CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Size: px
Start display at page:

Download "CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin"

Transcription

1 Electronic Control Manual 216 Sensors and Stats Section S Technical Bulletin CDS-2000 Issue Date 0393 CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Introduction 3 Pre-Verification Information 3 Verification 5 Equipment Needed 5 Procedure 6 Calibration 7 Single-Point Calibration 7 Full Calibration 9 Troubleshooting 11 Maintenance and Repair 12 Recommended Maintenance 12 Servicing the Sensor 12 Accessory and Replacement Data Johnson Controls, Inc. 1 Part No Rev.

2 2 S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

3 Introduction When properly installed, the CDS-2000 is accurate to ±100 ppm per year. The calibration should be verified after the sensor has been in operation for one full year and annually thereafter. Whenever an accuracy problem is detected via the verification procedure, the sensor should be recalibrated. Pre-Verification Information Before beginning the verification procedure, note the following: The CDS-2000 should be allowed to warm-up for at least one hour before checking calibration or before calibrating. This will ensure that the infrared sensor has reached an equilibrium temperature thereby delivering maximum calibration accuracy. Recalibration of the device should not be attempted until after 14 days of operation as the output may not have stabilized due to the drift compensation algorithm. This variation should not be confused with drift it is the self-tuning of the algorithm. The microprocessor-based compensation algorithm is based on the average of readings taken during a 30 minute or more unoccupied period (less than 500 ppm) once every 24 hours during the previous 14 days. Readings are taken and stored in non-volatile memory. If the power is removed, the readings will be retained. If after 14 days of operation the CDS-2000 appears to need recalibration, obtain a CDS Calibration Kit, and perform the single-point calibration procedure. S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin 3

4 Modes The CDS-2000 has four modes of operation: Warm-Up Primary Alarm Threshold Adjustment Calibration Notes: If it is ever unclear what mode the sensor is in during an adjustment procedure, refer to the Troubleshooting section of this bulletin. Pushbuttons SW-1 and SW-2, which are used to switch between operating modes, are recessed (Figure 1) on the front of the unit. The pushbuttons can be accessed with a small, non-metallic probe. (A metallic probe could short-circuit the unit.) When SW-1 or SW-2 is pressed, the green LED gives three quick flashes and a momentary dark pause to indicate a mode change. 4 S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

5 Verification Before verifying or calibrating the CDS-2000, make sure that the silicone blanket and metal clips are installed. With the duct-mounted unit (CDS ), the blanket and clips should already be in place. To install the silicone blanket: remove the sensor guard and wrap the blanket around the sensor. Line up the blanket s rubber dimples with the holes on the sensing chamber and press down firmly. Fasten the blanket in place with the metal clips provided. Refer to Figure 1 for proper placement of the blanket on the sensor. Zero Adjustment SW-1 SW-2 Silicone Blanket Regulator with Gage Wiring Access Mounting Clips Green LED Tubing Adapter Alarm Threshold Adjustment 1/4 in. ID Tubing verify.drw Figure 1: Verifying Operation of the CDS-2000 Equipment Needed a digital voltmeter with 10 mvdc resolution over a range of at least 0 to 10 VDC the CDS Calibration Kit which includes: a source of air with a known concentration of CO2 between 400 and 1200 ppm (bottled gas) a regulator with gage (see Figure 1) that can maintain flow at 400 to 800 ml/min. (24 to 48 SCIM) silicone blanket with mounting clips, tubing adapter, and tubing S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin 5

6 Procedure 1. Disconnect the source of power to the CDS Connect a Digital Voltmeter (DVM) between Terminals 1 (-) and 2 (+) at the terminal block on the back of the CDS-2000, and power-up the unit. 3. Allow the CDS-2000 approximately one hour to warm-up. In the first few minutes of the warm-up, the green LED on the front of the unit below the sensor will remain dark. After a few minutes the green LED will flash continuously, which indicates that the unit is in the Primary mode. The unit will be ready for verification once the one-hour warm-up period is over. (The green LED will still be flashing.) While the unit is warming up, check the CO2 concentration of the bottled gas, (from its label) and determine the corresponding voltage level by dividing the ppm value by 200. Example: A 1000 ppm CO2 concentration corresponds to 5 VDC (1000/200). 4. Attach the regulator to the gas bottle, and connect tubing to the regulator outlet. Attach the other end of the tubing to the tubing adapter, and firmly place it into the port extending from the blanket as shown in Figure 1. With the CDS-2000 in the Primary mode, slowly adjust the regulator to flow air from the gas bottle through the tubing and into the sensor. The regulator in the Calibration Kit will deliver 500 ml/min. of gas at 10 PSI. The tank can maintain this flow for two minutes.! CAUTION: To avoid depleting the gas, slowly open the regulator valve. Close the valve as soon as verification is complete. 5. Compare the voltage reading on the DVM to the value determined in Step 3. If the two voltage levels differ by more than ±0.5 volts, proceed to the single-point calibration instructions that follow. 6. Reconnect the flow pickup device if using the duct-mounted CDS sensor. Remove the silicone blanket and metal clips if using the CDS S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

