TEST REPORT BEAMSCAN WITH HALCYON

1. BEAM MODIFIERS IN RADIOTHERAPY

Evaluation of six TPS algorithms in computing entrance and exit doses

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

IMRT in clinical practice at the UMC-Utrecht. Utrecht

Total Ionization Dose (TID) Test Results of the RH117H Positive Adjustable Low Dose Rate (LDR) LDR = 10 mrads(si)/s

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

MIT Analytical X-Ray Safety Policies and Procedures

IAEA-TECDOC-1274 Calibration of photon and beta ray sources used in brachytherapy

3D Treatment Planning and verification with hand calculations

Quality control of breast ultrasound: How to use a breast ultrasound quality assurance phantom?

The dose distribution of medium energy electron boosts to the laryngectomy stoma

Study of Mixed Gas Flow Pattern Inside RPC

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Cryo-Evaporator Operation

High Speed Direct SAD Radiosurgery Beam Scanner

Disclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant

Perturbation Correction of a Cylindrical Thimble-type Chamber. in a Graphite Phantom for 60Co Gamma Rays*

Jefferson Lab Bubble Chamber Experiment Update and Future Plans. C(a,g) 16 O. Claudio Ugalde

DESIGN VERIFICATION AND QUALIFICATION OF NEW POPPET AND BUSHING ON GMA FLIGHT SOLENOID VALVES

Iranian Journal of Medical Physics

Total Ionization Dose (TID) Test Results of the RH1028MW Ultralow Noise Precision High Speed Operational Low Dose Rate (LDR)

CALIBRATION PRODUCTS & PHANTOMS

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

REPORT. A comparative study of the mechanical and biomechanical behaviour of natural turf and hybrid turf for the practise of sports

In air calibration of an HDR 192 Ir brachytherapy source using therapy ion chambers

Assessment of correlations between NDE parameters and tube structural integrity for PWSCC at U-bends

Analysis of RPC Performance with Different Gas Mixture

Transcription:

STANDARD WEDGE AND TRAY TRANSMISSION VALUES FOR VARIAN, SEIMENS, ELEKTA/PHILIPS ACCELERATORS; A QUALITY ASSURANCE TOOL David S. Followill, Nadia Hernandez, and William F. Hanson Introduction The Radiological Physics Center (RPC), through it s on-site dosimetry review visits to institutions participating in NCI cooperative clinical trials, has accumulated wedge and tray transmission data for approximately 5 wedges on nearly 11 Varian, Siemens and Elekta/Philips accelerators since 1985 and nearly 35 tray transmission measurements since 1995. The majority of the wedges used for specific make and models of accelerator are manufactured by the vendor and are similar in design. From time to time, however, the vendor will change its wedge design, which in turn affects the wedge transmission value. The vast majority of the blocking trays, regardless of make and model of accelerator, can be classified as thin or thick trays whose transmission depends solely on the photon beam energy. We believe we can predict the wedge and tray transmission value for a particular nominal energy, wedge angle, tray thickness and make and model of accelerator. The data presented here can be used to predict the wedge and tray transmission for most wedges and trays in clinical use today to within 2%. These data also serve as a redundant QA tool that may identify potential problems. Materials and Methods Measurements were made on a series of Varian, Siemens and Elekta/Philips accelerators. The make and models of those listed here reflect those in current use and most often seen by the RPC. 1. Varian Clinac 4/8, 4/1, 6/1, 6, 6C, 18, 18, 2, 21C, 23C, and 25. 2. Siemens Mevatron 6, 12, 2, 6, 63, 64, 67, 674, 74, 77, KD series, MD series, and MX series. 3. Elekta/Philips SL 75-5, 18, 2 and 25. Nominal energies from 4-25 MV, wedge angles from 15 6 and thin and thick trays. The RPC tray transmission factor (TF) is defined as: TF (1x1, 5* cm) = tray ioniz. (1x1, 5 cm) open ioniz. (1x1, 5 cm) The RPC wedge transmission factor (WTF) is defined as: WTF(1x1, 5*cm) = wedged ioniz. (1x1, 5cm) open ioniz. (1x1, 5cm) At 1 cm SSD/SAD in a water phantom. The WTF s ionization readings are averaged over heel-in and heel-out wedge orientations (if applicable) and centering of ion chamber is assured by measurements in multiple collimator orientations. *For energies>15 MV, depth is 7 cm and after 1/2 for all energies, depth is 1 cm.

