INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Similar documents
INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

DIGIVAC 22W Series INSTRUCTION MANUAL. Wide Range Linearized Vacuum Transmitter and Controller YOU MUST READ THIS MANUAL BEFORE USE

92831 TEL: (714) FAX:

SDS -Series 4C-D201/-D202 Supplemental Drying System User s Manual

INSTRUCTION MANUAL. DIGIVAC Model 801 Series. Digital Vacuum Gauge. 801W Ranges to 760 Torr 1 to 9999 microns.001 to 999 millibar

Operating Instructions for BAIR22-6 AIR-POWERED CRIMPING TOOL

User s Guide Temperature Sensor Converter TSC-599

6 digital caliper with case

Model ASSEMBLY and OPERATING INSTRUCTIONS

LUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS

RAM 4021 Operation Manual

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

KJL-205A. Operation Manual YOU MUST READ THIS MANUAL BEFORE USE

PM105 PRESSURE METER USER S MANUAL

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

CSA Sample Draw Aspirator Adapter Operator s Manual

Yoke Block Instruction Manual

DPC-30 DPC-100. Reference Manual

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

! Warning, refer to accompanying documents.

INSTRUCTION MANUAL. Digivac Model 276. Digital Vacuum Gauge. Digivac Models Model Ranges.001 to 760 Torr.001 to 1013 millibar

Instruction Manual LIMITED 1 YEAR WARRANTY. Hydraulic Punch Driver Read this material before using this product.

Operation Manual. O2 Quickstick. Oxygen Analyzer

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

Operation Manual. Pro CO 2 Analyzer. Carbon Dioxide Analyzer. Rev

COOPER POWER SERIES. D-73P3 bypass switches installation instructions. Switches MN008005EN. Effective April 2016 Supersedes S June 2013

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

DEH Instructions. Three-Phase, Platform-Mounted Distribution Transformers

T EK-SUB 4800C 19 mm Submersible Level Transmitter

92831 TEL: (714) FAX:

Instruction Manual Differential Pressure Transmitter

RADROVER REAR RACK INSTALLATION MANUAL

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

YOU MUST READ THIS MANUAL BEFORE USE

RK-IR Sample Draw Aspirator Adapter Operator s Manual

USER MANUAL FLOW SELECTOR PM1000 SAVE THESE INSTRUCTIONS. Federal (USA) law restricts this device to sale by or on the order of a physician.

Model FS5001B VA.1. MEMS Mass Flow Sensors ISO 9001 ISO OHSAS 18001

INSTRUCTION MANUAL. Lesker Model KJL615TC-K Lesker Model KJL615TC-H. Digital Vacuum Gauge/Controller. KJL615TC K/H Ranges

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

Pressure Relief Valve Instruction Manual

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

FLANGED TWO-PIECE BALL VALVES

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

COOPER POWER SERIES. Manual or Hostick Operable, Padlockable 150 A Tap-Changer Switch Installation Instructions. Tap-Changer Switch MN800015EN

Compact Triple Cabinet Outlet Station Model B Installation and Operating Instructions

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

Digital Vacuum Regulator

ASDX Series Silicon Pressure Sensors

D Series Air and Water Kit Part# 02550

Electropneumatic Transducer (I/P) Installation, Operation and

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

Model DPC3500 Continuous Low Range Dew Point Analyzer. Operations Manual

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

PERSONAL AIR BREATHING UNIT

J Air and Water Kit Instructions Part# 02584

Norrsken Family Booklet

2000 lb manual winch

COOPER POWER SERIES. QD8 Quik-Drive tap-changer VR or VL reversing switch stationary contact assembly kit number A48 installation instructions

Media-Isolated Mag Plus Probe Low Pressure-ISO Kit

GASGUARD VENT LINE3 Ammonia Sensor OPERATING & INSTALLATION MANUAL

ISQB1 INTRINSICALLY SAFE ELECTRO-PNEUMATIC CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS HAZARDOUS AREA CLASSIFICATION

