GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY

Similar documents
THE development of more advanced techniques in radiotherapy,

CHAPTER 4 PRE TREATMENT PATIENT SPECIFIC QUALITY ASSURANCE OF RAPIDARC PLANS

13 QUALITY ASSURANCE OF A LINEAR ACCELERATOR 13.1 COLLIMATOR ISOCENTER, JAWS, LIGHT FIELD VS INDICATORS, COLLIMATOR ANGLE INDICATORS.

Relative Dosimetry. Photons

Commissioning and quality assurance of a commercial intensity modulated radiotherapy (IMRT) treatment planning system PrecisePLAN

Commissioning an IMRT System for MLC Delivery. Gary A. Ezzell., Ph.D. Mayo Clinic Scottsdale

Constancy checks of well-type ionization chambers with external-beam radiation units

TG-119 IMRT Commissioning Tests Instructions for Planning, Measurement, and Analysis Version 10/21/2009

Review of fundamental photon dosimetry quantities

The Royal Australian and New Zealand College of Radiologists. FRANZCR Examination Part I Radiation Oncology. Radiotherapeutic Physics.

Monitor Unit Calculations Part 1. Return to our first patient. Purpose. 62 yr old woman with Stage IIIB (T1N3M0) NSCLC rt lower lobe Dose prescription

Only 8% to go. TOTAL 1494 of 1623 ACTIVE INSTITUTIONS (92%) May-09. May-04 Nov-04. May-07. Nov-02 May-03. Nov-05. Nov-06. Nov-07 May-08.

Clinical Implementation of the TG-51 Protocol. David Followill Radiological Physics Center Houston Texas

Proton Therapy QA Tools

A beam-matching concept for medical linear accelerators

Monitor Unit Verification for Small Fields

Dosimetric Calculations. Lonny Trestrail

Investigation of Buildup Dose for Therapeutic Intensity Modulated Photon Beams in Radiation Therapy

IMRT in clinical practice at the UMC-Utrecht. Utrecht

Outline. Chapter 11 Treatment Planning Single Beams. Patient dose calculation. Patient dose calculation. Effect of the curved contour surface

ANALYSIS OF OFF-AXIS ENHANCDED DYNAMIC WEDGE DOSIMETRY USING A 2D DIODE ARRAY A CREATIVE PROJECT (3 SEMESTER HOURS) SUBMITTED TO THE GRADUATE SCHOOL

Zoubir Ouhib Lynn Cancer Institute

Field size and depth dependence of wedge factor for internal wedge of dual energy linear accelerator

INTERCOMPARISON OF IONIZATION CHAMBER CALIBRATION FACTORS IN THE IAEA/WHO NETWORK OF SSDLS

SnapShot IMRT with compensators and FFF beams

The MC commissioning of CyberKnife with MLC (Tips and Tricks)

Comparison of ionization chambers of various volumes for IMRT absolute dose verification

Commissioning of Elekta 6MV FFF Versa HD and Pinnacle

Exradin Ion Chambers. What attributes make Exradin the smart choice? EXRADIN Ion Chambers

Commissioning and periodic tests of the Esteya electronic brachytherapy system

CLINICAL IMPLEMENTATION OF RAPIDARC Treatment Planning Strategies to Improve Dose Distributions

Practical Course in Reference Dosimetry, National Physical Laboratory February 2014 MV Photon Dosimetry in the Clinic Page 1 of 11

Abstract: Introduction:

Validation of Treatment Planning Dose Calculations: Experience Working with Medical Physics Practice Guideline 5.a.

Original Article. Effective Point of Measurement in Cylindrical Ion Chamber for Megavoltage Photon Beams

Dynamic wedges dosimetry and quality control

Absorption measurements for a carbon fiber couch top and its modelling in a treatment planning system

Trial calibrations of therapy level electron beam ionisation chambers in terms of absorbed dose to water

TG-61 deals with: This part of the refresher course: Phantoms. Chambers. Practical Implementation of TG-61:

Scientific Journal of Nuclear Medicine & Radiation Therapy

Surface buildup dose dependence on photon field delivery technique for IMRT

The Effect of Immobilisation Devices on Radiotherapy Dose Distributions. Alison Gray August 2007

Badan Jadrowych Nuclear Research Institute. Report INR No. 739/XIX/D CERN LIBRARIES, GENEVA CM-P Differential Recombination Chamber

