Plasma Etching: Safety Operating Procedures

Similar documents
SPUTTER STATION STANDARD OPERATING PROCEDURE

Approved by Principal Investigator Date: Approved by Super User: Date:

5.1.3 Mechanical Hazards Drive assemblies have sufficient power to cause injury. Keep hands, fingers, clothing and tools clear of moving parts.

MMRC PVD Evaporator Instructions

Vacuum Pumpdown and Venting Procedure, CRaTER Thermal Vacuum System. Dwg. No

1.1 Equipment: substrate, wafer tweezers, metal targets 1.2 Personal Protective Equipment: nitrile gloves, safety glasses 1.

March CS-1701F Reactive Ion Etcher

Lesson 6: Flow Control Valves

Warnings: Notes: Revised: October 5, 2015

Angstrom Dielectric Sputterer Operation Manual

Standard Operating Manual

Warnings: Notes: Revised: January 8,

JANIS OPERATING INSTRUCTIONS FOR SUPERCONDUCTING MAGNET CRYOSTATS

Solvent System Walkthrough

SRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006

Oerlikon Sputtering Evaporator SOP

Warning: NEVER ALTER TEST EQUIPMENT

LNG Regulator Test Kit

KARL SUSS MJB3 MASK ALIGNER STANDARD OPERATING PROCEDURE

THERMAL EVAPORATION UNIT (for Al evaporation)

Nanofabrication Facility: Sputter Evaporator SOP Rev. 04, June 2007

256 Pneumatic Pressure Indicator

WARNING: READ THE GENERAL INFORMATION MANUAL INCLUDED FOR OPERATING AND SAFETY PRECAUTIONS AND OTHER IMPORTANT INFORMATION.

Chromat Calibration Updated October 27th, 2017

Glove Box Installation Manual

PANACEA P200 STAINLESS STEEL USER MANUAL

MJB4 Mask Aligner Operating Procedure. Effective Date: 07/12/2012 Author(s): Jiong Hua Phone:

Cryo-Evaporator Operation

Standard Operating Procedure Inert Vacuum Line. Ryan Mulvenna, May 2012

Pump out insert (condenser and still lines) overnight with turbo pump

Revised: June 7, 2017

Introductory Lab: Vacuum Methods

Trion PECVD SOP IMPORTANT: NO PLASTIC, TAPE, RESISTS, OR THERMAL PASTE ARE ALLOWED IN THE CHAMBER

WELLS JOHNSON HIGH VOLUME CANISTERS

Installation Procedure for your Disco Double Kit, by Airgun Lab, LLC.

Disassembly and Assembly of FDAR High-pressure Manual Pump (v. 1.0) IMPORTANT What s included:

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

MANUAL BE SERIES Test Benches

OPERATION MANUAL NTF-60 Plus

The Discussion of this exercise covers the following points:

R I T. Title: STS ASE Semiconductor & Microsystems Fabrication Laboratory Revision: Original Rev Date: 01/21/ SCOPE 2 REFERENCE DOCUMENTS

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL!

REPRODUCIBILITY IN OPTICAL THIN FILM PROCESSING PART 1, THE VACUUM AND PUMPING

Standard Operating Manual

5 Gallon Pressure Pot with HVLP Spray Gun and Hose

Usage Policies Notebook for AMST Molecular Vapor Deposition System MVD 100

Nordiko Metal Sputtering System Standard Operating Procedure

Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions

Standard Operating Manual

Memorandum Background: Results and Discussion:

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual

User Instruction Manual

Columbia Taping Tools. Hydra Reach Handle. Manual

OPERATION MANUAL NTF-15

OPERATION MANUAL Please read this Operation Manual carefully before use, and file for future reference.

Operation & Service Manual

How to use Solvent Purification System A brief manual

Nanofabrication Facility: PECVD SOP Rev. 00, April 24

Compressed Gas Properties. Introduction. Compressed Gas Properties. Definition of a Compressed Gas

This educational seminar discusses creating, measuring, and troubleshooting Rough Vacuum.

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

Angstrom E-Beam Instructions. PROCESS CHECKS: the tooling factors of each metal; Ti/Au layer for wire bonding pull test.

PANACEA P100 LOW PRESSURE USER MANUAL

AIR DRIVE ROLL FEED CARTON STAPLER MODEL: H-1031 MANUAL

LEO 1525 FEG SEM Standard Operating Procedures Mar. 6, 2012 For additional assistance, please contact the facility manager.

