MODEL GT820 OXYGEN SENSOR

Similar documents
Sensoric 4-20 ma Transmitter Board Operation Manual

HLM-2000-TX Gas Transmitter User Manual

Transmitter CS 21 Operation Manual

HLM-2000-TX Gas Transmitter User Manual

210 Series Transmitter with External Electrochemical Sensor

Model TX-FM/FN Sensor Head

O3 3E 1 F Gas Sensor Module

Columbus Instruments

Operating Instructions

Electrochemical Sensor

Electrochemical Sensor

Air Check Lite Toxic Gas Monitor

Operation and Maintenance Manual

Transmitter CS 21 Operation Manual

Dissolved Oxygen Guide

ISQB1 INTRINSICALLY SAFE ELECTRO-PNEUMATIC CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS HAZARDOUS AREA CLASSIFICATION

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

Technical Data Sheet MF010-O-LC

Best Practice Guide, Servomex 2700

ME3-HCL Electrochemical sensor

Electrochemical Carbon Monoxide Gas Sensor (Model:ME3-CO) Manual. Version 1.2 Valid from May 1 st 2014

CC Operation Manual Oak Valley Drive, Suite 20, Ann Arbor, Michigan USA fax

OSAT Series. Intrinsically Safe Infrared Temperature Sensor Operator's Guide

Model B Wire Wet Oxidant Gas Transmitter. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

EASIDEW TRANSMITTER with Current Source Output

All Sixth Sense cells for toxic gas detection are based on electrochemical principles and can be classified as amperometric fuel cells.

GAS FUEL VALVE FORM AGV5 OM 8-03

Electropneumatic Transducer (I/P) Installation, Operation and

Preferred Instruments Danbury, CT USA

12mm Dissolved Oxygen Sensor Care and Use Instructions

! Warning, refer to accompanying documents.

INSTRUCTION MANUAL Detcon Model DM-200

User Manual. PolyGard AT-1130 V3. Electrochemical Nitrogen Dioxide Transmitter. July, 2011

Design Features. General Description

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

Operation and Maintenance Manual

User Manual. PolyGard DT V4. Electrochemical Carbon Monoxide Transmitter Serial No. DT December, 2008

SDX Submersible Depth Transmitter User Manual

Nemoto Sensor Engineering Company Ltd

DPC-30 DPC-100. Reference Manual

FLAMMABLE GAS SENSOR/TRANSMITTER

Operation Manual. Rev. C LPT Low Power Transmitter (Electrochemical Sensor)

PURA. Pure Gas Dewpoint Transmitter. Users Guide

with O 2 Controller Instruction Manual

Intelligent SUNTEX DC-5310(RS) Dissolved Oxygen Transmitter

MANUAL KPS Pressure Control Valve

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

PyroCouple Series Operators Guide

Detcon MicroSafe Model DM-534C Oxygen Deficiency Sensor (0-25% O2)

OEM Manual. MODEL ½ Digit DRUM SCALE

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

MEC OEM. User Manual. Commercial in Confidence

Type N-10, N-11 Hazardous Area Non-incendive Transmitters

Nemoto Sensor Engineering Company Ltd

KING-GAGE. Calibration Instruction Manual. D/P LiquiSeal Purge Control and D/P Purge Control

Electro-Pneumatic Converter YT-940 SERIES

KJL-205A. Operation Manual YOU MUST READ THIS MANUAL BEFORE USE

GG-VL2-NH3 AMMONIA VENT LINE SENSOR. Installation and Operation Manual

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

Instruction Manual. Digital Thermocouple Gauge Control Units Models DTC & DTC Models DTC-06M-115 & DTC-06M-230

