Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached

Similar documents
FX134, FX1234, FX3030 Problem and Solutions

FAULT CODE TROUBLESHOOTING INDEX

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Armfield Distillation Column Operation Guidelines

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

BASIC Z-STACK AND TIME SERIES SCAN ON THE ZEISS LIGHTSHEET Z. 1

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Section 3- Repair and Calibration Procedures

UsER manual for Watersens ph -REDOX

AKTA MC SOP Page 1 9/27/04 AKTA METAL CHELATING SOP

Napa Technology Trouble Shooting. For Premier & Premier PLUS Models

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

CircuFlow Quick Setup Guide

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

NEULOG OXYGEN LOGGER SENSOR GUIDE

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

AKTA ION EXCHANGE CHROMATOGRAPHY SOP Date: 2/02/05 Author: A DeGiovanni Edited by: C. Huang Reviewed by:

IMPORTANT PLEASE READ BEFORE COMMENCING INSTALLATION

UNIVERSITY OF ROCHESTER ENVIRONMENTAL HEALTH & SAFETY

Navigator 600 Silica analyzers

AKTA 3D SOP. Click on the System Control tab. This screen has 4 windows.

KIV-SACS Safe automatic calibration system:

Read this first. Zetasizer nano series Self installation and Quick start guide MRK825-02

Boyle s Law: Pressure-Volume Relationship in Gases

QUICK START GUIDE TO DRINKING WATER MONITORING

User Manual for the Mars Calibration Bench

Physics Experiment 17 Ideal Gas Law Qualitative Study

Met One E-BAM Particulate Monitor

Installation Instructions

Using the Akta Prime plus October 22, 2012

to the LI-6400 Sensor Head

Gas Absorption Draft Standard Operating Procedure

Operator Quick Guide ORBISPHERE 3654

with O 2 Controller Instruction Manual

XpressFill XF2500 Counter Pressure Filler. Operating Instructions

WIDDER TOOLS. HPICAL Calibration Station 1 (203)

OPERATING INSTRUCTIONS

Installation Instructions

BR-2444 ROWING MACHINE

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

Glove Box Installation Manual

RM-80 respiration monitor

Overview. Front Panel: Keypad and Display

RAM 4021-DPX Operation Manual

patient education program 4156 South 52 nd Street, Omaha, NE ChildrensOmaha.org/HomeHealthcare

ShellPa. Standard Mechanical Cell Stretch System Model No: NNMS Serial #: User Manual

Operating Instructions

Model 130M Pneumatic Controller

Using the UltraRAE. Firmware 2.35

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Microfluidic Demonstration

MP15 Jockey Pump Controller

UNIVERSITY OF WATERLOO

icreasepro Creaser Operators Manual

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Quick Start Guide. A. Hardware installation B. Software installation C. Start the software for the first time D. Do your first measurement

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

Laboratory Mortar Mixer (Testing)

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

300 Held Drive Northampton Pa

R E D I C O N T R O L S

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

TR Electronic Pressure Regulator. User s Manual

256 Pneumatic Pressure Indicator

LABORATORY INVESTIGATION

SS32L Dissolved O 2 Probe

User s Guide. Pressure Controller for liquid delivery systems

FX IMPACT OWNER S MANUAL

2. Determine how the mass transfer rate is affected by gas flow rate and liquid flow rate.

Instruction Manual Dräger MSI P7 and MSI P7 plus

Quick Installation Guide Doc nr.: C Date:

ETT Olfactometer: Getting started guide 2013a. ETT Olfactometer quick start

HZ-300 HAZER MACHINE USER MANUAL. Congratulations on the purchase of your new ANTARI HZ-300 Haze Machine.

Installation Instructions

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

User s Guide. Vacuum Controller for liquid delivery systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

RAM 4021 Operation Manual

NanoSight NS300. NanoSight NS300. Operation instructions. Laser Spectroscopy Labs, UCI

Operating Instructions for Cobra and Lynx Duct Leakage Testers

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

ACV-10 Automatic Control Valve

Installation and Use Manual

RAM Operation Manual

BIOGAS COMBUSTION PLANT MAINTENANCE MANUAL

Leica TCS SP5 Confocal Live Cell Imaging Set Up User Guide

Bante821 Portable Dissolved Oxygen Meter Instruction Manual

SENSUS PRO MANAGER (for SENSUS or SENSUS PRO devices) User s Guide -- Windows. Version 2.0 Published October 17, ReefNet Inc.

