Victoreen and

Similar documents
Victoreen B. Operators Manual. Image Intensifier Ion Chamber

RADIOLOGY/ FLUOROSCOPY 15 RAD-CHECK PLUS X-RAY EXPOSURE METER

Nuclear Associates

DPC-30 DPC-100. Reference Manual

! Warning, refer to accompanying documents.

E X R A D I N Spherical Ion Chambers A3. A4 REF DOC #

User s Guide Temperature Sensor Converter TSC-599

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

Operator s Manual 701 Test Kit for Static Control Surfaces (Megohmmeter)

Advanced Test Equipment Rentals ATEC (2832)

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

EC214 EC215 - EC215R Bench Conductivity Meters

Powers Controls RC 195 Multiple Input Receiver- Controller

Operation Manual. Pro CO 2 Analyzer. Carbon Dioxide Analyzer. Rev

SC505 Sof Care Inflator. Operating Instructions and Service Manual

User Manual PTW-DensiX. D / Di/Zi

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Overview. Front Panel: Keypad and Display

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

Constant Pressure Inlet (CCN) Operator Manual

Instruction Manual BA90 Precision Barometer

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-DPX Operation Manual

Instruction Manual EMA 84 Digital Pressure Gauge

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale

HI 2314 HI 2315 HI 23151

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Model 130M Pneumatic Controller

E X R A D I N. Parallel Plate Ion Chambers User Manual DOC #

RADIATION PROCEDURES MANUAL Procedure Cover Sheet

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User Manual. Pro CO LP Alarm. Carbon Monoxide Analyzer and Low Pressure Monitor with Alarm REV: 03.19

Exposure System Selection

ANALOX O2 PORTABLE OPERATORS MANUAL

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

DTG - LCD Digital Temperature Gauge USER MANUAL

Sensoric 4-20 ma Transmitter Board Operation Manual

Columbus Instruments

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

The Ins and Outs of I/P Transducers

A3 / A4 / A5 / A6 / A8 Spherical Ionization Chambers REF / / / / User Manual

KERN EG/EW Version /02

Datasheet: K-30 ASCII Sensor

Powers Controls RC 195 Single Input Receiver-Controller

Technical Data Sheet MF010-O-LC

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

StockWeigh 500 D3632 US-B

RAM Operation Manual

Heavy Duty Dissolved Oxygen Meter

VISO Manual. LARSEN & BRUSGAARD Mosevej Kirke Hyllinge, Denmark Phone: Fax:

I/P Transducer Positioner Module

Instruction and Maintenance Manual

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor

Digital Melting Point Apparatus

American Weigh Scales, Inc. BLADE Series D i g i t a l P o c k e t S c a l e BLADE-550. User Manual BLADE-400 BLADE-1000

Operation Manual. Pro O 2 Alarm TM. Oxygen Analyzer. Rev

Technical Information Conducal CLY421

Minnesota Rules, Chapter 4732 X-ray Revision

The Univentor 1250 Anaesthesia Unit

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

OEM Manual. MODEL ½ Digit DRUM SCALE

O2 Portable. User Manual

INSTALLATION & MAINTENANCE INSTRUCTIONS

Operation Manual. Pro CO Analyzer TM. Pro CO with High Temp Alarm Carbon Monoxide Analyzer

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

GasSense NDIR User Manual

SDX Submersible Depth Transmitter User Manual

Research Grade Blood Pressure Transducer User's Manual

ASDX Series Silicon Pressure Sensors

MODEL 100 NITROGEN INFLATION CART

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

SDX Submersible Depth Transmitter User Manual

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Do Not Print This Page.

Instruction Manual. PLT Series. 1. General Specification. 2. Installation. 3. Wiring. 4. Function. 5. Setting & Calibration

Instruction Manual. Digital Thermocouple Gauge Control Units Models DTC & DTC Models DTC-06M-115 & DTC-06M-230

DRAFT U S E R M A N U A L CAUTION. Model: 19MFA1001 Series. Federal (USA) law restricts this device to sale by or on the order of a physician.

