Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

Similar documents
CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

THE development of more advanced techniques in radiotherapy,

Relative Dosimetry. Photons

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

Review of fundamental photon dosimetry quantities

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Monitor Unit Verification for Small Fields

SnapShot IMRT with compensators and FFF beams

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Clinical Implementation of Volumetric Modulated Arc Therapy

GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

Constancy checks of well-type ionization chambers with external-beam radiation units

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Surface buildup dose dependence on photon field delivery technique for IMRT

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

IMRT in clinical practice at the UMC-Utrecht. Utrecht

Dynamic wedges dosimetry and quality control

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

A beam-matching concept for medical linear accelerators

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

The dose distribution of medium energy electron boosts to the laryngectomy stoma

Dosimetric Calculations. Lonny Trestrail

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

Proton Therapy QA Tools

Scientific Journal of Nuclear Medicine & Radiation Therapy

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

Math Review. Overview

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Zoubir Ouhib Lynn Cancer Institute

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

Abstract: Introduction:

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Planning Assignment (3 field rectum)

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

Commissioning of a new total body irradiation protocol

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques

Disclosures. What Is So Hard? End of ITV: Gating is the Best ITV Killer 8/3/2016. Daniel A. Low, Ph.D. UCLA. Varian Grant Siemens Grant Accuray Grant

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Commissioning and periodic tests of the Esteya electronic brachytherapy system

VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Iranian Journal of Medical Physics

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

Radiotherapy physics & Equipments

Evaluation of six TPS algorithms in computing entrance and exit doses

High Speed Direct SAD Radiosurgery Beam Scanner

1. Question Answer cgy / MU cgy / MU 2. Question Answer

Commissioning and quality assurance for the treatment delivery components of the AccuBoost system

Electronic Oncology Systems the Management and the Safety Measures. Hansen Chen, Director, TDSI

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT

Treatment Planning Techniques for Larger Body Habitus Patients for Breast/Chestwall and Regional Lymph Nodal Irradiation

Review of backscatter measurement in kilovoltage radiotherapy using novel detectors and reduction from lack of underlying scattering material

TG-51: Experience from 150 institutions, common errors, and helpful hints

VariSource High Dose Rate Afterloader Procedures. For Performing Breast Brachytherapy with. The SAVI TM Applicator

Methods to model and predict the ViewRay treatment deliveries to aid patient scheduling and treatment planning

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Resist round robin ELETTRA

kvue IGRT Couch top One couch top, multiple treatment solutions For Varian, Elekta, Siemens, Proton and more The Best of Both Worlds

Beam modeling and VMAT performance with the Agility 160-leaf multileaf collimator

WEDGE FILTERS FOR MEGAVOLTAGE ROENTGEN RAY BEAMS

Breast Treatment System REFERENCE MANUAL

User Manual. RW3 Slab Phantom T29672 and T

The Use of ILI In Dent Assessments

3D Treatment Planning and verification with hand calculations

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

Kill-painting of hypoxic tumors in charged particle therapy

Improved Sputter Depth Resolution in Auger Thin Film Analysis Using In Situ Low Angle Cross-Sections

EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY

TEST REPORT BEAMSCAN WITH HALCYON

Modeling of Estimated Respiratory Waveform

The Gas Attenuator of FLASH

Failure Mode and Effect Analysis (FMEA) for a DMLC Tracking System

Hardware Triaxial and Consolidation Testing Systems Pressure Measurement and Control

Introduction. Materials and Methods

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber

CHAPTER-2. Wedge filters are the most important beam modifiers used in

Variable Depth Bragg Peak Test Method

Bubble-bubble interactions and wall pressures/temperatures produced by the collapse of a bubble pair near a rigid surface

External Pressure Coefficients on Saw-tooth and Mono-sloped Roofs

Influence of HALO and drain-extension doping gradients on transistor performance

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Transcription:

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media N Tyagi 1, B H Curran, P L Roberson, J M Moran, E Acosta and B A Fraass Department of Radiation Oncology, The University of Michigan, Ann Arbor, MI, USA, 48109-0010 E-mail: Currently at University of California, San Diego: ntyagi@ucsd.edu Abstract. IMRT often requires delivering small fields which may suffer from electronic disequilibrium effects. The presence of heterogeneities, particularly low-density tissues in patients, complicates such situations. In this study, we report on verification of the DPM MC code for IMRT treatment planning in heterogeneous media, using a previously developed model of the Varian 120-leaf MLC. The purpose of this study is twofold: (a) design a comprehensive list of experiments in heterogeneous media for verification of any dose calculation algorithm and (b) verify our MLC model in these heterogeneous type geometries that mimic an actual patient geometry for IMRT treatment. The measurements have been done using an IMRT head and neck phantom (CIRS phantom) and slab phantom geometries. Verification of the MLC model has been carried out using point doses measured with an A14 slim line (SL) ion chamber inside a tissue-equivalent and a bone-equivalent material using the CIRS phantom. Planar doses using lung and bone equivalent slabs have been measured and compared using EDR films (Kodak, Rochester, NY). 1. Introduction Dose calculations for MLC-based IMRT planning are quite challenging primarily because IMRT beams consist of a large number of small segments which may suffer from electronic disequilibrium. In addition, for a complex intensity pattern the dose distributions can be very sensitive to the detailed structure of the multileaf collimator (MLC).[1] The presence of an inhomogeneity exacerbates the existing electronic disequilibrium. The existing literature contains many studies showing the superior performance of Monte Carlo methods in heterogeneous media,[2-7] especially at tissue interfaces.[8-11] Most of these verification studies are, however, limited to conformal beam planning in the presence of inhomogeneities. In this article, we report on verification of the DPM MC code [12] for IMRT treatment planning in heterogeneous media using a previously developed model of the Varian 120-leaf MLC (Varian Medical Systems, Palo Alto, CA).[13] The accuracy of the model has been verified previously against ion chamber and film measurements for several SMLC and DMLC based IMRT treatment plans in a homogeneous Solid Water phantom.[13] We now compare our MLC model in heterogeneous media using an IMRT head and neck phantom that represents human head and neck anatomy and consists of interchangeable inserts for tissue equivalent and heterogeneous media. In addition, we have compared film measurements and Monte Carlo simulations at multiple depths 1 To whom any correspondence should be addressed. c 2008 Ltd 1

