SOP-I-101 Operation and Calibration of the Analytical Balance

Similar documents
QAM-I-117 Volumetric Equipment Calibration Verification

OPERATING PROCEDURE. Number: V2. Authors. Approvals

Technical Procedure for General Laboratory Equipment

Title: INSTRUMENT AND EQUIPMENT DOCUMENTATION AND RECORDS

Technical Procedure for Instrument Calibration and Maintenance

Standard Operating Procedure. Air Displacement Pipette Calibration

20. Technical Equipment. Archived. This section applies to all equipment used in forensic examinations, such as

User-Calibration of Mettler AT200 Analytical Balance

ENVIRONMENTAL LABORATORY APPROVAL PROGRAM CERTIFICATION MANUAL

Quantos Automated Dosing Solution Preparation Precise concentrations Process compliance Minimize out-of-specs

Predicted Dispense Volume vs. Gravimetric Measurement for the MICROLAB 600. November 2010

II. MATERIALS REQUIRED FOR PERFORMANCE TESTING

STD-3-V1M4_1.7.1_AND_ /15/2015 page 1 of 6. TNI Standard. EL-V1M4 Sections and September 2015

User Manual. Precision and Analytical Balances ME-T

Aug 17, 2001 LAB MANUAL MAXIMUM SPECIFIC GRAVITY (RICE VOIDS TEST) OF PAVING MIXTURES AASHTO Designation T 209 (Mn/DOT Modified))

Protocol Gas Verification For Compressed Gas Cylinders Containing Either SO 2, NO or CO. Quality Assurance Plan/Standard Operating Procedure

Routine Testing. SOP Temperature Calibration of Halogen Moisture Analyzer. Applicable for HR / HG / HB / MJ Moisture Analyzers

TEXAS DEPARTMENT OF PUBLIC SAFETY Breath Alcohol Laboratory

DENSITY CALIBRATIONS AT MIKES. Heikki Kajastie, Jorma Manninen and Kari Riski Centre for Metrology and Accreditation (MIKES) ABSTRACT

Liquid Handling Systems

INDUSTRIAL HYGIENE GROUP Standard Operating Procedure. TVA-1000 PID/FID Portable Toxic Vapor analyzer. 1 of 9. Contents

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

General Accreditation Guidance. User checks and maintenance of laboratory balances

Standard Operating Procedure

Oct 1, 2002 LAB MANUAL SPECIFIC GRAVITY & ABSORPTION of FINE AGGREGATE AASHTO Designation T 84 (Mn/DOT Modified)

PROTOCOL FOR COMPRESSED AIR PROFICIENCY TESTING (CAPT) PROGRAM SAMPLE ANALYSIS

Chemical Industry. School experiments. Weighing. Helper Booklet. Tips and Tricks for Weighing. Natural science laws experience live learn easily

Beamex. Calibration White Paper. Weighing scale calibration - How to calibrate weighing instruments

ali-q TM Gravimetric Verification Procedure

v /07/17 Page 1 of 18

English. Quick Guide MS-TS Precision and Analytical Balances

Attachment C. Standard Operating Procedures. AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT

How to Achieve Optimal Weighing Performance

Instrument Installation and Performance Verification

CHM 100 / Introductory Laboratory Experiment (r10) 1/11

Laboratory Activity Measurement and Density. Average deviation = Sum of absolute values of all deviations Number of trials

Ti-Pure TITANIUM DIOXIDE DETERMINATION OF SLURRY RHEOLOGY METHOD: T WP

If you have any questions concerning this report, please feel free to contact me. REPORT OF LABORATORY ANALYSIS

CHM250 Calibration and Measurement Lab. Balance Calibration

CH2250: Techniques in Laboratory Chemistry. Outline Measuring Mass Measuring Volume Significant figures. Mass Measurement

Pipetting Small Volumes

Sr. No. General points CPCB Comments 1. Print option, acknowledgement of submission for CPCB OCEMS Protocol

CHM Introductory Laboratory Experiment (r17sd) 1/13

Inspection and Approval of Asphalt Mix Design Laboratories

MILITARY SPECIFICATION POWDER, BLACK

Inspection and Acceptance of Asphalt Mix Design Laboratories

FULL OPERATING INSTRUCTIONS FOR MODELS AE 50/100 and/or AE 160/200

Quality Assurance Charting for QC Data

PERFORM Operating Document

STANDARD OPERATING PROCEDURE FOR MAINTENANCE AND CLEANING OF PULMONARY FUNCTION LABORATORY AND EQUIPMENT

Toxicology Gas Chromatography-Mass Spectrometry (GC-MS)

Measurements. Metric System

DRAFT LOCKOUT/TAG OUT PROGRAM (CONTROL OF HAZARDOUS ENERGY)

PERFORMANCE TEST INSTRUCTIONS FOR BIOHIT PROLINE PIPETTORS

STANDARD OPERATING PROCEDURES. Andersen Instruments Federal Reference Method (FRM) Ambient Particulate (PM 10 /PM 2.

