DIGIVAC 22W Series INSTRUCTION MANUAL. Wide Range Linearized Vacuum Transmitter and Controller YOU MUST READ THIS MANUAL BEFORE USE

Similar documents
INSTRUCTION MANUAL. DIGIVAC Model 801 Series. Digital Vacuum Gauge. 801W Ranges to 760 Torr 1 to 9999 microns.001 to 999 millibar

YOU MUST READ THIS MANUAL BEFORE USE

INSTRUCTION MANUAL. Digivac Model 276. Digital Vacuum Gauge. Digivac Models Model Ranges.001 to 760 Torr.001 to 1013 millibar

KJL-205A. Operation Manual YOU MUST READ THIS MANUAL BEFORE USE

INSTRUCTION MANUAL. DIGIVAC Model 200 Pirani. Digital Vacuum Gauge Revsion K13A YOU MUST READ THIS MANUAL BEFORE USE. Readings from millitorr

INSTRUCTION MANUAL. Lesker Model KJL615TC-K Lesker Model KJL615TC-H. Digital Vacuum Gauge/Controller. KJL615TC K/H Ranges

INSTRUCTION MANUAL MP4AR Remote Convection Gauge Range: 1 x 10-3 Torr to 1 x 10+3 Torr

105 B C h u r c h S t, M a t a w a n, N J s a l e d i g i v a c. c o m w w w. d i g i v a c. c o m Copyright 2015,

YOU MUST READ THIS MANUAL BEFORE USE

Instruction Manual. Digital Thermocouple Gauge Control Units Models DTC & DTC Models DTC-06M-115 & DTC-06M-230

DPC-30 DPC-100. Reference Manual

PRESSURE SENSOR - ABSOLUTE (0 TO 700 kpa)

URC Voltage Sensor SEN-VOLT for use with MRX units containing sensor ports

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

INSTRUCTION MANUAL. FLOW CONTROL DRAWERS MANUAL / PLC CONTROL SERIES Model Version Perma Pure LLC Tel:

Advanced Test Equipment Rentals ATEC (2832)

UNIVERSAL MASK TEST SYSTEM

INSTALLATION & MAINTENANCE INSTRUCTIONS

E8AA. Pressure Sensor of Stainless Steel Construction Is Ideal for a Wide Range of Applications. Pressure Sensor (Stainless Steel Diaphragm)

CSA Sample Draw Aspirator Adapter Operator s Manual

In Vivo Scientific, LLC INSTRUCTION MANUAL

Model DPC3500 Continuous Low Range Dew Point Analyzer. Operations Manual

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION SPECIFICATIONS. CE (EMC) Compliant

RAM 4021 Operation Manual

92831 TEL: (714) FAX:

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

PM105 PRESSURE METER USER S MANUAL

Model ASSEMBLY and OPERATING INSTRUCTIONS

TELEDYNE HASTINGS INSTRUMENTS MODEL 40 POWER SUPPLY INSTRUCTION MANUAL

Columbus Instruments

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

Sterile Visual Flow Indicator

Installation and Operation Manual

92831 TEL: (714) FAX:

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

T EK-SUB 4800C 19 mm Submersible Level Transmitter

RK-IR Sample Draw Aspirator Adapter Operator s Manual

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

E2K-L. Liquid Level Sensor That Is Unaffected by the Color of the Pipe or Liquid. Liquid Level Sensor. Ordering Information

Instruction Manual BA90 Precision Barometer

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

Design Features. General Description

ROPV R40 E Series User Manual

PC7-C Flowmeter O2/N2O Sedation Unit Instruction Manual

Yoke Block Instruction Manual

ASDX Series Silicon Pressure Sensors

P5523. Users Manual. Liquid to Gas Separator. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

MASS FLOW SYSTEMS MASS FLOW MEASURING, CONTROLLING AND BLENDING SYSTEMS

Basic Nitriding Sampling System Hydrogen Analyzer with Calculated % DA, % NH 3, and K N Values. Operations Manual

SPECIFICATIONS APCEPH1

BAROMETER PRESSURE STANDARD PRESSURE CONTROLLER

MODEL GT820 OXYGEN SENSOR

4-in-1 Professional Inflator Kit

Instruction Manual AD 1000 / BA 1000 Absolute Pressure Sensor

2600T Series Pressure Transmitter Models 264DC Differential and 264HC Gage Level Transmitters. Kent-Taylor

Digital Vacuum Regulator

Yoke Block Instruction Manual

WATER HEATER THERMAL EXPANSION TANKS Owner s Manual. Safety Instructions Installation Maintenance Warranty. Models: 2-5 Gallon Capacity