(including interfaces) for slab phantom geometries consisting of lung and bone equivalent slabs embedded within Solid Water for 6 MV IMRT treatment beams. Such verification measurement is necessary to document the accuracy of MC-based dose calculations in anatomical sites such as the lung, head and neck, and prostate where tissue heterogeneities make accurate dose calculation challenging. 2. Methods and Materials 2.1. A Phase-space modeling of the linear accelerator A Varian 21EX linear accelerator equipped with a 120-leaf (Millennium) multileaf collimator was simulated using the BEAMnrc[14] and DPM MC codes. The details of the treatment head simulation have been described in our previous publication.[13] In our model, the patient-independent components (target, primary collimator, flattening filter, ion chamber, and mirror) were simulated once using the BEAMnrc MC code and stored in 4 phase-space (PS) files containing approximately 200 million particles. The PS files were read into the DPM MC code and used for the patient-specific calculations which depend on the MLC, jaws and patient geometry. While multiple photon scattering is accounted for, electrons generated within the jaws and MLC deposit their energy locally.[13] 2.2. Measurement and Phantom set-up Measurements were made in flat and cylindrical geometries. First, verification measurements were made using a Solid Water (Gammex RMI, Middleton WI) phantom of dimensions 30x30x20 cm 3 at 5 cm depth, 95 cm SSD for a 6 MV photon beam as shown in Figure 1. Kodak EDR2 (Extended Dose Range) film were placed at various depths in the phantom. Lung-ICRU (ρ = 0.26 g/cm 3 ) and Cortical bone (ρ = 1.85 g/cm 3 ) slab inhomogeneities of thickness 6 cm were placed 4 cm from the top of the phantom. All the processed films were digitized using a laser film digitizer (Lumisys Lumiscan LS75, Kodak, Rochester, NY) and analyzed using a software analysis package, IGOR pro (Interactive Graphics Oriented Research) that reads the Lumisys file formats. The film measurements were taken at the following depths: 3 cm, 4 cm (water/lung or water/bone interface), 8 cm (within the inhomogeneity), 10 cm (lung/water or bone/water interface), and 15 cm (within Solid Water) as illustrated in Figure 1. The films were exposed at a dose rate of 400 MU/min. The IMRT treatment plan sequenced using a partial breast irradiation protocol (intensity pattern shown in Figure 1) was delivered using SMLC (222 segments, 345 MUs and field size = 5x8 cm 2 ) and DMLC (150 segments and 375 MUs). A dose calibration curve was measured at the same time and used to convert the film response from optical density to dose. In addition, dose differences between calculations and measurements were analyzed using a dose-gradient analysis tool or a gradient compensation method [15] that evaluates local dose differences as a function of the dose gradient at each point in the dose distribution. In gradient compensation method a distance parameter, which is typically the size of the geometric uncertainty (1mm in all our examples), was chosen and the dose differences that might have been caused by this geometric shift were removed. Additional measurements were performed using a CIRS head and neck IMRT phantom that represents the human head and neck anatomy as a cylindrical geometry. The cylindrical phantom, shown in Figure 2, is 16 cm in diameter and 16 cm in height and allows interchangeable rod/inserts of tissue equivalent and inhomogeneity material. Each insert is 2.5 cm in diameter and can be placed at any of the 5 locations as shown in Figure 2. The phantom includes measurement plugs for point dose measurements using ion chamber, TLDs, diodes, and other small detectors. The phantom also permits the use of film and gel dosimetry, however, their use was not investigated here. The head and neck phantom, with the chamber in position, was scanned using a GE Lightspeed 4 slice CT scanner in various configurations. Helical scans were obtained using a 1.25 mm slice thickness. The scans were transferred to our in-house planning system (UMPLAN), and contours were drawn for the external phantom and chamber volumes for 3-D dose calculation. 2

Anterior IMRT beam 4 cm 6 cm 10 cm Solid Water Lung or Bone slab Solid Water 3 cm 4 cm 8 cm 10 cm 15 cm Right Insert 6 cm 5 cm 3 cm 4 cm Left Posterior Figure 1.Slab Phantom geometry. Intensity pattern of the incident IMRT beam is also shown. Figure 2. IMRT Head and Neck Phantom consisting of 5 interchangeable inserts for tissueequivalent and inhomogeneity material. In this study we have used an A14 SL ion chamber with an active volume of 0.009 cm 3 for point dose measurements. The collected charge was read using an Inovision 35040 electrometer (Standard Imaging). Tissue equivalent, bone equivalent (ρ = 1.6g/cm 3 ) and air (no insert, ρ = 1.203e-03 g/cm 3 ) inserts were placed in different locations and with different beam arrangements as described in Table 1. Plans were generated using a 6 MV photon beam in the anterior-posterior (AP) and left-right (LR) direction (IEC gantry angle = 90º) using a 3x3 cm 2 square field formed by jaws and MLC. In addition, the IMRT treatment plan described in the previous section was also delivered using SMLC and DMLC delivery techniques. The isocenter for each plan was assumed to be at the center of the chamber. Each beam was delivered using 200 MU and a dose rate of 600 MU/min for each set-up. Absolute dose calibration was performed by referencing all readings to our standard calibration geometry (SAD = 100 cm, depth = 10 cm, and a 10 x 10 cm 2 field size) in a water phantom. Measured point doses were converted to absolute doses and compared with the MC simulations performed using the DPM MC code. Table 1. Ion chamber (IC) placement and beam arrangement for the CIRS head & neck phantom. Case # IC placement Isocenter location SSD (cm) Beam angles (degrees) 1 Inside Tissue Origin 92.2 0 2 Inside Tissue Origin 92.2 90 3 Inside Tissue 2.9 cm Anterior 95.1 0 4 Inside Tissue 5.1 cm Posterior 87.1 0 5 Inside Bone Origin 92.2 0 6 Inside Bone 2.9 cm Anterior 95.2 0 7 Inside Bone 5.1 cm Posterior 87.1 0 8 Inside Tissue (bone to the left) Origin 92.2 90 9 Inside Tissue (air to the left) Origin 92.2 90 10 Inside Tissue (bone to the left and air at the center) 6 cm right 86.1 90 2.3. Monte Carlo Simulations The DPM MC code, integrated with UMPLAN, was used to calculate dose in CT-based geometries for clinical-like situations. Six different materials (air, lung, adipose tissue, water, muscle and bone) were identified and the corresponding cross-sections were generated. DPM calculations were performed 3