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

ABRASIVE REGULATOR II

The Discussion of this exercise covers the following points:

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

Exploring the Properties of Gases. Evaluation copy. 10 cm in diameter and 25 cm high)

Underwater Housing for Panasonic Lumix DMC-ZS60, DMC-TZ80

Transcription:

Running the LiCor 6400 Infra Red Gas Analyzer (IRGA) with 4 whole-plant cuvettes attached Note: You must read the Primer manual Sections 1-6 before using the IRGA. The manual is located in the wall cabinet with glass doors to the left of the fridge. This protocol will be referring to two components of the 6400 system, the main unit and the IRGA head. The main unit is the greenish box with the liquid crystal display on its top surface. The IRGA head is the hand-held unit connected by tubes and cables to the main unit. The hand-held unit actually contains the infra-red sensors, hence it is called the IRGA. For this experiment the IRGA head must be connected to the main unit according to instructions in the Primer until CALIBRATION is completed. In this (the standard operating) mode, the Sample tubing connection from the IRGA head (the hand-held unit), labeled with a black ring of tape, is connected to the gas port labeled Sample on the bottom right of the right side of the main unit. Preparing the 4-cuvette system: 1. Tubing: When running the 4 whole-plant cuvettes, these tubing connections to the main Li6400 unit will remain undisturbed. At the other end of the tubes, where they attach to the hand-held IRGA head unit, you will disconnect them and connect instead the Sample and Reference tubes from the cuvette array. These are labeled Sample and Reference and physically cannot be connected to the incorrect input tubes to the IRGA head. Do this after calibration (if necessary), and after the Open System software is up and running. See below. 2. Leaf Thermocouple: The leaf thermocouple plug from the IRGA head s built-in cuvette must be detached. It is a purple plug connected to wires that originate under the IRGA head s built-in cuvette. Unplug it and plug in the matching plug that is attached to the 4- cuvette frame. 3. PAR (photosynthetically active radiation) sensor: The PAR light sensor plug is a small white plug on the right hand side of the IRGA head with white wires striped with yellow and red. Use a small pair of pliers to gently pull this off its connection. Replace with the matching plug from the 4-cuvette frame. Refer to section 2 page 2-12 for pictures of the thermocouple and PAR connections. 4. Cuvette preparation: a. Clean the exterior of the bottles with a soft cloth, removing any smears or dust. Any bottles that are scratched or otherwise compromised should be replaced. b. Clean the black gaskets that line the interior edges of the articulating cuvette bases. Replace foam central gasket strip and/or black gasket with material in the Green metal LiCor tool box in the glass-fronted cabinet to the left of the refrigerator. Alert the lab manager if you are using up most of the gasket material so that more can be obtained. c. Adjust the angle of the leaf thermocouple on the sensor rod (attached to the bottle cap) so that is angles downward and to the side of the rod (i.e. is not directly below the rod or the PAR sensor). d. Adjust the angle of the PAR sensor (the little rectangular block on the sensor rod so that it is approximately parallel to the counter top. e. Adjust the white/beige plastic tension ring on the inside of the bottle base and the black gasket material on the outside of the bottle base so that they are directly aligned.

