USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve

Size: px
Start display at page:

Download "USER'S GUIDE. Nanoflow Probe. Red Stripe. 3-4mm. ptfe Sleeve. To Injector. ZU1XC Fused Silica TTF115. Fused Silica. ptfe Sleeve"

Transcription

1 USER'S GUIDE Nanoflow Probe 3-4mm Red Stripe ptfe Sleeve To Injector ZU1XC Fused Silica 20µm 90µm TTF115 ptfe Sleeve Fused Silica 25µm 375µm

2 Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: Fax: Tudor Road Altrincham WA14 5RZ Tel: Fax: Warning There are high voltages and hot surfaces present throughout the mass spectrometer. Samples analysed may be toxic. Many of the instrument's components are susceptible to contamination by toxic deposits. Operational and maintenance procedures may require the use of corrosive and flammable agents. All procedures must be performed only by suitably trained personnel, aware of the inherent hazards and observing every precaution. This manual is a companion to the instrument's and Parts List, and to the MassLynx User s Guide. All information contained in these manuals is believed to be correct at the time of publication. The publishers and their agents shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance or use of this material. All product specifications, as well as the information contained in this manual, are subject to change without notice. Issue 2 Micromass Ltd Page 1

3 Contents Introduction 3 Preparation 4 Loading the Sample 6 Fused Silica Syringe Needle 6 Gel Loader Tip 6 Operation 7 Calculation of Flow Rate 9 Fused Silica Capillary Preparation 10 Operation 13 Consumable Spares 14 Fused Silica Capillary 14 Contents Page 2

4 Introduction The Micromass nanoflow probe allows electrospray ionisation to be performed in the flow rate range 5 to 1000 nanolitres per minute. There are two options for the spraying capillary, which can be alternately fitted to the probe: Borosilicate metal coated glass capillaries. Metal coated glass capillaries allow the lowest flow rates to be obtained although they are used for one sample only and then must be discarded. Fused silica capillaries. This option is suitable for flow injection analyses or for coupling to nano-hplc, and uses a pump to regulate the flow rate down to 100 nl/min. If a syringe pump is to be used, a gas -tight syringe is necessary to obtain correct flow rates without leakage. For a given sample concentration, the ion currents observed in nanoflow are comparable to those seen in normal flow rate electrospray. Great sensitivity gains are therefore observed when similar scan parameters are used, due to the great reductions in sample consumption. A list of consumable spare parts is included at the end of this document. Introduction Page 3

5 Preparation Referring to the instrument's, remove the counter electrode from the API source and clean the sample cone. Attach the alignment insert to the inner source. Reduce the source temperature to 30 C. Connect the source's drying gas inlet tube to the instrument's nebuliser gas supply. On Quattro 1, use the 1 /8" - 1 /16" adaptor to connect the bath gas line to the nebuliser flow regulator. Ensure that the exhaust tube from the inner source is not connected. Carefully remove the capillary from its case by lifting vertically while pressing down on the foam with two fingers. Caution: It is important to take great care when handling the tips as they are extremely fragile. Always handle using the square end of the capillary. Foam Capillary Page 4

6 Locate the Swagelok union in which the capillary is to be held and remove the knurled nut. Cut a 5mm length of conductive elastomer, if not already available. Pass, in turn, the knurled nut, the elastomer and the union, over the square end of the capillary. Secure the nut, finger tight, such that approximately 15mm of capillary protrudes from the nut as shown. Glass Capillary 15mm Blue Conductive Elastomer Blanking Nut 6mm ptfe 6-4mm Reducer Page 5

7 Loading the Sample There are five types of capillary available, each with a slightly different tip profile: Types A, C and E have been found to work optimally when using, for example, peptide samples. Types B and D work optimally for larger molecules or different solvent systems, such as 100% water + 2mM ammonium acetate. The capillary may be reused if the same sample is to be analysed, but to avoid cross contamination it is recommended that a new capillary be used for each sample. If possible, filter the sample using a 0.2µm Anopore filter. It has been found that beam stability improves when 50% methanol is used instead of 50% acetonitrile. The sample can now be loaded into the capillary using either a fused silica syringe needle or gel loader tips, as follows. Fused Silica Syringe Needle Gel Loader Tip Push the needle to the base of the tip before depressing the syringe plunger. A fused silica needle syringe is commercially available (Hamilton 75RNFS) which allows accurate volumes to be injected. Inject the sample to about half way. Hold the capillary vertically until the sample runs to the pointed end of the capillary. Gel loader tips allow accurate volumes to be injected without the possibility of cross contamination. Page 6

8 Operation Once the sample has been loaded into the capillary: Secure the Swagelok union to the probe. For Quattro 1, check that the distance from the probe stop to the probe tip is less than 257mm. For other instruments, check that the distance from the front of the probe body to the tip is less than 171mm. Referring to the diagram: On Quattro 1, attach the back pressure gas to the 1 /16" tubing using the ¼" to 1 /16" union provided. On other instruments, attach the nitrogen gas line to the red coupling at the back of the probe (in normal electrospray this would be the nebulising gas supply coupling). 0 to 2 bar (30 psi) Regulator 6mm ptfe Valve Tee 6mm ptfe 6mm - 4mm Reducer To Probe Nitrogen Quattro 1 6mm - 1/16" Reducer To Probe 1/16" ptfe Page 7