RC 195 Receiver-Controller

RAM 4021 Operation Manual

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark

Radiation Producing Machines Radiation Safety Training for X-ray Diffraction Units

Stand-Alone Bubble Detection System

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

WARNING! Jump Number Main Window Exit Altitude Alti-Meter Mode Deployment Altitude Speed-Meter Mode...

Transmitter CS 21 Operation Manual

In Vivo Scientific, LLC INSTRUCTION MANUAL

Operating instructions Safety Rope Emergency Stop Switches ZB0052 / ZB0053 ZB0072 / ZB0073

42045 Heavy Duty ADA Base Model Kit: 85/105 PSI (ADA Compressor Only) Heavy Duty ADA Base Model Kit: 110/145 PSI (ADA Compressor Only)

Electropneumatic Transducer (I/P) Installation, Operation and

Pressure & Vacuum Measurement. Series 370. Solutions. Stabil-Ion Vacuum gauge and Controller

NOVALYNX CORPORATION MODEL WEATHER STATION INSTRUCTION MANUAL

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in:

Transcription:

Victoreen 06-526-5240 and 06-526-5242 RAD-CHECK MICRO-R Operators Manual February 2005 Manual No. 06-526-5240-1 Rev. 2 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies

Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms

Table of Contents Section 1: General Information... 1-1 1.1 Product Description... 1-1 1.2 Specifications... 1-1 1.3 Receiving Inspection... 1-2 1.4 Storage... 1-2 1.5 Routine Cleaning... 1-3 1.6 Procedures, Warnings and Cautions... 1-3 Section 2: Operation... 2-1 2.1 Connecting the Chamber... 2-1 2.2 Applying Power... 2-1 2.3 Exposure Mode... 2-1 2.4 Exposure Rate Mode... 2-2 2.5 Applications... 2-2 2.5.1 Radiographic Output (mr/mas) Dose Measurements... 2-2 2.5.2 Determining Minimum Filtration Requirements (Beam Quality - HVL)... 2-3 2.5.3 mas Reciprocity... 2-4 2.5.4 Fluoroscopic Exposure Rate... 2-4 2.6 Accuracy Considerations... 2-5 2.7 Sources of Error... 2-7 2.8 Precautions... 2-7 2.9 Operational Checks... 2-7 2.10 Drift Rate... 2-7 2.11 Battery Voltage... 2-7 2.12 Zero Set... 2-8 Section 3: Service Information... 3-1 3.1 General... 3-1 3.2 Circuit Description... 3-1 3.3 Calibration and Adjustments... 3-1 Section 4: Troubleshooting... 4-1 4.1 Precautions... 4-1 4.2 Troubleshooting... 4-1 Appendix A: Replacement Parts...A-1 A.1 Replacement Parts... A-1 i

(Blank page)

General Information Product Description 1 Section 1 General Information 1.1 Product Description The RAD CHECK MICRO-R is a battery operated, portable unit which measures the output radiation of diagnostic x-ray equipment. The exposure is displayed on a 3 ½ digit Liquid Crystal Display (LCD) as either: 1. Exposure, Roentgens (0.01 mr to 19.99 R) or SI units of milligrays (0.1 μgy to 199.9 mgy). (Refer to specifications for ranges.) 2. Rate, in Roentgens per minute (0.1 mr/min to 199.9 R/min.) or SI units of milligrays per minute (1 μgy/min to 1999.0 mgy/min). (Refer Section 1.2, Specifications, for ranges.) Radiation is detected with an image intensifier ionization chamber P/N 06-524-3000 with a 15-foot cable assembly. 1.2 Specifications Ranges: Model 06-526-5240: (Dose) Low Range 00.00-19.99 mr; Resolution: 0.01 mr (Dose) High Range 00.00-19.99 R; Resolution: 0.01 R (Rate) Low Range 000.0-199.9 mr/min; Resolution: 0.1 mr/min (Rate) High Range 000.0-199.9 R/min; Resolution: 0.1 R/min Model 06-526-5242: (Dose) Low Range 000.0-199.9 μgy; Resolution: 0.1 μgy (Dose) High Range 000.0-199.9 mgy; Resolution: 0.1 mgy (Rate) Low Range 0000-1999 μgy/min; Resolution: 1.0 μgy/min (Rate) High Range 0000-1999 mgy/min; Resolution: 1.0 mgy/min Standard Cal.: Reproducibility: Electrometer Drift: Maximum Exposure Rate: Manual Reset: At 75 kvp with 4 mm AI filtration at 22 C and one atmosphere using Model 06-524-3000 chamber (chamber optional). Within 2% short-term over 100 mr to 2 R range. (1 mgy to 20 mgy) <2 digits in 5 minutes in exposure dose mode. Min. 90% collection at 20 R/sec. Resets display to zero. (Dose Mode only) 1-1