COMPACT METAL bender 99826

Model 4140/4143 General Purpose Thermal Mass Flowmeter

Installation and Operation Manual

TOP VALVE. Pat. #5,857,486 & 5,944,050. High Temperature: max. 300 F (149 C) Back Pressure And Pressure Relief Valves. Instruction Manual

5 Channel Calibrator

RAM 4021-DPX Operation Manual

URC Voltage Sensor SEN-VOLT for use with MRX units containing sensor ports

Float Operated Level Controllers

INSTALLATION & MAINTENANCE INSTRUCTIONS

RAM Operation Manual

BS Series Basket Strainer

INTRODUCTION BILL OF MATERIALS SPECIFICATIONS & TOOLS PROCEDURES OPERATION OF SEATPOST GIANT LIMITED WARRANTY

700 SERIES BMS II BUBBLER MONITORING SYSTEM

AIR INLINE METAL SHEAR

Instruction Manual. Digital Thermocouple Gauge Control Units Models DTC & DTC Models DTC-06M-115 & DTC-06M-230

Model 106 DPI "Micro-switch" Installation and Operating Instructions

UE- 90- X X X X X X O P E R AT I N G M A N UA L 0 TO 90 HYDRAULIC UPENDER

Operation Manual. Pro CO Analyzer TM. Pro CO with High Temp Alarm Carbon Monoxide Analyzer

RGC-IR Remote Gas Calibrator for IR400

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

GRAVITY BIKE RACK ASSEMBLY & OPERATING INSTRUCTIONS

Mini-Pro CO2 Sensor User s Manual - ANALOG MODEL -

PRESSURE, VACUUM AND TEMPERATURE

Instruction Manual BA90 Precision Barometer

COOPER POWER SERIES. 200 A 15, 25, and 35 kv class portable feedthru installation instructions. Loadbreak Apparatus Connectors MN650037EN

Installation Instructions Air Compressor Part # 2780, 2781

Model 4040/4043/4045 General Purpose Thermal Mass Flowmeter

DISASSEMBLING & REASSEMBLING CARTRIDGE INSTALLING SEATPOST ON BIKE

SCOPE OF THIS GUIDE STARTING-UP. 1 Check mini CORI-FLOW functional properties

GILMONT ACCUCAL FLOWMETERS

200 A, 15 and 25 kv class loadbreak bushing insert installation instructions

10% OFF HHPROMO. Havahart Spray Away Animal Repellent Sprinkler Owner s Manual Model #5266. Caught a critter in the act?

COOPER POWER SERIES. 600 A 15, 25 and 35 kv class bushing adapter installation instructions. Deadbreak Apparatus Connectors MN650058EN

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

U S E R M A N U A L AIR FLOWMETER. MODELS: 1MFA2001 (shown) 1MFA9001 CAUTION. ISO Certified

Transcription:

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr A DIVISION OF THE FREDERICKS COMPANY 2400 PHILMONT AVE. HUNTINGDONVALLEY, PA 19006

PARTS LIST 1 3 4 2 # QTY ITEM DESCRIPTION 1 1 4AR Remote Convection Electronics 2 1 Cable (specified at ordering) 3 1 Instruction Manual 4 1 Sensor phone: (215) 947-2500 fax: (215) 947-7464 e-mail: sales@televac.com web site: www.televac.com

TABLE OF CONTENTS SECTION TITLE PAGE 01. DESCRIPTION and INSTRUMENT OPERATION.2 02. SAFETY.3 03. PRINCIPLES OF SENSOR OPERATION 4 04. SENSOR OPERATION AND INSTALLATION 5 05. RECORDER OUTPUTS..7 06. MAINTENANCE 8 07. CALIBRATION.8 08. TROUBLESHOOTING. 8 09. SPECIFICATIONS 9 10. DIMENSIONS..10 11. UNPACKING AND INSPECTION.11 12. WARRANTY INFORMATION 11 PLEASE READ THIS MANUAL THOROUGHLY BEFORE USING GAUGE. REPORT ANY PROBLEMS IMMEDIATELY. 1

