Products Designed to Solve Nagging GC Problems
|
|
- Ashlie Bryan
- 5 years ago
- Views:
Transcription
1 Products Designed to Solve Nagging GC Problems Abby Folk Chromatography Technical Specialist Page 1
2 Contaminated Gas Fouls the Whole GC Dirty Lines, Inlet, Column, Detector More Frequent Maintenance More Down Time Replacements Cost Money Solution = Use only Clean Gas! Page 2
3 Gas Clean Filters The Insurance Policy Page 3
4 Gas-Clean Filters Design Sensitive indicators for moistureand oxygen Note where indicators change Color! Top or Bottom Short filter equilibration time Fast, leak-free filter replacement without turning off the gas NO tools needed Page 4
5 Oxygen Damages GC columns Repair of the GC column is impossible Replacements means re-calibration Replacement means delay Replacement column costs money Damage to electron multiplier, filaments Solution = No Leaks! Page 5
6 Leak Detector G3388B Overview
7 Key Features The full list of features is included below: Detects thermal conductivity differences Audible and visual alerts Minimum detection limit ml/min hydrogen and helium Lightweight only 105 g Small roughly 50 (W) 19.5(D) 111(H) mm Rechargeable lithium ion battery with over 5 hours of life Includes a USB cable for battery charging One year warranty from Agilent
8 How do you make better GC column connections? 1. Start with the right tools and supplies Column nuts Material : stainless but need brass for MS interface Choice of right ferrules 2. Proper assembly process Make a good clean column cut every time After ferrule is installed Fused silica tubing cutters : ceramic, diamond tipped etc. Magnifier to inspect the cut - cracked fused silica and flaps of polyimide are active sites that ruin chromatography. 3. Ensure the proper and consistent length of column into the fitting 8
9 Supply Selection: Which Capillary Ferrules to use? polyimide polyimide/graphite graphite flexible metal Composition Re-use Max T Use Limitation Polyimide (Vespel) yes 280 Easy seal Shrink after heating causing leaks after thermal cycle; isothermal only Graphite yes 450 FID, NPD Inlets Contamination, permeable to air not for oxygen sensitive detectors Polyimide/graphite (85% / 15%) limited 350 MS, ECD, Inlets Still shrink after thermal cycles creating leaks; need to retighten regularly Flexible metal no 450 Capillary Flow Technology (backflush, splitters, ) May not seal well with damaged fittings or rough surfaces Short ferrules for inlet and detector configurations on Agilent GC s Long ferrules for MS transfer lines and MS interface nut 9
10 Ensure the proper length Column Installation Tools Follow the manufacturer s recommended procedure Optimized GC performance and reproducibility requires ensuring the proper length of column into the fittings, every time Column Installation pre-swaging tools for the Split/Splitless, Multi Mode and Purge Packed Inlets available for Agilent GC models 7890/7820 and 6890/6850, for graphite or metal ferrules G Graphite Ferrules Metal Ferrules G G MS Install Tool 10
11 Graphite Ferrules Have a Down Side Too! Pros High temperature range (450C) Low Cost Soft, easily conforms Con Can flake, fall apart or extrude Permeable Active sites in the flow path Extensive inlet maintenance needed NOT recommended with MS or ECD Page 11
12 Supply Selection: Graphite/polyimide blend ferrules Graphite / polyimide (vespel) blend ferrules are very popular Recommend Long style for Mass Spec Short for Inlets, and other oxygen sensitive detectors, like ECD Soft enough to make seal against rough surfaces Fairly high temperature stability (350C) Don t extrude beyond the fitting when tightened (like graphite) Lower air permeability than pure graphite 12
13 Flexi-Metal Ferrules: Touchless Dial Packaging o Easy column insertion o No lost ferrules o No contamination risks 13
14 Capillary column nuts Column nuts are determined by the instrument fitting Basic mechanical fittings with little enhancements Brass for the MS Some finger tight designs for ease of use New design addresses ease of use and productivity issues of leaking and over tightening 14
15 What if you Could install the column easily Finger tighten, without tools that cause over tightening And make leak free connections Which don t need to be retightened Every time?
