Model 951A NO/NOx Analyzer

Similar documents
Model 951C NO x Analyzer

Model 951C NOx Analyzer

MODEL 400 CLD INSTRUCTION MANUAL

Rosemount Analytical MODEL 951C NOX ANALYZER INSTRUCTION MANUAL PN

DPC-30 DPC-100. Reference Manual

Rosemount Analytical MODEL 755 OXYGEN ANALYZER INSTRUCTION MANUAL J

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Columbus Instruments

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Rosemount Analytical MODEL 755R OXYGEN ANALYZER INSTRUCTION MANUAL

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

! Warning, refer to accompanying documents.

RAM 4021-DPX Operation Manual

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

RAM Operation Manual

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

CSA Sample Draw Aspirator Adapter Operator s Manual

ACV-10 Automatic Control Valve

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

INSTALLATION & MAINTENANCE INSTRUCTIONS

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Design Features. General Description

21700 Nordhoff St. Chatsworth, CA Phone

Model 400A Hydrocarbon Analyzer

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

Generating Calibration Gas Standards

Inert Air (N2) Systems Manual

RC 195 Receiver-Controller

RAM 4021 Operation Manual

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

36C/36D HSI, DSI Proven Pilot Gas Valves INSTALLATION INSTRUCTIONS

Stand-Alone Bubble Detection System

SPECIFICATIONS ATTENTION

Autocalibration Systems For In Situ Oxygen Analyzers

OPERATING INSTRUCTIONS FOR MODEL

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

CHAPTER 16 %UHDWKLQJ*DV0L[LQJ3URFHGXUHV

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

Thermo Scientific Model 146i Principle of Operation

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Model PDT Dewpoint Transmitter

Best Practice Guide, Servomex 2700

RK-IR Sample Draw Aspirator Adapter Operator s Manual

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

The HumiSys. RH Generator. Operation. Applications. Designed, built, and supported by InstruQuest Inc.

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

Powers Controls RC 195 Multiple Input Receiver- Controller

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

TR Electronic Pressure Regulator. User s Manual

Autocalibration Systems for In Situ Oxygen Analyzers

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

GAS FUEL VALVE FORM AGV5 OM 8-03

Your safety and the safety of others are very important.

User s Guide Temperature Sensor Converter TSC-599

GM Series Dual-Block Multi-Function Gas Control Valves

H-DIL Hydrogen/Oxygen Sample-Draw Detector Operator s Manual

Model 130M Pneumatic Controller

Technical Information

In Vivo Scientific, LLC INSTRUCTION MANUAL

Overview. Front Panel: Keypad and Display

TECHNICAL INFORMATION

DF-550E PROCESS ANALYSERS APPLICATIONS FEATURES

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Workhorse-15 Oxygen Generator

INSTRUCTION MANUAL. Lesker Model KJL615TC-K Lesker Model KJL615TC-H. Digital Vacuum Gauge/Controller. KJL615TC K/H Ranges

AFC. SDPROC and AFC Analog Mass Flow Controller ANALOG MASS FLOW CONTROLLERS. Principles of Operation. Design Features

Powers Controls RC 195 Single Input Receiver-Controller

3-Way Decoking Valve Option - Installation and Operation Instructions Accessory G Background

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

PERFORM Operating Document

CAPITAL CONTROLS SERIES NXT3000

Title: Standard Operating Procedure for Measurement of Ethylene (C 2 H 4 ) in Ambient Air by Reduced Gas Detection (RGD)

Model EXAxt AV550G. Averaging Oxygen Analyzer

This test shall be carried out on all vehicles equipped with open type traction batteries.

Nuclear Associates

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

Capital Controls Series NXT3000 Modular design gas feed system with self-contained automatic switchover capability.

METHOD 3A - DETERMINATION OF OXYGEN AND CARBON DIOXIDE CONCENTRATIONS IN EMISSIONS FROM STATIONARY SOURCES (INSTRUMENTAL ANALYZER PROCEDURE)

N2-14, N2-22, N2-35, N2-14A, N2-22A, N2-35A

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

RK LEL Sample Draw Aspirator Adapter Operator s Manual

PERFORM Operating Document

TECHNICAL DATA MAINTENANCE AIR COMPRESSOR MODEL G-1

Flow meter. bellow vaporizer. APL valve. Scavenging system

Features. Main Voltage: Vdc. Input Control Signal: 0-5 Vdc or 4-20 madc (Sourcing) Monitor Output Voltage: 0-5 Vdc or 4-20 madc (Sourcing)

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

White Paper. Chemical Sensor vs NDIR - Overview: NDIR Technology:

OPERATION MANUAL NTF-15

Rejuvenation Instructions

INSTRUCTION MANUAL FOR MODEL 7360V VERTICAL CURING CHAMBER Revision E May

200 PSI FAST-FILL AIR SOURCE KIT

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

HumiSys HF High Flow RH Generator

Transcription:

Instruction Manual Model 951A NO/NOx Analyzer http://www.processanalytic.com

ESSENTIAL INSTRUCTIONS READ THIS PAGE BEFORE PROCEEDING! Rosemount Analytical designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you MUST properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions MUST be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount Analytical products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation. Read all instructions prior to installing, operating, and servicing the product. If you do not understand any of the instructions, contact your Rosemount Analytical representative for clarification. Follow all warnings, cautions, and instructions marked on and supplied with the product. Inform and educate your personnel in the proper installation, operation, and maintenance of the product. Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources. To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product. When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product s performance, place the safe operation of your process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire, electrical hazards, or improper operation. Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury. The information contained in this document is subject to change without notice. Teflon is a registered trademark of E. I. dupont de Nemours and Co., Inc. Alconox is a registered trademark of Alconox, Inc. Emerson Process Management Rosemount Analytical Inc. Process Analytic Division 1201 N. Main St. Orrville, OH 44667-0901 T (330) 682-9010 F (330) 684-4434 e-mail: gas.csc@emersonprocess.com http://www.processanalytic.com

Model 951A Instruction Manual TABLE OF CONTENTS PREFACE...P-1 Definitions...P-1 Intended Use Statement...P-2 Safety Summary...P-2 General Precautions For Handling And Storing High Pressure Gas Cylinders...P-4 Documentation...P-5 Compliances...P-5 1-0 DESCRIPTION AND SPECIFICATIONS...1-1 1-1 Overview...1-1 1-2 Options...1-2 a. Range I.D....1-2 b. Remote Range Change...1-2 c. Range I.D. and Remote Range Change...1-2 d. Sample Pump...1-2 1-3 Specifications...1-3 2-0 INSTALLATION...2-1 2-1 Facility Preparation...2-1 a. Outline and Mounting Dimensions...2-1 b. Location...2-1 c. Power Requirements...2-1 2-2 Unpacking...2-1 2-3 Gas Requirements...2-1 a. Air Or Oxygen...2-1 b. Span Gas...2-2 2-4 Sample Requirements...2-2 2-5 Gas Connections...2-2 2-6 Electrical Connections...2-6 a. Recorder Output...2-6 b. Thermocouple Connections for Measuring Temperature of NO 2 to NO Converter...2-6 c. Remote Range Change...2-6 d. Connections of the Range I. D. Kit...2-6 e. Power Connections...2-6 3-0 STARTUP AND OPERATION...3-1 3-1 Startup Procedure...3-1 3-2 Calibration...3-3 3-3 Measuring Efficiency Of No 2 To No Converter And Adjusting Temperature Setpoint...3-4 a. Test Setup for Measurement of Conversion Efficiency...3-4 b. Test Procedure...3-4 c. Subnormal Conversion Efficiency...3-8 d. Replacement of Converter...3-8 e. Capillaries...3-8 f. TEA Scrubber...3-9 3-4 Routine Operation...3-9 3-5 Recommended Calibration Frequency...3-9 Rosemount Analytical Inc. A Division of Emerson Process Management Contents i

Instruction Manual Model 951A 4-0 THEORY...4-1 4-1 Principles Of Operation...4-1 a. Nitric Oxide Determination By Chemiluminescense Method...4-1 b. NO X Determination...4-1 c. Ozone Generation...4-1 4-2 Analyzer Flow System...4-1 a. Flow Of Sample Or Standard Gas To Reaction Chamber...4-1 b. Flow of Air or Oxygen...4-3 c. Flow System Operating Modes...4-3 d. Converter Bleed Flow...4-4 4-3 Electronic Circuitry...4-4 a. Amplifier Board and Associated Circuitry...4-5 b. Valve Control Board, Front Panel Mode Switch and Associated Circuitry...4-6 c. ±15 Volt Power Supply...4-6 d. High Voltage Power Supply...4-7 e. Converter Temperature Control Board and Associated Elements...4-7 f. Fan Control Circuit...4-8 g. Remote Operation Option...4-8 h. Range I.D. Option...4-8 5-0 ROUTINE SERVICING...5-1 5-1 System Checks And Adjustments...5-1 a. Meter Mechanical Zero...5-1 b. Amplifier Zero Adjustments...5-1 c. Inter-Range Attenuation Correlation adjustments...5-1 d. Meter Fullscale Span Adjustment...5-2 e. Overall Sensitivity...5-3 f. Ozone Output...5-3 g. Background Current...5-3 h. NO 2 to NO Converter Temperature Adjustment...5-4 i. Flow Balance...5-4 5-2 Servicing Flow System...5-5 a. Sample Capillary...5-5 b. Ozone Restrictor and Capillary...5-6 5-3 NO 2 To NO Converter...5-7 a. Replacing...5-7 5-4 Photomultiplier/Reaction Chamber...5-8 a. Cleaning Reaction Chamber...5-8 b. Photomultiplier Tube and Housing...5-10 5-5 Servicing Electronic Circuitry...5-11 6-0 REPLACEMENT PARTS...6-1 6-1 Circuit Board Replacement Policy...6-1 6-2 Matrix...6-1 6-3 Replacement Parts...6-2 a. Pneumatics...6-7 b. Converter Assembly...6-8 c. Low Tempco Option...6-10 7-0 RETURN OF MATERIAL...7-1 7-1 Return Of Material...7-1 7-2 Customer Service...7-1 7-3 Training...7-1 ii Contents Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual LIST OF ILLUSTRATIONS Figure 1-1. Model 951A NO/NO X Analyzer... 1-1 Figure 2-1. Model 951A Outline and Mounting Dimensions... 2-3 Figure 2-2. Front Panel Indicators and Controls... 2-4 Figure 2-3. Controls and Adjustments Located Behind Swing Out Front Panel... 2-4 Figure 2-4. Rear Panel... 2-4 Figure 2-5. Remote Range Kit Installed... 2-7 Figure 2-6. Typical Interconnection of Remote Range Kit... 2-7 Figure 3-1. Amplifier Board Adjustments... 3-3 Figure 3-2. Measuring Efficiency of NO 2 to NO Converter... 3-7 Figure 3-3. Conversion Efficiency as a Function of Converter Temperature... 3-8 Figure 4-1. Schematic Flow Diagram of Model 951A... 4-2 Figure 4-2. Functional Schematic Diagram of Electronic Signal Circuitry... 4-5 Figure 5-1. Amplifier Board...5-2 Figure 5-2. NO 2 to NO Converter Assembly... 5-7 Figure 5-3. Reaction Chamber/Photomultiplier Assembly... 5-9 Figure 5-4. Reaction Chamber Assembly and Phototube Housing... 5-9 Figure 5-5. Terminal Chassis Assembly Wiring Diagram... 5-12 Figure 6-1. Model 951A Exploded View... 6-3 Figure 6-2. Model 951A Exploded View (continued)... 6-4 Figure 6-3. Model 951A Exploded View (continued)... 6-5 Figure 6-4. Electronics... 6-6 Figure 6-5. Front Panel Pneumatic Components... 6-7 Figure 6-6. Converter Components... 6-8 Figure 6-7. Ozone Generator Components... 6-9 Figure 6-8. Tempco Retrofit Component Location... 6-10 LIST OF TABLES Table 2-1. Model 951A Controls and Adjustments... 2-5 Table 3-1. Proper Gas Supply Pressures for Various Levels of Sample NO X... 3-2 Rosemount Analytical Inc. A Division of Emerson Process Management Contents iii

Instruction Manual Model 951A DRAWINGS (Located In Rear Of Manual) 619604 Schematic Diagram, Computer Interface 619710 Schematic Diagram, ±15V Power Supply 641871 Schematic Diagram, Temp Control 649819 Schematic Diagram, Amplifier Board 649822 Diagram, Tubing - Model 95A 649834 Schematic Diagram, 951A NO/NOx Analyzer 649835 Pictorial Wiring Diagram, Model 951A NO/NOx Analyzer 649958 Flow Diagram, Model 951A 652423 Schematic Diagram, Power Supply 652834 Schematic Diagram, Power Supply - Thermocooler 652838 Pictorial Wiring Diagram, Low Tempco Option 654348 Schematic Diagram, Hi Voltage Board Assembly 656313 Schematic Diagram, Valve Control Board 780350 Schematic Diagram, 4-20mA 0-5V Option 780726 Wiring Diagram, 4-20mA 0-5V Option 780727 Installation Drawing, Model 951A w/4-20ma 0-5V Option 780809 Wiring Diagram, 4-20mA 0-5V Option and Low Tempco iv Contents Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual PREFACE The purpose of this manual is to provide information concerning the components, functions, installation and maintenance of the Model 951A. Some sections may describe equipment not used in your configuration. The user should become thoroughly familiar with the operation of this analyzer before operating it. Read this instruction manual completely. DEFINITIONS The following definitions apply to DANGERS, WARNINGS, CAUTIONS and NOTES found throughout this publication. DANGER. Highlights the presence of a hazard which will cause severe personal injury, death, or substantial property damage if the warning is ignored. WARNING. Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in injury, death, or long-term health hazards of personnel. CAUTION. Highlights an operation or maintenance procedure, practice, condition, statement, etc. If not strictly observed, could result in damage to or destruction of equipment, or loss of effectiveness. NOTE Highlights an essential operating procedure, condition or statement. Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-1

