SIKA Ba_ODWT15_en 05/2018. Operating manual for oil deadweight tester 0.015% accuracy ODWT15

Similar documents
COP700 / COP1400 Operating manual for comparison test pump COP700 / COP1400

COP700 / COP1400 Operating manual for comparison test pump COP700 / COP1400

Pressure calibration equipment

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

Deadweight Tester / Pressure Balance High pressure up to 5,000 bar / 70,000 psi CPB5000-HP. Applications. Special features. Description. Rel.

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

Hydraulic pressure balance Model CPB5800

P3000 Series. Pneumatic Deadweight Testers Model P3000. Technical Data. Features

Dead-weight tester High-pressure version Model CPB3800HP

Pneumatic dead-weight tester Model CPB3500

Guide for the uncertainty analysis in pressure when using P3000 Series Deadweight Testers

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

FPG8601 Force Balanced Piston Gauge

Organisation Internationale de Métrologie Légale

BT 800-OX Operating Manual

Primary standard differential pressure balance Model CPB6000DP

Vibration isolation system 1VIS10W. User manual

Operating manual for the pressure transducer unit CCS 1000

ACCURACY, PERFORMANCE, AND HANDLING OF OIL-FILLED DIGIQUARTZ PRESSURE INSTRUMENTATION

Operating Procedure TITON SUBSEA BOLT TENSIONER

Hydraulic hand spindle pump Models CPP1000-M, CPP1000-L

Assembly-, installation- and maintenance instruction manual for bladder accumulators IBV / EBV , top reparable

P3800 Series. Users Manual. High Pressure Hydraulic Deadweight Tester

Hydraulic comparison test pump Models CPP1000-X, CPP1600-X

Beamex PG PRESSURE GENERATORS. Beamex PGM PGV PGC PGHH PGPH PGL

Mounting and Operating Instructions EB 2558 EN. Self-operated Pressure Regulators. Type Pressure Build-up Regulator

Air Operated Hydraulic Pumping Systems to 50,000 psi

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

A New Piston Gauge to Improve the Definition of High Gas Pressure and to Facilitate the Gas to Oil Transition in a Pressure Calibration Chain

USER INSTRUCTIONS. NAF Duball DL Ball Valves. Installation Operation Maintenance. Experience In Motion. flowserve.com

Hi-Force Limited Prospect Way Daventry Northants NN11 8PL United Kingdom Tel: +44(0) : Fax: +44(0) : Website:

PG7000 high line differential mode for defining differential pressure at line pressures above atmospheric pressure (PG7000 V2.

VALVCHEQ BACKFLOW PREVENTERS FIGURE RP03

Components for air preparation and pressure adjustment. OUT port position ( ) connected Rear side. of IN port. Air tank. directly.

THE BP-301 SERIES. Operating and Service Manual. Series includes all variants of BP-301 (LF 0.1Cv / MF 0.5Cv)

TECHNICAL MANUAL CALIBRATION PROCEDURE FOR PRESSURE GAUGE. This publication replaces T.O. 33K dated 30 January 2012.

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Reduce pressure zone device suitable for high and medium hazard rated applications BSP screwed connections

Inflatable Packer Single & Double. Single & Double Packer Dimension. Wireline Packer. Water Testing Packer (WTP) Packer

VPPL VARIABLE DISPLACEMENT AXIAL-PISTON PUMPS FOR INTERMEDIATE PRESSURE SERIES 10

User Manual Pressure Calibration System C and Hydraulic Pressure Pump Models T-620/T-620H Copyright 2008 AMETEK Denmark A/S

HYDRAULIC MOBILE LIFTING TABLE

Pressure calibration. 20

VACUUM REGULATORS CONTENTS

Level MEASUREMENT 1/2016

VLI Visible Level Interface Operation & Maintenance Manual

UIB30 and UIB30H Sealed Inverted Bucket Steam Traps for use with Pipeline Connectors

PNEUMATIC PRESSURE CONTROLLERS

INSTRUCTION MANUAL LOW PRESSURE CALIBRATION PUMP MODEL PGL

Coriolis Mass Flow Meter

Gripping rotary modules

INSTALLATION GUIDE MODEL NUMBER X. UniMaxx-Plus-SC-500-AC-V3 UNIMAXX SOLAR PUMP STATION

THE BP-690 SERIES. Operating and Service Manual. Series includes all variants of BP-LF/MF-690/691

FILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS

Technical Information Conducal CLY421

Instruction Manual. AlfaLavalSBAntiVacuumValve ESE02960-EN Original manual

for Building facilities Industrial facilities etc.,multipurpose Pilot operated type(high capacity)

SPG. Application example. Pneumatic 2-Finger Parallel Gripper Heavy-load Gripper. Weight 35 kg. Stroke per finger 100 mm.