with a maximum electron condensed history step size of 1 mm and low energy electron and photon cutoffs of 200 kev and 50 kev, respectively. All calculations are reported in absolute dose (cgy) as described in our previous publication. [13] The slab phantom geometry was simulated outside the planning system using the DPM MC code and a voxel size 2x2x1 mm 3. Sufficient histories (~10-20 billion) were simulated for each run such that the uncertainty (1σ) in the average dose over all voxels greater than Dose max /2 was less than 1 %. The comparison between calculated and measured doses in heterogeneous media was performed using point-dose comparisons for head and neck phantom geometry and 2-D dose difference maps, gradient compensated dose difference maps and 1-D line profiles at various depths for the slab phantom geometry. In addition issues related to dose reporting by Monte Carlo calculations and film measurements are also discussed. MC reports dose to medium whereas, the film is calibrated with respect to dose to water and is subsequently used to measure doses in different density material. The effect due to discrepancy in dose reporting has been addressed in the future sections. 3. Results and Discussion 3.1 Slab Phantom set-up: The IMRT beam was delivered using SMLC in the heterogeneous lung phantom. Figures 3 (a) (e) show the dose difference maps (top graphs) between film measurements and MC calculations (DPM- Film) at depths of 3 cm (within the Solid Water), 4 cm (water/lung interface), 8 cm (within the inhomogeneity), 10 cm (lung/water interface), and 15 cm (within Solid Water) respectively. The middle graphs show the corresponding gradient compensated dose difference maps at these depths and the bottom graphs show the 1-D profiles along the lines shown on the dose difference map for the corresponding dose maps at the same depths. The film data is represented using solid lines and the MC calculation is shown as dashed lines in the 1-D line profiles. The dose difference map shows differences of up to 30-40 cgy, located primarily in the penumbral regions. One possible reason for these differences is the much larger voxel size resolution of the DPM calculation (2 mm) as compared with film (0.1 mm) which could lead to averaging errors in the high dose gradient regions. In the penumbral region, shifts of 1 mm can result in dose differences of 10-20%. When a gradient compensation of 1 mm is applied, many regions of large dose difference disappear. In the out-of-field regions for the IMRT beams, the film over-responds due to its energy response to low-energy scattered photons.[16] Overall, the film dose is initially higher than MC up until the distal inhomogeneity interface and lower than MC below that depth. The gradient compensated dose difference maps (middle row) represent the differences due to limitations in the MC model or film measurement error. The differences between film and MC (DPM-film) at 3 cm (~ -6 cgy), 4 cm (~ -6 cgy) and 8 cm (~ -3 cgy) depths are within 2 to 3%. The scatter from the MLC mostly consists of low energy photons which may result in an over-response in the film measurements above the inhomogeneity. At the distal interface (10 cm), differences of the order of -3 cgy are seen once again. At a depth of 15 cm inside water, differences of the order of 2.5% (~ (3-4) cgy) are seen between film and MC. The presence of low density media causes a decrease in beam attenuation, changes in photon scatter, and an increased lateral electron range for small fields. The differences seen at lower depths could be due to a variety of factors such as: measurement uncertainties in the positioning of the film (~1 mm), energy response of the film (~2%), uncertainties in film processing (~2%), MC statistical uncertainties (~1%) and uncertainties in the model selection parameters such as the mean electron-on-target energy. The SMLC-sequenced beam (from the IMRT breast plan) was delivered in a bone equivalent medium as shown in Figures 4 (a)-(e). At depths of 3 cm and 15 cm in water and at the water-bone and bone-water interface (4 cm and 10 cm), the agreement between calculation and film measurement is within 2% with a maximum local difference of 4%. The DPM MC accurately accounts for the backscattered electrons from the bone medium at the proximal interface. Systematic differences of the order of 3.5% were seen between calculation and measurement at 8 cm depth (within the bone 4

