STANDARD OPERATING PROCEDURE ARTSANA MERCURY SPHYGMOMANOMETER

Similar documents
HOME BLOOD PRESSURE KIT INSTRUCTIONS TRILINGUAL INSTRUCTION GUIDE MODEL UA-100

STANDARD OPERATING PROCEDURE

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION

STANDARD OPERATING PROCEDURE

STANDARD OPERATING PROCEDURE VORTEX-GENIE 2

Wallmax Professional Aneroid Sphygmomanometer Model 222 User Manual

Circulation and Respiration: Vital Signs Student Version

STANDARD OPERATING PROCEDURE FORCE DIAL FDN 100

USE, CARE & MAINTENANCE DIAGNOSTIX TM 972 MERCURIAL BLOOD PRESSURE INSTRUMENT

Blood Pressure Monitor

STANDARD OPERATING PROCEDURE FORCE DIAL FDN 100

Model AUTOMATIC UPPER ARM Blood Pressure Monitor

DIAGNOSTIX TM 922 MERCURIAL BLOOD PRESSURE INSTRUMENT. Use, Care, & Maintenance

BLOOD PRESSURE SENSOR BT17i USER S GUIDE

INSTRUCTION MANUAL. Automatic Blood Pressure Monitor with Fit Cuff. =Fit Cuff=!"#$% IA1B. Model

Aneroid Sphygmomanometer. Use, Care, & Maintenance

3.1.3 Weight. Frequency. Weight is obtained at the baseline examination and annually. Equipment

VER: A00. Model No:BA2010

TECHNICAL CHARACTERISTIC CT 071

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

Semi-Automatic Blood Pressure Monitor with Memory

STANDARD OPERATING PROCEDURE ADVANCED INTRAVENOUS TRAINING ARM (S )

MANUAL BLOOD PRESSURE MONITOR BPM 168B Contents Blood Pressure Monitor Intended Use What is blood pressure?

Instruction Manual. Auto Inflate Blood Pressure Monitor

User Manual. Hopkins IMPACT Digital Wrist BP RELATED ITEMS. Item #526014

Digital Blood Pressure Monitor. Model UA-705. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni 1WMPD B

Model UB-328. Wrist Digital Blood Pressure Monitor. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Oregon Scientific Wrist Blood Pressure Monitor (BPW211)

REF CH-302B IDENTIFICATION OF PARTS

MODEL CH-432B INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR. - Eng 1 - English Español Português Deutch. Italiano Français CH-432B

Auto Arm Blood Pressure Monitor

Blood Pressure Monitor

Contents. Safety Notice. Wrist Type

Objectives: Assisting with Medication, Checking vital Signs

PIMA MEDICAL INSTITUTE RADIOGRAPHY PROGRAM ARRT GENERAL PATIENT CARE COMPETENCIES CPR

User manual STARLOCK. Careways International - Innovation Quality Performance

PHY100 s Lab: The Relationship Between Liquid Pressure and Depth as it applies to Blood Pressure. Course & Sec:

BLOOD PRESSURE SENSOR 0377i

Introduction. Table of Contents. Automatic Wrist Blood Pressure Monitor With Voice-Guided Operation. Model No.: BP5K

REF CH-602B IDENTIFICATION OF PARTS

Lesson 12: Fluid statics, Continuity equation (Sections ) Chapter 9 Fluids

English Español Português Italiano Deutch Français INSTRUCTION MANUAL FOR DIGITAL BLOOD PRESSURE MONITOR MODEL CH-656C. - Eng 1 -

BOYLE S / CHARLES LAW APPARATUS - 1m long

Digital Blood Pressure Monitor for the Upper Arm INSTRUCTION MANUAL

Digital Blood Pressure Monitor. Model UA-787

Talking Wrist Style Blood Pressure Monitor Model INS-LAB-RevB08

Talking Wrist TypeBlood Pressure Monitor. Talking Wrist Type Blood Pressure Monitor Model: BPW810. User Manual

Wrist Type Blood Pressure Monitor

MEDICARE 60. Semi-automatic upper arm blood pressure monitor. Instruction manual

The RICH LIFE Project: Blood Pressure Training Certification Checklists

G622. APPLIED SCIENCE Monitoring the Activity of the Human Body ADVANCED SUBSIDIARY GCE. Thursday 27 May 2010 Afternoon. Duration: 1 hour 30 minutes

Sports Wrist Digital Blood Pressure Monitor Item #

Soft-Over-Ball. Exercise Chart.

Jackson Heart Study Protocol. Manual 3. Blood Pressure. Visit 2. Version 2.0. September 9, For Copies, Please Contact:

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

ADC. Sphygmomanometer Accessories. Directions for Use, Care, & Maintenance AMERICAN DIAGNOSTIC CORPORATION

limbsandthings.com Advanced Venepuncture Arm User Guide For more skills training products visit Limbs & Things Ltd.