Chapter I Standard application in photon dosimetry José Guilherme Pereira Peixoto Maria da Penha Potiens

Math Review. Overview

TG-51: Experience from 150 institutions, common errors, and helpful hints

Clinical Implementation of Volumetric Modulated Arc Therapy

High Speed Direct SAD Radiosurgery Beam Scanner

Experimental verification of a Monte Carlo-based MLC simulation model for IMRT dose calculations in heterogeneous media

JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, VOLUME 3, NUMBER 4, FALL 2002

1. Question Answer cgy / MU cgy / MU 2. Question Answer

Clinical Implementation of the IPEM 2003 Code of Practice for Electron Dosimetry

AAPM s TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

TG219: IT'S USE, STRENGTHS AND WEAKNESSES

Technical protocol of APMP key comparison for measurement of air kerma for 60 Co (APMP.RI(I)-K1.1)

A dose delivery verification method for conventional and intensitymodulated radiation therapy using measured field fluence distributions

Monitor Unit Calculations Part 2. Calculation of machine setting. Collimator setting

Commissioning of a new total body irradiation protocol

TEST REPORT BEAMSCAN WITH HALCYON

A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation

Implementation of respiratory-gated VMAT on a Versa HD linear accelerator

Measurement and simulation of the flow field around a triangular lattice meteorological mast

Radiotherapy physics & Equipments

Evaluation of six TPS algorithms in computing entrance and exit doses

Proceedings of Meetings on Acoustics

IPEM Working Party: D IThwaites(Chair), A R DuSautoy, T Jordan, MRMcEwen, A Nisbet, A E Nahum and W G Pitchford

Technical Note: On the use of cylindrical ionization chambers for electron beam reference dosimetry

Geant4 Electromagnetic Physics for HEP Applications

arxiv: v1 [physics.ins-det] 16 Aug 2014

QUALIFICATION TESTS FOR A TYPE B (U) PACKAGE

Guideline for RMO Key comparison for Air kerma rate in 60 Co gamma radiation

Introduction. Materials and Methods

COMPASS. Dr. Lutz Müller. AK IMRT Bamberg April D IMRT Verification in Patient Anatomy

Wind Flow Validation Summary

Larry A. DeWerd, Ph.D., FAAPM UW ADCL & Dept. of Medical Physics University of Wisconsin

Application of pushover analysis in estimating seismic demands for large-span spatial structure

Vacuum Simulations of the KATRIN Experiment

Numerical Tools for LCLS-II Vacuum Systems

ELEVATION CORRECTION FACTORS FOR RADON MONITORS*

7.65 ±0.05 mm of lead. The maximum energy of the bremsstrahlung photons was calibrated using the threshold for the D(-y,n)H reaction at

The Gas Attenuator of FLASH

How to Design Medical Accelerator Systems for Reliability: IBA PT System

The dose distribution of medium energy electron boosts to the laryngectomy stoma

Introduction. Introduction. The CyberKnife SRS/SBRT System. Introduction. Contour Structures. CyberKnife Components

IAEA-TECDOC-1455 Implementation of the International Code of Practice on Dosimetry in Radiotherapy (TRS 398): Review of testing results

Basics of Proton Therapy. Proton Treatment Planning and Beam Optimization

Modeling of Hydraulic Hose Paths

PHYSICAL AND NUMERICAL MODELLING OF WAVE FIELD IN FRONT OF THE CONTAINER TERMINAL PEAR - PORT OF RIJEKA (ADRIATIC SEA)

Treatment plan complexity metrics for predicting IMRT pretreatment. assurance results. Scott Crowe, QUT

Development of Technology to Estimate the Flow Field around Ship Hull Considering Wave Making and Propeller Rotating Effects

SPIRIT III Radiometer Saturation Effect

SUBMERGED VENTURI FLUME. Tom Gill 1 Robert Einhellig 2 ABSTRACT

WIND-INDUCED LOADS OVER DOUBLE CANTILEVER BRIDGES UNDER CONSTRUCTION

A clinical comparison and analysis between conventional MLC based and solid compensator based IMRT treatment techniques

Agilent Dimension Software for ELSD User Manual

Folding Reticulated Shell Structure Wind Pressure Coefficient Prediction Research based on RBF Neural Network

Numerical and Experimental Investigation of the Possibility of Forming the Wake Flow of Large Ships by Using the Vortex Generators