Victoreen 06-526-5240 and 06-526-5242 Operators Manual Operating Conditions: 10-40 C max.; 90% relative humidity, non-condensing Display: 3 ½ digit LCD, ½" high, low-battery indicator Controls: Display zero reset button, dose or dose rate output selector, and On/Off switch. Tilt stand to adjust unit visibility. Power: 9 V Alkaline battery, Duracell MN1604 or equal >50 hour life Size: 2.75" high x 6" wide x 6.25" deep. Net 18 oz. (7 cm x 15.25 cm x 15.9 cm, Net 0.51 kg) Accessories Available: 06-524-3000: Image intensifier Ion Chamber, 100 cc volume, energy response: within 7% from 15 to 65 kvp (30 to 150 kvp); 15 ft. (4.5 meter) cable; 5.5 in. (14 cm) diameter x.72 in. (1.8 cm) thick. Nominal sensitivity: 30 nc/r. 89-525: Carrying and Storage Case. Holds "RAD-CHECK MICRO-R" and accessories Figure 1-1. Model 06-526-5240 RAD-CHECK MICRO-R 1.3 Receiving Inspection Upon receipt of the package: 1. Inspect the cartons (s) and contents for damage. If damage is evident, file a claim with the carrier and notify Fluke Biomedical at 440.248.9300. 2. Remove the contents from the packing material. 3. Verify that all items listed on the packing list have been received and are in good order. 1.4 Storage If the unit is to be stored prior to use, pack it in the original container if possible, and store in an environment free of corrosive materials, fluctuations in temperature, humidity, vibration and shock. Remove batteries prior to storage. Refer to Section 2.11, Battery Voltage, for further details. 1-2

General Information Routine Cleaning 1 1.5 Routine Cleaning Do not immerse the RAD-CHECK. The unit is not waterproof. Liquid could damage the circuits. The unit should be kept clean and free from dirt and contamination. The unit may be cleaned by wiping with a damp cloth, using any commercially available cleaning or decontaminating agent. 1.6 Procedures, Warnings and Cautions WARNING Extreme caution should be used when making connections with the chamber and rear panel, an electrical shock hazard exists on the ion chamber bias connector (HV out). NOTE In the event of a transient induced lockup of the unit. It is necessary to reset the unit by cycling the power (turning the unit off and back on). After power-up, the unit will be in its normal operating mode. The equipment described in this manual is intended to be used for the detection and measurement of ionizing radiation. It should be used only by persons who have been trained in the proper interpretation of its readings and the appropriate safety procedures to be followed in the presence of ionizing radiation. Although the equipment described in this manual is designed and manufactured in compliance with all applicable safety standards, certain hazards are inherent in the use of electronic and radiometric equipment. WARNINGS and CAUTIONS are presented throughout this document to alert the user to potentially hazardous situations. A WARNING is a precautionary message preceding an operation which has the potential to cause personal injury or death. A CAUTION is a precautionary message preceding an operation which has the potential to cause permanent damage to the equipment and/or loss of data. Failure to comply with WARNINGS and CAUTIONS is at the user's own risk and is sufficient cause to terminate the warranty agreement between Fluke Biomedical, Radiation Management Services and the customer. Adequate warnings are included in this manual and on the product itself to cover hazards that may be encountered in normal use and servicing of this equipment. No other procedures are warranted by Fluke Biomedical. It shall be the owner's or user's responsibility to see to it that the procedures described here are meticulously followed, and especially that WARNINGS and CAUTIONS are heeded. Failure on the part of the owner or user in any way to follow the prescribed procedures shall absolve Fluke Biomedical and its agents from any resulting liability. Indicated battery and other operational tests must be performed prior to each use to assure that the instrument is functioning properly. If applicable, failure to conduct periodic performance tests in 1-3