16 Self-Tightening Column Nuts Reliable performance: Innovative spring-driven piston continuously presses against ferrule maintaining a leak-free seal Less wasted time: No retightening needed after repeated thermal cycles Ease of use: Finger-tight, consistent connections without tools Faster maintenance: Low-torque seal prevents sticking or crumbling during removal Leak Free = Lower column bleed: Longer column life Video at agilent.com/chem/stnutvideo Page 16
17 Self Tightening Column Nuts For inlet or detector p/n For mass spec transfer line p/n Short ferrule exposes more thread of the fitting for better sealing also no need for a short and long ferrule Page 17
18 Self-Tightening Column Nuts Page 18
19 The BIGGEST Problem in GC is There are more things that DON T go through a GC than DO!.therefore, don t inject anything and you ll never have problems. OK, inject, but realize that everything just got dirty deal with it! Page 19
20 Where Does it Get Dirty? Here Here Here Here Here Here Page 20
21 What Are You Doing!? Page 21
22 It s in the Inlet! Do Inlet Maintenance - Trim the Column - Replace the Liner - Replace the Gold Seal - Replace the Syringe - Replace the Split Vent Trap, Split Vent Line - Rinse the Inlet Do not aggressively scrub the inlet!! Start as clean and inert as possible Page 22
23 Bonus Peaks Where do they come from? CONTAMINATION in INJECTOR or FLOW (carrier gas) -Contaminated consumables -Carryover from a backflash or previous sample -Bad tank of gas or traps have expired -Septum bleed *TIP = Run a blank run it should be blank! Page 23
24 Bonus Peaks - Ferrule Contamination pa 28 Vespel/Graphite Ferrule from a bag Agilent Page 24
25 Bonus Peaks - O-Ring Contamination ALLTech O-Ring outgases pa Agilent coated O-Ring ougases Competitor A Head Space Analysis Vial equilibrium time: 60 minutes Vial equilibrium temperature: 150 C pa min Competitor R Agilent s Non-Stick O-ring min Page 25
26 Off-Brand O-Ring Issues Controlled Substances Analysis, H2 Carrier Residue on top of inlet weldment Problem Resolution: Agilent Non-Stick Liner O-Ring p/n , 10PK Page 26
27 Peak Tailing What causes this? 1. ACTIVITY either in inlet or column -Reversible adsorption of active compounds (-OH, -NH, -SH) 2. FLOW PATH PROBLEM - dead volume, obstruction, poor installation, or severe column contamination Miscellaneous temperature issues for late eluters, overloading of PLOT columns, co-elution, polarity mismatch between phase, solute or solvent, and some compounds always tail *Tip = Inject a light hydrocarbon, should not tail unless flow path problem. Page 27
28 Symptom Tailing of Active Compounds Sample: 0.5 ng on column loading with ISTD Column: 20m 0.18mm 0.18µm Carrier: Helium 37cm/sec, Ramped flow; 0.7ml/min (0.1min) to 1.3ml/min (15ml/min 2 ) Oven: 35 o C (2.5 min) to 80 o C (40 o C/min), 15 o C/min to 200 o C, 8 o C/min to 275 o C (2 min) Injection: 0.5µl, Splitless, 280 o C, purge flow 30ml/min at 0.75 min MSD: Transfer Line 290 o C, Source 300 o C, Quad 180 o C n-nitrosodimethylamine 2. Aniline 3. 1,4-Dichlorobenzene-d4 4. Benzoic Acid 5. Naphthalene-d8 6. Acenaphthene-d ,4-Dinitrophenol 8. 4-Nitrophenol 9. 2-Me-4,6-dinitrophenol Aminobiphenyl 11. Pentachlorophenol 12. Phenanthrene-d Benzidine 14. Chrysene-d ,3 -Dichlorobenzidine 16. Benzo[b]fluoranthene 17. Benzo[k]fluoranthene 18. Perylene-d Abundance Time (min) Page 28
29 Solution Ultra Inert GC Column Sample: 0.5 ng on column loading with ISTD Column: 20m 0.18mm 0.18µm Carrier: Helium 37cm/sec, Ramped flow; 0.7ml/min (0.1min) to 1.3ml/min (15ml/min 2 ) Oven: 35 o C (2.5 min) to 80 o C (40 o C/min), 15 o C/min to 200 o C, 8 o C/min to 275 o C (2 min) Injection: 0.5µl, Splitless, 280 o C, purge flow 30ml/min at 0.75 min MSD: Transfer Line 290 o C, Source 300 o C, Quad 180 o C n-nitrosodimethylamine 2. Aniline 3. 1,4-Dichlorobenzene-d4 4. Benzoic Acid 5. Naphthalene-d8 6. Acenaphthene-d ,4-Dinitrophenol 8. 4-Nitrophenol 9. 2-Me-4,6-dinitrophenol Aminobiphenyl 11. Pentachlorophenol 12. Phenanthrene-d Benzidine 14. Chrysene-d ,3 -Dichlorobenzidine 16. Benzo[b]fluoranthene 17. Benzo[k]fluoranthene 18. Perylene-d Abundance Time (min) Page 29
30 AND Ultra Inert Liners Touchless packaging Easy installation of new, clean liner without risk of contamination from touching Includes non-stick plasma treated O-ring Page 30
31 More and More Touchless Packaging Page 31
32 Agilent UI Gold Seal: Deactivated Gold Surface Soft gold plating is essential for proper sealing Ultra Inert chemistry blocks active sites (gold is NOT inert) Smooth surface doesn't leak Agilent MIM seal Competitor s machined seal Reliable ppb and ppt measurements require attention to the little things! Page 32
33 New Product: Agilent Inert Inlet UltiMetal Plus treatment creates inert surface 7890 inlet weldment & shell Limits adsorption/degradation of active analytes Optimum for Trace GC/MS and GC-ECD pesticide analyses Agilent s proven proprietary Ultimetal Plus surface treatment Page 33
