Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Similar documents
Operating Procedures for Metal Evaporator I

Arizona State University Center for Solid State Electronic Research. Table of Contents. Issue: C Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8

OXFORD PLASMALAB 80PLUS (CLOEY)

Cryo-Evaporator Operation

Standard Operating Procedure. For. PVD E-Beam

Temescal BJD-1800 E-Beam Evaporator 1 System Overview - Cryo-pumped for typical base pressures in the low 10e-7 Torr range. - Four pockets in the

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

R I T. Title: Amray 1830 SEM Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 09/29/03 1 SCOPE 2 REFERENCE DOCUMENTS

Approved by Principal Investigator Date: Approved by Super User: Date:

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

SPUTTER STATION STANDARD OPERATING PROCEDURE

Standard Operating Manual

Oerlikon Sputtering Evaporator SOP

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

Standard Operating Manual

THERMAL EVAPORATION UNIT (for Al evaporation)

Usage Policies Notebook for Parylene Coating System

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

Revised: June 7, 2017

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

OPERATION OF THE DIMPLER

Basic ICP Operating Procedures

Cambridge NanoTech: Savannah S100. Table of Contents

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

The SPI Sputter Coater Handbook

QUORUM TECH 150RES THE FIRST AND THIRD WEEK WILL BE SET UP FOR CARBON COATING THE SECOND AND LAST WEEK WILL BE SET UP FOR GOLD COATING

MMRC PVD Evaporator Instructions

Standard Operating Manual

Nordiko Metal Sputtering System Standard Operating Procedure

Nanofabrication Facility: Sputter Evaporator SOP Rev. 04, June 2007

Arizona State University NanoFab PLASMATHERM 790 RIE. Version A

Unaxis PECVD. SiH4 (5% in He)

Usage Policies Notebook for STS DRIE System

Standard Operating Manual

Arizona State University NanoFab XACTIX ETCHER. Rev A

KARL SUSS MJB3 UV400 Mask Aligner Standard Operating Procedure

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

ABM MASK ALIGNERS. NanoFab 26 March 2009 A Micro Machining & Nanofabrication Facility

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

March CS-1701F Reactive Ion Etcher

OLED vacuum deposition cluster system Sunicel Plus 400 / Plus 400L (Sunic Systems)

Usage Policies Notebook for Trion RIE / ICP Dry Etch

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

Login to ilab Kiosk. Revised 05/22/2018. Load your sample:

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Notes-PECVD: Chamber 1

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

STS ICP-RIE. Scott Munro (2-4826,

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

LESKER #1 STANDARD OPERATION PROCEDURE

Mass Spec will not Autotune

2 Switching off ECRIS 3. 4 Acquire Beam Spectrum 4. 6 Vent Instrument Chamber 5. 7 Pump down Chamber 5. 8 Baking out Chamber 5.

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

COBILT CA-800 Mask Aligner Equipment Operation

Angstrom Dielectric Sputterer Operation Manual

VACUUM TECHNIQUES. Why? How else might you make the measurement?

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

1)! DO NOT PROCEED BEYOND THIS MARK

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

Savannah S100 ALD at SCIF, UC Merced Standard operating Procedure

Usage Policies Notebook for Thermco Atmospheric Diffusion Furnace system

Standard Operating Manual

University of MN, Minnesota Nano Center Standard Operating Procedure

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Chapter 2 General description of the system

BASICS of MAINTENANCE and TROUBLESHOOTING of VACUUM DIFFUSION PUMPS

Plasma Asher: March PX-500 User guide (May-30, 2017)

Equipment Standard Operating Procedure Greg Allion and Kimberly Appel

Warnings: Notes: Revised: October 5, 2015

Mass Spec will not pump down

QUARTZ HOUSING COATING PROCEDURE

STS PECVD Instructions

University of MN, Minnesota Nano Center Standard Operating Procedure

Usage Policies Notebook for BMR HIDEP Low Temperature Plasma Enhanced CVD (LTPECVD)

Usage Policies Notebook for Karl Suss MA6 Mid / Deep UV Mask Aligner

Approved by Principal Investigator Date: Approved by Super User: Date:

DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3

Operating Procedures for the. SAMCO ICP RIE System

CHEMICAL ENGINEERING SENIOR LABORATORY CHEG Initiated Chemical Vapor Deposition

Plasma Cleaner. Yamato Scientific America. Contents. Innovating Science for Over 125 Years. Gas Plasma Dry Cleaner PDC200/210/510 PDC610G.

