CS Controlled Dissolved Gas assembly with optional CS Air Aspirator assembly. Installation and Operation information

Similar documents
product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

1020 Industrial Drive, Orlinda, TN fax

BS Series Basket Strainer

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

NGP-250/500 Nitrogen Generator Quick Start Guide

Pure Water Power Purification System 4-Stage RODI Operations Manual

INSTALLATION INSTRUCTIONS. CVS 67CFR Pressure Reducing Instrument Supply Regulator INTRODUCTION

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

FlashGARD HP Reverse Osmosis Filtration System Model Number FSTMO75 Part Number

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

INSTALLATION & MAINTENANCE INSTRUCTIONS

Interface Devices, Inc. Hydraulic Mini Mule

Operators Manual. Supera Floor Stand Anesthesia Machine M1000. Models: M1000 Basic Non-rebreathing M1200 Basic M1200

Avenger Model 33M Installation & Operation Instructions

CART FOR CENTURION 1500+

LUDLUM MODEL 239-1F FLOOR MONITOR. Revised December 2010

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

Instruction Manual Updated 7/26/2011 Ver. 2.2

SERVICE MANUAL MODEL BA600AMST-S1 BREATHING AIR PANEL

DEOX FPS 1 CORRISION INHIBITING SYSTEM O&M MANUAL. Revision 2 Date 02/04/16

SERVICE MANUAL MODEL BA020AMST BREATHING AIR PANEL

Installation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series

Model PDT Dewpoint Transmitter

FloWash. Industrial Filter Cart

Float Operated Level Controllers

LNG Regulator Test Kit

Model PSI Compressor with 3-Gallon Air Tank 12VDC

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 4075 INSTRUCTION MANUAL

Installation, Operation and Maintenance Manual for Back Pressure Regulator

Index Table. Model 794. Installation, Operating and Maintenance Instructions

5 Channel Calibrator

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

Model 106 DPI "Micro-switch" Installation and Operating Instructions

RK-IR Sample Draw Aspirator Adapter Operator s Manual

WIDDER TOOLS. HPICAL Calibration Station 1 (203)

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

SERVICE MANUAL MODEL BA200AA-S1 BREATHING AIR PANEL

INSTRUCTIONS FOR MODELS APC & APG HIGH PURITY, CORROSIVE GAS REGULATORS

TITAN FLOW CONTROL, INC.

CSA Sample Draw Aspirator Adapter Operator s Manual

3M Liqui-Cel EXF-10x28 Series Membrane Contactor

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

ECS Protector Nitrogen Generator PGEN-30 (PGEN-30E)

LRS(H)4 Pressure-Reducing Regulator User Manual

Yoke Block Instruction Manual

SERVICE MANUAL MODEL BA020ANM-S1 BREATHING AIR PANEL

Installation, Operation, and Maintenance Manual

TECHNICAL INFORMATION

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Marschalk Model # 94000

RK LEL Sample Draw Aspirator Adapter Operator s Manual

PNEUMATIC PRESSURE CONTROLLERS

SERVICE MANUAL MODEL BA200AMST-S1 BREATHING AIR PANEL

UltRo Dual Flow Reverse Osmosis Water System

HIGH RATE BASKET STRAINER

Installation Manual. Model # SF-3W 9/16/03 REV 1. Arch Chemicals, Inc Lower River Road. P.O. Box 800 Charleston, TN PULSAR

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 3035 INSTRUCTION MANUAL

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator

Operation & Service Manual

Standard Operating and Maintenance Instructions for Pumping System Model PS-90

CARTRIDGE FILTERS. Models: CF50, CF75, CF100, CF150,

3M Liqui-Cel EXF-10x28 Series Membrane Contactor with ANSI or JIS Connections with Integrated End Caps

PORTABLE AFTERCOOLER MODEL CA-700 PORTABLE AFTERCOOLER/FILTER AFTERCOOLER OPERATION MANUAL ORIGINAL INSTRUCTIONS

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

Halsey Taylor Owners Manual STOP!

Installation Operation Maintenance. Bermad Level Control Valve with Modulating Horizontal Float Pilot valve One Way Flow IOM.

