IEP SENSOR BRACKET CHECKOUT AND VERIFICATION. This is a non-hazardous test.

Similar documents
LOW PRESSURE VOLTAGE BREAKDOWN TEST PROCEDURE FOR GSS FSU FLIGHT SPARE HVA BOARD October 28, P0938 Rev

GMA FUNCTIONAL TEST SETTING & TESTING OF SPIN UP PARAMETERS. P0586 REV -- 20, June 2000 GPB SCIENCE MISSION PROCEDURE

BLEED DOWN OF THE HIGH GMA PRESSURE

PROCEDURE TO MEASURE READOUT LOOP RESISTANCE

Coda cx1 Thermal Cycle Test Procedure

CALIBRATING THE GMA PRESSURE SENSORS AND THE ECU ENGINEERING UNIT

Proton Monitor Thermal Vacuum Test Procedure GP-B P0635

INTERNAL LEAKAGE CHECK OF THE FLIGHT GMA

TEST OF GMA SPINUP ASSEMBLY PRESSURE SENSOR CALIBRATION

DESIGN VERIFICATION AND QUALIFICATION OF NEW POPPET AND BUSHING ON GMA FLIGHT SOLENOID VALVES

GSE CAGING AND SPINUP BUFFER LINE INSTALLATION

Vibration/leak Test 2.5 and 6 Vatterfly Valves (3179 and 3223) spare units. GP-B Engineering Procedure P0761 Rev.-

STANFORD UNIVERSITY W.W. HANSEN EXPERIMENTAL PHYSICS LABORATORY GRAVITY PROBE B, RELATIVITY GYROSCOPE EXPERIMENT STANFORD, CALIFORNIA

CHECK READOUT LOOP CONTINUITY IN PROBE C

Proto-qual Vibration Test Procedure 2.5 and 6 Vacuum Valves (3179 and 3223) Retest of Flight Spare Valves. GP-B P0735 Rev.-

POROUS PLUG CHARACTERIZATION TEST

Stand-Alone Bubble Detection System

RPS900W Redundant Power Supply. Installation Guide.

GDS OPERATING PROCEDURE

GMA SLEEP PROCEDURE AT VAFB

GSE CAGING BALLAST SYSTEM INSTALLATION

DigiTRON Electrical Flying Leads, Jumpers and Harness Assemblies Site Received Test Manual

TR Electronic Pressure Regulator. User s Manual

BLEED DOWN OF THE GMA HIGH PRESSURE AT VAFB

MIL-STD-883G METHOD

BGA244 Binary Gas Analyzer

Lockout/Tagout Training Overview. Safety Fest 2013

SCRUB LAUNCH RECOVERY PROCEDURE FOR THE GMA

Rejuvenation Instructions

Operating Manual. Model 25200B/25202B Refrigerant Recovery Unit. (Not for use in automotive R-134a applications)

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

400C & 450C DUAL PERFORMANCE VALUE PACKS

WBEA Standard Operating Procedure

GAS FUEL VALVE FORM AGV5 OM 8-03

GMA VENT EXTENSION SERVICE PROCEDURE

GMA PREPARATION FOR E28 TESTING

STANFORD UNIVERSITY W.W. HANSEN EXPERIMENTAL PHYSICS LABORATORY GRAVITY PROBE B, RELATIVITY GYROSCOPE EXPERIMENT STANFORD, CALIFORNIA

WATER HEATER THERMAL EXPANSION TANKS Owner s Manual. Safety Instructions Installation Maintenance Warranty. Models: 2-5 Gallon Capacity

MASSACHUSETTS STATE POLICE CRIME LABORATORY. Certificate of Calibration Procedure for the Alcotest 9510

CLEANLINESS TEST OF PROBE SPINUP EXHAUST LINES

FX134, FX1234, FX3030 Problem and Solutions

OC Panel High Limit Aquastat Kit, Manual Reset p/n

Constant Pressure Inlet (CCN) Operator Manual

444C DUAL PERFORMANCE VALUE PACK

INSTALLATION & MAINTENANCE INSTRUCTIONS

DK46 - DK800 Supplementary instructions

PROCEDURE FOR THERMAL CYCLING IN LIQUID NITROGEN

Model DPC3500 Continuous Low Range Dew Point Analyzer. Operations Manual

QPEO2/037N Credit Value: 15 QCF Level: 2 GLH: 68 Maintaining electrical equipment/systems