Victoreen 06-526-5240 and 06-526-5242 Operators Manual accordance with ANSI N323-1978 (R1983) Radiation Protection Instrumentation Test and Calibration, paragraphs 4.6 and 5.4, and to keep records thereof in accordance with paragraph 4.5 of the same standard, could result in erroneous readings or potential danger. ANSI N323-1978 becomes, by this reference, a part of this operating procedure. 1-4

Operation Connecting the Chamber 2 Section 2 Operation Extreme caution should be used when making connections with the chamber and rear panel, an electrical shock hazard exists on the ion chamber bias connector (HV out). 2.1 Connecting the Chamber WARNING 1. Connect the cable of the chamber to the rear panel connectors. a. The coaxial BNC on the rear panel is the signal input. b. The recessed banana plug is the high voltage connection for the chamber bias. 2.2 Applying Power RAD-CHECK MICRO-R performs to specifications immediately on application of power. However, a five-minute warm-up period is suggested to minimize the drift associated with surface charge distribution. 1. Turn the power switch to the ON position. a. If "LOW BAT" appears in the upper left corner of the display, the battery voltage is low. Refer to Section 2.11, Battery Voltage, for battery replacement procedures. b. If no display is present, the battery may be fully discharged. Refer to Section 2.11, Battery Voltage, for battery replacement procedures. 2.3 Exposure Mode NOTE 1. Set the front panel selector switch to DOSE. (Button is in the OUT position.) 2. Select High Exposure (19.99 R Max) or Low Exposure (19.99 mr Max.). 3. Using the collimator light field, adjust the field size so that the entire chamber is in the field. 4. Position the image intensifier ion chamber, Model 06-524-3000, so that the primary beam is centered on, and perpendicular to, the chamber. 5. Press the RESET button to zero the display; make the x-ray exposure. 6. Read the exposure in roentgens (R), high exposure, or in roentgens (μr), low "exposure. To minimize drift effects, readings should be recorded immediately. 2-1

Victoreen 06-526-5240 and 06-526-5242 Operators Manual Extreme caution should be used when making connections with the chamber and rear panel, an electrical shock hazard exists on the ion chamber bias connector (HV out). 7. When finished, turn the unit OFF before disconnecting the cables to the chamber. 2.4 Exposure Rate Mode 1. Perform 2.1 and 2.2. 2. Set the front panel selector switch to RATE (Button is in the IN position.). 3. Select High Exposure (button IN) or Low Exposure (button OUT). 4. Using the collimator light field, adjust the field size so that the entire chamber is in the field. 5. Position the chamber so that the primary beam is centered on, and perpendicular to the chamber. 6. The RESET button has no effect on the zero. Zero the instrument from the rear if necessary; make the x-ray exposure. 7. Read the exposure rate (R/min or mr/min). Extreme caution should be used when making connections with the chamber and rear panel, an electrical shock hazard exists on the ion chamber bias connector (HV out). 8. When finished, turn the unit OFF before disconnecting the cables to the chamber. 2.5 Applications WARNING WARNING The RAD-CHECK MICRO-R can be used to measure radiation output of diagnostic x-ray equipment. Several applications are discussed in the following paragraphs. In each of these applications, it is important that all data pertaining to tests performed be recorded for later comparison. Records should include the date, technique factors used, and readings obtained. 2.5.1 Radiographic Output (mr/mas) Dose Measurements 1. Set up x-ray field as follows: a. Using the collimator light field, adjust the field size so that the entire chamber is in the field. b. Position the chamber so that the primary beam is centered on, and perpendicular to the chamber. 2. Select the desired x-ray technique (kvp, ma, and time). 3. Make an exposure per Section 2.3, Exposure Mode, of this manual. 4. Record the reading in R or mr. 5. Calculate the mr/mas value. 2-2

