Xactix XeF2 OPERATION MANUAL

Similar documents
Xactix Xenon Difluoride Etcher

Arizona State University NanoFab XACTIX ETCHER. Rev A

TABLE OF CONTENTS 2 INTRODUCTION 4 DELEGATION OF RESPONSIBILITIES 5 INSTALLATION 6

Title: Xactix XeF2 Etcher Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 03/23/2016

Usage Policies Notebook for Xenon Difluoride (XeF 2 ) Isotropic Si Etch

University of MN, Minnesota Nano Center Standard Operating Procedure

Arizona State University NanoFab PLASMATHERM 790 RIE. Version A

The MRL Furnaces USED FOR THIS MANUAL COVERS

PLASMA ENHANCED CHEMICAL VAPOR DEPOSITION (PECVD) SOP OXFORD PLASMALAB SYSTEM 100

Arizona State University Center for Solid State Electronic Research. Table of Contents. Issue: C Title: Oxford Plasmalab 80plus (Floey) Page 1 of 8

University of MN, Minnesota Nano Center Standard Operating Procedure

OXFORD PLASMALAB 80PLUS (CLOEY)

Savannah S100 ALD at SCIF, UC Merced Standard operating Procedure

Unaxis PECVD. SiH4 (5% in He)

JETFIRST 150 RTA SYSTEM OPERATING MANUAL Version: 2 Feb 2012

STS ICP-RIE. Scott Munro (2-4826,

SSI Solaris 150 RTA Revision /27/2016 Page 1 of 9. SSI Solaris 150 RTA

Edge Isolation Tool. Standard Operating Procedure. Version 1.1. Date: Prepared by, Sandeep S S. Department of Electrical Engineering

Standard Operating Manual

STS PECVD Instructions

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

University of Minnesota, MN Nano Center Standard Operating Procedure

Usage Policies Notebook for Trion RIE / ICP Dry Etch

March CS-1701F Reactive Ion Etcher

Cambridge NanoTech: Savannah S100. Table of Contents

Usage Policies Notebook for STS DRIE System

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

Arizona State University Center for Solid State Electronics Research Issue: E Title: Heat Pulse 610 Operating Procedure Page 1 of 7

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

STS DRIE User SOP Location: NRF Cleanroon, Dry Etch Bay

Standard Operating Manual

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

Notes-PECVD: Chamber 1

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

Standard Operating Manual

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

LEO SEM SOP Page 1 of 9 Revision 1.4 LEO 440 SEM SOP. Leica Leo Stereoscan 440i

Operating Procedures for Metal Evaporator I

AX5000 Operational Manual

MANUAL FOR SPTS APS (DIELECTRICS ETCHER)

Operating Procedures for the. SAMCO ICP RIE System

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

KIOSK: SCORE CENTRE USER MANUAL RCGA Network Kiosk: Score Centre User Manual

Standard Operating Manual

Unaxis ICP/RIE SOP Revision 8 09/30/16 Page 1 of 5. NRF Unaxis ICP/RIE Etch SOP

HPICAL Operation & Data Logging Procedures. Click spacebar to advance through slides 1

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

RAPID THERMAL PROCESSOR (Annealsys AS-ONE 150) Lab Manual

NNCI ETCH WORKSHOP SI DRIE IN PLASMATHERM DEEP SILICON ETCHER. Usha Raghuram Stanford Nanofabrication Facility Stanford, CA May 25, 2016

Plasma Asher: March PX-500 User guide (May-30, 2017)

Unifilm Technology PVD-300 Sputter Deposition Operation Instructions

To Logon On to your tee sheet, start by opening your browser. (NOTE: Internet Explorer V. 6.0 or greater is required.)

Roster Maintenance. How-to Access Your Club Roster

NORDSON MARCH PX-1000 PLASMA ASHER STANDARD OPERATING PROCEDURE Version: 1.0 July 2016

Access will be via the same Player Registration tab via the Player Registrations Officer role section.