Model GFC 7000T CO 2 Analyzer (Addendum to GFC 7000TA Manual, PN 07272)

How to specify a product. Process Sensors and Mechanical Instruments

BAPI Pressure Line of Products - FAQs

Precision pressure sensor Basic version Model CPT6020

TOP VALVE. Pat. #5,857,486 & 5,944,050. Mid-Range Pressure PSIG Back Pressure and Pressure Relief Valves. Instruction Manual

Acid- Rite ph Adjustment System 2500 by Axiall, a Westlake Company

RK LEL Sample Draw Aspirator Adapter Operator s Manual

DeltaSpan Pressure Transmitters LD10 Series Owner s Manual

Instruction Manual 742 5/1/2009. Eclipse Ratio Regulators ES Series Version 1

BI-680 Online Dissolved Oxygen Controller Instruction Manual

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

P9000 DESCRIPTION. For parts requiring RoHS compliance, please contact factory. P February /6

TL-12 FLUSH TANK. Manual Ref. # MN-??? Polyurethane Machinery Corp. REVISION 1.0

GFM. Typical Aluminum GFM Mass Flow Meter NIST MASS FLOW METERS. Principles of Operation. Design Features

Model 4140/4143 General Purpose Thermal Mass Flowmeter

User s Guide Temperature Sensor Converter TSC-599

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

KJ4000 Operating Instructions & Parts Manual NSN

Compact Triple Cabinet Outlet Station Model Installation and Operating Instructions

PRESSURE COMPARATOR 10000PSI 700bar. Operator Instructions CRYSTAL. engineering corporation

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

6 digital caliper with case

Instruction Manual Differential Pressure Transmitter

ACCURA D80 User Manual

Transmitter CS 21 Operation Manual

Instruction Manual No. 742, 8/98

Digital Vacuum Regulator

210 Series Transmitter with External Electrochemical Sensor

TOP VALVE. Pat. #5,857,486 & 5,944,050. High Temperature: max. 300 F (149 C) Back Pressure And Pressure Relief Valves. Instruction Manual

Model 4040/4043/4045 General Purpose Thermal Mass Flowmeter

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

Installation, operating and maintenance Instructions for Seemag bypass level indicator

3800 SERIES END SUCTION PUMP REPAIR PARTS INDEX. Model Model 3804

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

RDK-408D2 Cold Head. Technical Manual. SHI-APD Cryogenics Inc Vultee Street Allentown, PA U.S.A. Revision A: September 2005

Research Grade Blood Pressure Transducer User's Manual

HLM-2000-TX Gas Transmitter User Manual

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

RAM 4021-DPX Operation Manual

Model Series 62 Constant Differential Relay

O3 3E 1 F Gas Sensor Module

FTS SUBMERSIBLE PRESSURE TRANSMITTER USER S MANUAL

Transcription:

DIGIVAC 22W Series INSTRUCTION MANUAL Wide Range Linearized Vacuum Transmitter and Controller YOU MUST READ THIS MANUAL BEFORE USE The Digivac Company 105B Church Street Matawan, NJ 07747 Email: sales@digivac.com (732) 765-0900 (732) 765-1800 FAX www.digivac.com

CONTENTS 1.0 Description and Principle of Operation 2.0 Construction 3.0 Unpacking and Inspection 4.0 Installation 5.0 Operation 6.0 Servicing, Maintenance and Calibration 6.1 Factory Repair and Calibration 6.2 Field Calibration 7.0 Notes on Calibration 8.0 Understanding Torr 9.0 Accessories and Modifications 10.0 Attachments, Illustrations and Specifications 11.0 Terms and Conditions

1.0 DESCRIPTION AND PRINCIPLE OF OPERATION The DIGIVAC Model 22W (LCD) Vacuum Transmitter is an active gauge compact digital vacuum sensing instrument. It works in conjunction with a DC powered Vacuum Gauge tube to produce a linear and RS-232 output proportional to the sensed vacuum. This instrument has a set point enabling it to control external equipment. The DigiVac Model 22W operates when the unit s vacuum gauge tube produces a millivolt output proportional to its temperature rise above ambient. This millivolt signal is amplified and conditioned by the microprocessor. A control algorithm in the processor adjusts the excitation of the gauge tube so as to hold the tube at an essentially constant temperature. The pressure is communicated with the user as a number on the digital display, a signal on the unit s serial port, and as an analog voltage. Consult the DigiVac website www.digivac.com for information about other DigiVac vacuum controllers and gauges. 2.0 CONSTRUCTION The Model 22W consists of indicating and controlling instrument, and the gauge tube. The gauge tube houses the various thermocouple sensing, heating and compensating elements and terminates in an octal connector. Regulating circuitry in the compact controller provides proper current for gauge tube excitation, and thus compensates for resistance in the probe leads. The Model 22W LCD option includes a.62 x 1.49 inch LCD screen that provides a digital readout of real time vacuum pressure.

