Operating manual Analogue Brix Refractometer

Similar documents
Operating manual Analogue refractometer

Manual Refractometer PCE-ALK

Operating Manual Body fat balance

Abbe 5 Refractometer

WXG-4 Manual Polarimeter. Instruction Manual BANTE INSTRUMENTS CO., LTD

Fruit Testing during Long Term Storage

SAUTER FK Version /2016 GB

NovelBright LED Illuminator

The pipettes cover a volume range from 0.1µl to 10ml.

KERN YDB-03 Version /2015 GB

RayPette Plus Autoclavable Pipette. User Manual

Pipettor. User Manual

KERN CM Version /2004

Installation and Maintenance Instruction Manual

Type BBS-03, BBS-05, BBS-06, BBS-25

Operating Instruction English. Revision A Date author NH This operating instruction is not subject to updating. Airgun

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Instruction Manual BA90 Precision Barometer

User Manual PTW-DensiX. D / Di/Zi

Info Counting chamber (haemacytometer)

Operating Instructions Hand Test Pump BHP 700

Instrument Operating Manual

Operating Instructions. ATMOS Chair Comfort. English GA1GB Index 01

Instrument Installation and Performance Verification

Notes Before Use...2 Notes on charging your watch...2. Your Watch...4 Name of Components...6 About the crown...7

Storage Shelving. User Manual [Revision 3.0 May 2017]

SMV-001 Thermostatic Shower Mixer Valve Installation Instructions & Adjustment Settings

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor


Operating Manual. Manual Blaine Air-Permeability Apparatus for determining the fineness of cement EN 196-6

Transferpette -8/-12

Instruction Manual EMA 84 Digital Pressure Gauge

KERN EG/EW Version /02

UNCC Biotechnology and Bioinformatics Camp. Dr. Jennifer Weller Summer 2010

Instruction Manual Contact Pressure Vacuum Gauge

Digital Melting Point Apparatus

Siphon vessel / Priming aid

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Contents. Operating instructions. Pressure switch, heavy-duty version Model PSM-520

Elcometer 138. Bresle Kit and Patches. Operating Instructions

Tex-414-A, Air Content of Freshly Mixed Concrete by the Volumetric Method

Instructions for use Compressed air injector with water trap

Underwater Housing for Canon G7X Mark II

OxyScan Graphic. Operating Instructions. UMS Micro-oxygen sensor 501. Microprocessor instrument

HYDRAULIC MOBILE LIFTING TABLE

Please read these instructions carefully before use! Please consider the safety instructions! Please keep for future reference!

3M Liqui-Cel EXF-14x28 and 14x40 Series Membrane Contactor

Installation and Maintenance Instruction Manual

FIELD TIP SHEET. Mark the bag with black permanent marker before filling it with water.

Elcometer 138. Bresle Kit and Patches. Operating Instructions

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

Underwater Fiber-Optic Plane Irradiance Collector

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

INSTALLATION MANUAL JAPANESE ENGLISH GERMAN FRENCH ITALIAN SPANISH. Before installation. Installation INSTALLATION MANUAL: B700

Any laboratory is equipped with specific tools, equipment,

Polarimeter tubes and accessories

Transferpette. Testing Instructions (SOP) 1. Introduction. May 2009

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

MSDS MATERIAL SAFETY DATA SHEET AMYRIS ESSENTIAL OIL. 341 Christian Street, Oxford, CT 06478, USA

VWR UHP j :01 Page 1. ENGLISH 1 22 DEUTSCH FRANÇAIS ESPAπOL ITALIANO

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE:

Operating Instructions Part No

Underwater Housing for Panasonic Lumix DMC-ZS60, DMC-TZ80

UsER manual for Watersens ph -REDOX

Hygienic Bubble Trap Installation and Operations Manual

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

BOYLE S / CHARLES LAW APPARATUS - 1m long

Harvard Trip Balance BA710. Impact Test Equipment Ltd & User Guide. User Guide

Peak Pressure Indicator MSI-3

GI Series Diaphragm Pressure Gauge Guard

VITLAB micropipette. Standard Operating Procedure (SOP)

WARNING NOTES TABLE OF CONTENTS

BI-680 Online Dissolved Oxygen Controller Instruction Manual

book : t95575.fm Seite 1 Mittwoch, September 20, :10 AM. Avensys. Exposed Single Lever Mixer