Photolithography. Operating Instructions

MP15 Jockey Pump Controller

GA-300 Gas Analyzer. Technical Note. Overview. Front Panel. iworx Systems, Inc. GA-300

University of MN, Minnesota Nano Center Standard Operating Procedure

Nanofabrication Facility: ECR Etcher SOP Rev. 01b, March 06. Standard Operating Procedure for PlasmaQuest ECR II Etching

Figure SM1: Front panel of the multipatcher software graphic user interface (GUI) at the beginning of multipatcher operation.

Standard Operating Manual

GN21 Frequently Asked Questions For Golfers

Standard Operating Manual

University of Minnesota Nano Center Standard Operating Procedure

Touch Screen Guide. OG-1500 and OG Part # T011

Basic ICP Operating Procedures

Quickstart Installation Checklist (Please refer to operation manual for complete installation instructions)

COMELEC C-30-S Parylene Coating System

1)! DO NOT PROCEED BEYOND THIS MARK

Standard Operating Procedure. For. PVD E-Beam

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

GN21 Frequently Asked Questions For Golfers

March Asher Operation

Warnings: Notes: Revised: October 5, 2015

Standard Operating Manual

The HumiPyc ( Model 2) - Gas Pycnometer; Density, Moisture, Permeation Analyzer; Filter Integrity Tester; RH sensor Calibrator

Karl Suss MA6 Mask Aligner SOP

FireHawk M7 Interface Module Software Instructions OPERATION AND INSTRUCTIONS

Thermcraft Tube Furnace General Use

Procedures for operation of the TA Instruments DSC

Vacuum Pumpdown and Venting Procedure, CRaTER Thermal Vacuum System. Dwg. No

σ = force / surface area force act upon In the image above, the surface area would be (Face height) * (Face width).

Plasma 1 (Technics) Standard Operating Procedure. Revision: 1.0 Last Updated: Feb.5/2013, Revised by Grace Li

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Using the Akta Prime plus October 22, 2012

Online League Management lta.tournamentsoftware.com. User Manual. Further support is available online at

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

ACV-10 Automatic Control Valve

Plasma II (AXIC) Standard Operating Procedure. Revision: 1.0 Last Updated: Feb.6/2013, Revised by Grace Li

UNIVERSITY OF ROCHESTER MEDICAL CENTER

FAQ RCGA Network. The RCGA is dedicated to offering you the best customer support possible. Our goal is to respond to your requests within 24hrs.

SCW Web Portal Instructions

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Hydro-Control V User Guide

Operation of the Perkin Elmer TGA-GC/MS

Approved by Principal Investigator Date: Approved by Super User: Date:

GOLF LIFE MANAGEMENT SYSTEM Member Manual

Experiment AMe-1: Small Animal Respiratory Exchange Ratio (RER)

Transcription:

General Information The Xactix e-1 is a xenon difluoride (XeF 2) isotropic silicon etcher. XeF 2 is a vapor phase etch, which exhibits very high selectivity of silicon to photo-resist, silicon dioxide, silicon nitride and aluminum. Safety Hazards WARNING: Proper use and safe operating of the equipment is the responsibility of the lab member. All lab members who work with, or exposed to this equipment must take precautions to protect themselves against possible serious injury. Do not be careless around this equipment. 1. Electrical: 120 VAC, 60 Hz voltages are present throughout this unit. Observe standard electrical safety precautions to avoid personal injury or damage to the equipment. 2. Covers: DO NOT remove any covers or touch any lead wires to avoid electrical shock. 3. Moving Parts: Keep hands, hair, tools and loose clothing away from any moving parts to avoid personal injury or damage to the equipment. Revision Jan. 2017 Page 1 of 8