March Asher Operation

Operation of the Perkin Elmer TGA-GC/MS

NORDSON MARCH PX-1000 PLASMA ASHER STANDARD OPERATING PROCEDURE Version: 1.0 July 2016

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

Xactix XeF2 OPERATION MANUAL

STS DRIE User SOP Location: NRF Cleanroon, Dry Etch Bay

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

University of Minnesota Nano Center Standard Operating Procedure

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Operating Instructions

RAPID THERMAL PROCESSOR (Annealsys AS-ONE 150) Lab Manual

Operation of the mask aligner MJB-55

MINIBRUTE ANNEAL TUBE STANDARD OPERATION PROCEDURE

BEST KNOWN METHODS. Transpector XPR3 Gas Analysis System. 1 of 6 DESCRIPTION XPR3 APPLICATIONS PHYSICAL INSTALLATION

Introduction of Vacuum Science & Technology. Diffusion pumps used on the Calutron mass spectrometers during the Manhattan Project.

Transcription:

Title: CHA E-Beam Evaporator Page 1 of 7 Table of Contents Purpose/Scope... 2 2.0 Reference Documents... 2 3.0 Equipment/Supplies/Material... 2 4.0 Safety... 2 5.0 Set Up Procedures... 2 5.1 PC Logon and Chamber/Source Preparation... 2 6.0 Procedure... 3 6.1 Chamber pump down... 3 6.2 TELEMARK POWER SUPPLY CONTROLLER... 4 6.3 Deposition and shut down... 4 7.0 Forms... 6 7.1 Log Sheet... 6

Title: CHA E-Beam Evaporator Page 2 of 7 Purpose/Scope 1.1 This document covers the procedure that should be followed for normal operation of the CHA E-beam evaporator for the purpose of depositing ferrous, non-ferrous & dielectrics on substrate materials that might be used for research purposes. It is suggested that you review this document thoroughly before proceeding with the operation of this tool & always check with staff when using a new target material, to verify if special precautions need to be taken to reduce cross contamination. If an error condition occurs at any time during your process immediately shut off the power supplies if they are on & place the tool in a safe mode. Place the sign on the tool to down, contact CSSER staff & submit a service request. 2.0 Reference Documents 2.1 Chemical Safety & Hazardous Waste Management Rules & Procedures Handbook 3.0 Equipment/Supplies/Material 3.1 Isopropyl Alcohol 3.2 Nilfisk clean room vacuum 3.3 Cleanroom wipes 3.4 Source Material 3.5 Graphite Crucible Liner 4.0 Safety 4.1 Follow all safety procedures outlined in the CSSER Handbook mentioned in section 2.1 4.2 Follow safety procedures for high voltage when working with high voltage or RF energy. 4.3 Follow safety and handling procedures when working with high pressures, pyroforic, or toxic gases. 5.0 Set Up Procedures 5.1 PC Logon and Chamber/Source Preparation 5.1.1 Log into the PC, using your ASURITE/ASUAD ID, 5.1.2 Ensure that the Q-pod software is running. There is a Q-pod shortcut on the desktop. Select the DataLog Off option on the Q-pod display. Also, if XTAL FAIL is displayed the crystal will have to be replaced before any further processing. 5.1.3 Refer to the ON-BOARD display; ensure the MONITOR button is lit. The display should read CRYO ON 12K. The 12K indicates the cryp pump temperature. It has an operating range of 12 to 17 degrees. If the temperature is above 19 K, then submit a service request and contact CSSER staff. 5.1.4 Fill out machine logbook with name, date, time, and material you intend to deposit. 5.1.5 Ensure that FILAMENT is off on the ION GAUGE CONTROLLER. 5.1.6 On the VACUUM CONTROLLER, select STANDBY with the Auto Control switch. Wait until all of the valves are closed. Then select VENT with the Auto Control switch. This will bring the chamber to atmospheric pressure. It will take approximately 2 to 3 minutes. When the chamber is vented, select STANDBY to close the nitrogen vent valve. 5.1.7 Switch Hoist toggle switch to RAISE to lift the bell jar. (Rocker switch will only switch one position at a time. Place at center position, and then switch to desired function. Hoist will not