J Air and Water Kit Instructions Part# 02584

Light Commercial Reverse Osmosis System. EE-1000 Manual

TWO POST RAM PACKAGE 5 GALLON READ THIS MANUAL CAREFULLY BEFORE INSTALLING, OPERATING OR SERVICING THIS EQUIPMENT.

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50-P-Mini Portable Systems Outfitted with ABM-725 Monitor

Technical Information

DuraSite 3000 Sample Vessel User's Guide. Version Rev-1

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

Installation Operating Instructions for Duplex Manual Manifolds PX-TMD Series

DuraSite 500(1) Portable Sample Vessel User's Guide. Version (1) = 90 Duro Wetted O-Rings

SWIRLFLO Refrigerated fountains with FLEXI-GUARD

ISQB1 INTRINSICALLY SAFE ELECTRO-PNEUMATIC CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS HAZARDOUS AREA CLASSIFICATION

M5400 Operators Manual

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

INDUSTRIAL VALVES MODELS: C62-A; C62-D. INSTRUCTION MANUAL Installation Operation Parts Service DIAPHRAGM BYPASS PRESSURE REGULATING VALVES

Manual. Bypass Feeder. Installation Maintenance Repair Manual. Chemical Addition and Filtering

INSTRUCTIONS FOR MODELS SG3897 AND SG3898 CROSS PURGE ASSEMBLIES

Accu-Tab Systems 2000 P Series by Axiall Corporation

1200 Series Pressure Regulators

with O 2 Controller Instruction Manual

CentriFlow Meter Integrated Air Entrainment System

INSTALLATION AND OPERATION MANUAL CHEMICAL BYPASS FEEDER General Instructions Regarding All Models

Operators Manual. Supera Wall Mount Anesthesia Machines. Models: M2300 M2500. (All shown with optional equipment) M2300 M2500

Pressure Dump Valve Service Kit for Series 2300 Units

Installing ASI Umbilical Delivery Systems

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions

Installation, Operation, and Maintenance Manual

OPERATING AND MAINTENANCE MANUAL

Tri-Tech Medical Inc.

TECHNICAL INFORMATION

D Series Air and Water Kit Part# 02550

INSTRUCTIONS FOR MODELS UPH & UPHS HIGH FLOW, HIGH PURITY GAS REGULATORS

Transcription:

CS-425-01 Controlled Dissolved Gas assembly with optional CS-603-01 Air Aspirator assembly Installation and Operation information Serial # CS-XXX Date of shipment:, 2005 Shipped to: Purchase order #: xxxxxxx Attention: Table of Contents General information ---------------------------------------------------------------------------------- page 2 System configuration check list --------------------------------------------------------------------- page 3 Consumables and Dissolved oxygen meter information---------------------------------------- page 4 Simplified P&ID--------------------------------------------------------------------------------------- page 4 Un-packaging instructions --------------------------------------------------------------------------- page 5 CS-425-01 facility requirements--------------------------------------------------------------------- page 5 CS-603-01 facility requirements--------------------------------------------------------------------- page 5 Installation information ------------------------------------------------------------------------------ page 6 System commissioning ------------------------------------------------------------------------------- page 7-9 System operation -------------------------------------------------------------------------------------- page 10 Warranty ----------------------------------------------------------------------------------------------- page 10 End user (customer) notes on dissolved oxygen testing ---------------------------------------- page 11

General, but important information (CS-425-01 with optional CS-603-01) First, it is very important that all of the following instructions and support documentation is read and understood prior to installation and use. Failure to follow the System Commissioning instructions will affect performance, shorten the life of the de-gasification cells and void OEM warranties. If there are any questions regarding operation or use, contact Custom Fab Solutions prior to taking action. The CS-425-01 / CS-603-01 Controlled Dissolved Gas assembly does not include a dissolved oxygen meter to measure the dissolved oxygen in DI water. See page 4 for ordering information. Important information. Prior to installing the two (2) DI water filter elements and the two (2) de-gasification cells on the CS-425-01 assembly it is very important that the system is flushed with DI water as outlined in the System Commissioning section of this document. Failure to follow these instructions will dramatically decrease the life of the de-gasification cells, with pre-mature failure very likely, and void the manufacturers warranty. Consumables may be ordered directly from Custom Fab Solutions or from the original manufacturer. OEM documentation on the components used to manufacture the CS-425-01 and the CS-603-01 products is included in section 5 of the manual. This is important information and should be read, understood and followed before use. Thank you for selecting the CS-425-01/CS-603-01 products and your interest in Custom Fab Solutions. Feel free to contact us if we can be of any further service to you. The company web site is located at www.customfabsolutions.com. Page 2 of 12

