Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display.

Size: px
Start display at page:

Download "Thermo Probe, Inc. TP-5 TP-7 TP-8. Instruction Manual English. Switch. Display. Carry Strap. Probe. Ground Clip. Switch. Display."

Transcription

1 Display Thermo, Inc. Instruction Manual English TP-5 Display TP-7 Display TP-8 1-1

2 Thermo, Inc. Instructions for use- Models TP-5, TP-7, and TP-8 The manual describes basic function and use of Thermo instruments and basic safety instructions. The Thermo Portable Electronic Thermometer is a portable digital temperature instrument. Intended use of Thermo is both in hazardous (flammable) and non-hazardous areas. Typical areas for use: a) Tankfarms or tanks containing oil or chemical products. b) Refineries, chemical plants, hot-water producing plants, etc. c) Marine transportations such as; tankers, barges or other transport facilities. d) Food processing plants. e) Other industrial plants where temperature measuring is needed. FUNCTION Thermo case enclosures are built upon knowledge from previous models and constructed to withstand the rigors of field use. Measuring cable is made with a Teflon FEP or PFA outer layer with carbon blended to reduce static charge. DuPont Kevlar reinforcement is included in cables for select models in order to resist cuts. Cable and sensor probe storage is integral with the case. Electronic circuits are placed inside the case, well protected. To reach the electronic circuits and battery, the faceplate, side-plates, or battery cover must be removed depending on model. Measured temperature is displayed via Liquid Crystal Display (LCD) visible through a window on the case. Electronic is CMOS technology, and powered by one non-rechargeable 9 volt battery. Cable is connected at one end to electronic circuits and to the probe containing a solid-state type chip or Platinum Resistive Temperature Device (RTD) on TP-7 and TP-8. Depress the sealed ON button on the case to read measured temperature on the display. For use in hazardous areas a grounding cable is included with the instrument. Also a carrying strap is enclosed for easy carrying. USE OF INSTRUMENT As Thermos are intended to be used in hazardous locations, user must have a working knowledge of appropriate safety instructions. SEE MARKINGS ON INSTRUMENT AND READ FOLLOWING INSTRUCTIONS PROCEDURES BEFORE USE a) The inspector must have a thorough knowledge of the products to be measured and must know of the safety measures to be taken when working with the material to be measured. b) Instrument shall be checked concerning severe defects; check that instrument is complete (including grounding cable), has good batteries, etc. If necessary, check measurement accuracy. If any defects are found, the instrument should not be used until repairs have been made. c) Instrument, especially cable and probe, should be clean both for safety reasons and ease of use. d) Estimation should be carried out to validate if physical place for measuring is considered a primary or secondary, or both, risk source. MEASURING PROCEDURES a) Check wind direction or other air-movements at place for measuring to avoid inhaling vapors. b) Attach grounding cable/bonding cable to bare metal grounding facilities associated with the vessel. c) Lower cable and probe into intended product at desired depth. 1-2

3 d) Press the ON button. Raise and lower the sensor about 1 foot in each direction to decrease the time for temperature stabilization. TP-7 and TP-8 units will turn off automatically after 1 minute, to save battery charge (Press OFF button on model TP-5). e) Wipe the cable and probe clean while winding up so that it will be clean for the next use. f) Remove grounding cable from grounding facilities and wrap around instrument to store. STANDARDS SIGNIFICATION Thermo petroleum gauging instruments are classified as intrinsically safe to North American, European and Japanese standards. Depending on the particular instrument and regional requirements, certifications are provided by the following testing laboratories: UL (US), KEMA (Denmark), NEMKO (Norway) and TIIS (Japan). Please check instrument identification plates for necessary designation. Safety Classifications: North America: Class 1, Division I, Groups CD - Model TP-5 North America: Class 1, Division I, Groups ABCD - Model TP-7,8 Europe: EExia IIB T3 Model TP-5 Europe EExia IIB T3/T4/T6 Model TP-7,8 Japan: EExia IIB T4 Model TP-7,8 Applicable Standards are: UL 913 Intrinsic Safety CSA 22.2 Intrinsic Safety CENELEC EN50014 Intrinsic Safety CENELEC EN50020 Intrinsic Safety CENELEC EN50081 EMC Emissions CENELEC EN55022 EMC Immunity Japan uses CENELEC Standards Agency or Safety Designation US: UL Canada: ULc Europe: Ex Europe: Ex Europe: CE Europe: CE Japan: Only Text Designation STANDARDS SIGNIFICATION Intrinsic Safety Standard UL913 is written by Underwriters Laboratories with reference to NEC (National Electric Code) and NFPA (National Fire Protection Association). This standard addresses the use of devices in Hazardous Locations (Flammable) and is recognized in the US and other parts of North America. UL normally uses this standard for safety evaluations. Intrinsic Safety Standard CSA 22.2 is written by Canadian Standards Association. This standard addresses the use of devices in Hazardous Locations (Flammable) and is recognized in the Canada and other parts of North America. UL also does safety evaluations to this standard. Intrinsic Safety Standards EN50014 and EN50020 are written by the European standards organization CENELEC, and these standards are recognized by European countries, Japan, and other countries as well. These standards address the use of devices in Zone 0 locations (Flammable). KEMA, NEMKO and TIIS use these standards for their evaluations. INTRINSIC SAFETY Intrinsically safe equipment is defined as "equipment and wiring which is incapable of releasing sufficient electrical or thermal energy under normal or abnormal conditions to cause ignition of a specific hazardous atmospheric mixture in its most easily ignited concentration." (ISA-RP12.6) This is achieved by limiting the amount of power available to the electrical equipment in the hazardous area to a level below that which will ignite the gases. In order to have a fire or explosion, fuel, oxygen and a source of ignition must be present. An intrinsically safe system assumes the fuel and oxygen is present in the atmosphere, but the system is designed so the electrical energy or thermal energy of a particular instrument loop can never be great enough to cause ignition. 1-3