Introduction. Part one: Identify the Hydraulic Trainer Components

Mass Spec will not pump down

Usage Policies Notebook for NanoFurnace Furnace (EasyTube 3000 System)

Hot Compressed Air Lance Operation Manual. LAB Manufacturing 9483 Reading Road Cincinnati, OH (800)

PANACEA P125 USER MANUAL

Exercise 5-2. Bubblers EXERCISE OBJECTIVE DISCUSSION OUTLINE. Bubblers DISCUSSION. Learn to measure the level in a vessel using a bubbler.

University of MN, Minnesota Nano Center Standard Operating Procedure

Table of Contents Want full page pictures with most important parts labeled in the beginning.

MEDICAL REGULATOR RANGE (MEDIREG RANGE)

MoveRoll Conveyor Operating and Maintenance Manual

Issue: H Title: CHA E-Beam Evaporator Page 1 of 7. Table of Contents

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Plasma-Therm PECVD. Operating Characteristics. Operating Instructions. Typical Processes. I. Loading. II. Operating

QUARTZ HOUSING COATING PROCEDURE

Safe Work Instructions For Oxygen Acetylene Equipment

TRADE OF HEAVY VEHICLE MECHANIC

OPERATING INSTRUCTIONS FOR THE MODEL 85B

S & M Water Consultants

Standard Operating Procedure. For. PVD E-Beam

Dilution Refrigerator manual

Rejuvenation Instructions

MODEL NUMBER: AS18-2 USER GUIDE

1)! DO NOT PROCEED BEYOND THIS MARK

The Cyclone Tire Deflation & Inflation Tool

Operation & Maintenance Manual Place this manual with valve or person responsible for maintenance of the valve

Section 3- Repair and Calibration Procedures

COMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS

Plasma Asher: March PX-500 User guide (May-30, 2017)

TWIST-IN STEPPER SF-S0636 USER MANUAL

Karl Suss MJB4 Mask Aligner

Standard Operating Manual

Model: 720-UL INSTALLATION OPERATION MAINTENANCE. Bermad Pressure Reducing Valve IOM. Model: FP -720-UL Sizes: 2"-12" BERMAD. Application Engineering

CRYOGENIC TANK SERVICES - CHART RECORDER MANUAL

INSTALLATION INSTRUCTIONS

Remote Plasma Cleaning from a TEM Sample Holder with an Evactron De-Contaminator ABSTRACT

Transcription:

Plasma Etching: Safety Operating Procedures Amber Cai and Caleb Eades Profs. Donnelly and Van Ryswyk Harvey Mudd College, Claremont, CA

Contents 1 Introduction 1 1.1 Goal of Etching.................................. 1 2 Setup and Terminology 2 3 Operating Procedures 5 3.1 Procedure: Pumping Down............................ 5 3.2 Procedure: First-Time Oxygen Settings..................... 6 3.3 Procedure: Using Previous Oxygen Settings.................. 7 3.4 Procedure: Venting................................ 7 4 Safety Hazards 8 5 Contacts 9 i

1 Introduction The purpose of this document is to provide a comprehensive guide to properly and safely plasma etch polystyrene samples with an oxygen gas in vacuum. It will be structured in a hopefully reasonable format. First, we will discuss the setup and terminology in Section 2 such that a reader can become familiar with the equipment and feel more comfortable around it. This discussion will also make the procedures more understandable, so we encourage looking at this section carefully. We will follow this discussion with detailed procedures in Section 3 for effectively and safely completing a plasma etch. Finally, we will briefly recap the important safety hazards to be aware of in Section 4. If at any time you feel uncomfortable with the setup or any of the procedures, do not operate the equipment until you have sorted it out. Do not hesitate to contact a knowledgeable individual for help in understanding any part of this document, or for assistance in running the etching itself. See Section 5 for details on people available for contact. 1.1 Goal of Etching Before jumping into the first section of this SOP, it may be useful to a reader to know why we are using a plasma etching setup at all. As of November 2015, the Donnelly Research Group is attempting to develop a method of delivering individual polystyrene spheres to the focus of a high-intensity laser. The current method involves depositing these spheres onto a silicon substrate in a deposition process. Unfortunately, through this process, the spheres wind up bonding to each other. This is not acceptable because they appear to remain bonded after being ejected from the substrate via laser ablation. That would cause the spheres to enter the focus of a high-intensity laser in clusters, not as isolated individuals. The details of our experiment are not important to understand, however, to be able to carry out a plasma etch. The key point is that we want the spheres to not be bonded on the substrate after our depositions. So, with this in mind, the purpose of the plasma etching system is to break the bonds between polystyrene spheres in order to create isolated spherical targets. When a sample (a silicon substrate with polystyrene spheres deposited on it) is in 1

