OEM Manual. MODEL ½ Digit DRUM SCALE

Similar documents
OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

Manual Weighingblock VB2 series and Uniscale

DPC-30 DPC-100. Reference Manual

Preferred Instruments Danbury, CT USA

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Alphasense 4-20 ma transmitters offer convenience and easy maintenance for toxic sensors:

Metham Aviation Design Limited, Station Approach, Four Marks, Alton. Hants. GU34 5HN Tel +44 (0) Fax +44 (0)

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

222 Schwinn Recumbent Exercise Bike Parts List Full Size Hardware Chart Product Illustration Assembly Instructions

I.H.S INSTALLATION INSTRUCTIONS

StockWeigh 500 D3632 US-B

Operating Instructions Part No

MODEL 9875 STADIUM BASEBALL SCOREBOARD. Instruction Manual

REGAL ELECTRONIC CYLINDER SCALES

Budget Range Operators Handbook

Installation and Operation Manual

24L AIR COMPRESSOR OPERATION & MAINTENANCE INSTRUCTIONS MODEL NO: RANGER 7/240 PART NO: LS0913

Columbus Instruments

24L OIL FREE AIR COMPRESSOR MODEL NO: TIGER 7/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS10/13

ACV-10 Automatic Control Valve

Universal Valve Company Inc

MODEL WEIGH MODULE

200 PSI COMPRESSORS - MODEL NUMBERS

BA307C & BA308C Intrinsically safe loop-powered 3½ digit panel mounting indicators issue 9

TELEDYNE HASTINGS INSTRUMENTS MODEL 40 POWER SUPPLY INSTRUCTION MANUAL

24L AIR COMPRESSOR MODEL NO: TIGER 11/250 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

Model B Wire Wet Oxidant Gas Transmitter. 6 Iron Bridge Drive Unit 1 & 2 Gatehead Business Park

100L AIR COMPRESSOR MODEL NO: TIGER 16/1010 PART NO: OPERATION & MAINTENANCE INSTRUCTIONS LS01/13

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

VERTICAL AIR COMPRESSORS

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

100C Air Compressor Kit

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Instruction Manual. PLT Series. 1. General Specification. 2. Installation. 3. Wiring. 4. Function. 5. Setting & Calibration

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

Universal Valve Company Inc

EcoPRO-Series 4-20mA SUBMERSIBLE PRESSURE TRANSMITTER INSTRUCTIONS

Deep Submersible Level Transducer Series 300DS

Installation, Operation and Maintenance Manual for Back Pressure Regulator

Installation, Operation and Maintenance Instructions for Electronically Controlled Pressurisation Units

Model 106 DPI "Micro-switch" Installation and Operating Instructions

Product Manual. Cyl-Tel and Tank-Tel Liquid Level Gauges. Designed and Built by: Chart Inc th Street NW New Prague, MN USA (800)

Plankton pump Model ,2 KWatt. Manual. KC Denmark. Research Equipment Limnology Oceanography Hydrobiology

Service Instructions. Syntron Control Model: CSCR-1-B (115V)

BRILLIANT WONDERS LED BUBBLERS INSTALLATION INSTRUCTIONS & PRODUCT MANUAL CMP SERIES

MODEL 154XST INSTALLATION AND SERVICE INSTRUCTIONS FOR MODEL 154XST STROBE LIGHT FOR USE IN HARSH ENVIRONMENTS/ HAZARDOUS LOCATIONS

WARNING: READ THE GENERAL INFORMATION MANUAL INCLUDED FOR OPERATING AND SAFETY PRECAUTIONS AND OTHER IMPORTANT INFORMATION.