Instruction Manual Model 951A INTENDED USE STATEMENT The Model 951A NO/NOx Analyzer is intended for use as an industrial process measurement device only. It is not intended for use in medical, diagnostic, or life support applications, and no independent agency certifications or approvals are to be implied as covering such applications. SAFETY SUMMARY If this equipment is used in a manner not specified in these instructions, protective systems may be impaired. AUTHORIZED PERSONNEL To avoid explosion, loss of life, personal injury and damage to this equipment and on-site property, do not operate or service this instrument before reading and understanding this instruction manual and receiving appropriate training. Save these instructions. DANGER. ELECTRICAL SHOCK HAZARD Do not operate without doors and covers secure.. Installation requires access to live parts which can cause death or serious injury. For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power. DANGER. ULTRA VIOLET LIGHT HAZARD UV light from the ozone generator can cause permanent eye damage. DO NOT LOOK DIRECTLY AT THE UV SOURCE IN THE OZONE GENERATOR. Use of UV filtering glasses is recommended. P-2 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual DANGER. OZONE HAZARD This instrument generates ozone while operating. Ozone is toxic by inhalation and is a strong irritant to the throat and lungs. Ozone is also a strong oxidizing agent. Its presence is detected by a characteristic pungent odor. The instrument EXHAUST outlet contains ozone and nitrogen dioxide which is toxic by inhalation. The BYPASS outlet contains various oxides of nitrogen, and if the sample source is from the exhaust of an internal combustion engine, it may contain unburned hydrocarbons and carbon monoxide which is highly toxic and, depending on duration of exposure, can cause headache, nausea, loss of consciousness and death. Avoid any inhalation of the internally generated ozone, sample, EXHAUST and BYPASS discharge. Keep all tubing fittings checked for tightness to avoid internal leaks. Connect rear panel EXHAUST and BYPASS outlets to outside vent via separate lines 1/4 inch (6.3mm) or larger. Use only Teflon or stainless steel tubing. WARNING. PARTS INTEGRITY Tampering with or unauthorized substitution of components may adversely affect the safety of this instrument. Use only factory approved components for repair. CAUTION AIR FLOW Do not operate instrument without air flow to the ozonator; plugging of the filter will result. CAUTION. PRESSURIZED GAS This unit requires periodic calibration with a known standard gas. It also may utilizes a pressurized carrier gas, such as helium, hydrogen, or nitrogen. See General Precautions for Handling and Storing High Pressure Gas Cylinders, page P-4. Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-3

Instruction Manual Model 951A GENERAL PRECAUTIONS FOR HANDLING AND STORING HIGH PRESSURE GAS CYLINDERS Edited from selected paragraphs of the Compressed Gas Association's "Handbook of Compressed Gases" published in 1981 Compressed Gas Association 1235 Jefferson Davis Highway Arlington, Virginia 22202 Used by Permission 1. Never drop cylinders or permit them to strike each other violently. 2. Cylinders may be stored in the open, but in such cases, should be protected against extremes of weather and, to prevent rusting, from the dampness of the ground. Cylinders should be stored in the shade when located in areas where extreme temperatures are prevalent. 3. The valve protection cap should be left on each cylinder until it has been secured against a wall or bench, or placed in a cylinder stand, and is ready to be used. 4. Avoid dragging, rolling, or sliding cylinders, even for a short distance; they should be moved by using a suitable hand-truck. 5. Never tamper with safety devices in valves or cylinders. 6. Do not store full and empty cylinders together. Serious suckback can occur when an empty cylinder is attached to a pressurized system. 7. No part of cylinder should be subjected to a temperature higher than 125 F (52 C). A flame should never be permitted to come in contact with any part of a compressed gas cylinder. 8. Do not place cylinders where they may become part of an electric circuit. When electric arc welding, precautions must be taken to prevent striking an arc against the cylinder. P-4 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual DOCUMENTATION Contact Customer Serv- The following Model 951A NO/NOx Analyzer instruction materials are available. ice Center or the local representative to order. 556383 Instruction Manual (this document) COMPLIANCES This product may carry approvals from several certifying agencies, for use in non-hazardous, indoor locations. 97-C218 Rosemount Analytical Inc. A Division of Emerson Process Management Preface P-5

Instruction Manual Model 951A P-6 Preface Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual SECTION 1 DESCRIPTION AND SPECIFICATIONS 1-1 OVERVIEW The Model 951A NO/NO X Analyzer continuously analyzes a flowing gas sample performing one of two switch selectable determinations: 1. Nitric oxide (NO); or 2. Combined nitric oxide (NO) and nitrogen dioxide (NO 2 ) designated NO X. By definition [NO X ] = [NO] + [NO 2 ]. Typical applications include analyzing vehicular exhaust emissions from internal combustion engines and monitoring the effluent from stationary (stack) sources The analyzer utilizes the chemiluminescent method of detection In the nitric oxide determination, sample NO is quantitatively converted into NO 2 by gas-phase oxidation with molecular ozone produced within the analyzer from air or oxygen supplied by an external cylinder. A characteristic of this reaction is the elevation of approximately 10% of the NO 2 molecules to an electronically-excited state, followed by immediate reversion to the non-excited state accompanied by emission of photons. The emitted photons impinge on a photomultiplier detector generating a low-level DC current. The current is amplified to drive a front panel meter and an accessory potentiometric recorder if desired. To minimize noise and reduce dark current, the photomultiplier tube is mounted in a thermoelectrically-cooled housing with temperature held constant at about 59 o F (15 o C). Control circuitry is contained in the 652831 Cooler Temperature Controller/Power Supply Assembly. The power supply circuit provides a high-current source of DC voltage. A thermistor sensor attached to the housing, and an associated switching transistor, control a pass transistor, providing closely regulated on-off control. Attached to the cooler housing is thermal fuse F3, setpoint 150 o F (65 o C). This fuse protects the thermoelectric cooler against the overheating that otherwise could occur as a result of excessively high ambient temperature or failure of a fan. SPAN ZERO ON OFF NO MODE NOx PPM RANGE 250 1000 CONVERTER 100 2500 25 10000 10 AUTO ADJUST Rosemount Analytical Model 951A NO/NOx Analyzer Figure 1-1. Model 951A NO/NO X Analyzer Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1

Instruction Manual Model 951A CAUTION EXCESSIVE HEAT Do not operate this analyzer without the air duct covering the cooling fins of the thermoelectric cooler. Excessive heat may damage the cooling devices. The cooler indicator lamp DS1 (mounted on the upper left side of the inner flow regulator panel and can be viewed through the window) will cycle on and off until a control point is reached. Analyzer functioning for the NO X determination is identical to that described above for the NO determination except that, before entry into the reaction chamber, the sample is routed through a converter where the NO 2 component is dissociated to form NO. Instrument response is proportional to total NO in the converted sample, that is the sum of the NO present in the original sample plus the NO produced by dissociation of NO 2. To minimize system response time an internal sample-bypass feature provides high-velocity sample flow through the analyzer. The electronic circuitry is modularized, utilizing plug-in printed circuit boards with solid state components and test jacks for easy troubleshooting and replacement. If desired, the Model 951A may be factory equipped with various optional features in addition to the standard features of the basic instrument. Brief descriptions of the principal options are given in the following section. 1-2 OPTIONS a. Range I.D. This option provides contact closure signals that enable a computer or other external device to identify the setting of the front panel PPM RANGE Switch. The Model 951A is designed to continuously measure the concentration of a single component of interest in a flowing gas mixture. The measurement is based on the different thermal conductivity's of the individual components of the sample stream. The method is especially well suited to analysis of two-component sample streams. However, analysis of multi-component streams is possible if the various components of the background gas occur in relatively constant ratio, or have similar thermal conductivity's. b. Remote Range Change For applications where remote operation of the analyzer is desired, as in an emission test console, the Remote Range Change Option may be used. This option permits either the operator or a computer to override and disable the front panel MODE and PPM RANGE Switches and thus to control selection of: (a) NO or NO X mode, and (b) ppm range. The unit consists of an electrical plug connector with plug-in logic card and harness for connection to a rear panel terminal strip on the analyzer. c. Range I.D. and Remote Range Change The Range I.D./Remote Range Change Option is a combination of the Range I.D. and Remote Range Change options. This option is compatible with a user supplied remote control system employing a 24 VDC digital output and input for analyzer range control and analyzer range sense, respectively. It is completely integral within the analyzer and provides a terminal strip on the rear of the analyzer for connections of the user cable. d. Sample Pump The basic Model 951A is designed to accept pressurized samples. To permit analysis of gases at atmospheric or subatmospheric pressure the analyzer may be equipped with an optional, internally mounted sample pump. This option is not available on Low Tempco versions of the 951A. An external, accessory pump can be ordered instead. 1-2 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual 1-3 SPECIFICATIONS Catalog Number... 193702 Model 951A NO/NO X Analyzer Ranges... Selectable fullscale range of 10, 25, 100, 250, 1000, 2500 and 10,000 parts per million Sensitivity... 0.1 ppm on 10 ppm range Linearity... ±1% of fullscale The specified linearity is obtainable throughout the operating range, contingent upon use of an appropriate combination of oxygen source gas, gas pressure settings and electronic adjustments. Response Time (Electronic Plus Flow) Standard Sample Capillary... Approximately one second on all ranges except 10 ppm. Approximately three seconds on 10 ppm range. Auxiliary Sample Capillary... Five seconds on all ranges. For such applications as monitoring stack sources, where comparatively slow response is desired, an internal switch provides an optional electronic response time of approximately 10 seconds to 90% of fullscale on all ranges. Precision... ±5% of fullscale Stability Zero... 1% of fullscale in 24 hours Span... 1% of fullscale in 24 hours Detector Operating Temperature.. Atmospheric Recorder Output... Selectable output of 10 millivolts, 100 millivolts, 1 volt or 5 volts Ambient Temperature... 40 F to 100 F (4.4 C to 37.7 C) Electrical Power Requirements... 107 to 127 VAC, 50/60 Hz, 1000 watts Dimensions... 9.0 x 17.8 x 22.0 inches (228.6 x 450.9 x 558.8mm) HxWxD Weight... 76 lbs. (34.5 kg) Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3

Instruction Manual Model 951A 1-4 Description and Specifications Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual SECTION 2 INSTALLATION 2-1 FACILITY PREPARATION Sections 2-1a through 2-1c provide information that may be required prior to installation of the analyzer. a. Outline and Mounting Dimensions Significant dimensions are shown in Figure 2-1. b. Location Install analyzer in a clean area, not subject to excessive vibration or extreme temperature variations. Preferably, the analyzer should be mounted near the sample stream, to minimize sample transport time. A circuit controlled by a thermal switch holds internal temperature of the analyzer to the correct operating temperature for ambient temperatures in the range 40 o F to 100 o F (4.4 o C to 37.7 o C). Temperatures outside these limits necessitate use of special temperature controlling equipment or environmental protection. Preferably, the cylinders of air (or oxygen) and span gas should be located in an area of relatively constant ambient temperature. c. Power Requirements 2-2 UNPACKING Electrical power requirements are 107 to 127 VAC, 50/60 Hz, 1000 watts. Unpack instrument carefully. Preparatory to shipment, the photomultiplier tube housing and the sample pump (if instrument is so equipped) are immobilized with hold down screws, inserted from the bottom of the instrument and marked with red paint for identification. The hold down screws must be removed prior to operation of the instrument. In the event the instrument is ever returned to the factory, these screws must be replaced to ensure safe shipment. 2-3 GAS REQUIREMENTS CAUTION HIGH PRESSURE GAS CYLINDERS This instrument requires use of oxygen and a known standard gas in high pressure cylinders. Refer to Handling and Storing High Pressure Gas Cylinders located on page P-4. a. Air or Oxygen This is used as both (a) oxygen source for generation of the ozone required for the chemiluminescent reaction, and (b) standard gas for zero calibration. Gas for both purposes may be supplied from a single cylinder and routed through a tee. Alternatively, two separate cylinders may be used. Oxygen is usable in all applications. Air is suitable for the oxygen source only if the desired fullscale operating range is 2500 ppm or less. Breathing grade oxygen or air is recommended. Clean, dried ambient air containing less than 0.1 ppm nitric oxide may be used, provided that its dewpoint is below -10 o F (-23 o C). If air is insufficiently dried, or contains excessive nitric oxide, instrument response will not be linear to 2500 ppm. Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1