Available online at ScienceDirect. Procedia Engineering 147 (2016 ) 74 80

CHAPTER-2. Wedge filters are the most important beam modifiers used in

Numerical Simulations of Bubbling Fluidized Beds

Transcription:

The Monte Carlo Method: Versatility Unbounded in a Dynamic Computing World Chattanooga, Tennessee, April 17-21, 2005, on CD-ROM, American Nuclear Society, LaGrange Park, IL (2005) GEANT4 SIMULATION OF AN ACCELERATOR HEAD FOR INTENSITY MODULATED RADIOTHERAPY F. Foppiano National Cancer Research Institute (IST) Largo Rosanna Benzi10, 16132 Genova - Italy franca.foppiano@istge.it. B. Mascialino, M.G. Pia and M. Piergentili Italian Institute for Nuclear Research (INFN) via Dodecaneso 33, 16146 Genova -Italy Barbara.Mascialino@ge.infn.it; MariaGrazia.Pia@ge.infn.it; Michela.Piergentili@ge.infn.it ABSTRACT We present a Geant4-based application for the simulation of the absorbed dose distribution given by a medical linac used for intensity modulated radiation therapy (IMRT). The linac geometry is accurately described in Monte Carlo code using the accelerator s manifacturer s specifications. The flexible design of this object-oriented system allows for an easy configuration of the geometry of the treatment head for various types of medical accelerators used in clinical practice. The precision of the software system relies on the application of Geant4 Low Energy Electromagnetic models, extending the treatment of electron and photon interaction down to low energies for precise dosimetry. The capability of the software to evaluate dose distribution has been verified by comparison with measurements in water phantom; the comparisons were performed for percent depth dose (PDD) and for flatness at 15, 50 and 100 mm depth for various field size, for a 6 MV electron beam. The source-surface distance (SSD) was 100 cm. We show the comparisons between simulation results and experimental measurements. The code is fully parallelized, allowing to achieve the high statistics Monte Carlo production required fastly. The parallelization scheme adopted allows to run the code trasparently either in a local computing farm or with geografically distributed resources, such as a computing GRID. This software system is released to the public as an open source software system, for the benefit of the scientific community. Key Words: IMRT, Geant4, Radiotherapy 1 INTRODUCTION The development of advanced techniques in radiotherapy, such as intensity modulation, brings about the need for more sophisticated and precise methods to calculate dose distributions. In particular, Intensity Modulated Radiation Therapy (IMRT) involves a series of small fields, to obtain a specific treatment, with a complex dose distribution, for an individual patient. The most commonly treatment planning systems are based on calculation algorithms, that do not account for electron transport; and, therefore, they do not predict the dose accurately in small fields, where lateral electronic equilibrium is not achieved. The project described in this paper developed a dosimetric system, based on a Monte Carlo method, that calculates accurately the dose distributions for IMRT.

F. Foppiano, B. Mascialino, M.G. Pia and M. Piergentili 2.1 Monte Carlo calculation 2 MATERIALS AND METHODS GEANT4 [1] is an object-oriented toolkit developed to simulate the passage of particles through matter. It contains a large variety of physics models, covering the interactions of electrons, muons, hadrons and ions with matter from 250 ev up to several PeV. For these features and for its versatility, Geant4 applications are developed not only in particle physics, but also in many other fields, where high accuracy and precision of simulations are needed, such as radiation physics, space science and radiotherapy. Specifically for our application, of particular interest is the Low Energy Package [2], developed to extend electromagnetic interaction of particles with matter down to very low energy: 250 ev for electrons and photons. We developed a Geant4 application, which simulates the linear accelerator s head used at National Cancer Research Institute in Genova for the IMRT. The components in the simulated linac head were the following: target, primary collimator, vacuum windows, flattening filter, ion chamber, mirror, jaws on the X and Y coordinates, multileaf collimator, with rounded leaf tips, and the phantom. Each pair of jaws can be rotated through an axis that is perpendicular to the beam axis. The user can select the position of each jaw and of every single leaf of the multileaf collimator. The phantom was divided in voxels, in which we collected the energy deposit to calculate the relative dose absorbed. In the case of squared fields the phantom was made of water, in the case of the intensity modulated fields the phantom was made of plexiglass. We have assumed a Gaussian distribution for energy and momentum of the primary electrons. 2.2 Experimental measurements The measurements of both depth dose curves and lateral dose profiles for the squared fields were made using an ion chamber (31002 Semiflex, PTW, Freiburg Germany) with an active volume of 0.125 cm 3 and a cavity diameter of 0.55 cm. The ion chamber was mounted on a computer controlled scanning system (MP3-System, PTW, Freiburg Germany). The uncertainty of the position accuracy of the scanning system provided by the manufacturer was ±0.1 mm. All the measurements were obtained at a source-surface distance (SSD) of 100 cm. The effective point of measurement for the ion chamber was taken to be 0.60r cav, where r cav is the cavity radius of the chamber, consistent with the International Atomic Energy Agency (IAEA) protocol [3]. The estimated total uncertainty of an ionization measurement at one depth was 0.5%, deriving from the positioning repeatability of the ion chamber and short term fluctuations of the chamber, electrometer, air pressure and temperature, during the time frame of one scan. The estimated total uncertainty of a relative dose measurement was taken to be 1.8%; this estimation takes in account the relative uncertainty on the calibration of the dosimeter, on the correction terms for influence quantities, on the beam quality correction, the long term stability of the dosimeter, on the dosimeter reading, the long term stability of user dosimeter and the establishment of reference conditions. The measurements of absorbed dose in a plane, for the fields obtained with the multileaf collimator, were made using Kodak X-Omat V films and the film dosimetry system RIT113 with the scanner VXR-16 DosimetryPRO. American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 2/10