IMPORTANT SAFETY INSTRUCTIONS

THE UNIVERSITY OF NORTH CAROLINA

100C Air Compressor Kit

Lab 1c Isentropic Blow-down Process and Discharge Coefficient

A C Sipper. Installation and Operation Manual. Rev Part #

LOCK-OUT/TAG-OUT (LO/TO) SAFETY PROGRAM

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

E2.14 Control of Hazardous Energy. Effective Date: 03/01/2018

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

210 Series Transmitter with External Electrochemical Sensor

May08 Rev B

Installation Procedure Esco Miri Multi-Room Incubator

HYDROSTATIC LEAK TEST PROCEDURE

DPC-30 DPC-100. Reference Manual

AC : ACCURATE CRYOCHAMBER FOR A SMALL LABORATORY WITH SMALL BUDGET

EC214 EC215 - EC215R Bench Conductivity Meters

G92 Series BASOTROL Automatic Pilot Gas Valve

- Repeat calibration until desired results are reached otherwise set SW2 and SW1-8 open.

Model 106 DPI "Micro-switch" Installation and Operating Instructions

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

BGA158 Series CE Approved Class A Shutoff Gas Valve

SEMEM3-65 Servicing anaesthetic and ventilation equipment

PRODUCT SPECIFICATION. SFP+ Cage and Cage Assembly 1 of 8 A

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Warning: NEVER ALTER TEST EQUIPMENT

SEMEM3-78 Testing and calibrating instrumentation and control equipment and circuits

Model PSI Compressor with 3-Gallon Air Tank 12VDC

SAFETY DIRECTIVE 2.0 DEPARTMENTS AFFECTED. This Administrative Directive shall apply to all Town of Marana departments and employees.

LOCKOUT/TAGOUT PROGRAM

! Warning, refer to accompanying documents.

OPERATION MANUAL BFX-2

Gas Network Craftsperson

OMS P/N: Rev.: B. Description OMS: ASSY BACKUP POD 6.0. DCR# 3137 Date: Operation: Special Instructions: TQC. Torque.

GMA TO SPACE VEHICLE PLUMBING HOOKUP

SAPCON. User Manual. Capacitance Continuous Level Indicator. . Introduction. . General Description. . Principle of Operation. .

OEM Manual. MODEL ½ Digit DRUM SCALE

BGA158 Series CE Approved Class B Shutoff Gas Valve

WARNING FOR YOUR SAFETY

ARKANSAS TECH UNIVERSITY FACILITIES MANAGEMENT HEALTH AND SAFETY MANUAL (LOCKOUT/TAGOUT) 30.0

Technical Data Sheet MF010-O-LC

Transmitter CS 21 Operation Manual

Significant Change to Dairy Heat Treatment Equipment and Systems

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

400H HARDMOUNT AIR COMPRESSOR KIT PART NO H HARDMOUNT AIR COMPRESSOR KIT PART NO

200 PSI COMPRESSORS - MODEL NUMBERS

Dual Solenoid Gas Valve Installation

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions

Additions, Revisions, or Updates

SCIENCE LAB UTILITY CONTROLS

User s Guide Temperature Sensor Converter TSC-599

PART Q CONTROL OF HAZARDOUS ENERGY (LOCKOUT-TAGOUT)

Transcription:

Page 1 of 20 NEXT GENERATION GAS SYSTEM (NGGS ) IEP SENSOR BRACKET CHECKOUT AND VERIFICATION PROCEDURE 870-PROC-623 Revision A Goddard Space Flight Center Greenbelt, Maryland 20771 FEBRUARY 13, 2002 This is a non-hazardous test. As run date: Sensor Bracket P/N: Rev.: S/N: Technical Services Division 5010 Herzel Place Beltsville, Maryland 20705 V: 301-931-0388 F: 301-931-0396