34 Putting It All Together Inert Flow Path vs. Standard Flow Path Drugs of Abuse 250 ppb using HP-5ms UI Green = Inert Flow Path Blue = Standard Flow Path 1. Dronabinol 2. Oxycodone 3. Temazepam 4. Flunitrazepam 5. Diacetyl morphine Column: Agilent J&W HP-5ms UI 30 m x 0.25 mm x 0.25 um Oven: 100 C 4 min hold, 10 /min to 280 C, 6 /min to 300 C (4.67 min hold), Carrier : Helium 52.7 cm/s (2 ml/min) set at 100 C, EPC-Constant Flow Inlet: Pulsed Splitless 35 PSI pulse until 0.73 min, 0.75 min purge 50 ml/min, gas saver 20, ml/min at 2 min Inlet liner: Ultra Inert with wool / Standard single taper liner with wool (p/n ) Gold Seal: UI Gold Seal / Standard gold seal Detector: MSD Scan mode 40 t0 450 m/z, 230 C source temp, 150 C Quad temp, 310 C transfer line Page 34
35 Broad Peaks Why? INJECTOR -Poor Installation -Change in injection solvent -Change in settings (temps/flows) -Poor sample focusing -Large change in sample concentration -Breakdown or adsorption FLOW -Change in gas velocity -Constant Flow vs Constant Pressure COLUMN -Contamination -Damaged/old stationary phase -Breakdown or adsorption -Reverse Solvent Effect Page 35
36 Splitless Injection Reverse Solvent Effect Hexane 5 6 Methanol Time (min.) Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 50 C for 1 min, C at 20 /min; Helium at 30 cm/sec 1. 1,3-DCP 2. 3-hexanol 3. butyl acetate 4. 1-heptanol 5. 3-octanone 6. 1,2-dichlorobenzene Page 36
37 Retention Gap Also Called A Guard Column Injector Detector Deactivated Fused Silica Tubing Union Column Usually 2-10 meters long and same diameter as the column (or larger if needed) Page 37
38 Fixing the Reverse Solvent Effect 3 m x 0.25 mm I.D. Retention Gap Hexane 1 Methanol Time (min.) Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 50 C for 1 min, C at 20 /min; Helium at 30 cm/sec 1. 1,3-DCP 2. 3-hexanol 3. butyl acetate 4. 1-heptanol 5. 3-octanone 6. 1,2-dichlorobenzene Page 38
39 Press Fit Connectors Competitor Press-Tight Agilent Ultra Inert Press Fit Inertness, seal and transparency are key parameters
40 Maximum Force Required to pull apart 2.5 Pull Value, 0.25 mm I.D. Columns Pounds (force) Strong bond between FS polyimide & press fit UI deactivation layer Ultra Inert Quartz Fused Silica Glass Press Fit Connector Type Page 40
41 Press Fit Connector Ultra Inert Universal Press Fit Union for 0.10 to 0.75mm, 10/pk Ultra Inert Universal Press Fit Y-Splitter Page 41
42 Blood Alcohol Application: Agilent Solution S/S Inlet HSS Transfer Line FID CH 1 FID CH 2 Column 1: DB- ALC1 0.4 m x 0.45 mm Retention Gap 7890 GC Un-Purged Splitter Column 2: DB-ALC2 7697A Headspace Sampler 42
43 Better Connections: Capillary Flow Technology Devices UltiMetal Plus Ultimate Union/UltiMetal Plus Tee Stainless Steel Micro Fluidic plates technology Deactivation essential to block active sites Column connection easy to assemble Release hole for stuck ferrules Using Flexible Metal ferrules to overcome issues UltiMetal Plus surface chemistry prevents activity Flexible design reduces risk of over tightening or column breaks Leak free seal remains after repeated temperature cycles 43
44 Flexi-Metal Ferrules: Touchless Dial Packaging o Easy column insertion o No lost ferrules o No contamination risks 44
45 Compare current vs new Agilent ferrule Current Ferrule Stiff, prone to cause damage to column or fitting Agilent Flexi-Metal Ferrule Computer aided Design Finite Element Analysis More flexible Small ID compression (40 to 60um) leaks Tolerances poorly controlled Less column damage Large range of ID compression (100 to 120um) less leaks issues
46 Impact of ferrule surface on inertness Very small amount of ferrule surface exposed to active pesticides x10 4 Untreated FMF Post column Std 500ppb 2 1 Acephate Omethoate Demeton-S 0 4 x10 2 FMF + UltiMetal Plus Post column Std 500ppb Acephate Omethoate Demeton-S Counts vs. Acquisition Time (min) 46
47 Conclusions for a Problem Free GC Start With Ultra Inert/High Quality Consumables Never Have a Leak (no oxygen) Use Gas Clean Filters Never Inject Anything (dirty or reactive) Don t Touch Anything Do Routine Inlet Maintenance Often Page 47
48 TECHNICAL SUPPORT , #3 Page 48
49 Troubleshooting Resources Online Video Maintenance Manual Troubleshooting Guide Publication G pdf Page 49
50 NOW IT S EASIER THAN EVER TO GET RELIABLE GC RESULTS THROUGH BETTER GC CONNECTIONS FROM AGILENT Order the poster View the video 50
Recent Advances to Ensure Simple, Leak Free GC Column Connections. Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division
Recent Advances to Ensure Simple, Leak Free GC Column Connections Ken Lynam R&D/Application Chemist Agilent Consumables and Supplies Division 1 GC column installation is a challenge A routine, basic procedure
More informationBest Practice for Identifying Leaks in GC and GC/MS Systems
Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,
More informationConsiderations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port
Considerations and Best Practices for Maintaining Your Agilent GC s Split/Splitless Injection Port So, Why Do I Have To Do Maintenance? 1. Things wear out septa, syringes, nuts, ferrules, o-rings, etc.