System Configuration check list: Model CS-425-01 Controlled dissolved gas assembly including: Label Qty. Description Ordered 1 ea. 304 Stainless Steel Component mounting frame; electro-polished 1 ea. 8 feet of ½ PFA tubing for DI water source connection W-1 1 ea. PFA On/Off valve; ¼ turn W-2 1 ea. Gauge isolator with pressure gauge for DI water W-3 1 ea. Polypropylene (5 ) filter housing; 0.1 element included 1 ea. Polypropylene filter housing wrench W-4 1 ea. Liqui-Cel model G420 (includes ready to install plumbing package) W-5 1 ea. Liqui-Cel model G420 (includes ready to install plumbing package) W-6 1 ea. Gauge isolator with pressure gauge for DI water W-7 1 ea. DI water flow meter W-8 1 ea. PFA (5 ) filter housing; 0.05 element included 1 ea. 16 feet of ¼ outlet tubing; PFA G-1 1 ea. Gas On/Off valve; ¼ turn G-2 1 ea. Gas filter housing; 0.3 filter element included G-3 1 ea. Gas regulator G-4 1 ea. 15 psi gauge G-5 1 ea. Gas flow meter G-6 1 ea. 0.003 disposable (in-line) gas filter V-1 1 ea. 0 30 Hg vacuum gauge 1 ea. Vacuum interface with 3/8 ID vinyl tubing Liqui-Cel(s) and the DI water filter elements are supplied un-installed; as received from the OEM supplier. Model CS-603-01 Air Aspirator assembly including: 1 ea. 304 Stainless Steel Component mounting frame; electro-polished A-1 1 ea. Gas On/Off valve; ¼ turn A-2 1 ea. Gas filter housing; 5 micron filter element included A-3 1 ea. Gas regulator A-4 1 ea. 100 psi gauge V-2 1 ea. PIAB air aspirator with vacuum gauge and muffler V-3 1 ea. PIAB air aspirator with vacuum gauge and muffler 1 ea. 12 feet of 2 flexible, wire re-enforced PVC tubing Page 3 of 12

Consumables & Accessories W-3 1 ea. 0.1 micron DI water (5 ) filter element (Pall; P/N VFSG100-04M3V) W-8 1 ea. 0.05 micron DI water (5 ) filter element (Pall; P/N VFSG050-04M3F) G-2 1 ea. 0.3 CDA pre filter element (SMC; P/N 630611) G-6 1 ea. 0.003 disposable gas filter (Pall; P/N GDF6406E5) W-4 1 ea. Liqui-Cel (Liqui-Cel; P/N G420) W-5 1 ea. Liqui-Cel (Liqui-Cel; P/N G420) 1 ea. PFA filter housing wrench Dissolved oxygen meter information Label Qty. Description Manufacturer Part # D-1 1 ea. OxyGuard Handy Alpha (ppm / % saturation) OxyGuard Alpha Or D-1 1 ea. OxyGuard Handy Beta (ppm / % saturation / temp / display) OxyGuard Beta Model CS-425-01 Controlled Dissolved Gas assembly with optional CS-603-01 Air Aspirator assembly G-1 G-2 G-3 G-4 G-5 G-6 (Simplified P&ID ) October 2003 ` CDA End User Responsibility CDA D-1 Exhaust DI W-1 W-2 W-3 A-1 Beaker & Oxygen meter A-2 A-3 V-3 A-4 Output V-2 W-8 LiquiCel (A) W-4 W-5 Custom Fab Solutions, LLC confidential product information; material may not be disclosed to third parties without written approval from Custom Fab Solutions. LiquiCel (B) W-7 flow W-6 V-1 Page 4 of 12