Page 2 of 20

Page 3 of 20

Page 4 of 20 TABLE OF CONTENTS 1.0 SCOPE...5 1.1 Purpose...5 2.0 APPLICABLE DOCUMENTS...5 2.1 Standards...5 2.2 GAS Program References...5 3.0 QUALITY PROVISIONS...6 3.1 Test Conductor...6 3.2 Test Prerequisites...6 3.3 Failure Criteria...6 3.4 Failure Reports...6 3.5 Test Conditions...6 3.6 Procedural Changes...6 4.0 REQUIREMENTS...7 4.1 General Requirements...7 4.2 Test Personnel...7 4.3 Test Equipment...8 4.4 Pre-Power up Visual Test...8 4.5 Test Preparation...9 5.0 ISOLATION AND CONTINUITY...10 6.0 GSE SETUP...11 7.0 THERMISTOR ALIVENESS TEST...12 7.1 Ambient Test...12 7.2 Hot Test...12 7.3 Cold Test...13 8.0 PRESSURE TRANSDUCER ALIVENESS TEST...15 8.1 Ambient Pressure Test...15 8.2 Low Pressure Test...15 8.3 High Pressure Test...16 9.0 PRESSURE TRANSDUCER CHARACTERIZATION...17 9.1 Plumbing Setup...17 9.2 Low Pressure Test...17 9.3 High Pressure Test...18 10.0 END OF TEST...20

Page 5 of 20 1.0 SCOPE This procedure tests the NGGS IEP sensor bracket assembly in a stand-alone configuration. This test is to be performed on the assembly prior to integration the next higher assembly. This procedure can also be used to retest the assembly as required. 1.1 PURPOSE The goal of this procedure is to verify the electrical operation of the NGGS Sensor Bracket Assembly. The Sensor Bracket will be mounted to the NGGS IEP. The bracket contains a pressure transducer and a thermistor. The pressure transducer is a modified, off the shelf Kavlico (p/n P155-30A-E2A) pressure transducer. A portion of this procedure will be to characterize the pressure transducer. 2.0 APPLICABLE DOCUMENTS 2.1 STANDARDS Unless otherwise specified, the latest revision of the documents below should be used. FED-STD-209B MIL-STD-45662A NASA-STD-8739.7 Requirements for Cleanliness Calibration System Requirements Requirements for Electrostatic Discharge Control 2.2 GAS PROGRAM REFERENCES Unless otherwise specified, the latest revision of the documents below should be used. 2035521 Assembly, IEP Sensor Bracket, NGGS 2035524 Altered Part, Pressure Transducer, NGGS 2035503 Wiring Diagram, Standard IEP, NGGS 2035660 Assembly, Sensor Bracket EGSE 2035661 Assembly, Sensor Bracket Test Valve

Page 6 of 20 3.0 QUALITY PROVISIONS Stated below are the Quality Provisions for Quality Assurance (QA) as they relate to the assembly under test. 3.1 TEST CONDUCTOR The performance of operations specified in this procedure does not require the supervision of the appropriate Quality Assurance representative. Notify QA in advance of the execution of this test procedure. QA has the option of monitoring test activities. 3.2 TEST PREREQUISITES The Test Engineer shall certify the item is complete and ready for testing by verification of the following: a) During performance of this test all special and general test equipment shall be within prescribed certification or calibration intervals as applicable. b) Upon completion of the test, the resultant test data shall be attached or included within the item s build data. 3.3 FAILURE CRITERIA The inability of the item to perform its required functions, as specified, during test operation, or mechanical degradation, which will impair its mounting and/or alignment, shall be considered a failure. 3.4 FAILURE REPORTS Malfunction reporting shall be in accordance with the GSFC NCR System. Failure reporting shall be in accordance with ISO 9000. 3.5 TEST CONDITIONS The test area shall comply with the following restrictions for bench level testing: a) Relative Humidity: 50 20%. b) Cleanliness: As defined by FED-STD-209B c) ESD Protection: As defined by NASA-STD-8739.7 3.6 PROCEDURAL CHANGES To facilitate the operation of this procedure, other procedures, or applicable parts may be performed out of sequence before this procedure. Deviations and/or redlines from this procedure during test shall be recorded on the appropriate page and approved prior to performance of the deviation by the Test Engineer and/or the Design Engineer. All redlines shall be initialed by the Test Engineer.