9 Insert the probe into the source. Set the back pressure to 70 to 140 mbar (1 to 2 psi). On the tune page, set Capillary and HV Lens to zero. Select. Set Capillary on the tune page to 1.5kV. An ion beam should now be present. Use the source parameters to tune as in normal flow rate electrospray. The normal operating flow rate for the bath gas is approximately 40 litres/hour. To improve beam stability and sensitivity: Adjust Capillary and/or the nitrogen back pressure. The capillary voltage should not need to be raised above 2.5kV and may even require reducing to achieve a stable beam. The nitrogen back pressure should not be raised above 1.4 bar (20 psi). If no ion beam is obtained, it may be possible to initiate the spray as follows: Reduce any back pressure to zero using the valve on the gas line and reset the regulator pressure to 70 to 140 mbar (1 to 2 psi). Remove the probe from the source. Using a tissue, touch very lightly on the tip. A drop of liquid may appear. Reinsert the probe and repeat the above procedure. Page 8

10 Calculation of Flow Rate If an unknown amount of sample is placed in the capillary, the amount of sample consumed and the average flow rate may be calculated as follows: Measure the internal diameter of the capillary (approximately 0.58mm). Measure the length (in mm) of liquid used during the run. By holding the capillary up to light it is possible to see the meniscus level of the liquid. Calculate the amount of sample consumed. mm³ is equivalent to µl. Knowing the duration of the run, calculate the average flow rate. Page 9

11 Fused Silica Capillary 3-4mm Red Stripe ptfe Sleeve To Injector Fused Silica 20µm 90µm ZU1XC TTF115 ptfe Sleeve Fused Silica 25µm 375µm Preparation Warning: Always wear eye protection when handling fused silica. Referring to the instrument's, remove the counter electrode from the API source and clean the sample cone. Attach the alignment insert to the inner source. Reduce the source temperature to 30 C. Ensure that the exhaust tube from the inner source is not connected. Cut approximately 600mm of the 25µm i.d. fused silica. The 25µm i.d. silica links the Valco injector on the top of the probe to the low dead volume union. Take care not to confuse the 25µm i.d. fused silica with the 75µm i.d. fused silica which is used to link the pump to the injector. Cut approximately 40mm of the 20µm i.d. fused silica. The 20µm i.d. silica is used as the probe tip. Fused Silica Capillary Page 10

12 It is recommended a white background is used when handling the 20µm i.d. fused silica. Cut approximately 30mm of the ptfe tubing. Cut approximately 30mm of the red stripe peek tubing. Locate the ultra low dead volume Valco union and the Valco plug cap. Remove the probe cover. Loosen the fingertight nuts in the union at the back of the probe. If fitted, remove the blanking nut from the Swagelok tee in the probe and the Swagelok adapter used in the borosilicate capillary option. Thread one end of the 25µm i.d. fused silica through the ptfe sleeve (TTF115). Set the correct pilot depth for the ptfe sleeve and a Valco ferrule using the plug cap provided. There is no need to overtighten when setting the pilot depth. Remove the ptfe sleeve from the plug cap and the nut from the ptfe sleeve. With a small amount of fused silica protruding, cut a small piece (approximately 2mm) off the end and then ensure that the fused silica is flush with the sleeve, using a scalpel blade to push it back. This ensures that the fused silica has a clean end. Using the plug cap again, set the correct pilot depth for the red stripe peek sleeving and a Valco ferrule. Remove the peek from the plug cap and the nut from the peek. Cut the peek so that approximately 5mm is protruding from the back end of the ferrule, and hence none can be seen when the nut is replaced. Thread the 20µm i.d. fused silica through the peek and ensure that the fused silica is flush with the peek sleeve using a scalpel blade. Attach the peek sleeved fused silica to one end of the low dead volume union using a normal Valco nut and attach the ptfe sleeved fused silica to the other end using the Valco adapter. Ensure both pieces of fused silica are tightly held. Thread the 25µm fused silica through the probe shaft and through the 0.023" i.d. ptfe sleeves (in 2 places) in the fingertight union at the probe end plate. Fused Silica Capillary Page 11

13 Screw the Valco adapter into the probe shaft (fingertight is adequate) and tighten the fingertight nuts to secure the fused silica. Cut down the 20µm i.d. fused silica such that 3 to 4mm protrudes from the nut. Attach the injector valve to the bracket and then the bracket to the probe. Replace the probe cover. The ports on the injector valve are as follows: C P W S column pump waste sample (attach the VISF port). Connect the pump to the injector using 75µm i.d. fused silica. For Quattro 1, adjust the probe stop such that approximately 10mm of the stop protrudes from the bracket. For other instruments, set the distance between the probe stop and the fused silica tip to 167mm. Ensure that any air bubbles are purged from the plumbing. This is effectively done by first purging up to and including the valve at a high flow rate (approximately 5 µl/min) with the 25µm i.d. fused silica disconnected. Reduce the flow to the required rate and attach the 25µm fused silica to the valve output. Insert the probe into the source. Fused Silica Capillary Page 12