3.0 UNPACKING AND INSPECTION After the DIGIVAC Model 22W is received, it should be carefully unpacked and inspected for damage during shipment and for completeness. In the event of a loss shipment, a claim should immediately be made to the common carrier or the postal service, as applicable. The DigiVac warranty pertains only to the instrument, and does not cover losses in shipping. For more information regarding our warranty visit www.digivac.com Each Model 22W should come with: -A vacuum gauge controller - A thermocouple vacuum gauge tube (vacuum sensor) -An AC adapter that runs on 100-230VAC, 50/60 Hz with line cord adapter -Pre-tested under actual vacuum against a NIST standard It is also available with an optional output vacuum reading to LCD display.

4.0 INSTALLATION The instrument should be mounted into a clean, dry part of a vacuum system. The electrical connections can be made as follows: Pin 1 Analog Input (Set point) Pin 2 RS232 Transmit Output (9600 N81) Pin 3 No connection Pin 4 Analog Output Pin 5 Power and Signal ground Pin 6 Relay Normally Open Pin 7 Relay Common Pin 8 Relay Normally Closed Pin 9 See packing list. 5 to 24 VCD (adjacent to pin 5) The Gauge Tube should be installed stem down in a clean, dry vacuum system to facilitate drainage The axis of the tube should be vertical, although angles of up to 45 degrees are generally well tolerated. The unit should not be mounted directly on a vacuum pump It is best to avoid mounting the unit where it will be subjected to excessive vibration or oil vapors If it must be placed close to a vacuum pump, it is desirable to have at least two 90 Degree turns in the piping, this way, there will be no direct optical path for oil vapors, and gauge tube life and calibration will be maximized The vacuum sensor has a 1/8 pipe thread connection. If any significant toque is to be applied to this connection, Use a 7/16 open-end wrench Do not use the body of the instrument as a wrench!

5.0 OPERATION After installation, the instrument is ready for immediate operation. The unit consists of the instrument housing with interface connector and a Vacuum Gauge tube. It can be powered by an external source of 5.0V or with the included external 115/230 VAC power adapter. Analog Output The unit has an analog output, which can be used to transmit vacuum to external equipment such as PLC s or chart recorders. This analog output is normally scaled by 5V5T volts. LBD volts are also available, as described below. The settings are scaled, linear-by-decade, as follows: 0.0 to 0.999 volts = 0 to 999 millitorr 1.1 to 1.99 volts = 1.0 to 9.9 torr 2.1 to 2.99 volts = 10.0 to 99 torr 3.1 to 3.76 volts = 100 to 760 torr Torr Display Scale LBD Volts 5V5T Volts.000 0 0 0 0 0.000 0.000V.100 1 0 0 0 0.100 0.100V.200 2 0 0 0 0.200 0.200V.500 5 0 0 0 0.500 0.500V.925 9 2 5 0 0.925 0.925V 1.00 1.0 0 1 1.100 1.000V 2.20 2.2 0 1 2.200 2.200V 5.18 5.1 8 1 1.518 5.000V 8.86 8.8 6 1 1.886 5.000V 11.9 11. 9 2 2.119 5.000V 34.7 34. 7 2 2.347 5.000V 55.3 55. 3 2 2.553 5.000V 71.9 71. 9 2 2.719 5.000V 96.4 96. 4 2 2.964 5.000V 110 110 3 3.110 5.000V 340 340 3 3.340 5.000V 550 550 3 3.550 5.000V 760 760 3 3.760 5.000V

6.0 SERVICING AND MAINTENANCE GAUGE TUBE SERVICING In many cases, a gauge tube may become fouled with oil or other foreign matter. It is often possible to restore the functionality of contaminated probes with cleaning. If the contaminant is known, the tube should be filled with a fluid that is known to be a solvent to that contaminant. As an example, Acetone is often effective in removing residues of some oils. Commercial carburetor cleaners are very powerful solvents and are highly effective against some contaminants. After cleaning with solvents, the gauge tube should be completely dried or flushed with a volatile solvent to assure that it is dry prior to re-installing it. If this is not done, it may result in contamination of the system. MAINTENANCE Your vacuum instrument should give you many years of trouble free service. There are no regularly scheduled maintenance intervals. If consistent accuracy is required, it is recommended that the gauge, tube and power supple be returned for a yearly calibration check. 6.1 FACTORY REPAIR AND CALIBRATION The vacuum gauge assembly is designed to provide years of trouble-free service, and the liberal internal use of plug-in components make it easily repairable. No field servicing of the unit is recommended, other than replacement of the gauge tube, but factory servicing and calibration are available at a nominal cost and fast turn-around times. 6.2 FIELD CALIBRATION Each DigiVac vacuum gauge controller is calibrated to the particular vacuum gauge sensor that is shipped with the unit. While changing the gauge tube is possible, it will result in a slightly different reading as all