Instruction Manual Model 951A b. Span Gas This is a standard gas of accurately known composition, used to set an upscale calibration point. Alternative span gases are: 1. The usual span gas is NO in a background of nitrogen. 2. For span check or adjustment in the NO X mode, the span gas may be NO 2 in a background of air or nitrogen. 3. For convenient calibration in either the NO or NO X mode, the span gas may be a cylinder gas mixture consisting of known concentrations of both NO and NO 2 in a background of nitrogen. NOTE For maximum calibration accuracy, the concentration of NO and/or NO 2 in the span gas should be as near as possible to that in the sample gas. Also, the span gas should be supplied to the rear panel SAMPLE inlet at the same pressure as the sample gas. To ensure constant pressure, a pressure regulator may be utilized immediately upstream from the SAMPLE inlet. Preferably, each gas used should be supplied from a tank or cylinder equipped with a clean, non-corrosive type two stage regulator. In addition, a shut-off valve is desirable. If possible, install the gas cylinders in an area of relatively constant ambient temperature. 2-4 SAMPLE REQUIREMENTS The Sample must be relatively clean and dry before entering the analyzer. In general, before admission to the analyzer, the sample should be filtered to 2 microns and should have a dewpoint below 90 o F (32 o C). The factory can provide technical assistance if desired. Proper supply pressures for sample and span gases depend on whether or not the analyzer is equipped with the optional 632748 Sample Pump: For the basic Model 951A, without sample pump, sample must be supplied to the SAM- PLE inlet at a pressure of 5 to 10 psig (34.5 to 69 kpa). This ensures that the normal bypass flow of two liters per minute will be obtainable by adjustment of the BYPASS Needle Valve. (Proper bypass flow is essential for rapid system response and stable flow into reaction chamber.) For an analyzer equipped with sample pump, the acceptable pressure range at the SAM- PLE inlet is 0 to 5 psig (0 to 34.5 kpa). The pump pressurizes the sample to between 5 and 10 psig (34.5 to 69 kpa) for supply to the internal flow system. 2-5 GAS CONNECTIONS 1. Remove plugs and caps from all inlet and outlet fittings. See Figure 2-4. 2. Connect EXHAUST outlet to outside vent via tubing with O.D. of 1/4 inch (6.3mm) or larger. 3. Connect external lines from air (or oxygen) cylinder and sample source to corresponding rear panel inlet ports. For sample line, stainless steel tubing is recommended. 4. Adjust regulator on air (or oxygen) cylinder for output pressure of 35 to 40 psig (251 to 276 kpa). 5. Supply sample gas to rear panel SAMPLE inlet at appropriate pressure: 5 to 10 psig (34.5 to 69 kpa) for basic analyzer without sample pump; 0 to 5 psig (0 to 34.5 kpa) for analyzer with pump. 2-2 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual 19.0 [482] 8.7 [222] 1.25 [32] 20.2 [513] 9.0 [228] 1.0 [25.4] 6.0 [152] 1.25 [32] 8.7 [221] 7.0 [178] 4.5 [114] 4.6 [118] 6.6 [168] DIMENSIONS Inch mm 3. ALLOW 16 INCHES IN FRONT OF INSTRU- MENT FOR DOOR SWING. 2. ALLOW 12 INCHES MINIMUM CLEARANCE ABOVE INSTRUMENT FOR MAINTENANCE.. 1. RECOMMENDED MOUNTING HARDWARD: 10-32 MACHINE SCREWS (SUPPLIED BY CUSTOMER). 2.0 [51] 18.3 [465] 18.3 [465] 18.3 [465] 18.3 [465] 18.3 [465] RECOMMENDED PANEL CUTOUT Figure 2-1. Model 951A Outline and Mounting Dimensions 2.8 [70] 8.5 [216].25 [63] 6 PLC'S Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3

Instruction Manual Model 951A Zero Adjust Span Adjust Sample Pressure Gauge Meter SPAN Ozone Pressure Gauge AC Power On/off Switch Mode Switch MODE ON NO NOx OFF Rosemount Analytical ZERO PPM RANGE 250 1000 CONVERTER 100 2500 25 10000 10 AUTO ADJUST B OZONE Y P FLOW A BALANCE S S OZONE OFF Model 951A NO/NOx Analyzer Bypass Flowmeter PPM Range Switch Converter Adjust Converter Indicator Figure 2-2. Front Panel Indicators and Controls Response Time Selector Switch Bypass Needle Valve Bypass Flowmeter Photomultiplier High Voltage Interlock Switch 10 100 MV REC 5 FAST SLOW 1 RESPONSE V OUTPUT Ozone Pressure Regulator Recorder Output Select Switch R6 R22 R34 CALIB R16 R33 CALIB R31 CALIB INPUT R32 CALIB FLOW BALANCE B Y P A S S OZONE Ozone Off Pushbutton Switch OZONE OFF Rosemount Analytical Model 951A NO/NOx Analyzer Flow Balance Needle Valve Sample Backpressure Regulator Figure 2-3. Controls and Adjustments Located Behind Swing Out Front Panel Sample Inlet Recorder Connection TB4-2 Recorder Connection TB4-1 Thermocouple Connector TB9 Range I.D. Connector J10 (option) Cable Gland For Recorder Cable Air Inlet Exhaust AC Power Cable Fuse F1 Fuse F2 Figure 2-4. Rear Panel 2-4 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual CONTROL FUNCTION ON/OFF Controls AC power to all components in instrument MODE Switch For selection of NO or NO X Readout of NO or NO X in ppm as selected with MODE switch. Upper scale is Meter graduated 0 to 100 for use with 10, 100, 1000, and 10,000 ppm ranges. Lower scale is graduated 0 to 25 for use with 25 and 2500 ppm ranges. PPM RANGE Switch Select fullscale range for meter and recorder LED illuminates during application of power to converter heater. When Converter reaches temperature equilibrium, LED will go off and on at intervals of CONVERTER about 2 seconds, indicating correct temperature control. Through-panel screwdriver adjustment of converter temperature. Should be set for temperature that yields optimum combination of high efficiency for the NO ADJUST 2 to NO conversion and extended life for the catalytic converter. Optimum temperature differs from one instrument to another, see Section 3-3, page 3-4. Set zero point on meter scale or recorder chart. With MODE switch in NO position and zero air supplied to SAMPLE inlet, ZERO Control is adjusted for zero ZERO reading. Set upscale calibration point on meter scale or recorder chart. MODE switch is at NO or NO X and suitably pressurized standard gas of accurately known SPAN NO/NO X content is supplied to SAMPLE inlet. With PPM RANGE switch set to the appropriate range for the span gas, the SPAN control is adjusted to correct reading on meter or recorder. Adjustment and indication of pressure (and therefore flow) of sample or standard SAMPLE backpressure regulator and pressure gauge gas routed through sample capillary, and into reaction chamber. Proper setting dependent on operating range. See Table 3-1, page 3-2. Bypass flow indication and adjustment for sample or standard gas to SAMPLE BYPASS Flowmeter and Needle Valve inlet. Setting of 2 L/min. is recommended to ensure rapid response and optimum functioning of sample backpressure regulator Pressure adjustment and indication for air or oxygen supplied to AIR inlet for use OZONE pressure regulator and as oxygen source for internal ozone generator. Oxygen is usable for all ranges. gauge Air is usable only for 2500 ppm range or lower. Proper pressure setting dependent on operating range, see Table 3-1, page 3-2. Removes AC power to ultraviolet source lamp in ozone generator. Power is OZONE OFF switch OFF when switch indicator is red. Use to equalize pressure drop for NO and NO X legs of the internal flow system. Criterion for correct adjustment is an NO2 free nitric oxide standard gas should FLOW BALANCE needle valve give equal readings of ppm for NO and NO X modes. An alternate flow balance check, if NO 2 free standard gas is unavailable, is to verify that the reading on the SAMPLE pressure gauge is the same for NO and NO X modes. Select either FAST or SLOW electronic response. When set to FAST, electronic response time (for 0 to 90% of fullscale) is then approximately one second for all RESPONSE TIME selector ranges except 10 ppm, which is approximately three seconds. When set to switch SLOW (as in monitoring stack sources), electronic response time 0 to 90% of fullscale) is approximately ten seconds for all ranges. RECORDER OUTPUT selector To select output of 10 mv, 100 mv, 1V or 5 V for a potentiometric recorder. switch Located on rear of meter. With AC power OFF, meter should read zero. If not Meter mechanical zero use screw adjustment on meter. Photomultiplier High Voltage When analyzer front panel is open, photomultiplier high voltage power supply is Interlock Switch automatically shut off. Table 2-1. Model 951A Controls and Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5

Instruction Manual Model 951A 2-6 ELECTRICAL CONNECTIONS a. Recorder Output If a recorder is used, connect leads to terminals marked REC 1 (+) and 2 (-) on TB4 (Figure 2-4, page 2-4). (Note that, within the analyzer, the negative recorder output terminal is connected to ground.) Set REC OUTPUT Selector Switch SW1 (Figure 2-3, page 2-4) to the recorder span: 10 mv, 100 mv, 1 VDC or 5 VDC. b. Thermocouple Connections for Measuring Temperature of NO 2 to NO Converter Temperature of the NO 2 to NO converter may be monitored by connecting a customer supplied millivolt meter to thermocouple connector TB9 located on the rear panel (Figure 2-4, page 2-4). The position of the thermocouple in the converter bundle will influence the actual temperature readout. c. Remote Range Change The Remote Range Change Kit, Figure 2-5 page 2-7, consists of a plug-in Adapter Board plus attached harness and rear terminal plate. The Adapter Board has two two-position slide switches: the MAN/AUTO MODE Switch and the MAN/AUTO RANGE Switch. The MAN/AUTO MODE Switch provides the choice of local or remote control of the NO/NO X Mode Switching Solenoid Valve, associated with the NO 2 to NO converter. With switch at MAN, the function is under control of the front panel MODE Switch. With switch at AUTO, the front panel MODE Switch is disabled; the function is then remotely controlled, by either the operator or a computer, via contact closure signals applied to terminals on TB5. The MAN/AUTO RANGE Switch provides the choice of local or remote selection of operating range. With switch at MAN, range selection is under control of the front panel PPM RANGE Switch. With switch at AUTO, the front panel PPM RANGE Switch is disabled; range selection is remotely controlled, by either the operator or a computer, via contact closure signals applied to terminals on TB6. Control at TB5 and TB6 is accomplished by a contact closure from the terminal marked COM (for common) to the terminal marked with the name of the desired function. Contact closure requirements are 25 ma at 15 VDC. d. Connections of the Range I. D. Kit The Range I.D. option provides contact closure signals that permit a computer or other external device to identify the position selected manually with the front panel PPM RANGE Switch. The cable is connected to PPM RANGE Switch SW1 and extends to connector J10 mounted on the rear of the case (Figure 2-4, page 2-4). The pin-out connections of J10 are as follows: Pin A Pin B Pin C Pin D Pin E Pin F Pin H Pin J 10 ppm 25 ppm 100 ppm 250 ppm 1000 ppm 2500 ppm 10000 ppm Wiper e. Power Connections Connect power cord to an AC source of 107 to 127 volts, 50/60 Hz. If power outlet does not have third (ground) contact, use an adapter to provide proper grounding. 2-6 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual Computer Adapter Board 10 5 FAST SLOW 100 1 RESPONSE MV V REC OUTPUT R22 R34 CALIB R16 R33 CALIB R31 CALIB R6 INPUT R32 CALIB Electronics Assembly DC Harness (J1) NOX SPAN ZERO COM 10 25 100 250 1000 2 500 Auto/Man Range Switch RANGE MODE AUTO AUTO MAN MAN Rosemount Analytical Auto/Man Mode Switch 10,000 Computer Adapter Board installed on Interconnect Board of the Electronics Assembly Terminal Assembly mounted on rear panel of Model 951A Figure 2-5. Remote Range Kit Installed Terminal Assembly Remote Mode Select Switch NOx 10 NOx SPAN ZERO COM 25 100 250 1000 Remote Range Select Switch 2500 10,000 COM With all contacts of Range Select Switch open, analyzer is on 250 ppm range. Figure 2-6. Typical Interconnection of Remote Range Kit Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7

Instruction Manual Model 951A 2-8 Installation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual SECTION 3 STARTUP AND OPERATION Preparatory to startup and operation, a thorough familiarization with Table 2-1 (page 2-5), Figure 2-2 (page 2-4), Figure 2-3 (page 2-4) and Figure 2-4 (page 2-4) is recommended. These figures give locations and brief descriptions of operating controls. For more detailed information on controls, refer to Section 4. 3-1 STARTUP PROCEDURE Following are detailed stepwise instructions on startup and calibration. For convenience, condensed instructions for startup, routine calibration and normal operation are provided at the front of this manual. WARNING OZONE HAZARD When instrument power is on, the ultraviolet source sample is energized, converting a portion of the oxygen contained with the ozonator into ozone. With normal flow through the ozonator, the ozonated air or oxygen is continuously swept though the ozonator and into the reaction chamber. If flow is stopped, however, the ozone will diffuse and will attack the sintered metallic restrictor element in the tee fitting at the upstream end of the ozonator. Operation of the analyzer with the restrictor element thus rusted will result in the following consequences: Reduced flow of air or oxygen through the ozonator, ozone deprivation within the reaction chamber and non-linear response of the analyzer to NO/NO X. To prevent such damage, verify that the front panel ozone ON/OFF switch is turned off if flow of feed gas to the air inlet is terminated. 1. With power removed from analyzer, check front panel meter. It should read zero; if not, adjust Mechanical Zero Screw at rear of meter for zero reading. 2. Place front panel PPM RANGE Switch at 1000. 3. Set front panel MODE Switch at NO or NO X. Place POWER Switch ON. Electrical power is now being supplied to all circuits, including sample pump if analyzer is so equipped. Analyzer will now require approximately one hour for temperature equilibration before ready for calibration. 4. Establish correct pressure for air oxygen: a. Verify that pressure regulator on cylinder of air or oxygen is set for supply pressure of 35 to 40 psig (242 to 276 kpa). b. On internal gas control panel, Figure 2-3. adjust OZONE Pressure Regulator so that OZONE Pressure Gauge indicates either 20 psig (138 kpa) or 30 psig (207 kpa), depending on the desired operating range. 5. Establish correct flow of sample gas: a. Supply sample gas to rear panel SAMPLE inlet. b. Adjust SAMPLE Backpressure Regulator so SAMPLE Pressure Gauge indicates the value appropriate to the desired operating range. c. Adjust BYPASS Needle Valve for reading of two liters per minute on BYPASS Flowmeter. Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-1