Pressure Calibration Solutions

INSTALLATION & MAINTENANCE INSTRUCTIONS - ORIGINAL VERSION AVK KNIFE GATE VALVES

AIR/OVER HYDRAULIC JACK 20 TON

Ball valve HKSF-W100. Ball valve HKSF-W100. RMA Kehl GmbH & Co. KG Oststrasse 17 D Kehl / Germany

Contents. Overview Setup Safety Instructions Operating Instructions Connection Diagram Specifications...

TEST FOR STABILOMETER VALUE OF BITUMINOUS MIXTURES

VALFONTA INSTRUCTIONS: OPERATION AND INSTALLATION PRESSURE REDUCING VALVE MODEL PRV44

STAINLESS STEEL LOW PRESSURE REDUCING VALVE LPRV ELITE

WARNING: READ THE GENERAL INFORMATION MANUAL INCLUDED FOR OPERATING AND SAFETY PRECAUTIONS AND OTHER IMPORTANT INFORMATION.

P3000 Deadweight Tester Setup, Part 1: Limited Partial Correction Method

Hydraulic hand test pump Models CPP700-H and CPP1000-H

RS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.

Establishing Primary Gas Pressure Standard facility (0.140 to 70) bar Gas Piston Gauge at FCRI and its applications

Differential Pressure Regulator Type Type 45-6 (0.1 to 1 bar, DN 15) Mounting and Operating Instructions EB 3226 EN

Operation and Maintenance Instructions

PRESSURE COMPARATOR 10000PSI 700bar. Operator Instructions CRYSTAL. engineering corporation

Anchor Load Cells. Among the different types that are used as anchor load cells

Discontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves.

Flow control valves Series RFU - RFO 2/

Gripping modules Pneumatic 2-finger parallel gripper Long-stroke gripper for small components Gripping force 120 N..

P5523. Users Manual. Liquid to Gas Separator. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

A3S Bellows Sealed Stop Valve Installation and Maintenance Instructions

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

BSA6T and BSA64T Stainless Steel Bellows Sealed Stop Valves Installation and Maintenance Instructions

APP pumps APP and APP Disassembling and assembling

WIKA INSTRUMENT CORPORATION

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF PRESSURE METROLOGY

PO Box 370 Phone: Swansea NSW 2281 Fax: AUSTRALIA Mobile: Web:

STAINLESS STEEL OVERFLOW VALVE TYPE 417

Mounting and Operating Instructions EB 3007 EN. Self-operated Pressure Regulators. Differential Pressure Regulators (opening) Type Type 42-25

E-DWT-H Electronic Deadweight Tester

Product Information. Gripper for small components MPC

Test pumps. Pressure sources. Good reasons for proper and reliable calibration

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Technical Data Sheet TI-F50 Locking Units series KFH

MAINTANCE MANUAL CNC ROTARY TABLE MODEL: DMNC-5C YUASA INTERNATIONAL.

Standard Operating and Maintenance Instructions for Pumping System Model PS-150

256 Pneumatic Pressure Indicator

Digital deadweight tester Model CPD8500

Instructions. Control Solution ICF 20-4, ICF 25-4, ICF ICF 40-4, ICF 20-6, ICF 25-6, ICF 32-6, ICF 40-6, Direction and position.

Transcription:

SIKA Ba_ODWT15_en 05/2018 Operating manual for oil deadweight tester 0.015% accuracy ODWT15

Table of Contents 1 introduction...1 1.1 general product description... 1 1.2 operating principle... 1 2 general specifications... 2 2.1 hydraulic scheme...2 2.2 instrument outline... 3 2.3 vent valve...3 3 installation... 4 3.1 site requirements... 4 3.2 filling up...4 4 operating instruction...5 4.1 working instruction... 5 4.2 determine of the pressure... 6 5 maintenance...7 5.1 changing the liquid...7 5.2 recalibration... 7 6 parts list...8

1 introduction 1.1 general product description The model ODWT15 Pressure Standard is an oil operated deadweight tester used for calibrating test gauges, transducers and transmitters. The system consists of the following main components : deadweight tester platform 1 piston cylinder assembly Tungsten carbide, built in mass set 1 set pressure control manually oil operated variable volume The oil lubricated / oil operated pistons of the ODWT15 series are specially designed to have superior performance and a high range ability. 1.2 operating principle The primary function of the deadweight tester is to combine two primary metro logical quantities: 1. the piston-cylinder which defines an effective area, A. 2. the masses, value m, which press on the piston with a force F. force (F) effective area (A) pressure (p) page 1