Geant4 Simulation of an Accelerator Head for Intensity Modulated RadioTherapy 2.3 Normalization and comparison of data Data concerning the depth-dose curves were normalized to their maximum value. Lateral profiles obtained with experimental measurements were normalized to the dose value on the beam axe, while the ones obtained with the dosimetric system were normalized to the mean value in the flat zone. Moreover in the specific case of the lateral profiles, we rebinned data in the central flat region, with the aim of reducing statistical fluctuations. Statistical comparisons between experimental measurements and Geant4 simulations have been performed by means of a Goodness-of-Fit statistical toolkit [4], specialized in the comparison of data distributions. Among the ones available in the toolkit, we have selected the Kolmogorov-Smirnov [5] test. In the particular case of lateral dose profiles, we divided both experimental data and Geant4 simulation results in five regions of interest, having a relevance in the medical-physics domain: - the central flat zone, - the two high gradient regions, - the tails, and we tested the agreement between the two distributions within each of these regions of interest. For any analysis performed, the confidence level was set to the value 0.05; a p-value higher than 0.05 led to the acceptance of the null hypothesis, stating the equivalence between reference data and Geant4 simulations. 3 RESULTS In all the simulations, we reproduced with the dosimetric system the real experimental set-up, such as the position of the jaws and of every single leaf of the multileaf collimator, the energy of the beam, the source-surface distance and the material of the phantom. 3.1 Field 10x10 cm We start this study from the 10x10 cm field, that is often used as a reference field to calculate relevant dosimetric quantities. In Figure 1 we show the lateral dose profiles at 50 mm depth for the 10x10 cm field, with the 6 MV beam (SSD=100 cm). Table I shows the quantitative statistical results obtained in each region of interest: the first and the second columns indicate the region of interest under study, in the third column we reported the distance between the two distributions (Kolmogorov-Smirnov test statistics) and in the fourth column we listed the corresponding p-value. In any region considered, the Kolmogorov-Smirnov test pointed out a good agreement between experimental data and Geant4 simulations (p-values > 0.05 in any case). In Figure 2 we show the lateral dose profiles at 100 mm depth, for the 10x10 cm field, with the 6 MV beam (SSD=100 cm). Table II shows the quantitative statistical results obtained in each region of interest: the first and the second columns indicate the region of interest under study, in the third column we reported the distance between the two distributions (Kolmogorov-Smirnov test statistics) and in the fourth column we listed the corresponding p-value. In any region American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 3/10

F. Foppiano, B. Mascialino, M.G. Pia and M. Piergentili flatness campo 10x10 cm; 50 mm di profondita ; 6 MV 120 sistema dosimetrico misure sperimentali 100 80 dose percentuale 60 40 20 0-150 -100-50 0 50 100 150 distanza (mm) Figure 1: Lateral dose profiles at 50 mm depth, for the 10x10 cm field, with the 6 MV beam. The red symbols correspond to the experimental measurements, the dark symbols correspond to the Geant4 simulation. Table I: Results of the Kolmogorov-Smirnov test for the lateral dose profiles at of 50 mm depth, for the 10x10 cm field, with the 6 MV beam. region of interest range distance p-value tail -84, -60 mm 0.39 0.23 high gradient zone -59, -48 mm 0.27 0.90 central flat zone -47, 47 mm 0.43 0.19 high gradient zone 48, 59 mm 0.30 0.82 tail 60, 84 mm 0.40 0.10 considered, the Kolmogorov-Smirnov test pointed out a good agreement between experimental data and Geant4 simulations (p-values > 0.05 in any case). 3.2 Field 5x5 cm After the 10x10 cm field, we studied the 5x5 cm field, that have dimensions similar to the fields used in IMRT. In Figure 3 we show the lateral dose profiles at 15 mm depth, for the 5x5 cm field, with the 6 MV beam (SSD=100 cm). Table III shows the quantitative statistical results obtained in each region of interest: the first and the second columns indicate the region of interest under study, in the third column we reported the distance between the two distributions (Kolmogorov-Smirnov test statistics) and in American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 4/10

Geant4 Simulation of an Accelerator Head for Intensity Modulated RadioTherapy 120 sistema dosimetrico misure flatness campo 10x10 cm; profondita = 100 mm; 6 MV 100 80 dose percentuale 60 40 20 0-150 -100-50 0 50 100 150 distanza (mm) Figure 2: Lateral dose profiles at 100 mm depth, for the 10x10 cm field, with the 6 MV beam. The pink symbols correspond to the experimental measurements, the dark symbols correspond to the Geant4 simulation. Table II: Results of the Kolmogorov-Smirnov test for the lateral dose profiles at 100 mm depth, for the 10x10 cm field, with the 6 MV beam. region of interest range distance p-value tail -84, -64 mm 0.43 0.25 high gradient zone -63, -46 mm 0.33 0.60 central flat zone -45, 50 mm 0.52 0.17 high gradient zone 51, 56 mm 0.52 0.65 tail 57, 84 mm 0.38 0.40 the fourth column we listed the corresponding p-value. In any region considered, the Kolmogorov-Smirnov test pointed out a good agreement between experimental data and Geant4 simulations (p-values > 0.05 in any case). In Figure 4 we show the lateral dose profiles at 100 mm depth, for the 5x5 cm field, with the 6 MV beam (SSD=100 cm). Table IV shows the quantitative statistical results obtained in each region of interest: the first and the second columns indicate the region of interest under study, in the third column we reported the distance between the two distributions (Kolmogorov-Smirnov test statistics) and in the fourth column we listed the corresponding p-value. In any region considered, the Kolmogorov-Smirnov test pointed out a good agreement between experimental data and Geant4 simulations (p-values > 0.05 in any case). American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 5/10

F. Foppiano, B. Mascialino, M.G. Pia and M. Piergentili 100 sistema dosimetrico dati sperimentali flatness campo 5x5 cm, 6 MV, profondita =15 mm 80 dose percentuale 60 40 20 0-150 -100-50 0 50 100 150 distanza (mm) Figure 3: Lateral dose profiles at 15 mm depth, for the 5x5 cm field, with the 6 MV beam. The violet symbols correspond to the experimental measurements, the dark symbols correspond to the Geant4 simulation. Table III: Results of the Kolmogorov-Smirnov test for the lateral dose profiles at 15 mm depth, for the 5x5 cm field, with the 6 MV beam. region of interest range distance p-value tail -56, -35 mm 0.26 0.89 high gradient zone -34, -22 mm 0.43 0.42 central flat zone -21, 21 mm 0.38 0.08 high gradient zone 22, 32 mm 0.26 0.98 tail 33, 56 mm 0.57 0.13 Table IV: Results of the Kolmogorov-Smirnov test for the lateral dose profiles at 100 mm depth, for the 5x5 cm field, with the 6 MV beam. region of interest range distance p-value tail -61, -35 mm 0.33 0.5 high gradient zone -34, -26 mm 0.40 0.75 central flat zone -25, 21 mm 0.32 0.24 high gradient zone 22, 32 mm 0.27 0.75 tail 33, 61 mm 0.21 0.91 American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 6/10