KAYAK ERGOMETER. Operating Instructions (Rev 3.0)

Operation Manual. O2 Quickstick. Oxygen Analyzer

VITLAB Volumetric instruments

MULTISCALE REFRACTOMETER

ITT RICHTER CHEMIE-TECHNIK

Sunrise Fountain Water Wonders Assembly & Mounting Instructions

SPS-5. Rate Computing Scale. Operation Manual

Short manual testo

7130 Lancer Rear Drive Magnetic Commercial Indoor Cycling Bike

Instructions for use EN

3M Liqui-Cel EXF-4x13 and 4x28 Series Membrane Contactors

State of Nevada Department of Transportation Materials Division METHOD OF TEST FOR DETERMINING THE LIQUID LIMIT OF SOIL

Tripod Series OPERATOR S MANUAL

New AQF Filter Polymer Filtration System

model 304 ShrinQ PEDOMETER

Type C-CUT. Quickstart. English Deutsch Français

ANALOX O2 PORTABLE OPERATORS MANUAL

Model 405 &407 Small Animal Ventilator. User Manual v3.5

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual

HemoCue Hb Procedure Template

Type 131. Roller base. Information and Operating Instructions for Dewar Vessels Type 131 up to 135 boxed shaped KGW-ISOTHERM

Operating Instructions - Rescue Equipment. RAM Support LRS-T EN Edition (translation of the original operating instructions)

Operating Instructions Melt Pressure Sensor

TQC Bresle Chloride test 2008

Mounting instructions for digger tollo

Blood Pressure Monitor

Transcription:

KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen, Germany E-mail: info@kern-sohn.com Operating manual Analogue Brix Refractometer Tel: +49-[0]7433-9933-0 Fax: +49 (0)7433-9933-149 Internet: www.kern-sohn.com KERN ORA 10 BB ORA 18 BB ORA 20 BB ORA 32 BB ORA 80 BB Version 1.0 01/2015 GB ORA_10_18_20_32_80BB-BA-e-1510

GB KERN ORA Version 1.0 01/2015 Operating instructions Analogue refractometer Contents 1 Technical data... 3 2 Description... 4 3 General information... 5 3.1 Intended use... 5 3.2 Warranty... 5 4 Basic safety information... 6 4.1 Follow the instructions in the operating manual... 6 4.2 Warning... 6 5 Supplied items... 7 6 Before the first use... 7 7 Use/measurement... 7 7.1 Zero point calibration... 8 7.2 Additional advice... 9 7.3 Measuring procedure... 9 8 Cleaning and maintenance... 9 9 Storage... 10 10 Service... 10 11 Disposal... 10 12 Additional information... 10 13 Brix to refractive index (nd) conversion table... 11 14 Annex... 1 ORA_10_18_20_32_80BB-BA-e-1510 2

1 Technical data Models KERN Measuring range and Scales Scale graduations Accuracy Dimensions Product Net weight ORA 10BB 0-10 % Brix 0.1% Brix 189x40x40mm 0.155kg ORA 18BB 0-18 % Brix 0.1% Brix 189x40x40mm 0.155kg ORA 20BB 0-20 % Brix 0.1% Brix 189x40x40mm 0.155kg ORA 32BB 0-32 % Brix 0.2 % Brix 178x40x40mm 0.145kg ORA 80BB 0-80 % Brix 0.5 % Brix 137x40x40mm 0.135 kg Example scale on the ORA 18 BB 3 ORA_10_18_20_32_80BA-e-1510

2 Description 3 4 2 1 5 6 1. Prism cover 2. Prism surface 3. Adjustment screw 4. Eyepiece with rubber eyeshade 5. Diopter adjustment ring 6. Optical tubes with rubber grip 9 8 10 11 7 12 7. Storage box 8. Pipette 9. Refractometer 10. Distilled water / calibration liquid 11. Adjustment tool 12. Cleaning cloth ORA_10_18_20_32_80BB-BA-e-1510 4