Operation Applications 2 6. Record the data calculated above and the technique factors selected in Step 2 (referenced at a later date). 2.5.2 Determining Minimum Filtration Requirements (Beam Quality - HVL) 1. Select a tube potential which is commonly used and is in the highest kvp range of the x-ray machine. 2. Position the x-ray tube to the center of the chamber. 3. With no added filtration in the beam, make an exposure per Section 2.3, Exposure Mode, of this manual and record the reading. 4. Using the optional P/N 07-436 Half-Value Layer Kit (or equivalent), tape the increments of filtration to the face of the collimator. 5. Make an exposure and record the reading for each total thickness of filtration as indicated in Table 2-1. NOTE The information contained in Table 2-1 was extracted from DHEW Publication (FDA) 76-8014 "Suggested Optimum Survey Procedures for Diagnostic X-Ray Equipment. 6. Plot the exposure reading (log scale) versus the total added filtration thickness on semilog paper. 7. Determine the exposure value which is 50% of the reading recorded in Step 3. The corresponding thickness is the HVL value. Refer to Section 2-1 for the minimum HVL values required for that particular kvp. To check the minimum filtration requirement at a particular kvp: 1. At a pre-selected kvp, record an exposure reading and an exposure reading with no filtration. 2. Record an exposure reading with the minimum filtration requirement taped to the collimator. The minimum filtration requirement has been met if the second exposure is less than or equal to one-half of the first reading. Table 2-1. Minimum Filtration Requirements kvp Range Total Added Filtration Steps Measured Below 50 0.5, 1.0, 1.5, 2.0 30 0.3 40 0.4 49 0.5 50 to 70 1.0, 1.5, 2.5, 3.5 50 1.2 60 1.3 70 1.5 Above 70 1.5, 2.5, 3.5, 4.5 70 2.1 80 2.3 90 2.5 100 2.7 110 3.0 kvp HVL (mm AL*) 2-3

Victoreen 06-526-5240 and 06-526-5242 Operators Manual 120 3.2 130 3.5 140 3.8 150 4.1 *Type 1100 aluminum alloy as give in "Aluminum Standards and Data." 2.5.3 mas Reciprocity At any given kvp, combinations of time and ma that yield equal ma values should produce equal exposure (output) values. 1. Make an exposure at each ma station, varying the time to maintain mas constant. Variations exceeding the 10% average value may indicate the need for equipment recalibration. 2.5.4 Fluoroscopic Exposure Rate 1. Position the chamber so that the center of the chamber is centered to the central beam. 2. Using the optional P/N 07-706 Patient Phantom (or equivalent), place two ¾ inch aluminum attenuators between the chamber and the image intensifier. 3. Center the detector in the fluoroscopic beam as follows: a. Turn on the x-ray beam at a low ma setting. b. Use the image intensifier to view the image intensifier ion chamber image. c. Press the rate button on the RAD-CHECK MICRO- R front panel. (Select High or Low). d. Turn the fluoroscope on and read the display. CAUTION To prevent damage to the image intensifier, the aluminum attenuators must be in place while making an exposure. 4. Place a 1/8 inch lead beam stop plate (also included in the 07-706 Patient Phantom) or equivalent (folded lead apron) between the MICRO chamber and the image intensifier tube. This drives the automatic brightness control to maximum output. NOTE On equipment without automatic brightness controls, select maximum output settings. 21 cfr 1020 requires that the maximum exposure rate be less than 10 R/min (Cg/min). 2-4