Geant4 Simulation of an Accelerator Head for Intensity Modulated RadioTherapy 3.3 Field 40x40 cm We have started to study the 40x40 cm field, that can help us to determine if we have used the correct parameters characterizing the initial electron beam. In Figure 5 we show the depth dose curves at 100 mm depth, for the 40x40 cm field, with the 6 MV beam (SSD=100 cm). Table V contains the results of the Kolmogorov-Smirnov test, the test demonstrate that the dosimetric system can simulate with great accuracy the experimental measurements. Table V: Results of the Kolmogorov-Smirnov test for the depth-dose curves, for the 40x40 cm field, with the 6 MV beam. range distance p-value 0, 300 mm 0.005 1 3.4 Field obtained with intensity modulated beams We studied an intensity modulated field used to treat a patient with prostate cancer. This field was obtained superimposing two fields shaped with the multileaf collimator, the first field is shown in figure 6; this figure represents the dose distribution in a plane, obtained with the simulation system; the sizes of the symbols in the picture are proportional to the absorbed dose. Figure 7 shows the whole intensity modulated field; this figure illustrates the dose flatness campo 5x5 cm, 6 MV, 100 mm di profondita 120 100 dose percentuale 80 60 40 20 0-150 -100-50 0 50 100 150 distanza (mm) Figure 4: Lateral dose profiles at 100 mm depth, for the 5x5 cm field, with the 6 MV beam. The azure symbols correspond to the experimental measurements, the dark symbols correspond to the Geant4 simulation. American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 7/10

F. Foppiano, B. Mascialino, M.G. Pia and M. Piergentili PDD campo 40x40 cm; 6 MV 100 sistema dosimetrico dati sperimentali 80 dose percentuale 60 40 20 0-300 -200-100 0 100 200 300 distanza (mm) Figure 5: Depth dose distribution, for the 40x40 cm field, with the 6 MV beam. The green symbols correspond to the experimental measurements, the dark symbols correspond to the Geant4 simulation. Figure 6: Dose distribution in a plane given by the modulated beam, this distribution was obtained with the simulation system; the dimensions of the symbols in the picture are proportional to the absorbed dose. distribution in a plane, obtained with the simulation system. In both the figures 7 and 6we can appreciate the regions with different dose intensity, and we can see that the simulation system is able to reproduce the dose distribution, and in particular the dose interleaf transmission due to the rounded leaf tips. American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 8/10

Geant4 Simulation of an Accelerator Head for Intensity Modulated RadioTherapy Figure 7: Dose distribution in a plane given by the modulated beam, this distribution was obtained with the simulation system; the dimensions of the symbols in the picture are proportional to the absorbed dose. 4 CONCLUSIONS The results obtained with Geant4 simulation for the fields in homogeneous phantoms show a excellent agreement with the measured dose data. All the statistical comparisons state that there is not a statistical difference between the experimental measurements and the data obtained with the Geant4 simulations. This method of dose calculation can be a dosimetry tool in addition to the traditional analytical treatment planning systems. This simulation can be used, changing some parameters in the modelization of the geometry, to study also others medical linac s heads. The linac simulation system tool is publicly available with the public release of Geant4 as an Advanced Example. 5 REFERENCES 1. Geant4 Collaboration, Geant4 - a simulation toolkit, NIM A, vol. 506, pp. 250-303 (2003). 2. S. Chauvie, G. Depaola, V. Ivanchenko, F. Longo, P. Nieminen, M.G. Pia, Geant4 Low Energy Electromagnetic Physics Proceedings of CHEP 2001, Beijing, China, 2001, pp. 337-340 (2001). 3. International Atomic Energy Agency, Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water, Technical Reports Series No. 398, IAEA, Vienna (2000). 4. G.A.P. Cirrone, S. Donadio, S. Guatelli, A. Mantero, B. Mascialino, S. Parlati M.G. Pia, A. Pfeiffer, A. Ribon, P. Viarengo A Goodness-of-Fit Statistical Toolkit Transactions on Nuclear Science, vol. 51 (5), pp. 2056-2063 (2004). American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 9/10

F. Foppiano, B. Mascialino, M.G. Pia and M. Piergentili 5. A.N. Kolmogorov, Sulla determinazione empirica di una legge di distribuzione, Giornale dell istituto italiano degli attuari, vol.4, pp. 83-91 (1933). American Nuclear Society Topical Meeting in Monte Carlo, Chattanooga, TN, 2005 10/10