CTG Certification Protocol -- Version 3.0 SUMMARY

Similar documents
NuAire Model NU PR797 Positive Pressure Recirculating Compounding Aseptic Isolator Performance Evaluation Compliance to USP797

HPI-4N1 NEGATIVE PRESSURE COMPOUNDING ASEPTIC ISOLATOR PERFORMANCE EVALUATION

NuAire Model NU NR797 Negative Pressure Recirculating Compounding Aseptic Containment Isolator Performance Evaluation Compliance to USP797

NuAire Model NU NTE797 Negative Pressure Total Exhaust Compounding Aseptic Containment Isolator Performance Evaluation Compliance to USP797

CETA Compounding Isolator Testing Guide CAG Revised December 8, 2008

PERFORMANCE QUALIFICATION AND OCCUPATIONAL EXPOSURE LIMIT (OEL) FOR ISOLATOR

STANDARDS FOR COMPRESSED GAS TESTING

EuroBioConcept s.a.s Hospital Pharmacy Isolator Systems

Laminar Flow LFGI (CAI or CACI) User Manual

Whitepaper Technical Opinion: Grade A - Air Supply and Local Protection

VALITEQ DMS v2.o Document Management Software Tutorial

Kansas State University Fume Hood Operation

Facility design for containment. Gordon Farquharson, July 2016

Micro Dial-Flowmeter. Instructions for Use

DESIGN AND INSTALLATION CONSIDERATIONS FOR TSI AEROTRAK VHP-RESISTANT REMOTE PARTICLE COUNTERS

CLEANLINESS TEST OF PROBE SPINUP EXHAUST LINES

Draft. Equipment/Instrument

IC Sentinel Applications Guide for Healthcare Environmental Air Quality Monitoring

Regulatory Basics for Facility Design (WHO GMP): Current GMP Requirements

Oxygen Dialflow Meter. Instructions for Use

The routine maintenance and inspection of local exhaust ventilation (LEV)

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

Aseptic Techniques. Techniques for Sterile Compounding. Pharmacy Technician Training Systems Passassured, LLC

Outside Air Nonresidential HVAC Stakeholder Meeting #2 California Statewide Utility Codes and Standards Program

HVAC Air Duct Leakage. Tech University 2013 Presented By: Mark Terzigni Project Manager Technical Services SMACNA

Piped Air SCBA Refilling System Standard

Safety Services. Introduction

Building Official's Guide to Duct Testing based on 2009 IECC

TECHNICAL INFORMATION

LABORATORY SAFETY SERIES: Laboratory Hoods

Respirator Fit Testing DEPARTMENT OF LABOR Information on the Requirements Types of Respirators & Filters

Leaded Negative Pressure Isolator User/Service Manual. Biodex Model & Germfree LFGI Series (CACI)

Met One E-BAM Particulate Monitor

Group Quality Certification System of Performance Test for Clean Rooms (CR Mark)

Pressure Regulator. Instructions for Use

An enterprise of United McGill Corporation Family owned and operated since System Pressure Testing for Leaks McGill AirFlow LLC

Technical Information

PERFORM Operating Document

PERFORM Operating Document

GSE CAGING BALLAST SYSTEM INSTALLATION

CORESTA RECOMMENDED METHOD N 6

"LABORATORY SAFETY SERIES"

The primary purpose of laboratory fume hoods is to keep toxic or irritating vapors out of the general laboratory working area.

Biography. Abstract. Keywords. Body box, garments, validation, calibration, clean garments, IEST-RP-CC003.3, garment. Introduction and Scope

Aseptic processing. the starting material to the finished product, should be controlled by Validation Control of the aseptic process

Oxygen Dialflow Meter. Instructions for Use

Guidance on room integrity testing and its interpretation

Significant Change to Dairy Heat Treatment Equipment and Systems

STANFORD UNIVERSITY W.W. HANSEN EXPERIMENTAL PHYSICS LABORATORY GRAVITY PROBE B, RELATIVITY GYROSCOPE EXPERIMENT STANFORD, CALIFORNIA

Glove Management. PDA Ireland Chapter Dublin 04 th Oct Dr. Aidan Harrington DPS Group

VINAYAK PHARMA TECHNOLOGY

Design DSA Steam-Atomized Desuperheater

Integrity Testing LifeASSURE BNA045 and BNA065 Series Filter Cartridges

TECHNICAL INFORMATION

Instruction Bulletin Addendum For Pressure-Balanced In Situ Oxygen Probes

Sole reliance for sterility or other quality aspects must not be placed on finished product test.