the presence of an oxygen plasma, the polystyrene will be etched out, which will dissolve the bonds between spheres. It will also reduce the sphere sizes because the plasma will etch into every area of contact with polystyrene at equal rates, presumably. This is another important factor to consider when using plasma etching. 2 Setup and Terminology In this section, we will describe the setup for plasma etching and define some of the terms that will be used frequently in the procedures section. We will first begin with the overall setup of the vacuum pumps interactions with the etching chamber. Figure 2.1: View of the vacuum pump system and etching chamber setup, with pertinent labels. From this figure, there are several things to note. First is that the labeled entities will 2

be referenced throughout the procedural sections of this document. They are important to be familiar with. In the picture itself, the plasma etching chamber is the brown-faced box with a gray cover and a leather handle on top to the right of the image. Connected to this chamber is a hose connecting it to the vacuum system, which is contained in the cart that occupies the remainder of the image. Notice as well in Figure 2.1 that the black chamber cap has a copper tube connected to it and directed out of the frame to the right of the image. This tube connects to the oxygen canister that is used to produce the oxygen plasma in vacuum. Figure 2.2: View of the canister containing the oxygen gas used for etching, with pertinent labels. This figure shows the various components of the regulator on the oxygen gas canister. Note that the lowermost black knob visible in the picture (just below the Silver Needle Knob) is another way to control the oxygen flow through the tubing. This may be important when 3

fiddling with the knobs and rods to get the flow rate and pressure to the desired values, but mostly you will just have that open and adjust with the other components. The final figure, Figure 2.3, below, shows the switchboard for turning on the Rough Pump and the Molecular Drag Pump, as well as their respective gauges, the Thermocouple Gauge and the Cold Cathode Gauge. The LED switches light up when in the On position, as shown in the image. Figure 2.3: Photo of the switches which control the pumps and gauges. 4

3 Operating Procedures The procedures in this section will detail steps to safely complete all plasma etching in our current setup with our current equipment. 3.1 Procedure: Pumping Down Follow this procedure for starting a pumpdown cycle for an etching trial. (1) Before pumping down, ensure the following: (i) Have the Black Oxygen Knob completely tightened such that oxygen is not flowing from the canister. (ii) Have the Green Vacuum Knob completely tightened as well such that air is not getting into the vacuum system. NOTE: This means to have the knob finger-tight. Too much tightening could cause damage. (iii) Put in the sample on a clean microscope slide and center the slide with sample in the plasma etching chamber so that the sample is between the two coil rings. (2) Press the Black Chamber Cap on the etching chamber opening, flip the Rough Pump and Thermocouple Gauge switches. You can cease pressing the Black Chamber Cap with your hand when the pressure is sufficient to hold it in place. (3) Once the Thermocouple Gauge reads about 10 microns, flip the Molecular Drag Pump switch on. Then flick the Start Pin to the left and wait until the green LED is visible. NOTE: The light takes a while to turn on after flicking the Start Pin. (4) Flick the Cold Cathode Gauge switch. (5) Once the Cold Cathode Gauge reads about 2 10 4 Torr, oxygen can be introduced. Continue to the secondary procedures as appropriate. 5

NOTE: In steps 3-4, both the Thermocouple Gauge and the Cold Cathode Gauge could potentially increase slightly, but this is fine. These pressures will decrease again once the Molecular Drag Pump is warmed up. 3.2 Procedure: First-Time Oxygen Settings Follow this procedure for introducing oxygen to the system if previous knob settings have not been established or if the Silver Needle Knob had to be tightened for the Rough Pump. (1) Make sure the main canister valve is open (the Main Canister Valve should read a large PSI number; in November 2015, it reads about 2500 PSI when opened). Also have the Black Oxygen Knob and Silver Needle Knob completely tightened. (2) Make sure the Brown Oxygen Rod is untightened and feels loose (the Secondary Canister Gauge should read zero and the needle should not be raised at all). Slowly tighten the rod until the needle just pops above zero. (3) Open the Black Oxygen Knob slowly while observing the Cold Cathode Gauge to make sure the pressure does not increase too drastically. It should open all the way and still have the pressure under 2 mtorr. (4) Open the Silver Needle Knob VERY carefully because it is extremely responsive. You do not want the pressure to jump too quickly or too largely here as it could damage the Rough Pump. The goal is to get the system to a set pressure with the oxygen flowing in at a given pressure. As of 11/20/2015, our target pressure is 5 mtorr on the Cold Cathode Gauge. (5) Once the reading on the Cold Cathode Gauge is steady at the target pressure, turn the plasma etching chamber power on. Turn the RF Level Knob on the chamber to the Hi setting and make sure there is a light purple-pink color in the chamber indicating a plasma. Simultaneously, if a plasma is observed, start a stopwatch. Otherwise, increase the pressure slightly with the oxygen controls and try to initiate the plasma again. 6