7 Calibration The Calibration mode has two sub-modes: the Single-Point mode, which is only for adjustment of the offset, and the Full Calibration mode, which is for adjustment of the span and offset. The Calibration modes cannot be used to verify settings. To verify the settings, refer to the Verification section. If the alarm threshold needs verification, refer to the Adjusting the Alarm Threshold section in the CDS-2000 Series Carbon Dioxide Sensors Technical Bulletin. Always use the verification procedure in this bulletin to check operation before calibrating the sensor. Single-Point Calibration Equipment Needed Procedure a digital voltmeter with 10 mvdc resolution over a range of at least 0 to 10 VDC a small (0.07 in. or less), flat-blade jeweler s screwdriver a non-metallic probe for accessing the pushbuttons the CDS Calibration Kit, which includes: a source of air with a known concentration of CO2 between 400 and 1200 ppm a regulator that can maintain flow between 400 to 800 ml/min. (24 to 48 SCIM) a calibration blanket with mounting clips, adapter, and tubing To do a single-point calibration, perform the following steps: 1. Disconnect the power supply to the unit. Connect a Digital Voltmeter (DVM) between Terminal 2 (-) and Terminal 1 (+) at the terminal block on the back of the CDS-2000, and power up the unit. 2. Allow the CDS-2000 approximately one hour to warm-up. In the first few minutes of the warm-up, the green LED on the front of the unit below the sensor will remain dark. After a few minutes, the green LED will flash continuously, which indicates that the unit is in the Primary mode. The unit will be ready for calibration once the one-hour, warm-up period is over. (The green LED will still be flashing.) 3. Using a non-metallic probe, press and hold down the recessed SW-1 button, until the green LED gives three quick flashes and goes dark; then, immediately release the button. The sensor is now in the Calibration mode, and the LED remains off. S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin 7

8 4. Press and hold down SW-1 again until the LED gives three quick flashes; then, immediately release the button to switch the unit into the Single-Point Calibration mode. The voltage reading on the DVM will indicate approximately 2 VDC (400 ppm), and the LED remains off.! CAUTION: If the SW-2 button is accidentally pressed at this point, the span setting is permanently erased, and the sensor must be recalibrated using the Full Calibration Mode procedure and bottled gas at 2000 ppm of CO2. 5. Check the CO2 concentration of the bottled gas, and determine the corresponding voltage by dividing the ppm value by 200. (For example, a 1000 ppm CO2 concentration corresponds to 5 VDC (1000/200). Press SW-1 until reading at the DVM corresponds with the value calculated. Note: Repeatedly pressing SW-1 increases the voltage in 0.1 VDC (20 ppm) increments up to 6 VDC (1200 ppm). Continuously holding down SW-1 causes the voltage to increment from 2 to 6 VDC in cycles. If SW-1 is held long enough, the voltage will increment back to 0, and so on. 6. Use tubing to connect the bottled gas to the CDS-2000 as shown in Figure 11. (Refer to the instructions in the Verification Procedure section as needed.) 7. Press SW-2, and slowly adjust the regulator valve (pressure gage at 10 PSI) to flow gas into the sensor. Continue the gas flow until a stable reading is reached at the DVM. The gas will flow at 500 ml/min. for about 2 minutes with the regulator at 10 PSI. 8. When a stable reading is achieved (not necessarily the calculated voltage), press SW-1 to return the unit to the Primary mode. The LED now flashes continuously indicating that the CDS-2000 has resumed operation using the new settings. The reading at the DVM should be the same value (within ±0.1 volts) as that calculated in Step Disconnect the DVM. Reconnect the flow pickup device if using the CDS Remove the silicone blanket and metal clips if using the CDS S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

9 Full Calibration Equipment Needed Procedure a digital voltmeter with 10 mvdc resolution and a range of at least 0 to 10 VDC a small (0.07 in. or less), flat-blade jeweler s screwdriver a small, non-metallic probe for accessing the pushbuttons a source of dry nitrogen gas from which all CO2 has been removed (0 ppm) a source of air with a known concentration of 2000 ppm (±40 ppm) CO2 a regulator that can maintain flow between 400 and 800 ml/min. (24 to 48 SCIM) a silicone blanket and tubing Use the following procedure only when it is necessary to completely reset the sensor s offset and span.! CAUTION: Make sure bottled gas at 2000 ppm CO2 and a bottle at 0 ppm CO2 concentration is available before beginning calibration. The previous span setting is automatically erased when the Full Calibration Mode procedure is initiated. 1. Connect the Digital Voltmeter (DVM) between Terminal 8 (+) and Terminal 2 (-) on the terminal block (back of the unit). Power up the unit. 2. Allow the CDS-2000 approximately one hour to warm-up. In the first few minutes of the warm-up, the green LED on the front of the unit below the sensor will remain dark. After a few minutes, the green LED will flash continuously, which indicates that the unit is in the Primary mode. The unit will be ready for verification once the onehour warm-up period is over. (The green LED will still be flashing.) 3. Using a non-metallic probe, press and hold down the recessed SW-1 button until the green LED gives three quick flashes and goes dark; then, immediately release the button. The unit is now in the Calibration mode, and the LED remains off. 4. Press and hold down the SW-2 button (to the right of SW-1, see Figure 1) until the LED gives three quick flashes and a momentary dark pause; then, immediately release the button to switch the unit into the Full Calibration mode (LED will be flashing continuously). S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin 9