14 Operation It has been found that beam stability improves when 50% methanol is used instead of 50% acetonitrile. The normal operating flow rate for the bath gas is approximately 40 litres/hour. On the tune page, set Capillary and HV Lens to zero. Select. A stable beam should be obtained with a Capillary voltage of ~2.5kV. It may be necessary to inject a known sample, for example gramicidin, to see a stable beam. However, there will be an upper and lower limit where the beam starts to go unstable with a difference between the two of approximately 800V. It has been found that the system operates efficiently when the capillary voltage is set to the mid point of these values. With time, the Capillary voltage needed to obtain a stable beam will start to rise. This usually means that the sample cone requires cleaning again with methanol. Fused Silica Capillary Page 13

15 Description Consumable Spares Ref. No. Borosilicate glass capillary, type A Borosilicate glass capillary, type B Borosilicate glass capillary, type C Borosilicate glass capillary, type D Borosilicate glass capillary, type E Blue conductive elastomer Fused Silica Capillary Description Ref. No. 25µm i.d. fused silica capillary µm i.d. fused silica capillary ptfe tubing TTF Red stripe peek tubing Ultra low dead volume Valco union Valco plug cap ptfe sleeve 0.023" i.d VISF port µm i.d. fused silica Consumable Spares Page 14

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column

UWPR. LC-MS setup. Before you connect your trap and column. Connecting your trap and column Before you connect your trap and column You need to provide your own trap and column (Note our fittings accommodate trap and column made with fused silica 360 µm OD). Check our guide for packing traps

More information

Addendum to the Waters Micromass Quattro Premier XE Mass Spectrometer Operator s Guide

Addendum to the Waters Micromass Quattro Premier XE Mass Spectrometer Operator s Guide Addendum to the Waters Micromass Quattro Premier XE Mass Spectrometer Operator s Guide 715001216/Revision A Copyright Waters Corporation 2006. All rights reserved. Copyright notice 2006 WATERS CORPORATION.

More information

Connectors. SECTION 2 OF 3 SECTIONS ON CONNECTORS VHP Tees and Crosses. Top Seller SEE STARRED PRODUCTS. VHP Tees & Crosses for Capillary Tubing

Connectors. SECTION 2 OF 3 SECTIONS ON CONNECTORS VHP Tees and Crosses. Top Seller SEE STARRED PRODUCTS. VHP Tees & Crosses for Capillary Tubing SECTION 2 OF 3 SECTIONS ON CONNECTORS VHP Tees and Crosses VHP Tees & Crosses for Capillary Tubing VHP Tees & Crosses for 1/16 OD Tubing IDEX Health & Science offers Very High Pressure, designed to bring

More information

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All

More information

PicoChip. Nanospray System. Column Care & Use Manual

PicoChip. Nanospray System. Column Care & Use Manual PicoChip Nanospray System Column DOCUMENT: 34414 RELEASE DATE: December 22, 2014 Read This First Read This First This document is provided to customers who have purchased New Objective equipment to use

More information

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule

UP-F Delrin TM Winged Nut. UP-F PEEK TM Nut. with F-142 PEEK TM Ferrule One Piece or Two? A one-piece fitting is more convenient and less cumbersome, since the ferrule cannot stick in a receiving port and the fitting is more easily found if dropped. With two-piece Fingertight,

More information

The Open-Split Capillary Interface. Part No

The Open-Split Capillary Interface. Part No The Open-Split Capillary Interface Part No. 113532 2 INTRODUCTION With the advent of fused silica capillary columns and their high degree of flexibility, it is obviously attractive to be able to take the

More information

Connecting a SSL/PTV Backflush System

Connecting a SSL/PTV Backflush System This section provides instructions for connecting your SSLBKF/PTVBKF injector modules with pre-column and analytical capillary column into the oven. Refer to the chapters SSLBKF Injector Module and PTVBKF

More information

DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3

DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3 DART-Bruker Vapur Interface Manual SI-410-A-R Rev 1-3 For All Bruker Mass Spectrometer Models (Except Bruker Esquire) Rev 1 Document # 7.5.060 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright

More information

Valve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve

Valve INV-907. Instructions. 1 Introduction. 2 Installation. Safety. Unpacking. Installing the valve Instructions 1 Introduction is a motorised rotary 7 port valve. The valve is used in ÄKTA design chromatography systems. It is powered from the ÄKTAdesign P-900 series system pump, and is controlled from

More information

Agilent 1290 Infinity Pump Head Maintenance

Agilent 1290 Infinity Pump Head Maintenance Agilent 1290 Infinity Pump Head Maintenance Technical Note Agilent Technologies Notices Agilent Technologies, Inc. 2012-2015, 2016 No part of this manual may be reproduced in any form or by any means (including

More information

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany

Manual. Online Degasser DDG-75. Manual Degasser DDG-75 April DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE Hockenheim Germany Manual Degasser DDG-75 April 2007 1 Manual Online Degasser DDG-75 DURATEC Analysentechnik GmbH Rheinauer Strasse 4 DE-68766 Hockenheim Germany Tel. +49-6205-9450-0 Fax. +49-6205-9450-33 email info@duratec.com

More information

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct

More information

LC, RSLC, Bio RS and Dual Gradient Systems

LC, RSLC, Bio RS and Dual Gradient Systems Thermo Scientific Dionex UltiMate 3000 LC, RSLC, Bio RS and Dual Gradient Systems P/N 4823.0205 Revision B December 2014 Automated Method Scouting Considerations All information in this manual is subject