More informationTips and Tricks of Injector Maintenance. Abby Folk Agilent GC Technical Support August 2009
Tips and Tricks of Injector Maintenance Abby Folk Agilent GC Technical Support August 2009 Typical Gas Chromatographic System Mol-Sieve Traps Fixed Restrictors Have to be able to get things to their gas
More informationA Column-Flow Independent Configuration for QuickSwap. Application. Authors. Abstract. Introduction
A Column-Flow Independent Configuration for QuickSwap Application Authors Matthew S. Klee and Bruce Quimby Agilent Technologies, Inc. 80 Centerville Road Wilmington, DE 08 USA Abstract As can be seen from
More information6890 GC Site Preparation
6890 GC Site Preparation (a16011) This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information
More informationno peaks missing peaks too small peaks increasing retention decreasing retention
GC Troubleshooting We can help you. 1. Observation: no peaks no peaks missing peaks too small peaks increasing retention decreasing retention declining baseline rising baseline bleeding plateaus interfering
More informationImpact of Air Leaks on the Productivity of GC and GC/MS Systems
Impact of Air Leaks on the Productivity of GC and GC/MS Systems Application Note Environmental Author Ken Lynam Agilent Technologies, Inc. Abstract Agilent 7890A GC dual-channel FID, Agilent 7890B GC,
More informationSimplified Backflush Using Agilent 6890 GC Post Run Command Application Note
Simplified Backflush Using Agilent 6890 GC Post Run Command Application Note Gas Chromatography Author Matthew S. Klee Agilent Technologies 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract
More informationCarrier Gases in Capillary GC
Carrier Gases in Capillary GC LC Columns and Consumables Mark Sinnott Application Engineer January 15, 2009 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences
More informationSite Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)
Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 7820MSD Series Mass Selective Detector (MSD). This checklist is
More informationUsing the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer
application Note Gas Chromatography/ Mass Spectrometry Using the PreVent System in Time-Saver Mode with the AutoSystem XL Gas Chromatograph and the TurboMass Mass Spectrometer Overview A mass spectrometer
More informationGhost Peaks: How to Fix a Haunting Problem. Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager
Ghost Peaks: How to Fix a Haunting Problem Jacob A. Rebholz Teledyne Tekmar VOC Product Line Manager Outline of Topics What is a ghost peak? How to identify ghost peaks and their source How to correct
More informationAgilent G3180B Two-Way Splitter Kit
Agilent G3180B Two-Way Splitter Kit With Makeup Gas Installation and Operation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or
More informationGas Clean. Filters. Delivering Clean Gases for GC and GC/MS Operation
Gas Clean Filters Delivering Clean Gases for GC and GC/MS Operation Gas Clean Filters Fast, leak-free replacement without tools! Each Gas Clean Filter features a unique quick disconnect design which allows
More informationAgilent G Rebuilt Electron Capture Detector
Agilent G1533-90790 Rebuilt Electron Capture Detector User Information Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means
More informationEnvironmental Petrochemical, Pharmaceutical, Foods & Flavors, Chemicals, Personal Care, Forensics,...and more!
