Procedure for Field Calibration Check of Digital Pressure Gauges (Using a DG-700 or DG-500 as the Reference Standard)

Similar documents
Getting the Most From Your Diagnostic Equipment

CONSERVATORY DIAGNOSTIC TOOLS TO MEASURE BUILDING PERFORMANCE

Building Official's Guide to Duct Testing based on 2009 IECC

What is Pressure? Residential Pressure Measurement

DPC-30 DPC-100. Reference Manual

Getting the Most From Your Duct Blaster

Model 130M Pneumatic Controller

Pneumator. Testo Industrial Services More assurance, better service. Pressure calibrator and measuring instrument.

American Weigh Scales, Inc. BLADE Series D i g i t a l P o c k e t S c a l e BLADE-550. User Manual BLADE-400 BLADE-1000

300 Held Drive Northampton Pa

Digital Tourniquets. Product Details. For electricity VAC, 50/60 Hz

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

7250 Sys. Ruska Multi-Range Pressure Calibration System. GE Sensing & Inspection Technologies. Features. Applications

TOPIC: KNOWLEDGE: K1.01 [2.5/2.7] An atmospheric pressure of 15.0 psia is equivalent to... A psig. B psig. C psig.

BioPak 240 Revolution

Operating Manual Body fat balance

COMMON PRECAUTIONS Data Sheet - Be Sure To Read Before Handling Page 1 of 6

The Discussion of this exercise covers the following points:

Cygnus Underwater. UT Meters & Spare Parts Sales. Searchwise Aberdeen Ltd Unit 6, Broomiesburn, Ellon Ind Estate, Ellon, Aberdeenshire, UK, AB41 9RD

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

How to specify a product. Process Sensors and Mechanical Instruments

PTG100 Precision Test Gauge

AQUARIUS. Operating Instructions for the. Type 70 Electrolytic Gas Soldering / Welding Unit. Table of Contents

COSMO INSTRUMENTS CO., LTD.

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

Operation Manual. O2 Quickstick. Oxygen Analyzer

MOYNO FLUID DETECTION CONTROL SERVICE MANUAL. Section: ACCESSORIES. Page: 1 of 8 Date: May 1, 1998 GENERAL

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

MET 335W Fluid Mechanics Laboratory. Lab 1: Bourdon Tube Calibration. Nick Peak

Operating Instructions for Cobra and Lynx Duct Leakage Testers

BS12N Piezoresistive OEM Pressure Sensor

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

Gas Laws: Boyle s and Amonton s Laws Minneapolis Community and Technical College v.9.08

Agilent Flowmeter ADM1000

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

BioPak 240 Revolution

ENVIRONMENTAL CONTROL SYSTEM (ECS)

Dissolved Oxygen/Temp Waterproof Tester

Powers Controls RC 195 Multiple Input Receiver- Controller

Pneumatic service kit Model CPG-KITP

Important Notices. Mouthpiece. Used properly, POWERbreathe can be used safely by almost anyone without any harmful side effects.

DPI 610/615 Series. Druck Portable. Pressure Calibrators. GE Sensing. Features. Setting the Standard for Portable. Pressure Calibrators

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

Background information. normal force on a surface area of the surface

Handi+N2 OPERATING MANUAL & INSTRUCTIONS FOR USE. R218M06 Industrial. R218M06 Rev. E

Operation Manual. rev Pressure Gauges DM-2

Instruction Manual. Digital Thermocouple Gauge Control Units Models DTC & DTC Models DTC-06M-115 & DTC-06M-230

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

Hand-held pressure calibrator with integrated pump Model CPH6600

UsER manual for Watersens ph -REDOX

BS12 Piezoresistive OEM Pressure Sensor

MPM280 Piezoresistive OEM Pressure Sensor

Instruction Manual Dräger MSI P7 and MSI P7 plus

Pneumatic high-pressure controller Model CPC7000

P-5215 Differential Pressure Transmitter

Budget Range Operators Handbook

SAUTER FK Version /2016 GB

OPERATING AND MAINTENANCE MANUAL

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

256 Pneumatic Pressure Indicator

CSA Sample Draw Aspirator Adapter Operator s Manual

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

MPM280 Piezoresistive OEM Pressure Sensor

Calibration Procedure

User Manual. Heads-Up Display (HUD) DiveCAN. Mechanical Button Version

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

A little technical and safety...