4 ADDITIONAL SAFETY INSTRUCTIONS a) Battery must be changed in Non-hazardous area. Battery must be of correct type, see instrument identification plates on instrument. b) Exchange of components other than original components may compromise the safety and is not recommended. BATTERY CHECK Thermo models TP-7 and TP-8 will indicate a flashing readout of Lo in the display. Replace battery if this occurs in order to prevent inaccurate readings. TP-5 model battery is considered good as long as a measured value appears on the display when pressing the ON button. However, if backlight will not light when in darkness or if measured value is unstable, battery replacement is recommended. REPLACING BATTERY a) Move into non-hazardous area. b) Remove the four screws holding the battery cover on the TP-5 and TP-8 models, or the three faceplate screws on the TP-7 model. c) Slide the battery cover out on the TP-5 and TP-8. Remove the circuit board/faceplate assembly on the TP-7. d) Lift battery at the outer end and pull the battery away from the plus/minus connections. e) Mount the new battery by pressing appropriate (+)/(-) connections together and push the battery to its correct position. f) (TP-5 and TP-8) Mount the battery cover and fasten the four screws. (TP-7) Reinstall the circuit assembly and fasten the three screws. CALIBRATION PROCEDURE Refer to API 7.3 for Calibration Verification Procedures: Calibration is not to be performed in any environment considered to be hazardous. Thermos should be calibrated at least annually to keep up to the standard of high accuracy. Please check for drift of electronic circuit, against proper temperature stable bath. Calibration requires two precision temperature reference points, for example; 0.0 C (32.0 F) and 90.0 C (194 F) in stable liquid calibration baths or similar certified temperature calibration equipment. Equipment needed: Ice Bath with certified reference thermometer. Warm to hot fluid bath between at 20 C (approx 68 F) or higher up to 90 C (approx 194 F) with certified reference thermometer. (see Note)* Small Phillips head or standard blade screwdriver to gain access to circuit board or calibration adjusters. Micro-style bladed screwdriver for electronics adjustment or similar tiny flat blade screwdriver for turning the adjusters. *Note for field calibration: If entire range capability of instrument is not required, the high adjustment can be made at a temperature relatively close to the common temperature of the liquid measured and accuracy will be maintained within the limited range. For example: If liquid to be measured is commonly less than 38 C (approx. 100 F), then a span or high point calibration can be made with a stable bath at room temperature of 21 C (70 F) to 27 C (80 F) using a certified thermometer for reference. Measurement accuracy will be maintained as long as liquid temperature to be measured does not exceed approx. 40 C (105 F). 1-4

5 To calibrate proceed with the following steps: The following instructions specify the TP-5, TP-7 and TP-8, other models are similar. **Refer to Appendix A for special instruction variation for Thermo model TP-2. a) For model TP-5, remove the two screws covering zero and span adjustment screws on the outside of front side of the case. The zero is the lower most adjuster and the span is located directly above. In order to access the adjustment screws on the TP-7 remove the circuit board assembly from case as described in Replacing Battery section above. TP-8 adjusting screw access requires removal of bottom cover plate by removing four screws. The adjusters on the TP-7 and TP-8 are found side by side on the lower edge of the circuit board. Refer to markings on the circuit board to verify which is zero and which is span. b) Lower the probe into the 0.0 C temperature bath and read temperature on the Thermo display. If deviation occurs, adjust the zero adjustment screw to get the same temperature as certifiable bath temperature. c) Lower the probe into the 90.0 C temperature stable bath and read temperature on the display. If deviation occurs, adjust the span adjustment screw to the same temperature as the certifiable bath temperature. d) Repeat step b) and step c) until same accordance between the Thermo and both bath temperatures. Span is now set to 90.0 C. e) Replace the covering screw on zero and span places on the TP-5 or reassemble the TP-7 or TP-8. APPENDIX A. The following refers to the special calibration procedure for the high zero adjustment on the TP- 2 model. This model is currently out of production. Many of these instruments remain in the field and these instructions are provided as a courtesy in order to assist in proper calibration procedures. The TP-2 differs from all other Thermo models due to its having three adjusters rather than two. The third adjuster (called the high Zero ) is used to correct calibration on Fahrenheit models only and only for temperatures over 200 F. Remove bottom cover to access the three adjusters. Follow the calibration procedures listed in the body of this document for temps under 200 F. After the zero and span have been set, then place the probe sensor in a heated circulating bath at between F and F. Allow the reading to stabilize. Flip the range switch on the right side front of the TP-2 so that it is in the up position over 200 F marking. Adjust the high zero adjuster until the display reads according to the following chart and then reassemble the instrument: High Range Table for TP-2 Bath Temp. F High Zero Reading

DTG - LCD Digital Temperature Gauge USER MANUAL

DTG - LCD Digital Temperature Gauge USER MANUAL DTG - LCD USER MANUAL Page 1 of 7 USERS GUIDE page 1. Description 3 2. Function 3 3. Safety Instruction 3 3.1. Safety Conventions 3 3.2. Proper Use 3 4. Installation 4 4.1. Unpacking 4 4.2. Storage 4 4.3.