After years of development and testing, DPS Instruments is pleased to present the newest, most expandable and versatile Gas Chromatography Systems in history. The DPS 600 Series GC systems are the world
More informationCapillary Conversion System CCS-4/NC & CCS-05/NC. Installation and Operating Instructions. Operating Instructions
Operating Instructions Check the detector for the correct operating conditions. If make-up gas is being used a flow of 30ml/min. of carrier plus make-up is normally recommended. This can most effectively
More informationSite Preparation Specification for GCMS system. Dimensions and Weight. Agilent 5975C MSD Series G3170A, G3171A, G3172A, G3174A
Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 5975C Series Mass Selective Detector (MSD). This checklist is designed
More informationOMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT
OMCL Network of the Council of Europe QUALITY ASSURANCE DOCUMENT PA/PH/OMCL (16) 17 R QUALIFICATION OF EQUIPMENT ANNEX 2: QUALIFICATION OF GC EQUIPMENT Full document title and reference Document type Legislative
More informationCarrier Gases in Capillary GC
Carrier Gases in Capillary GC GC Columns and Consumables Abby Folk Application Scientist January 15, 2008 CARRIER GAS Mobile Phase Carries the solutes down the column Selection and velocity influences
More informationConverting Helium Carrier Gas GC Methods Nitrogen and Hydrogen
Converting Helium Carrier Gas GC Methods Nitrogen and Hydrogen Jim McCurry Agilent Technologies Wilmington, DE 18951 USA 1 Market Situation The world of He supply is not reliable, prices are increasing
More informationJaap de Zeeuw Restek Corporation, Middelburg, The Netherlands. Copyrights: Restek Corporation
Using Nitrogen instead of Helium as carrier gas, while maintaining exact the same separation efficiency, retention times, peak elution order, without changing the oven temperature conditions. Jaap de Zeeuw
More informationGas Chromatography. MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass Gold, and Clarus 500 GC/MS Systems
Inorganic Analysis Chromatography Molecular Spectroscopy Thermal/Elemental Analysis Informatics Gas Chromatography MS Vent Enhancement to the PreVent System to Optimize Operation with the TurboMass, TurboMass
More informationThe Open-Split Capillary Interface. Part No
The Open-Split Capillary Interface Part No. 113532 2 INTRODUCTION With the advent of fused silica capillary columns and their high degree of flexibility, it is obviously attractive to be able to take the
More informationGeneral Specifications: Forensic. Petrochemical Environmental. Lab Quality Analyses in the Field, It Goes with you Anywhere!
The portable Gas Chromatograph is now more versatile than ever. DPS Instruments is pleased to present the newest Portable Companion 2 Gas Chromatograph with room for 2 Detectors. The Companion 2 GC was
More informationGC Accessories. Agilent Consumables. Agilent Instrument Specific Replacement Parts. SGE Replacement Parts by Instrument Vendor.
GC Accessories Agilent Consumables GC and GC/MS Maintenance Schedule Bulk GC Supplies Inlet Septa Inlet Liners Specialty Injection Liners Gold Seals for Split/Splitless Inlets Liner O-Rings Capillary Column
More informationGCMSD-Headspace Analysis SOP
Before you start GCMSD-Headspace Analysis SOP Method and Sequences names are restricted to 50 characters (MSD program will crash otherwise) The GC oven and Injection Ports need to be cooled to 50 oc to
More informationOptimizing Vial Pressurization Parameters for the Analysis of <USP 467> Residual Solvents Using the 7697A Headspace Sampler
Optimizing Vial Pressurization Parameters for the Analysis of Residual Solvents Using the 7697A Headspace Sampler Application Note Pharmaceuticals Author Roger L Firor Agilent Technologies, Inc.
More informationMass Spec will not Autotune
Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems
More informationAdvances in Capillary Column Backflushing
Advances in Capillary Column Backflushing Matthew S. Klee, Ph.D. ISCCE 2010 1 Capillary Flow Technology Traits Ease of use - ruggedness Inertness Usability to > 350 oc Excellent exclusion of air No unpurged
More informationRetention Time Locking: Concepts and Applications. Application
Retention Time Locking: Concepts and Applications Application Gas Chromatography December 1997 Authors Vince Giarrocco Bruce Quimby Matthew Klee Agilent Technologies, Inc. 2850 Centerville Road Wilmington,
More informationCustomer Responsibilities. Important Customer Information GCMS QQQ Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationAgilent G3185B QuickSwap Accessory
Agilent G3185B QuickSwap Accessory Reference Manual Agilent Technologies Notices Agilent Technologies, Inc. 2005 No part of this manual may be reproduced in any form or by any means (including electronic
More informationLow-Pressure Retention Time Locking with the 7890A GC. Application. Authors. Abstract. Introduction HPI
Low-Pressure Retention Locking with the 789A GC Application HPI Authors Courtney Milner and Russell Kinghorn BST International 41 Greenaway Street Bulleen, VIC 315 Australia Matthew S. Klee Agilent Technologies,
More informationDetector Carrier Gas Comments Detector anode purge or reference gas. Electron Capture Nitrogen Maximum sensitivity Nitrogen Argon/Methane
Gas requirements Gases for packed columns The carrier gas you use depends upon the type of detector and the performance requirements. Table 520-1 lists gas recommendations for packed column use. In general,
More informationAgilent 7697A Headspace Sampler
Agilent 7697A Headspace Sampler Data Sheet Overview The Agilent 7697A Headspace Sampler is part of Agilent s line of gas chromatography products. The 7697A based on the architecture of the market-leading
More informationCP-3800 GC Getting Started Manual