f. Close the cuvette with the bottle in place and tighten the knurled knob on the left of the closing mechanism. 5. Gas flow rate: Gas flow rate is not controlled by the IRGA console when sampling from the whole plant cuvettes. Instead, the manual flow meters at the back of the cuvette fram are used to control flow rate. There are two manual flow meters. The left one is labeled Sample and the right one is labeled Reference. Labels are also attached to indicate the appropriate settings to achieve 250 ml and 500 ml. per minute. a. Turn on the outlet strip to the left of the cuvette frame. You will hear the pump come on in the cabinet under the cuvette frame. Open the cabinet. Observe the 50- litre mixing bottle to the right and the small pump to the left. Observe the four white plastic 1/2 tubes. One of these is labeled A and one B. They originate in chambers 101 and 102 respectively. The chambers can be used to supply a large volume of gas with constant CO2 concentration. Make sure the correct hose (as determined by experimental design) is connected so that the correct gas is supplied. b. Adjust the Reference flow rate to the level specified in your experimental design. c. Now check the Sample flow rates from the individual cuvettes. Note the black cuvette selector switch box on the left side of the cuvette frame. Each switch is labeled with a number on colored tape. The cuvettes are labeled with a corresponding number and tape color. i. Push each of the four cuvette selector buttons in succession, noting the Sample flow rate on the left-hand flow controller (labeled Sample ) as you do this. The cuvettes with highest Sample flow rates should have flow rates above the flow rate specified in your experimental design., The flow rate will be adjusted downward later, but first we want all cuvettes supplied with a higher level through their individual adjustment valves. Then we will adjust the Sample flow controller downward. This insures positive pressure in the system. ii. If any cuvette has a notably lower flow rate, check the tightness of the knob on the closing mechanism and check the gaskets for cleanliness and integrity. Also check that the connector tubing at the base and top is properly connected. If you are convinced that the seals are good, go on the the next step. iii. Adjust the small brass valve on the incoming gas line to increase the flow. This valve sits under the cross-piece that supports the cuvettes. Note the two 1/4 tubes that feed gas to the two sides of the cuvette base. These tubes originate at a T under and behind the cuvette base. On the incoming 1/4 gas tube, just back from this brass T you will find the brass valve. VERY SLIGHT adjustments are sufficient. If VERY SLIGHT adjustments do help, the problem lies elsewhere. Seek help. When all cuvettes have equal flow rates, go to the next step. iv. Adjust the Sample flow controller downward to the flow level specified in your experimental design. Do this by turning the knob at the base of the flow controller to the right until the silver ball rests at the appropriate mark. v. Check to see that all cuvettes produce this same flow rate. 6. Light levels: Using the LiCor Li1000 Data Logger, check the light level in the center bottom of each cuvette. Do this by opening the cuvette and holding the quantum flux probe so that it is in the center of the cuvette with its top surface horizontal. Adjust the halogen lights attached to the top of the cuvette frame to achieve the intended light level. Starting the LiCor 6400

1. Turn on the main unit. The power switch is on the right-hand side of the main unit of the Li6400. Remember to have the external power source or batteries plugged in. Even with the external power supply, you will need to have a charged battery plugged in if you want to power the IRGA head. Without either two batteries, or the power supply cord and one battery, the main unit can be powered, but the IRGA head cannot. You will not get an error message. The machine will instead just repeatedly re-start the set-up sequence. The LCD will display the message "initializing", then "Loading Open System 3.4...". A window will appear requesting that you choose a configuration. Choose "Factory Default" After a few minutes, you will be asked if the IRGA is connected. Press "Y". You should hear fans starting inside the main unit and a clicking noise coming from the IRGA head. If instead you get the message "initializing..." then you need to replace the battery with a freshly charged one. The battery is in an open compartment on the bottom right of the main unit. After the whirring and clicking noises start, the LCD will display a screen entitled "LI- 6400 Photosynthesis System". Wait 30 minutes for the IRGAs to warm up. At the bottom of the screen, you will see a series of options. At the start of each week's work, choose "Calib(ration) Menu". Follow the directions in the Primer for both the "IRGA zero" and "IRGA span" options. Section 4 page 7.. These two procedures set the zero point and the gain. This programs the IRGA to calibrate the slope of the relationship between current output of the infra-red light sensor in the IRGAs and the concentration of CO 2 displayed on the LCD and recorded in the data set. IRGA zero option: turn the front-most knurled knob on the left top of the main unit all the way to back (labeled "scrub"). This will scrub all CO 2 from the air entering the IRGA. Follow the directions in the Primer/ on the screen to calibrate the zero point. IRGA span option: Return to the Calib Menu screen and select "IRGA span". Turn the same knurled knob toward you all the way (the "bypass" direction). A tank of 500 ppm CO 2 is attached to a glass and metal flow meter located behind the the main unit. This flow meter is labeled "CALIBRATION". Open the main valve on the attached gas tank until the dial shows gas pressure. Open the fine adjustment valve on the left of the gas tank regulator until the sliver ball on the calibration flow meter behind the unit reads at least 11. Locate the black plastic 1/4" tube that comes out of the flow meter. Push the end onto the air input port on the right side of the main unit. This is the small silvercolored port below the sample and reference ports. If you hear the pumps in the main unit speed up and get louder, open the fine adjustment on the gas tank regulator a little more until the pumps in the main unit quiet down again. Follow the directions in the primer/on the screen for adjusting the span calibration. When both Sample and Reference CO2 have been adjusted, quit the IRGA span program and return to the LI 6400 operating system screen. Disconnect the calibration gas from the inlet port and close the main valve on the calibration gas tank.