Un-packaging instructions The CS-425-01 and the CS-603-01 have been Clean-room packaged (class 1000) and are ready for installation. Inside the shipping crate is: The CS-425-01 assembly. Two (2) Liqui-Cel* contactor cells. Liqui-Cel Part # G420. A 0.1 micron DI water filter element. Pall filter element, Part # VFSG100-04M3V. A 0.05 micron DI water filter element. Pall filter element, Part # VFSG050-04M3F. 8 feet of ½ PFA tubing (flared connection on one end) for the incoming DI water connection. 16 feet of ¼ PFA tubing (flared connection on one end) for the DI water output connection. The CS-603-01 air aspirator assembly. 12 feet of 2 wire re-enforced PVC tubing; for the air aspirator exhaust connection(s). Installation and Operation manual. * Liqui-Cel contactor information: Cartridge: 000000L000000 Product S/N: L000000 Cartridge: 000000L000000 Product S/N: L000000 o The two (2) DI water filter elements and the two (2) Liqui-Cels have been supplied as received from the respective manufacturer. o CDA (clean dry air) filtration (G-2, G-6 and A2) has been factory installed; ready for use. CS-425-01 facility requirements/connections DI water: 20 60 psig; four (4+) liters per minute ½ tubing connection CDA* 15 90 psig; ten (10) liters per minute ¼ FNPT connection *CDA = Clean Dry Air used to add oxygen to DI water CS-603-01 facility requirements/connections CDA 90 psig; twelve (12) scfm. ¼ FNPT connection Exhaust; 2 each 10 scfm per connection 2 tubing connection(s) Page 5 of 12

Installation information Mount the CS-603-01 aspirator assembly to the CS-425-01 frame using the three (3) ¼ -20 screws supplied with the air aspirator assembly. See the CS-425 / CS-603 installation drawing and CS-603-01 ref. photos 1, 4 and 5 of 9 for further detail. Installation note. Each installation is unique and different and the CS-425-01 assembly should only be installed by a qualified technician. Installation further assumes the technician installing the system is competent in general plumbing practices and procedures. Mount the CS-425-01 frame to a wall or a surface capable of supporting the weight (approximately 50 lbs.) of the assembly utilizing the pre-drilled 3/8 mounting holes on the top and the bottom of the frame. It is recommended that a minimum of four (4) bolts and/or lag bolts, two (2) on the top and two (2) on the bottom, are used to secure the system. See CS-425 with optional CS-603 Installation Drawing (Section 1 in this manual) for dimensional detail. Custom Fab Solutions suggests the CS-425-01/CS-603-01 assembly be located in close proximity (12 feet or less) from the sink (use) area. Mount the assembly in a suitable area where controls and gauges are accessible, visible and protected. (Note, clear visibility of V1 is critical to operation/controlling dissolved oxygen.) After Mounting: Remove the ½ PFA Nut and the PFA cap from valve W-1. Discard the sealing cap and slide the nut on to the ½ PFA tubing. Hand tighten the nut to the fitting on W-1. Cut excess tubing to length and connect to the DI water source. o See ½ DI inlet capped and ½ DI inlet connection photos, photo s 7 and 8 of 16 for further detail. Remove the ¼ PFA Nut and the PFA cap from filter W-8. Discard the sealing cap and slide the nut on to the ¼ PFA tubing. Hand tighten the nut to the fitting on W-8. Direct the outlet tubing to the designated sink; secure tubing as necessary. o See ¼ DI outlet connection photo 9 of 16 for further detail. Remove and discard the ¼ plastic pipe plug from G-1. Connect to CDA. o See Gas Input photo 12 of 16 for further detail. Remove and discard the ¼ plastic pipe plug from A-1. Connect to CDA; 90 psig min. o See CS-603-01 ref. photo 7 of 9 for further detail. Attach PVC flanges and mufflers to the air aspirators V2 and V3. o See CS-603-01 ref. photos 2 and 3 of 9 for further detail. Page 6 of 12