Instruction Manual Model 951A NOTE Inability to obtain a bypass flow of two liters per minute by adjustment of the BYPASS Needle Valve usually indicates insufficient sample supply pressure at the SAMPLE inlet. 6. Establish correct flow of zero air: a. Supply zero air to rear panel SAM- PLE inlet. b. Note reading on SAMPLE Pressure Gauge. It should be the same as in Step 5b. If not, adjust output pressure regulator on air cylinder as required. 7. Establish correct flow of upscale standard gas: a. Supply upscale standard gas to rear panel SAMPLE inlet. b. Note reading on SAMPLE Pressure Gauge. It should be the same as in Step 5b. If not, adjust output pressure regulator on cylinder of upscale standard gas as required. NOTE Supply pressures for sample and upscale standard gases must be the same; otherwise, readout will be in error. The Analyzer is now ready for calibration per Section 3-2, page 3-3. Max. NO X Level (ppm) Gas Supplied to Air Inlet 1000 Air 2500 Air 10,000 Oxygen Ozone Pressure Gauge Setting 20 psig (138 kpa) Provides flow of approx. 500 cc/min to ozone generator 30 psig (207 kpa) Provides flow of approx. 1000 cc/min to ozone generator 30 psig (207 kpa) Provides flow of approx. 1000 cc/min to ozone generator Sample Pressure Gauge Setting 4 psig (27.6 kpa) Provides flow of approx. 60 cc/min to reaction chamber 1.5 psig (10.3 kpa) Provides flow of approx. 20 cc/min to reaction chamber Table 3-1. Proper Gas Supply Pressures for Various Levels of Sample NO X 3-2 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual 3-2 CALIBRATION 1. Zero Calibration. a. Set PPM RANGE Switch for the same range that will be used during sample analysis. Set SPAN Control at about midrange. b. Supply zero gas to rear panel SAMPLE inlet. c. Adjust ZERO Control for zero reading on meter or recorder, then lock ZERO Control knob. 2. Upscale Calibration. a. Set PPM RANGE Switch at the position appropriate to the particular span gas. b. Supply upscale standard gas of accurately known NO/NO X content to rear panel SAMPLE inlet. c. Place MODE Switch at NO if nitric oxide span gas is used. d. Adjust SPAN Control so that reading on meter or recorder is equal to the known ppm concentration of NO or NO X in the span gas. In operation at NO X levels in the range of 2500 to 10000 ppm, the correct reading may not be obtainable, initially, by adjustment of the SPAN Control. The cause is that the reduced sample flow required for linearity results in lowered sensitivity and slower response. To compensate for these effects, make the electronic adjustments of Steps e and f below. e. If necessary, increase sensitivity by raising photomultiplier voltage per Section 5-1e, page 5-3. f. To reduce observed noise, select SLOW position of Response Time Selector Switch SW2, Figure 3-1. 3. When correct upscale reading is obtained, lock SPAN Control knob. Calibration is now complete. Before Placing analyzer in operation, however, measure efficiency of the NO 2 to NO converter per Section 3-3, page 3-4. 10 100 MV REC 5 FAST SLOW 1 RESPONSE V OUTPUT R22 R34 CALIB R16 RESPONSE TIME SWITCH R6 INPUT R33 CALIB R31 CALIB R32 CALIB Figure 3-1. Amplifier Board Adjustments Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-3

Instruction Manual Model 951A 3-3 MEASURING EFFICIENCY OF NO 2 TO NO CONVERTER AND ADJUSTING TEM- PERATURE SETPOINT It is the responsibility of the user to measure efficiency of the NO 2 to NO converter during initial startup and thereafter at intervals appropriate to the application, normally once a month. The reactant material used in the converter provides the optimum combination of high conversion efficiency and low ammonia interference. Unlike most competitive analyzers, the Model 951A utilizes a reactant material that gradually becomes more efficient at a given temperature, and thus after a period of use may permit operation with a lower temperature setpoint than that required initially. Conversion efficiency in the Model 951A is typically 95% to 98% +, considerably above the 90% minimum accepted by the EPA. Refer to 40 CFR 60, App. A, Method 20, Section 5.6. a. Test Setup for Measurement of Conversion Efficiency A typical setup for measurement of conversion efficiency is shown in Figure 3-2A, page 3-7. The test setup includes: 1. A cylinder of nitric oxide standard gas consisting of NO in N 2. The concentration of NO in the standard gas should be about the fullscale value of the range under test. The test sample supplied to the analyzer should contain a concentration of NO comparable to that in the samples that are to be analyzed. Alternately, a higher concentration NO standard may be used if the test setup includes provision for diluting it appropriately with zero air. Suitable standard gases are available from various suppliers. Stainless steel cylinders are commonly used, but specially treated aluminum is preferred for low ppm NO samples. 2. An ozone generator utilizing an ultraviolet lamp, not a corona discharge. A corona discharge ozone generator is undesirable as it may produce oxygen atoms, which can then combine with atmospheric nitrogen to form NO. The result can be an erroneously high value for the measured conversion efficiency. b. Test Procedure 1. Measure converter temperature by connecting a millivolt meter to terminals 5 (+) and 6 (-) on TB9. Within the analyzer, these terminals are connected to a Type J thermocouple. Note present reading as a reference for comparison with subsequent measurements. The optimum temperature differs from one analyzer to another. but typically is in the range of 660 o F to 750 o F (350 o C to 400 o C). 2. Turn front panel CONVERTER AD- JUST one turn counterclockwise, to lower temperature setpoint slightly from the original factory setting. Wait 15 minutes for temperature equilibration, then proceed to Step 3. 3. Refer to Figure 3-2, page 3-7. Connect the Model 958 NO2 Converter Efficiency Tester to the Model 951A, and use the following procedure, adapted from 40 CFR 60, to determine the values to be used in Equation (A): a. Attach the NO/N 2 supply at C2. the air supply at C1, and the analyzer inlet connection at C3. b. With the variable transformer off, place Model 951A in NO mode and close valve MV1. Open valve MV2 until Model 951A SAMPLE Pressure Gauge reaches operating pressure and 3-4 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual BYPASS Flowmeter indicates some bypass flow. Wait until stable readings are obtained at analyzer. Zero and span the analyzer output to indicate the value of the NO concentration being used. This value should be approximately 80% of fullscale. Record this concentration. c. Open valve MV1 (air supply metering valve) and adjust to blend enough air to lower the NO concentration (ii) about 10%. Record this concentration. d. Turn on the ozonator and increase its supply voltage until the NO concentration of (iii) is reduced to about 20 percent of (ii). NO 2 is now being formed from the NO + O 3 reaction. There must always be at least 10 percent un-reacted NO at this point. Record this concentration. e. When a stable reading has been obtained from (iv), place Model 951A in NO X mode. It will now indicate the total NO X concentration. Record this concentration. f. Turn off the ozonator and allow the analyzer reading to stabilize. This reading is the total NO X concentration of the dilute NO span gas used at step (iii). Record this concentration. g. Close valve MV1. The NO concentration should be equal to or greater than the reading of (ii). Indicating whether the NO contains any NO 2. h. Calculate the efficiency of the NO converter by substituting the concentrations obtained during the test into the following equation. Equation (A) % Efficiency = 1 + v-vi iii-iv X 100 In the example of Figure 3-2B (page 3-7), 85-90 % Efficiency = 1 + X 100% = 92% 80-20 Efficiency checks should be made on each analyzer range, using an NO span gas concentration appropriate to the instrument range. NOTE In the initial measurement, following lowering of the temperature setpoint in Step 2., the efficiency will normally be less than 92%. 4. Turn CONVERTER ADJUST one quarter turn clockwise, thus raising temperature setpoint slightly. Wait 15 minutes for temperature equilibration. Again measure conversion efficiency per Step 3. ; it should be somewhat improved. 5. Repeat Step 4. until either (a) 95% efficiency is attained, or (b) the final one quarter turn clockwise adjustment yields an increase in efficiency of less than 1%. Then proceed to Step 6.. 6. Turn CONVERTER ADJUST one-eighth turn counterclockwise, thus lowering temperature setpoint very slightly. Converter temperature is now set at a point on the front edge of the plateau on the efficiency-vs.-temperature curve, Figure 3-3, page 3-8. This setting should provide the optimum combination of high conversion efficiency and low ammonia interference. Wait about fifteen minutes for temperature equilibration, then again measure converter temperature as in Step 1.. Compare present temperature with original value. Normally, converter temperature should be in the range of 660 o F to 750 o F (350 o C to 400 o C). Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-5

Instruction Manual Model 951A NOTE Converter temperature is not a direct measure or conversion efficiency. Temperature measurement is for reference purposes only. 3-6 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual A. TYPICAL TEST SETUP CONNECTOR C3 FLOWMETER FM2 METERING VALVE MV2 115 VAC ON OFF T1 VARIABLE TRANSFORMER OZONATOR UTILIZING ULTRAVIOLET LAMP FLOWMETER FM1 TO SAMPLE INLET OF MODEL 951A FLOW APPROXIMATELY 3 LITERS PER MINUTE CONNECTOR C2 METERING VALVE MV1 CONNECTOR C2 CONNECTOR C1 STANDARD GAS: NO IN N2 BACKGROUND ZERO AIR ALL LINES AND FITTINGS STAINLESS STEEL OR TEFLON B. TYPICAL TEST RESULTS 90 85 80 NO in N2 AIR ADDED OZONATOR OFF a-b OZONATOR ON c c-d a b % Efficiency = 100 1+ v - vi iii - iv 20 d TO NOX MODE 0 NO Mode NOx Mode Figure 3-2. Measuring Efficiency of NO 2 to NO Converter Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-7

Instruction Manual Model 951A About 95% Final 1/4 Turn Clockwise adjustment of CONVERTER ADJUST Converter Efficiency Final 1/8-Turn Counterclockwise Adjustment of CONVERTER ADJUST Converter Temperature Figure 3-3. Conversion Efficiency as a Function of Converter Temperature c. Subnormal Conversion Efficiency If a measured conversion efficiency of 95% is unobtainable within the normal temperature range, the most probable cause is depletion of the catalytic material within the converter. However, before concluding that the converter is defective, make sure that the conversion efficiency measurement is accurate. Even though the measured efficiency is less than 95%, the actual efficiency may be somewhat higher. An apparent subnormal efficiency can be due to a problem external to the analyzer, located either within the test setup, or between it and the analyzer. Check for the following: 1. Leakage. 2. Loss of NO 2 between test setup and analyzer. Such loss can occur by reaction with a rubber diaphragm in a pressure regulator or flow controller. Stainless steel diaphragms are preferred. Loss can also occur during passage through filter media. d. Replacement of Converter If it is determined that the subnormal conversion efficiency is real, and not due to measurement error introduced by the test setup, the converter must be replaced per Section 5-3a, page 5-7. The usual cause of converter failure is destruction of a large part of the catalytic material by excessive heat, due either to excessively high temperature setpoint or failure of the converter temperature control circuit. e. Capillaries Replacement capillaries should be installed finger-tight. Use of a wrench can constrict capillaries, changing flow rate. 3-8 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual f. TEA Scrubber The TEA Scrubber accessory (PN 635741) can be used to remove residual NO 2 from the NO cylinders. Use of this accessory allows a true calibration of the analyzer. White crystal deposits on the windows of the reaction chamber and plugging of capillaries and vent are usually due to sample contaminates reacting with the high ozone levels. To eliminate the contaminates, the sampling system should be reworked or a preventive maintenance program developed (if dropout is not excessive). Another source of crystalline formation is contaminated air. 3-4 ROUTINE OPERATION After calibrating analyzer per Section 3-2 (page 3-3), supply sample to SAMPLE inlet. Place MODE Switch in NO or NO X position, depending on the desired determination. Set PPM RANGE Switch in appropriate position. The instrument will now continuously analyze the sample stream. The Model 951A is designed for continuous operation. Normally, it is never turned off except for servicing or for a prolonged shutdown. However, if the instrument is to remain idle for an extended period with power on, the recommended practice is to leave the MODE Switch in NO position to de-energize the NO/NO X Mode Switching Solenoid Valve. 3-5 RECOMMENDED CALIBRATION FRE- QUENCY After initial startup, or startup following a shutdown, the analyzer requires about one hour for stabilization before it is ready for calibration. Maximum permissible interval between calibrations depends on the analytical accuracy required, and therefore cannot be specified. It is recommended that initially the instrument be calibrated at least once every eight hours, and that this practice be continued until experience or requirements indicate that some other interval is more appropriate. Rosemount Analytical Inc. A Division of Emerson Process Management Startup and Operation 3-9