More information

User Manual. Quantos Automated Dosing Liquid Module

User Manual. Quantos Automated Dosing Liquid Module User Manual Liquid Module 1 Safety Information 1.1 Definition of warnings and symbols Signal Words WARNING for a hazardous situation with medium risk, possibly resulting in severe injuries or death if

More information

Fiber Optic SMA Z-Flow Cell Manual Version 2.1

Fiber Optic SMA Z-Flow Cell Manual Version 2.1 Fiber Optic SMA Z-Flow Cell Manual Version 2.1 The line of fiber optic SMA Z-Cells is the top choice for many absorbance measurements. SMA-Z-Cell path lengths are available between 2.5mm - 100mm and can

More information

LCT User's Guide. Micromass UK Limited. Floats Road Wythenshawe M23 9LZ Tel: Fax:

LCT User's Guide. Micromass UK Limited. Floats Road Wythenshawe M23 9LZ Tel: Fax: Micromass UK Limited Floats Road Wythenshawe M23 9LZ Tel: +44 161 945 4170 Fax: +44 161 998 8915 Tudor Road Altrincham WA14 5RZ Tel: +44 161 282 9666 Fax: +44 161 282 4400 http://www.micromass.co.uk The

More information

6900 Maintenance Instruction System Flush

6900 Maintenance Instruction System Flush Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex

More information

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A.

Gas Lines. Technical Manual. Sumitomo (SHI) Cryogenics of America, Inc Vultee Street Allentown, PA U.S.A. Gas Lines Technical Manual Sumitomo (SHI) Cryogenics of America, Inc. 1833 Vultee Street Allentown, PA 18103-4783 U.S.A. Revision F: April 2008 261320A TABLE OF CONTENTS Page DESCRIPTION...1 SPECIFICATIONS...2

More information

Constant Pressure Crude Oil Container Model CPCCP

Constant Pressure Crude Oil Container Model CPCCP Installation, Operations, and Maintenance Manual Constant Pressure Crude Oil Container Model CPCCP The information in this manual has been carefully checked for accuracy and is intended to be used as a

More information

Pressure Injection Cell Operator s Manual

Pressure Injection Cell Operator s Manual Pressure Injection Cell Operator s Manual Operator s Manual Pressure Injection Cell CONGRATULATIONS! Congratulations on your purchase of a Next Advance Pressure Injection Cell. Please read this operator

More information

INSTRUCTION MANUAL. January 23, 2003, Revision 0

INSTRUCTION MANUAL. January 23, 2003, Revision 0 INSTRUCTION MANUAL Model 810A In-Vitro Test Apparatus for 310B Muscle Lever January 23, 2003, Revision 0 Copyright 2003 Aurora Scientific Inc. Aurora Scientific Inc. 360 Industrial Parkway S., Unit 4 Aurora,

More information

Vibration isolation system 1VIS10W. User manual

Vibration isolation system 1VIS10W. User manual Vibration isolation system 1VIS10W User manual Standa 2014 Table of contents 1. General information 3 1.1. Introduction 3 1.1.1. Safety 5 1.2. Location of the table 5 1.3. Air supply requirements 5 2.

More information

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection

User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection User guide for operating the Agilent 6420A QqQ mass spectrometer by direct infusion/injection This user guide does not have the intention to introduce you to any of the theory and/or principles behind

More information

Pressure Dump Valve Service Kit for Series 2300 Units

Pressure Dump Valve Service Kit for Series 2300 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 00 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00 applicator tanks when the unit

More information

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL!

Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL! Parallel Reactor Standard Operating Procedure THIS SOP IS NOT MEANT TO REPLACE HANDS-ON TRAINING FROM QUALIFIED PERSONNEL! This SOP describes the basic procedure for using the HEL Cat-24 parallel reactor,

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

Agilent Auxiliary Gas Module AGM 2. User s Guide

Agilent Auxiliary Gas Module AGM 2. User s Guide Agilent Auxiliary Gas Module AGM 2 User s Guide Notices Agilent Technologies, Inc. 1992, 1996, 1999, 2010 No part of this manual may be reproduced in any form or by any means (including electronic storage

More information

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Probe Instrument Regulator Model The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for

More information

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301

THE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301 THE HF-300 SERIES Operating and Service Manual Series includes all variants of HF-300/301 Issue A July 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 3.1. Spring Loaded...

More information

Bermad Pressure Reducing. Model: 42T

Bermad Pressure Reducing. Model: 42T Bermad Pressure Reducing Pilot Operated Pressure Control Valve Model: 42T Installation Operation Maintenance Manual (IOM) REV. 27.7.17 Page 1 of 12 Safety First BERMAD believes that the safety of personnel

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

accidents which arise due to non-observance of these instructions and the safety information herein. SPECIFICATIONS

accidents which arise due to non-observance of these instructions and the safety information herein. SPECIFICATIONS 18 GAUGE 2 INCH BRAD NAILER Model: 7555 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement, or other masonry.

More information

Customer Responsibilities

Customer Responsibilities Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation

More information

accidents which arise due to non-observance of these instructions and the safety information herein. SPECIFICATIONS

accidents which arise due to non-observance of these instructions and the safety information herein. SPECIFICATIONS 18 GAUGE 1-1/4 INCH BRAD NAILER Model: 7611 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement, or other masonry.