gauge tubes are not created equal. Although it is preferable that all calibration be performed at DigiVac, field calibration can be accomplished. Before re-calibrating the instrument, it should be ascertained that the instrument is in fact incorrect. In many cases, the problem will be with a tube that is fouled, or a system that is operating improperly. It is recommended that a spare tube be kept on hand and stored in a clean, dry place. Then, in cases of suspect readings, the tube should be changed before proceeding further. Calibration: The unit arrives to the factory calibrated, and re-calibrations are best performed by the DigiVac Company. User calibrations are possible, however using the following procedure: Remove the four screws holding the top housing in place and carefully remove the top and bottom exposing the printed circuit board. High End: At atmosphere, adjust the potentiometer adjacent to the INA 122 and set it so that pin 2 of the microprocessor, (same as Pin 6 of the INA 122) is at 4.25 VDC at atmospheric pressure. Low End: Locate the markings on the printed circuit board adjacent to pins 26, 27, and 28 of the microprocessor. It reads +Z - Connect the serial output of the unit to a PC set at 9600 baud Observe the readings on the PC screen Operate the unit at the lowest attainable vacuum value. (Must be 100 mtorr or lower) Short pin 27 ( Z ) to pin 28 ( + ) to increase the reading Short pin 27 ( Z ) to pin 26 ( - ) to decrease the reading 7.0 NOTES ON CALIBRATION The instrument is calibrated in nitrogen, which has thermal properties virtually identical to air. Other gasses will affect the readings by an amount

proportional to the thermal conductivity of the gases. In most cases, the gases present in a vacuum system will be air, nitrogen, or oxygen, and no appreciable errors will occur. Certain other gases, however, have thermal conductivity significantly greater than air and will cause the instrument to read higher than the actual amount of pressure. Examples of such gasses are water vapor, fluorocarbon refrigerants, and acetone. Conversely, other gasses have thermal conductivity significantly lower than air and will cause the instrument to read lower than actual pressure. Examples of such gasses include helium, oxygen and to a lesser extent, CO2. When interpreting readings using gasses other than air, it should be borne in mind that the DIGIVAC Model 22W reads Torr, which is a measure of absolute pressure - that is the opposite of vacuum. Thus, a lower numerical reading actually is a higher level of vacuum. For more information, refer to section 8.0. When in doubt, consult DigiVac. 8.0 UNDERSTANDING TORR The DIGIVAC vacuum instrument and many similar instruments are calibrated in microns or "millitorr." It is appropriate to discuss what microns are and to relate microns to other measures of pressure and vacuum. Microns are not really a measure of vacuum at all, but rather of absolute pressure. It will be recalled that the pressure of the atmosphere is 14.696 or approximately 14.7 pounds per square inch at sea level. This pressure is due to the weight of all of the air in the earth's atmosphere above any particular square inch. This 14.696 psi is equivalent to the pressure produced by a mercury column of approximately 29.92 inches high or.76 meters (about 3/4 of a yard) or 760 millimeters of mercury. Atmospheric pressure varies greatly with altitude. It decreases approximately 1 inch of mercury per thousand feet of altitude. It also varies widely with local weather conditions. (Variations of one half inch in a single day are common.)

The word vacuum means pressure lower than atmospheric or "suction," but, in describing negative pressure, the atmosphere is only a satisfactory reference if we are dealing with values of vacuum down to about 27 inches of mercury. Below that, it is much more useful to talk in terms of absolute pressure, starting from absolute zero. The Model 22W measures from 1 micron (1x10-3 Torr) all the way up to 760,000 Microns (760 Torr) or Atmosphere. Since this gauge measures through the full range of rough vacuum, you can easily pinpoint either gross or fine process errors or leaks. One TORR, a commonly used unit, is an absolute pressure of one millimeter of mercury. A millitorr is equal to one thousandth of a TORR. A MICRON is the same as a millitorr. 9.0 Accessories and Modifications Please consult the product guide and website for the latest available 10.0 SPECIFICATIONS Range.001-760 Units Torr or mbar Vac Interface 1/8 inch MNPT or KF/NW Sensor Varian 531 or 536 Display 16 digit character display Dimensions 1.6 x 1.8 x 2.4 Analog Output 0-5 VDC Power 100-240VAC 50/60Hz CE rated