Pressure is our Business

The Univentor 1250 Anaesthesia Unit

Series IRV1000/2000/3000

Additel 761 Automated Pressure Calibrators Selection Guide

911A Summa Canister Sampler

Additel 761 Automated Pressure Calibrators

Additel 761 Automated Pressure Calibrators

Experiment 18 Properties of Gases

Variable Temperature Operation

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Model EXAxt AV550G. Averaging Oxygen Analyzer

THE UNIVENTOR 400 ANAESTHESIA UNIT

Operating Instructions Part No

CÜBE. User Manual. American Weigh Scales. CUBE-100 (100g x 0.01g) CUBE-650 (650g x 0.1g)

Additel 761 Automated Pressure Calibrators Selection Guide

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Model PDT Dewpoint Transmitter

Digital Vacuum Regulator

Status update of Beam pipe. S. Tanaka KEK

ChlorTainer is the safest, most reliable technology available for processing and preventing a chlorine gas release.

OPERATION MANUAL RTI RHS650 RTI TECHNOLOGIES, INC East Market Street York, PA Manual P/N

Model Description

Digital Display large color coded LED display is easy to read precise measuring shows actual cuff pressure

EXTREME PROCESSOR UNIT

Gas Laws: Boyle s and Amonton s Laws MCTC Chemistry v.9.17

Pneumatic high-pressure controller Model CPC7000

Compact Manometer. Pressure measurements can easily be taken any time, anywhere. POWER LIGHT

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

Fresenius Medical Care North America Service Bulletin

Appendix: Simplified Instructions for Operating the HP 5890 GC/FID and Air Sampling Systems

Powers Controls RC 195 Single Input Receiver-Controller

Transcription:

Procedure for Field Calibration Check of Digital s (Using a or DG-500 as the Reference Standard) 1. Testing Apparatus Set up the (or DG-500) Standard and the gauge to be tested as shown in the diagrams below. Always push the syringe fully in before connecting the tubing to the digital gauges. The tubing to the digital gauges should be connected to the Reference taps. The purpose of the Magnehelic gauge is to prevent pressure changes introduced by the syringe from easily overpressurizing the digital gauges. Note: The Standard should be recalibrated by The Energy Conservatory once every 6 months and should only be used for calibration checks of other pressure gauges. In lieu of having a separate in-house standard, it is acceptable to test your gauges against another gauge that has just been returned from calibration at The Energy Conservatory. Magnehelic Standard or DG-500 to be tested Yellow tubing Clear tubing Syringe Magnehelic Standard DG-3 or DG-2 to be tested Yellow tubing Clear tubing Syringe January 2007-1 -

2. Comparing Measurements to the Digital Standard This procedure will involve comparing pressure readings from the Standard with the gauge being tested at a number of different pressures. These pressures will be created by adjusting the syringe. At each comparison point (called a Station) the difference in the pressure reading between the two gauges will be determined and compared with the maximum acceptable difference for that Station shown in Table 1. If the difference in readings is less than or equal to the maximum acceptable error shown in Table 1, the gauge being tested is still within the manufacturer s calibration specification. If the difference is greater than the maximum acceptable error for any of the Stations, the digital gauge being tested should be returned to The Energy Conservatory for recalibration. a. Procedure for Checking Calibration of (or DG-500) : 1) Turn on both the gauge to be tested and the Standard and leave them both in the PR/PR mode. 2) SLOWLY pull out on the syringe. The pressure readings should increase on both digital gauges. Adjust the syringe until on the Standard reads about +500 Pa (+/- 30 Pa). Place the syringe on the table. 3) After the pressure readings have stabilized, determine the difference in the pressure reading between the Standard and both and on the gauge being tested. (Note: If the pressure readings are dropping too quickly to make an accurate comparison between the 2 gauges, there is a leak somewhere in the pressure testing apparatus. First check the tubing connections, the syringe, and the tubing itself for leaks. If you are unable to eliminate the air leak, it may be located in the gauge being tested, or possibly the standard. Call TEC at 612-827-1117 for further help in diagnosing the problem). Table 1: Acceptable Error Max Acceptable +500 5 +50 0.5-500 5-50 0.5 4) Compare the pressure difference found with the maximum acceptable difference shown in Table 1 above for the +500 Station. (Note: If the difference is less than or equal to the maximum allowable error, the gauge is still within it s calibration specification.) 5) Slowly push in the syringe until the Standard reads approximately +50 Pa (+/- 5 Pa). After the pressure readings have stabilized, determine the pressure difference between the Standard and both channels on the gauge being tested. Now compare this difference with the maximum acceptable difference found in Table 1 for the +50 Station. 6) Remove the clear tubing from the digital gauge being tested (keep the yellow tubing connected to the Standard). Push the syringe completely in. Reconnect the clear tubing to both Input taps on the gauge being tested. January 2007-2 -

