Guidance and history. Why MEIGaN? Medical Locations & Healthcare Electrics Part 1

Similar documents
CHAPTER 2: ELECTRICAL SAFETY

C&G 2395 Exam Paper June Section A-All questions carry equal marks. Answer all three questions. Show all calculations.

C&G Level 3 Award in the Periodic Inspection, Testing and Certification of Electrical Installations

5 th year (Control department)

C&G 2395 Exam Paper April Section A - All questions carry equal marks. Answer all three questions. Show all calculations.

Electrical Safety in Hospitals

Inspection and Testing Pocket Notes

Electrical Safety in the Food Industry

Inspection and Testing

SAFETY QUALITY TECHNOLOGY. Guidance on Safe Isolation Procedures

The definition of a competent person given in the Electricity at Work Regulations 1989 (EWR) is as follows:

Bulletin Swimming pool, hot tub and spa installations Rules , , , , , and

XMPA12B2142 pressure sensor XMP - 12 bar - G 1/4 female - 2 NC - without control type

Florida Building Code 6 th Edition (2017) Residential. Broward County Edition Loose-leaf Supplement

Electrical Safety. Toolbox Talk

Guidelines to the standard IEC60601

INSPECTION OF MEDICAL DEVICES

SG33KTL-M Quick Installation Guide. 1 Unpacking and Inspection

Electrical. OSHA Office of Training and Education 1

Dealing with Electrical Hazards in the Workplace

Check with local zoning official for property line distance requirements.

Code of Practice for Electrical Safety in Beer Dispense in Licensed Premises 2011 Guidance for Landlords. Fourth Edition April 2011

Zelio Relays RSL Slim Interface Plug-In Relays

Aligning EDIS with Guidance Note 3

Leakage Current Testing Is it right for your application?

Essential Performance for MED rd ed. PSES San Diego chapter meeting December 11, 2012

Low Voltage Electricity System Safety Rules & Associated Safety Guidance

Sikkerhed og certifisering for sundhedsteknologisk udstyr

DX 3 RCDs. Technical data Nature and consequences of electrical risks Direct and indirect contact. Direct contacts

AUTOMATIC GAS MANIFOLDS

Electromechanical pressure switches

Portable Appliance Testing

CERTIFICATION IMPACT ANALYSIS: IEC 60065, SEVENTH EDITION vs. GB Audio, video and similar electronic apparatus Safety requirements

Swimming Pool Requirements

IEC April Authorized Sole Agent : PT. Wisma Niagatama Perkasa. Phone : (031) Fax. : (031)

Engineering World Health Youth EWH Design Projects That Matter

Reference Number PDS 10 (RIC Standard: EP SP)

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0

This training session covers electrical safety for unqualified persons. Unqualified persons are machine operators, operators of powered industrial

Circuit breakers PR 60

Technical Data. Dimensions

SWIMMING POOL REQUIREMENTS

ARC:Gear from Unipart Rail

Gas density monitor With integrated transmitter Model GDM-100-TI

Electrical Safety. Introduction

PECIMEN IMM6. Minor Electrical Installation Works Certificates. Amd 3: In accordance with BS 7671

SG36KTL-M Quick Installation Guide. 1 Unpacking and Inspection

AutoChanger Installation & User Guide Issue 2

Spasticity / Dystonia

CCSSL35 Carry out the wiring and testing of electrical circuits and equipment

Definitions found in several locations:

ServicePlus S21R & S27R Series 2 Installation Instructions

QPEO2/037N Credit Value: 15 QCF Level: 2 GLH: 68 Maintaining electrical equipment/systems

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

ARC:Gear from Unipart Rail

PERIODIC INSPECTION REPORT FOR AN ELECTRICAL INSTALLATION

PSSI 3 High Voltage Metal-Enclosed Switchgear

ES18 Security Theory/Regulations Answer Schedule

REPORTING AND TREATMENT REQUIRED AS A RESULT OF AN ELECTRICAL INCIDENT CS-OHS-5

Technical Data. Dimensions

Circuit model of medical equipment for electrical safety purposes

Figure Source:

MEDIANA HEARTON AED A15. Mediana Automated External Defibrillator A15

Installation, operating and maintenance Instructions for Seemag bypass level indicator

London Underground - Section 12 approved Switch Disconnectors

Further technical information is available on

Code of Practice for the Inspection and Testing of Portable Electrical Equipment (PAT)

Technical Data. General specifications Switching element function DC Dual NC Rated operating distance s n 3 mm. Short-circuit protection

CS ENERGY PROCEDURE FOR ELECTRIC SHOCK TREATMENT AND REPORTING CS-OHS-05

Dimensions. Technical Data General specifications Switching element function Rated operating distance s n 5 mm

PSSI 6 Demarcation of Work Areas in Substations

G622. APPLIED SCIENCE Monitoring the Activity of the Human Body ADVANCED SUBSIDIARY GCE. Thursday 27 May 2010 Afternoon. Duration: 1 hour 30 minutes

Technical Datasheet. Rigid & Retractable Pendants. Rigid Pendant

SM6. instructions for use. connection with compartment for incoming cable from the top. MV distribution factory built assemblies at your service

Health & Safety: Electrical safety

Ground Fault Circuit Interrupter(GFCI) Policy

Technical Data. General specifications. Rated operating distance s n 5 mm

Network Regulator Maintenance Operatives LS (Non-Accredited)

PLEASE READ CAREFULLY BEFORE INSTALLING OR USING MEGA POOL SAVER MPS 1100

Metham Aviation Design Limited, Station Approach, Four Marks, Alton. Hants. GU34 5HN Tel +44 (0) Fax +44 (0)

Technical Data. Dimensions

SEMEMI SQA Unit Code H2AP 04 Carrying out fault location on electrical equipment and circuits

MAKING ELECTRICAL A GUIDE ON HOW TO INSPECT AND TEST PORTABLE APPLIANCES SAFELY AND EFFECTIVELY

Technical Data. Dimensions

Ground Fault Protection for Electrical Tools

[Instruments for vacuum measurement, checking and adjustment] 3

POWR-BLOKS Distribution/Splicer Blocks and Covers

Policies, Procedures, Guidelines and Protocols

Data sheet Pressure switch type CS December 2001 DKACT.PD.P10.A B1119

Evaluation of Ground Ring vs. Equipotential Mat at a Swimming Pool in Buford, Georgia

TeSys protection components

AREA VALVE SERVICE UNITS

Copper Bonded Ground Rod / Grounding Rods / Copper Bonded Earth Rod / Earthing rods

PRODUCT SPECIFICATION

Electromechanical pressure switches

PERMANENT POOLS. Location

Lightning and surge protection for yachts

15 Appendix 15 Notes on periodic inspecting and testing

Electrical Shore Connections

Technical Data. General specifications Switching element function Rated operating distance s n 2 mm

Transcription:

Medical Locations & Healthcare Electrics Part 1 Emergency Isolation Room Paediatric Isolation Room Paediatric HDU Medical Locations The risk, Earthing, IPS and UPS IEC 60364, MEIGaN and IET GN7 2015 Richard Knight C Eng. MCIBSE, FIHEEM Village Hotels, Coryton, Cardiff Monday 6 th March 2017 Patient Monitor 1 And now me in 2015! Medical Locations And BS 7671:2008, BS 7671:2008 AMD 1, & BS 7671:2008 Corrigendum, June 2013 Safety issues and review of guidance and developments Guidance and history HSR 25 Memorandum of guidance on the Electricity at Work Regs 1989:2015 Third ed, first published 1989 MEIGaN was published to replace TRS 89: (Technical Requirements for the supply and installation of Equipment for Diagnostic Imaging and Radiotherapy (1989)) MEIGaN first published 2005. with annex and updated in 2007 to take account of the changes in procedures undertaken in X-ray rooms and other similar locations. Why MEIGaN? The IEC 60364 document makes certain performance requirements, but does not define well how to deliver it in the field MEIGaN, requires additional performance tests that are easier to measure in the field e.g. Touch voltages and when to measure them It also defines better the practical installation measures to improve chance of passing the tests e.g. ERB design 1