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

Gas density monitor With integrated transmitter Model GDM-100-TI

STERILIZABLE DISSOLVED OXYGEN SENSORS

Sample Draw Oxygen Monitoring System (OMS) User Manual

OxiQuant S. Operating Instructions

RAI-M series RUDDER ANGLE INDICATOR SYSTEM

QBS Electronic Pressure Regulator

FTC130 Transmitter. Operating Manual

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

DF-310E PROCESS ANALYSERS APPLICATIONS FEATURES

BI-680 Online Dissolved Oxygen Controller Instruction Manual

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

User s Guide Temperature Sensor Converter TSC-599

SITRANS P measuring instruments for pressure

High-performance submersible pressure transmitter For level measurement Model LH-10

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

CCS240 and CCS241. Technical Information. Sensors for chlorine dioxide Amperometric, membrane-covered sensors for installation in the CCA250 assembly

Model DM-100 DM-100 Toxic Gas Sensors DM-100 O2 Deficiency Sensors

SDX Submersible Depth Transmitter User Manual

User Manual. µ-gard MD Electrochemical Carbon Monoxide Transmitter Serial No. MD April, 2011

Pressure Measurement Single-range transmitters for general applications

OXYGEN MONITORING SYSTEM OMS

INSTALLATION & MAINTENANCE INSTRUCTIONS

Powers Controls RC 195 Multiple Input Receiver- Controller

Heavy Duty Dissolved Oxygen Meter

Oxymax W COS41. Technical Information. Potentiostatic amperometric two-electrode sensor Long-term stable sensor for water and wastewater applications

LK-SX VOC. Application. Security Advice Caution. Notes on Disposal. Combined sensor mixed gas. Data sheet

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI

I/P Transducer Positioner Module

an ISO9001 company TECHNICAL INFORMATION FOR SK-25F Technical Information for Figaro Oxygen Sensor SK-25F

OPERATING INSTRUCTIONS FOR LXT-280. Dissolved Oxygen Transmitter DANGER

1 Overview. Pressure Measurement Single-range transmitters for general applications. 1/22 Siemens FI US Edition

Dynisco SPX Industrial (2280/2281) Series

Instruction Manual BA90 Precision Barometer

MIL-STD-883G METHOD

JUMO dtrans p30 Pressure Transmitter

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

Specifications of Oxygen Sensor Module V-8344A

Transcription:

INSTRUCTION MANUAL MODEL GT820 OXYGEN SENSOR 70046 The information and technical data disclosed by this document may be used and disseminated only for the purposes and to the extent specifically authorized by SAFETY S YSTEMS TECHNOLOGY in writing. Such information and technical data are proprietary to SAFETY S YSTEMS T ECHNOLOGY and may not be used or disseminated except as provided in the foregoing sentence. S AFETY SYSTEMS TECHNOLOGY 23282 Mill Creek Drive, Suite 215 Laguna Hills, California 92653 USA Phone (949) 583-1857 Fax (949) 340-6643 http://www.safetysys.com

DESCRIPTION MODEL GT820 OXYGEN DEFICIENCY/ENRICHMENT GAS SENSOR MODEL GT820 OXYGEN DEFICIENCY/ENRICHMENT GAS SENSOR DESCRIPTION Model GT820 Oxygen Deficiency/Enrichment Gas Sensors measure the concentration of oxygen in a protected area and transmit this information to a central control point. The sensor includes a loop powered transmitter which converts the sensor output to a standard 4-20 ma signal. This signal may be connected to any device with a standard 4-20 ma input. PRINCIPLE OF OPERATION Figure 820-1 Mounting Dimensions The Oxygen Deficiency/Enrichment gas sensor uses an electrochemical fuel cell to sense the percentage of oxygen present in the atmosphere of the protected area. The sample enters the cell through a diffusion barrier and gas permeable membrane. A reaction takes place at the working electrode (anode), releasing electrons which flow to the counter electrode (cathode), where a counter reaction occurs. The number of electrons released in the chemical reaction is linearly dependent on the oxygen concentration. The gas diffusion barrier only allows a small quantity of sample atmosphere to reach the electrode. As a result, only a small fraction of the active electrolyte is reacted, leaving a large quantity in reserve for higher concentrations of oxygen. Cell output remains linear at concentrations of 30% and above. The currents generated by the oxidation reaction are quite low, typically 0.4 ua/%o2. However, the intrinsically low background current and low noise output of the cell result in an excellent sensitivity to the gas. DESIGN MANUAL December, 2005 820-1