medium). A possible reason for the discrepancy was the medium in which dose is reported: film is calibrated with respect to water and hence reports dose to water where as MC calculation in bone reports dose to the bone. To illustrate this effect a thin layer of water (1mm) was inserted inside the bone medium from 8 cm to 8.1 cm and MC simulations were performed. The dose was evaluated at the depth of 8.05 cm (center of the voxel in z-direction) and the results are shown in Figure 5. The dose difference map, gradient compensated map and profiles now show excellent agreement between measurement and calculation. The 3.5% difference represents the correction that one needs to perform in converting dose-to-medium to dose-to-water or vice versa. This difference, especially significant for high Z material, arises from the difference in stopping power of the electron spectrum in the two media. This difference, is however, not consistent with the stopping power ratios (~10-12%) reported by Siebers et al for the cortical bone and is quite puzzling.[17] At this point we do not have a clear explanation for this discrepancy. We are still investigating this issue. (a) (b) (c) (d) (e) (4-7) cgy (4-7) cgy (2-3) cgy (2-3) cgy < 3 cgy Dose (cgy) cm Figure 3: IMRT beam simulated using SMLC delivered in a lung slab medium (ρ = 0.26 g/cm3) Dose difference map (DPM-film), Gradient compensated dose difference map and 1-D profiles (along the lines shown on the dose difference map) at depths of (a) 3 cm (b) 4 cm(c) 8 cm (d) 10 cm and (e) 15 cm. Profile comparisons between film and MC; film is shown in solid and MC in dashed lines at depths of (a) 3 cm (b) 4 cm (c) 8 cm (d) 10 cm and (e) 15 cm. 3.2. IMRT Head and Neck set-up Table 2 shows the absolute dose comparison between measured data and DPM MC calculations for various situations described in Table 1. All the comparisons on an average are within 2% of the measured data. Cases 1-4 represent the situation where ion chamber was placed inside a tissue equivalent medium at various SSDs and different gantry angles. The agreement between measurement and simulations is excellent. For the situation where ion chamber is placed inside bone at various SSDs (cases 5,6 and 7) slightly higher discrepancies are seen. The differences are within the combined statistical uncertainty associated with MC simulations and measurements. As reported by Bouchard et 5

al, the reference calibration is not valid for IMRT fields or presence of inhomogeneities due to lateral disequilibrium effect that may affect the chamber fluence correction factor. [18] For a dynamic IMRT field the correction factors could be as high as 10% and requires a field dependent correction factor. In addition, ion chamber is reporting dose-to-water in bone and like film would require a stopping power correction which is not very significant in the situation (< 2%) as the beam passes through only 1.25 cm radius of bone. Also, the center of the measurement plug inside the bone insert was found to be off by 2 mm during the scans. Although the position of the chamber was marked on the bone insert after the scan, a slight difference in set-up error may account for slight differences. (a) (b) (c) (d) (e) < 5 cgy < 2 cgy (3-4) cgy < 7 cgy (2-3) cgy Dose (cgy) cm Figure 4. IMRT beam simulated using SMLC delivered in a bone slab medium (ρ = 1.85 g/cm3) Dose difference map (DPM-film), Gradient compensated dose difference map and 1-D profiles (along the lines shown on the dose difference map) at depths of (a) 3 cm (b) 4 cm(c) 8 cm (d) 10 cm and (e) 15 cm. Profile comparisons between film and MC; film is shown in solid and MC in dashed lines at depths of (a) 3 cm (b) 4 cm (c) 8 cm (d) 10 cm and (e) 15 cm. (a) (b) (c) Bone slab Dose (cgy) Figure5. Effect of reporting dose-to-water in a bone medium. Simulation carried out by inserting a 1mm layer of water (from 8-8.1 cm) in a bone equivalent medium. (a) Dose difference map (DPMfilm) (b) Gradient compensated dose difference map and 1-D line profiles at the depth of 8 cm. cm 6