The IEE then the IET Guidance and history GN7 :1998 GN7 :2003 GN7 :2004 reprinted 2006 incorporating amendments to imprints verso GN7 :2009 BS 7671:2008 AMD 1: 2011 GN7 :2012 BS 7671:2008 Corrigendum June 2013 BS 7671:2008 AMD 3: 2015 GN7 :2015 Why IEE/IET? The IEC 60364 document makes certain performance requirements, but does not define well how to deliver it in the field GN7 and then BS 7671 are intended to put into a UK context the requirements of IEC 60364 The variants of GN7 and BS 7671 (8 of them) are intended to cover IEC 60364-7-710:2002 (now HD:March 2012) EaWR 1989 guidance on regulation 3 Persons on whom duties are imposed by these Regulations A person who may cause danger to others by their actions In the healthcare setting this might be with withdrawal of power from an ICU (similar has occurred with medical gasses) Or for example a PFI on an existing site sharing electrical infrastructure (often exasperated by poor understanding of risk and of the system) EaWR 1989 guidance on regulation 3 Persons on whom duties are imposed by these Regulations Absolute/reasonably practicable 'reasonably practicable'. Where qualifying terms are absent the requirement in the regulation is said to be absolute. By definition if the clause is absolute the requirement must be met regardless of cost or any other consideration EaWR 1989 guidance on regulation 3 Persons on whom duties are imposed by these Regulations Reasonably practicable Clause 58 gives some guidance The greater the degree of risk, the less weight that can be given to the cost of measures needed to prevent that risk. Regulation 29 can give some defence, but very little 2

EaWR 1989 guidance on regulation 8 Earthing or other suitable precautions Precautions are to be taken by earthing or by preventing contact with exposed conductive parts this being achieved by earthing, double insulation, safe voltages... This regulation is absolute EaWR 1989 guidance on regulation 8 Earthing or other suitable precautions (a) double insulation; (b) earthing; (c) connection to a common voltage reference point on the system; (d) equipotential bonding; (e.g. ERB) (e) use of safe voltages; (f) earth-free non-conducting environments; (g) current/energy limitation; and (h) separated or isolated systems (e.g.ips) The other requirements of a Medical Electrical Installation. Protection against an electric shock is normally based on a hand to hand or hand to foot shock. Current is assumed to be limited by skin resistance. The type of shock experienced is classified as a MACRO Shock. Classification of Macro-Shock. Can t let go (Child) Can t let go (Adult) Suffocation Ventricular fibrillation Perception 5 ma 10 ma 35 ma 100 ma ~1 ma In a medical location we must also consider the possibility of Micro-Shock. A Micro-Shock is a shock below the threshold of perception, and is therefore not felt as a shock.(~1ma) All patients undergoing procedures which involve placing a conductor in the central circulatory system which is accessible outside of the patient are at risk. In a medical location we must also consider the possibility of Micro-Shock. Such procedures are increasingly used in treatment and diagnosis. The diagram shows the potential gradient across the heart for a hand to hand shock, and for a hand to catheter shock. A Micro-Shock applied to the heart can trigger Ventricular Fibrillation, in which the upper and lower chambers of the heart loose synchronisation The conductor could be an endoscope used in Key-Hole Surgery, a pacing lead, or most common of all, a catheter, a plastic tube filled with saline. 3

