AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS

Similar documents
Prestige Endoscopic Graspers Instructions for Use

Shukla Medical. TITLE: Manual Surgical Orthopedic Instruments Recommendations for Care, Cleaning, Maintenance and Sterilization

Suction Regulators. Recommendations for Cleaning, Disinfection and Reprocessing: Boehringer 3800 &3900 Series Suction Regulators

TJF-Q180V Cleaning and Disinfection Checklist

TJF-Q180V Cleaning and Disinfection Checklist

Hercules 3 Universal Stabilizer Arm. Catalog Number

Competency Record for Sonic Cleaning

Cavitron DualSelect. Introduction ENGLISH. Directions For Use. Cavitron DualSelect Dispensing System ENGLISH 1

Laboratory Safety Training. Environmental Health and Safety

Turnout gear is comprised of boots, coat, pants, suspenders, hood, gloves, helmet, and SCBA Facepiece.

User Manual POINT SETTER MODEL:

SUGGESTED POLICY FOR THE MONITORING OF THE CLEANING EFFICIENCY OF AN ULTRASONIC CLEANERS( SONIC & TOSI) WEEKLY

COMMON PRECAUTIONS Data Sheet - Be Sure To Read Before Handling Page 1 of 6

Wolfson Campus Natural Science Department. Biology Laboratory. Student Safety Contract

STAINLESS STEEL TABLES

GUIDELINES FOR HANDLING EXTERNALLY CONTAMINATED MEDICAL CYLINDERS IN A PANDEMIC SITUATION

McLane / Black Lake Fire Department Policy & Procedural Guideline PPG #

MONT BLANC 10400X XL XLED 10401X XL XLED 10403X XL XLED

Air-Mate High Efficiency Powered Air Purifying Respirator (PAPR) Staff Training Module

General Warnings! A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately could lead

BIOSAFETY SELF AUDIT PRINCIPAL INVESTIGATOR/PERMIT HOLDER: CONTAINMENT LEVEL: 1 2 (Shaded) RISK GROUP: 1 2 AUDIT TO BE COMPLETED BY (DATE):

Use and Care Guide. PMU and PMU w/over the Patient. Swing Mount. Over the Patient. Side Delivery

F-330 Instruction manual. Magic Glove Bio Skin Smoother High Frequency Ultrasound Galvanic

Vertebrate Animal Biosafety Level 1 Criteria

O P E R A T O R S M A N U A L

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

Recommendations - USA

Frigitronics CRYO-PLUS

NUS Safety & Health Guidelines Procedure No: OSHE/SOP/GL/08

GLIDERITE RIGID STYLET. Operations & Maintenance Manual

WELLS JOHNSON HIGH VOLUME CANISTERS

ASAHI Neurovascular Guide Wire

Cleaning, Sanitizing, and Disinfecting in Child Care

INFUSION SYSTEM (110V) (220V) USER S MANUAL

25+ instruments answer the most common barrier to 25G instrument flexibility.

POOL CARE GUIDE. The E-Z Clor System To Perfect Pool Water.

RECONSTITUTION AND ADMINISTRATION INSTRUCTIONS FOR HEALTHCARE PROFESSIONALS

Guidelines for Safe Use and Handling in Health Care Facilities DIVISION OF ENVIRONMENTAL AND OCCUPATIONAL HEALTH SERVICES

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

To provide adequate disinfection for reusable semi-critical equipment other than scopes. Processing of scopes is covered in

RECONSTITUTION AND ADMINISTRATION INSTRUCTIONS FOR HEALTHCARE PROFESSIONALS. Hereditary Angioedema

Standard Operating Procedure (SOP)

INSTALLATION & MAINTENANCE INSTRUCTION

Hygea Ultrasonic Validation Kit

AFT25 Facemask with Two-Way Non-Rebreathing Valve

Urgent Field Safety Notice (Field Safety Corrective Action) for CORAIL AMT Neck Trials Surgical Instruments

CART FOR CENTURION 1500+

EKC International FZE. Safety Instructions For Handling & Inspection Manual For Refillable Seamless Steel Gas Cylinders

FLANGED TWO-PIECE BALL VALVES

WARNING! DO NOT THROW AWAY THESE INSTRUCTIONS! READ AND UNDERSTAND BEFORE USING EQUIPMENT!

User Guide for. UTR Turbo Air Lappers Pneumatic Air Grinders & Lappers UTR-30, UTR-70, UR , ,

The Outer Seal is highly abrasion resistant and provides mechanical stiffness to the splice in order to better match the cable stiffness.