Connect the Cuvette system Disconnect the the cuvette gas supply from the main unit and connect the gas lines from the cuvette frame according to the directions in step 1 of "Preparing the 4-cuvette system" above. From the "LI-6400 Photosynthesis System" menu, select "New Measurements". This should bring up a screen with all the output measurements displayed, including RCO2, SCO2, etc. If instead you get a small window displaying "CORRUPT ARRAY press enter", then press "enter". This is an error that we cannot fix until we replace the mother board. This will take you to the 'Welcome Menu", showing the following choices along the top: "Editor Filer Run Shell Xchng". Press "r" for Run. The default program name will appear. Press enter or select. The program will begin reloading. In my experience, the second time around, you will get the new measurements window without a hitch. At the bottom of the New Measurements screen, there are a number of options that can be selected by pressing the red keys labeled f1 - f(6?). Towards the left is an option labeled "open log file" or maybe "new log file". Press the red function key below it. You will be prompted to enter a file name. Files should be named according to the treatment, run and tray. For example, RHH41. If more than one day is required to finish a tray, the second day's file would be named RHH41a. Begin loading plants into the cuvettes: consult an experienced person first. Plants cannot not have water on them, but the soil should stay moist. Plants should be held in the same light levels in which they will be measured until they are queued into the cuvette series. Do not bring more plants out of the chamber than can be run in 20 minutes. When the system is run at a flow rate of 250 ml/min., a plant must be in its cuvette for 10-12 minutes before it is measured. You can only effectively make use of three cuvettes at this flow rate. The thermocouple must be in contact with a leaf and par sensor should be placed as close to the center of the cuvette as possible without shading any leaves. Plants should be placed into cuvettes at six minute intervals. The table below illustrates the timing of events in each cuvette. "Switch #1" means that you should depress the #1 selector button on black selector box on the left of the cuvette frame, thereby directing the gas flow from cuvette #1 to the IRGA and connecting the cuvette #1 sensors to the sensor connections. Time since start (min) Cuvette1 Cuvette2 Cuvette3 Cuvette4 0 Load empty empty empty 6 switch #1 Load empty empty 12 record switch #2 Load empty 18 Load record switch #3 empty 24 switch #1 Load record " 30 record switch #2 Load " 36 Load record switch #3 " etc. With larger plants, the flow rate can be increased, and the time between loading and recording reduced proportionately. For example, if the flow rate is doubled to 500 ml/min, the time intervals in the table can be reduced by half to three minutes intervals. Before recording the measurements for a given plant, its pot number and cuvette number must be recorded. Press the red function button f4 underneath the "remark" option. In the remark entry window, type the pot number, a comma, then the cuvette number. For example, for plant 2164 in cuvette 3, type "2164,3" and press "enter".

Six minutes after you depress the selector switch for the cuvette in question, if the plant is producing a stable ΔCO2, press the function button under the "LOG" option. The unit will beep and the small number under the word "LOG" will increment by 1. If the ΔCO2 reading stabilizes before six minutes, you may log the measurement sooner than six minutes and move on. Once logged, the plant can be removed from the cuvette and transferred to the holding rack for fluorometry measurement. Constantly check the display to be sure that the measures seem realistic. If readings do not appear to be realistic (if there is no net CO2 uptake, or a large, healthy-looking plant is giving small net CO2 uptake), or there is little or no change in water content between the sample and reference air streams, contact the lab manager or Steve for troubleshooting. When you are finished, follow the instructions in section 3 page 7 for shutting the 6400, with the IRGAs off. Disconnect the tubing and cabling for the IRGA head. The main unit may now be carried to the Mac for data transfer. Now transfer the data to the Macintosh. See Downloading IRGA protocol.