Page 7 of 20 4.0 REQUIREMENTS 4.1 GENERAL REQUIREMENTS The following general requirements shall be followed while performing this procedure. 4.1.1 Test Responsibility a) The Test Engineer has authority over the execution of the test and is responsible for approving or disapproving any deviations from this procedure. b) The Test Conductor is responsible for adhering to the test procedure, conducting the test, recording all test data, and writing deviation, as required. 4.1.2 Precautions and Operating Instructions! CAUTION! - USE EXTREME CAUTION WHEN PROBING FLIGHT HARDWARE a) The requirements of handling flight hardware properly shall be followed. Personnel handling the flight hardware shall be ESD certified. b) All breast pockets shall be emptied, and badges tucked in or put away, prior to working above the flight hardware. c) Extreme caution shall be used when handling the flight hardware to prevent damage to thermal surfaces, finishes, or electrical parts. d) The test area shall be kept as clean and neat as possible during the test. 4.1.3 Test Records a) All data shall be recorded in pen. b) All major steps are verified complete by the test conductor initialing where indicated by the blank at the end of the step. c) All data taken during this test shall be recorded in, and/or attached to this procedure. d) All hardcopies of waveform plot shall be labeled with the date, time, plot number, this procedure number, test section number, and the unit under test serial number. All waveform traces shall be labeled. e) If deemed necessary by the Test Engineer, a summary report highlighting significant data should be generated and attached to this procedure. 4.2 TEST PERSONNEL Indicate the personnel participating in this test. Name Initials

Page 8 of 20 4.3 TEST EQUIPMENT The following list of test equipment is required to perform this procedure. The equipment shall be in accordance with MIL-STD-45662A or equivalent. If additional equipment is used during the procedure, record the information to the list. Table 4-1 Equipment List Nomenclature Manufacturer Model No. ECN Cal Due DMM #1 DMM #2 Power supply (+9 to 12V) Control pressure transducer Sensotec 990607RA0900 Sensor bracket EGSE GSFC 2035660 N/A N/A Break out box, DB9 N/A N/A N/A N/A DB9 straight cable N/A N/A N/A N/A Wire leads and jumpers N/A N/A N/A N/A Thermometer N/A N/A N/A N/A Heat gun N/A N/A N/A N/A Freeze Spray N/A N/A N/A N/A Pressure/vacuum chamber N/A N/A Vacuum pump N/A N/A Pressure source N/A N/A Pressure Sensor GSE GSFC 2035661 N/A N/A 4.4 PRE-POWER UP VISUAL TEST a) A visual check shall be performed on the item under test to ensure each of the required steps (as listed on the WOA) to prepare the unit for test has been implemented correctly. b) A visual check shall be performed to ensure the item to be tested is free of defects, foreign objects or debris, and in the proper configuration for this test.

Page 9 of 20 4.5 TEST PREPARATION 4.5.1 Record the sensor bracket assembly information below. Assembly Part Number: Assembly Revision Level: Assembly Serial Number: Pressure Transducer Manufacturer: Pressure Transducer Part Number: Pressure Transducer Serial Number: Work Order Number: ( OSC / SAI / GSFC ) Work Order Event Number:

Page 10 of 20 5.0 ISOLATION AND CONTINUITY This section should be performed prior to the first power up of the assembly. 5.1.1 Connect the 9-pin break-out-box to the sensor bracket J301. 5.1.2 Using DMM #1, measure and record the isolation and continuity resistance measurements in the Table 5-1 below. 5.1.3 Verify that all signals are isolated from chassis. Do not proceed any further if any of the isolation measurements have failed. Table 5-1 J301 Isolation & Continuity Measurements Plus Lead Measurement Minus Lead Expected Value Measured Value Pass/Fail Chassis Pin-1, Therm RTN > 10 M Ohms Chassis Pin-2, IEP Therm > 10 M Ohms Chassis Pin-3, IEP Press > 10 M Ohms Chassis Pin-4, IEP RTN > 10 M Ohms Chassis Pin-5, Sensor PWR > 10 M Ohms Pin-1, Therm RTN Chassis > 10 M Ohms Pin-2, IEP Therm Chassis > 10 M Ohms Pin-3, IEP Press Chassis > 10 M Ohms Pin-4, IEP RTN Chassis > 10 M Ohms Pin-5, Sensor PWR Chassis > 10 M Ohms Pin-2, IEP Therm Pin-1, Therm RTN 10.5K +/- 2K * Pin-5, Sensor PWR Pin-4, IEP RTN 10.5K +/- 2K Pin-3, IEP Press Pin-4, IEP RTN 35K +/- 3K * Varies with temperature.

Page 11 of 20 6.0 GSE SETUP The sensor bracket EGSE is used for the remainder of the tests in this procedure. 6.1.1 Refer to Figure 6-1 below for the electrical GSE setup using the sensor bracket EGSE P/N 2035660. 6.1.2 Set the power supply to 10VDC and then verify the power supply is turned OFF. Connect the power supply to the Sensor Bracket EGSE. 6.1.3 Connect DMM #2 setup to measure current across TP1 (+) and TP2 (-). Set the current monitor switch to ON. 6.1.4 Connect DMM #1 setup to measure voltage. Connect the minus lead to TP5 (-). The plus lead will be connected to either TP3 or TP4 to measure the pressure or thermistor voltages respectively. 6.1.5 Using a DB-9 straight cable, connect the Sensor Bracket EGSE to the sensor bracket. 6.1.6 Setup and power on the calibrated pressure transducer. Current DMM #2 + - Voltage DMM #1 - + SW1 Current Monitor ON/OFF TP1 Current Monitor + TP2 Current Monitor TP3 Pressure Voltage TP4 Thermistor Voltage TP5 Voltage Ground Power Supply TP1 TP2 TP3 TP4 SW1 TP5 Sensor Bracket EGSE DB9 J301 Flight Sensor Bracket Sensor Display Unit Calibrated Pressure Sensor Figure 6-1 - Sensor Bracket Electrical GSE Setup

Page 12 of 20 7.0 THERMISTOR ALIVENESS TEST This section verifies the general operation of the sensor bracket thermistor. The thermistor is a YSI- 44006. This test is not a characterization or calibration of the performance of the thermistor. 7.1 AMBIENT TEST 7.1.1 Turn ON the GSE power supply. 7.1.2 Record the ambient temperature using a thermometer in Table 7-1 below. 7.1.3 Record the thermistor voltage observed in Table 7-1 below. 7.1.4 Using the chart in Figure 7-1 and the thermistor voltage, determine the thermistor temperature and record it in Table 7-1 below. 7.1.5 Turn OFF the GSE power supply. 7.2 HOT TEST This test requires the use of the heat gun. 7.2.1 Turn ON the GSE power supply. 7.2.2 Apply heat to the thermistor, being careful not to over heat the unit. 7.2.3 Verify that the thermistor voltage is decreasing. It is not necessary to make the voltage drop more than 1V. 7.2.4 Record the minimum thermistor voltage observed in Table 7-1 below. 7.2.5 Using the chart in Figure 7-1 and the thermistor voltage, determine the thermistor temperature and record it in Table 7-1 below. 7.2.6 Turn OFF the GSE power supply.

Page 13 of 20 7.3 COLD TEST This test requires the use of the freeze spray. Note: The freeze spray will make the affected suface cold in a short amount of time. Use in a well ventilated area. 7.3.1 Turn ON the GSE power supply. 7.3.2 Apply freeze spray to the thermistor, being careful not to over freeze the unit. 7.3.3 Verify that the thermistor voltage is increasing. It is not necessary to make the voltage increase more than 1V. 7.3.4 Record the maximum thermistor voltage observed in Table 7-1 below. 7.3.5 Using the chart in Figure 7-1 and the thermistor voltage, determine the thermistor temperature and record it in Table 7-1 below. 7.3.6 Turn OFF the GSE power supply. Table 7-1 Thermistor Aliveness Test Data Measurement Expected Value Measured Value Pass/Fail Thermometer temperature N/A C N/A Amb. Thermistor Voltage DMM #1 N/A VDC N/A Thermistor temperature Thermometer +/-2 C C Cold Test Hot Test Thermistor Voltage DMM #1 0V < V < Amb. Volt. VDC Thermistor temperature < Thermometer C Thermistor Voltage DMM #1 Amb. Volt. < V < 5V VDC Thermistor temperature > Thermometer C