Cumulative Percentage Micro-shock The drawing shows the structure of the heart and the electrical activity during one cycle. The pumping action is initiated by a nerve impulse at a point called the Sinus Node(1). This impulse spreads out, causing the contraction of both atria(2), and gives rise to the P wave. The atria contract pumping the blood into the ventricles. Micro-shock The excitation wave passes from the atria through the Atrioventricular node and the Bundle of His nerve pathway (3) to the ventricles. As soon as the excitation reaches the ventricles, their activity, shown by the QRS complex, begins. The ventricles contract forcing blood out into the aorta and pulmonary arteries. This is followed by the re-polarisation of the ventricles, shown as the T wave. This is the most vulnerable point in the cardiac cycle. Micro-shock The ECG and Blood pressure curve below shows that when ventricular fibrillation occurs, the synchronised contractions of the atria and ventricles become disorganised, so that pumping action no longer takes place. This causes the blood pressure to drop, and the blood stops circulating. The result is that the brain becomes starved of oxygen and begins to shut down. Unless this situation can be quickly reversed death will ensue. Micro-shock Mechanical stimulation of the heart carries some small risk of triggering ventricular fibrillation. The risk remains at about 0.2% for currents below 10 A, but increases sharply above this point, At 50 A the probability will have risen to about 1%. Probability of inducing ventricular fibrillation Probability of inducing ventricular fibrillation Pub. Association for the Advancement of Medical Instrumentation ANSVAAMI ES1-1993 Published by. Association for the Advancement of Medical Instrumentation ANSVAAMI ES1-1993 Some papers show the risk is higher Microamperes 4

Micro shock in renal units How can we ensure that there are no significant differences in potential between various earthed surfaces? Providing a correctly designed Earth Reference Bar is installed, there should be no difficulty in achieving a potential difference of less that 10 mv between the various earthed surfaces. European Dialysis and Transplant Nurses Association/Europe an Renal Care Association (EDTNA/ERCA) Journal 2006 XXXII nr3 This will ensure that a current of less than 10 A will flow into the heart. The Earth Reference Bar The only way of preventing micro-shock is to ensure that all earthed surfaces are at the same potential. In practical terms this means that every medical location where interventional procedures take place should have an Earth Reference Bar. An Earth Reference Bar is defined as: One or more copper connection bars installed in an enclosure, and forming part of the protective earth system in a room and designated as a reference or datum for the purpose of defining and measuring resistance values. The Earth Reference Bar The copper connecting bars should have a CSA sufficient to carry the peak short circuit current in the event of a short circuit from phase to earth. The bars should be housed in a lockable enclosure mounted in an accessible position and Labelled Earth Reference Bar or ERB. In order to facilitate fault finding, earths should be grouped together as shown in the diagram. There should be sufficient bolts available for each earth conductor to be individually connected. Brass bolts, washers, and nuts should be used. The earth conductor should terminate on the ERB, ensuring the lowest possible resistance in the earth path. Earth reference bar as suggested by MEIGaN Similar for GN7 This installation was live when this photograph was taken Earth conductors should terminate at the ERB not on the isolator or bus bar The earth reference bar should be arranged as MEIGaN, this facilitates testing GN7 and section 710 also are similar but slightly less defined 5

The photograph below shows an example of an inadequate ERB Earthing All equipotential bonding and protective earths should be returned to the ERB, together with the earths from all the mains sockets. Bolt and lug connections are important if these are not carried out correctly these often cause earth resistances (and touch voltages) to increase from what is a very low level The maximum resistance measured from the ERB to the earth connection of all installed devices, and the resistance from the ERB to the earth pin of all the mains sockets should be less than 100 milliohms. GN 7 and section 710 have similar but higher values Earthing All accessible conductive surfaces should be earthed to the ERB. A touch voltage check should be made to ensure that there are no touch voltages greater than 10 millivolts present in locations where interventional procedures will take place. Phase All of the mains sockets in the location should be on the same phase. In most locations they should also be fed from an IPS So what if both sockets were on the same phase in a group 2 location? If socket a is fed from UPS A circuit L1 and another socket is fed from UPS B circuit L1 Supplementary Earth Points Where: Are they really on the same phase? Only if the UPSs frequency is tied together by the UPS controls and interlinking cables what if the generator is supplying UPS A, and UPSB is in by pass? If not that will be a touch voltage over 25volts then!! Perhaps UPS A & B installations should not have a by pass? In 2012 Renal group 1 IET BS 7671:2008 Corrigendum June 2013 6