More information

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity

Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Agilent ICP-MS Interactive Troubleshooting tool for Low sensitivity Rev. B 2017,Janyary,20 Next > Preconditions 7900, 7800, 7700, or 8800 ICP-MS is used. Sample introduction type is PeriPump. If you are

More information

Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter. Instructions for Use

Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure Reverse Cleaning Adapter. Instructions for Use Penn-Century, Inc. 7670 Queen Street Suite 200 Wyndmoor, PA 19038-8031 USA 215-753-6540 Tel 215-220-3040 Fax www.penncentury.com Unblocking the MicroSprayer Aerosolizer Model IA-1C With the High Pressure

More information

Techcon Systems TS1254 Pressure Pot

Techcon Systems TS1254 Pressure Pot Techcon Systems TS1254 Pressure Pot User Guide Copyright OK International, Inc. TABLE OF CONTENT 1. SAFETY. 3 1.1 Intended Use 3 1.2 Safety Precaution 3 2. FEATURES. 5 3. SPECIFICATIONS 4 4. INSTALLATION

More information

Service and Repair Manual

Service and Repair Manual II stage R2 Ice/ Special, II stage R 1 Pro DOWNSTREAM 2 nd STAGE REGULATOR Service and Repair Manual Introduction Safety Precautions...4 General Procedures, Maintenance Schedules...5 Initial Inspection

More information

AIR-OPERATED DOUBLE DIAPHRAGM PUMP USER S MANUAL

AIR-OPERATED DOUBLE DIAPHRAGM PUMP USER S MANUAL 00, 0, 000 00, 000, 00 A. TECHNICAL INFORMATION Model 00 Inlet/Outlet " Air Inlet /" 0 / 000 /" /" 00 /" /" 000 /" /" 00 /" /" Flow Rate GPM/ 0LPM GPM/ 0LPM GPM/ LPM GPM/ LPM GPM/ 0LPM GPM/ 0LPM Maximum

More information

Neotecha Tank Sampling Sapro

Neotecha Tank Sampling Sapro Tank sampling system for representative sampling without process interruption Features The sampling system assures a true representative sample every time without process interruption. The sampling system

More information

B. Assemble the Method-25 sample train with a heated probe/filter.

B. Assemble the Method-25 sample train with a heated probe/filter. 6/30/2015 VOC Reporting, Inc. Method 25 Heated Controller Operating Instructions. pg. 1 of 5 INSTRUCTIONS FOR EPA METHOD-25 FIELD SAMPLE COLLECTION USING A VRi HEATED PROBE AND FILTER ASSEMBLY Introduction:

More information

6890 GC Site Preparation

6890 GC Site Preparation 6890 GC Site Preparation (a16011) This document is believed to be accurate and up-to-date. However, Agilent Technologies, Inc. cannot assume responsibility for the use of this material. The information

More information

operator s manual plus system system nanolc Operator s Manual Revision A1

operator s manual plus system system nanolc Operator s Manual Revision A1 operator s manual plus system system Notice Support At Eksigent Technologies, we are committed to providing the highest level of support to our users. To obtain answers to questions about any of our products,

More information

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT

Syringe, Distribution Valve and Infusion Pump Removal/Replacement ATTENTION SYRINGE REPLACEMENT ATTENTION SYRINGE REPLACEMENT Please read through the document completely before starting any repairs. Refer to the proper section in the service manual for complete removal and replacement procedures.

More information

4400 TwinHybrid Gas Seal

4400 TwinHybrid Gas Seal 4400 TwinHybrid Gas Seal EQUIPMENT PREPARATION MECHANICAL SEAL INSTALLATION INSTRUCTIONS 1 2 200.005" 0,13 mm 125 µ" 3,2 µm R a 3 4 32 µ" 0,8 µm R a 1000 ±.002" 0,05mm CAUTIONS These instructions are general

More information

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware

Water Weir Flow Controller. Introduction. Safety Precautions. Mounting the Hardware 57007-88 Introduction Safety Precautions This instruction sheet describes how to set up and use the Hach (Figure 1). A water weir is a device that raises or diverts water to regulate the flow. Hach s water

More information

Best Practice for Identifying Leaks in GC and GC/MS Systems

Best Practice for Identifying Leaks in GC and GC/MS Systems Best Practice for Identifying Leaks in GC and GC/MS Systems Technical Overview Maintaining a leak-free GC or GC/MS system is critical for obtaining optimal system performance with reliable, reproducible,

More information

Making OTR Measurements in Bottles using the. BottlePerm Fixture

Making OTR Measurements in Bottles using the. BottlePerm Fixture Application Note: Dated: June 24, 2013 Introduction Making OTR Measurements in Bottles using the BottlePerm Fixture Measurement of oxygen transmission rate (OTR) of bottles can be complicated and time

More information

Sample Concentrator SBH CONC/1 Instructions for use

Sample Concentrator SBH CONC/1 Instructions for use Sample Concentrator SBH CONC/1 Instructions for use Thank you for purchasing this piece of Barloworld Scientific equipment. To get the best performance from the equipment, and for your own safety, please

More information

CRYOGENIC TANK SERVICES - CHART RECORDER MANUAL

CRYOGENIC TANK SERVICES - CHART RECORDER MANUAL CRYOGENIC TANK SERVICES - CHART RECORDER MANUAL Pressure Chart Recorder / Temperature Chart Recorder / Dual Recorder Portable Mechanical Chart Recorder for the accurate measurement and recording of Pressure