Instruction Manual Model 951A 3-10 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual SECTION 4 THEORY 4-1 PRINCIPLES OF OPERATION a. Nitric Oxide Determination By Chemiluminescense Method The chemiluminescence method for detection of nitric oxide (NO) is based on its reaction with ozone (O 3 ) to produce nitrogen dioxide (NO 2 ) and oxygen (O 2 ). Some of the nitrogen dioxide molecules thus produced are initially in an electronically excited state (NO 2 *). These revert immediately to the ground state, with emission of photons. The reactions involved are: where NO + O 3 NO 2 * + O 2 NO 2 * NO 2 + hv h = Planck's constant v = frequency, Hz As NO and O 3 mix in the reaction chamber, the chemiluminescence reaction produces light emission that is directly proportional to the concentration of NO. This emission is measured by a photomultiplier tube and associated electronic circuitry. Refer to Section 4-3, page 4-4. b. NO X Determination The NO X determination is identical to the NO determination described in Section 4-1a except that, prior to entry into the reaction chamber, the sample is routed through a converter where the NO 2 component is converted into NO. The reaction is: 2NO 2 2NO + O 2 Instrument response is proportional to total NO in the converted sample, that is, to the sum of the NO originally present in the sample plus the NO resulting from conversion of NO 2. This sum of NO and NO 2 is commonly termed NO X, i.e., ([NO X ] = [NO] + [NO 2 ]). c. Ozone Generation Ozone for the chemiluminescence reaction is produced in a flow chamber where a stream of air or oxygen from an external cylinder is exposed to ultraviolet radiation from a source lamp. The reaction is: hv 3O 2 2O 3 4-2 ANALYZER FLOW SYSTEM The analyzer flow system is shown schematically in Figure 4-1, page 4-2. Its basic function is to deliver regulated flows of sample or calibration gas and ozonized air or oxygen to the reaction chamber. The discharge from the reaction chamber exits the analyzer via the EXHAUST outlet. a. Flow of Sample or Standard Gas to Reaction Chamber Suitably pressurized sample or standard gas is supplied to the rear panel SAMPLE inlet. Within the analyzer, flow rate of the selected gas into the reaction chamber is controlled by a back-pressure regulator. It provides an adjustable, controlled pressure on the upstream side, where gas is supplied to the calibrated, flow-limiting sample capillary. The regulator is adjusted for appropriate reading on the SAMPLE Pressure Gauge. For operation at NO X levels below 2,500 ppm, correct setting on the SAMPLE Pressure Gauge is 4 psig (27.6 kpa), resulting in a flow of approximately 60 cc/min to the reaction chamber. For operation at NO X levels in the range of 2,500 to 10,000 ppm, setting on SAMPLE Pressure Gauge should be Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-1

Instruction Manual Model 951A 1.5 psig (10.3 kpa), resulting in a flow of approximately 20 cc/min to the reaction chamber. Gas in excess of the flow required for the reaction chamber is discharged through the BYPASS Flowmeter and out the BY- PASS outlet. Bypass flow should be two liters per minute to ensure optimum functioning of the SAMPLE Pressure Regulator and rapid system response. Total flow of sample or standard gas into the SAM- PLE inlet is adjusted by means of the BYPASS Needle Valve (or optional flow controller). Excessive changes, on the order of ±5 psig (±34.5 kpa), in the pressure of the sample or standard gas will affect the bypass flow rate and can affect accuracy. OZONE Pressure Regulator OZONE Pressure Gauge Ozone Generator Ozonized Air or Oxygen for Chemiluminescent Reaction Window End-On Photomultiplier Tube AIR Inlet Restrictor Reaction Chamber Air or Oxygen SAMPLE Inlet Flow Controller (option) Sample Pump (option) Bypass Valve NO2 to NO Converter Flow Balance Valve NO/NOx Mode Switching Solenoid Valve K3 Sample Pressure Gauge Sample Capillary Pump Capillary Signal to Electronic Circuitry Bleed Capillary Exhaust Bypass Flowmeter Sample Backpressure Regulator Figure 4-1. Schematic Flow Diagram of Model 951A 4-2 Theory Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual b. Flow of Air or Oxygen WARNING OZONE HAZARD When instrument power is on, the ultraviolet source sample is energized, converting a portion of the oxygen contained with the ozonator into ozone. With normal flow through the ozonator, the ozonized air or oxygen is continuously swept though the ozonator and into the reaction chamber. If flow is stopped, however, the ozone will diffuse and will attack the sintered metallic restrictor element in the tee fitting at the upstream end of the ozonator. Operation of the analyzer with the restrictor element thus rusted will result in the following consequences: Reduced flow of air or oxygen through the ozonator, ozone deprivation within the reaction chamber and non-linear response of the analyzer to NO/NO X. To prevent such damage, verify that the front panel ozone ON/OFF switch is turned off if flow of feed gas to the air inlet is terminated. Suitably pressurized air or oxygen from an external cylinder is supplied to the rear panel AIR inlet. Immediately downstream from the inlet, a regulator and associated gauge provide pressure adjustment and indication for air or oxygen. Proper pressure setting. and the corresponding flow to the ozone generator, will depend on sample NO X level. Refer to Table 3-1, page 3-2. 1. For operation at NO X levels below 1,000 ppm, correct air pressure setting on OZONE Pressure Gauge is 20 psig (138 kpa), resulting in a flow of approximately 500 cc/min of air to the ozone generator. 2. For operation at NO X levels in the range of 1,000 to 2,500 ppm, air pressure setting on OZONE Pressure Gauge should be 30 psig (207 kpa), resulting in a flow of approximately 1 liter per minute of air to the ozone generator. 3. For operation at NO X levels in the range of 2,500 to 10,000 ppm, oxygen pressure setting on OZONE Pressure Gauge should be 30 psig (207 kpa), resulting in a flow of approximately 1 liter per minute of oxygen to the ozone generator. Within the ozone generator, a portion of the oxygen in the stream is converted into ozone by exposure to ultraviolet radiation from a lamp source. From the generator, the flow passes into the reaction chamber for use in the chemiluminescence reaction. c. Flow System Operating Modes Downstream from the BYPASS Needle Valve, the sample or standard gas flows through either the FLOW BALANCE Needle Valve or the NO 2 to NO converter, depending on the status of NO/NO X Mode Switching Solenoid Valve K3. Status of this valve depends on whether the front panel MODE Switch is in NO or NO X position. Operation for NO and NO X determination is as follows: NO Determination Sample or standard gas is supplied to the rear panel SAMPLE inlet, passes in turn through the BYPASS Needle Valve, NO/NO X Mode Switching Solenoid Valve K3, the sample capillary, and on to the reaction chamber. Since the NO 2 to NO converter is bypassed, any NO 2 component originally present in the sample remains in the NO 2 form and does not contribute to the output signal. (During startup, the FLOW BALANCE Needle Valve is adjusted to equalize flow rated for NO and NO X flow modes.) Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-3

Instruction Manual Model 951A NO X Determination Sample or standard gas supplied to the rear panel SAMPLE inlet passes in turn through the BYPASS Needle Valve, the NO 2 to NO converter, the solenoid valve, the sample capillary and on to the reaction chamber. With the sample now routed through the converter, the NO 2 component is converted into NO prior to entry into the reaction chamber, and thus will contribute to the output signal. During startup, the FLOW BALANCE Needle Valve is adjusted to equalize flow rates for NO and NO X flow modes. Criterion for correct adjustment is that an NO 2 free nitric oxide standard gas should give equal readings of ppm for NO and NO X modes. An alternate method of checking flow balance, usable if NO 2 free standard gas is unavailable, is to verify that the reading on the SAMPLE Pressure Gauge is the same for NO and NO X modes. If not, adjust FLOW BALANCE Needle Valve as required. d. Converter Bleed Flow Even in converter bypass condition (i.e., de-energized state of the solenoid valve), a flow of approximately 100 cc/min of sample or standard gas passes continuously through the converter and the bleed capillary. This bleed flow provides faster response and ensures that the sample within the converter is current. Bleed flow need not be precisely 100 cc/min. However, an excessively low bleed flow will result in impaired response. Conversely, use of excessively high bleed flow may result in failure to obtain the required flow of 60 cc/min for the reaction chamber. Thus the nominal value of 100 cc/min represents a compromise between response time requirements and sample supply. 4-3 ELECTRONIC CIRCUITRY Photons emitted during the chemiluminescence reaction (Section 4-1a) impinge on the end-on photomultiplier tube, generating a signal current of approximately 3 x 10-10 amperes per ppm concentration of NO in the reaction chamber. This current is measured by an electrometer amplifier. A functional diagram of the electronic signal circuitry is shown in Figure 4-2, page 4-5. For a more detailed presentation of the signal circuitry, plus control circuitry for actuation of NO/NO X Mode Switching Solenoid Valve, etc., refer to the overall schematic and pictorial diagrams in the rear of this manual, respectively. Individual circuit boards plug into the Interconnect Board, shown in Figure 6-2, page 6-4. As shown in Figure 4-2 (page 4-5), the anode of the photomultiplier tube is connected to the inverting input of AR1, and integrated circuit MOSFET amplifier with extremely high input impedance to minimize loading effects, and comparatively low output impedance to match the following stage. The signal from AR1 is directed to integrated circuit amplifier AR2 for additional gain. The front panel ZERO Control applies an adjustable zero-biasing signal to the input of AR1, to permit setting the zero point on the meter scale or recorder chart. For adjustment of overall sensitivity, the current measuring circuitry utilizes two front panel controls: 1. The PPM RANGE Switch, associated with AR1, provides a choice of seven fullscale sensitivities, obtained by a combination of high/low feedback selection and output attenuation. Potentiometers R31, R32, R33 and R34 are adjustable to provide inter-range correlation of exactly adjusted integral multiples. Refer to Section 5-1c, Step 3. 2. The SPAN Control, associated with AR2, provides continuously variable gain through feedback adjustment. A voltage divider network associated with AR2 provides a selectable output for a potentiometric recorder. The following sections describe electronic circuitry of individual circuit boards. 4-4 Theory Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual +15 ZERO PHOTOMULTIPLIER TUBE - AR1 TP1 R33 TP2 + R16 REACTION CHAMBER - HIGH VOLTAGE POWER SUPPLY + + R6 R32 R31 AR2 - S P A N 5V 1V R34 PPM RANGE R22 100mV 10mV ZERO AND SPAN POTENTIOMETERS LOCATED ON FRONT PANEL. R6, R16, R22, R31, R32, R33 AND R34 LOCATED ON AMPLIFIER BOARD AND ACCESSIBLE THROUGH HOLES IN AMPLIFIER BOARD COVER. + REC - Figure 4-2. Functional Schematic Diagram of Electronic Signal Circuitry a. Amplifier Board and Associated Circuitry The Amplifier Board provides two stages of amplification utilizing integrated circuit amplifiers AR1 and AR2. Functions Associated with AR1: AR1 Zero Adjust Potentiometer R6: This screwdriver adjusted trimming potentiometer, accessible through a hole in the amplifier cover, is used to eliminate voltage offset with AR1. When input signal is zero, output signal also should be zero. Refer to Section 5-1, page 5-1 for proper adjustment procedure. Front Panel Zero Control: This control applies an adjustable zero-biasing signal to the input of AR1. to permit setting the zero point on the meter scale or recorder chart. The adjustment compensates for a residual amplifier input signal resulting from a combination of background signal and photomultiplier dark current. Front Panel PPM RANGE: Switch and Associated Resistor Network. The PPM RANGE Switch controls sensitivity by a combination of two methods: Selection of high or low feedback resistance via a transistor switching circuit. Potentiometer R34, adjustable by insertion of a screwdriver through a hole in the amplifier cover, permits establishing exactly the correct ratio between the high and low feedback resistances. Output attenuation via resistors associated with relays U1, U2 and U3 on the amplifier board. Potentiometers R31, R32 and R33, adjustable by insertion of a screwdriver through the corresponding holes in the amplifier cover, permit inter-range correlation of exactly adjusted integral multiples. Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-5