Page 14 of 20 A/D Input Voltage (V) V out V in R 1 T 5 0 A/D Input Voltage vs. Temperature 5 4.9 4.8 4.7 4.6 4.5 4.4 4.3 4.2 4.1 4 3.9 3.8 3.7 3.6 3.5 3.4 3.3 3.2 3.1 3 2.9 2.8 2.7 2.6 2.5 2.4 2.3 2.2 2.1 2 1.9 1.8 1.7 1.6 1.5 1.4 1.3 1.2 1.1 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0 60 50 40 30 20 10 0 10 20 30 40 50 60 70 80 90 100 110 120 60 C T 120 Environment Temperature (deg. C) Figure 7-1 YSI-44006 Thermistor Temperature Conversion

Page 15 of 20 8.0 PRESSURE TRANSDUCER ALIVENESS TEST This section performs an aliveness test on the pressure transducer. This is not a calibration or characterization test. 8.1 AMBIENT PRESSURE TEST 8.1.1 Use the control pressure transducer then measure and record the ambient pressure in Table 8-1 below. 8.1.2 Turn ON the GSE power supply. 8.1.3 Record the sensor current on DMM #2 and the sensor output voltage on DMM #1 in Table 8-1 below. 8.1.4 Turn OFF the GSE power supply. 8.2 LOW PRESSURE TEST This test requires the use of a vacuum source. The control pressure transducer and the sensor bracket must see the same vacuum. This can be accomplished with plumbing or by using a small vacuum chamber. 8.2.1 Pull a slight vacuum (~3 psia) on the system. 8.2.2 Using the control pressure transducer, verify that the chamber is holding the pressure and that the reading has settled out. This may take a few minutes. 8.2.3 Use the control pressure transducer then measure and record the pressure in Table 8-1 below. 8.2.4 Turn ON the GSE power supply. 8.2.5 Record the sensor current on DMM #2 and the sensor output voltage on DMM #1 in Table 8-1 below. 8.2.6 Verify that the sensor output voltage is less than the ambient voltage. 8.2.7 Turn OFF the GSE power supply.

Page 16 of 20 8.3 HIGH PRESSURE TEST This test requires the use of a pressure source. The pressure presented to the sensor must never exceed 31 psia or permanent damage to the unit will result. The control pressure transducer and the sensor bracket must see the same pressure. This can be accomplished with plumbing or by using a small pressure chamber. 8.3.1 Increase the pressure (~3 psia) on the system. 8.3.2 Using the control pressure transducer, verify that the system is holding the pressure and that the reading has settled out. This may take a few minutes. 8.3.3 Use the control pressure transducer then measure and record the pressure in Table 8-1 below. 8.3.4 Turn ON the GSE power supply. 8.3.5 Record the sensor current on DMM #2 and the sensor output voltage on DMM #1 in Table 8-1 below. 8.3.6 Verify that the sensor output voltage is greater than the ambient voltage. 8.3.7 Turn OFF the GSE power supply. Table 8-1 Pressure Transducer Aliveness Test Data Measurement Expected Value Measured Value Pass/Fail Calibrated pressure sensor N/A psia N/A Amb. Current DMM #2 3.0 ma +/- 0.5 ma ma Sensor output voltage DMM #1 N/A VDC N/A Low Press High Press Calibrated pressure sensor N/A psia N/A Current DMM #2 3.0 ma +/- 0.5 ma ma Sensor output voltage DMM #1 N/A VDC N/A Calibrated pressure sensor N/A psia N/A Current DMM #2 3.0 ma +/- 0.5 ma ma Sensor output voltage DMM #1 N/A VDC N/A