Supplementary Earth Points Where: Supplementary Earth Points How many: Supplementary equipotential bonding connection points for the connection of medical electrical equipment shall be provided in each medical Location, as follows: vi Group 1: a minimum of one per patient location IET GN 7:2015 vii Group 2: a minimum of four but not less than 25% of the number medical IT socket-outlet provided per medical location. BS 7671:2008 Corrigendum June 2013, 710.415.2.1 and now GN 7:2015, 9.12 Supplementary Earth Points Supplementary Earth Points How many are needed? 3 here How many are needed? 1 here Supplementary Earth Points How many are needed? Supplementary Earth Points How many are needed? 3 here to many to count lets see how many... 7

Supplementary Earth Points How many are needed? and another here Supplementary Earth Points How many are needed? and another here From a community hospital with out a CCU All from a fairly old stack system Good socket and stud arrangement Look out for these, not allowed In a cardiac theatre Shame about the plastic socket Earth reference Bars Earth reference Bars 8

Equipotential Bonding points Equipotential Bonding points in BS 7671:2008 So how are they to be identified? 514.13 Warning notices: earthing and bonding connections States:... Equipotential Bonding points Equipotential Bonding points, labels 514.13.1 A durable label to BS 951 with the words "Safety Electrical Connection Do Not Remove" shall be permanently fixed in a visible position at or near: (i) the point of connection of every earthing conductor to an earth electrode, and (ii) the point of connection of every bonding conductor to an extraneous-conductive-part, and (iii) the main earth terminal, where separate from main switchgear. Equipotential Bonding points, labels Equipotential Bonding points, labels 9

Cabling Equipotential Bonding points, labels Testing Test all metal surfaces And the supplementary earth studs Including the fixed medical equipment 10

Testing Taps as well That will be a bit high then! Not all that seems is metal is metal except when it is glued together Mad socket locations IPS/UPS systems Most Cat 4/5 (or group 2) rooms will include an IPS/UPS system, the purpose of which is to increase supply resilience. IPS/UPS systems The maximum rating of an IPS system is 10kVA. Where there is the possibility of the need for more output than this, two IPS systems must be installed, but both can be supplied from the same UPS. IPS/UPS systems It is common practice where there are more than one IPSs supplying a theatre suite, to split the load in each theatre between two IPS units, so that should one fail, critical items of equipment can be transferred to sockets that are still live. 11

IPS/UPS systems Sockets that are supplied by the IPS should be blue in colour, engraved Medical Devices Only in white. They can be either fitted with a double pole switch or unswitched. Sufficient, conveniently located sockets should be installed to ensure that extension mains leads are not needed. Spot the mistake, plastic sockets not allowed 16.50 Metal finished sockets Metalfinished sockets should be installed within a clinical risk Category 3 area and above in order to limit the effect of electromagnetic interference and the increased mechanical protection. 16.50 Note BS 7671:2008 AMD 3:2015 Section 512 now makes reference to the EMC directive and includes requirements for the designer of the fixed installation Circuit protection (IET touch voltage) Electrical medical locations, are defined in GN7 as a special location, like many special locations circuits The designer should check that the touch voltage cannot exceed 25 V a.c... (GN 7:2012 & 2015 9.12.1) What does this mean for typical circuits? From HTM 06 01 Part A Socket-outlets in clinical risk category 4 and 5 (e.g. Group 2) areas and connected to the medical IT system (IPS circuits) will be connected in a radial format. The protective device for such circuits should be a 20 A MCB with a Type B characteristic. 16.45 What is this guy on!! Richard Knight Mobile +44 (0) 77 949 14 211 Email millham.orchard@tiscali.co.uk Thank you PDF and Slides Richard Knight PDF and Slides Richard Knight 12