MODEL GT820 OXYGEN DEFICIENCY/ENRICHMENT GAS SENSOR ELECTRICAL SPECIFICATIONS ELECTRICAL SPECIFICATIONS Power Supply required: 24 volts DC nominal. Will operate within specifications at any supply voltage between 16 and 32 volts. Sensor output: Integral transmitter will supply standard 4-20 ma signal into a load resistance of 700 ohms or lower. ENVIRONMENTAL SPECIFICATIONS Changes in humidity or direct exposure to moisture have little effect on the SST electrochemical cell. All reactions take place at the working electrode, where moisture is continually present. The electrolyte reservoir has sufficient overcapacity to allow the cell to accommodate all but the most prolonged periods of very high or very low humidity. The cell has been tested for three (3) months at a relative humidity of 0% without ill effects. In addition, once the RH was increased, the cell reabsorbed all lost moisture. LOCATION OF DETECTORS Location of the sensors is important. The sensor should be located roughly at eye level for a typical person in areas occupied by personnel. Do not locate sensors closer than one foot to floor to prevent damage from water, dust, etc. You must also consider air circulation, the pressure of leaking gases, and obstructions that could trap pockets of gas or air when locating sensors. Preferred orientation of sensor is with screen pointing down. If necessary, it may be installed at an angle or horizontally. The sensor must never be installed pointing upwards. INSTALLING AND WIRING THE DETECTORS A junction box with 3/4 inch hubs for the sensor and conduit connections, SST part number 850-2, is required for mounting and terminating the sensor. The Model GT820 uses a self-contained two wire transmitter. The operating power source for the transmitter electronics is obtained from the same 2 wires that are used to transmit the 4-20 ma output signal. Use two-conductor shielded cable, or two wires installed in metal conduit. These wires carry a maximum of 20 ma current. Wire size of 22 AWG is suitable for up to 6000 feet. Wire size of 0.35 mm 2 is suitable for distances up to 1800 meters. Typical connections are shown in Figure 820-2. Note that you can use the 4-20 ma indicator, the 1-5 volt indicator, or both. The total loop resistance of the circuit, including the field wiring to the sensor, must not exceed 700 ohms. 820-2 December, 2005 DESIGN MANUAL

DETECTOR ACCEPTANCE TESTS MODEL GT820 OXYGEN DEFICIENCY/ENRICHMENT GAS SENSOR Figure 820-2 Typical Connections to GT820 Sensor DETECTOR ACCEPTANCE TESTS The current loop output of the sensor is normally between 4 ma and 20 ma, and is a direct linear read-out of oxygen concentration. The output is 4 ma when 0% Oxygen is being detected and 20 ma with 25% Oxygen. Thus, the output loop will generate 17.37 ma at normal Oxygen levels of 20.9%. During normal operation, the loop current can be determined with the following formula: I (ma) = 4 + 0.64 x %Oxygen Should a malfunction occur in the sensor head, the output will be 4 ma (equivalent to 0% Oxygen). Faults of the transmitter or malfunction of the current loop wiring will result in 0 ma loop current (-6.25% Oxygen). Any current outside the 4-20 ma range indicates improper hookup. Check the indicator for proper reading; if not correct, perform the calibration procedures below. MAINTENANCE The linear nature of the SST sensor, along with its absolute zero and excellent repeatability, allow single point calibration at six-month intervals. This is primarily to check that the sensor screen has not become physically blocked by dust, paint or oil. SST sensors have been operating for up to three years with unchanged calibration. Calibration Gas Oxygen Deficiency/Enrichment Gas sensors require no special calibration gas. DESIGN MANUAL December, 2005 820-3