More information

215 Liquid Handler Up-Grade Manual

215 Liquid Handler Up-Grade Manual BRUKER NMR 220 25 Liquid Handler Up-Grade Manual For Up-Grade to Run with XWINNMR Vers.. Manual # H977 The kit will include: H9770 BEST NMR Up-Grade Kit to No. Article Bruker # Comment 500mm Capillary

More information

Pressure Dump Valve Service Kit for Series 3000 Units

Pressure Dump Valve Service Kit for Series 3000 Units Instruction Sheet Pressure Dump Valve Service Kit for Series 000 Units. Overview The Nordson pressure dump valve is used to relieve hydraulic pressure instantly in Series 00, 400, 500, and 700 applicator

More information

MB-SP Series Solvent Purification System

MB-SP Series Solvent Purification System Operation Manual MB-SP Series Solvent Purification System Table of contents 1. System overview 2 2. Column activation.. 5 3. System purging.. 6 4. Glove box integration.... 7 5. Solvent reservoir installation....

More information

Model: 720-UL INSTALLATION OPERATION MAINTENANCE. Bermad Pressure Reducing Valve IOM. Model: FP -720-UL Sizes: 2"-12" BERMAD. Application Engineering

Model: 720-UL INSTALLATION OPERATION MAINTENANCE. Bermad Pressure Reducing Valve IOM. Model: FP -720-UL Sizes: 2-12 BERMAD. Application Engineering Bermad Pressure Reducing Valve Model: 720-UL INSTALLATION OPERATION MAINTENANCE Application Engineering BERMAD 1. Safety First BERMAD believes that the safety of personnel working with and around our equipment

More information

Disassembling and assembling APP and APP 16-22

Disassembling and assembling APP and APP 16-22 MAKING MODERN LIVING POSSIBLE Instruction Disassembling and assembling APP 11-13 and APP 16-22 ro-solutions.com Table of Contents 1. Disassembling...3 2. Disassembling the pump...3 3. Assembling the pump....7

More information

RK-IR Sample Draw Aspirator Adapter Operator s Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual 30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

Neotecha Tank Sampling Sapro

Neotecha Tank Sampling Sapro E Neotecha Tank Sampling Sapro Tank sampling system for representative sampling without process interruption Features The sampling system assures a true representative sample every time without process

More information

AP100 series ph/redox (ORP) cartridge sensors with water-wash option

AP100 series ph/redox (ORP) cartridge sensors with water-wash option ABB MEASUREMENT & ANALYTICS INSTRUCTION MANUAL IM/AP100 REV. D AP100 series ph/redox (ORP) cartridge sensors with water-wash option Measurement made easy AP100 series ph/redox (ORP) cartridge sensors For

More information

PFC Analysis Kit for ACQUITY UPLC System Guide

PFC Analysis Kit for ACQUITY UPLC System Guide PFC Analysis Kit for ACQUITY UPLC System Guide 71500183002/Revision A Copyright Waters Corporation 2009 All rights reserved Copyright notice 2009 WATERS CORPORATION. PRINTED IN THE UNITED STATES OF AMERICA

More information

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE

NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE NB/NBR NITROGEN BOOSTER FOR AVIATION SERVICE INSTALLATION, OPERATION & MAINTENANCE MANUAL INTERFACE DEVICES, INC. 230 Depot Road, Milford, CT 06460 Ph: (203) 878-4648, Fx: (203) 882-0885, E-mail: info@interfacedevices.com

More information

Installation of Your SprayMaster System

Installation of Your SprayMaster System Installation of Your SprayMaster System 1. At the installation site, remove all equipment from the corrugated box and the polyethylene drum and replace the drum lid. Check the picture to identify each

More information

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:

EASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax: 2 Table Of Contents 1. Introduction 3 2. About this Manual 3 3. Contacting YZ Systems 3 4. Vessel Components 4 5. Specifications 5 6. Application 6 7. Theory of Operation 7 8. DuraSite Installation & Use

More information

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight Technical Specifications Operating Instructions Maintenance Information Troubleshooting Guide Parts Diagrams AC1810 / AC1810-A THE EVOLUTION OF PERFECTION CAUTION: Before attempting to use or service this

More information

ACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS

ACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS ACCUMULATOR OPERATING & MAINTENANCE INSTRUCTIONS READ ALL INSTRUCTIONS PRIOR TO INSTALLATION AND OPERATION TO AVOID POSSIBLE INJURY Warning: Always consider any accumulator to contain pressure until proven

More information

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #: ID-Cube Installation and Operations Manual For ID-Cube Source and Power Supply Revision 2 Document #: 7.1.002 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright 2014 by IonSense Inc. All rights

More information

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 4075 INSTRUCTION MANUAL

REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 4075 INSTRUCTION MANUAL REVERSE OSMOSIS WATER FILTRATION SYSTEM MODEL EWR 4075 INSTRUCTION MANUAL Excalibur Water Systems 142 Commerce Park Drive, Unit M & N Barrie, Ontario L4N 8W8 CANADA www.excaliburwater.com 2013.05.7446

More information

GE Healthcare. Valve INV-907. Instructions ÄKTA

GE Healthcare. Valve INV-907. Instructions ÄKTA GE Healthcare Valve INV-907 Instructions ÄKTA Important user information All users must read this entire manual to fully understand the safe use of Valve INV-907. WARNING! CAUTION! Note CE Certifying

More information

Installation and Use Manual

Installation and Use Manual Installation and Use Manual EMASMB & LMASMB Surface Mount Bottle Filling Stations Model EMASMB IMPORTANT THIS IS AN INDOOR APPLICATION ONLY! ALL SERVICE TO BE PERFORMED BY AN AUTHORIZED SERVICE PERSONNEL.