10 5. Use tubing to connect the 0 ppm gas to the CDS-2000 s tubing port shown in Figure 1. (Refer to the instructions in the Verification Procedure section as needed.) Slowly adjust the regulator valve to flow gas into the sensor. Continue flowing the gas until a stable reading is reached at the DVM. (This will take about 2 minutes at 500 ml/min.) When a stable reading is achieved at the DVM, use the small-blade screwdriver to slowly adjust the recessed zero adjustment potentiometer (to the left of SW-1, see Figure 1) until the DVM reads between 2.75 and 2.85 VDC. Turn Clockwise (CW) to decrease or Counterclockwise (CCW) to increase. Note: A 0.2 V, 1 Hz signal is superimposed on the DC level. The 2.8 VDC value should correspond with the highest of the two values, which alternate on the DVM readout. 6. Press the SW-1 button until the LED gives three quick flashes and a momentary dark pause (the LED will then flash continuously). Release the button, and flow 2000 ppm CO 2 gas into the sensor until a stable reading is reached (between 0.4 and 0.8 volts). 7. When a stable reading is achieved (will take about 2 to 3 minutes), press the SW-1 button to return the unit to the Primary mode (LED flashes continuously). If the 8 (+) lead of the DVM is moved to Terminal 1 at this point, the DVM should read between 9.5 and 10.5 VDC. (If not, refer to the Troubleshooting section.) Disconnect the DVM. 8. Reconnect the flow pickup device if using the CDS Remove the silicone blanket and metal clips if using the CDS S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

11 Troubleshooting If the SW-1 and SW-2 buttons are tampered with or are pressed out of sequence during an adjustment or calibration procedure, the CDS-2000 may not operate properly. The flowchart below and the green LED can be used to determine what status/mode the CDS-2000 is in and how to return to the Primary mode. Use it if there is ever any confusion about what mode the sensor is in or to recover from an unsuccessful calibration attempt. IMPORTANT: Once the Single-Point or Full Calibration modes are selected, all previous calibration settings are erased. PRESS SW-2 LED Gives 3 Quick Flashes, Then Stays Dark PRIMARY MODE LED Flashes Continuously PRESS SW-1 Then Stays Dark ALARM THRESHOLD ADJUSTMENT MODE Adjust Threshold Potentiometer to Desired Setting Press SW-2 to Store Desired Setting * Do NOT proceed from this point unless the appropriate supply of bottled gas at the indicated CO 2 concentration is available. Once the Full Calibration Mode Procedure is initiated, the procedure must be completed. SINGLE POINT CALIBRATION Press SW-1 to Select Output Voltage PRESS SW-1 CALIBRATION SELECT MODE Then Stays Dark Press SW-2 After Voltage Selection Flow Gas Press SW-1 After Reading Stabilizes PRESS SW-2 * FULL CALIBRATION Flow 0 ppm Gas and Adjust Zero Potentiometer Press SW-1 Flow 2000 ppm Gas Press SW-1 cdsflow.drw Figure 2: Troubleshooting Guide S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin 11

12 Maintenance and Repair Recommended Maintenance Servicing the Sensor Accessory and Replacement Data To ensure maximum performance from the CDS-2000: keep the air vents on the unit clean and unobstructed verify sensor calibration annually. If necessary, recalibrate as indicated in the Calibration section of this bulletin to prevent inaccuracy due to drift. If the CDS-2000 fails verification and subsequent field adjustment of the calibration settings does not correct the problem, contact your local Johnson Controls branch or distributor office. Note: The CDS-2000 cannot be repaired in the field. Contact your local Johnson Controls representative to arrange for a unit replacement. If a replacement unit is required, specify the complete product code number from Table 1, and contact the nearest Johnson Controls branch office or wholesale distributor. A Single-Point Calibration Kit for the CDS-2000 can also be ordered using Table 1. Table 1: CDS-2000 Ordering Information PRODUCT CODE NUMBER PRODUCT CDS Duct-Mount CO 2 Sensor CDS Wall-Mount CO 2 Sensor CDS Single-Point Calibration Kit DESCRIPTION With round base, ppm CO 2 range, 0-10 VDC output and SPST/NO alarm relay With telephone-style base, ppm CO 2 range, 0-10 VDC output and SPST/NO alarm relay For verification and single-point calibration of CDS-2000 Series Carbon Dioxide Sensors A Air Filter Use with the CDS in duct applications where the flow pickup tube is installed Controls Group FAN E. Michigan Street Electronic Control Manual P.O. Box 423 Issue Date 0393 Milwaukee, WI Printed in USA 12 S Sensors and Stats CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2.

Table of Contents. Sensor Calibration and Troubleshooting CDS4000 CO 2. Introduction 1. Handling Information. Calibration 2. FANs 216, 1628.3 Technical Bulletin CDS4000 Issue Date 0797 CDS4000 CO 2 Sensor Calibration and Troubleshooting Table of Contents Introduction 1 Handling Information 1 Calibration 2 Preparation 2 Cautions

More information

P-5215 Differential Pressure Transmitter

P-5215 Differential Pressure Transmitter P-5215 Differential Pressure Transmitter Pneumatic Control Manual 717.1 Pressure Section Product Bulletin P-5215 Issue Date 0891 Features and Benefits Ultra Sensitive Feedback Circuit - Enhances System

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

User manual CF8-D/W-IN

User manual CF8-D/W-IN User manual CF8-D/W-IN General The sensor CF8-D/W-IN is used to measure the carbon dioxide concentration inside incubators. All functions can be modified from a PC with the communication cable. UMA 48

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

TEL/jlRE" Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes

TEL/jlRE Introduction. Display Features and Modes. Startup Procedure. Power-Up Procedure. Adjustment Modes TEL/jlRE" Introduction The Telaire 7001 CO 2 /T emperature monitor (shown in Fi gt u e 1 below) is an easy to use hand-held instnunent, which provides stable and highly accurate readings due to Telaire