2 general specifications measurement uncertainty p* 1.5 10-4 p e (of reading) certification standard delivered with company certificate (EA 1 traceable) pressure connection 3/8 BSP LH with BSP adapters footprint base plate 340 (w) x 225 (d) mm maximum size 340 (w) x 430 (d) mm overall height (excl. weights) 220 mm overall height (incl. weights) Max. 450 (depending on range) mm pressure media oil DWT oil Miglyol 812N Sasol * when using the formula of chapter 4.2. 2.1 hydraulic scheme 1 European Accreditation, see http://www.european-accreditation.org page 2

2.2 instrument outline weight column reservoir valve piston cylinder assembly pressure connection bubble level leveling nut (4x) screw pump 2.3 vent valve Only open the valve when the pressure is below 5 bar. page 3

3 installation 3.1 site requirements The ODWT15 is delivered with a mounting plate which is designed to contain the platform and variable volume pump. BSP adapters are also included in the scope of delivery. The room in which the instrument is placed should have proper founding, no vibrations are allowed during operation of the ODWT15 as this results in unpredictable errors. The ODWT15 should be placed on a rugged table which is rated for at least 140 kg without deforming. The table should be horizontally leveled. The ODWT15 can be places on the table without fixation. Or can be fit the a table using the extra set of leveling nuts. Use the stencil to mark the screw holes. Air movement around the ODWT15 should be avoided. Room temperature needs to be stable between 18 22 C during the time the ODWT15 is used to avoid uncertainties due to the thermal expansion coefficients of the piston cylinder and adiabatic effects in the measuring system. 3.2 filling up Before connecting any external device it is good practice to fill up the ODWT15 first. make sure there is sufficient liquid (Miglyol 812N oil) in the reservoir 1. open the valve (clockwise, till its mechanical stop) 2. rotate the screw pump clockwise until it hits its end stop 3. block the pressure connection with a plug or a finger 4. rotate the screw pump anti clockwise until it hits its end stop (the pump is filled with liquid) 5. close the valve (anti clockwise, max 2 turns) 6. unplug the pressure connection 7. carefully operate the screw pump until the liquid level is at the sealing of the pressure connection 8. mount the pressure instrument (instrument under test) 9. carefully operate the screw pump (clockwise) up to maximum 5 bar 10. carefully open the valve, the pressure will be reduced and air will be pressed out 11. repeat step 10 and 11 until no air bubbles escape when the valve is opened 12. the deadweight tester is ready for use Mount the table as shown below. It is not necessary to fasten it, it will tighten itself by the weights put on later. page 4

4 operating instruction 4.1 working instruction 1. before using the deadweight tester, check oil reservoir and if necessary fill up with Miglyol 812Noil (see chapter 3.2 for filling up) 2. place the instrument on a firm level base 3. check if the bubble level is in the center, if not: level the instrument with the leveling nuts 4. the instrument under test may be connected to the tester, using one of the adapters supplied 5. the correct number of weights to give the pressure desired may be placed on the piston cylinder 6. open isolation valve (clockwise, till its mechanical stop) 7. turn screw pump clockwise till end of its stroke 8. turn screw pump contra-clockwise till end of its stroke 9. close isolation valve (anti-clockwise, max. 2 turns) 10.turn screw pump slowly clockwise until the carrying table floats about 3mm higher than the pillar 11. rotate the weights and piston cylinder assembly clockwise 12.collect the readout of the instrument under test together with the mass load of the deadweight tester and file them 13. carefully change the mass load to realize the next pressure step 14. continue with 10 15.after the test, turn the screw pump contra clockwise until the piston has reached its end stroke and open the isolation valve 16. remove the weights and remove the pressure gauge Piston in lower position Piston starting to float Piston in highest position page 5