CHAPTER 1 Description and Instrument Operation The MP4AR series Remote Module provides a compact stand-alone measurement solution for applications that do not require traditional panel-mounted instrument readouts. These modules can provide local readout and control or can be interfaced directly with a PLC, chart recorder, or data acquisition system. The MP4AR series of instruments use the time proven reliable and rugged 4A Sensor, which can be changed in seconds. These instruments operate from low voltage DC and are easily integrated into the largest process system with a minimal effort. Likewise, the MP4AR Remote Gauge can be used as a small stand-alone vacuum measurement solution. Features Wide range 1 x 10-3 Torr to 1 x 10-3 Torr Sensors can easily and quickly be replaced if required Compact & rugged self-contained package Fast system control with one setpoint 2

CHAPTER 2 Safety Instructions START BY READING THESE IMPORTANT SAFETY INSTRUCTIONS AND NOTES collected here for your convenience and repeated with additional information at appropriate points in these instructions These safety alert symbols in this manual or on the Product rear panel mean cautions - personal safety, property damage or danger from electrical shock. Read these instructions carefully. In these instructions the word product refers to the MP4AR and all of its approved parts and accessories. NOTE: These instructions do not and cannot provide for every contingency that may arise in connection with the installation, operation, or maintenance of this product. Should you require further assistance, please contact Televac at the address on the title page of this manual. This product has been designed and tested to offer reasonably safe service provided in it s installed, operated and serviced in strict accordance with these safety instructions. Failure to comply with these instructions may result in serious personal injury, including death, or property damage. These safety precautions must be observed during all phases of operation, installation, and service of this product. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Televac disclaims all liability for the customer s failure to comply with these requirements. The service and repair information in this manual is for the use of Qualified Service Personnel. To avoid shock, do not perform any procedures in this manual or perform any Servicing on this product unless you are qualified to do so. READ Instructions Read all safety and operating instructions before operating the product. RETAIN instructions Retain the Safety and Operating Instructions for future reference. HEED warnings Adhere to all warnings on the product and in the operating instructions. FOLLOW instructions Follow all operating and maintenance instructions. ACCESORIES Do not use accessories not recommended in this manual as they may require a technician to restore the product to its normal operation. To reduce risk of fire or electric shock, do not expose this product to rain or moisture. 3

CHAPTER 2 (cont.) Safety Instructions (cont.) ` Objects and Liquid Entry Never push objects of any kind into this product through openings as they may touch dangerous voltage points or short out parts that could result in a fire or electric shock. Be careful not to spill liquid of any kind onto the products. Do not substitute parts or modify instrument. Because of the danger on introducing additional hazards, do not install substitute parts or perform any unauthorized modifications to the product. Return the product to Televac for service and repair to ensure that safety features are maintained. Do not use this product if it has unauthorized modifications. CHAPTER 3 Principles of Sensor Operation The convection gauge measures absolute pressures by determining the heat loss from a thermocouple heated to a constant temperature by a current source. A second thermocouple compensates for ambient temperature changes and extends the vacuum measurement range above 2 Torr by measuring the convection effects. The convection sensor must be mounted vertically for accurate measurements. Sensors are available in stainless steel or nickel-plated brass and can withstand 150-psi overpressure. 4