More information

GM Series Dual-Block Multi-Function Gas Control Valves

GM Series Dual-Block Multi-Function Gas Control Valves Installation Sheets Manual 121 Gas Combustion Combination Controls and Systems Section G Technical Bulletin GM Issue Date 0297 GM Series Dual-Block Multi-Function Gas Control Valves Figure 1: GM Series

More information

Model 106 DPI "Micro-switch" Installation and Operating Instructions

Model 106 DPI Micro-switch Installation and Operating Instructions Mid-West Instrument Model 106 DPI "Micro-switch" Installation and Operating Instructions BULLETIN NO. IM116DPImicro/09A Replaces --- INSPECTION Before installation carefully check the Model Number on each

More information

Sensoric 4-20 ma Transmitter Board Operation Manual

Sensoric 4-20 ma Transmitter Board Operation Manual Sensoric 4-20 ma Transmitter Board Operation Manual 1 Content Features Operation Manual Technical Data Mechanical Dimensions Remarks & Contact information 2 Features Soldered sensor cell (non replaceable)

More information

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev

Operation Manual. Pro CO. Carbon Monoxide Analyzer. Rev Operation Manual Pro CO Carbon Monoxide Analyzer Rev. 10.17 Quick Reference Guide READ ENTIRE MANUAL BEFORE USE 1. To switch on, hold the On/Off button until the display powers up. 2. To turn off, hold

More information

Operation Manual. Pro CO 2 Analyzer. Carbon Dioxide Analyzer. Rev

Operation Manual. Pro CO 2 Analyzer. Carbon Dioxide Analyzer. Rev Operation Manual Pro CO 2 Analyzer Carbon Dioxide Analyzer Rev. 10.17 Quick Reference Guide READ ENTIRE MANUAL BEFORE USE 1. To switch on, hold the On/Off button until the display powers up. 2. To turn

More information

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel

TANK MANAGER FOR TWO TANKS OPERATING MANUAL. 10/31/11 C-More T6C L color touch panel TANK MANAGER FOR TWO TANKS OPERATING MANUAL 10/31/11 C-More T6C L color touch panel 1 TABLE OF CONTENTS GENERAL...3 INSTALLATION...4 STONE TEST PROCEDURE...7 OPERATIONAL SUMMARY...7 AUTO CARBONATION...10

More information

HS653 Negative Pressure Ventilation Controller

HS653 Negative Pressure Ventilation Controller HS653 Negative Pressure Ventilation Controller General Installation Notes: Make sure that power is disconnected from system prior to servicing. Installation of this equipment and related OEM equipment

More information

CSC-2000 SERIES. Reset Volume Controllers MADE IN U.S.A. DESCRIPTION MODELS SPECIFICATIONS ORDERING

CSC-2000 SERIES. Reset Volume Controllers MADE IN U.S.A. DESCRIPTION MODELS SPECIFICATIONS ORDERING CSC-2000 SERIES Reset Volume Controllers MADE IN U.S.A. DESCRIPTION The CSC-2000 series are designed for use on VAV terminal units in HVAC systems. These are submaster air velocity controllers whose velocity

More information

Universal Valve Company Inc

Universal Valve Company Inc 1975-FA34 (Air Tower Troubleshooting Manual) Overview This manual has been arranged as a tool for diagnosing, and repairing Universal Air Towers Please read closely to identify the components that are

More information

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V O2100C Oxygen Measurement Module The O2100C module measures the partial pressure of O2 and thus the module output is proportional to the pressure in the sample cell. Gas sampled must be free of liquids

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

Exercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE

Exercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE Exercise 8 Closed-Loop Pressure Control, EXERCISE OBJECTIVE To understand open and closed-loop pressure control; To learn how to sense the pressure in a pneumatic circuit; To control the pressure in a

More information

RK-IR Sample Draw Aspirator Adapter Operator s Manual

RK-IR Sample Draw Aspirator Adapter Operator s Manual 30-0951RK-IR Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0018RK Revision: A Released: 6/2/10 www.rkiinstruments.com Product Warranty RKI Instruments, Inc. warrants gas alarm equipment

More information

CSC-2000 SERIES. Reset Volume Controllers MADE IN U.S.A. DESCRIPTION MODELS SPECIFICATIONS ORDERING

CSC-2000 SERIES. Reset Volume Controllers MADE IN U.S.A. DESCRIPTION MODELS SPECIFICATIONS ORDERING CSC-2000 SERIES Reset Volume Controllers MADE IN U.S.A. DESCRIPTION The CSC-2000 series are designed for use on VAV terminal units in HVAC systems. These are submaster air velocity controllers whose velocity

More information

SL G3 - Fuel Conversion to Propane (NG to LP) P-Kit 302

SL G3 - Fuel Conversion to Propane (NG to LP) P-Kit 302 SL 40-399 G3 - Fuel Conversion to Propane (NG to LP) P-Kit 302 If converting from Propane to Natural Gas, order the Natural Gas Conversion Parts Kit (IBC Part # P-303) from your authorized IBC Distributor.

More information

Manual Weighingblock VB2 series and Uniscale

Manual Weighingblock VB2 series and Uniscale Manual Weighingblock VB2 series and Uniscale Note: At page 8 in this manual you will find a short form instruction. Normally the only instruction shipped together with the Scale. Overview different ranges.