Emergency Shutdown Procedure Depressing the Emergency Off (EMO) will instantly de-energize and remove all power to the system. System Features The XeF2 system etches silicon with high selectivity to most masking materials. The samples can be pieces, pixelated die, or any other silicon structures up to 6 diameter wafers. The etch process is computer controlled and the resulting etch details are captured by the control software. The etch progress can be visually monitored through the transparent chamber lid. Tool Start-Up If the system goes in to a gauge calibration routine, the total time is just over one hour. It is best to let the system complete the gauge calibration. Contact the CNM2 staff and they can adjust your Coral time to remove the gauge calibration charges. 1. Status Check: Check Badger for reservations, Shutdown, Problems or Comments, and to see if the etcher currently enabled by another lab member. Logging on the System 1. Each user is required to log in to the system using their XeF2 Username and ID. 2. The KB button beside the Username and Password field can be used to access the onscreen keyboard. 3. Press the LOGIN button to access the MAIN MENU screen. Revision Jan. 2017 Page 2 of 8

NOTE: If the system goes in to a gauge calibration routine, the total time is just over one hour. It is best to let the system complete the gauge calibration. Contact the CNM2 staff and they can adjust your Badger time to remove the gauge calibration charges. Loading a Sample 1. Press Load/Unload sample on the Main Menu screen. NOTE: If the chamber ventilation shroud covering the chamber is not fully forward, a request to close the shroud will display before venting the chamber. Move the shroud forward and acknowledge the prompt. 2. When the chamber is vented, the dialog box below will appear and you can open the chamber lid. Do not select the Examine or Done icon at this time. The lid will rest open on the stop behind chamber. 3. Inspect the O-ring. If necessary, wipe down the O-ring with a dry wipe. 4. Load your sample and close the lid. 4.1. Make sure the lid is fully closed and the lid is seated against the O-ring. 5. Press Done on Pop-up screen. 5.1. The system will go through a purging cycle prior to chamber pump-down. As a side note, the Examine button only pumps the chamber down, without purges, so that the system can be quickly vented to load the sample. This is very useful when examining a sample away from the system to prevent moisture from accumulating in the chamber. However, it is always necessary to press Done before etching the sample. Revision Jan. 2017 Page 3 of 8

6. During the pumping cycle, a click sound will be heard that indicates that the ventilation shroud can be moved. Running a Recipe Select ETCH MENU from the Main Menu Screen. A prompt asking for a lot number will appear. Enter an alpha-numerical lot number. Press the DONE Button. The following Etch Menu will appear. 1. Under the CURRENT RECIPE pull-down, select the required recipe. 2. Verify the # of Cycles, Etch Time and XeF2 Pressure values are correct. 3. If any values are not correct, change them by either moving the slider or typing in the correct value in the required field. 4. Verify that NORMAL is selected under the ETCH MODE pull-down. 5. Press the START ETCH Button. 6. Progress of the etch can be monitored following the Etch Time Completed and Cycles Completed bars along with the Elapsed Cycle Time and the ETA. Changing the Number of Cycles During an Etch. 1. The CHANGE CYCLES button can be used at any time during the etch to modify the number of cycles to the etch. 2. The top row of arrows will add cycles in this order, (right to left); ones, tens, hundreds, and thousands; the lower row of arrows, having the same values, will remove cycles. Revision Jan. 2017 Page 4 of 8

3. Click on the designated arrow for the desired number of cycles needed to be added to the etch in-progress. 4. Example, an extra 125 cycles have been added to the etch by clicking the right most arrow (ones) five times, the next right most arrow (tens) twice, and the 2nd from left most arrow (hundreds) once. 5. Press OK to return to the ETCH MENU screen. Unloading a Sample 1. Press Load/Unload sample on Main Menu screen. 2. Press YES if you re sure, NO if you re not. The system begins chamber purges and flushing cycles to evacuate the chamber, shown below. 3. NOTE: If the chamber ventilation shroud covering the chamber is not pulled fully forward, a request to close the shroud will be displayed before venting the chamber. Move the shroud forward and acknowledge the prompt. 4. When the chamber is vented, the dialog box below will appear and you can open the chamber lid. The lid will rest open on the stop behind chamber. Unload your sample. 1. Inspect the O-ring. If necessary, wipe down the O-ring with a dry wipe. 2. Close the lid. 3. Make sure the lid is fully closed and the lid is seated against the O-ring. Revision Jan. 2017 Page 5 of 8