Control of Nuclear Gloveboxes and Enclosures Using the No- Moving-Part Vortex Amplifier (VXA)

2005 CERTIFICATE OF ACCEPTANCE (Part 1 of 3) MECH-1-A

R2S Standard Sampling Assembly OPERATION PROCEDURE

1. The laboratory supervisor must enforce the institutional policies that control access to the laboratory.

Oxygen Usage Best Practice Guide

Walk in Modular Material Transfer Chamber Including Integration of a Bioquell Clarus L2 hydrogen peroxide vapour (HPV) generator

Installation, Operation, and Maintenance Manual Parker Balston Model N2-04 Nitrogen Generator

University of North Florida Performance Requirements For Laboratory Fume Hoods. As Manufactured Performance Tests. As Installed Performance Tests

Summary of Substantive Changes between the 2004 e1 and 2017 editions of ASSE 1011, Hose Connection Vacuum Breakers

BERNER FlowSafe I-[MaxPro] ³

CentriFlow Meter Integrated Air Entrainment System

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

Discussion and guidance on the definition and qualification of porous loads

A. Evaluation of the written program

Vertebrate Animal Biosafety Level 1 Criteria

1 5 2 Definition of Isolator Technology Isolator Types and Applications 3 Critical Design Considerations 4 Biodecontamination Conclusion and Summary 2

Pneumator. Testo Industrial Services More assurance, better service. Pressure calibrator and measuring instrument.

AIR/OXYGEN BLENDER INSTRUCTION MANUAL

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

OPERATOR, PRODUCT AND ENVIRONMENTAL PROTECTION

OPERATOR, PRODUCT AND ENVIRONMENTAL PROTECTION

Air Handling Unit (AHU) Air handling unit which serves to condition the air and provide the required air movement within a facility.

Perform Pressure & Leak Test, Tubing & Piping. Module 12306

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

IH Report: SUFMO 08/09-CC21 1/27/2009 Stanford University Gas Cabinet Installation and Performance Criteria

Mott Manufacturing Limited

Title: Standard Operating Procedure for Elemental and Organic Carbon (EC and OC) using Non-Dispersive Infrared Detection (NDIR)

Installing a Probe. Typical Installation

METHOD 21 - DETERMINATION OF VOLATILE ORGANIC COMPOUND LEAKS. 1.2 Scope. This method is applicable for the

USM21 Sealed Bimetallic Steam Trap for use with Pipeline Connectors Installation and Maintenance Instructions

Model PDT Dewpoint Transmitter

ANDERSON GREENWOOD SERIES 9000 POSRV INSTALLATION AND MAINTENANCE INSTRUCTIONS

ACCESSORY KIT INSTALLATION INSTRUCTIONS

ATTACHMENT O WIPP MINE VENTILATION RATE MONITORING PLAN

Gas Injection Systems

1.0 PURPOSE The purpose of this procedure is to describe the use of basic aseptic technique.

product manual HM-4140, HM-4150, HM-4160 HM-4160A HM-4150 Humboldt FlexPanels

with O 2 Controller Instruction Manual

American Society of Sanitary Engineering PRODUCT (SEAL) LISTING PROGRAM

EMT EnvironmentalMonitoringTECHNOLOGIES

ProSys Laboratory Analytical Isolators

Facility management and equipment. Institut Pasteur du Maroc

Calibration Procedure. PosiTest DFT Combo

Phalanx Alpha Phalanx Beta and Ultra Elite

Transcription:

Date of Test: Test Report Number: SUMMARY Date of Prior Certification: Certified by: Re-certification Due Date: Unit Description: Make : Model: Serial Number: Location: CERTIFIER INFORMATION Name: Client Information: Client: Contact: Address: City, State: Phone: E-mail: Company: Phone: Results 4.1 Particulate Test (non-viable) ISO Class 5 Pass Fail 4.2 First Air Test - ISO Class 5 Pass Fail 4.3 Egress Test - ISO Class 5 Pass Fail 4.4 Pressure Differential Test Pass Fail 4.5 HEPA Filter Integrity Test Pass Fail 5.0 Viable Samples Taken Airborne PEC Airborne Area Surface PEC Finger Tip Yes or No 6.0 Unidirectional Airflow Smoke Test CTG Certification Protocol Version 3.1 (2-5-2018) Unidirectional Airflow Yes No CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 1 of 16

Signature below verifies that you have read and understand the protocols set forth by Containment Technologies Group, Inc. for certification of CTG products. The protocols are based on compliance to FDA guidelines and USP standards*. Questions are comments should be addressed hrahe@mic4.com Protocol Approvals Date: Protocol Executed by Protocol Prepared by: Technical and Regulatory Consultant Hank Rahe *USP <797> contains tests that represents a potential health hazard if preform in a pharmacy environment. These tests is excluded from consideration when preparing this protocol CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 2 of 16

Section Number Contents Page Number Summary Results of Certification Protocol 1 Protocol Acknowledgement 2 Table of Contents 3 1.0 Purpose 4 1.1 Purpose of the certification protocol 4 1.2 Verification Approach 5 1.3 Testing Condition 6 2.0 System Overview 7 3.0 References 8 4.0 Performance Test Procedures 9 4.1 Particle Counts 9 4.2 First Air 10 4.3 Egress 11 4.4 Pressure Differential 12 4.5 HEPA Filter Integrity Test 13-14 5.0 Viable Sampling 15 6.0 Unidirectional Airflow Smoke Test CTG Certification Protocol Version 3.1 (2-5-2018) 16 CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 3 of 16

1.0 Purpose 1.1 The purpose of the certification protocol is to verify that the isolator is operating as designed after installation. The design criteria for the isolator is to meet federal and state regulations and standards. Verification of the internal environment is accomplished by specified testing procedures described in this document. After testing is complete the results of the test are documented on form Appendix A. Note: It is acceptable to substitute different formats for certification given the information is required and referenced by regulatory agencies. The current standards for air quality particulate measurement is the International Organization for Standardization (ISO). ISO Class 5 particulate air quality is required for aseptic manipulations in the critical preparation zone by both FDA and state boards of pharmacy. The following ISO standards apply to the particulate air quality specifications, test methods and acceptance criteria. 1. ISO 14644-1:2015(E) Cleanroom and associate controlled environments 2. ISO 14644-2: 2015 (E)-Specifications for testing and monitoring to prove continued compliance with ISO 14644-1:2015 (E) 3. ISO 14644-3:2005(E) Test methods 4. ISO 14644-7:2004(E) Separative devices ( clean air hoods, gloveboxes, isolators and mini-environments Note: The ISO documents are the basis for particulate air quality test described. Other guidance and standards requirements utilized in the certification process are: 1. FDA Guidance for Industry - Sterile Drug Products Produced by Aseptic Processing Current Good Manufacturing Practice; 2004 2. United States Pharmacopeia Standards*: USP <797> 2008; USP <800> 2018 CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 4 of 16

1.2 Verification Approach Verification will consist of testing isolators to verify that both particulate air quality, FDA guidance and required standards are met. Critical parameters for the isolator are meeting ISO 5 conditions in the critical zone where compounding occurs, first air requirements, airflow design, and egress requirements per USP <797>. The isolators are designed with HEPA filtration for both inlet and exhaust air. Airflow design is based on regulatory requirements and falls into three categories: 1. Re-circulating 2. Once through 3. Once through exhausted In the re-circulating system the air passes through two HEPA filters. In the once through the inlet air is HEPA but not the exhaust and the once through exhausted is HEPA filtered both inlet and exhaust. The inlet HEPA filter is the critical filter for controlling particulate and this filter is tested per the HEPA filter integrity test. For facilities requiring once through exhaust the exhaust filter should also be tested. Testing frequency for ISO 5 environments or when the chamber is moved or modified in such a way that the modification would impact the filter system. Test that involve particulate air quality are to be valuated against the ISO 14644-1:2015(E) standard. CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 5 of 16