(6) Turn the RF Level Knob back to the Off setting once the desired etching time has elapsed, as determined by the stopwatch. Continue to the venting procedure. 3.3 Procedure: Using Previous Oxygen Settings Follow this procedure for introducing oxygen if the Brown Oxygen Rod and Silver Needle Knob have already been set to the proper positions for the appropriate oxygen concentration in the etching chamber. (1) Make sure the Main Canister Valve is open (the Main Canister Valve should read a large PSI number; in November 2015, it reads about 2500 PSI when opened). (2) Open the Black Oxygen Knob slowly while observing the Cold Cathode Gauge to make sure the pressure does not increase too drastically. Theoretically, if the Brown Oxygen Rod and Silver Needle Knob are set from a previous trial at a previous target, opening this Black Oxygen Knob should bring the pressure to that target. As of 11/20/2015, our target pressure is 5 mtorr on the Cold Cathode Gauge. (3) If the target pressure is not reached just from opening the Black Oxygen Knob, see steps 2-4 of the first time oxygen settings procedure and apply as appropriate to reach the target pressure. (4) Once the reading on the Cold Cathode Gauge is steady at the target pressure, turn the plasma etching chamber power on. Turn the RF Level Knob on the chamber to the Hi setting and make sure there is a light purple-pink color in the chamber indicating a plasma. Simultaneously, if a plasma is observed, start a stopwatch. (5) Turn the RF Level Knob back to the Off setting once the desired etching time has elapsed, as determined by the stopwatch. Continue to the venting procedure. 3.4 Procedure: Venting Follow this procedure to vent the system back to air. 7

(1) Tighten the Black Oxygen Knob to stop the flow of oxygen. NOTE: the Silver Needle Knob may also need to be tightened to completely stop the flow of oxygen. If this is the case, then for the next run, the first time settings procedure will have to be followed again. (2) Flick the Stop Pin to the left on the Molecular Drag Pump and wait for the pump to spin down. Then, flick the Cold Cathode Gauge and Molecular Drag Pump switches to the off position. (3) Flick the Rough Pump switch to the off position. (4) Slowly loosen the Green Vacuum Knob to allow air back into the system. (5) Once the hissing noise ceases (or the Black Chamber Cap to the etching chamber falls off), you know the system is at air. At this point, venting is complete. NOTE: When venting, one should not simply allow the Black Chamber Cap to fall. Rather, hold it lightly with your hand until there is no longer a vacuum pressure holding it on, at which point you can gently place it to the side wherever is appropriate. (6) If you are completely done with all etching trials for the day, close the Main Canister Valve on the oxygen tank. Otherwise, return to the pumping down procedure and continue with more trials. 4 Safety Hazards Most of the things to watch out for have been noted in Section 3 above, so we encourage the reader to look through these sections in some detail: they will provide the best sense of what can go wrong and at what times. During the etching process, there are really two main hazards that one must take precaution against and seek to minimize, mostly through acting carefully, thoughtfully and deliberately at every step. The two main hazards are: (i) Flooding the Rough Pump and Molecular Drag Pump when introducing oxygen to the system, and (ii) Damaging the main oxygen canister. 8

For (i), care should be taken when introducing oxygen to the system so that the system pressure does not increase too rapidly. If such a pressure change did occur, both the Drag Pump and the Molecular Gauge Pump could potentially fail, which in the worst case could induce an explosion. For (ii), the main oxygen canister is pretty hefty and well-built already, but it is important to keep in mind that it could easily become a dangerous projectile if damaged. Therefore, specific care should be taken to not damage the cap of the canister where the Main Canister Valve is located. This includes being extra careful when removing and fixing any of the components in Figure 2.2 as these could potentially affect the Main Canister Valve. 5 Contacts For more information on this document or plasma etching with our equipment in general, feel free to contact any of the following individuals. Chemistry Faculty Prof. Van Ryswyk hal vanryswyk@hmc.edu Physics Faculty Prof. Donnelly donnelly@hmc.edu Physics Undergraduates Amber Cai Caleb Eades xcai@hmc.edu keades@hmc.edu 9