Completed by: Anthony Jones (BE Design Laboratory Manager) Staff/Student number: School/Unit: BE Design Laboratory. Current version: V_01

Standards of Practice Emergency Shower/Eyewash Installation, Use, Testing & Maintenance

Improve Process Reliability

Tank and Silo Weighing. 5 Safety Tips For Tanks and Silos

PQAO Training 2017 Karl Tupper, San Luis Obispo APCD

Experiment #2. Density and Measurements

Vaisala Pressure, RH & Temp. Transmitter. The instrument was operational upon receipt. The instrument was adjusted and calibrated.

PREPARED BY: Marshall K. Cheung, Ph.D., Laboratory Director. REVISED BY: Marshall K. Cheung, Ph.D., Laboratory Director

QUICK START GUIDE TO DRINKING WATER MONITORING

METHOD 204F--VOLATILE ORGANIC COMPOUNDS CONTENT IN LIQUID INPUT STREAM (DISTILLATION APPROACH) 1.1 Applicability. This procedure is applicable for

Pressure Metrology in Mexico

Standard Operating Procedure Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES) - Thermo Scientific icap 6300

Lab Skills Practice: Pipetting Small Volumes. B3 Summer Science Camp at Olympic High School 2016

SURVEYING THE HAZARDOUS MATERIALS INCIDENT

SmartCal. SmartCal Troubleshooting Checklist

SAUTER FK Version /2016 GB

User Manual. Portable Balances PL-E

STANDARD OPERATING PROCEDURES

CBMTG Laboratory Committee-Laboratory Guidance

Transferpette. Testing Instructions (SOP) 1. Introduction. October 1998

Information here is intended to assist the investigator in the

Experiment 1 Basic Laboratory Operations

User Manual. Quantos Automated Dosing Powder Module

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

General European OMCL Network (GEON) QUALITY MANAGEMENT DOCUMENT

Calibration of Volumetric Glassware

Module 14: Blood Collection and Handling DBS

December 16, I. Introduction

Safety In the Science Lab. Safety First

Yeo Soh Bee A/Prof Dan Yock Young 28/03/2016 Prepared by Approved By Issue Date

Service Calibration are you doing it properly?

Name:. Correct Questions = Wrong Questions =.. Unattempt Questions = Marks =

Requirements and Methods for Calibration of Equipment in Clinical Laboratories. NK Chan (HKRCBTS) 24 August 2007

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Best Practice for Calibrating LTH Conductivity Instruments

SOP for use of the SX19 Stopped-Flow Instrument

ANALOG PRESSURE GAUGE

Specific Accreditation Criteria Calibration ISO IEC Annex. Mass and related quantities

CALIBRATION CERTIFICATE # 500

Lab 1. Instrumentation Familiarity: Using Micropipetters and Serological Pipettes

Calibration Procedure

OPERATION OF LIQUID PARTICLE COUNTER. GP-B P0045 Rev -

Department of Laboratories St. Louis, MO ISSUE DATE: June 2007 REVISION DATE: July 2010 REVIEWED DATE: May 2017 PRINCIPILE:

Picus NxT Electronic Pipettes Ergonomic, Efficient, & Safe

Navigator 600 Silica analyzers

Transcription:

1. Applicability and Purpose This procedure applies to the operation and calibration of the analytical and top loading balances used for weighing. The purpose of this procedure is to provide a method for the operation and calibration of the Sartorius MC1 AC 210P, Denver Instrument Company XL-3K and Mettler PL-602 balances. The guidelines assure quality of data and uniformity of techniques between analysts. 2. Definitions a. Class 1: A standardized weight that is traceable to the National Institute of Standards and Technology (NIST). 3. Equipment, Reagents, and Standards a. Equipment i. Sartorius MC1 AC 210P analytical balances ii. Denver Instrument Company XL-3K top loading balance iii. Mettler Toledo PL-602 top loading balance iv. Plastic forceps v. Desiccator b. Reagents: Not applicable c. Standards: Store weights in their containers in a desiccator i. Troemner Inc. Class 1, 100 mg S/S Wt. or equivalent ii. Troemner Inc. Class 1, 10 g S/S Wt. or equivalent 4. Procedure a. Sartorius MC1 AC 210P i. Leveling (daily when in use) 1. If necessary, level the balance by adjusting the feet while observing the level indicator on the back of the balance. When the balance is level the bubble is in the center of the level indicator. 2. Press the power key to turn on the display (if off). ii. Internal Calibration (each time the instrument is turned on) 1. After the display reads 0.0000g, press the F1 key to initiate an internal calibration. 2. When the display reads 0.0000g, the internal calibration is complete. iii. Calibration Check (daily when in use and prior to making measurements on samples or chemicals) Rev. 10 Page 2 of 6 TIAER