Operation Accuracy Considerations 2 2.6 Accuracy Considerations Electrometer Drift Each instrument has a constant drift which can be measured in a zero radiation field. Drift is expressed as mr/min and has a typical value of 10 mr/min or less. By recording the drift rate (see Table 2-2), this error can be corrected. Table 2-2. Periodic Drift Rate Record Date Drift (mr) Time (Min) Rate (mr/min) * / / * The first reading is performed at the factory for future reference over the lift of the instrument. Air Density Corrections Temperature and pressure have a definite effect on the response on the ion chamber; they determine the number and density of molecules present for ionization. Figure 2-1 shows the correction factors used for changes in temperature and altitude (from 22 C and sea level). To achieve maximum accuracy, use the following equation: Corrected Exposure = (Measured Exposure - Drift x Exposure Time) ( CF T ) (CF P ) Where (CF T ) and (CF P ) are the correction factors for temperature and pressure respectively. Note that no more than 2% errors result over the temperature range of 16 C to 28 C (60 F to 82 F), so that the temperature correction can normally be disregarded. In the above equation, (CF T ) and (CF P ) can be calculated (rather than approximating using Figure 2-1). CF T = CF P = Where: T( C)+273 295 760 P(mmHg) 2-5

Victoreen 06-526-5240 and 06-526-5242 Operators Manual T ( C) = temperature measured in degrees C, and P (mmhg) = pressure measured in mm Hg. Figure 2-1 Temperature Altitude Correction Graphs 2-6

Operation Sources of Error 2 2.7 Sources of Error If an incorrect measurement is suspected, check the items listed below as they are potential sources of error: 1. Low battery voltage. 2. Failure to properly zero the unit before/between measurements. 3. Uncorrected drift error over long exposure times. 4. Unusual atmospheric conditions requiring a correction factor. 5. Extended low energy backscatter. 6. Capacitive effects (Refer to Section 2.8, Precautions). 7. Static discharge (Refer to Section 2.8, Precautions). 8. A damaged cable or connector on the chamber. 9. Operating the unit beyond specified limits of temperature, x-ray energy, etc. 2.8 Precautions Extreme caution should be used when making connections with the chamber and rear panel, an electrical shock hazard exists on the ion chamber bias connector (HV out). 1. Due to the high voltage inside the ion chamber and its delicate nature, puncturing or removing the chamber cover may result in damage to the unit or a change in calibration. 2. If a large amount of static electricity is present, small-to-moderate absolute errors may occur as a result of touching the exposed window. This is most likely to occur when the ambient humidity is very low. 3. Because of the highly sensitive nature of the electrometer input circuit, changing the position of the instrument can cause small changes in the reading due to capacitive effects. This is a reversible error and can be corrected by returning the unit to its original position or manually resetting the unit. 2.9 Operational Checks WARNING Regular monitoring of the instrument's performance is necessary to achieve maximum accuracy. 2.10 Drift Rate Periodic measurements of the drift rate should be recorded for reference when long exposure periods are used (Table 3-1). 2.11 Battery Voltage Correct battery voltage is important for the proper operation of the unit. Installation and replacement of the battery requires the removal of the rear access panel. A MN1604 alkaline battery (P/N 16-29 or 2-7

Victoreen 06-526-5240 and 06-526-5242 Operators Manual equivalent) is recommended. However, any standard 9 V transistor battery will work, but with shorter life. The battery should be replaced after a maximum of one year's use. WARNING Extreme caution should be used when making battery changes. An electrical shock hazard exists on the ion chamber bias connector (HV out). 2.12 Zero Set NOTE Turn off power before replacing the battery. Due to aging of circuit components, small changes in the zero reading can be expected over the lifetime of the unit. A Zero Adjustment, located on the rear panel, is provided to permit field adjustment as follows: 1. Turn the unit ON. 2. Wait five minutes. 3. Set the front panel switch to DOSE. 4. Press RESET. 5. Adjust ZERO if necessary so that the display shows 0.000. 2-8