MODEL GT820 OXYGEN DEFICIENCY/ENRICHMENT GAS SENSOR MAINTENANCE Calibration Instructions Calibration requires access to the potentiometer on the transmitter PC board located inside the GT820 housing. Unscrew the front cover from the sensor housing to locate the pot. Some housings have a setscrew in the front of the housing which must be loosened before the front cover can be unscrewed.! If the sensor is installed in a hazardous area, be sure that no combustible gasses are present before the front cover is removed from the housing. Once the front cover is removed, you must identify the type of transmitter pc board that is used in your detector. Look for the identifying part number 1647 or 10256 on the pc board. Calibrating transmitter type 1647 A small potentiometer for routine span adjustment is located on the transmitter circuit board. For calibration purposes, the output signal may be monitored using either a standard milliammeter inserted into the 4-20 ma circuit, by connecting a millivolt meter to connector CN2 (which is across a 10 Ohm precision resistance on the circuit board), or by observing the reading on the associated control equipment.! If using CN2, accessory calibration cable, SST part no. 20185-1, is required. To adjust the potentiometer a small, non-metallic screwdriver is required. Perform the following procedure to calibrate the sensor/transmitter combination: 1. Unscrew the front of the sensor housing to gain access to the calibration potentiometer. 2. Ensure the sensor is in pure ambient air, free from any other gases. 3. Adjust the span potentiometer until a reading of 17.37 ma output current, or 173.7 mv across CN2, is obtained. This corresponds to a reading of 20.9% oxygen on a scale of 0 to 25%. Routine Calibrating transmitter type 10256 Potentiometers for span and zero adjustment are located on the rear side of the transmitter circuit board. Look for the markings span and zero on the printed circuit board. For calibration purposes, the signal must be monitored using either a standard milliammeter inserted into the 4-20 ma circuit, or by observing the reading on the associated display equipment. To adjust the potentiometers a small screwdriver is required. In most cases, the routine adjustments, which do not require any calibrating gas, will be sufficient. Adjust as follows: 1. Ensure the sensor is located in pure atmospheric air with no other gases present. Apply power (positive power goes to connector pin nearest to edge of pc board) and wait for the output reading from the transmitter to stabilize. This could require up to five minutes, although required time is usually much shorter. 2. Adjust the span potentiometer until a reading of 20.9% oxygen is shown in the measuring system, or until the current drawn by the sensor is exactly 17.37 ma. Turn the 820-4 December, 2005 DESIGN MANUAL

MAINTENANCE MODEL GT820 OXYGEN DEFICIENCY/ENRICHMENT GAS SENSOR span pot clockwise to increase the output. If the required output cannot be obtained, the cell is beyond its useful life and must be replaced. Complete Calibration Instructions transmitter type 10256 These calibrations are performed at the factory, and need to be repeated only if the transmitter is inadvertently misadjusted so far out of range that the routine instructions will not work. Read the transmitter output current with a milliammeter or read on the associated display equipment. As each test is made, be sure to wait for output to stabalize. 1. Place sensor in pure atmospheric air with no other gases present. Apply power (positive power goes to connector pin nearest to edge of pc board). Output will stabalize in 5 minutes or less. 2. Unplug the purple oxygen fuel cell from the printed circuit board. Adjust the zero potentiometer for an output of 4.0 milliamps on the milliammeter or zero percent (0%) on the display equipment. Then reinstall the fuel cell on the PC board. 3. Adjust the span potentiometer for an output of 17.37 ma on the milliammeter or for an output of 20.9% oxygen on display equipment. 4. Repeat steps 2 and 3 until no further adjustments are needed. Replacing the fuel cell The operating life of the fuel cell is affected by temperature, relative humidity and the amount of gas to which it has been exposed. After several years of operation, it may no longer be possible to calibrate the transmitter to the output of the cell. In that case, order replacement cells from SST. Unscrew the cover from the GT820 housing, remove the old fuel cell and replace with a new one. After replacement, recalibrate the transmitter output. DESIGN MANUAL December, 2005 820-5

Blank Page