4.2 determine of the pressure The reference level of the deadweight tester is at the top of the sealing of the pressure connection. If the weight set is marked with a pressure unit (bar; psi or kg/cm²) the pressure can be determined by the formula: p e = ( p 0 + Σp c ) g 1 g n where p e * : Gauge pressure at reference level [bar; psi; kg/cm²] p 0 : Starting pressure piston + plate [bar; psi; kg/cm²] Σp c : Summarized pressure equivalent of the [bar; psi; kg/cm²] weights g 1 : Local gravity [N/kg] g n : Normal gravity (9.80665) [N/kg] * The value of the pressure p e which puts the piston into equilibrium. When using this formula the maximum error will not exceed 0.03% of reading. To reach the maximum accuracy of 0.015% of reading the following formula has to be used: p e = ( m c (1- ρ a / ρ m ) g l + π Τ d A 0 (1+ p) (1+(α p+α c) (t-20)) - ρ oil g l h ) 10-6 [MPa] p e * : gauge pressure at reference level [MPa] m c : conventional mass [kg] 1- ρ a / ρ m : air buoyancy correction (= 0,99985) [-] g l : local gravity [N/kg] A 0 : effective area at zero pressure [m 2 ] Τ : surfase tension of Miglyol 812N oil 0.031 [N/m] d : diameter piston low: 140 bar (2000 psi) : 0.0071 medium: 150 bar (3000 psi) ; 700 bar (10000 psi) : 0.0032 [m] high: 700 bar (10000psi) : 0.00225 : pressure distortion coefficient piston + cylinder [MPa -1 ] p : nominal line pressure [MPa α p+α c : thermal expansion coefficient piston + cylinder (= 9.0.10-6 ) [1/ C] t : piston temperature [ C] ρ oil : specific weight Miglyol 812N 950 [kg/m 2 ] h : height difference between the seal of the pressure connection (reference level) and the position of the bottom of the piston in mid floating situation [m] * The value of the pressure p e which puts the piston into equilibrium. The conventional masses m c, the coefficient, the effective area A 0 and the height difference between the seal of the pressure connection (reference level) and the position of the bottom of the piston in mid floating situation are listed in the calibration certificate. page 6

5 maintenance The entire tester should be kept clean as dirt and grit will cause rapid wear. As instruments to be tested come from an unknown process it is likely that the inside of the instrument is contaminated, this will in time contaminate the inside of the deadweight tester. To guarantee the performance of the deadweight tester it is good practice to change the Miglyol 812N oil once a year (dependent on the use and local conditions). In case of poor running performance it is possible to clean the piston and guidance piston. Disassemble the bronze cylindrical part. When taking off the bronze part take hold off the guidance piston else it could fall. Clean the measuring piston/cylinder, bearings, guidance piston and bushing with Loc-tite 7063. Grease the piston and guidance piston with a little Miglyol 812N oil. Mount the cylinder and close the oil reservoir valve and turn the hand pump clockwise until oil comes out of the cylinder. Put the measuring piston in the cylinder and move the piston up an down several times to spread the oil. Assemble the bronze part with the guidance piston and move the piston up an down several times to spread the oil. Mount the table as shown in chapter 3.2 5.1 changing the liquid 1. open the valve 2. rotate the screw pump clockwise until it hits its end stop 3. take out the liquid 4. blow through the pressure connection with low pressure air 5. clean the reservoir 6. see chapter 3.2 purging 5.2 recalibration Although the ODWT15 is designed to have a very good long term stability, a first recalibration at 2 years after purchase is recommended both for piston cylinder and mass set. The results of the recalibration can be used as a guideline for future recalibration. Depending on the environment and frequency of use a recalibration interval of 2 to 5 years is normal. page 7

6 parts list Part Code Qty. Remark Base plate 340 x 225 mm 1 Piston cylinder assembly 1 Including start weight Oil hand pump 1 16 mm piston U-cup Merkel T20 8-16-5.7 1 1 In piston of oil pump Back-up ring 041-959-01 PTFE 1 In piston of oil pump for U-cup Multi seal under cylinder Ø10,35x16 x2 1 Mounting of cylinder Bearings (degreased)* Ø21xØ12 x 5 2 For guidance piston Bearing Ø9.5 x Ø5.2 x 1.2 1 Under measuring piston Oil reservoir and valve 1 O-ring in valve Ø5.23x2.62 mm 90 shore 1 Leveling nut M10 x 55 mm 4 For base plate Fixed level nut M10 x 55 mm 4 For base plate O-ring Ø23 x 2,4 mm 90 shore NBR 1 For mounting oil reservoir O-ring Ø15,88 x 2,62 mm 90 shore NBR 1 For mounting hand pump O-ring Ø18,6 x 2,4 mm 90 shore NBR 3 For mounting pressure connections Adapter 1/2 bsp 027-411-26 1 Adapter 3/8 bsp 027-411-14 1 Adapter 1/4 bsp 027-411-25 1 Adapter 1/8 bsp 027-411-15 1 Multi seal 065-260-09 2 Ø13,2 x 6,9 x 1,3 mm, for adapters Certificate 1 Traceble to European standard Weight box 1...3 for weight set Deadweight tester oil Miglyol 812N 0.5 ltr. Sasol * only for mid and high piston-cilinder. page 8