More information

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin

CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Electronic Control Manual 216 Sensors and Stats Section S Technical Bulletin CDS-2000 Issue Date 0393 CDS-2000 CO 2 Sensor Verification, Calibration, and Troubleshooting Bulletin Introduction 3 Pre-Verification

More information

Instruction Manual for Intrinsically Safe - Reed Switches

Instruction Manual for Intrinsically Safe - Reed Switches Instruction Manual for Intrinsically Safe - Reed Switches Option Code: -1S2-IS-LED or -2S2-IS-LED -1S2-IS or -2S2-IS ATEX RCM Industries, Inc. Direct Reading Differential Pressure Flowmeters 110 Mason

More information

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact

More information

High-performance submersible pressure transmitter For level measurement Model LH-10

High-performance submersible pressure transmitter For level measurement Model LH-10 Electronic pressure measurement High-performance submersible pressure transmitter For level measurement Model LH-10 WIKA data sheet PE 81.09 Applications Level measurement in rivers and lakes Deep well

More information

1.8 INDUSTRIAL PROCESS WEIGHING IN HAZARDOUS AREAS

1.8 INDUSTRIAL PROCESS WEIGHING IN HAZARDOUS AREAS 1.8 INDUSTRIAL PROCESS WEIGHING IN HAZARDOUS AREAS EXPLOSION PROTECTION In addition to the type approval and certification of industrial weighing systems concerned with accuracy, equipment that is also

More information

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual

OxyCheq Expedition-X Oxygen Analyzer. Operator s Manual OxyCheq Expedition-X Oxygen Analyzer Operator s Manual OxyCheq 3528 Russell Road Marianna, Florida 32446 USA Tel: 850.482.0385 Email: oxy@oxycheq.com Web: http://oxycheq.com CONTENTS 1.0 Introduction..

More information

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A

More information

Calibration Requirements for Direct Reading Confined Space Gas Detectors

Calibration Requirements for Direct Reading Confined Space Gas Detectors : Calibration Requirements for Direct Reading Confined Space Gas Detectors However, the definition of bump test has always been a little slippery. Some manufacturers differentiate between a bump test that

More information

How to specify a product. Process Sensors and Mechanical Instruments

How to specify a product. Process Sensors and Mechanical Instruments How to specify a product Process Sensors and Mechanical Instruments Keep the overview. Here is some guideline information on how to specify our products. Intended as supplementary help to specification

More information

BA307C & BA308C Intrinsically safe loop-powered 3½ digit panel mounting indicators issue 9

BA307C & BA308C Intrinsically safe loop-powered 3½ digit panel mounting indicators issue 9 BA307C & BA308C Intrinsically safe loop-powered 3½ digit panel mounting indicators issue 9 Issue: 9 22 nd March 2013 2 CONTENTS 1. Description 2. Operation 3. Intrinsic Safety Certification 3.1 ATEX certificate

More information

API MPMS Chapter 17.6 Guidelines for Determining the Fullness of Pipelines between Vessels and Shore Tanks

API MPMS Chapter 17.6 Guidelines for Determining the Fullness of Pipelines between Vessels and Shore Tanks API MPMS Chapter 17.6 Guidelines for Determining the Fullness of Pipelines between Vessels and Shore Tanks 1. Scope This document describes procedures for determining or confirming the fill condition of

More information

Intrinsically Safe Acoustic Instrument Used in Troubleshooting Gas Lift Wells

Intrinsically Safe Acoustic Instrument Used in Troubleshooting Gas Lift Wells Gas Well Deliquification Workshop Sheraton Hotel, Denver, Colorado February 17 20, 2013 Intrinsically Safe Acoustic Instrument Used in Troubleshooting Gas Lift Wells Lynn Rowlan, Echometer Company Carrie-Anne

More information

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure

1 Overview. Pressure Measurement Transmitters for basic requirements. 1/16 Siemens FI SITRANS P220 for gauge pressure Siemens AG 204 Overview The pressure transmitter SITRANS P220 measures the gauge pressure of liquids, gases and vapors. Stainless steel measuring cell, fully welded Measuring ranges 2.5 to 600 bar (36.3

More information

High-performance submersible pressure transmitter For level measurement Model LH-10

High-performance submersible pressure transmitter For level measurement Model LH-10 Electronic pressure measurement High-performance submersible pressure transmitter For level measurement Model LH-10 WIKA data sheet PE 81.09 Applications Level measurement in rivers and lakes Deep well

More information

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C) Product Data Sheet PDS 106-581.A01 March, 2004 Model 5081FG Two-Wire In Situ Oxygen Analyzer (550 to 1400 C) Intrinsically safe: - CENELEC EEx ia IIC - Class I Div. I Gr. B, C, D (pending) Operates at