RC 195 Receiver-Controller

CP10K. 10,000 lb. Cable Puller Instruction Sheet. SAFETY ISSUES

WINCH USER GUIDE WARN 3000ACI Utility Winch Part Number: 93000

Operating Instructions Model and Hydrostatic Test Pump

Fisher Lever Lock Handlever Actuators for POSI SEAL A81 Rotary Valves

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

AUTOMATIC GAS MANIFOLDS

Misaligned Folds Paper Feed Problems Double Feeds Won t Feed FLYER Won t Run iii

VERTICAL AIR COMPRESSORS

Instruction Manual BA90 Precision Barometer

MODEL GT820 OXYGEN SENSOR

IMPORTANT SAFETY INSTRUCTIONS

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

Data sheet Pressure switch type CS December 2001 DKACT.PD.P10.A B1119

420C AIR COMPRESSOR KIT PART NO C AIR COMPRESSOR KIT PART NO

For use with select HPC Fire Pits ONLY- refer to product catalog or website. 100K BTU Maximum WARNING: FOR OUTDOOR USE ONLY

INSTALLATION & MAINTENANCE INSTRUCTIONS

444C DUAL PERFORMANCE VALUE PACK

97C COMPRESSOR KIT 12V PART NO C COMPRESSOR KIT 24V PART NO C COMPRESSOR KIT PART NO

Operating Instructions Part No

250C-IG COMPRESSOR KIT 12V PART NO C-IG COMPRESSOR KIT 24V PART NO

Rigel 601 CHECKBOX. Instruction Manual. 348A551 Issue 2.0. April Seaward Electronic Ltd. Issue 2.0

WINCH USER GUIDE WARN 3700 Utility Winch Part Number: 93700

Instruction and Practice Manual

ECONORESS ELECTRONIC EPS & EPT - ENHANCED PRESSURISATION SET INSTALLATION OPERATION & MAINTENANCE DOCUMENTATION

RAI-M series RUDDER ANGLE INDICATOR SYSTEM

P-5215 Differential Pressure Transmitter

Fisher Lever-Lock Handlever Actuators for POSI-SEAL A81 Rotary Valves

210 Series Transmitter with External Electrochemical Sensor

PRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.

Fiber Cable Puller with Tuf-Lugger lite

INSTRUCTION MANUAL TIPPING BUCKET RAINGAUGE MODEL TB305-1MM

Teledyne Oil & Gas North Williamson Blvd. Daytona Beach, FL USA (phone) (fax)

User Manual GRI- 1500Li

GAS FUEL VALVE FORM AGV5 OM 8-03

Troubleshooting Guide: 640 Pediatric Exam Table with Midmark Scale

APPS-160 Gen 2 User Manual

NOVALYNX CORPORATION MODEL BAROMETRIC PRESSURE SENSOR PCB MODEL BAROMETRIC PRESSURE SENSOR IN NEMA ENCLOSURE INSTRUCTION MANUAL

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

SILENTAIRE TECHNOLOGY

INSTALLATION. and INSTRUCTION MANUAL. for QUALITY AIR BREATHING SYSTEMS. Model 50 Systems Outfitted with ABM-725 Monitor C O M P A N Y

Instructions for Use

KING-GAGE. Calibration Instruction Manual. D/P LiquiSeal Purge Control and D/P Purge Control

Precision Liquid Settlement Array Manual

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

CÜBE. User Manual. American Weigh Scales. CUBE-100 (100g x 0.01g) CUBE-650 (650g x 0.1g)

5 Channel Calibrator

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

3.0 Pressure Transmitter Selection

LRS(H)4 USER MANUAL. Read the complete manual before installing and using the regulator.

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

Transcription:

OEM Manual MODEL 4020-3 ½ Digit DRUM SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA & Canada) Phone: (+1) 215-766-2670 (international) Fax: (+1) 215-766-2672 E-mail: info@scaletronscales.com Website: www.scaletronscales.com