3 General information 3.1 Intended use The refractometer is a measuring instrument for determining the refractive index of transparent substances in liquid or in some cases also in the solid state. It is used to observe the behaviour of light as it passes from a prism with known properties to the substance being tested. Use of the refractometer for other purposes is contrary to its intended use and may be hazardous. The manufacturer shall not be liable for any damages caused by improper use. 3.2 Warranty The warranty shall be void in the event of: Failure to observe the instructions in the operating manual Use for purposes other than those described Modifications or opening the device housing Mechanical damage and/or damage resulting from media, liquids, natural wear and tear 5 ORA_10_18_20_32_80BA-e-1510

4 Basic safety information 4.1 Follow the instructions in the operating manual Carefully read through the operating manual even if you have prior experience with KERN refractometers. Every language version includes a non-authoritative translation. The original German document is the definitive version. 4.2 Warning Do not let acids come into contact with skin or eyes. If acid comes into contact with skin, flush with copious amounts of water. Shower if larger areas of skin are affected. If acid comes into contact with eyes, keep the eyelid open and flush the eye with running lukewarm water from the outer corner to the inner corner. Flush eyes for at least 15 minutes. Then consult a doctor or ophthalmologist immediately. Thoroughly clean the refractometer after each use. The refractometer must not be exposed to extreme temperatures, high mechanical stresses, strong direct sunlight or high humidity. This refractometer is not a toy. Keep out of reach of children. Make sure that you will not be hit by anything else while you are using the refractometer, as this could cause serious eye injuries The rubber eyeshade may cause irritation when in prolonged contact with the skin. If this happens, consult your doctor. Do not touch the lenses with your fingers. ORA_10_18_20_32_80BB-BA-e-1510 6

5 Supplied items After unpacking and before using the device for the first time, check that all listed parts have been supplied. Replace damaged or faulty parts immediately and do not put them into operation. Refractometer Storage box Pipette Adjustment tool Cleaning cloth Distilled water / calibration liquid 6 Before the first use Remove the protective film (if present) from the prism surface [2] and check that the rubber eye-cup [4] is fitted correctly. 7 Use/measurement The refractometer can be used to quickly and accurately determine the refractive index of transparent substances, liquid or solid ones. To ensure correct measurement, the measuring device should be adjusted before measurements are carried out. Please make sure your hands are dry before handling the measuring device. 7 ORA_10_18_20_32_80BA-e-1510

7.1 Zero point calibration Hold up the refractometer against a sufficiently bright light source and look through the eyepiece [5], holding the rubber eyeshade [4] closely against your eye/glasses. Rotate the eyepiece [5] to adjust it for your eyesight, until you can see the scale sharply. Open the prism cover [1]. Thoroughly clean the prism [2] and the underside of the prism cover [1] using a soft cloth or soft paper (with alcohol if necessary) and wipe dry. Water should only ever be applied to the measuring prism, not to other parts of the device. Now apply a few drops of the calibration liquid [10] onto the prism surface [2]. close the prism cover [1]. The amount of fluid should be sufficient to moisten most of the prism surface. There should not be any air bubbles between the measuring prism [2] and prism cover [1]. Wait about 30 seconds to allow the temperatures of the test fluid and prism to equalise. Look through the eyepiece [4] while pointing the refractometer's prism surface [2] at a bright light source. Through the eyepiece [4], you will see a bright and a blue field. The boundary line between them shows the refractive index value on the scale which is also visible through the eyepiece [4]. Use the supplied adjustment tool [11] to turn the adjustment screw [3] behind the prism surface [2] (under the rubber cap), and adjust the scale so that the measuring device is optimally set up. The bright/dark boundary line should be at 0% Brix. For refractometers which are not adjusted at 0% Brix using distilled water and have a special calibration liquid, set the bright/dark boundary line to the calibration line in the field of view. The room temperature must be +20 C. For refractometers without automatic temperature compensation (ATC): Convert the measured value to take account of the deviation in room and sample temperature from +20 C. (See the table at the end of this manual, Point 13.) Repeat step 4 (cleaning). Important! The ambient/room temperature and the sample temperature influence the refractometer measuring result. The scales are designed for an ambient temperature of +20 C! If the measurements are carried out at a temperature other than +20 C, the results must be correspondingly corrected. A correction table can be found in the annex, Point 13. ORA_10_18_20_32_80BB-BA-e-1510 8

7.2 Additional advice It is important that the samples being measured are representative samples. For samples that evaporate easily, measurements should be carried out quickly. The samples should be at the same temperature as the measuring instrument in order to achieve an accurate result. 7.3 Measuring procedure Make sure your hands are dry before handling the refractometer. Open the prism cover [1] and use the supplied pipette [8] to apply a few drops of the sample liquid [8] onto the prism surface [2], then close the prism cover [1]. Spread the liquid evenly by pressing down on the prism cover [1] and eliminate any air bubbles present. Hold the device horizontally and wait about 30 seconds (for optimal temperature equalisation between the sample and device). View the measurement scale through the eyepiece [4]. Point the prism surface [2] of the refractometer at a bright light source while doing this. Rotate the adjustment ring [5] on the eyepiece [4] to adjust the focus. The boundary line will move on the measurement scale depending on the concentration. This bright/dark boundary line shows the result on the scale. If the temperature deviates from +20 C, correct the measured result using the corresponding value from the temperature correction table [13]. Carefully clean the supplied pipette [8] and the refractometer after carrying out the measurement. Important! After every measurement, use an absorbent, lint-free cloth to remove the fluids from the prism surface [2] and prism cover [1]. Then carefully clean the prism and prism cover using a cloth moistened with water or if necessary alcohol, and dry both parts using a soft, dry and lint-free cloth. Avoid rubbing the prism [2]. 8 Cleaning and maintenance Clean the refractometer using a clean, lint-free cloth moistened with water or if necessary alcohol. Do not use any aggressive or abrasive cleaning agents. Never im- 9 ORA_10_18_20_32_80BA-e-1510

merse the device in water or hold it under running water. Never handle the device with wet or damp hands. Never touch the measuring prism [2] with hard tools made from plastic, wood, rubber, metal, glass etc. Hard objects can quickly damage the relatively soft prism glass, resulting in measurement errors. The refractometer is maintenance-free. Cleaning should be carried out immediately before and after each use of the refractometer to maximise its life and optimise measurement results. 9 Storage Store the refractometer in a dry, non-corrosive environment, preferably between 10 C and 30 C. 10 Service After reading this operating manual, if you have any questions about setting up or using the refractometer, or if any unexpected problem occurs, please contact your dealer. The device housing may only be opened by trained service technicians authorised by KERN. 11 Disposal The packaging consists of environmentally friendly materials which can be disposed of via local recycling facilities. The device and storage box should be disposed of by the operator in accordance with applicable national or regional regulations at the place of use. 12 Additional information The product may differ slightly from the illustrations. Avoid exposing the refractometer to direct sunlight. Never bring the refractometer into contact with solvents. ORA_10_18_20_32_80BB-BA-e-1510 10

13 Brix to refractive index (nd) conversion table BRIX Refractive index BRIX Refractive index BRIX Refractive index % nd % nd % nd 0 1.33299 30 1.38115 60 1.44193 1 1.33442 31 1.38296 61 1.44420 2 1.33586 32 1.38478 62 1.44650 3 1.33732 33 1.38661 63 1.44881 4 1.33879 34 1.38846 64 1.45113 5 1.34026 35 1.39032 65 1.45348 6 1.34175 36 1.39220 66 1.45584 7 1.34325 37 1.39409 67 1.45822 8 1.34476 38 1.39600 68 1.46061 9 1.34629 39 1.39792 69 1.46303 10 1.34782 40 1.39986 70 1.46546 11 1.34937 41 1.40181 71 1.46792 12 1.35093 42 1.40378 72 1.47037 13 1.35250 43 1.40576 73 1.47285 14 1.35408 44 1.40776 74 1.47535 15 1.35568 45 1.40978 75 1.47787 16 1.35729 46 1.41181 76 1.48040 17 1.35891 47 1.41385 77 1.48295 18 1.36054 48 1.41592 78 1.48552 19 1.36218 49 1.41799 79 1.4881 20 1.36384 50 1.42009 80 1.49071 21 1.36551 51 1.42220 81 1.49333 22 1.36720 52 1.42432 82 1.49597 23 1.36889 53 1.42647 83 1.49862 24 1.37060 54 1.42862 84 1.50129 25 1.37233 55 1.43080 85 1.50398 26 1.37406 56 1.43299 27 1.37582 57 1.43520 28 1.37758 58 1.43743 29 1.37936 59 1.43967 Data from "ICUMSA" International Commission for Uniform Methods of Sugar Analysis, at 20 C and 589 nm wavelength. 11 ORA_10_18_20_32_80BA-e-1510