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

Understanding Tracheostomy Care

LM-900 Cryosurgery System

Service and Repair Manual

Instruction Manual. PE-25/32/37 Tapping Machines. POLYTAPP Valves

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

LION ERS Glove System

9.0 Equipment Cleaning and Repair

BioAerosol Nebulizing Generator. Operation and Maintenance User Manual

Instructions for use Compressed air injector with water trap

Specimen Collection. Lead Screen Filter Paper Testing Dried Blood Spots (DBS) Laboratory Services February 2017

Client Care Items and Equipment

For Establishing and Maintaining Your Spa using PROtech Spa Chemicals

Rubber Dock Hose care, Use & Maintenance

Parts identification:

OPERATION AND INSTRUCTION MANUAL

Spirometry. Operating Manual. Part 4: Hygiene and disinfection for custo spiro mobile. Operating characteristics:

User s Guide. VetPro 5000 Wall / Cabinet & Table Mount Dental Delivery System. For Models: Rev L (11/10/15)

Decontamination Procedures for Gilson Pipettes

Sterile Visual Flow Indicator

A. APPLICATIONS 1. WARNING TO USER

UNIVERSITY OF ROCHESTER MEDICAL CENTER

Phalanx Alpha Phalanx Beta and Ultra Elite

Varithena Delivery System Instructions for Use

Tools to Enhance Laboratory Biosafety and Biosecurity Performing Risk Assessments

GATOR Battery-powered Pistol Grip Cable Cutters

Air Removal from the Sterilization Chamber

14500 Coy Drive, Grass Lake, Michigan Isolation Glove Box Instruction Manual

User Instruction Manual For Davit Rescue System

Ontario ENGINEERING STANDARD SPECIFICATION 1.0 PURPOSE

CRANE WITH WINCH USE AND MAINTENANCE MANUAL. Code Rev. Release MD /03

1. The laboratory supervisor must enforce the institutional policies that control access to the laboratory.

Operating manual Analogue refractometer

INSTRUCTIONS. Rotatable Clip Fixing Device HX-110LR HX-110QR HX-110UR. Clip HX HX Long Clip HX L.

Viral Haemorrhagic Fever (VHF) - The correct order for donning and the safe order for removal and disposal of Personal Protective Equipment (PPE)

Ultima Salt System Overview

Fingertip Pulse Oximeter

OPERATION AND INSTRUCTION MANUAL Swivel Anchor Model: SWY100N

Varithena Delivery System Instructions for Use INSTRUCTIONS FOR USE Please read the Product Monograph before using the product.

Tips for Proper Pipette handling and Maintenance. D. Muruganand, Ph.D ILQA - Bangalore, Feb 2014

COATINGS SPRAY GUIDE Urethane 520

Safety Management System. Initial Revision Date: Version Revision No. 02 GENERAL SAFETY HEALTH PROVISION

Cleaning rod: spring steel, stainless steel or carbon fibre cleaning rod - only use a one-piece rod. Avoid using snakes.

OPERATION MANUAL. Hand Held Leak Tester Model # ZUTR-10002

Standard Operating Procedure (SOP)

A N C H O R T A B S N O R T H W E S T ANCHOR TAB VF. Patented. User Instructions

THE UNIVERSITY OF NEWCASTLE - DISCIPLINE OF MEDICAL BIOCHEMISTRY

Compliance Requirements

Transcription:

AMBLER SURGICAL INSTRUCTIONS FOR USE - OPHTHALMIC DUET-STYLE SCISSORS Ambler Item # 6112E Duet-style MICS Packer/Chang IOL cutter, 1 3/8", 19 gauge, straight shaft, for use with Ambler #6100T INDICATIONS FOR USE A hand-held ophthalmic surgical tip used with a handpiece to cut, dissect, and/or divide tissue in the eye and/or surrounding tissues during the surgical treatment, mitigation, prevention, and/or diagnosis of ophthalmic disease or conditions. These instruments are reusable. GENERAL INFORMATION Federal (U.S.A.) law restricts this device to sale, distribution, and use, by, or on the order of a physician. These instructions are intended for use only by persons with the required knowledge and training in a health care facility. Ophthalmology procedures should be performed only by persons having adequate training and familiarity with ophthalmologic surgical techniques. All cleaning and sterilization processes provided are general guidelines and any deviation by the processor should be properly evaluated for effectiveness and potential adverse consequences. Any sterilization process will still require validation by the end user at the point of use. The end user should also routinely monitor the validation process as its effectiveness can vary dependent on multiple factors. Description of the detail of which function configurations micro scissors were designed for. Luer fittings are made to the ANSI National Standard ISO 594/1-1986 (E) with conical fittings, 6% Luer taper for syringes, needles, and certain other instruments. CONTRAINDICATIONS Damaged or broken instruments may result if the instruments are used improperly during transport, handling, surgical use, or reprocessing. WARNINGS FOR REPROCESSING The following instructions are for all NON-POWERED surgical instruments supplied by Ambler Surgical, unless stated otherwise with the packaging of the product. The surgical instruments are provided NON-STERILE and must be inspected, cleaned and sterilized before first use and before every reuse. Tip covers and other protective packaging material must be removed from the instruments prior to the first use. Long narrow cannulations and blind holes require particular attention during cleaning. Automated or manual flushing should be performed thoroughly during cleaning. Immediate use steam sterilization (IUSS) should only be used for emergency reprocessing and should not be used for routine sterilization processing of the instruments. IUSS instruments should be used immediately and not stored for later use. DO NOT REPROCESS SINGLE USE ITEMS. Do not use this process for diamond knives, cannulae, or silicone tubing and /or bulbs. If this instrument is/was used in a patient with, or suspected of having Creutzfeldt-Jakob Disease (CJD), the instrument cannot be reused and must be destroyed due to the inability to reprocess or sterilize to eliminate the risk of cross- contamination. Consult WHO and local regulations for further information. REPROCESSING PRECAUTIONS When reprocessing surgical instruments, always handle them with care, wearing personal protective equipment: impervious apron, shoe protection, gloves, and face shield in accordance with Universal Precautions (OSHA) and your facility's policies. Delicate surgical instruments require special handling to prevent damage to the tips. Use caution during cleaning and sterilization. Do not use excessive stress or strain at joints; misuse will result in misalignment or cracks at the box locks or jaws. Titanium instruments that are color anodized may lose their color over time through normal use and reprocessing. Saline, cleaning / disinfection agents containing aldehyde, mercury, active chlorine, chloride, bromine, bromide, iodine or iodide are corrosive and should not be used. Instruments must not be placed or soaked in Ringers Solution. Manual scrubbing with brushes should always be performed with the instrument below the surface of the cleaning solution to prevent generation of aerosols and splashing which may spread contaminants. Cleaning agents must be completely rinsed from device surfaces to prevent accumulation of detergent residue. Do not soak instruments in hot water, alcohol, disinfectants or antiseptics to avoid coagulation of mucus, blood or other body fluids. Do not exceed 2 hours soaking in any solution. Do not use steel wool, wire brushes, pipe cleaners or abrasive detergents. Do not use high acid (ph 4.0 or lower) or high alkaline (ph 12 or higher) products for disinfection. Neutral ph detergents (at or near 7.0) are

preferred.

LIMITATIONS ON REPROCESSING Repeated reprocessing has minimal effect on the instrument life. End of useful life for metal surgical instruments is normally determined by wear and damage due to the intended surgical use. INSTRUCTIONS Point of Use 1. Following use, the instrument should be cleaned of excess soil using a disposable cloth/merocel wipe and flushed or immersed in sterile water as soon as possible. 2. Ophthalmic viscosurgical device solution(ovd) should not be allowed to dry on instruments or in cannulae. OVD will dry and harden quickly and render the instrument useless. 3. The instrument should be kept moist with sterile water to prevent soil from drying on the instrument. Containment and transport 1. Surgical instruments should be reprocessed as soon as possible. 2. Whether used or not, opened instruments should be placed in a suitable sealed/closed container labeled as biohazard to protect personnel from contamination during transport to the decontamination area. 3. Frequent retrieval and transport of containers of instruments to the decontamination area is recommended. Preparation for decontamination and cleaning 1. Universal precautions should be followed including the use of suitable personal protective equipment (impervious apron, shoe protection, gloves, and face shield, etc.) according to Universal Precautions (OSHA) and your facility's policies. 2. Cleaning of instruments should be performed immediately upon receipt in decontamination area. WARNING: AUTOMATED CLEANING AND THERMAL DISINFECTION ARE NOT RECOMMENDED FOR MICRO SCISSORS. MANUAL AND ULTRASONIC CLEANING ONLY ARE RECOMMENDED. Manual and Ultrasonic Cleaning 1. Disassemble the micro scissors if applicable and inspect for damage or corrosion. Discard and replace if found. 2. Pre-rinse the micro scissors by holding it under deionized or distilled water, rotating the micro scissors to expose all surfaces and cavities to flowing water. 3. Micro scissors with should be flushed using a Luer lock syringe filled with 10cc of deionized or distilled water and neutral ph (at or near 7.0) cleaning solution mixed to cleaning solution manufacturers specifications precisely. Repeat until micro scissors flush easily. If unable to flush micro scissors discard and replace. 4. Flush should be repeated three times. 5. Place the micro scissors on a silicone mat in an ultrasonic machine filled with deionized or distilled water and neutral ph (at or near 7.0) cleaning solution mixed to cleaning solution manufacturers specifications precisely and sonicate for 3 minutes. 6. Ensure that the micro scissors are fully immersed in the deionized or distilled water and neutral ph (at or near 7.0) cleaning solution. 7. When using an ultrasonic machine, the solution should be drained after each use to avoid retaining bioburden on the micro scissors. The ultrasonic machine should be cleaned daily, if not after each use following the ultrasonic machine manufacturer's instructions. 8. Rinse the instrument by holding it under deionized or distilled water for at least 30 seconds, rotating the micro scissors to expose all surfaces and cavities to flowing water. Additional rinsing may be necessary to entirely remove cleaning solution. 9. Micro scissors should then be flushed using a Luer lock syringe filled with 10cc of deionized or distilled water. 10. Flush should be repeated three times. Drying After ultrasonic bath, dry outside of the micro scissors with a lint free surgical wipe. Blow the micro scissors dry with micro filtered pressurized medical grade air. When blowing dry with pressurized air, ensure secure grip on instrument to avoid damage to instrument from air pressure. If micro filtered pressurized medical grade air is not available, the I/A tip should have a minimum of 10 cc of air pushed through the tip. Repeat three times. Maintenance, Inspection and Testing 1. It is very important to carefully examine each micro scissors for breaks, cracks or tip burs before use. Attach to handpiece and ensure smooth open and closing action. A microscope should be used whenever possible. Packaging 1. Package the instrument in a suitable sterilization pouch or sterilizing instrument tray. 2. Double peel pack packing is not recommended unless pouch manufacturer validates the product is acceptable for this use. 3. Sterilizing trays should be of proper size lined with soft silicone mats. 4. Instruments should not be touching each other. 5. Protective tips made of soft silicone of the proper size and thickness are recommended.