1.3 Testing Conditions ISO 14644-1:2015 (E) Cleanrooms and associated controlled environments PART 1 Classification of air cleanliness by particulate concentration specifies a standard method of testing to determine particulate cleanliness. The test methods described in ISO 14644-3-2005(E) describes three Occupancy states or conditions under which testing is to take place. 1. As-built 2. At-rest 3. Operational USP standards call for dynamic testing which is the same as to the operational testing conditions described in the ISO standard. The specific air quality tests in this protocol to demonstrate compliance with both ISO standards, FDA, USP and state pharmacy board regulations are as follows: 1. Chamber particulate counts (Direct Compounding Area (DCA)) 2. First air 3. Egress The air quality results of each of these tests are to be ISO class 5 not more than 3520 particle of 0.5 microns or larger per m 3. The following sequence of compounding operations should be followed to create dynamic compounding conditions 1. Load materials for manipulations into trays 2. Place trays into airlock/ transition chamber 3. Spray decontaminating agent (hydrogen peroxide or sterile alcohol into chamber 4. Close door on chamber 5. Enter gloves/ sleeves 6. Open inter chamber door 7. Remove tray 8. Close door 9. Perform manipulations 10. Place finished compound into tray and place into airlock/transition chamber CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 6 of 16

2.0 System Overview 2.1 System Boundaries The MIC / MIC/EDU isolator consist of the following components: 1. Main chamber 2. Airlock / transition chamber 3. Glove ports for personnel interaction 4. Air handling system a. Inlet HEPA filter b. Pressure control valve c. Blower d. Pressure control valve e. Outlet HEPA filter The 797P consist of the following components 1. Main chamber 2. Airlock / transition chamber 3. Glove ports for personnel interaction 4. Air handling system a. Inlet HEPA filter b. Blower c. Pressure control plate The 800N consist of the following components 1. Main chamber 2. Airlock / transition chamber 3. Glove ports for personnel interaction 4. Air handling system a. Inlet HEPA filter b. Pressure control plate c. Blower d. Outlet HEPA filter e. Gate Valve CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 7 of 16

3.0 References 3.1 Owner s Manual 3.2 Industry Guidance and Standards: a. ISO 14644-1:2015 (E) Cleanrooms and associated controlled environments PART 1 Classification of air cleanliness by particulate concentration b. ISO 14644-2:2015 (E) Specifications for testing and monitoring to prove continued compliance with ISO 14644-1:2015 (E) c. ISO 14644-3:2005(E) Test methods d. ISO 14644-7:2004(E) Separative Devices ( clean air hoods, gloveboxes, isolators and mini-environments e. FDA Guidance for Industry - Sterile Drug Products Produced by Aseptic Processing Current Good Manufacturing Practice; 2004 f. United States Pharmacopeia Standards: USP <797> 2008 *; USP <800> 2018 g. IEST-RP-CC034 CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 8 of 16