Standard Wedge and Tray Transmission Values for Varian, Seimens, etc... Page 2 Results The wedge transmission data for the manufacturer s standard wedges on most makes and models of accelerators exhibit a gaussian distribution with a standard deviation of ± 2%. The distribution of the wedge transmission factor data is often bimodal exhibiting more than one standard factor for each wedge angle. Some of the wedge transmission data for differing makes and models of accelerators for a specific manufacturer, dependent on energy and wedge angle, are in good agreement. The tray transmission data, is not dependent on make and model of accelerator, but rather the beam energy. There appear to be two types of trays in clinical use today; thin and thick trays. Conclusions 1. For each combination of nominal wedge angle, energy, and accelerator make and model, a standard wedge transmission factor can be assigned. 2. Frequently, there is more than one standard wedge transmission factor for a given make and model of accelerator. (The RPC has not captured data to determine whether the different wedges represent different sizes, composition or location of wedges). 3. The same or similar wedges are used on different makes and models of accelerators by the same manufacturer. 4. There are two types of blocking trays in clinical use today, thin and thick trays. 5. These mean standard wedge and tray transmission values provide a good redundant check of an institution s own measured values. 6. With few exceptions, these data can be used to predict the wedge and tray transmission factors for most wedges and trays in clinical use today to within ± 2%.

Standard Wedge and Tray Transmission Values for Varian, Seimens, etc... Page 3 Table I: Standard Values for Varian Accelerators 15 deg wedge 3 deg wedge 45 deg wedge 6 Machine Model Energy Ion Ratio Std Dev N WTF Std Dev N WTF Std Dev N WTF Std Dev N WTF Cl 4 4 MV.619.9 9.795.6 21.755.9 77.578.12 89.489 Cl 4/1 4 MV.631.6 31.776.8 7.736.4 6.495.11 1.491.85.6 16.764.6 16.589.5 16 Cl 6C 4 MV.632.5 8.769.11 4 Insufficent Data.477.4 5.394 Cl 6, Cl 6X 6 MV.664.12 38.78.8 3.627.3 5.617.8 9.425.827.3 8.777.9 17.651.8 11.459.873.4 6 Cl 6/1 6 MV.661.6 14.78.2 9.643.15 27.481.1 32.418.823.6 8.74.5 4.569.6 6.455.783.7 78.614.8 98 Cl 6C 6 MV.665.5 66.78.2 8.545.3 8.499.7 62.418.78.2 27.631.4 37 Cl 18 1 MV.733.5 63.892.7 45.752.9 8.696.7 2.511.83.5 5 Cl-18 6 MV.687.24 89.788.13 49.638.6 52.485.6 72.419.76.9 9.572.8 13 Cl-18 1 MV.739.3 16.814.4 6.681.2 6.536.6 1.451.85.4 5.751.5 4.626.7 6 Cl-18 15 MV.76.2 18.824.3 12.697.3 12.556.3 15.449 Cl-18 18 MV.784.2 52.83.4 25.76.4 28.514.1 7.443.57.4 4.652.7 5. Cl-2 15 MV.758.7 2.854.4 1.758.4 11.642.6 17.452 Cl-2 18 MV.78.4 14.859.2 1.767.3 1.65.3 13.442 Cl-21, 6 MV.674.4 18.711.4 14.547.3 6.493.9 178.414-21 C, 21 CD.783.7 53.632.8 94.57.6 24 Cl-21 1 MV.735.3 38.751.2 3.62.2 3.53.6 34.448-21 C, 21 CD.813.2 16.68.3 16 Cl-21 15 MV.761.2 28.82.8 18.692.8 23.527.4 23.445-21 C, 21 CD.554.3 5 Cl-21 18 MV.782.9 11.768.3 12.629.2 15.521.5 88.439-21 C, 21 CD.827.6 42.73.5 54.566.3 19 Cl-23 6 MV.674.4 35.773.7 12.625.8 15.493.12 33.48 l-23c, 23 CD Cl-23 15 MV.762.3 1.813.5 6.684.5 6.531.14 9 Ins l-23c, 23 CD Cl-23 18 MV.783.2 15 Insufficent Data.7.5 6.519.3 8.43 l-23c, 23 CD Cl-25, 25C 6 MV.675.5 23.783.4 11.635.5 12.489.6 13.419.821.3 6.79.5 7.577.5 8 Cl-25, 25C 24 MV.85.3 22.832.5 11.71.4 11.58.4 5.428.861.4 6.77.6 7.578.3 7.658.6 9