More information

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model

User's Manual. Heavy Duty Dissolved Oxygen Meter. Model User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Introduction Congratulations on your purchase of Extech's Heavy Duty Dissolved Oxygen / Temperature Meter which simultaneously displays Dissolved

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

780S Ultra High Purity Series

780S Ultra High Purity Series Smart Ultra High Purity Thermal Gas Mass Flow Meter Features Measures mass flow directly, no seperate temperature or pressure inputs required Field adjustment of critical flow meter settings via on-board

More information

TC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL

TC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL TC65M/15M TEMPERATURE CALIBRATOR USER S MANUAL Thank you for purchasing a Scan-Sense temperature calibrator. The products are manufactured by Scan-Sense AS in accordance with our high quality standards

More information

Hand-held pressure calibrator with integrated pump Model CPH6600

Hand-held pressure calibrator with integrated pump Model CPH6600 Calibration technology Hand-held pressure calibrator with integrated pump Model CPH6600 WIKA data sheet CT 16.01 Applications Calibration service companies and service industry Measurement and control

More information

DM01. Please visit our website: Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

DM01. Please visit our website:  Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05 DM0 Battery Powered Stainless Steel Sensor class 0.05 Nominal pressure from 0 00 mbar up to 0... 00 bar Special characteristics modular sensor concept data logger graphic display stainless steel housing

More information

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. Type: ETd 04. class 0.05

ETd 04. Battery Powered Precision Digital Gauge. Stainless Steel Sensor.  Type: ETd 04. class 0.05 Battery Powered Stainless Steel Sensor class 0.05 Nominal pressure from 0 100 mbar up to 0... 400 bar Special characteristics modular sensor concept data logger graphic display stainless steel housing

More information

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions

L 100. Bubble-Tube Level System. Installation, Operation and Maintenance Instructions L 100 Bubble-Tube Level System Installation, Operation and Maintenance Instructions Figure 1 Contents Section Description Page 1.0 Introduction 2 2.0 Specifications 3 3.0 Installation 3 4.0 Warranty 6

More information

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet:

OXY Integral. INTERCON ENTERPRISES INC Tel: Fax: Internet: OXY Integral INTERCON ENTERPRISES INC Tel: 800 665 6655 Fax: 604 946 5340 E-Mail: sales@intercononline.com Internet: www.intercononline.com Manual Integral 2006 1 INDEX 2-3 PREFACE 4 INTRODUCTION 5 Principle

More information

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20

High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20 Electronic pressure measurement High-performance submersible pressure transmitter For level measurement in hazardous areas Model LH-20 WIKA data sheet PE 81.56 Applications Deep well and borehole measurements

More information

Pressure switch KP-E. Data sheet

Pressure switch KP-E. Data sheet Data sheet Pressure switch KP-E KP-E pressure switches for use in refrigeration and air conditioning systems are equipped with SPDT gold plated contacts and stainless steel bellows. The high pressure switches

More information

Heavy Duty Dissolved Oxygen Meter

Heavy Duty Dissolved Oxygen Meter User's Manual Heavy Duty Dissolved Oxygen Meter Model 407510 Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 FAX 781.665.0780 - TestEquipmentDepot.com Introduction Congratulations

More information

I/P Transducer Positioner Module

I/P Transducer Positioner Module I/P Transducer Positioner Module VRC P/N: 7958032 Installation, Operation and Maintenance Instructions 2.09 [53.1] 1.30 [32.9] 2.97 [75.5] 0.78 [19.8] 0.18 [4.4] 1.42 [36.1] n.18 MOUNTING HOLES (2) PLACES

More information

Oxford. Manufacturer s Contact Details. Contents. Oxford Low Profile Weigh Scale. English

Oxford. Manufacturer s Contact Details. Contents. Oxford Low Profile Weigh Scale. English Manufacturer s Contact Details EUROPE Joerns Healthcare Limited Drakes Broughton Business Park, Worcester Road Drakes Broughton, Pershore, Worcestershire WR10 2AG United Kingdom Tel: 0844 811 1156 Fax:

More information

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05

DM01. Battery Powered Precision Digital Gauge. Stainless Steel Sensor. class 0.05 Battery Powered Stainless Steel Sensor class 0.05 Nominal pressure from 0 00 mbar up to 0... 00 bar Special characteristics modular sensor concept data logger graphic display stainless steel housing Ø

More information

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in:

XMO2. Panametrics Smart Oxygen Analyzer. Applications. Features. bhge.com. An oxygen transmitter for use in: XMO2 Panametrics Smart Oxygen Analyzer Applications An oxygen transmitter for use in: Inerting/blanketing liquid storage tanks Reactor feed gases Centrifuge gases Catalyst regeneration Solvent recovery

More information

DPC-30 DPC-100. Reference Manual

DPC-30 DPC-100. Reference Manual DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,

More information

Technical Data Sheet MF010-O-LC

Technical Data Sheet MF010-O-LC Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the

More information

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex.