14 Annex Table 1: International Temperature Correction Table for Brix (% sugar gradient) Correct the result by the following values (refractometer must be correctly calibrated at 20 C) % Brix reading Temperature C 0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0 40.0 45.0 50.0 55.0 60.0 65.0 70.0 75.0 80.0 85.0 10.0-0.53-0.56-0.59-0.62-0.65-0.67-0.69-0.71-0.72-0.73-0.74-0.75-0.75-0.75-0.75-0.75-0.74-0.73 11.0-0.49-0.52-0.54-0.57-0.59-0.61-0.63-0.64-0.65-0.66-0.67-0.68-0.68-0.68-0.68-0.67-0.67-0.66 12.0-0.44-0.47-0.49-0.51-0.53-0.55-0.56-0.57-0.58-0.59-0.60-0.60-0.61-0.61-0.60-0.60-0.60-0.59 13.0-0.40-0.41-0.43-0.45-0.47-0.48-0.50-0.51-0.52-0.52-0.53-0.53-0.53-0.53-0.53-0.53-0.52-0.52 14.0-0.34-0.36-0.38-0.39-0.40-0.42-0.43-0.44-0.44-0.45-0.45-0.46-0.46-0.46-0.46-0.45-0.45-0.44 15.0-0.29-0.31-0.32-0.33-0.34-0.35-0.36-0.37-0.37-0.38-0.38-0.38-0.38-0.38-0.38-0.38-0.37-0.37 16.0-0.24-0.25-0.26-0.27-0.28-0.28-0.29-0.30-0.30-0.30-0.31-0.31-0.31-0.31-0.31-0.30-0.30-0.30 17.0-0.18-0.19-0.20-0.20-0.21-0.21-0.22-0.22-0.23-0.23-0.23-0.23-0.23-0.23-0.23-0.23-0.23-0.22 18.0-0.12-0.13-0.13-0.14-0.14-0.14-0.15-0.15-0.15-0.15-0.15-0.15-0.15-0.15-0.15-0.15-0.15-0.15 19.0-0.06-0.06-0.07-0.07-0.07-0.07-0.07-0.08-0.08-0.08-0.08-0.08-0.08-0.08-0.08-0.08-0.08-0.07 20.0 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 0.00 21.0 0.06 0.07 0.07 0.07 0.07 0.07 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.08 0.07 22.0 0.13 0.14 0.14 0.14 0.15 0.15 0.15 0.15 0.16 0.16 0.16 0.16 0.16 0.16 0.15 0.15 0.15 0.15 23.0 0.20 0.21 0.21 0.22 0.22 0.23 0.23 0.23 0.23 0.24 0.24 0.24 0.24 0.23 0.23 0.23 0.23 0.22 24.0 0.27 0.28 0.29 0.29 0.30 0.30 0.31 0.31 0.31 0.32 0.32 0.32 0.32 0.31 0.31 0.31 0.30 0.30 25.0 0.34 0.35 0.36 0.37 0.38 0.38 0.39 0.39 0.40 0.40 0.40 0.40 0.40 0.39 0.39 0.39 0.38 0.37 26.0 0.42 0.43 0.44 0.45 0.46 0.46 0.47 0.47 0.48 0.48 0.48 0.48 0.48 0.47 0.47 0.46 0.46 0.46 27.0 0.50 0.51 0.52 0.53 0.54 0.55 0.55 0.56 0.56 0.56 0.56 0.56 0.56 0.55 0.55 0.54 0.53 0.52 28.0 0.58 0.59 0.60 0.61 0.62 0.63 0.64 0.64 0.64 0.65 0.65 0.64 0.64 0.64 0.63 0.62 0.61 0.60 29.0 0.66 0.67 0.68 0.69 0.70 0.71 0.72 0.73 0.73 0.73 0.73 0.73 0.72 0.72 0.71 0.70 0.69 0.68 30.0 0.74 0.75 0.77 0.78 0.79 0.80 0.81 0.81 0.81 0.82 0.81 0.81 0.81 0.80 0.79 0.78 0.77 0.75 ORA_10_18_20_32_80BB-BA-e-1510 1