When the beam passes through a 2.5 cm of high density bone (case # 8) it results in roughly a 5% reduction in dose to the ion chamber at the center (compared to the homogeneous situation) while the presence of air (case #9) enhances the dose to the chamber by approximately 10%. As shown in Table 2, DPM simulations accurately accounts for these effects. A single beam in the AP direction from a breast IMRT plan was delivered using SMLC and DMLC delivery techniques. The results are shown in Table 3. The ion chamber was located at the center of the phantom inside a tissue equivalent and bone equivalent medium. The main purpose of this test was to investigate the sensitivity of IMRT beams to the curvature/contour of the phantom. The difference between ion chamber measurements and DPM MC calculations is 2% or less in both homogeneous and heterogeneous medium. The IMRT ion chamber measurements are very sensitive to the chamber position within the phantom and require a field dependent correction factor.[18] The differences are within measurement and calculation uncertainty. Table 2. Absolute dose comparison between IC measurement and DPM MC calculation for the IMRT head and neck phantom set-up. A 6MV, 3x3 cm 2 beam was delivered to the phantom. Case # IC placement Measured Data (cgy) Monte Carlo (cgy) % difference IC MC ( ) x 100 IC 1 Inside Tissue 72.33 72.43-0.1 2 Inside Tissue 73.04 72.40 0.9 3 Inside Tissue 83.09 82.19 1.1 4 Inside Tissue 57.47 58.02-1.0 5 Inside Bone 72.22 73.68-2.0 6 Inside Bone 82.51 83.68-1.4 7 Inside Bone 57.64 56.50 2.0 8 Inside Tissue (bone to the left) 68.76 68.52 0.4 9 Inside Tissue (air to the left) 80.4 80.3 0.1 10 Inside Tissue (bone to the left and air at the center) 57.74 57.89-0.3 Table 3. Absolute dose comparison between IC measurement and DPM MC calculation for the IMRT head and neck phantom. A 6MV, IMRT beam was delivered to the phantom. Case # IC placement Measured Data (cgy) Monte Carlo (cgy) % difference IC MC ( ) x 100 IC SMLC Inside Tissue 179.43 176.75 1.5 DMLC Inside Tissue 181.59 183.75-1.2 SMLC Inside Bone 176.22 176.45-0.2 DMLC Inside Bone 177.4 180.55-2.0 4. Discussion and Conclusions In this study we have verified a previously developed model of a 120-leaf MLC (Varian 21 EX) within the DPM Monte Carlo code using ion chamber and film measurements in phantoms with high and low density inhomogeneities for square and IMRT fields. Lung and bone simulations in a slab phantom geometry show excellent agreement between film measurements and calculations. Bone simulations raise an important concern regarding reporting dose to water or to the medium. Measurement dosimetry (film, ion chamber or TLDs) is typically based on dose-to-water. Commissioning and verification of a dose calculation algorithm, especially MC-based thus requires a good understanding of measurement dosimetry and care must be taken to convert the simulated dose in the medium to dose 7