More information

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions

Anderson Greenwood Series 93 Positive Pressure POSRV Installation and Maintenance Instructions Before installation these instructions must be fully read and understood Installation and maintenance instructions for Series 93 Positive Pressure Pilot Operated Safety Relief Valves (POSRV). The intent

More information

User Manual. Quantos Automated Dosing Powder Module

User Manual. Quantos Automated Dosing Powder Module User Manual Powder Module 1 Introduction Thank you for purchasing the Quantos dosing system. Quantos is based on the renowned and reliable XPE series balances from METTLER TOLEDO. Besides the highly accurate

More information

abc INSTALLATION, OPERATING & MAINTENANCE MANUAL FOR HYDRAULIC EXTRACTOR

abc INSTALLATION, OPERATING & MAINTENANCE MANUAL FOR HYDRAULIC EXTRACTOR abc INSTALLATION, OPERATING & MAINTENANCE MANUAL FOR HYDRAULIC EXTRACTOR This Jiskoot Product is designed to provide outstanding service if correctly installed, used and maintained recognising the effects

More information

Instruction Manual Updated 7/26/2011 Ver. 2.2

Instruction Manual Updated 7/26/2011 Ver. 2.2 4-Unit Model MB HTHP Filter Press #171-50-4: 115-Volt #171-51-4: 230-Volt Instruction Manual Updated 7/26/2011 Ver. 2.2 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele:

More information

Endura AZ40 Sensor assembly High temperature filter assembly kit

Endura AZ40 Sensor assembly High temperature filter assembly kit Instruction replacement procedure INS/ANAINST/026 EN Endura AZ40 Sensor assembly High temperature filter assembly kit Kit reference: AZ400-761, -771, -772 Measurement made easy 1 Introduction This publication

More information

BCV31 DN20 - Blowdown Control Valve

BCV31 DN20 - Blowdown Control Valve 4034650/8 IM-P403-37 AB Issue 8 BCV31 DN20 - Blowdown Control Valve Installation and Maintenance Instructions 1. Safety information 2. Application 3. Technical data 4. Operation 5. Installation 6. Flow

More information

Installation, Operation, and Maintenance Manual

Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker The information in this manual has been carefully checked for accuracy and is intended to be used as a guide for the installation, operation, and

More information

MoveRoll Conveyor Operating and Maintenance Manual

MoveRoll Conveyor Operating and Maintenance Manual MoveRoll Conveyor Operating and Maintenance Manual 1. Read this first! This manual contains information for protection of personnel in the roll handling area from possible injury and/or equipment damage.

More information

Halsey Taylor Owners Manual

Halsey Taylor Owners Manual Halsey Taylor Owners Manual Refrigerated Marblyte Fountains with Back Panel HRFG - ER HRFG - SR HRFG - SER Installer To assure you install this model easily and correctly, PLEASE READ THESE SIMPLE INSTRUCTIONS

More information

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM IMI CORNELIUS INC One Cornelius Place Anoka, MN 55303-623 Telephone (800) 238-3600 Facsimile (612) 22-326 INSTALLATION INSTRUCTIONS FOR INSTALLING CO 2 BLENDER KIT (P/N 111612000 IN BEER SYSTEM SECONDARY

More information

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0)

Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW. Tel +44 (0) Allspeeds Ltd. Royal Works, Atlas St Clayton le Moors Accrington Lancashire England BB5 5LW Tel +44 (0)1254 615100 www.allspeeds.co.uk HP690A Intensifier Panel including the following versions: Model Intensification

More information

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401

THE MF-400 SERIES. Operating and Service Manual. Series includes all variants of MF-400/401 THE MF-400 SERIES Operating and Service Manual Series includes all variants of MF-400/401 Issue A October 2013 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 4. Special

More information

Operator: Save these instructions for future use!

Operator: Save these instructions for future use! WHITE-RODGERS Type 36C67 Combination (24 Volt Models) INSTALLATION INSTRUCTIONS Operator: Save these instructions for future use! FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING

More information

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION

Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Blue River Technologies Port-A-Poly Mixer w/2.5 GPH LMI Pump And Secondary Water Dilution Line INSTALLATION AND OPERATION Install your Blue River Technologies Port-A-Poly mixing system in a clean dry area.