Instruction Manual Model 951A Functions Associated with AR2: AR2 Zero Adjust Potentiometer R16: This screwdriver adjusted trimming potentiometer, accessible through a hole in the amplifier cover, is used to eliminate voltage offset with AR2. Front Panel SPAN Control: Provides continuously variable adjustment of AR2 gain. Response Time Selection Switch SW2: Provides choice of fast or slow response: With switch in FAST position, electronic response time (for 0 to 90% of fullscale) is approximately one second for all ranges except 10 ppm, approximately three seconds for 10 ppm range. For such applications as monitoring of stack sources, where slow response is desired, switch is placed in SLOW position. Electronic response time (0 to 90% of fullscale) is then approximately 10 seconds for all ranges. Closure of the switch connects capacitors C7 and C8 across resistor R17, thus increasing the RC time constant. Recorder Output Selection Switch SW1: The desired recorder output is obtained by closing the corresponding contact on Switch SW1, thus selecting the appropriate tap on a voltage divider connected across the output of AR2. Meter Span Adjustment Potentiometer R22: Potentiometer R22, adjustable by insertion of a screwdriver through the hole in the amplifier cover, permits adjusting the fullscale span of the meter so that meter readout agrees with recorder readout. b. Valve Control Board, Front Panel Mode Switch and Associated Circuitry As shown in Drawings 649834 and 649835, actuation of NO/NO X Switching Solenoid Valve K3, associated with the NO 2 to NO converter, is controlled by front panel MODE Switch SW2 via the valve driver circuit on the Valve Control Board. With MODE Switch in NO position, the solenoid valve is de-energized, placing the converter in bypass condition. With switch in NO X position, the solenoid valve is energized, placing the converter in in-line condition. Drawing 656313 shows internal circuitry of the Valve Control Board. The valve driver circuit consists of a single-pole, single-throw relay and associated TRIAC element. The relay is used only to switch the gate current for the TRIAC; therefore, power carried by the relay contacts is very small. Each relay is connected to AC neutral while the associated TRIAC is connected to 115 VAC. When the relay closes, in response to closure of the associated contacts on the MODE Switch, gate current flows through the relay contacts and causes the TRIAC to fire. Resistance of the TRIAC becomes very small, causing application of 115 VAC across the solenoid. When the relay contacts open, the gate current is interrupted and the TRIAC stops conducting. Thus the relay controls the power applied to the solenoid. c. ±15 Volt Power Supply The ±15 Volt Power Supply provides power for the various circuits. As shown in Drawing 619710, power transformer T1 has three secondarys that are used as follows: 38 VAC center-tapped secondary - Powers both 15 volt supplies through diode bridge CR1 and filter capacitors C1 and C4. The adjustable positive regulator, VR1, is set by voltage divider R1, R2 and R3; and its output is applied to pin A of the circuit board and to test point TP1. Potentiometer R2 should be adjusted to +15.5 VDC ±50 mvdc. The negative DC, regulated by VR2, is applied to pin D of the circuit board. 4-6 Theory Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual The center tap is the common reference for both the +15 and the -15 volt supplies and is applied to pin R of the circuit board and test point TP2. Both outputs are used for individual amplifiers on the various circuit boards and for the zero-biasing circuit associated with the front panel ZERO Control (Section 4-3a, page 4-5). 90 V secondary - Not used in this instrument. 9.5 VAC secondary - Drives a +5 V supply not used in this instrument. d. High Voltage Power Supply As shown in Drawings 649834 and 649835, DC high voltage for the photomultiplier tube is provided by the 630922 High Voltage Power Supply. It operates from the ±15 Volt Power Supply (Section 4-3c, page 4-6). Drawings 652423 and 654348 shows internal circuitry of the high voltage power supply. It consists of a voltage regulator circuit utilizing non-inverting operational amplifier AR1 and a flux-switching DC to AC converter. Potentiometer R4 provides an adjustment range of approximately -650 V to -2100 V for the regulated DC voltage supplied to the photomultiplier. Nominal setting is for -900 V; however, during factory checkout of the individual instrument, R4 is adjusted as required for overall system sensitivity. If necessary, R4 is readjusted per Section 5-1e, page 5-3. e. Converter Temperature Control Board and Associated Elements The temperature control circuit for the NO 2 to NO converter consists of: Temperature sensor R3, a platinum resistance element mounted in the converter. CONVERTER ADJUST potentiometer R6 mounted on the front panel, Figure 2-2 page 2-4. Resistive heating element within a lace-on jacket. During startup, temperature of the platinum sensor is initially below the setpoint selected with the CONVERTER ADJUST potentiometer. Controller now switches to on condition, and applies power to the header via TRIAC element Q5. The controller remains in on condition until temperature of the platinum sensor has risen sufficiently to come within the range of the proportional band of the controller. The controller then switches alternately off and on during a two-second time base. to maintain the temperature of the heater at the setpoint. Clockwise rotation of the CONVERTER ADJUST raises the setpoint; counterclockwise rotation lowers it. Setpoint adjustment is explained in Section 3-3, page 3-4. NOTE When installing or reinstalling the lace-on healer jacket on the converter, make sure that the yellow fiberglass sleeve on the temperature sensor leads is exposed and is not enclosed within the jacket. Otherwise, the insulation may burn off the sensor leads, causing short-circuiting of the sensor and temperature runaway or the converter heater. Converter Temperature Readout Thermocouple The Converter has an associated Type J thermocouple to permit temperature monitoring. The thermocouple connections, i.e., terminals 5(+) and 6(-) on TB9, are used to measure converter temperature in the range of 660 o F to 750 o F (350 o C to 400 o C). The optimum temperature differs from one analyzer to another. Refer to Section 3-3, page 3-4. Rosemount Analytical Inc. A Division of Emerson Process Management Theory 4-7

Instruction Manual Model 951A NOTE Converter temperature is not a direct measure of conversion efficiency. Temperature measurement is for reference purposes only. An exact determination of conversion efficiency is performed by the procedure given in Section 3-3, page 3-4. f. Fan Control Circuit The fans in current production instruments run continuously. In older instruments, the fan control circuit consists of thermistor R4, TRIAC switching circuit and blower fan B. When internal temperature of the analyzer reaches approximately 100 o F (38 o C), the fan is turned on. g. Remote Operation Option The Remote Operation Option, Figure 2-5 (page 2-7), consists of a Plug-In Adapter Board with attached harness that terminates on a rear panel plate. The adapter board has two two-position slide switches: the MAN/AUTO MODE Switch and the MAN/AUTO RANGE Switch. The MAN/AUTO MODE Switch provides the choice of local or remote control of the NO/NO X Mode Switching Solenoid Valve, associated with the NO 2 to NO converter. With switch at MAN, the function is under control of the front panel MODE Switch (Section 4-3b, page 4-6). With switch at AUTO, the front panel MODE Switch is disabled; the function is then remotely controlled, by either the operator or a computer, via contact-closure signals applied to terminals on TB5. The MAN/AUTO RANGE Switch provides the choice of local or remote selection of operating range. With switch at MAN, range selection is under control of the front panel PPM RANGE Switch. With switch at AUTO, the front panel PPM RANGE Switch is disabled; range selection is remotely controlled by either the operator or a computer, via contact-closure signals applied to terminals on TB6. Control at TB5 and TB6 is accomplished by a contact closure from the terminal marked COM (for common) to the terminal marked with the name of the desired function. Contact closure requirements are 25 ma at 15 VDC. h. Range I.D. Option The Range I.D. Option provides contact closure signals that permit a computer or other external device to identify the position selected on the front panel PPM Switch. This cable is connected to PPM Switch SW1 and extends to connector J10 mounted on the rear of the case. 4-8 Theory Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual SECTION 5 ROUTINE SERVICING DANGER ELECTRICAL SHOCK HAZARD Do not operate without doors and covers secure. Servicing requires access to live parts which can cause death or serious injury. Refer servicing to qualified personnel. For safety and proper performance this instrument must be connected to a properly grounded three-wire source of power. WARNING ULTRAVIOLET LIGHT HAZARD UV light from the ozone generator can cause permanent eye damage. DO NOT LOOK DI- RECTLY AT THE UV SOURCE IN THE OZONE GENERATOR. Use of UV filtering glasses is recommended. WARNING LIGHT EXPOSURE The photomultiplier tube must not be exposed to ambient light. Unless appropriate precautions are observed, light can strike the tube upon removal of fittings from the reaction chamber. 5-1 SYSTEM CHECKS AND ADJUSTMENTS The following procedures may be used to determine the cause of unsatisfactory instrument performance, or to make adjustments following replacement of components. If a recorder is available, use it for maximum convenience and accuracy in the various tests. a. Meter Mechanical Zero With electrical power removed from the analyzer, connect a shorting jumper across the terminals of the front panel readout meter. Meter should read zero; if not, adjust screw at rear of meter as required. Remove jumper. b. Amplifier Zero Adjustments Refer to Figure 5-1 (page 5-2) for location of Amplifier Board components noted in the following procedure. 1. Open front panel to turn off high voltage; set SPAN Control at 1000; set ZERO Control at 1000. Then perform the following checks and adjustments: 2. With PPM RANGE Switch at 250, obtain zero reading on meter or recorder by screwdriver adjustment of R16, accessible through hole in cover of amplifier board. 3. With PPM RANGE Switch at 1000, obtain reading of 0 (±1 division) on meter or recorder by screwdriver adjustment of R6, accessible through hole in cover of amplifier board. 4. Repeat Step 1. If the specified zero readings are unobtainable by adjustment of the corresponding potentiometers, replace amplifier board. c. Inter-Range Attenuation Correlation adjustments Refer to Figure 5-1 (page 5-2) for location of Amplifier Board components noted in the following procedure. 1. During the following adjustments, observe the analyzer output on a digital voltmeter or strip-chart recorder. If neither is available, use the front panel meter. Perform the following adjustments: 2. Supply 10 ppm standard to analyzer. Set PPM RANGE Switch at 10. Obtain Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-1

Instruction Manual Model 951A fullscale reading by adjustment of front panel SPAN Control. 3. Move PPM RANGE Switch to 250. Obtain reading of 4% of fullscale by screwdriver adjustment of R33, accessible through hole in cover of amplifier board. 4. Move PPM RANGE Switch to 25. Obtain reading of 40% of fullscale by screwdriver adjustment of R31, accessible through hole in cover of amplifier board. 5. Move PPM RANGE Switch to 100. Obtain reading of 10% of fullscale by screwdriver adjustment of R32, accessible through hole in cover of amplifier board. 6. Supply 100 ppm standard to analyzer. Adjust front panel SPAN Control for fullscale reading. 7. Move PPM RANGE Switch to 1000. Obtain reading of 10% of fullscale by screwdriver adjustment of R34, accessible through hole in cover of amplifier board. d. Meter Fullscale Span Adjustment If a recorder is used, and has been properly zeroed, it should agree with the meter reading. If not, obtain agreement by screwdriver adjustment of R22, accessible through hole in cover of amplifier board (Figure 5-1, page 5-2). If agreement cannot be obtained, check recorder. If recorder is functioning properly, replace amplifier board. 10 100 MV REC 5 FAST SLOW 1 RESPONSE V OUTPUT R22 R34 CALIB R16 R6 INPUT R33 CALIB R31 CALIB R32 CALIB Figure 5-1. Amplifier Board 5-2 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual e. Overall Sensitivity Principal factors that determine overall sensitivity of the analyzer are: (a) sample flow rate to the reaction chamber, and (b) sensitivity of the photomultiplier tube. If specified fullscale readings are unobtainable by adjustment of the SPAN Control, sensitivity is subnormal. The cause of the reduced sensitivity may be in either the flow system (Section 5-2, page 5-5) or the electronic circuitry (Section 5-5, page 5-11). If either the High Voltage Board or the Phototube/Reaction Chamber Assembly has been replaced, a readjustment of R4 on the High Voltage Board will probably be required to obtain the correct overall sensitivity. Turn the adjustment in the appropriate direction: clockwise to increase photomultiplier high voltage and therefore sensitivity; or, counterclockwise to decrease voltage and sensitivity. Adjustment range is approximately -650 volts to -2100 volts for the regulated DC voltage applied to the photomultiplier. Nominal setting is for -1100 volts; however, the adjustment should be set as required for overall system sensitivity. f. Ozone Output Maximum concentration of NO or NO X which the analyzer will measure with linearity depends on both the ozone generator output and the sample flow rate. If the analyzer seems to be ozone limited at high concentrations of NO or NO X, possible causes are: incorrect sample flow rate, incorrect flow rate through ozone generator, malfunction in ozone generator or associated transformer. WARNING TOXIC GAS Use extreme caution in troubleshooting ozone generator. Ozone is toxic. CAUTION OZONE DAMAGE When instrument power is on, the ultraviolet source sample is energized, converting a portion of the oxygen contained with the ozonator into ozone. With normal flow through the ozonator, the ozonized air or oxygen is continuously swept though the ozonator and into the reaction chamber. If flow is stopped, however, the ozone will diffuse and will attack the sintered metallic restrictor element in the tee fitting at the upstream end of the ozonator. Operation of the analyzer with the restrictor element thus rusted will result in the following consequences: Reduced flow of air or oxygen through the ozonator, ozone deprivation within the reaction chamber and non-linear response of the analyzer to NO/NO X. To prevent such damage, verify that the front panel ozone ON/OFF switch is turned off if flow of feed gas to the air inlet is terminated. g. Background Current With zero air supplied to rear panel SAM- PLE inlet, and MODE Switch at either NO or NO X, excessive background current is evidenced by inability to obtain zero meter reading by adjustment of the ZERO Control. If background current is excessive, the cause must be found and corrected. The fault may be in either the electronic circuitry or the sample flow system. First, establish proper performance of the electronic circuitry. Turn off high voltage by opening door. Turn on analyzer power. Verify that ZERO Control and amplifier are functioning properly. Close door and check for the following: 1. Excessive photomultiplier dark current. To check, shut off all flow to the ozone generator. Supply cylinder air to rear panel SAMPLE inlet. With MODE switch at NO, note response on meter or recorder. If background is still excessive, possible causes are: (1) leak Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-3