Sterilization Follow the sterilizer manufacturers' instructions for operation and loading of steam sterilizers. There must be direct steam exposure to all surfaces of the instruments being sterilized including the internal surface and tubes channels. The instrument and/or instrument tray should be processed through a complete sterilization drying cycle as residual moisture from sterilizers can promote staining, discoloration, and rust. NOTE The tables below represent variations in sterilizer manufactures recommendations for exposure at different temperatures per ANSI/AAMI ST79:2010 and A1:2010 & A2:2011. Other time and steam temperature cycles may also be used. However, user must validate any deviation from the recommended time and temperature. Contact the manufacturer of your steam sterilizer to confirm appropriate temperatures and sterilization times. Minimum cycle times for gravity-displacement steam sterilization cycles Item Exposure time at 121º C (250º F) Exposure time at 132ºC (270º F) Exposure time at 135ºC (275º F) Drying times Wrapped instruments 30 minutes 15 minutes 15-30 minutes Textile packs 30 minutes 25 minutes 15 minutes Wrapped utensils 30 minutes 15 minutes 15-30 minutes items (e.g., instruments) and porous items in mixed load 3 minutes 3 minutes 0-1 minute 10 minutes 10 minutes 0-1 minute Minimum cycle times for dynamic-air-removal steam sterilization cycles Item Exposure time at 132º C (270º F) Exposure time at 135ºC (275º F) Drying times Wrapped instruments 4 minutes 20-30 minutes 3 minutes 16 minutes Textile packs 4 minutes 5-20 minutes 3 minutes 3 minutes Wrapped utensils 4 minutes 20 minutes items (e.g., instruments) and porous items in mixed load 3 minutes 16 minutes 3 minutes 3 minutes N/A 4 minutes 3 minutes N/A

Storage Following sterilization processing packaged instruments may be stored in a clean area free of temperature and humidity extremes in accordance with your facility s policies. For additional information regarding the reprocessing of ophthalmic instruments see: ASCRS/ASORN Special Report Recommendations for the cleaning and sterilization of intraocular cataract surgical equipment. J. Cataract Refract Surg. 2007; 33(6):1095-1100. Association for the Advancement of Medical Instrumentation. ANSI/AAMI ST79: 2010 & A1:2010 Comprehensive Guide to Steam Sterilization and Sterility Assurance in Health Care Facilities. Arlington, VA: Association for the Advancement of Medical Instrumentation, 2010. Centers for Disease Control, Guideline for Disinfection and Sterilization in Healthcare Facilities, 2008. Association for the Advancement of Medical Instrumentation (AAMI), Accreditation Association for Ambulatory Health Care (AAAHC), Association of perioperative Registered Nurses (AORN), Association for Professionals in Infection Control and Epidemiology (APIC), ASC Quality Collaboration, Association of Surgical Technologists (AST), and International Association of Healthcare Central Service Materiel Management (IAHCSMM). Position Paper Immediate Use Steam Sterilization, 2011 Ambler Surgical - 730 Springdale Drive - Exton, PA 19341 Last Updated 12/19/2016