Title: CHA E-Beam Evaporator Page 3 of 7 6.0 Procedure operate until the chamber pressure switch senses the chamber has reached atmosphere.) Hoist will stop automatically. 5.1.8 Remove carousel from track and place upside down to load samples. The carousel will be loaded with dummy wafers. The dummy wafers can be removed and placed in a wafer box that is provided. 5.1.9 Open the shutter and place material into a source pocket. Aluminum is provided by CSSER, typically in position 2. If other than Aluminum, use the crucible indexer knob on the left side front of the tools tank cover, rotate to desired pocket 2, 3, or 4. Use tweezers to place and fill each crucible liner. (Source material must be filled to where beam is visible in window, 50% minimum, covering the bottom of the Crucible liner.) 5.1.10 Close the shutter and replace carousel with samples onto the track. 5.1.11 Wipe down bell jar O-ring and associated chamber base plate with isopropyl alcohol only! Ensure there is no particulate generated from handling chamber hardware. (If so, vacuum out before starting pump down.) Chamber O-ring and base plate should be cleaned every time chamber is opened. 5.1.12 Switch Hoist toggle switch to LOWER position, keeping hands away from bell jar as it lowers to chamber base plate. 5.1.13 Enter the Density and Z-ratio for your material and the tooling factor in their respective fields. Note: There is an extensive material data table in the help section of the Q-pod software. Not all of these materials are allowed in the tool. Please verify with CSSER staff before running any new material. 5.1.14 Ensure that the tooling factor is equal to 120% 6.1 Chamber pump down 6.1.1 On the Q-pod display, select the START button to start the Run Time. The Run Time can be used to monitor the time it takes to evacuate the chamber. 6.1.2 On the VACUUM CONTROLLER, select START with the Auto-Control switch. This will begin the process of evacuating the chamber. 6.1.3 After the controller switches to high vacuum (indicated by light on display), ensure that power to the ION GAUGE CONTROLLER is ON. 6.1.4 On the ION GAUGE CONTROLLER, turn rate switch to the 4 position. Press the FILAMENT switch. Refer to the IONIZATION GAUGE, when the value drops below 1 than the range switch can be increased to a higher range position. 6.1.5 At 5x10-5 Torr, press DeGas switch to turn ON, you may leave it on as long as 30 minutes but it is recommended at least 2 minutes. Press switch to turn OFF. (IONIZATION GAUGE may swing to full scale position and shut off the filament. This is normal, simply turn the range switch to a lower setting, (i.e. if in 5 then switch to 4), and switch FILAMENT on again to monitor chamber pressure. 6.1.6 Allow time for the vacuum system to pump the chamber down to 3x10-6 Torr. Enter the chamber pressure into the log sheet. 6.1.7 Place the Motor Control switch to the lo setting. The speed is preset for optimum deposition so you should not have to adjust. (Carousel rotates samples around to obtain a more uniform film; the speed may be monitored by watching motor/chain drive at top of chamber.)