LABORATORY TECHNIQUES. Pouring Liquids

Fully Fuzzy Auto Digital Wrist Blood Pressure Monitor. tensio control. INSTRUCTION MANUAL Model GP-6220

Diagnostix TM Aneroids

MODEL 1329 Tank Gauge

Canine Endotracheal Intubation

Oregon Scientific Wrist Blood Pressure Monitor (BPW120)

INSTRUCTIONS. Manual Home Blood Pressure Monitor MODEL Stethoscope Chestpiece. Cuff. Bulb

A designated area will have been assigned for all blood specimens to be taken. All blood should then be taken in that area only.

W. D. A. SMITH Research Department of Anaesthetics, Royal College of Surgeons of England, London

VOLUMETRIC TECHNIQUES

NEW EMPLOYEE ERGONOMIC WORKSHEET

Blood Pressure Monitor

Vital Signs Monitoring

Blood Pressure Monitor

6833_INSTRUCTIONS MANUAL DIGITAL WIRST BLOOD PRESSURE

premium wrist blood pressure monitor

INDEX. Specifications 1. Important Safety Precautions 2. Operating Instructions 3, 4. Optional Extras (Meal Tray & Footrest) 5.

Operating Instructions Instrucciones de funcionamiento

BLOOD PRESSURE SIMULATOR LF01095U INSTRUCTION MANUAL

Model UA-705. Digital Blood Pressure Monitor 使用手冊翻譯. Instruction Manual. Manuel d instructions. Manual de Instrucciones. Manuale di Istruzioni

Noninvasive Blood Pressure Measurement. D. J. McMahon rev cewood

Understanding Tracheostomy Care

User Manual DMD1001. Arm Type. Digital Blood Pressure Monitor. version:1.0

STANDARD OPERATING PROCEDURE

Morning Prime Time. Practice taught by Elizabeth Integral Real IntegralReal.com

Air Plus User Manual

User Manual DMD1001. Arm Type. Digital Blood Pressure Monitor. version:1.0

Fully Automatic Blood Pressure Monitor

ESTABLISHED OVER 150 YEARS FROM Accuracy starts in the name SPHYGMOMANOMETERS

Wolfson Campus Natural Science Department. Biology Laboratory. Student Safety Contract

Blood Pressure Simulator LF01095U Instruction Manual

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

Understanding Tracheostomy Care for your Child

Blood Pressure Monitor

Lab: Safety Due: Lab # 0

High School Safety Acknowledgment Form

Chemical Spill Clean-up Protocol

Declaration of Equivalence Form

Standard Operating Procedure (SOP)

SBC 21. Electromagnetic Compatibility Information

Katy Independent School District Science Safety Rules

Basic Sanitary CPR Dog LF01156U Instruction Manual

Centrifugation: High Speed Floor Centrifuges

Transcription:

Page 1 of 8

1. Scope This Standard Operating Procedure (SOP) applies to the staff and students using the Artsana Mercury Sphygmomanometer in the Pharmacy Practice Resource Unit (PPRU) at the Pharmacy Department, University of Malta. 2. Objective To describe the procedure for the operation and maintenance of the Artsana Mercury Sphygmomanometer. 3. Definitions 3.1. Blocking Lever: A safety valve that prevents the mercury from being released out of its storage reservoir during device transportation. 3.2. Diastolic Blood Pressure (DBP): Minimum pressure in the arteries when the heart is at rest. 3.3. Mercury Sphygmomanometer: A device used to measure BP. It consists of an inflatable cuff, a hand bulb with a deflator valve and a mercury column to display the obtained BP readings. 3.4. Systolic Blood Pressure (SBP): Maximum pressure being exerted when the heart is contracting. 4. Responsibility 4.1. The members of the Department of Pharmacy (staff and students) are responsible for following this SOP. 4.2. The designated Laboratory Officer or Laboratory Assistant is responsible for ensuring that this SOP is followed. 5. Procedure 5.1. Operation 5.1.1. Open the lid of the device and set it in the proper upright position. 5.1.2. Push the blocking lever to the On position and wait for the mercury level to settle at 0. 5.1.3. Place the inflatable cuff 1-2cm above the bend of the elbow on the bare left/right upper arm. Page 2 of 8