System commissioning SYSTEM FLUSHING --- The DI water plumbing must be thoroughly flushed with DI water before use! Failure to flush the system will result in pre-mature failure of the LiquiCel contactors and void the manufacturers warranty. Do not install the filter elements or the LiquiCel s at this time. The following instructions assume: a) the CS-603-01 is attached to the CS-425-01 frame. b) the DI water input is plumbed and valve W-1 is closed. c) CDA is plumbed to G-1 and the valve is closed. d) CDA is plumbed to valve A-1 (90 psig minimum pressure) and valve is closed. e) the DI water output from W-8 (1/4 PFA tubing) has been plumbed to the designated sink (use) area and the tubing has been secured. f) both filter housings, W-3 & W-8 are installed and hand tightened. g) the Flushing tubing is connected between the W-3 filter housing and the W-6 DI water pressure gauge. See the Flushing tube photo 10 of 16 for details. Assuming the above is satisfied proceed with the following steps: 1. Slowly open valve W-1 and adjust the DI water flow rate (W-7) to the maximum (3000+) adjustment on the flow meter. Flush the system for approximately 10 minutes. During the flushing period, check for leaks and correct as necessary. After 10 minutes turn valve W-1 off. 2. Take appropriate action to drain and remove the two filter housings and install the two (2) DI water filter elements per the manufacturers instructions in the box. o The DI water filter housings are filled with DI water; take the necessary steps to prevent spilling when removing the housing from the head assembly. o Be sure the 0.1 micron filter element (p/n VFSG100-04M3V) is installed in filter W-3 and the 0.05 element (p/n VFSG050-04M3F) is installed in the W-8 filter housing. 3. With the filter elements installed in W-3 & W-6 housings, open valve W-1 and flush for 5 minutes at the maximum flow of the flow meter. Check for leaks and correct as necessary. o During this step it is necessary to bleed trapped gas from the W-3 DI water filter. This is accomplished by losing the filter vent fitting. See CS-425-01 DI water filter vent photo; 16 of 16. 4. After flushing, take the appropriate steps to remove the flushing tube. See photo 10 of 16. Page 7 of 12

5. Remove the LiquiCel(s) from the manufacturers package and install per the manufacturers (LiquiCel) instructions. The water ports are labeled SHELLSIDE. The vacuum/gas ports are unlabeled! SHELLSIDE ports may or may not be oriented as shown in the following illustration or as in the photos attached to this document. Check and then re-check Shellside identification prior to installation. o Prior to installation it is necessary to remove the ¼ MNPT shipping plugs; (2 each) per cell. Water SHELLSIDE Note - The SHELLSIDE marking is VERY difficult to see. The letters are transparent (same color as the LiquiCel) and slightly raised. Vacuum or Gas 6. Installing the LiquiCel(s) is performed by the following steps. Prior to installation refer to the LiquiCel Connections and the Overview photos. Reference photos 2, 13, 14 & 15 of 16 a) Vacuum fitting connection -- W-4 Reference photo 14 of 16 Install the ¼ NPT plug as shown in the attached photo. Install the barbed fitting as shown in the photo. b) Vacuum gauge and fitting connection -- W-5 Reference photo 15 of 16 a. Install the vacuum gauge assembly as shown. c) DI water in -- W-4 Reference photo 14 of 16 ¼ NPT x ½ tubing connection. (DI water filter connection) Page 8 of 12

d) LiquiCel to LiquiCel connection -- W-4 & W-5 Reference photo 14 & 15 of 16 Install the two (2) ¼ NPT x 3/8 tubing fittings. Install the 3/8 PFA connection tube. e) DI water W5 to W6 connection -- W-5 Reference photo 15 of 16 ¼ NPT x 3/8 tubing connection. (LiquiCel to DI pressure gauge connection) f) Secure LiquiCel(s) using mounting brackets as shown in LiquiCel Connection photo. Reference photo 13 of 16 g) Gas input -- W-5 Open valve G-1 and adjust the pressure on G-3/4 to 15 psig. Open the gas flow meter valve G-5 set to 10. Purge the line for two (2) minutes Install the ¼ NPT x 3/8 tube fitting on the Liqui-Cel and attach the tubing from G-6 to W-5 as shown in photo. Reference photo 15 of 16 The CDA regulator supplied with the CS-425-01 and the CS-603-01 is a locking style regulator. Unlocking the regulator is accomplished by pulling down (out) on the adjustment knob; push up or in to lock. CS-603-01 Connections I. Vacuum source connections a. Connect the vacuum source from V2 to LiquiCel W-4 using the 3/8 supplied (vinyl) tubing. b. Connect the vacuum source from V3 to LiquiCel W-5 using the 3/8 supplied (vinyl) tubing. See CS-603-01 ref. photo 6 of 9 for further detail. II. CDA (air aspirator control) connection a. Connect the 3/8 tubing from regulator A-3 to V-3; fitting supplied on tubing. See CS-603-01 ref. photo 6 of 9 for further detail. III. Air Aspirator (CDA) discharge (Exhaust) 2 PVC connection(s) a. Cut 2 flexible tubing to size and connect tubing to the PVC muffler bushing; 2 connections V2 and V3. See CS-603-01 ref. photos 8 & 9 of 9 for further detail. Plumb CDA (air aspirator) discharge to an exhaust or a suitable area. Note: 2 tubing should be adequately supported to avoid stress on the PVC flange.. Page 9 of 12