Service Information General 3 Section 3 Service Information WARNING Extreme caution should be used when making connections with the chamber and rear panel, an electrical shock hazard exists on the ion chamber bias connector (HV out). 3.1 General The RAD-CHECK MICRO-R contains two functional elements: the electrometer circuit and the Voltage Ratio Module (VRM). The external ion chamber consists of electrically charged plates enclosing a constant volume. Under operation (no radiation), no free ions exist in this volume. When exposed to an ionization source, ions are created in direct relationship to field strength. The ions are collected, producing an output current. 3.2 Circuit Description The electrometer circuit is either a charge integration circuit or a current to voltage circuit depending on the selection of DOSE or RATE respectively. The DOSE or RATE mode is selectable using the front panel switch. When operating in the DOSE mode, the electrometer circuit is a charge integration circuit. When operating in the RATE mode, the electrometer circuit is a current to voltage converter. The input is a high impedance operational amplifier which provides minimum drift when used as an integrator. The charge from the ion chamber is collected by the electrometer amplifier producing a voltage across a capacitor. This voltage is a variable percentage of the operational amplifier output. A dose calibration potentiometer is used to calibrate the unit to give a direct radiation reading represented by this voltage. The front panel RESET switch closes a relay, which shorts the charge across a capacitor, resetting the operational amplifier to 0. The zero offset adjustment potentiometer is used to adjust the electrometer input. It can be found on the rear panel. 3.3 Calibration and Adjustments Standard Calibration: Calibrated @ Technique M75 3-1

Victoreen 06-526-5240 and 06-526-5242 Operators Manual NOTE The user should be aware that changes in altitude and temperature will affect the reading (Refer Air Density Corrections in Section 2.6). The voltages listed below should be checked before continuing with adjustments or calibration. 1. TP8 - TP12 = +7 to +9 V 2. TP9 TP12 = +5 V ± 5% 3 TP10 - TP12 = -270 V ± 10% (Use an electrostatic voltmeter). 4. TP4 - TP6 = 1.1 V ± 0.2 V Zero Adjustment This adjustment eliminates offset and aging errors common to operational amplifiers. 1. Set the front panel switch to DOSE. 2. Reset the display, observe for several minutes, and adjust the zero adjustment located on the rear panel as required. Offset (R30, U3) WARNING Use care, 270 Vdc is present inside the unit. 1. With the cover on the unit, set the front panel switch to DOSE. 2. Press the RESET button on the front panel. The display should read 0.00. 3. If necessary, adjust R30, located on the printed circuit board, to display a 0.00 reading. NOTE Use a voltmeter between TP5 and TP12 for accuracy. 3-2

Troubleshooting Precautions 4 Section 4 Troubleshooting 4.1 Precautions WARNING Extreme caution should be used when making connections with the chamber and rear panel, an electrical shock hazard exists on the ion chamber bias connector (HV out). 4.2 Troubleshooting CAUTION Many components on the printed circuit board are static sensitive. ESD precautions should be observed when handling the printed circuit board assembly. If there is a problem with the unit, refer to the table below for possible causes and corrective actions. Symptom Possible Cause Corrective Action No display Dead Battery Replace battery (See 2.11) & No Reset Light Broken battery snap lead Replace lead Defective ON/OFF switch Replace switch Inaccurate Readings zero Does not zero Adjust rear panel (see ZERO SET) Low battery voltage Defective/damaged ion chamber Replace battery (See 2.11) Recalibrate chamber Return unit to factory for chamber repair/replacement 4-1

Victoreen 06-526-5240 and 06-526-5242 Operators Manual (Blank page)

Appendix Replacement Parts A Appendix A Replacement Parts A.1 Replacement Parts There are no user replaceable parts on the RAD-CHECK MICRO-R circuit board. If a part is suspected of being faulty, contact Fluke Biomedical at 440.248.9300. A-1

Fluke Biomedical Radiation Management Services 6045 Cochran Road Cleveland, Ohio 44139 440.498.2564 www.flukebiomedical.com/rms