5.1.4. Close the cuff around the arm. 5.1.5. Check that two fingers can still be inserted between the arm and the cuff. 5.1.6. Position the cuff marking on the artery. 5.1.7. Instruct the patient to lay his or her arm on a table with palm turned upwards and to keep still during the measurement. 5.1.8. Place the stethoscope chest piece below the cuff so that the membrane of the chest piece is resting against the inside of the bicep, where the artery is located (refer to SOP/PD/203_02). 5.1.9. Hold the bulb with its tubing facing upwards. 5.1.10. Check that the deflating valve on the bulb is tightly closed by rotating it clockwise. 5.1.11. Squeeze the hand bulb until the cuff pressure is 20-30 mmhg higher than the maximum expected BP. 5.1.12. Slowly start opening the deflator valve by rotating it anti-clockwise to allow air to escape uniformly at a rate of approximately 2-3 mmhg per second. 5.1.13. Record the level of the mercury column when the first rhythmical thumping sound is heard. This corresponds to the SBP. 5.1.14. Record the level of the mercury column when the thumping sound stops. This corresponds to the DBP. 5.1.15. Open the deflating valve completely to deflate the cuff. 5.1.16. Tilt the device about 45 o to the right to allow the mercury to flow back into its storage reservoir. 5.1.17. Push the blocking lever to the Off position. 5.1.18. Securely close the lid of the device. 5.2. Maintenance 5.2.1. Clean the device with a clean soft cloth. 5.2.2. Contact an authorised supplier if the glass column blackens. 5.2.3. Contact an authorised supplier if the meniscus of the mercury lies either above or below the zero mark after allowing the device sufficient time to stabilise. 5.2.4. Service the device every 2 years by an authorised supplier. Page 3 of 8

5.3. Flow Charts 5.3.1. Operation Start Open lid of device and set it in the proper upright position Push blocking lever to the On position and wait for mercury level to settle at 0 Place inflatable cuff 1-2cm above bend of elbow on bare left/right upper arm Close cuff around arm Check that two fingers can still be inserted between arm and cuff Position cuff marking on artery Instruct patient to lay his or her arm on a table with palm turned upwards and to keep still during measurement Place stethoscope chest piece below cuff so that membrane of chest piece is resting against inside of bicep, where artery is located (refer to SOP/PD/203_02) Hold bulb with tubing facing upwards Check that deflating valve on bulb is tightly closed by rotating it clockwise Squeeze hand bulb until cuff pressure is 20-30 mmhg higher than the maximum expected BP 1 Page 4 of 8

1 Slowly start opening deflator valve by rotating anti-clockwise to allow air to escape uniformly at a rate of approx. 2-3 mmhg per second Record level of mercury column when first rhythmical thumping sound is heard. This corresponds to SBP. Record level of mercury column when thumping sound stops. This corresponds to DBP. Open deflating valve completely to deflate cuff Tilt device about 45 o to the right to allow mercury to flow back into its reservoir Push blocking lever to Off position Securely close lid of device End Page 5 of 8

5.3.2. Maintenance Start Clean device with a clean soft cloth Glass column blackened No Mercury meniscus lying above or below zero mark after allowing device sufficient time to stabilise No Yes Yes Contact an authorised supplier Service the device every 2 years by an authorised supplier End Page 6 of 8

6. Precautions 6.1. Medical Device 6.1.1. Do not inflate the cuff by more than 300 mmhg since this can cause a mercury spillage. 6.1.2. When storing the device, always place the bulb with its metal deflator valve in the right case compartment to prevent accidental breakage of the glass column. 6.1.3. Avoid damaging the rubber tubings with sharp, cutting objects. 6.1.4. Do not expose the device to extremes of temperature. 6.1.5. Always store the device at a temperature range of 10 o C to 40 o C. 6.1.6. Do not use any chemical products to clean the device. 6.1.7. Do not scrub, machine wash, iron or dry clean the cuff. 6.1.8. In the case of a mercury spillage: 6.1.8.1. Ventilate room to prevent inhalation of mercury vapours. 6.1.8.2. Wear appropriate latex gloves to handle the broken material. 6.1.8.3. Gather all the spilled mercury droplets, transfer them to a glass container, close tightly and dispose off as hazardous waste. 6.2. Accurate Measurement 6.2.1. Ensure that the correct cuff size is being used to obtain precise BP readings. 6.2.2. Ensure that the patient is relaxed, still and is not speaking when measurement is being taken. 6.2.3. Wait for 5 minutes until the pulse rate returns to normal, if the patient had been active, before taking readings. 6.2.4. Ensure that the patient is sitting down and that the cuff is positioned at heart level when taking readings. 6.2.5. Ensure that the measurements are performed on a bare arm, as circulation could be restricted. 6.2.6. Ensure that the cuff is not too tight or too loose fitting as inaccurate readings may result. 6.2.7. Wait 3-4 minutes between successive readings. 6.2.8. Ensure that the membrane of the chest piece is placed on the artery. Page 7 of 8

7. Reference Artsana S.p.A. Artsana Mercury Sphygmomanometer. Grandate: Artsana. 8. Appendices N/A 9. Revision History Version Number Amendments/ Reasons for change 01 Initial Release 02 Inclusion of subheading titles of Procedure section in Table of Contents Page 8 of 8