Instruction Manual Model 951A age of ambient light to photomultiplier tube, or (2) defective photomultiplier tube. 2. Contamination of reaction chamber or sample flow system. Refer to Section 5-2, page 5-5. h. NO 2 to NO Converter Temperature Adjustment Use procedure of Section 5-1i, page 5-4. The FLOW BALANCE Control must now be adjusted to equalize pressure drops for the NO and NO X legs of the internal flow system. In the usual method, NO 2 free nitric oxide standard gas is supplied to the analyzer and the FLOW BALANCE Control is adjusted for equal reading of ppm for the NO and NO X modes. An alternate method of flow balance, useable if NO 2 free nitric oxide standard gas is unavailable consists of checking SAMPLE pressure reading for NO and NO X modes. i. Flow Balance Using Front Panel Meter Indication with NO 2 Free Nitric Oxide Standard Gas 1. Supply to the SAMPLE inlet a suitable NO 2 free nitric oxide span gas, for example 2500 ppm NO. And NO span gas may be used, since balance is independent of concentration. However, it is important that the span gas be entirely free of NO 2 ; otherwise, the supposed flow balance achieved will be erroneous. A NO span gas known to be free of NO 2 can be purchased as a certified standard gas. Set SAMPLE Pressure Regulator so that SAMPLE Pressure Gauge indicated the value appropriate to the desired operating range. Refer to Figure 3-3, page 3-8. Set BYPASS Needle Valve for reading of 2 liters per minute on BYPASS Flowmeter. 2. Supply to the AIR inlet suitable air or oxygen, depending on the operating range. Refer to Section 2-3, page 2-1. Adjust OZONE Pressure Regulator so that OZONE Pressure Gauge indicates the value appropriate to the desired operating range. Refer to Figure 3-3, page 3-8. 3. Set MODE Switch at NO. Leave ZERO Control at previous setting. Turn PPM RANGE Switch to a setting appropriate to the nitric oxide content of the particular span gas. Adjust SPAN Control for reading of approximately 90% on meter or recorder. 4. Move MODE Switch from NO to NO X ; compare meter or recorder readings for the two modes. The NO X reading should be equal to the NO reading. If the readings are within acceptable agreement, the flows are properly balanced. If the readings for the two modes are not in acceptable agreement, adjust the front panel FLOW BALANCE Control as required. Alternate Flow Balance Method, Using Sample Pressure Reading for NO and NO X Modes If NO 2 free nitric oxide standard gas is unavailable, verify that reading on SAMPLE Pressure Gauge is the same for NO and NO X modes. If not, adjust FLOW BALANCE Needle Valve as required. 5-4 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual 5-2 SERVICING FLOW SYSTEM To facilitate servicing and testing, the Model 951A has front, rear and top access. Refer to Drawings 649958 and 649842 in the rear of this manual for details of flow system. a. Sample Capillary 3-2), adjust SAMPLE Pressure Regulator to normal operating setting, and verify that flowmeter indicates appropriate flow. If flow is correct, restore analyzer to normal operation. If flow is low, the capillary requires cleaning; proceed with the following steps: Depending on when the instrument was made, one of two types of capillaries may be present: a) a black tubing with white Teflon inner sleeve, or b) a one-piece black Teflon tubing. If clogging of sample capillary is suspected, measure flow rate as described below: 1. Turn off instrument power and shut off all gases. 2. Cover and shade the fittings on the reaction chamber with a dark cloth or other light-shielding material. While maintaining the area properly darkened, remove the fitting associated with the sample capillary (red-banded tube) and place a cap over the open fitting to prevent entry of stray light. NOTE If the opened fitting is inadvertently exposed to ambient light, the instrument will temporarily give a highly noisy background reading. If so, this condition may be corrected by leaving the instrument on, with high voltage on, for several hours. 3. With instrument power off, supply suitable test gas (dry nitrogen or air) to rear panel SAMPLE inlet. 4. Connect a flowmeter to open end of sample capillary. Per Table 3-1 (page 5. On NO/NO X Mode Switching Solenoid Valve K3, remove fitting from port associated with sample capillary. 6. Clean capillary with alcohol, then purge with dry nitrogen for one minute. Reconnect capillary to K3. 7. With the photomultiplier still covered, slowly insert the free end of the capillary into the corresponding fitting on the reaction chamber. (If capillary is of the two-piece variety, the bare Teflon tube at the end of the capillary must be inserted blindly and rotated until it enters a channeled internal passage, which can be detected by feel. The end of the Teflon tube should not project into the interior of the chamber; it should be slightly below the surface of the inner face. If tube is too long, trim back slightly as required.) Push the capillary in until it bottoms against the internal fitting, then tighten fitting. NOTE After capillary cleaning, instrument readout may be noisy for a short time, but should then become quiet. However, if bare Teflon tube is improperly inserted in Step 7, noisy readout will persist. Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-5

Instruction Manual Model 951A b. Ozone Restrictor and Capillary CAUTION OZONE DAMAGE When instrument power is on, the ultraviolet source sample is energized, converting a portion of the oxygen contained with the ozonator into ozone. With normal flow through the ozonator, the ozonized air or oxygen is continuously swept though the ozonator and into the reaction chamber. If flow is stopped, however, the ozone will diffuse and will attack the sintered metallic restrictor element in the tee fitting at the upstream end of the ozonator. Operation of the analyzer with the restrictor element thus rusted will result in the following consequences: Reduced flow of air or oxygen through the ozonator, ozone deprivation within the reaction chamber and non-linear response of the analyzer to NO/NO X. To prevent such damage, verify that the front panel ozone ON/OFF switch is turned off if flow of feed gas to the air inlet is terminated. With instrument power off, supply suitable test gas (dry nitrogen or air) to rear panel AIR inlet. Cover photomultiplier housing with a dark cloth. At the fitting on the reaction chamber, disconnect the ozone tube (yellow banded) and place a cap over the open fitting to prevent entry of ambient light. Connect a flowmeter to open end of ozone tube. Adjust the OZONE Pressure Regulator so that OZONE Pressure Gauge indicates normal operating pressure, as listed in Table 3-1 (page 3-2); verify that test flowmeter indicates appropriate flow, as listed in Table 3-1. Subnormal flow indicates clogging in the flow path that supplies air or oxygen to the ozone generator. This path contains two pressure reducing elements: 1. A Restrictor, consisting of a metal fitting with internal fritted (metal membrane) restrictor, the fitting is upstream from the inlet port of the ozone generator. If the internal restrictor becomes plugged, the Restrictor must be replaced, as it cannot normally be cleaned satisfactorily. 2. A capillary connected from the outlet of the ozone generator to the ozone inlet of the reaction chamber (coded with a yellow dot). This capillary may be cleaned if desired. Remove the ozone capillary by a procedure similar to that given for the sample capillary in Section 5-2a (page 5-5), observing all cautions cited. Clean ozone capillary by same method used for sample capillary. Test for Ozone Efficiency Perform the following steps: 1. Span the instrument with sample at 5 lb. 2. Decrease the pressure in 1 lb. steps. The response should be equal steps. If the steps are not equal, there is an ozone deficiency. The ozone transformer lamp should be checked. 5-6 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual 5-3 NO 2 TO NO CONVERTER a. Replacing Converter replacement is recommended upon failure to achieve a conversion efficiency of 95% during the test of Section 3-3b, page 3-4 To remove and replace converter, refer to Figure 5-2, page 5-7. Within the outer lace on jacket heater, the converter tube is wrapped in aluminum foil, with the temperature sensor in contact with the glass A. EXPLODED VIEW converter tube. In removing foil, note sensor and thermocouple position to ensure correct repositioning during reassembly. Assemble sensor to replacement converter, as shown in Figure 5-2 (page 5-7), using new aluminum foil. NOTE When installing or reinstalling the lace on heater jacket on the converter, make sure that the yellow fiberglass sleeve on the temperature sensor leads is exposed and is not enclosed within the jacket. See Figure 5-2, page 5-7. B. ATTACHING TEMPERATURE SENSOR TO CONVERTER TUBE Lace-On Heater Thermocouple Temperature Sensor Converter Tube Connector Tube (2) Top of Converter Assembly Temperature Sensor NOx -to-no2 Converter Tube Thermocouple NOTES: Sensor leads to be outside heater jacket. Orient converter tube as shown and install sensor flush with outside edge of tube as shown. C. HEATER JACKET PROPERLY INSTALLED ON CONVERTER TUBE Temperature Sensor Leads Yellow Fiberglass Sleeving Exposed Outside of Jacket Thermocouple Lead Lace-On Heater Jacket Converter Tube Inlet/Outlet Connections Figure 5-2. NO 2 to NO Converter Assembly Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-7

Instruction Manual Model 951A 5-4 PHOTOMULTIPLIER/REACTION CHAMBER a. Cleaning Reaction Chamber CAUTION LIGHT DAMAGE Photomultiplier will develop temporary electronic noise if exposed to ambient light with high voltage on. Damage may be permanent if exposure occurs with high voltage off. A temporary noisy condition may be corrected by leaving instrument on, with high voltage on, for several hours. The required recovery time depends on intensity and duration of the exposure. Noise level on the most sensitive range usually drops to normal within 24 hours. CAUTION HAND PROTECTION Always wear surgical rubber gloves when handling reaction chamber. If sample gas is properly filtered, the reaction chamber should not require cleaning. In event of carryover or contamination, however, the chamber should be disassembled to permit cleaning the quartz window and the optical filter. The following procedure is recommended. 1. Turn off instrument power and shut off all gases. 2. Cover and shade the Reaction Chamber Photomultiplier Assembly, Figure 5-3 (page 5-9), with a dark cloth or other light shielding material. 3. At the front of the Reaction Chamber Photomultiplier Assembly, remove the three screws that secure the reaction chamber mounting bracket. See Figure 5-3 (page 5-9). The middle screw is a 1/4-turn quick-release type; the two adjacent screws are the conventional continuous thread type. 4. With reaction chamber now free, detach its three tube fittings. See Figure 5-3 (page 5-9). 5. Slide off and remove the reaction chamber from the housing. Take care not to let light enter. 6. Unscrew plastic end cap, Figure 5-4 (page 5-9), thus freeing the quartz window and the red plastic optical filter. 7. Clean the reaction chamber by the appropriate one of the following two methods, a or b. Method a is applicable in most cases. Alternate method b is applicable when the instrument has shown high residual fluorescence, as indicated by high residual currents on a zero gas and high differentials between zero gas readings obtained with the ozone lamp on and off. CAUTION SOLVENT USAGE For a solvent, use only distilled water, do not use alcohol or acetone. Standard Cleaning Procedure Using a stiff plastic bristle brush, such as a toothbrush, scrub the Teflon surface and gas ports of the reaction chamber with clean distilled water and Alconox detergent (634929). Alconox detergent is included in the shipping kit provided with the Model 951A NO/NO X Analyzer and is also available from Sargent-Welch Scientific Company under its Catalog Number S-19650-A. Using Alconox and clean, soft facial tissue (NOT industrial wipes), carefully clean the quartz window, vigorously flush reaction chamber and quartz window with clean distilled water. Blow out all possible water from internal passages of reaction chamber. Dry reaction chamber and quartz 5-8 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual window in a warm oven at 125 o F to 150 o F (52 o C to 66 o C) for 30 to 45 minutes or purge dry the parts with dry cylinder air or nitrogen to eliminate all moisture. Alternate Cleaning Procedure - For High Residual Fluorescence Holding the reaction chamber by the tube fittings, and using appropriate caution. immerse the white Teflon part of the chamber in 50% concentrated Reagent Grade HCl. After five minutes, rinse thoroughly with deionized water, then air dry as in method a above. Place parts in position and press on end cap so that mating threads engage properly, without cross-threading. Turn mating parts in one continuous motion until the parts bottom. Do not over torque. With reaction chamber now assembled, replace and reconnect it in the reverse order of that used for removal. Reaction Chamber Assembly Phototube Housing Phototube Magnetic Shield 1 Phototube Socket Assembly O-rings (1 or 2, depending on Phototube length. (Lubricate with silicone grease) 1 Proper orientation of photomultiplier in socket may be determined by measuring the resistance from the center pin to ground (should be approximately 11 megohm). Figure 5-3. Reaction Chamber/Photomultiplier Assembly Reaction Chamber Assembly Phototube Housing Optical Filter (Red Plastic) 876351 O-Ring 836486 O-Ring Cap Quartz Window Figure 5-4. Reaction Chamber Assembly and Phototube Housing Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-9

Instruction Manual Model 951A b. Photomultiplier Tube and Housing The photomultiplier tube operates at high DC voltages (nominal range 600 to 1200 volts) and generates small currents that are highly amplified by the signal conditioning circuitry. It is therefore important that ambient humidity and condensed water vapor be excluded from the interior of the photomultiplier housing. Ambient humidity can result in electrical leakage, evidenced as abnormally high dark current. Water vapor or condensed moisture in contact with the thermoelectrically-cooled photomultiplier may be evidenced by abnormally high noise level during instrument readout on zero air or upscale standard gas. The Reaction Chamber Photomultiplier Assembly, Figure 5-3 (page 5-9), incorporates several features for exclusion of humidity and moisture. The photomultiplier socket assembly is potted with a high impedance compound and is sealed from external influences with epoxy and rubber gasket material. The socket assembly and the reaction chamber are sealed, with o-rings, into opposite ends of the tubular photomultiplier housing. The socket end of the housing may be sealed with either one or two o-rings, depending on the length of the phototube. In addition, a rolled-up piece of blotter-like desiccant paper is inserted between the shield external to the housing. The desiccant paper ensures hermetic isolation of the housing despite possible minor leakage around the o-rings, and eliminates the effects of any humidity possibly trapped during assembly of the socket and the reaction chamber to the housing. If the photomultiplier housing is exposed to ambient air for a period of tie, the desiccant paper will absorb water. Before reassembly, the paper must then be heated in an oven at about 300 o F (149 o C) for at least two hours. Within the housing is a humidity indicator which turns pink in the presence of humidity. Teflon Liner in Lamp Housing of Ozone Generator WARNING MERCURY HAZARD Handle lamp assembly with extreme caution. Lamp contains mercury vapor. CAUTION LAMP DAMAGE Do not touch lamp. Fingerprints could cause a decrease in lamp output. A Teflon liner was used inside the ozone lamp housing during early production runs of the Model 951A. A bare lamp housing is now used in lieu of the liner. For replacement of lamp only, order Ozone Lamp Kit 658156 (includes O- Ring). For replacement of complete Ozone Generator, order Kit 654793. Refer to instructions provided with kit. 5-10 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual 5-5 SERVICING ELECTRONIC CIRCUITRY Refer to the appropriate schematic and pictorial diagrams in the rear of this manual when servicing. The electronic system, Figure 6-4 (page 6-6), utilizes plug in printed circuit boards with solid state components. After a malfunction is traced to a particular board, the recommended procedure is to return it to the factory for repair. The Customer Service Center Representative will arrange a Returned Goods Authorization. Rosemount Analytical Inc. A Division of Emerson Process Management Routine Servicing 5-11