Title: CHA E-Beam Evaporator Page 4 of 7 6.2 TELEMARK POWER SUPPLY CONTROLLER 6.2.1 Turn the Main Power Key to ON. 6.2.2 Press the Main Power ON button. (located under Main Power on the controller). 6.2.3 Wait a few moments for the cooling fans and power supply to warm up. 6.2.4 Press the RESET button (located under High Voltage on the controller). 6.2.5 Verify all interlock LED s are on (located under HIGH VOLTAGE). 6.2.6 Press the ON button (located under High Voltage on the controller). 6.2.7 Ensure that the HIGH VOLTAGE display reads -10.00 ± 0.07 kv. 6.2.8 Verify that the emission current control knob is at zero (located under Source1 on the controller). 6.2.9 Turn sweep control switch ON (located next to the writing that says Power Supply Controller). 6.2.10 Press the Source1 ON button. (it is located under Source1 on the High Voltage controller) 6.2.11 SLOWLY raise the emission current by no more than 10 ma per 10 seconds to the desired current. Wait 1-2 minutes for metal to melt. While metal is heating, ensure that the beam is not hitting the hearth. Also ensure that the metal is not overheating and splattering out of the pocket 6.3 Deposition and shut down 6.3.1 Reset Q-pod to zero thickness value and open the shutter. Ensure MOTOR CONTROL switch is set at lo 6.3.2 During the deposition, watch the beam position through the sight glass to ensure that the spot is not hitting the crucible or anywhere on the gun. If there is a problem with the beam position or oscillation, quickly reduce the beam current and contact CSSER staff. 6.3.3 Halfway thru to the desired thickness, record the emission current, deposition rate, and chamber pressure during evaporation in the logbook. 6.3.4 Close the shutter when the desired thickness is reached. 6.3.5 SLOWLY lower the emission current back to 0. 6.3.6 Press Source1 OFF button. 6.3.7 Press High Voltage OFF button. 6.3.8 Turn sweep control switch to OFF. 6.3.9 Press Power Supply MAIN OFF button and turn Key switch to OFF. 6.3.10 If another metal layer is needed before venting chamber, you may switch locations at this time by rotating the crucibles to the next source location (i.e. 1 4) that you have inserted a material to be deposited. At this time you will also need to re-program the Q-pod for that material & any successive materials that you wish to deposit as well. You will need to restart from section 6.2 for processing purposes. 6.3.11 Slide Motor Control switch to OFF 6.3.12 Press FILAMENT to turn the filament off, on the ION GAUGE CONTROLLER. 6.3.13 Wait several minutes for the crucible to cool before venting the chamber.

Title: CHA E-Beam Evaporator Page 5 of 7 6.3.14 Place the Auto Control rocker switch to its STANDBY position and allow time for valves to close, and then switch to VENT. Allow two to three minutes to vent chamber. Return to STANDBY position. 6.3.15 Switch Hoist rocker switch to RAISE to lift the chamber. Return switch to center position. 6.3.16 Remove carousel and remove your samples. 6.3.17 Remove any source materials you may have installed, and vacuum out any particles from the chamber. 6.3.18 Wipe down any chamber assembly that may have flakes or particulate on them, pay special attention to chamber base plate and wipe chamber seal O-ring every time it is opened with isopropyl alcohol. 6.3.19 Replace carousel and switch Hoist rocker switch to LOWER. 6.3.20 Once the chamber bell jar is seated, switch the Auto Control rocker switch to START. Ensure that the chamber pumps down and crosses over into high vacuum. 6.3.21 Close the Q-pod software, and log off the PC. 6.3.22 Tool should be left in a good state for the next student to use. If you have any problems with tool, make a note in the logbook, contact CSSER staff, and complete a Service Request form on the CSSER website. You must coat the chamber with aluminum after deposition of Gold or other mid gap materials; you may start that deposit at 5E-5 as opposed to a regular deposition.

Title: CHA E-Beam Evaporator Page 6 of 7 7.0 Forms 7.1 Log Sheet CHA Log Sheet Date Name Start Time End Time Material Density Z Ratio Tooling Factor Base Pressure Deposit Pressure Current Rat e Thickness Actual

Title: CHA E-Beam Evaporator Page 7 of 7 Effective Date Originator DESCRIPTION OF REVISION Issue 11/17/08 Paul Boland Change process to reflect manual operation of new power supply and supporting hardware controls. 7/22/09 Jon Martin Update Format D 1/14/11 Jon Martin Change to section 6.2.9 for sweep control to use all materials in same position 7/31/12 Paul Boland Include Q-pod rate monitor instructions. F 8/20/12 Jerry Eller Minor procedure and fix typo s G 11/8/13 Jerry Eller Add instructions for new cryo-pump H C E