INSTRUMENT ACCURACY.001 to.010 Torr +/-.001 Torr.010 to 2.00 Torr +/- 10% of reading 2.0 to 160 Torr +/- 50% of reading 160 to 760 Torr +/- 25% of reading For repair or recalibration, return gauges to: The DigiVac Company 105B Church Street Matawan, NJ 07747 Ph: 732.765.0900 Fax: 732.765.1800 E-mail: Direct from our website www.digivac.com The DigiVac Company manufactures a complete line of vacuum gauges and process computers. Contact us or your distributor if you wish for further information. See www.digivac.com for our latest offerings

11.0 Terms and Conditions TERMS OF USE, LIMITED WARRANTY & LIABILITY WAIVER THE DIGIVAC COMPANY ( DigiVac ) offers all of its products with the following terms and conditions and notices as follows. By accepting and/or using a DigiVac product, you hereby acknowledge and agree to the following terms and conditions, and acceptance of these terms and conditions are a condition precedent to any purchase/sale agreement between you and DigiVac. Exclusive Obligation: The DigiVac product you are purchasing has been designed for a specific use within a set of suitable operating conditions, as set forth in its User Manual, or as indicated otherwise by DigiVac. Any use of the DigiVac Product for any purpose or under any conditions, other than those specified, shall render any limited warranty void, and shall expressly invalidate any liability of DigiVac for damages as a result of such misuse. User limitation: You may not modify, copy, distribute, transmit, display, perform, reproduce, publish, license, create derivative works from, transfer, or sell, any information, software, products or services obtained from or created by DigiVac to any third party, without the express written consent of DigiVac to do otherwise. Any violation of this provision shall give rise to an indemnification of DigiVac by you, for any third party claims arising out of such violation. THIRTY (30) DAY LIMITED Warranty: All DigiVac products are warranted against any manufactured defect for a period of thirty (30) days from date of purchase, unless such product is a custom-work for you and not a standard DigiVac product. Any product qualifying as a custom-work shall not be warranted against any defects for any purpose, and your acceptance of such custom-work shall relieve DigiVac of any liability for any purpose. WITH THE EXCEPTION OF THE LIMITED WARRRANTY ABOVE, YOU AGREE ANY DIGIVAC PRODUCT IS PROVIDED AS IS, EXCLUSIVE OF ANY WARRANTY, INCLUDING, WITHOUT LIMITATION, ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE, NON-INFRINGEMENT, OR ANY OTHER WARRANTY, EXPRESSED OR IMPLIED. Limitation Of Liability: You agree and acknowledge, DigiVac shall have no liability to you whatsoever for any direct, indirect, punitive, incidental, special consequential damages arising out of or connected with the use or misuse of its products. In particular, given the nature of DigiVac products, you agree and acknowledge, under no circumstances whatsoever shall DigiVac be liable to you for any consequential damages for damage to any non-digivac product or service, arising from the failure, use or misuse of a DigiVac product, including, but not limited to, any vacuum system, engine, vehicle, factory, or the like. In the event, a court of law with proper jurisdiction finds DigiVac liable to you for any purpose, you agree and acknowledge DigiVac s maximum liability shall not exceed the purchase price of one unit of product giving rise to such liability, or $250.00, whichever is greater. Entire Obligation: These terms and conditions express the entire obligation of DigiVac with respect to its products. If any part of these terms and conditions are deemed void, invalid, unenforceable or illegal, including, but not limited to, the warranty disclaimers, liability disclaimers and liability limitations set forth above, then the unenforceable clause or sentence may be disregarded with the remainder of these terms and conditions valid and enforced. In the event the unenforceable clause or sentence leaves a void in these terms and conditions, a provision closely matching the intent of the unenforceable provision should be deemed inherent within these terms and conditions, slightly modified to render such provision valid and enforceable. General: These terms and conditions are governed by the laws of the State of New Jersey, USA. You hereby consent to the exclusive jurisdiction and venue of the Courts of New Jersey, in all disputes arising out of or relating to the use of this product. Use of this product is unauthorized in any jurisdiction that does not give effect to all provisions of these terms and conditions. Modification of Terms and Conditions: DigiVac reserves the right to change the terms, conditions, and notices under which their products are offered.