CP-3800 GC Getting Started Manual Getting Started Installation Basic Operation Maintenance Parts and Supplies Varian, Inc. 2700 Mitchell Drive Walnut Creek, CA 94598-1675/usa Varian, Inc. 1999 Printed
More informationAdvancements in Gas Chromatography Analyzers - Keeping up with New Technology. Chuck Runkle Gas Phase Product Specialist ASTS June 2013
Advancements in Gas Chromatography Analyzers - Keeping up with New Technology Chuck Runkle Gas Phase Product Specialist ASTS June 0 Capillary Flow Technology -- solves difficult application problems easily
More informationZebron Gas Management
Zebron www.phenomenex.com/gasmanagement What are the key benefits of Zebron filters and traps? The high adsorption capacity of the Zebron Filters removes contaminants from mobile phase gases, reducing
More information3 Flow and Pressure Control
3 Hydrogen shutdown Column shutdown Turning gas flows on and off EPC-controlled streams NonEPC-controlled streams Electronic Pneumatic Control (EPC) Interpreting flow and pressure readings Configuration
More informationAgilent 7890 Series Gas Chromatograph
Agilent 7890 Series Gas Chromatograph Advanced Operation Manual Agilent Technologies Notices Agilent Technologies, Inc. 2007-2013 No part of this manual may be reproduced in any form or by any means (including
More informationms NoVent II Instruction Booklet
ms NoVent II Instruction Booklet ms NoVent II The ms NoVent II interface is designed to allow column changing or column maintenance to be carried out without the need to shut down the mass spectrometer.
More informationFor additional information about our solutions, please visit our web site at US/Pages/HomePage.
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationInstant Connect Helium Saver Injector Module
Thermo Scientific Instant Connect Helium Saver Injector Module for TRACE 1300/1310 Gas Chromatographs User Guide PN 31709737 Revision B January 2016 2016 Thermo Fisher Scientific Inc. All rights reserved.
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek. Abstract: Discussion:
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: The RSK-175 standard operating procedure was developed in order to determine the amount of dissolved gas in water. Due to the expansion
More informationAgilent Inlet EPC Leak Test
Agilent Inlet EPC Leak Test Gas Chromatographs Instructions Agilent Technologies Notices Agilent Technologies, Inc. 2002 No part of this manual may be reproduced in any form or by any means (including
More informationExperiment GC : Analysis of BTEX by GC-FID
46 Experiment GC : Analysis of BTEX by GC-FID Learning Goals: Familiarity with gas chromatography Gain experience in temperature programming and method development Correctly use an internal standard, and
More informationVarian 3300 Gas Chromatograph Abbreviated Operating Instructions
Varian 3300 Gas Chromatograph Abbreviated Operating Instructions 1. Introduction and Startup 2 2. Basic Configurations for the GC 3 3. Sample Preparation 3 4. Building a Method 3 5. Split/Splitless Injection
More information7820A GC Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationConnecting a SSL/PTV Backflush System
This section provides instructions for connecting your SSLBKF/PTVBKF injector modules with pre-column and analytical capillary column into the oven. Refer to the chapters SSLBKF Injector Module and PTVBKF
More informationAgilent ICP-MS Interactive Troubleshooting tool for Low sensitivity
Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Rev. B 2017,Janyary,20 Next > Preconditions 7900, 7800, 7700, or 8800 ICP-MS is used. Sample introduction type is PeriPump. If you are
More informationMarkes Thermal Desorption Systems Site Preparation Page 1 of 8
Thermal Desorption Systems Site Preparation Purpose of site preparation Your site must meet this specification to assure a successful and timely installation of your Agilent thermal desorption equipment.
More informationA New GC/MS System Designed for Helium Carrier Gas Conservation
A New GC/MS System Designed for Helium Carrier Gas Conservation Approaches for the Smarter Use of Helium Dale R. Walker GCMS Applications Specialist August, 7, 2013 1 Helium Facts How it's produced Where
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Engineered Solutions, Guaranteed Results.
Laboratory Hardware Custom Gas Chromatography Solutions Engineered Solutions, Guaranteed Results. WASSON - ECE INSTRUMENTATION Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions
More informationLaboratory Hardware. Custom Gas Chromatography Solutions WASSON - ECE INSTRUMENTATION. Custom solutions for your analytical needs.
Laboratory Hardware Custom Gas Chromatography Solutions Custom solutions for your analytical needs. Laboratory Hardware Wasson-ECE Instrumentation offers hardware-only solutions for advanced chromatography
More informationUWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column
Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps
More informationAgilent 7890 / 220 Ion Trap GCMS Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationAgilent 7697A Headspace Sampler
Agilent 7697A Headspace Sampler Site Preparation Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic
More informationMultiple Gas#5 GC configuration Jan 2016
History: Unfortunately there is no single column that can separate: Hydrogen Oxygen Nitrogen Methane CO CO2 Ethane Water Propane Butane Pentane Over the years SRI Instruments has devised several solutions
More information6850 Series Control Module User Information
6850 Series Control Module User Information Contents Introduction Important Information...4 Overview...4 Control Module elements...5 Navigating the screens...8 Configuring the instrument...12 To adjust
More informationDemystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards.