CHAPTER 4 SENSOR OPERATION AND INSTALLATION 1.0 Connect the gauge to the vacuum system. It must be mounted vertically with the stem side down. NOTE: DO NOT TURN INSTRUMENTATION WHEN MOUNTING ON A THREADED INSTALLATION TO INSTALL, USE A WRENCH ON THE FLATTED PORTION OF THE SENSOR TUBE 2.0 Assure that there are no leaks. 3.0 Apply 24 DC volt power between pin # 3 (+) and pin # 4 (-) of the connector cable. 4.0 Configure output lines; pin # 5 (for linear 1 T or log 1000T output), pin # 8 (signal gnd) and pin # 7 (for non linear output). The table of voltage versus pressure is shown in chapter # 5. The linear signal output is 10 volts at 1000 microns and decreases linearly to zero at zero microns. The nonlinear signal output is 10 volts at atmospheric pressure and decreases non-linearly to zero volts at zero microns. If it is a log 1000T unit, the output decreases logarithmically from 6.0 volts at 1000Torr to zero volts at zero microns. 5.0 Pump down the vacuum system while observing the LED display and the signal voltage reading on the output. (see chapter # 5 for output values) 6.0 While the pressure is above 1000 Torr the display will show HI. Below 1000 Torr the signal output will correspond to values given in chapter # 5, depending upon chosen output. As the pressure drops below 1 Torr the red MICRONS LED lights and the displayed readings now drop from 990 down to zero while the linear output voltage drops correspondingly from 9.90 Volts to zero. 7.0 The trip pressure for the setpoint relay can be displayed by pressing the setpoint display button on the front panel of the instrument or by reading the Set Point voltage at pin # 9 of the cable connector. The display will blink the reading until the button is pushed again or until it times out automatically after 60 seconds. If the Set Point adjustment screw is turned to change the setting the new value displayed will be retained in memory. The setpoint adjustment range is 0 to 1 Torr for the units with linear recorder output and 0-1000 Torr for units with logarithmic recorder outputs. The relay contact connections are as follows; to the common # 6, # 1, normally open, and # 2, normally closed. 8.0 If the display shows -1 or LO then the gauge is out of adjustment. With a known vacuum of at least 1 X 10-5 Torr, adjust the Zero adjustment screw on the front panel. 9.0 With the sensor tube exposed to atmospheric pressure, the display reading should be near 760 Torr (which may vary with geographic location). Adjustment for the correct reading is done with the front panel ATM settings. Make sure that the tube is vertical, with stem down. SEE NEXT PAGE FOR GAUGE DIAGRAMS 5

CHAPTER 4 cont. SENSOR OPERATION AND INSTALLATION (cont.) PIN# PIN OUTS 1 Relay normally open contact 2 Relay normally closed contact Pin #s 3 +24 volt power 4 Power common 5 Log / Linear signal output 6 Relay common contact 7 Non-linear signal output 8 Signal return 9 Setpoint output Setpoint Adjust Potentiometer Setpoint Display Potentiometer Phillips Head screw SET POINT RELAY ON MP4AR REMOTE DIGITAL 4A Atmosphere Adjustment Potentiometer LED display Zero Adjustment Potentiometer LED lights for microns 6

CHAPTER 5 Recorder Outputs LINEAR Pressure (microns) Voltage 1.01 5.05 10.1 20.2 30.3 40.4 50.5 60.6 70.7 80.8 90.9 100 1 150 1.5 200 2 300 3 400 4 500 5 800 8 1000 10 2000 13.5 Non-linear & Logarithmic Pressure (TORR) Voltage (nonlinear) Voltage Logarithmic 1 x 10-3.01 0 5 x 10-3.06.70 2 x 10-2.24 1.3 4 x 10-2.45 1.6 6 x 10-2.64 1.78 8 x 10-2.82 1.9.1.98 2.2 1.67 2.3.5 2.99 2.7.8 3.84 2.9 2 5.45 3.3 10 7.32 4 30 7.84 4.48 50 7.93 4.7 80 8.01 4.9 120 8.08 5.08 160 8.19 5.2 180 8.24 5.26 300 8.76 5.48 500 9.57 5.7 760 10 5.88 1000 10.22 6 VOLTS VOLTS 4AR (linear) 4AR (linear) 4A (Logarithmic) 4A (Logarithmic) 12 10 12 8 10 8 6 4 6 4 2 2 0 0 200 400 600 800 1000 0 200 400 600 800 1000 PRESSURE (microns) PRESSURE (microns) VOLTS VOLTS 4A (non-linear) 4A (non-linear) VOLTS VOLTS 12 1112 1011 910 8 9 7 8 6 7 5 6 4 5 3 4 2 3 1 2 0 1 0 0.001 0.01 0.1 1 10 100 1000 0.001 0.01 0.1 1 10 100 1000 PRESSURE (TORR) PRESSURE (TORR) 8 8 6 6 4 4 2 2 0 0 0.001 0.1 10 1000 0.001 0.1 10 1000 PRESSURE (TORR) PRESSURE (TORR) 7