4.0 Performance Test Procedures 4.1 Particulate Test Non-viable Item 4.1 Particulate Test Purpose: Determine compliance to ISO 14644-1:2015(E) Class 5 Particulate count less than 3520 particles per cubic meter 0.5 microns or greater Regulatory Requirement USP <797>*, USP <800>*, FDA guidance 1 Acceptance Criteria: Equipment required: Particulate count less than 3520 particles per cubic meter 0.5 microns or greater ISO class 5 Calibrated Particulate Counter 1 FDA Guidance for Industry - Sterile Drug Products Produced by Aseptic Processing 2004 Test Instructions 1. Establish background air quality where the isolator is located. Take initial air samples with the zero filter in place (verifies the sample tubing is clean and free of leaks) 2. Place sample probe on top of the airlock, take two (2) samples and record results on Summary Requalification Report. If area background exceeds 18,500,000 particles contact CTG for additional instructions. Sample 1 Sample 2 Average 3. Test will consist of three locations within the critical compounding zone which is defined as between the glove ports (outside from the center of the zone). Location A- Centerline of left gloveport and twelve inches from back wall; Location B Twenty inches from left wall and twelve inches from back wall is twelve inches from the back wall and Location C is twelve inches from the back wall and six inches from the left wall. 4. Place probe into the isolator through the airlock. The particle counter is external to the isolator chamber (Note when operating the particle counter will remove one CFM from the chamber causing the chamber to go negative). Balance the pressure valves to reduce the negative pressure as must as possible. 5. Connect the particle counter tubing by cutting off a glove finger and inserting the tubing through the hole in the glove. Tape the tubing to the glove making a gas tight seal. 6. Place the glove over the tube opening for the trash container. Check Pressure. 7. Caution when moving probe do not slide 8. Take a sample count clear tubing. Take three counts at each location. Record each count Location Sample 1 Sample 2 Sample 3 Average Location A Location B CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 9 of 16

4.0 Performance Test Procedures 4.2 Particulate Test First Air. Item 4.2 Purpose: First Air Test Verify air exiting the HEPA filter in a unidirectional air stream is essentially particle free. Regulatory Requirement USP <797>*, USP <800>*, FDA guidance 1 Acceptance Criteria: Equipment required to perform this test: Particulate count less than 35 particles per cubic meter 0.5 microns or greater ISO class 5 Reference USP < 788. Calibrated Particulate Counter Test Instructions 1. Test will consist of one location 2. Place probe into the isolator through the airlock. The particle counter is external to the isolator chamber (Note when operating the particle counter will remove one CFM from the chamber causing the chamber to go negative). Balance the pressure valves to reduce the negative pressure as must as possible. 3. Connect the particle counter tubing by cutting off a glove finger and inserting the tubing through the hole in the glove. Tape the tubing to the glove making a gas tight seal. 4. Place the glove over the tube opening for the trash container. 5. Take a sample count at location one to assure any contamination is clear in the tubing. 6. Take two counts at approximately 6 to 12 inched below the center point of the inlet HEPA filter. Record each count and average at each location. Location Sample 1 Sample 2 Average Center Inlet HEPA CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 10 of 16 4.0 Performance Test Procedures

4.3 Particulate Test Egress Test Item 4.3 Purpose: Egress Test Verify during transfer of materials from airlock to main chamber particulate count will not exceed ISO 5. Particulate count less than 3520 particles per cubic meter 0.5 microns or greater Regulatory Requirement USP <797>*, USP <800>* Acceptance Criteria: Equipment required to perform this test: Particulate count less than 3520 particles per cubic meter 0.5 microns or greater during transfer of materials Calibrated Particulate Counter Test Instructions 1. Prepare transfer tray materials for compounding ( vial, syringe and piggy back) 2. Place particle counter probe fourteen inches from the center line of the inter airlock door with the probe facing toward the critical zone. 3. Load transfer tray into airlock by opening outside door and placing it into airlock. Spray down the transfer tray and airlock with decontaminating agent (3% hydrogen peroxide). Close the outside door and wait thirty seconds. 4. Enter the gloves and open the inside airlock door. Note: airlock is filled with decontamination aerosol wait sixty seconds for aerosol to clear before beginning counts. 5. Start particle counter. 6. Repeat test two times Sample 1 2 CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 11 of 16

4.0 Performance Test Procedures 4.4 Pressure Differential Test Item 4.4 Purpose: Pressure Differential Test Verify the isolator is operating within the proper pressure range Regulatory Requirement USP <797>*, USP <800>* Acceptance Criteria: Equipment required to perform this test: Greater than 0.02 inches of water column positive Less than 0.02 inches water column negative Visual Test Instructions The differential pressure acts as a secondary containment barrier either protecting the product or personnel. It is important that this feature is operating correctly. 1. Before checking the differential pressure, verify that the Isolator has been sealed properly. Check for leaks around the trash ports, sharps ports and glove ports. 2. After ensuring there are no leaks in the Isolator, turn the isolator off and verify that the pressure gauge reads zero (it may take up to two minutes for the pressure to reach zero). 3. After the zero pressure reading has been verified, turn the Isolator on and allow three minutes for it to return to proper operating range. Ensure that the pressure differential gauge reads in the range of +/- 0.2 to 0.5 inches of water column. Item Yes No Pressure verified within range CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 12 of 16

4.0 Performance Test Procedures 4.5 HEPA Filter Integrity Test Item 4.5 Purpose: HEPA Filter Integrity Test This test verifies the integrity of the supply HEPA filter, filter housing and gel seals are in situ operating conditions. Regulatory Requirement USP <797>*, USP <800>* Acceptance Criteria: Equipment required to perform this test: Leak rates greater than 0.01% (direct filter scan) of the up stream aerosol concentration are not acceptable. A leak rate greater than 0.01% is considered a leak (reference ISO 14644-3:2005, Annex B, B.6) Aerosol generator and aerosol photometer Note 1: Portions of the test methods have been adopted from IEST-RP-CC034. Note 2: Some Isolators may be set up where access for a direct filter scan is not possible. In these cases, it is acceptable to probe the membrane that separates the chamber from the filter duct. Test Instructions Introduce the aerosol challenge: 1.The output of one Laskin nozzle operating at 20 PSI will be introduced into the blower inlet. Remove the plug adjacent to the pressure control HEPA (PURPLE) on the left side of the Blower. Note: this is a 3/8 inch opening. 2. PAO ug/l of air 13,500 / airflow volume cfm 3. Determine supply air volume A. Measure velocity 2-3 inches below diffuser (3x3 grid nine locations 2 from edge) B. Q cfm = V fpm X A sq ft V fpm = Average of nine velocity readings A sq ft = (10 x 10 ) / 144 sq in / sq ft CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 13 of 16

4.6 HEPA Filter Integrity Test Continued 3. Recommended Test Method Full access for a direct scan of HEPA filter. 1. Make sure the Isolator is on. Gain access to inlet HEPA filter. 2. Using aerosol generator, add aerosol upstream of HEPA filter. 2. Setting the photometer gain: 1. Most digital aerosol photometers will only allow a maximum internal calibration of 100 ug of PAO/liter of air. 2. Set internal cal to 1/3 of challenge 3. You must now divide the leak reading size by 3 For example a.03% photometer reading needs to be divided by 3 and therefore equals.01% Note: Penetration is measured as a percent of the upstream aerosol challenge of cold poly dispersed PAO (Poly-alpha olefins). A forward light scattering aerosol photometer measures downstream penetration. 3. Scan 100% of the HEPA filter face or media. Note: To assure a homogenous mixing of the aerosol to the supply airflow the PAO should be added upstream of the blower. The upstream aerosol concentrations should not vary more than +/- 15% from the average value. Upstream concentration should be verified. Consult IEST-RP-CC034 for additional information. 1. With a photometer sampling rate set at 2.5 cm (one inch) from filter face, scan the downstream filter face of the HEPA filter. The sampling rate is 1.0 CFM (1.7 meter cube per hour). Leak rate greater than 0.01% of the upstream aerosol concentration is considered a leak. 2. If applicable, repair any leaks. Record results. CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 14 of 16

5.0 Viable Sampling Item 5.0 Purpose: Identify Viable Samples Taken Verify viable samples taken. Sample Taken Yes No Airborne PEC Airborne Area Surface PEC Finger Tip CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 15 of 16

6.0 Smoke Test Protocol Reference Unidirectional Airflow Smoke Test Protocol CTG Certification Protocol --- Version 3.1 (2-5-2018) CTG Certification Protocol ------ Version 3.0 (2-5-2018) Page 16 of 16