Calculating Total Uncertainty of Temperature Calibration with a Dry Block. Calibration White Paper. Beamex. Beamex Calibration White Paper info@beamex.com Calculating Total Uncertainty of Temperature Calibration with a Dry Block World-class calibration solutions Calculating Total Uncertainty of Temperature Calibration

More information

How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit:

How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit: How to calibrate the formaldehyde sensor within a PPMonitor Wireless Unit: Please read this document thoroughly before attempting to check or adjust the calibration of the formaldehyde sensor within the

More information

PM105 PRESSURE METER USER S MANUAL

PM105 PRESSURE METER USER S MANUAL PM105 PRESSURE METER USER S MANUAL Manual revision: 15 March, 2001 SCSMANP015 Copyrighted 2001 Thank you for purchasing a Tek Know pressure meter. The Tek Know products are manufactured by Scan-Sense AS

More information

Gas density monitor With integrated transmitter Model GDM-100-TI

Gas density monitor With integrated transmitter Model GDM-100-TI SF 6 gas solutions Gas density monitor With integrated transmitter Model GDM-100-TI grid Products WIKA data sheet SP 60.05 for further approvals see page 5 Applications Gas density monitoring of closed

More information

Installation, Operation, & Maintenance Manual. MODEL: Watlow Controller P53 Direct, P53 Remote

Installation, Operation, & Maintenance Manual. MODEL: Watlow Controller P53 Direct, P53 Remote Installation, Operation, & Maintenance Manual Mustang Sample Conditioning System MODEL: Watlow Controller P53 Direct, P53 Remote Conforms to UL STDS 61010-1 & 1203 Certified to CSA STD C22.2 Nos. 61010-1

More information

Advanced Test Equipment Rentals ATEC (2832)

Advanced Test Equipment Rentals ATEC (2832) Established 1981 Advanced Test Equipment Rentals www.atecorp.com 800-404-ATEC (2832) DOCUMENT NO: STP 2203 VERSION: 2008-07-01 Page 1 of 6 1.0 PURPOSE This test procedure is used by the Electrical Safety

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

9700 Transmitter Submersible hydrostatic level transmitter

9700 Transmitter Submersible hydrostatic level transmitter Specification sheet IP0078 October 2007 Level 9700 Transmitter Submersible hydrostatic level transmitter Description The 9700 series hydrostatic level transmitter is designed to perform in the most arduous

More information

BAPI Pressure Line of Products - FAQs

BAPI Pressure Line of Products - FAQs Table of Contents 1. Several manufacturers produce pressure transmitters, why should I purchase from BAPI?... p. 2 2. BAPI makes several styles of pressure transmitters. What are the features of each?...

More information

GEMS SENSORS & CONTROLS

GEMS SENSORS & CONTROLS GEMS SENSORS & CONTROLS OPERATING & INSTALLATION INSTRUCTIONS 31XX / 32XX SERIES PLEASE READ CAREFULLY BEFORE INSTALLING Part Number: 560550-0076 Issue: E INTRODUCTION Series 3100/3200 high output pressure

More information

HOT OILING OPERATIONS ALL HSE PRC 172. Approved By: Manager, HSE Performance Assurance. Table of Contents

HOT OILING OPERATIONS ALL HSE PRC 172. Approved By: Manager, HSE Performance Assurance. Table of Contents Owner: HSE Performance Assurance ALL HSE PRC 172 Approved By: Manager, HSE Performance Assurance Retention Code: CG01 CA Revised: March 2015 Review Frequency: Five years or less Table of Contents 1.0 Purpose...

More information

EC214 EC215 - EC215R Bench Conductivity Meters

EC214 EC215 - EC215R Bench Conductivity Meters Instruction Manual EC214 EC215 - EC215R Bench Conductivity Meters http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments

More information

Temperature gauge Installation Guide

Temperature gauge Installation Guide Temperature gauge Installation Guide improper functioning, or damage. Additional information can be found at: CAUTION: Read this installation guide carefully before unpacking the temperature gauge. Improper

More information

Installation Instructions

Installation Instructions Model: ACT-4100 ACT-4100-3 ACT-4100-6 SUN ELECTRIC CORPORATION Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation

More information

F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs

F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs F&J SPECIALTY PRODUCTS The nucleus of quality air monitoring programs AIR FLOW CALIBRATORS FOR MOST AIR MONITORING PROGRAMS World Calibrator PC Version World Calibrator VFD Version Compact Digital Calibrator

More information

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50

Operation manual Level sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50 Operation Manual Level Sensor DC-LS-50 Rev.02 (2014.06) OL Page 1 of 8 Table of contents I. STRUCTURE OF THE MANUAL / CLARIFICATION... 3 II. SAFETY AND HEALTH CONCERNS... 4 1 INTRODUCTION... 5 1.1 Use

More information

User s Guide Temperature Sensor Converter TSC-599

User s Guide Temperature Sensor Converter TSC-599 User s Guide Temperature Sensor Converter TSC-599 ILX Lightwave Corporation 31950 Frontage Road Bozeman, MT, U.S.A. 59715 U.S. & Canada: 1-800-459-9459 International Inquiries: 406-556-2481 Fax 406-586-9405

More information

Wireless I/O, Connecting Remote Sensors in a Wide Range of Environments

Wireless I/O, Connecting Remote Sensors in a Wide Range of Environments Gas Well Deliquification Workshop Sheraton Hotel, February 19 22, 2012 Wireless I/O, Connecting Remote Sensors in a Wide Range of Environments Glenn Longley, Sr. Product Manager FreeWave Technologies Agenda