Demystifying the Process of Calibrating your Thermal Desorption Gas Chromatography System using Compressed Gas Standards. Jamie Brown R&D Scientist Supelco 595 North Harrison RD. Bellefonte, PA 16823 National
More informationTechnical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS)
Technical Procedure for Drug Chemistry Gas Chromatograph/Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent 6890 GC interfaced
More informationAgilent 6890N, Option 652 Oral Fluids Analyzer
Agilent 6890N, Option 652 Oral Fluids Analyzer Installation and Operation Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or by any means
More informationinstrumentation Pulsed Discharge Detectors Non-Radioactive, Multiple Mode Electron Capture / Helium Photoionization
Instrumentation Most of the components we supply to the instrumentation industry are from our valve and fitting lines. The rest, from our R&D 100 Award-winning pulsed discharge detectors to our application-dedicated
More informationPlease refer to the Hydrogen Safety Guide which is shipped with the Instrument. 5977B Series GC/MS Systems Site Preparation Checklist
Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation
More informationSuper-Clean Gas Filters from Restek
Super-Clean Gas Filters from Restek Chromatography Products www.restek.com 800-356-688 84-353-300 regulator instrument-grade tubing on/off valve GC carrier gas moisture trap hydrocarbon trap high-capacity
More informationSystem Overview. TCD Detector temperature setpoint Regulator
System Overview POPULAR CONFIGURATION GCs Your educational TCD GC is configured on the compact 310 chassis. It is equipped with a TCD Detector, a temperature programmable Column Oven, a 3 Silica Gel packed
More informationUSER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve
USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44
More informationUP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule
One Piece or Two? A one-piece fitting is more convenient and less cumbersome, since the ferrule cannot stick in a receiving port and the fitting is more easily found if dropped. With two-piece Fingertight,
More informationAutomated Determination of Dissolved Gases in Water Anne Jurek
Automated Determination of Dissolved Gases in Water Anne Jurek Abstract: Tapping the natural gas reservoirs throughout the United States has long been a viable solution for energy independence; however
More informationToxicology Gas Chromatography-Mass Spectrometry (GC-MS)
Toxicology Gas Chromatography-Mass Spectrometry (GC-MS) 1.0 Purpose - This procedure specifies the required elements for the calibration and use of the Agilent Gas Chromatograph interfaced to the Agilent
More informationUNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008
UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method
More informationGC Accessories Inlet Liners Sky Inlet Liners Other Deactivations Inlet Liner Accessories
GC Accessories Inlet Liners Sky Inlet Liners... 175 184 Overview...175 178 For Agilent GCs...179 180 For Bruker/Varian GCs...181 For PerkinElmer GCs...182 For Shimadzu GCs...183 For Thermo Scientific GCs...184
More information6890 GC Not Ready Messages (A16010)
6890 GC Not Ready Messages (A16010) This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information
More informationASTM WK Standard Test Method for Dissolved Gases. Anne Jurek Applications Chemist
ASTM WK 43267 Standard Test Method for Dissolved Gases Anne Jurek Applications Chemist Rationale Hydraulic Fracturing is becoming more and more common. Nearby water well can be affected during the fracking
More informationMicroJet Cryo-Trap MODEL MJT-1030Ex OPERATION MANUAL
MicroJet Cryo-Trap MODEL MJT-1030Ex OPERATION MANUAL www.frontier-lab.com BEFORE USING THIS DEVICE Thank you for purchasing our MicroJet Cryo-Trap (MJT-1030Ex). We advise you to read this operation manual
More informationAcclaim RSLC. (Rapid Separation Liquid Chromatography)
Acclaim RSLC (Rapid Separation Liquid Chromatography) Product Manual for Acclaim RSLC Columns Page 1 of 9 Product Manual for Acclaim RSLC Columns (Rapid Separation Liquid Chromatography) Analytical Columns
More informationRapid and Reliable Detection of Dissolved Gases in Water
Rapid and Reliable Detection of Dissolved Gases in Water Andrea Caruso and Massimo Santoro Thermo Fisher Scientific, Milan, Italy Application Note 005 Key Words Chromeleon CDS, Environmental, Fracking,
More informationMETHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES
METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide
More informationGases&Technology. Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. FEATURE.