Standard Wedge and Tray Transmission Values for Varian, Seimens, etc... Page 4 Table 2: Standard Values for Siemens Accelerators 15 deg wedge 3 deg wedge 45 deg wedge 6 Machine Model Energy Ion Ratio Std Dev N WTF Std Dev N WTF Std Dev N WTF Std Dev N WTF Mev 12 8 MV.716.8 6 Insufficent Data Insufficent Data.694.7 5 Ins Mev 12 1 MV.736.8 13 Insufficent Data.719.4 9.696.8 11.685 Mev 2 15 MV.767.5 5 Insufficent Data Insufficent Data.64.6 5.611 Mev 6 6 MV.672.9 13 Insufficent Data.684.9 12.636.8 3 Ins Mev 6, 63, 64 4 MV.611.16 9.75.3 6.638.7 3.513.6 6.481 Mev 67, 674 6 MV.675.6 43.685.4 5.523.5 5.317.7 11.35.742.5 2.714.5 25.563.5 31.535 Mev 74, 77 1 MV.742.6 33.782.2 19.757.3 25.388.7 5.592.62.3 28 Mev 77 15 MV.766.3 12.794.2 1.77.3 11.637.5 12.68 Mev 77 18 MV.785.4 7.82.3 5.777.2 5.647.3 6.619 Mev KD, KD2, KDS 6 MV.682.7 57.683.4 13.518.7 21.316.9 35.342.745.3 13.716.4 17.565.5 21.538 Mev MD, MD2 6 MV.674.4 39.684.5 11.518.6 9.313.5 17.339.74.4 7.713.5 6.56.3 6.533 Mev MX, MX2, MXE 6 MV.678.5 22 Insufficent Data Insufficent Data.317.7 5.346.338.6 5 Table 3: Standard Values for Elekta/Philips Accelerators 6 deg wedge Machine Model Energy Ion Ratio Std Dev N WTF Std Dev N SL 18 6 MV.761.8 7.269.5 7 SL18 15 MV.761.8 7.283.4 7 SL 2 6 MV.68.8 19.271.7 13 SL 2 18 MV.776.4 12.279.7 11 SL 25 6 MV.681.1 15.27.4 14 SL 25 25 MV.8.3 12.268.5 13 SL 75-5 6 MV.68.3 15.261.1 14 Standard Average Mevatron and Philips Thin Tray Factors.98.975 Ave. Thin Tray Factor.97.965.96 y =.1517x +.855 R 2 =.954.955.95.66.68.7.72.74.76.78.8 Ionization Ratio

Standard Wedge and Tray Transmission Values for Varian, Seimens, etc... Page 5 Standard Average Clinac Thin Tray Factors.99.985 Ave. Thin Tray Factor.98.975.97.965 y =.115x +.8931 R 2 =.8729.96.955.6.65.7.75.8.85 Ionization Ratio 7 6 5 4 3 2 Varian Clinac 6/1, 6C, 18 21C, 23C, 25 6 MV 6 degree wedge.416.7 336 1.385.395.45.415.425.435.445.455.465.475

Standard Wedge and Tray Transmission Values for Varian, Seimens, etc... Page 6 18 16 14 12 1 8 6 Siemens Mevatron KD series 3 degree wedge.518.7 21.716.4 17 4 2.5.52.54.56.58.6.66.68.7.72.74 6 5 4 3 2 Philips/Elekta SL 25 6 degree wedge.27.4 14 1.259.262.265.268.271.274.277.28

Standard Wedge and Tray Transmission Values for Varian, Seimens, etc... Page 7 6 5 4 3 2 Varian Clinac 6/1 45 degree wedge.481.1 32.569.6 6.614.8 98 1.46.48.5.52.54.56.58.6.62.64.66 7 6 5 4 3 2 Varian Clinac 21C 6 MV 6 degree wedge.414.6 147 1.39.395.4.45.41.415.42.425.43