More information

Periodic Survey of Fuel Installations on Ships other than Liquefied Gas Carriers utilizing gas or other low flash point fuels

Periodic Survey of Fuel Installations on Ships other than Liquefied Gas Carriers utilizing gas or other low flash point fuels (Jan 2017) Periodic Survey of Fuel Installations on Ships other than Liquefied Gas Carriers utilizing gas or other low flash point fuels CONTENTS 1. Application 2. Special Survey 2.1 Schedule 2.2 Scope

More information

Model Description

Model Description LEAKHUNTER Plus LEAKHUNTER Plus Model 8066 Description The MATHESON LEAKHUNTER Plus is a truly universal leak detector. Extensive R&D has produced a multi-functional leak detector engineered to perform

More information

FLAMMABLE GAS SENSOR/TRANSMITTER

FLAMMABLE GAS SENSOR/TRANSMITTER FLAMMABLE GAS SENSOR/TRANSMITTER Plant safety protection for flammable gases. F E A T U R E S Fixed gas monitoring for point source hazards and perimeter protection in arduous duty and exposed locations.

More information

Mobrey 9700 Transmitter Submersible Hydrostatic Level Transmitters

Mobrey 9700 Transmitter Submersible Hydrostatic Level Transmitters Product Data Sheet Mobrey 9700 Transmitter Mobrey 9700 Transmitter Submersible Hydrostatic Level Transmitters Hydrostatic electronic level transmitters Factory sealed and tested for submersed duty 4 20

More information

North American Pressure Switches

North American Pressure Switches Combustion North American Pressure Switches Bulletin 8757 Pressure switches are used with Automatic Shutoff Valves or Solenoid Valves to interrupt fuel flow in the event of fuel or air failure. They are

More information

This test shall be carried out on all vehicles equipped with open type traction batteries.

This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following

More information

TECHNICAL NOTE. October 23, Procedures for Verifying Temperature Calibration

TECHNICAL NOTE. October 23, Procedures for Verifying Temperature Calibration TECHNICAL NOTE October 23, 2002 SUBJECT: Procedures for Verifying Temperature Calibration PROBLEM: DATATRACE users frequently need a procedure to verify that the Tracer calibration is in specification.

More information

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING

PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING PULSAR 5000 SERIES OPERATING & INSTALLATION INSTRUCTIONS SERIES 5000 PLEASE READ CAREFULLY BEFORE INSTALLING Please Note: Ranges above 500mbar are designed and manufactured in accordance with sound engineering

More information

Installation and Operation Manual

Installation and Operation Manual Installation and Operation Manual WIKA FLR-SBDF / BLR-SBDF Magnetic Level Transmitter (Please retain for future usage) Contact: Gayesco-WIKA USA, L.P. 229 Beltway Green Boulevard Pasadena, TX 77503 www.wika.com

More information

5 Function Indicator. Outside Air Temperature (C) Outside Air Temperature (F) Pressure Altitude Density Altitude Aircraft Voltage STD TEMP SL 15000

5 Function Indicator. Outside Air Temperature (C) Outside Air Temperature (F) Pressure Altitude Density Altitude Aircraft Voltage STD TEMP SL 15000 5 Function Indicator +20 +40-5000 STD TEMP SL 15000 Outside Air Temperature (C) Outside Air Temperature (F) Pressure Altitude Density Altitude Aircraft Voltage 427 HILLCREST WAY REDWOOD CITY, CA 94062

More information

The American Society for Nondestructive Testing

The American Society for Nondestructive Testing the NDT Technician FOCUS Vby Terry Tamutus* and Samuel J. Ternowchek Valve leaks are a common problem for refineries, processing plants and power generation plants. Studies conducted by several companies

More information

High-performance submersible pressure transmitter For level measurement Model LH-20

High-performance submersible pressure transmitter For level measurement Model LH-20 Electronic pressure measurement High-performance submersible pressure transmitter For level measurement Model LH-20 WIKA data sheet PE 81.56 Applications Deep well and borehole measurements Groundwater

More information

GAS ANALYSER OXYBABY M+ for O 2. or O 2 /CO 2. standard version

GAS ANALYSER OXYBABY M+ for O 2. or O 2 /CO 2. standard version GAS ANALYSER OXYBABY M+ for O 2 or O 2 standard version Cordless hand held oxygen or combined oxygen and carbon dioxide analyser for checking modified atmospheres in food packs. The ideal instrument for

More information

SITRANS P measuring instruments for pressure

SITRANS P measuring instruments for pressure SITRANS P measuring instruments for pressure Z series for gage pressure Siemens AG 008 Overview Design The main components of the pressure transmitter are: Brass housing with silicon measuring cell and

More information

Certificate of Compliance

Certificate of Compliance FF0002 Certificate of Compliance Certificate: 70024080 Date Issued: 2017-07-04 Issued to: ecom instruments GmbH Industriestrasse 2 Assamstadt, Baden Wurttemburg 97959 GERMANY Attention: Kevin Spence The

More information

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05

IDL01. Battery Powered Precision Digital Gauge for Leak Testing. Stainless Steel Sensor. class 0.05 IDL0 Battery Powered Precision Digital Gauge for Leak Testing Stainless Steel Sensor class 0.05 Nominal pressure from 0 00 mbar up to 0... 00 bar Special characteristics modular sensor concept data logger

More information

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter.