to water. MC simulations can thus potentially be used to calculate these correction factors as they depend on the beam energy and the measurement set-up. The agreement between ion chamber measurements and DPM MC calculations using a cylindrical head and neck phantom (CIRS) at varying SSDs is excellent when the chamber is placed inside a tissue equivalent or a high density bone inhomogeneity. In addition to the simulations, errors also lie in the manufacturing of the head and neck phantom and measurement plugs inside the inserts. The CT scans showed a shift in the measurement plug inside the bone insert and also a small amount of air in between the inserts and the chamber due to imperfections in the manufacturing of these rods. IMRT simulations carried out in the head and neck phantom geometry show good agreement between measurement and calculation when the chamber was placed at the center of the phantom inside a homogenous and bone equivalent medium. Such verification experiments are important to document the accuracy of any dose calculation algorithm in realistic patient-type geometry. A well benchmarked MC code in homogeneous and heterogeneous geometry provides an alternative QA tool by allowing dose calculation in full 3-D patient geometry. It also enables us to do a retrospective analysis of patients previously treated using IMRT on various protocols to investigate the effect of an improved dose calculation algorithm on clinical outcomes. Acknowledgements This work is supported by NIH grant: P01CA59872. The authors would like to thank Dr. Indrin Chetty for extremely useful comments on the paper. Thanks to Brian Tollenaar and Maria Dittman for their assistance in scanning the CIRS phantom. References [1] R. Mohan, M. Arnfield, S. Tong, Q. Wu, and J. Siebers, Med. Phys. 27 (6), 1226-37 (2000). [2] C. Martens, N. Reynaert, C. De Wagter, P. Nilsson, M. Coghe, H. Palmans, H. Thierens, and W. De Neve, Med. Phys. 29 (7), 1528-35 (2002). [3] W. U. Laub, A. Bakai, and F. Nusslin, Phys. Med. Biol. 46 (6), 1695-706 (2001). [4] F. C. du Plessis, C. A. Willemse, M. G. Lotter, and L. Goedhals, Med. Phys. 28 (4), 582-9 (2001). [5] M. Miften, M. Wiesmeyer, A. Kapur, and C. M. Ma, J Appl. Clin. Med. Phys. 2 (1), 21-31 (2001). [6] J. Seco, E. Adams, M. Bidmead, M. Partridge, and F. Verhaegen, Phys.Med. Biol. 50 (5), 817-30 (2005). [7] J. J. DeMarco, T. D. Solberg, and J. B. Smathers, Med. Phys. 25 (1), 1-11 (1998). [8] S. N. Rustgi, A. K. Rustgi, S. B. Jiang, and K. M. Ayyangar, Phys. Med. Biol. 43 (12), 3509-18 (1998). [9] O. Chibani and X. A. Li, Med. Phys. 29 (5), 835-47 (2002). [10] E. E. Klein, L. M. Chin, R. K. Rice, and B. J. Mijnheer, Int. J. of Radiat. Oncol. Biol. Phys. 27 (2), 419-27 (1993). [11] K. R. Shortt, C. K. Ross, A. F. Bielajew, and D. W. O. Rogers, Phys.Med.Biol. 31 (3), 235-49 (1986). [12] J. Sempau, S. J. Wilderman, and A. F. Bielajew, Phys. Med. Biol. 45 (8), 2263-91 (2000). [13] N. Tyagi, J. Moran, D. Litzenberg, A. Bielajew, B. Fraass, and I. Chetty, Med Phys 34, pp 651-663 (2007). [14] D.W.O. Rogers, B. Walters, and I. Kawrakow, BEAMnrc Users Manual: NRC Report PIRS 509(a)revH. (2004). [15] J. M. Moran, J. Radawski, and B. A. Fraass, J. Appl. Clin. Med. Phys. 6 (2), 62-73 (2005). [16] P. J. Muench, A. S. Meigooni, R. Nath, and W. L. McLaughlin, Med. Phys. 18 (4), 769-75 (1991). [17] J. V. Siebers, P J Keall, A E Nahum and R Mohan, Phys. Med. Biol. 45 (4), 983-995 (2000). [18] H. Bouchard and J. Seuntjens, Med. Phys. 31 (9), 2454-2465 (2004). 8