More information

User Manual. asense VAV. CO 2 / temperature sensor with built-in general purpose controller

User Manual. asense VAV. CO 2 / temperature sensor with built-in general purpose controller Gas and Air Sensors User Manual asense VAV CO / temperature sensor with built-in general purpose controller General The IAQ-sensor product asense VAV is used to measure indoor air carbon dioxide concentration

More information

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions L 100 Bubble-Tube Level System Installation, Operation and Maintenance Instructions Figure 1 Contents Section Description Page 1.0 Introduction 2 2.0 Specifications 3 3.0 Installation 3 4.0 Warranty 6

More information

with O 2 Controller Instruction Manual

with O 2 Controller Instruction Manual 14500 Coy Drive, Grass Lake, Michigan 49240 734-475-2200 E-mail: sales@coylab.com www.coylab.com Hypoxic Cabinets (In-Vivo / In-Vitro) with O 2 Controller Instruction Manual Index Page Warranty 2 Warnings

More information

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration

5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration Part No. 84468 5000TOC Sensor Standard Operating Procedure for TOC Calibration Conductivity Calibration Temperature Calibration Flow Rate Calibration This document contains proprietary information, which

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

Do Not Print This Page.

Do Not Print This Page. Do Not Print This Page. Load 1 sheet of cardstock paper. Print Page 2 - the cover For the body of the text, load 3 sheets of 28lb matte finish paper into the printer. Print pages 3, 5, 7 (Sheet 1, 2, 3)

More information

Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display.

Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display. Display Thermo, Inc. Instruction Manual English TP-5 Display TP-7 Display TP-8 1-1 Thermo, Inc. Instructions for use- Models TP-5, TP-7, and TP-8 The manual describes basic function and use of Thermo instruments

More information

SL G3. Warning

SL G3. Warning Important: If converting from Propane to Natural Gas - order the Natural Gas Conversion Parts Kit (IBC Part # P-307) from your authorized IBC Distributor. Use the Propane Gas conversion parts kit if the

More information

1020 Industrial Drive, Orlinda, TN fax

1020 Industrial Drive, Orlinda, TN fax Operation Manual Tank Distribution System A-UPT Series 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.2 Warnings and Cautions... 1 1.2 Theory of Operation...

More information

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP

AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP USER MANUEL AUTOMATIC TIRE INFLATOR # MW-60, MW-60-4WAY & MW-64HP TIRE EQUIPMENT MANUFACTURER 1.866.409.RACK WWW.MARTINSINDUSTRIES.COM info@martinsindustries.com PARTS Verify that the following components

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The HDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

DPC-30 DPC-100. Reference Manual

DPC-30 DPC-100. Reference Manual DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,

More information

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17

Operation Manual. Pro He Alarm TM. Helium Analyzer. Rev 08.17 Operation Manual Pro He Alarm TM Helium Analyzer Rev 08.17 If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: 805-815-4044

More information

Datasheet: K-30 ASCII Sensor

Datasheet: K-30 ASCII Sensor Datasheet: K-30 ASCII Sensor The K30 ASCII sensor is a low cost, infrared and maintenance free transmitter module intended to be built into different host devices that require CO2 monitoring data. The

More information

FAULT CODE TROUBLESHOOTING INDEX

FAULT CODE TROUBLESHOOTING INDEX FAULT CODE TROUBLESHOOTING INDEX 1. Display indicates Change Filters 2. Display indicates Drip Tray Full Continuous Alarm will Sound 3. Display indicates Cold Fault 4. Display indicates Hot Fault 5. Display

More information

Universal Valve Company Inc

Universal Valve Company Inc Universal Valve Company Inc 800-223-0741 www.universalvalve.com 1975-FA34 (Air Tower Troubleshooting Manual) Overview This manual has been arranged as a tool for diagnosing, and repairing Free Air Universal

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Nuclear Associates

Nuclear Associates Nuclear Associates 36-751 XenoGard 133 Xe Monitor Operators Manual March 2005 Manual No. 128011 Rev. 3 2004, 2005 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks

More information

CSA Sample Draw Aspirator Adapter Operator s Manual

CSA Sample Draw Aspirator Adapter Operator s Manual 30-0951-CSA Sample Draw Aspirator Adapter Operator s Manual Part Number: 71-0367 Revision: 0 Released: 4/30/15 www.rkiinstruments.com WARNING Read and understand this instruction manual before operating

More information

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual

Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Air Bubbler Depth Gauge DG2200 Installation and Reference Manual Rev. 3.2.2 06/2011 Installation and Reference Manual Page 2 Contents Chapter 1: Installation and Overview... 3 1.1 DG2200 Location Diagram...

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

ISQB1 INTRINSICALLY SAFE ELECTRO-PNEUMATIC CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS HAZARDOUS AREA CLASSIFICATION

ISQB1 INTRINSICALLY SAFE ELECTRO-PNEUMATIC CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS HAZARDOUS AREA CLASSIFICATION PROPORTION AIR ISQB1 INTRINSICALLY SAFE ELECTRO-PNEUMATIC CONTROL VALVE INSTALLATION & MAINTENANCE INSTRUCTIONS Hazardous area classification Description / identification General specifications Connection

More information

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions

UBEC 1AT. AUTO TANK Fill System Installation, Operation, & Setup Instructions Document Number: XE-ATA5PM-R1A UBEC 1AT AUTO TANK Fill System 08899155 Installation, Operation, & Setup Instructions Rev170906-EB-FRC PHYSICAL: 1302 WEST BEARDSLEY AVE ELKHART, IN 46514 WWW.ELKHARTBRASS.COM