Page 17 of 20 9.0 PRESSURE TRANSDUCER CHARACTERIZATION This section performs a detailed characterization test on the pressure transducer over its operational range. The control pressure transducer and the sensor bracket must see the same pressure. 9.1 PLUMBING SETUP Electrically the setup is the same as in section 6.0. Below is the plumbing setup recommended for this test. The leak valve in combination with the flow valve is used to control the system pressure. The plumbing uses the Sensor Bracket Test Valve Assembly GSE P/N 2035661. P/N 2035661 Sensor Display Unit Calibrated Pressure Sensor Flow Valve Vacuum Source T T Sensor Bracket EGSE Flight Sensor Bracket Leak Valve GN2 Pressure Source P/N 2035660 Figure 9-1 Plumbing Setup 9.2 LOW PRESSURE TEST This test requires the use of a vacuum source. The control pressure transducer and the sensor bracket must see the same vacuum. 9.2.1 Pull a vacuum on the system until is minimal pressure is reached. This may take a few minutes to 1 hour depending on the setup and equipment used. 9.2.2 Using the control pressure transducer, verify that the system is holding the pressure and that the reading has settled out. This may take a few minutes. 9.2.3 Turn ON the GSE power supply. 9.2.4 Adjust the system pressure using a valve to create a slow leak to allow the system to slow increase in pressure. Adjust the system pressure (as read on the control pressure transducer) as close as possible to each of the target pressures listed in Table 9-1 below. 9.2.5 At each target pressure measure and record the control pressure transducer, the sensor current on DMM #2, and the sensor output voltage on DMM #1 in Table 9-1 below. 9.2.6 Turn OFF the GSE power supply.

Page 18 of 20 9.3 HIGH PRESSURE TEST This test requires the use of a pressure source. The pressure presented to the sensor must never exceed 31 psia or permanent damage to the unit will result. 9.3.1 Place the system under pressure until maximum pressure of 30 psia is reached. DO NOT EXCEED 31 PSIA. 9.3.2 Using the control pressure transducer, verify that the system is holding the pressure and that the reading has settled out. This may take a few minutes. 9.3.3 Turn ON the GSE power supply. 9.3.4 Adjust the system pressure using a valve to create a slow leak to allow the system to slow decrease in pressure. Adjust the system pressure (as read on the control pressure transducer) as close as possible to each of the target pressures listed in Table 9-1 below. 9.3.5 At each target pressure measure and record the control pressure transducer, the sensor current on DMM #2, and the sensor output voltage on DMM #1 in Table 9-1 below. 9.3.6 Turn OFF the GSE power supply. Table 9-1 Pressure Transducer Characterization Test Data Target Pressure +/- 0.5 psia Calibrated Pressure Sensor Measurement Sensor Current DMM #1 Sensor Voltage DMM #1 1.0 psia psia ma VDC 3.0 psia psia ma VDC Low Press 6.0 psia psia ma VDC 9.0 psia psia ma VDC 12.0 psia psia ma VDC 14.7 psia psia ma VDC 27.0 psia psia ma VDC High Press 24.0 psia psia ma VDC 21.0 psia psia ma VDC 18.0 psia psia ma VDC 14.7 psia psia ma VDC

Page 19 of 20 9.3.7 Plot the data collected in Table 9-1 on the graph in Figure 9-2 below. Pressure (psia) 30 29 28 27 26 25 24 23 22 21 20 19 18 17 16 15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 P/N: Rev: S/N: Date: 0 0.0V 0.5V 1.0V 1.5V 2.0V 2.5V 3.0V 3.5V 4.0V 4.5V 5.0V Pressure Transducer Voltage (VDC) Figure 9-2 Voltage vs. Pressure

Page 20 of 20 10.0 END OF TEST 10.1.1 Turn OFF all power supplies. 10.1.2 Disconnect the test setup. 10.1.3 If hardcopy plots of waveforms were generated, attach them to this procedure. If files were made, print out hardcopies and attach them to this procedure. 10.1.4 The test conductor shall sign below indicating that this procedure has been completed. QA shall review the as run procedure and sign below. Test conductor Date Quality Assurance Date