1. Check to ensure the weights are within calibration by the label dates. Using the forceps, place the 100-mg Class 1 Wt. on the pan center (see Operation below). 2. Record the weight indicated in the display in the Analytical Balance Logbook or E-log (Attachment 1). 3. If weight recorded is not within ± 0.1% (0.0999-0.1001 g) of the 100 mg Class 1 Wt., initiate corrective action in accordance with SOP-Q-105, Corrective Actions. Note the CAR number in the comments column when appropriate. The balance is removed from service until the problem is investigated if not immediately corrected. iv. Operation 1. When the display indicates 0.0000 g, place the sample on the weighing pan and close the sliding doors. 2. Read the weight indicated in the display only after the weight unit appears as the stability symbol. 3. Use the tare control key, T, to zero the display or to zero the weight of a container, as needed, for next weighing. b. Denver Instrument Company XL-3K and Mettler PL-602 i. Leveling (daily when in use) 1. If necessary, level the balance by adjusting the feet while observing the level indicator on the back of the balance. When the balance is level the bubble is in the center of the level indicator. 2. Calibration Check (daily when in use) a. Check to ensure the weights are within calibration by the listed dates. Use the Class 1, 10 g weight to perform steps in Operation below. Rev. 10 Page 3 of 6 TIAER

b. Record the weight indicated in the display in the Analytical Balance Logbook (Attachment 1).If weight recorded is not within ± 0.2% (9.98-10.02 g) of the 10 g, Class 1 Wt., complete corrective actions in accordance with SOP-Q- 105, Corrective Actions. Note the CAR in the comments section when appropriate. The balance is removed from service until the problem is investigated and corrected, if not immediately corrected. 3. Operation a. When the display indicates 0.00g, place the sample or item on the weighing pan. b. Read the weight indicated in the display only after the motion indicator next to the weight unit has disappeared. c. Press the Tare control key to zero the display or to zero the weight of a container, if desired, for next weighing. 5. Quality Control and Safety Aspects a. All aspects of this procedure comply with QAM-Q-101, Laboratory Quality Control and QAM-S-101, Lab Safety. b. Calibration procedures and calibration checks are done daily before initial use of the balances. Calibration checks are recorded in the Analytical Balance Logbook (E-log). c. Instrument maintenance and internal calibrations of the balances are done annually by a contractor, or as needed, and recorded in the laboratory maintenance logbook or E-log. Contractor calibration records show the standards used bracket the range of normal use of the balance. These records are kept on file. d. Unique balance identification shall be noted in the front of the logbook as to which balance it refers to (example: S1 being the Sartorius). e. Do not handle the weights with fingers. Oily residue will change the weight. Weights should never need cleaning when properly stored in a desiccator. If the weight is dropped or gets dirty, rinse with DI water or alcohol and dry overnight in a desiccator prior to re-use. Scratched, dented or uncleanable weights are replaced. Rev. 10 Page 4 of 6 TIAER

f. The Laboratory Manager or his/her designee is responsible for ordering new weights, or having current weights certified, by an outside vendor on an annual basis (at a minimum). 6. References a. Sartorius MC1 Manual WAC 6009-e90011, Sartorius AG, Sartorius AG, Goettingen, 1991. b. XL Series Balances Operating Instructions 600248.3 Rev. F, Arvada, Colorado. c. Mettler Toledo PL-602 balance operating manual. d. National Environmental Laboratory Accreditation Conference (NELAC), The NELAP Institute, TNI standard, 2009. 7. Attachments a. Analytical Balance Log (or E-log) Rev. 10 Page 5 of 6 TIAER

Attachment 1: Analytical Balance Log Limits: for 0.100 g wt- 0.0999 to 0.1001g, for 10 g wt- 9.98 to 10.02g, for other wts- ±0.1% Elog I-101-1 rev. 9 Balance ID Date Time Initials True weight Observed Weight Pass/Fail Comments Weight ID QA check Rev. 10 Page 6 of 6 TIAER