More information

SL G3 - Fuel Conversion to Natural Gas (LP to NG) P-Kit 301

SL G3 - Fuel Conversion to Natural Gas (LP to NG) P-Kit 301 SL 26-260 G3 - Fuel Conversion to Natural Gas (LP to NG) P-Kit 301 Note If converting from Natural Gas to Propane, order the Natural Gas Conversion Parts Kit (IBC Part # P-300) from your authorized IBC

More information

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141

Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST Industrial Drive, Orlinda, TN 37141 0 Operation Manual Concentrate Distribution System (Stand Mounted) A-CDS-70-X & A-CDS-TANK-X-ST 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and

More information

RC 195 Receiver-Controller

RC 195 Receiver-Controller Document No. 129-082 RC 195 Receiver-Controller Product Description The POWERS RC 195 Receiver-Controller is a pneumatic instrument that receives one, two or three pneumatic inputs. It produces a pneumatic

More information

BAPI Pressure Line of Products - FAQs

BAPI Pressure Line of Products - FAQs Table of Contents 1. Several manufacturers produce pressure transmitters, why should I purchase from BAPI?... p. 2 2. BAPI makes several styles of pressure transmitters. What are the features of each?...

More information

G196 Series BASOTROL Redundant Combination Gas Valve with Manual Shutoff Valve

G196 Series BASOTROL Redundant Combination Gas Valve with Manual Shutoff Valve Installation Instructions Issue Date August 19, 2008 G196 Series BASOTROL Redundant Combination Gas Valve with Manual Shutoff Valve Application The G196 valves are suitable for use with natural gas, Liquefied

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0578 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

Operation Manual. Pro CO Analyzer TM. Pro CO with High Temp Alarm Carbon Monoxide Analyzer

Operation Manual. Pro CO Analyzer TM. Pro CO with High Temp Alarm Carbon Monoxide Analyzer Operation Manual Pro CO Analyzer TM & Pro CO with High Temp Alarm Carbon Monoxide Analyzer 02.1.12 If you have any questions on this equipment please contact Technical Support at: Nuvair 2949 West 5 th

More information

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev

Operation Manual. Pro CO O 2 Alarm. Carbon Monoxide and Oxygen Analyzer. Rev Operation Manual Pro CO O 2 Alarm Carbon Monoxide and Oxygen Analyzer Rev. 08.17 Quick Reference Guide for CO Analyzer READ ENTIRE MANUAL BEFORE USE 1. To switch on, hold the On/Off button until the display

More information

Maintenance of the Electronic ( EPC ) Pressure Controllers on GCs

Maintenance of the Electronic ( EPC ) Pressure Controllers on GCs Gas flows in all GCs are controlled by Electronic Pressure Controllers ( EPC ). There may be up to 7 EPCs mounted in one GC depending on which combination of injectors and detectors are installed. Each

More information

CALIBRATION KIT. Instruction Manual

CALIBRATION KIT. Instruction Manual CALIBRATION KIT Instruction Manual 2 CALIBRATION KIT CALIBRATION KIT 3 Table of Contents General description....4 Getting Started....4 GG-NH3....5 GG-NH3-2%....6 GG-CO....7 GG-CO2...8 GG-CL2...9 GG-H2-EC....

More information

G196 Series BASOTROL Redundant Combination Gas Valve with Manual Shutoff Valve

G196 Series BASOTROL Redundant Combination Gas Valve with Manual Shutoff Valve Installation Instructions Issue Date March 13, 2013 G196 Series BASOTROL Redundant Combination Gas Valve with Manual Shutoff Valve Application The G196 valves are suitable for use with natural gas, Liquefied

More information

MOYNO FLUID DETECTION CONTROL SERVICE MANUAL. Section: ACCESSORIES. Page: 1 of 8 Date: May 1, 1998 GENERAL

MOYNO FLUID DETECTION CONTROL SERVICE MANUAL. Section: ACCESSORIES. Page: 1 of 8 Date: May 1, 1998 GENERAL MOYNO Section: ACCESSORIES Page: 1 of 8 Date: May 1, 1998 SERVICE MANUAL FLUID DETECTION CONTROL GENERAL The Moyno Fluid Detection Control prevents unnecessary loss of production time or repair costs to

More information

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE

OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE OEM Manual MODEL 2305 ECONOMICAL DIGITAL SINGLE CYLINDER SCALE 1 These instructions generally describe the installation, operation, and maintenance of subject equipment. The manufacturer reserves the right

More information

BUBBLER CONTROL SYSTEM

BUBBLER CONTROL SYSTEM BUBBLER CONTROL SYSTEM Description: The LDBCS is a fully automatic bubbler system, which does liquid level measurements in water and wastewater applications. It is a dual air compressor system with, air

More information

CF8-W-Disp-CO. User manual. CO 2 / CO sensor with built-in general purpose controller

CF8-W-Disp-CO. User manual. CO 2 / CO sensor with built-in general purpose controller User manual CF8-W-Disp-CO CO 2 / CO sensor with built-in general purpose controller General The IAQ-sensor product CF8-W-Disp-CO is used to measure indoor air carbon dioxide and carbon monoxide concentrations.