Instruction Manual Model 951A CONV LED DS3 POWER SW4 ON OFF 18 WHT/BRN F1 (Top) 1 WHT/GRN 2 WHT/BLK TB4 9 WHT/ORN 11 WHT/RED 18 WHT/BRN 17 YEL NEUT 5 HOT 4 3 REC - 2 REC + 1 28 WHT 19 BLK 10 BLU 16 BLK 15 RED 3 GRN/YEL 19 BLK 5 PUR F2 K3 TB1 6 5 4 3 2 1 17 YEL 14 ORN 4 WHT 12 GRY 21 WHT 13 BRN 20 BLK 25 WHT 24 WHT TB2 21 WHT 26 WHT 6 HOT 5 NEUT S 15 16 BLK 15 RED 14 ORN 13 BRN 12 GRY 11 WHT/RED 10 BLU 649832 HARNESS 9 WHT/ORN 8WHT/ORN 7 WHT/BLU 6 YEL 5 PUR 4 WHT 3 GRN/YEL 1 WHT/GRN 2 WHT/BLK A 1 TB3 J7 6 YEL 23 BLK 28 WHT 8 WHT/ORN 7 WHT/BLU 6 5 4 3 2 1 CONV FAN THERM CONV 17 WHT 23 BLK 28 WHT 20 BLK 22 BLK 27 WHT 4 HOT 3 NEUT 2 HOT 1 NEUT 120 VAC WIRE NO. COLOR FROM TO 1 WHT/GRN DS3+ J7-1 2 WHT/BLK DS3 J7-A 3 GRN/YEL GND J7-C 4 WHT TB1-5 J7-D 5 PUR F2 Side J7-E 6 YEL TB3-6 J7-F 7 WHT/BLU TB3-1 J7-H 8 WHT/ORN TB3-2 J7-J 9 WHT/ORN R6 Slider J7-8 10 BLU TB4-3 J7-K 11 WHT/RED R6 CCW J7-9 12 GRY TB1-4 J7-L 13 BRN TB1-1 J7-M 14 ORN TB1-6 J7-P 15 RED TB4-1 J7-R 16 BLK TB4-2 J7-S 17 YEL F2 End SW4 18 WHT/BRN F1 Side SW4 19 BLK F1 End TB4-4 20 BLK F2 End TB2-2 21 BLK TB1-2 TB2-5 22 BLK TB2-2 TB2-6 23 BLK TB2-2 TB3-5 24 WHT TB1-2 TB1-3 25 WHT TB1-3 TB1-5 26 WHT TB2-3 TB2-5 27 WHT TB2-1 TB2-3 28 WHT TB2-1 TB3-4 29 Bare (Tinned) R6 Slider R6 CW Figure 5-5. Terminal Chassis Assembly Wiring Diagram 5-12 Routine Servicing Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual SECTION 6 REPLACEMENT PARTS The following parts are recommended for routine maintenance and troubleshooting of your instrument. If the troubleshooting procedures do not resolve the problem, contact your local service office. 6-1 CIRCUIT BOARD REPLACEMENT POLICY In most situations involving a malfunction of a circuit board, it is more practical to replace the board than to attempt isolation and replacement of the individual component, as the cost 6-2 MATRIX of test and replacement will exceed the cost of a rebuilt assembly. As standard policy, rebuilt boards are available on an exchange basis. Because of the exchange policy covering circuit boards, the following list does not include individual electronic components. If circumstances necessitate replacement of an individual component, which can be identified by inspection or from the schematic diagrams, obtain the replacement component from a local source of supply. 951A MODEL 951A NO/NO X ANALYZER Code Ranges 2 01 Standard Seven - 10 to 10,000 ppm 99 Special Code Remote Control/ID 00 None 01 Range I.D. 02 Remote Range Change 03 Isolated Remote Range Change 04 Range I.D. with Remote Range Change Code Output 01 Standard Voltage 02 Current 4-20 ma 99 Special Code Power Requirement 01 117V, 50/60 Hz 951A 01 00 01 01 01 01 Example Code Options 1 01 Standard Version 02 Pump (115 VAC, 60 Hz)1 03 Standard Version with Fan Thermostat 04 Pump (115 VAC, 60 Hz)1 with Fan Thermostat 05 Low Temperature Coefficient 99 Special Code Case 01 Standard 02 Standard w/ Tropicalization 99 Special Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-1

Instruction Manual Model 951A 6-3 REPLACEMENT PARTS 621095 ±15V Power Supply 649815 Amplifier Board 652842 Ballast Transformer 623719 Capillary, Sample 623743 Capillary, Pump 634398 Capillary, Bleed 630960 Coaxial Cable, Amplifier 632783 Converter Assembly 861273 Fan 809374 Fuse 3/4 Amp 777010 Fuse 5 Amp (Package of 15) 658992 High Voltage Power Supply Board 630920 Interconnect Board 630931 Meter 654793 Ozone Generator 658156 Ozone Generator Lamp 657842 Power Supply Board 632748 Pump (Option) 635664 Pump Repair Kit 636343 Reaction Chamber 902315 Signal Converter 4-20mA (Option) 641620 Temperature Control Board 624593 Valve Control Board 6-2 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual Pneumatics (see Figure 6-5) Pump (Option) Ozone Generator Interlock Switch Fan Meter Power Switch Ballast Transformer Power Switch Shield Converter Assembly (see Figure 6-6) Figure 6-1. Model 951A Exploded View Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-3

Instruction Manual Model 951A Electronics (see Figure 6-4) Terminal Chassis Assembly Coaxial Amplifier Cable Thermocouple Connector Reaction Chamber Figure 6-2. Model 951A Exploded View (continued) 6-4 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual Figure 6-3. Model 951A Exploded View (continued) Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-5

Instruction Manual Model 951A High Voltage Power Supply Board Amplifier Board ±15V Power Supply Interconnect Board Valve Control Board Temperature Control Board Figure 6-4. Electronics 6-6 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual a. Pneumatics 638614 Pressure Gauge, Ozone 644055 Pressure Gauge, Sample 839906 Flowmeter, Bypass 815187 Backpressure Regulator, Sample 634993 Needle Valve, Bypass 888692 Pressure Regulator, Ozone 652146 Tee Fitting Assembly 107542 Metering Valve, Flow Balance Metering Valve Needle Valve Sample Backpressure Regulator Sample Pressure Gauge Tee Fitting Assembly Ozone Pressure Regulator Ozone Pressure Gauge Bypass Flowmeter Figure 6-5. Front Panel Pneumatic Components Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-7

Instruction Manual Model 951A b. Converter Assembly 632782 Sensor 632784 Connector, Tube 632795 Tube, Packed - Aged 617124 Heater Jacket 623848 Thermocouple, Type J 876776 Solenoid Valve - 3-Way 879948 Nut, Teflon 3/32 Connector Tubes Sensor 3-Way Solenoid Valve Heater Jacket Teflon Nut 3-32 Type J Thermocouple Figure 6-6. Converter Components 6-8 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual Inlet Outlet O-Ring Ozone Lamp Assembly Connector Plug For replacement of lamp only, order Ozone Lamp Kit 658156 (includes O-Ring). For replacement of complete Ozone Generator, order Kit 654793. Refer to instructions provided with kit. Figure 6-7. Ozone Generator Components Rosemount Analytical Inc. A Division of Emerson Process Management Replacement Parts 6-9

Instruction Manual Model 951A c. Low Tempco Option 652831 Power Supply Assembly 654831 Reaction Chamber 861040 Sensor, Cooler Temperature 884791 Thermistor 654943 Thermoelectric Cooler Assembly 652836 Transformer, Thermoelectric Cooler TOROID TRAN SFORM ER INDUCTOR BALLAST OZONE GEN ERATOR THERM O COOLER POW ER SUPPLY THERM O COOLER INDICATOR FAN CON VERTER ASSEM BLY BAFFLE CHASSIS ASSEM BLY, TERM INAL THERM O COOLER ASSEM BLY ELECTRON ICS ASSEM BLY Figure 6-8. Tempco Retrofit Component Location 6-10 Replacement Parts Rosemount Analytical Inc. A Division of Emerson Process Management

Model 951A Instruction Manual SECTION 7 RETURN OF MATERIAL 7-1 RETURN OF MATERIAL If factory repair of defective equipment is required, proceed as follows: 1. Secure a return authorization from a Rosemount Analytical Inc. Sales Office or Representative before returning the equipment. Equipment must be returned with complete identification in accordance with Rosemount instructions or it will not be accepted. 2. In no event will Rosemount be responsible for equipment returned without proper authorization and identification. 3. Carefully pack the defective unit in a sturdy box with sufficient shock absorbing material to ensure no additional damage occurs during shipping. 4. In a cover letter, describe completely: a. The symptoms that determined the equipment is faulty. b. The environment in which the equipment was operating (housing, weather, vibration, dust, etc.). c. Site from where the equipment was removed. d. Whether warranty or non-warranty service is expected. e. Complete shipping instructions for the return of the equipment. 5. Enclose a cover letter and purchase order and ship the defective equipment according to instructions provided in the Rosemount Return Authorization, prepaid, to: If warranty service is expected, the defective unit will be carefully inspected and tested at the factory. If the failure was due to the conditions listed in the standard Rosemount warranty, the defective unit will be repaired or replaced at Rosemount s option, and an operating unit will be returned to the customer in accordance with the shipping instructions furnished in the cover letter. For equipment no longer under warranty, the equipment will be repaired at the factory and returned as directed by the purchase order and shipping instructions. 7-2 CUSTOMER SERVICE For order administration, replacement parts, application assistance, on-site or factory repair, service or maintenance contract information, contact: 7-3 TRAINING Rosemount Analytical Inc. Process Analytic Division Customer Service Center 1-800-433-6076 A comprehensive Factory Training Program of operator and service classes is available. For a copy of the Current Operator and Service Training Schedule, contact the Technical Services Department at: Rosemount Analytical Inc. Phone 1-714-986-7600 FAX 1-714-577-8006 Rosemount Analytical Inc. Process Analytic Division Customer Service Center 1-800-433-6076 Rosemount Analytical Inc. A Division of Emerson Process Management Return of Material 7-1

Instruction Manual Model 951A 7-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management

Model 864,865,867,951A,OM-11A Addendum 246510-D December 1997 ADDENDUM Addendum to the following instruction manuals: INSTRUMENT MANUAL PART NUMBER KIT 864/865 Infrared Analyzers 556872 619624 867 Infrared Analyzer 555986 619626 951A NO/NOx Analyzer 556383 619627 OM-11EA Oxygen Analyzer 555866 619613 24VDC Remote Range Select and Range I.D. Kit (Option) The 24 VDC remote range option is a combined remote range select/i.d. system that is compatible with a user supplied remote control system employing a 24 VDC digital output and digital input for analyzer range control and analyzer range sense respectively. It is completely integral within the analyzer and provides a terminal strip on the rear of the instrument for connection of the user's cable. The system consists of the connector block which contains a relay interface/isolation board as well as the terminal strips, a harness, and interconnect board and a modified rotary range switch on the front panel. The circuit is shown schematically in Figure 2. Specifications for the system are as follows: Range Select Signal: 18 to 20 VDC at 15 ma Relay Coil Resistance: 1750 ohms Relay Contact Rating: 2A, continuous Connections The terminal block is shown in Figure 1. The terminal strip on the left is for the user's digital output (range select) connections. Two terminals receive the 24 VDC and return continuously while only one of the upper seven terminals at a time receives the +24 VDC signal for range selection. For instruments having fewer than seven ranges, fewer than seven terminals are used. The terminal strip is also for the user's digital input (range I.D.) connections. The +24 V signal appears on the appropriate one of the upper seven terminals to indicate the selected range of the instrument. The +24 V signal also appears on the bottom terminal, marked AUTO, but only when the AUTO range select mode is activated. Operation The range switch on the front panel of the instrument may be used for normal, manual range selection. In this mode, remote selection is deactivated but remote I.D. is still active, providing that the 24 VDC and return are connected to the lower left terminals on the back of the instrument. When the range switch on the front panel of the instrument is placed in the AUTO position, the remote range selection function is activated and +24 VDC appears on the AUTO terminal. http://www.processanalytic.com