These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right to make engineering refinements that have not been described herein. Should any questions arise that may not be answered specifically by these instructions, they should be directed to SCALETRON INDUSTRIES, LTD., or our sales agent for a response. All possible precautions were taken in packaging each equipment item to prevent shipping damage. Carefully inspect each item and report damages immediately. Report damage claims to shipping agent involved for equipment shipped F.O.B. job site. Do not install any damaged equipment. All instructions given on any labels, or attached tags, should be followed. Carefully inspect all packing material before discarding it to prevent loss of accessories, mounting hardware, spare parts, or instructions. Table of Contents Page I. GENERAL DESCRIPTION 1 II. SPECIFICATIONS...... 1 III. ASSEMBLY AND START-UP.....2 IV. GENERAL USE INSTRUCTIONS 2-3 V. CALIBRATION PROCEDURE.3 VI. TROUBLESHOOTING.3 4 VII. BOARD DIAGRAM AND 4-20 CONNECTION DIAGRAM 5-6 I. General Description: The Model 4020 Digital Drum Scale is available with 12 or 16 discs and was designed to provide the accurate weight of cylinders the same diameter and smaller. The maximum net weight is 199.9 lbs./kg. for 0.1 lbs./kg. resolution. If maximum net weight is over 199.9, the decimal is blanked at the factory, and the display can read 1999 lbs./kg., but the maximum weight rating on the base is 600 lbs. and should not be exceeded. Maximum tare weight is 150 lbs. The low profile design of the platform, 1 ½, allows for safe and easy loading and unloading of cylinders. The printed circuit board contains a power supply section and a separate tare and span control. The weighmeter electronics are housed in a NEMA 4X enclosure for mounting on the wall. Standard Parts Available Options 1 Base assembly A. Remote mounting of standard enclosure 1 Weighmeter B. Loadcell Cable 1 Power Cord 1 Technical Manual II. Specifications: Display: 3 ½ Digit LCD with minus sign liquid crystal display, 0.5 high, seven segment Resolution: 199.9 lbs./kg. X 0.1 lbs./kg., 600 lbs./(272 kg.) X 1 lbs./kg. Load Cell Excitation: 12 VDC Power 3 Cells Zero Adjustment: Internal potentiometer, external tare knob Overload: Blanks with 1 displayed at far left of display Accuracy: +/- 0.5% Capacity Power: 85 265 VDC, 50/60 Hz single phase, 0.42 Amps. DEDICATED - ISOLATED POWER WITH EARTH GROUNDING 1

2 III. Assembly and Start-up: The Model 4020 Drum Scale is shipped disassembled. Assembly instructions are as follows: The load cells are secured to discs and mounted in the base housing. All wire connections in the base have been made and sealed at the factory. The digital weighmeter is assembled and calibrated to the base at the factory. The customer must mount the indicator on the wall. The wires must be run through the strain relief, connected to connector and plugged into indicator before scale is operated. Steps are as follows: Step 1: Remove all parts from box and inspect for damage, broken wires, cracks in indicator, etc. Any shipping damage must be reported to carrier. Step 2: Place the scale on a solid dry surface. Clear the area of all debris where scale will sit. Step 3: Weigh platform must be bolted to floor; clearance holes are provided in each corner for 3/8 anchor bolts. Make bolts snug, but do not over tighten. Step 4: Run conduit if required. Make hole in indicator box for conduit if used. If not, make ½ hole for liquid tight strain relief provided. Feed the wire up through the strain relief and tighten the fitting. Fasten indicator to the wall. Connect wires to 6 pin connector provided (TB 1) according to the following code: TB1 : Ret: Common, 4-20mA Return* Out: + 4-20mA Output* G: Green = + Signal W: White = - Signal Bk: Black = - Supply Voltage R: Red = + Supply Voltage *Connect your wires here if you are using 4-20mA output. Refer to the wiring diagram on page 5 and 6 for 4-20mA connections. Jumper J-4 will need jumper in the top two pins for loop powered, and the bottom two pins for scale powered (with board oriented as in drawing pg 5.) Close the weighmeter and apply power as required: Standard Unit requires a 120 Volt power source (power supply voltage range spans: 85 Volt to 265 Volt AC) Step 5: Allow approximately 15 minutes warm up time before using the scale. IV. General Use Instructions: TARE ADJUSTMENT: The tare knob, located on the front of weighmeter, allows either zeroing the scale to determine the gross weight of the drum, or presetting the known tare weight of a drum or cylinder before it is placed on the scale. With the scale platform empty, the operator can determine the tare weight (normally marked on each cylinder). When using a drum, weigh empty container and mark drum with weight. The operator would then turn the tare knob in the direction required to set a negative (-) reading on the display equivalent to tare weight of the cylinder. Example: 1. Cylinder or drum weight: 75 lbs. 2. Without cylinder or drum on scale, adjust tare knob until display reads -75.0. ** Clockwise rotation of knob increases tare. ** ** Counterclockwise rotation of knob decreases tare. ** Note: When other fittings, etc., are to be secured to the cylinder or drum, the weight of these items should be added to the tare weight of the cylinder or drum in order to have accurate tare setting. When a cylinder or drum is placed on the scale, the display will then indicate the actual net weight, or weight of its contents.