More information

Installation, operating and maintenance Instructions for Seemag bypass level indicator

Installation, operating and maintenance Instructions for Seemag bypass level indicator Issue: S Date: 05-09-14 Type G35 General information The Seetru bypass magnetic level indicator, abbreviate SEEMAG, serves to show the filling level of fluids in tanks, basins, tubes etc. The Seemag operates

More information

T-02b2: Set Up/Adjust Drive (DC Full Range GE-Valutr

T-02b2: Set Up/Adjust Drive (DC Full Range GE-Valutr Page 1 of 13 T-02b2: Set Up/Adjust Drive (DC Full Range GE-Valutr alutrol) SAFETY FIRST o Follow all Caterpillar facility safety standards when performing this task. o Dangerous voltage levels of AC and

More information

BI-680 Online Dissolved Oxygen Controller Instruction Manual

BI-680 Online Dissolved Oxygen Controller Instruction Manual BI-680 Online Dissolved Oxygen Controller Instruction Manual BANTE INSTRUMENTS CO., LTD BI-680 Online Dissolved Oxygen Controller 1 Introduction Thank you for selecting the BI-680 online dissolved oxygen

More information

MODEL 269DF-LCF-LCF/WS FILLING SYSTEM

MODEL 269DF-LCF-LCF/WS FILLING SYSTEM MODEL 269DF-LCF-LCF/WS FILLING SYSTEM 29975 Parkway Roseville, Michigan 48066 (586) 778-0426 Fax (586) 778-0479 www.gordinier.com Email: pat@gordinier.com MODEL 269DF-LCF-LCF/WS TABLE OF CONTENTS Receiving

More information

User Manual. asense miii. / CO sensor with built-in general purpose controller

User Manual. asense miii. / CO sensor with built-in general purpose controller Gas and Air Sensors User Manual asense miii CO 2 / CO sensor with built-in general purpose controller General The IAQ-sensor product asense miii is used to measure indoor air carbon dioxide and carbon

More information

G96 Series BASOTROL Dual Operator Valve

G96 Series BASOTROL Dual Operator Valve Installation Instructions 9. Issue Date February 22, 2013 G96 Series BASOTROL Dual Operator Valve Applications The G96 valves are combination, dual operator, automatic valves available with or without

More information

Workhorse-15 Oxygen Generator

Workhorse-15 Oxygen Generator 11436 Sorrento Valley Road, San Diego CA 92121 USA 858-558-0202 Fax 858-558-1915 Workhorse-15 Oxygen Generator Model No. 5719, 5727, 5735 for Domestic and International Units Part number 1432 Revision

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE15C150, VE18C150, VE25C150 PART NO: 2226010, 2226020, 2226025 OPERATION & MAINTENANCE INSTRUCTIONS LS0715 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

! Warning, refer to accompanying documents.

! Warning, refer to accompanying documents. About this Manual To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures accurate and adequate to operate this instrument

More information

OEM Manual. MODEL ½ Digit DRUM SCALE

OEM Manual. MODEL ½ Digit DRUM SCALE OEM Manual MODEL 4020-3 ½ Digit DRUM SCALE Scaletron Industries, Ltd. Bedminster Industrial Park 53 Apple Tree Lane P.O. Box 365 Plumsteadville, PA 18949 USA Toll Free: 1-800-257-5911 (USA & Canada) Phone:

More information

VERTICAL AIR COMPRESSORS

VERTICAL AIR COMPRESSORS VERTICAL AIR COMPRESSORS MODEL NO: VE11C150, VE15C150, VE18C150 PART NO: 2226005, 2226000, 2226015 OPERATION & MAINTENANCE INSTRUCTIONS LS0615 INTRODUCTION Thank you for purchasing this CLARKE Vertical

More information

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic

Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic Copyright 2004 by the Thomas G. Faria Corporation, Uncasville CT No part of this publication may by reproduced in any form, in an electronic retrieval system or otherwise, without the prior written permission

More information

GV Standard X-Vent. Setup, Commissioning & Installation Guide

GV Standard X-Vent. Setup, Commissioning & Installation Guide GV Standard X-Vent Setup, Commissioning & Installation Guide Technical experts in the design, manufacture and supply of precision engineered, architectural rooflights for residential and commercial buildings.

More information

NGP-250/500 Nitrogen Generator Quick Start Guide

NGP-250/500 Nitrogen Generator Quick Start Guide NGP-250/500 Nitrogen Generator Quick Start Guide Version: A July 2013 Potter Electric Signal Company, LLC 5757 Phantom Dr., Suite 125 P. O. Box 42037 Hazelwood, MO 63042 Phone: (314) 595-6900 Document

More information

Transmitter CS 21 Operation Manual

Transmitter CS 21 Operation Manual Transmitter CS 21 Operation Manual 1194 Oak Valley Drive, Suite 20, Ann Arbor, MI 48108 800-959-0573 734-769-1888 Content Page For your Safety 3 General Description 3 Detection Principle 4 Operation 4

More information

Instruction Manual. PLT Series. 1. General Specification. 2. Installation. 3. Wiring. 4. Function. 5. Setting & Calibration

Instruction Manual. PLT Series. 1. General Specification. 2. Installation. 3. Wiring. 4. Function. 5. Setting & Calibration 4 Digit Display, 4-20mA Output Instruction Manual (Digital Differential Pressure Transmitter) 1. General Specification 2. Installation 3. Wiring 4. Function 5. Setting & Calibration 1 4 20 1 Differential

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

NIV EAU MATIC ST User Manual Made in Canada

NIV EAU MATIC ST User Manual Made in Canada NIV EAU MATIC ST User Manual Made in Canada www.niveaumatic.com info@niveaumatic.com -2- Table of contents Introduction Section 1 Page 4 Installation Section 2 Pages 5 to 10 2.1 Control box 2.2 Sensor