More information

4 ANGLE GRINDER MODEL NO: CAT 52 PART

4 ANGLE GRINDER MODEL NO: CAT 52 PART 4 ANGLE GRINDER 4 ANGLE GRINDER MODEL NO: CAT 52 PART No: 3110685 OPERATION & MAINTENANCE INSTRUCTIONS 0807 Fig.1 SPECIFICATIONS Model:...CAG52 Part Number:...3110685 Rated Wheel...Capacity: 4 x 1/4 (type

More information

Unblocking the MicroSprayer Aerosolizer Model IA-1B With the High Pressure Reverse Cleaning Adapter. Instructions for Use

Unblocking the MicroSprayer Aerosolizer Model IA-1B With the High Pressure Reverse Cleaning Adapter. Instructions for Use Penn-Century, Inc. 7670 Queen Street Suite 200 Wyndmoor, PA 19038-8031 USA 215-753-6540 Tel 215-220-3040 Fax www.penncentury.com Unblocking the MicroSprayer Aerosolizer Model IA-1B With the High Pressure

More information

Description Pressure range Connection thread Device Page

Description Pressure range Connection thread Device Page Pressure Regulator Pressure range onnection thread Device Page max. 22 l/min miniature -850 0 mbar 1 8 NPT V800.02 max. 22 l/min miniature -850 0 mbar 10-32 and flange V900.02 max. 0 l/min precise -1 +0,.4

More information

Agilent G3180B Two-Way Splitter Kit

Agilent G3180B Two-Way Splitter Kit Agilent G3180B Two-Way Splitter Kit With Makeup Gas Installation and Operation Guide Agilent Technologies Notices Agilent Technologies, Inc. 2006 No part of this manual may be reproduced in any form or

More information

SP-400-M1/SP-403K-M1 INSTRUCTIONS

SP-400-M1/SP-403K-M1 INSTRUCTIONS SP-400-M1/SP-403K-M1 INSTRUCTIONS PACK YOUR OWN CAPILLARIES FOR HPLC AND LC-MS revision 3-15-15 TABLE OF CONTENT SP-400-M1 AND SP-403-M1 KIT COMPONENTS PAGE 2 COMPONENT ASSEMBLY DIAGRAM PAGE 3 COMPONENT

More information

APP pumps APP and APP Disassembling and assembling

APP pumps APP and APP Disassembling and assembling Service guide APP pumps APP 11-13 and APP 16-22 Disassembling and assembling hpp.danfoss.com Table of Contents Contents 1. Introduction... 2 2. Disassembling the pump... 3 3. Assembling the pump... 6 4.

More information

36H SERIES Combination Gas Valve

36H SERIES Combination Gas Valve FAILURE TO READ AND FOLLOW ALL INSTRUCTIONS CAREFULLY BEFORE INSTALLING OR OPERATING THIS CONTROL COULD CAUSE PERSONAL INJURY AND/OR PROPERTY DAMAGE. DESCRIPTION The 36H series combination gas valve is

More information

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A)

Site Preparation Specification for GCMS system. Dimensions and Weight. Agilent 7820MSD Series G3175A (G7020A), G3176A (G7021A) Purpose of Procedure Your site must meet this specification or set of requirements to assure a successful and timely installation of your 7820MSD Series Mass Selective Detector (MSD). This checklist is

More information

I T T Pressure Reducing Valve WARNING INSTALLATION, OPERATION, AND MAINTENANCE MANUAL

I T T Pressure Reducing Valve WARNING INSTALLATION, OPERATION, AND MAINTENANCE MANUAL INSTALLATION, OPERATION, AND MAINTENANCE MANUAL I-867-4T 867-4T Pressure Reducing Valve HANG THESE INSTRUCTIONS ON THE INSTALLED VALVE FOR FUTURE REFERENCE WARNING Read and understand all instructions

More information

Techcon Systems TS1258 Pressure Pot

Techcon Systems TS1258 Pressure Pot Techcon Systems TS1258 Pressure Pot User Guide TABLE OF CONTENT 1. SAFETY. 3 1.1 Intended Use 3 1.2 Safety Precaution 3 2. CERTIFICATE OF COMFORMANCE 4 3. FEATURES. 5 4. SPECIFICATIONS 5 5. INSTALLATION

More information

ms NoVent II Instruction Booklet

ms NoVent II Instruction Booklet ms NoVent II Instruction Booklet ms NoVent II The ms NoVent II interface is designed to allow column changing or column maintenance to be carried out without the need to shut down the mass spectrometer.

More information

List of Contents vma. Compressed Air Conditioning 2. System description 2. Individual Units Pre-Filters 3. Combinations 6

List of Contents vma. Compressed Air Conditioning 2. System description 2. Individual Units Pre-Filters 3. Combinations 6 Úprava stlačeného vzduchu EWO VMA List of Contents System description Individual Units Pre-Filters 3 Micro-Filters 4 Activated-Charcoal-Filters 5 Combinations 6 Accessories and Main spare parts 6 Service

More information

LRS(H)4 Pressure-Reducing Regulator User Manual

LRS(H)4 Pressure-Reducing Regulator User Manual LRS(H)4 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be considered

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310

Proudly serving laboratories worldwide since 1979 CALL for Refurbished & Certified Lab Equipment Varian 310 www.ietltd.com Proudly serving laboratories worldwide since 1979 CALL +847.913.0777 for Refurbished & Certified Lab Equipment Varian 310 310 LC/MS Manual System INCLUDES: 310-MS LC/MS triple quadrople

More information

Table of Contents. Instrument Description. Quattro II User's Guide

Table of Contents. Instrument Description. Quattro II User's Guide Instrument Description Table of Contents Introduction 11 The Vacuum System 12 Fine Pumping 12 Rotary Pumping 12 Automatic Control 12 Pressure Measurement 12 Vacuum Protection 12 Ionisation Techniques 14

More information