4. Press Done on the Pop-up screen. 5. The system will go through a purging cycle prior to chamber pump-down. NOTE: Examine button only pumps the chamber down, without purges, so that the system can be quickly vented to load the sample. This is very useful when examining a sample away from the system to prevent moisture from accumulating in the chamber. However, it is always necessary to press Done before etching the sample. 6. During the pumping cycle, a click sound will be heard that indicates that the ventilation shroud can be moved. Viewing Detailed Etch Information The Detailed Etch Information log is viewed by clicking on the VIEW LOG button on the Main Menu. The log file is a database that is queried by beginning date and ending date. The Today : button will automatically set the dates to the current month, day, and year. The lot number, username, recipe, note s keyword(s) and/or etching mode used can further specify your search. Wildcard characters (*,?) may be used to fully specify the search criteria. Creating or Changing a Recipe 1. Creating or changing recipes is done from the ETCH MENU screen. 2. Make the appropriate recipe changes 3. Save the recipe by clicking on the save button located near the top of the perform etch screen. 4. The save new recipe screen will appear. Revision Jan. 2017 Page 6 of 8

5. Global Recipes are starting recipe templates available to all users. 6. Use the toggle switch in the middle of the screen users can save the new recipe to their Personal Directory or the accessible Global Recipes. 7. Type in the name of your new recipe and Click on the SAVE button. Revision Jan. 2017 Page 7 of 8

APPENDIX Etch Recipe Parameters Number of cycles Since the e1 Series is a pulsed xenon difluoride etching system, the duration of etching is controlled by the number of cycles. A cycle consists of the xenon difluoride sublimating to the set pressure in the expansion chamber, etching for a set amount of time and evacuation of the main chamber and expansion chamber. Etch Time When the valve between the main chamber and expansion chamber is opened the pressure equilibrates and the etching process begins. The etch time is the time between the opening of the valve between the expansion chamber and the process chamber and the opening of the valve between the process chamber and the pump. XeF 2 Pressure In order to introduce the proper amount of xenon difluoride to the main chamber a set pressure of xenon difluoride is delivered to the expansion chamber. Because xenon difluoride has a vapor pressure of ~4T at room temperature the upper limit for the XeF2 pressure is approximately 4T. Due to the slightly elevated temperature inside of the etcher cabinet, you may be able to get considerably higher XeF 2 pressures, however. N2 Pressure Nitrogen is added to a recipe to improve selectivity. The pressure obtained in the expansion chamber likewise controls the amount of nitrogen introduced into the process chamber. The above variables can be set either by moving the white slider on the scroll bar or by tapping or depressing on the arrows at the top and bottom of the scroll bar to increment each value or by direct entry through the keyboard. Additionally, a range button is included for the number of cycles, etch time and nitrogen pressure so that the user can input higher values than the default range allows. Processing Tid-Bits 1. XeF 2 etching is a chemical diffusion process. The etch rate and etch profiles can be highly dependent on loading effects. For example, etch rate differences can be observed between the center and the edge for wafers with high open areas. 2. Exposed silicon at the edge of the wafer can lead to more etch rate non-uniformity. This can be from photoresist edge bead removal or simply form loss of photoresist on the vertical edge of the wafer. 3. Patterns with differences in feature sizes can show different etch rates dependant on the feature size or to the proximity of two adjacent features. 4. The surface preparation prior to etch is important in achieving smooth etch surfaces. Etch roughness occurs through a mechanism similar to micromasking in dry etch. 5. XeF 2 etches native silicon dioxide, but at a different rate than silicon. Any surface contamination or variation in removal of the native oxide can lead to surface roughness. 6. The XeF 2-Si reaction is very exothermic. A delay step can be used to cool the wafer between etch cycles to minimize any observed thermal issues. Revision Jan. 2017 Page 8 of 8