Instruction Manual HI Portable Waterproof Dissolved Oxygen Meter. Instruction Manual HI 9142 Portable Waterproof Dissolved Oxygen Meter www.hannainst.com 1 Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully

More information

Air Sensor. SAC Ex. Manual ATEX. AQ Elteknik AB

Air Sensor. SAC Ex. Manual ATEX. AQ Elteknik AB Air Sensor SAC Ex Manual AQ Elteknik AB ATEX Air Sensor SAC Ex ATEX Certified model Manual version 2.1 April 2014 AQ Elteknik AB 2 Table of contents 1. Manufacturer information... 4 Manufacture Declaration

More information

Optical Dissolved Oxygen Meter

Optical Dissolved Oxygen Meter Optical Dissolved Oxygen Meter HI764113 Rugged Optical Dissolved Oxygen Probe for Fresh and Saltwater Applications Digital probe Fast and stable readings No membranes Not affected by sunlight No electrolytes

More information

_ pressure transducers. User Manual

_ pressure transducers. User Manual _ pressure transducers User Manual summary introduction DescriPTION preliminary checks Installation taking measurements data management Troubleshooting maintenance Appendix 1 Page 4 Page 5 Page 6 Page

More information

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure

2 Overview. SITRANS P measuring instruments for pressure. Transmitters for gage and absolute pressure. Z series for gage pressure Z series for gage pressure Overview Design The main components of the pressure transmitter are: Brass housing with silicon measuring cell and electronics plate Process connection Electrical connection

More information

RTD temperature probes

RTD temperature probes RTD temperature probes Ex i for use in areas with an explosion hazard (Ex areas) Persons concerned: Experienced professional electricians as per EU Directive 1999/92/EC and trained personnel Operating

More information

HI 2314 HI 2315 HI 23151

HI 2314 HI 2315 HI 23151 Instruction Manual HI 2314 HI 2315 HI 23151 Multi-Range Conductivity Meters for Laboratories www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction

More information

User Manual. GPL 3000 e. gas detector

User Manual. GPL 3000 e. gas detector User Manual GPL 3000 e gas detector Table of contents Application...4 Unit ppm, Vol.%...5 Operation elements GPL 3000 e...6 Gas Sensor...7 Measurement range...8 LED allocation...8 Turning on the GPL 3000

More information

Optical Dissolved Oxygen Meter

Optical Dissolved Oxygen Meter Optical Dissolved Oxygen Meter HI764113 Rugged Optical Dissolved Oxygen Probe for Fresh and Saltwater Applications Digital, weighted probe No membranes No electrolytes No oxygen consumption No flow dependence

More information

S-D2 OPERATOR S MANUAL Lev A

S-D2 OPERATOR S MANUAL Lev A S-D2 OPERATOR S MANUAL 650132 Lev A 316 E. 9th St. Owensboro, KY 42303 866-835-0690 Fax: 270-685-6268 www.alcoholtest.com Operator s Manual S-D2 2002 by CMI, Inc. No part of this work covered by the copyrights

More information

Installation Instructions

Installation Instructions Model: ACT-4500 ACT-4500-3 ACT-4500-6 SUN ELECTRIC CORPORATION Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation

More information

Installation Instructions

Installation Instructions SUN ELECTRIC CORPORATION Model: MRC-450 MRC-450-3 MRC-450-6 MRC-450-GR Page: 1 of 5 Installation Instructions INSTALLATION MUST BE PERFORMED BY QUALIFIED SUN PERSONNEL ONLY INSTALLATION OVERVIEW: The Installation

More information

BGA158 Series CE Approved Class A Shutoff Gas Valve

BGA158 Series CE Approved Class A Shutoff Gas Valve Installation Instructions BGA158 Issue Date August 24, 2011 BGA158 Series CE Approved Class A Shutoff Gas Valve Applications The BGA158 Series shutoff gas valve is an electrically operated gas valve that

More information

200A Series ELECTRONIC LEVEL TRANSMITTER

200A Series ELECTRONIC LEVEL TRANSMITTER MARINE - WATER AND PROCESS INDUSTRIES LIQUID LEVEL AND CONTENTS APPLICATIONS HYDROSTATIC OPERATION 200A Series ELECTRONIC LEVEL TRANSMITTER INTRINSICALLY SUBMERSIBLE - SCREWED MARINE SAFE FLANGED - HYGIENIC

More information

RULES PUBLICATION NO. 119/P

RULES PUBLICATION NO. 119/P RULES PUBLICATION NO. 119/P PERIODIC SURVEYS OF FUEL INSTALLATIONS ON SHIPS OTHER THAN LIQUEFIED GAS CARRIERS UTILIZING GAS OR OTHER LOW FLASH POINT FUELS 2019 January Publications P (Additional Rule Requirements)

More information

Oxidation Stability of Gasoline and Aviation Fuels

Oxidation Stability of Gasoline and Aviation Fuels Oxidation Stability of Gasoline (Induction Period Method) Oxidation Stability of Aviation Fuels (Potential Residue Method) Test Method Provides an indication of the tendency of gasoline and aviation fuels