The system is now ready for use. To operate: Turn on the DI water, valve W-1 and set the W-7 flow meter to a flow rate of 2000cc (2 liters) per minute. Reference photos 7 of 16 & 11 of 16 Turn on valve A1, set regulator A3 to a minimum of 90 psig, and wait for V1 to reach approximately 29 Hg. Reference CS-603-01 ref. photo 7 of 9 and CS-425-01 photo 15 of 16 - - Vacuum gauge V1. Use the gas flow meter G-5 and oxygen meter to achieve desired settings. Reference photo 12 of 16 G-5 setting notation. In many cases (incoming water quality dependent) it may be necessary to adjust flow-meter G-5 to a +10 setting to achieve desired results. Use V1 as reference to quickly set the percentage (%) of dissolved oxygen in the DI water output. The reference is made by comparing the measurement of V1 with the readings from the dissolved oxygen (OxyGuard) meter. Reference photo 2 of 16 -- V1 gauge. Dissolved oxygen in the DI Water output can be easily changed by one or more or a combination of the following: o Increasing or decreasing the flow of CDA into the DI water flow stream; adjusting G5. o Increasing or decreasing the air pressure to the air aspirators; adjusting A-3 -- 40 -> 90 psig. See End user (customer) notes for additional details on dissolved oxygen and particle testing. Warranty I. Custom Fab Solutions warrants that at the time the goods are shipped to the purchaser, they will be free from defects in material and workmanship, under normal use and service, with proper maintenance, for a period of one (1) year. Said warranty is limited, at the option of Custom Fab Solutions, to repair or replacement of the defective item provided that such item is returned to Custom Fab Solutions, transportation prepaid, for inspection and approval. II. Components which are not manufactured by CFS, purchased items, are per the manufacturers warranty. Warranty claims for component items must be filed with the original manufacturer. Page 10 of 12

Dissolved oxygen test: End user (customer) notes Calibrate the meter according to manufacturer s instructions. Immerse the probe into a designated/clean 500ml beaker. Immerse DI water output from the CS-425-01 into the beaker. Observing change (time sensitive) in dissolved oxygen meter readings. Note: It will take several minutes for the system to reach an equilibrium of dissolved oxygen level after G-5 is changed. Adjusting the flow of G-5 will produce the desired dissolved oxygen level. Set the flow rate (W-7) to 2 L/min and check particle count using a clean beaker. The particle count >0.6 um/ml should be less than 0.5 particle/ml. If the count exceeds this limit, check the connections, continue flushing and retest with a clean beaker. Page 11 of 12

Packing List Items ordered: Customer Included in this shipment: PO # The CS-425-01 assembly. Two (2) Liqui-Cel* membrane contactor cells. Liqui-Cel Part # G420. A 0.1 micron DI water filter element. Pall filter element, Part # VFSG100-04M3V. A 0.05 micron DI water filter element. Pall filter element, Part # VFSG050-04M3F. Necessary fittings to interface the Liqui-Cel contactors to the CS-425-01 assembly. Optional CS-603-01 Air aspirator assembly. 12 feet of 2 wire re-enforced PVC tubing for Air Aspirator discharge/exhaust. 8 feet of ½ PFA tubing (flared connection on one end) for the incoming DI water connection. 16 feet of ¼ PFA tubing (flared connection on one end) for the DI water output connection. One (1) Manual including electronic (PDF) files on a CD-ROM disk. by date