More information

QBS Electronic Pressure Regulator

QBS Electronic Pressure Regulator INSTALLATION AND MAINTENANCE INSTRUCTIONS DESCRIPTION / IDENTIFICATION The QBS stainless steel series is an electronically controlled pressure regulator. The QBS controls the pressure on its output port

More information

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM

FOR INSTALLING CO 2 BLENDER KIT (P/N IN BEER SYSTEM IMI CORNELIUS INC One Cornelius Place Anoka, MN 55303-623 Telephone (800) 238-3600 Facsimile (612) 22-326 INSTALLATION INSTRUCTIONS FOR INSTALLING CO 2 BLENDER KIT (P/N 111612000 IN BEER SYSTEM SECONDARY

More information

KING-GAGE. Calibration Instruction Manual. D/P LiquiSeal Purge Control and D/P Purge Control

KING-GAGE. Calibration Instruction Manual. D/P LiquiSeal Purge Control and D/P Purge Control Effective: June 2010 KING-GAGE D/P LiquiSeal Purge Control and D/P Purge Control Calibration Instruction Manual for models 738 D/P LiquiSeal Purge Control and 788 D/P Purge Control D/P LiquiSeal Purge

More information

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

MODEL GENERA PURGE TURBINE INSTR DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM. 275R MODEL 275R POR ORTABLE TURBINE GENERA ENERATOR PURGE GAS ANAL NALYZER INSTR NSTRUCTION MANU ANUAL AL DANGER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.

More information

Over 20,000 Strain Gage Target flowmeters installed since 1952.

Over 20,000 Strain Gage Target flowmeters installed since 1952. Over 20,000 Strain Gage Target flowmeters installed since 1952. Liquid, gases, superheated and saturated steam from -320 to 500 O F and up to 15,000 psi. No frictional moving parts. Designed to withstand

More information

Operating instructions Capacitive sensors KG AC (M18) / / 2010

Operating instructions Capacitive sensors KG AC (M18) / / 2010 Operating instructions Capacitive sensors KG AC (M18) UK 704923 / 00 09 / 2010 Contents 1 Preliminary note 3 2 Safety instructions 3 3 Functions and features 3 3.1 Application examples 4 4 Installation

More information

Aquavar SOLO 2 Frequently Asked Questions

Aquavar SOLO 2 Frequently Asked Questions Aquavar SOLO 2 Frequently Asked Questions How do I size the Aquavar SOLO 2 for the appropriate pump/motor combination? Can I use a 208 Volt motor? Can I run the Aquavar SOLO 2 up to 80HZ? What are the

More information

O3 3E 1 F Gas Sensor Module

O3 3E 1 F Gas Sensor Module Product Information Pack O3 3E 1 F Gas Sensor Module (Ozone) CONTENTS Product Data Sheet Product Specification 2 Poisoning and Cross Sensitivities 3 Operating Instructions Introduction 4 Electrostatic

More information

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS APPLICATION NOTE LC-136 Introduction Specialized spaces often times must

More information

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000)

Operating Manual. Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000) Models AQT-02 & AQT-COO2 Manual No. MON004 (Rev 2 April 2000) Operating Manual AIR SYSTEMS INTERNATIONAL, INC. 829 Juniper Crescent, Chesapeake, Va., 23320 Telephone (757) 424-3967 Toll Free 1-800-866-8100

More information

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17

Operation Manual. Pro 4 Warn Analyzer TM Rev 08.17 Operation Manual Pro 4 Warn Analyzer TM Rev 08.17 If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: 805-815-4044 FAX: 805-486-0900

More information

OPERATION MANUAL ID900

OPERATION MANUAL ID900 OPERATION MANUAL ID900 Portable Refrigerant Identifier WARNING Liquid Samples will Damage Unit Do not operate if filter has red flecks. Replace filter every 3 months or after every 25 uses. Always have

More information

Operation Manual. Rev. C LPT Low Power Transmitter (Electrochemical Sensor)

Operation Manual. Rev. C LPT Low Power Transmitter (Electrochemical Sensor) Operation Manual Rev. C 2014.04 LPT Low Power Transmitter (Electrochemical Sensor) www.critical-environment.com LPT - Operation Manual Rev. C 2014.04 TABLE OF CONTENTS 1 POLICIES...4 1.1 Important Note...4

More information

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark

ECHO MANUAL WARNING. L B A ltim e te rs. ECHO is a trademark of LB Altimeters, Denmark ECHO MANUAL L B A ltim e te rs ECHO is a trademark of LB Altimeters, Denmark LB Altimeters operates a policy of continuous development Therefore, we reserve the right to make changes and improvements to

More information

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen

More information

T i m i n g S y s t e m s. RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA (408)

T i m i n g S y s t e m s. RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA (408) RACEAMERICA T i m i n g S y s t e m s Demo Tree Controller Owner s Manual Models 3204D, 3204DW & 3204DX Rev D RACEAMERICA, Inc. P.O. Box 3469 Santa Clara, CA 95055-3469 (408) 988-6188 http://www.raceamerica.com

More information

660 Multipoint Electronic Level Switch

660 Multipoint Electronic Level Switch 660 Multipoint Electronic Level Switch General Instructions The Multipoint Electronic Level Switch is a level sensing device which reads process level by capacitance measurement. Capacitance varies according

More information

Rejuvenation Instructions

Rejuvenation Instructions Rejuvenation Instructions #401 Air Systems UPR This NRI covers the following: Understanding the applications and operation of flow meters. Understand the application and operation of test pressure gauges.

More information