More information

DFX II Series. Digital Force Gauge

DFX II Series. Digital Force Gauge DFX II Series Digital Force Gauge Functions and Features Simple Operation User friendly four button operation for; Peak, Units, Zero and Info. PEAK- change from Normal to Peak mode UNITS- select desired

More information

A Rationale for Pressure Relief Device(s) Qualification Requirements (LH2)

A Rationale for Pressure Relief Device(s) Qualification Requirements (LH2) UN-GTR PART A INSERTION Stand 09.02.2011, 17:20 Uhr A.3.3 HYDROGEN STORAGE SYSTEM The hydrogen storage system consists of all components that form the primary pressure boundary of the stored hydrogen in

More information

White Paper August , Rev AA

White Paper August , Rev AA Section 1 Automatic Tank Gauging in Custody Transfer Applications Abstract Today Automatic Tank Gauges (ATG s) are commonly used world wide for custody transfer of petroleum products. International standards

More information

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C)

Two-Wire In Situ Oxygen Analyzer (550 to 1400 C) Product Data Sheet PDS 106-581.A01 September, 2007 Model 5081FG Two-Wire In Situ Oxygen Analyzer (550 to 1400 C) Intrinsically safe: ATEX, Zone 1, EEx ia IIC T4 Class I, Div. I, Gr. A, B, C, D T4 Operates

More information

Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors

Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors : Calibration and Bump Test Requirements for Direct Reading Portable Gas Monitors Manufacturers and regulatory agencies agree the safest and most conservative approach is to perform a functional test by

More information

GENERAL APPLICATION NOTES

GENERAL APPLICATION NOTES GENERAL APPLICATION NOTES There are common considerations when using our connectors. Additional considerations may apply based on the particular connector being used, the application, and conditions in

More information

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter

INSTRUCTION MANUAL. 840 Transmitter. Loop Powered Dissolved Oxygen Transmitter INSTRUCTION MANUAL Loop Powered Dissolved Oxygen Transmitter Contents Contents...2 Introduction...3 Range and Display Settings...4 Installation...5 Electrical Installation...6 Calibration...7 EasyCal...8

More information

ellison sensors international

ellison sensors international PR3900 HAZARDOUS AREA PRESSURE TRANSMITTER ATEX CERTIFIED AND ALL EQUIVALENTS FOR HAZARDOUS AREAS: ZONE 0 GAS GROUP IIC, TEMPERATURE CLASS T4 AND ZONE 20 DUST PROTECTION BY INTRINSIC SAFETY TO EEXIA IIC

More information

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use

Series 6517 Katharometers. The ideal equipment for Process Monitoring and Control. Robust no moving parts. Designed for continuous industrial use Data sheet Katharometers The ideal equipment for Process Monitoring and Control Designed for continuous industrial use Long working life Suitable for flammable gases CENELEC certified (ATEX) to EExia IIC

More information

BGA244 Binary Gas Analyzer

BGA244 Binary Gas Analyzer Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials

More information

Type Testing Procedure for Crankcase Explosion Relief Valves

Type Testing Procedure for Crankcase Explosion Relief Valves (Jan 2005) (Corr.1 Nov 2005) (Rev.1 Oct 2006) (Corr.1 Mar 2007) (Rev.2 Sept 2007) (Corr.1 Oct 2007) (Rev.3 Jan 2008) Type Testing Procedure for Crankcase Explosion Relief Valves 1. Scope 1.1 To specify

More information

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions

The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions The M-Series Eletta Flow Meter High accuracy DP Flow Meter with multiple functions Flow Meter with multiple functions for gases and liquids M3 The M-series Flow Meter, with its versatile and user-friendly

More information

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE

ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE ATEX VACUUM PUMPS AND GAUGES CHEMICALLY RESISTANT, OIL-FREE AND SAFE ATEX CHEMISTRY DIAPHRAGM PUMPS. ATEX VARIO CHEMISTRY DIAPHRAGM PUMPS. Chemistry diaphragm pumps, vacuum systems and gauges for use in

More information

APG. Intrinsically Safe, Heavy Duty Pressure Transducer Series: PT-400

APG. Intrinsically Safe, Heavy Duty Pressure Transducer Series: PT-400 Intrinsically Safe, Heavy Duty Pressure Transducer Series: PT-400 C US The PT-400 offers high accuracy and reliability over a wide range of pressures. The small size, integrated electronics, wide operating

More information

Compact Manometer. Pressure measurements can easily be taken any time, anywhere. POWER LIGHT

Compact Manometer. Pressure measurements can easily be taken any time, anywhere. POWER LIGHT Compact Manometer Series PPA Pressure measurements can easily be taken any time, anywhere. G GS PPA Compact and lightweight Portable type with a lightweight of only about 100 g (unit 50 g, battery 50 g)

More information

THERMAL MASS FLOW MEASUREMENT FOR GASES

THERMAL MASS FLOW MEASUREMENT FOR GASES THERMAL MASS MEASUREMENT FOR GASES Make the Wise Choice. Choose Sage Flow Meters. A Commitment to Higher Performance SAGE METERING is a manufacturer of high performance Thermal Mass Flow Meters which measure

More information