Gases&Technology FEATURE Measurement of Impurities in Helium Using the Dielectric Barrier Discharge Helium Ionization Detector. B Y M A T T H E W M O N A G L E Abstract Bulk gases are often delivered to
More informationASTM 3612/TOGA/Dissolved Gas GC Revised October 2013
The SRI ASTM 3612/TOGA/ Dissolved Gas GC configuration ( TOGA GC ) permits measurement of gasses dissolved in water, oil or other liquids, and is based on the requirements of ASTM method 3612, option C,
More informationAscentis Express HPLC Columns with Fused-Core Technology
Ascentis Express HPLC Columns with Fused-Core Technology Extreme Performance on Any LC System Twice the Speed and Efficiency Extreme Performance on ANY LC System Half the Backpressure Super-Sensitivity
More informationSimple and Worry Free Gas Filters
Simple and Worry Free Gas Filters Know When to Replace Your Filter The indicator on each Zebron Gas Management Filter will notify you when it is time to change your filter.* Attach a New Filter in Under
More informationSRI Model H2-40 Hydrogen Generator Operation and Maintenance Nov 2006
Nov 2006 Place the H2-40 on a benchtop surface which will not be damaged by water in the event of a leak. Fill a bottle with clean water. Tapwater is OK in most large cities, but bottled drinking water
More informationAgilent 7820A GC. Site Preparation Guide. Agilent Technologies
Agilent 7820A GC Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2009 No part of this manual may be reproduced in any form or by any means (including electronic storage and
More informationAgilent 7697A Headspace Sampler
Agilent 7697A Headspace Sampler Troubleshooting Agilent Technologies Notices Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage
More informationCORESTA RECOMMENDED METHOD Nº 67
CORESTA RECOMMENDED METHOD Nº 67 DETERMINATION OF WATER IN THE MAINSTREAM SMOKE OF CIGARS BY GAS CHROMATOGRAPHIC ANALYSIS (November 2005) 1. FIELD OF APPLICATION The method is applicable to the particulate
More informationAPPLICATION NOTE. GC Integrated Permeation Device
GC Integrated Permeation Device GC-integrated Design PPB to PPM Level Calibration No Cylinders or Regulators Required Option for Dual Ovens Cost Effective, Safe, Clean, Flexible, NIST Traceable Keywords:
More informationFigure 1: Plug in ion inlet of mass spec
Starting procedure: by Vicky - is the vacuum pump running (should run all time, otherwise it ll need 4-5 hours to have a adequate vacuum for the MS) otherwise something has gone wrong- F.T - Remove plug
More informationConnectors. SECTION 2 OF 3 SECTIONS ON CONNECTORS VHP Tees and Crosses. Top Seller SEE STARRED PRODUCTS. VHP Tees & Crosses for Capillary Tubing
SECTION 2 OF 3 SECTIONS ON CONNECTORS VHP Tees and Crosses VHP Tees & Crosses for Capillary Tubing VHP Tees & Crosses for 1/16 OD Tubing IDEX Health & Science offers Very High Pressure, designed to bring
More informationBuck Scientific, Inc.
Buck Scientific, Inc. 910/310 Gas Chromatographs The Buck 910/310 Gas Chromatographs In the lab, the 910-series GCs perform routine industrial analyses and complicated research separations easily, generating
More informationAgilent 5975T LTM GC/MSD
Agilent 5975T LTM GC/MSD Site Preparation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage
More informationGenerating Calibration Gas Standards
Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and
More informationRunning HPLC Methods on an ACQUITY UPLC System
Running HPLC Methods on an ACQUITY UPLC System Note: This document is an addendum to the ACQUITY UPLC System Operator s Guide. Contents Topic Page System considerations 1-1 Choosing fittings 1-3 System
More informationLOWER YOUR DETECTION LIMITS AND QUANTIFY ACTIVE ANALYTES WITH CONFIDENCE
UPDATED VERSION Agilent Inert Flow Path Solutions LOWER YOUR DETECTION LIMITS AND QUANTIFY ACTIVE ANALYTES WITH CONFIDENCE ENSURING AN INERT FLOW PATH FROM INJECTION THROUGH DETECTION IS CRITICAL AND NOW,
More informationYour local gas generation partner. Precision series Modular gas generation solution for GC.
Your local gas generation partner Precision series Modular gas generation solution for GC www.peakscientific.com Perform with Precision Specifically designed and engineered for GC laboratory applications,
More informationbe present throughout these services; otherwise, they will miss important operational, maintenance and safety information.
Thank you for purchasing an Agilent Instrument. To get you started and to assure a successful and timely installation of your 9000 GC, please refer to this site prep checklist. Correct site preparation
More informationDuring today's webinar we're going to look at the following topics:
Alternative Carrier Gas for GC and GC-MS Presentation Transcript Hello and welcome to the Peak scientific webinar which today will concern alternative carrier gas for GC and GC-MS. My name is Ed Connor,
More informationDETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY
DETERMINATION OF TETRAHYDROTHIOPHENE IN AMBIENT AIR BY GAS CHROMATOGRAPHY WITH A PFPD DETECTOR COUPLED TO A PRECONCENTRATION TECHNOLOGY Nengbing Xu, Hongmei Ying and Libo Zhu Ningbo Environmental Monitoring
More information