CHAPTER 6 Maintenance No maintenance is required beyond periodic replacement of a contaminated sensor. To replace the sensor, remove the Philips head screw in the center of the instrument, then gently rock and pull off the electronic assembly. Replace the old sensor with a new one, remount the electronics and replace the Phillips head screw. DO NOT OVER TIGHTEN THE SCREW! Adjust the ATM screw for the proper reading when the vacuum system is vented. CHAPTER 7 Calibration All instrumentation is calibrated at the factory. No further calibration at the customer s facility should be required, unless the gauge tube is replaced. In this instance the ATM adjustment should be done. No readjustment of the zero should be required unless it has been changed. For a reliable recalibration at a later date, it is necessary to have a reference standard vacuum system whose pressure is known to be accurate. For such a calibration, the instrument, sensor tube and cable should be returned to the factory for re-certification. However, if the vacuum calibration verification must be done at the customer s facility, a knowledgeable electronics technician trained in vacuum calibration should operate equipment of certifiable accuracy. CHAPTER 8 Troubleshooting 1. If display consistently displays HI, the atmosphere adjustment may be incorrect or there may be a broken filament in the sensor. If a broken filament is suspected, check for continuity between pins # 4 & # 8 and # 2 & # 6 on the sensor. 2. If the display constantly reads LO, the sensor may have become inoperative. Check for correct reassembly after bake out. NOTE: If, after doing some or all of the above, the units are still inoperative, return to the factory for repair. 8

CHAPTER 9 Specifications Measurement range Temperature 1 x 10-3 to 1 x 10+3 Torr Operating +15 to + 50 C Bake out at 200 C (with electronics removed) Calibration Medium Dry air or nitrogen Materials Stainless steel or nickel-plated brass Overpressure 150-psi Adjustment Potentiometers Zero, Atmosphere, Setpoint Power 24VDC @ 100mA (customer supplied) Set point Relay Form C contact 24 v.5 a rating Indicator lamp provided Accuracy +/- 1 micron 0 to 10 microns +/-10% of reading 10 microns to 10 Torr Response Time for a Pressure +/- 20% of reading 10 Torr to 100 Torr Change from 10-5 Torr to Atmosphere +/- 10% of reading 100 Torr to 1000 Torr < 500 milliseconds 9

CHAPTER 10 Dimensions 10

CHAPTER 11 Unpacking and Inspection Before each unit is installed or operated, a quick inspection should be performed and the following noted: a. damage to the unit (scratches, nicks, dents, cracks, etc.) b. missing: screws, switches or switch hardware c. broken barrier strips, etc. d. broken or loose components within instrument Should any of the above problems be encountered, contact the factory immediately. Any unauthorized repairs will void the warranty. CHAPTER 12 Warranty information The Televac division warrants instruments and components to be free of defects in material and workmanship for a period of one year after the date of shipment unless otherwise specified in the quotation pr product literature. No salesman, Representative or agent of the Fredericks Company, or its divisions is authorized to give any guarantee or warranty or make any representation in addition or contrary to those stated herein. other than those expressly stated herein, there are no other warranties of any kind, express or implied and specially excluded but not by way of limitation, are the implied warranties of fitness for a particular purpose and merchantability. It is understood and agreed the seller s liability whether in contract, in tort, under any warranty, in negligence or otherwise shall not exceed the return of the amount of the purchase price paid by the purchases and under no circumstances shall seller be liable for special, indirect, incidental or consequential damages. The price stated for the equipment is a consideration in limiting seller s liability. No action regardless of form, arising out of transactions of this agreement may be brought by purchase more than one year after the cause of action has accrued, seller s maximum liability shall not exceed and buyer s remedy is limited to either (1) repair or replacement. 11