7) SLOWLY pull out on the syringe. Adjust the syringe until the Standard is at about +500 Pa (+/- 30 Pa). The pressure readings on the digital gauge being tested will display a negative sign. Place the syringe on the table. 8) After the pressure readings have stabilized, determine the difference in the pressure reading between the Standard and both channels on the gauge being tested. (Note: When determining the reading difference, ignore the negative sign on the gauge being tested.) 9) Compare the pressure difference found with the maximum acceptable difference shown in Table 1 for the -500 Station. 10) Slowly push in the syringe until the Standard reads approximately +50 Pa (+/- 5 Pa). After the pressure readings have stabilized, determine the pressure difference between the gauges. Now compare this difference with the maximum acceptable difference found in Table 1 for the -50 Station. 11) The calibration check procedure is complete. Turn off both gauges and remove the clear tubing from the digital gauge being tested. b. Procedure for Checking Calibration of DG-3 or DG-2 : 1) Set up the DG-3 or DG-2 gauge to be tested by turning the CHANNEL knob to A, push the RANGE switch up to 2000 Pa (High Range 1 Pa resolution), and turn the gauge on in the PRESSURE mode with 1 second average readings. 2) Set up the Standard by turning the gauge on and leaving it in the PR/PR mode. 3) SLOWLY pull out on the syringe. The pressure readings should increase on both digital gauges. Adjust the syringe until on the Standard reads about +500 Pa (+/- 30 Pa). Place the syringe on the table. 4) After the pressure readings have stabilized, determine the difference in the pressure reading between the Standard and the gauge being tested. (Note: If the pressure readings are dropping too quickly to make an accurate comparison between the 2 gauges, there is a leak somewhere in the pressure testing apparatus. First check the tubing connections, the syringe, and the tubing itself for leaks. If you are unable to eliminate the air leak, it may be located in the gauge being tested, or possibly the standard. Call TEC at 612-827-1117 for further help in diagnosing the problem). 5) Compare the pressure difference found with the maximum acceptable difference shown in Table 1 below for the +500 Pa Station. (Note: If the difference is less than or equal to the maximum allowable error, the gauge is still within it s calibration specification.) 6) Slowly push in the syringe until the Standard reads approximately +50 Pa (+/- 5 Pa). Set the RANGE switch on the DG-3 or DG-2 gauge being tested down to 200.0 Pa (Low Range 0.1 Pa resolution). After the pressure readings have stabilized, determine the pressure difference between the gauges. Now compare this difference with the maximum acceptable difference found in Table 1 for the +50 Pa Station. 7) Remove the clear tubing from the digital gauge being tested (keep the yellow tubing connected to the Standard). Push the syringe completely in. Reconnect the clear tubing to the Input tap on the gauge being tested. 8) Set the RANGE switch on the gauge being tested back up to 2000 Pa. January 2007-3 -

9) SLOWLY pull out on the syringe. Adjust the syringe until the Standard is at about +500 Pa (+/- 30 Pa). The pressure reading on the digital gauge being tested will display a negative sign. Place the syringe on the table. 10) After the pressure readings have stabilized, determine the difference in the pressure reading between the Standard the gauge being tested. (Note: When determining the reading difference, ignore the negative sign on the gauge being tested.) 11) Compare the pressure difference found with the maximum acceptable difference shown in Table 1 for the -500 Pa Station. 12) Slowly push in the syringe until the Standard reads approximately +50 Pa (+/- 5 Pa). Set the RANGE switch on the DG-3 or DG-2 gauge being tested down to 200.0 Pa. After the pressure readings have stabilized, determine the pressure difference between the gauges. Now compare this difference with the maximum acceptable difference found in Table 1 for the -50 Pa Station. 13) The calibration check procedure is complete. Turn off both gauges and remove the clear tubing from the digital gauge being tested. 3. General Maintenance Information for Energy Conservatory Digital s Operating temperature range: 32 F to 120 F. Storage temperatures - 4 F to 160 F (best to keep it warm during cold weather). Avoid conditions where condensation could occur, for example taking a gauge from a hot humid environment into a cool environment. Do not store gauge in the same container as your chemical smoke. The smoke can and does cause corrosion. Use alkaline or rechargeable batteries. Avoid exposing the gauge to excessive pressures, such as caused by tubing slammed in a door. Check the tubing (used to connect the digital gauge to the fan, outdoors and/or duct system) for air leaks. Inspect both ends of the tubing to make sure they are not stretched out or split. Periodically trim off a small piece from each end of the tubing. To check the remainder of the tubing for leaks, seal off one end of the tubing by doubling it over on itself near the end. Now create a vacuum in the tubing by sucking on the open end. Let the end of the tubing stick to your tongue due to the vacuum. The tubing should stick to your tongue for at least 5 seconds. Replace the tubing if it fails this test. For more information or to send your gauge to TEC, visit www.energyconservatory.com. January 2007-4 -

Calibration Check Form Digital s ( ) Company: Being Being Being Being Being Being

Calibration Check Form Digital s (DG-3 or DG-2 ) Company: Model (circle): DG-3 DG-2 Being Being Being