3 ALTERNATE TARE METHOD: Cylinders are sometimes filled to a predetermined weight. If this is the case, the operator can place the cylinder on the scale, attach all hoses, fittings, etc., and then adjust the tare knob until the display reads that positive (+) predetermined weight. Example: Cylinder is known to have 150 lbs. of contents. The operator would place cylinder or drum, fittings, hoses, etc., on the scale, then adjust the tare knob until display reads 150.0 lbs. positive. Note to Operator: It is imperative that the cylinder be completely on the disc and that no part of the cylinder or drum touches the surrounding base plate. V. Calibration Procedures: The Model 4020 Drum Scale is calibrated at the factory to within specified accuracy and is calibrated to standards traceable to the Bureau of Weights and Measures. NO FURTHER CALIBRATION SHOULD BE NECESSARY. If the electronics, or load cells in the base, are replaced, the following calibration procedure should be used. VI. 1. Plug the AC cord into a well grounded receptacle. There is no power on-off switch, so power is applied as soon as it is plugged in. 2. Allow approximately 15 minutes warm up time before calibration. 3. While weighmeter is warming up, proceed to check under cylinder support disc to make sure there is no debris under platform. 4. Turn back tare knob clockwise until you reach the lock, and check display. There should be at least +10 lb. If the number is not close to this, please adjust R-6 on the pc board to allow for at least +10 lb. 5. Adjust tare knob on front of weighmeter until scale reads zero with no weight on platform. Adjust R-22 to set your 4 ma reading to represent 0 lb. on the display. (Your ma meter will be connected at the locations Out and Ret on the connector for this reading, with NO wires on the terminal.) 6. Place a known weight on disc as close to center as possible. Test weight should be equal to ½ of capacity or greater. 7. Adjust R-15 until the display shows the weight shows the amount of weight you are calibrating with. 8. You must now calculate the correct milliamp reading for this weight that you are using to calibrate. The formula is as follows: 16mA divided by the number of pounds in the maximum net weight of scale. (Ex: your scale is set for 0 to 500lb. net weight. You would take 16 and divide by 500. This will give you the number of milliamps per pound). Using the number of milliamps per pound, multiply it times the amount of weight you are calibrating with. (You are calibrating with 400 lb. Multiply 400 x the number of milliamps per pound.) To adjust this number, then add 4 milliamps to it. (Ex. In the example we use, the result is 16.8 ma. You would turn the pot R-19 until your meter displays 16.8mA). 9. Calibration is complete when specified accuracy is obtained. Troubleshooting: 1. If display reads -1 : A. Gross underload, turn tare knob clockwise. Numbers should reappear and can be adjusted to tare weight or zero as required. B. Check load cell connections at indicator and connect wires per chart on Page 3 & 4. 2. If display reads 1 :

4 A. Gross overload, if starting point was zero, the weight is more than 199.9 lbs. for.1 resolution, weight is more than 199.9 lbs., or 1999 lbs. for 1.0 resolution. B. If starting point was -150 lbs. then the weight is more than 349.9 lbs. Remove weight and recheck tare setting. C. Check load cell connections according to Page 3 & 4. 3. If no display: A. Check power to weighmeter. B. Check green light on power supply board. It should be lit. If not lit, it is shorted out. 4. Other Problems: A. Consult factory: USA & Canada Toll-Free: (800) 257-5911 Tel: (+1) 215-766-2670 Fax: (+1) 215-766-2672 Notice: Do not return any equipment without first contacting the factory. A return authorization number will be issued and it must be marked on all materials returned to the factory, accompanying a letter that explains the problem specifically. A Serial Number will also be required. It is located inside the indicator box. VII. Board Diagram:

5

6 4-20mA Diagram: