Supplemental Body Plethysmograph Service Guide Version/Revision 1.A ComPAS ver. 1.9

Size: px
Start display at page:

Download "Supplemental Body Plethysmograph Service Guide Version/Revision 1.A ComPAS ver. 1.9"

Transcription

1 Supplemental Body Plethysmograph Service Guide Version/Revision 1.A ComPAS ver. 1.9 Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 1

2 Information in this manual is subject to change without notice and does not represent a commitment on the part of Morgan Scientific, Inc. The software described in this document is furnished under a license agreement. The software may be used or copied only in accordance with the terms of the agreement. It is against the law to copy the software on any medium except as specifically allowed in the license or nondisclosure agreement. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying and recording, for any purpose without express written permission of Morgan Scientific, Inc. Copyright 2012 Morgan Scientific, Inc. Precautions Caution: Federal law restricts this device to sale by, or on the order of a physician Caution: Not suitable for use in the presence of flammable anesthetics Service of this instrumentation is restricted to factory trained personnel only Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 2

3 Table of Contents Chapter Section Details Page 1.0 Problem Guide Common problems with symptoms and solutions Diagnostics Body Box Diagnostic routines Changing Gas Cells 3-1 Replacing the O2 Cell Replacing the CO Cell Cleaning and Disassembling the Mouthpiece Shutter Assembly 5-0 Cleaning the Plethysmograph Cabin 4-1 General Recommendations Routine Practice Hygiene Cleaning Instructions Cleaning/Disinfecting- Recommendations Chart Disassembling the Pneumotachograph 15 Assembly 4-7 Disassembling the Valve and Shutter Assembly Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 3

4 1:0 Body Box Problem Guide Fault Symptom Solution The large inspirate bag will not empty for gas calibration The pump continues to run but the inspiratory bag will not empty. 1) Check the side gas analyzer absorber is seated properly. 2) Check that the gas cylinder output pressure is set to 50 psi 3) Replace the large inspirate bag. If the problem persists it is useful to try one of the small expirate bags to confirm that your spare large inspirate bag is leak free or not. 4) If you have completely ruled out any leaks in the bag, the problem will be resolved by adjusting the vacuum switch. This can be accessed on the rear of the box electronics module as shown: If the bag will not trigger the vacuum to indicate "empty", then turn the screw adjustment counterclockwise. Complete one full turn and try emptying the bag again. Continue until the vacuum trigger operates. No cylinder gas reaches the analyzers during calibration. The gas analyzers appear to be sluggish and the DLCO results are in error. The flow volume calibration will not be accepted after several attempts. Body Box appears to operate (i.e. valve functions and bag filling) with the exception of being able to calibrate the analyzers. System appears to be working, but the analyzer responses are slow. You may also notice that the inspirate and expirate bags are emptying simultaneously when only the expirate bag is supposed to be emptying. During the calibration sequence, the flow volume calibration is constantly rejected. The cylinder pressure is too high. Reduce the cylinder pressure to between 50 & 60 psi. Note: Sometimes old gas regulators do not display the correct pressure! Open the electronics module. It is very likely that a sample tube has been kinked or trapped. Check that the pneumotach is properly seated in the valve body. Check that the red and blue hose connections to the top of the pneumotach are connected properly. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 4

5 Fault Symptom Solution Communication is lost with the Body Box The analog flow signal appears to be working as do the digital controls, but all other analog channels are reading ) Administrative rights have been changed or taken away from ComPAS. In Regedit, set Full Administrative rights for the registry branch: HKEY_Local_Machine Software Morgan Scientific, Inc Navigate to C:\Program Files\Morgan Scientific\ and also the main program Full Administrative rights: C:\Program Files\Morgan Scientific\ 2) The Compatibility Mode properties for the ComPAS icon must be set to Windows XP Service Pack 3 for all users! Gas appears to be leaking from inside the electronics assembly. The cylinder of diffusion gas (model 5510) or compressed air (model 5500) is going down. When the main pressure is turned-off on the regulator, the gauge rapidly falls to zero. The AirTrol miniature precision regulator may have failed. Earlier models of Body Box 5500 and 5510 were fitted with a model 820 regulator. This should be replaced with the improved model 810. When received from Morgan Scientific, the output pressure will have been set to 50psi. Manometer fluid appears to climb when trying to calibrate Mouth Pressure. You cannot set zero on the manometer when trying to calibrate because the fluid keeps climbing. It is not possible therefore to calibrate Mouth Pressure. If you have not calibrated Mouth Pressure for some time, The problem would likely be seen as erroneous lung volume data. Typically a leak into the valve during a VTG maneuver will result in a falsely lower value for VTG, FRC and TLC. A low TLC value would most likely be recognized by an abnormally low RV (TLC -SVC= RV). One of the piston valves on the patient manifold is leaking and needs to be replaced. Please call Morgan Scientific for immediate help. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 5

6 Fault Symptom Solution No gas is collected in the expiratory bag at the end of a DLCO test. During either gas calibration or DLCO testing, the inspiratory bag will not inflate. All other test functions and valve operation seems to be normal. Items to check are as follows: 1) That the patient maintained a tight seal on the mouthpiece during expiration. 2) That the settings for Washout and Sample were appropriate. Note: Minimum setting for Sample is 400ml. Minimum setting for Washout is 200ml. 3) The cylinder pressure (used to operate the valves) is set at 50 psi. No gas is filling into the inspiratory bag. During either gas calibration or DLCO testing, the inspiratory bag will not inflate. All other test functions and valve operation seems to be normal. There is an over-pressure sensor that is used to prevent the inspiratory bag from being over-filled. The sensor shown in the illustration below has a screw adjustment on the side. A clockwise turn of the screw will decrease the sensitivity for detecting an over-pressure. The inspiratory bag is not filling with enough volume During DLCO tests, either too much or too little gas is entering the inspiratory bag. The control of the bag filling can be adjusted as follows: 1) Go to "Calibration" and then click on the Diagnostics button 2) From the side [Tabs] select "Configuration". The default setting for the "Inspirate Bag Filling Speed" is 18. Decreasing the value will increase the filling volume of the bag. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 6

7 Fault Symptom Solution The VA at BTPS values appear to be elevated. The VA (Alveolar Volume) at BTPS from the DLCO test are higher than TLC from either plethysmographic or helium dilution tests. 1) Assuming the TLC values are reliable and reproducible, the most likely cause is a minor leak in the expiratory bag. Sometimes the leak is so small that it cannot be detected by the vacuum switch test prior to completing any DLCO maneuver. 2) Check that the chemicals in the absorber are fresh. In particular, exhausted soda lime will cause significant changes in helium analysis. 3) Check that the absorber is not over-filled causing a leak. The gas analyzers are not calibrating properly from the inspiratory bag. If you did not recognize a problem in gas calibration, you may first see the symptom with erroneous Alveolar Volume values in DLCO. The helium analyzer wave form seen during calibration is clearly unusual (see diagram) The top coil of solenoid S-20 has failed and needs replacing. The flow/volume signal appears to have lost all gain. The flow signal can be zeroed but there was little or no gain when performing a test or calibration. The analog signals for the gas analyzers are normal. The internal differential pressure transducer has most likely failed. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 7

8 2.0 Body Box Diagnostics The diagnostics screen will show all the voltages coming from the Body Plethysmograph. At the same time, it will allow the service of biomedical engineer to test the digital control functions of the instrument. To access diagnostics, go to "Calibration" and then click on the Diagnostics button From here you will be able to view and check both the analog voltages and digital controls. The right-hand tabs provide an array of 'tools' and 'test functions' to help engineers with any service issues. Setup - is used for the initial installation of the Body Box and also for creating linearity tables and changing span factors in calibration. Gas - gives access to testing all functions where gas analysis is used. Digital - allows you to test all the digital commands and actions Modules - allows you to run any of the automated sequences used in testing Scope - provides a useful oscilloscope to view voltage changes Configuration - hardware control settings On all screens, the analog voltages can be viewed directly in the appropriate meter displays. To diagnose digital functions and Body Box solenoid operation, clicking on any of the Digital Controls will activate the command. The valve diagram will indicate which valve function is being activated by the "Change Valve" options. On the 5510, the valve has the following positions: Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 8

9 Patient to Atmosphere (the position for spirometry, flow volume and MVV studies) Patient to the Expirate Bag (activated in the DLCO test) Patient to the Inspirate Bag (activated in the DLCO test) Patient to Closed Shutter (used in VTG and Raw testing. Note: the "Shutter" button operates this position) Empty the Expirate Bag (also checks for any leaks) Empty the Inspirate Bag (also checks for any leaks) Fill the Inspirate Bag The [Configuration] tab allows you to change settings for both hardware function and test settings: Number of points to show on history reports DLCO % of VC to use for bag filling Number of points to show on history reports Gas sampling speed The default setting is 10 The number of past calibration dates to view on the calibration history displays The default setting is 100 The percent of measured VC to use when filling the bag for DLCO The default setting for the "Inspirate Bag Filling Speed" is 18. Increasing the value will reduce the volume of gas entering the bag and conversely, decreasing the value will increase the filling volume of the bag. The default setting is 0.6 L/min This is the pump speed for gas sampling and rarely if ever needs changing. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 9

10 Setting for oxygen zero Typically this value is very close to 4000 Flow noise filter The default setting is 3 This is a value that must be recorded (Set to Current) whenever the O2 cell is changed The filter is used in post processing of all flow data. Increasing the filter value will smooth data but care has to be taken not to affect peak flow readings. Mouth pressure calibration value The default setting is 20 cmh20 The value used when calibrating the mouth pressure transducer with a manometer. MIP MEP computation method The default setting is 0 Field not used at this time. AtoD Rev Body Box serial numbers after should all be set to 1 Swap A/B pistons on non- DLCO models Only used on a special research unit 3.0 Changing the O2 and CO Gas Cells 3-1 Changing the O2 Cell 1) Turn off the electronics and remove the power cord. 2) Remove the white O2 cell cap by simply pulling down. 3) Unplug the signal cable from the center of the sensor. 4) Unscrew the old O2 cell and replace with the new one; for now do not connect the signal cable to the O2 cell 5) Go to "Calibration" and then click on the Diagnostics button 6) From the side [Tabs] select "Configuration". Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 10

11 7) The current zero or open-circuit value will be shown. Click on [Set to Current] and be absolutely certain that you click the button! 8) Reconnect the O2 signal cable and replace the white cap over the cell. 9) Exit the diagnostics routines. 10) Once everything is replaced and the system has been powered-on, it is necessary to go to "Tools" and "Configuration" and "Devices" to tell the software that the cell(s) have been changed. Click the [Change] button to confirm the cell change. 11) Run a gas calibration to complete the cell change 3-2 Changing the CO Cell Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 11

12 1) Turn off the electronics and remove the power cord. 2) The CO sensor is mounted under the Body Box electronics module. 3) Pull off the large white cover. 4) Remove the plug and then unscrew the three retaining screws. 5) Once the new cell has been replaced, you will need to go to "Tools" and "Configuration" and "Devices" to inform ComPAS that new cells are being used. Click the [Change] button to confirm the cell change. 6) Run a gas calibration to complete the cell change 4.0 Cleaning and Disassembling the Mouthpiece Shutter Assembly Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 12

13 4-1 General Recommendation To reduce the risk of cross infection a new Bacterial Viral Filter should be used for each subject. A Risk Assessment should be carried out to assess the risks presented to both operator and subject, and an action plan devised by the facility to minimize the chance of cross infection occurring, particularly where known-infectious or immuno-deficient subjects are being tested. An assessment should be made of methods of decontamination available to the operator and their effectiveness against potential risks - a table of materials used in the Plethysmograph is provided below to assist in this. It is recommended in cases of high risk with no effective disinfection methods available that the contaminated parts are disposed of; for this device, these are the Pneumotach Assembly (pneumotach complete), Filter Adaptor and Interconnection Assembly. 4-2 Routine Practice A new Bacterial Viral Filter should be used for each subject. Between patients it is recommended that all exterior surfaces of the Plethysmograph (pneumotach assembly and valve) by wiped down using Sani Surface Wipes. A delay of at least 5 minutes should be allowed between subjects to allow settling of previously aerosolized particles in the measuring device. It is recommended that the Pneumotach Assembly be regularly cleaned according to the guidelines of the user s facility. In the event of visible contamination to the Pneumotach Assembly and Screen - these should be cleaned and disinfected. The frequency of cleaning is dependent on the Facilities Risk Assessment, usage, and test environment, but should be at least monthly or every 100 subjects (500 blows). 4-3 Hygiene It is vital for the user to set guidelines for protective hygiene measures whilst performing spirometry testing. There are three main potential sources of cross contamination, skin contact; aerosolized particles and saliva/body fluids. By far the most important is the last item a minimum requirement is to use a new disposable mouthpiece for each subject tested. We recommend using the Bacterial Viral Filter for all testing maneuvers. We also recommend that a delay of at least 5 minutes be allowed between subjects. This allows aerosolized organisms to be removed by gravitational sedimentation between tests. (Am J Respir Crit Care Med Vol 159. pp , 1999). 4-4 Cleaning Instructions All parts of the Plethysmograph require regular cleaning, i.e. the removal of visible particulate contamination. The parts of the Plethysmograph that make up the Pneumotach Assembly also require disinfecting. Morgan Scientific recommends that cleaning and disinfecting should normally be carried out on a bi-annual basis. In addition to a weekly cleaning and disinfection cycle, Morgan Scientific recommends cleaning and disinfection of equipment to be carried out after use on infected subjects or prior to use on immunocompromised subjects. The user must determine what level of disinfection is appropriate and acceptable in any particular circumstance. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 13

14 4-5 Cleaning/Disinfecting- Recommendations Chart Part Material Clean/Disinfect Autoclave Recommended Disinfectants Main Cabin Exterior Cabin Frame: Aluminum AlMg3, Polyurethane & aliphatic setting paint Clean No Wiping w ith a 70% Isopropyl Alcohol impregnated cloth provides a suitable form of cleaning and low -level disinfection. Cabin Door: Toughened security glass Sani Surface Wipes are recommended and w idely used in hospital settings. Chair Patient Valve Cabin Panels: Acrylic Wood and polyurethane coating OR acrylic (two types of chair are available) Polyoxymethylene (POM), Aluminum Clean No As Above Clean No As Above Interconnection assembly Nylon Clean No As Above Filter Adaptor Polyoxymethylene (POM) Clean Viable As Above Pneumotach Body Polyoxymethylene (POM) Clean Viable As Above Pneumotach Screen Aluminum, Stainless Steel Clean & Disinfect O Rings Viton fluoroelastomer Clean & Disinfect Viable Viable Disinfect by immersion in sodium dichloroisocyanurate solution at 1000ppm concentration of free chlorine for 15 minutes. The pneumotach screen w ill also w ithstand autoclaving at 134 C for 3 minutes. As Above Inspirate & Expirate Bags Polyoxymethylene (POM), Polyethylene Dispose No Dispose Definitions of cleaning and disinfection are as defined in Sterilization, Disinfection and Cleaning of Medical Equipment: Guidance on Decontamination from the Microbiology Committee to Department of Health Medical Devices Directorate, 1996 Recommendations for chemical disinfectants are derived from the PHLS publication Chemical Disinfection In Hospitals On the Body Box, the complete breathing pathways can be cleaned and sterilized. The full circuit includes the following components: The pneumotachograph assembly The patient valve Cleaning guidelines for each module will now be detailed: Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 14

15 4-6 The Pneumotachograph Assembly Step 1 Disconnect the heater cable and the red and blue hoses from the pneumotach. Carefully remove the pneumotach from the patient valve. Step 2 Place your thumbs on the two pressure fittings and gently turn against one another in an anti-clockwise direction. If the pneumotach is a later design, a special tool is available to help open the two halves. This will open the pneumotach and gain access to the Lilly-style screen. Up against a backlight, carefully examine the screen. Some bacterial/viral filters deposit a fine lint that can gradually clog the screen. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 15

16 Step 3 If the screen needs simple cleaning, preferably blow compressed air or O2 through the screen. If compressed air is not available, hold the screen beneath warm running water. Shake the screen vigorously to remove water droplets and then let it air dry before re-assembling the pneumotach. With each Plethysmograph there are spare screens to use while others are drying. If the screen needs to be sterilized, disinfect by immersion in sodium dichloroisocyanurate solution at 1000ppm concentration of free chlorine for 15 minutes. The pneumotach screen will also withstand autoclaving at 134 C for 3 minutes. Step 4 Once all the parts are cleaned and sterilized, apply a small amount of DuPont Krytox grease on the two O'ring seals that secure the pneumotach back into the patient valve. Krytox is available from Morgan Scientific: Cat No. CH71686 Krytox grease (1 oz) 4-7 The Valve and Shutter Assembly Step 1 Turn-off the cylinder pressure and use the diagnostics screen in ComPAS to bleed-off any pressure in the circuit. Remove the inspirate and expirate bags. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 16

17 Step 2 Disconnect the all the tubing connections. Each of the pneumatic tubing connections are labelled for correct re-assembly. To release the high-pressure connections, hold the collar against the manifold and pull on the hose. Regular soft hose connections simply pull away from the retaining nipple. Step 3 Remove the valve assembly from the patient arm. The valve simply slides out of the retaining slot Step 4 The valve can be further disassembled by removing the inspirate bag valve as shown. The valve body should either be cleaned with 70% Isopropyl Alcohol or immersed in Cidex as recommended by the manufacturer. Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 17

18 Step 5 Once the valve has been cleaned, it should be lubricated using the Krytox grease before use. The parts to lubricate are all the visible the O-rings. Step 6 Each of the pressure and tubing connections are labelled on the valve body with corresponding labels on the tubes. The pneumotach connections are labelled with red and blue identifiers. Having put everything back together, use the ComPAS Diagnostic screen to test that all the valve functions are performing properly. 5-0 Cleaning the Plethysmograph Cabin Since the Plethysmograph is designed to be an 'open' and friendly environment, the majority of the construction is made of glass and acrylic. These materials tend to show dirt and dust very easily, so we recommend that the instrument be cleaned on a regular basis. Cabin Exterior All the glass or acrylic surfaces should be cleaned with a commercial glass cleaner using a soft cloth or paper towel. For the interior panels, we recommend a glass cleaner that also contains a disinfectant or bacterial fighting agent. Having cleaned the interior panels, the door should be left open and the cabin ventilated for 20 minutes. Materials Data: Exterior Frame: Alluminium AlMg3 Frame Covering: Polyurethane & aliphatic setting paint Door: Toughened security glass Panels: Acrylic Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 18

19 Cabin Interior The cabin chair and floor surfaces should be cleaned with a commercial cleaning solution that contains a mild disinfectant or bacterial fighting agent. Having cleaned the interior, the door should be left open and the cabin ventilated for 20 minutes. Materials Data: Chair: Wood and polyurethane coating OR acrylic (two types of chair are available) Chair Post: Stainless Steel 3041 and 3161 Floor: Synthetic rubber (non latex) Morgan Scientific, Inc. Supplemental Body Plethysmograph Service Guide Page 19

Rejuvenation Instructions

Rejuvenation Instructions Rejuvenation Instructions #401 Air Systems UPR This NRI covers the following: Understanding the applications and operation of flow meters. Understand the application and operation of test pressure gauges.

More information

Breeze Suite Quick Start Operator Manual. An Introduction to Pulmonary Function Testing

Breeze Suite Quick Start Operator Manual. An Introduction to Pulmonary Function Testing Breeze Suite Quick Start Operator Manual An Introduction to Pulmonary Function Testing The MedGraphics Breeze Suite Quick Start Operator Manual is Part No. 142143-001 Rev A. Caution: Federal (USA) law

More information

Body Box 5500 Series Pulmonary Function Testing Systems

Body Box 5500 Series Pulmonary Function Testing Systems Body Box 5500 Series Pulmonary Function Testing Systems The highest quality of respiratory measurement Plethysmography is used to measure all the thoracic gas volume (VTG) whether in communication with

More information

256 Pneumatic Pressure Indicator

256 Pneumatic Pressure Indicator 256 Pneumatic Pressure Indicator 51425699 Copyright 2002 Slope Indicator Company. All Rights Reserved. This equipment should be installed, maintained, and operated by technically qualified personnel. Any

More information

OPERATION MANUAL BFX-2

OPERATION MANUAL BFX-2 OPERATION MANUAL BFX-2 Brake Fluid Exchanger IMPORTANT Test drive vehicle after service to verify proper brake system performance. MAHLE Aftermarket Inc., Service Solutions 10 Innovation Drive York, PA

More information

A Pulmonary Function Product of Rare Distinction

A Pulmonary Function Product of Rare Distinction A Pulmonary Function Product of Rare Distinction SpiroAir Fully Digital Positive Displacement PFT System How do you recognize PFT Excellence? The ultimate in volume and flow accuracy Reliable and reproducible

More information

Instruction Manual Updated 7/26/2011 Ver. 2.2

Instruction Manual Updated 7/26/2011 Ver. 2.2 4-Unit Model MB HTHP Filter Press #171-50-4: 115-Volt #171-51-4: 230-Volt Instruction Manual Updated 7/26/2011 Ver. 2.2 OFI Testing Equipment, Inc. 11302 Steeplecrest Dr. Houston, Texas 77065 U.S.A. Tele:

More information

Independent Health Facilities

Independent Health Facilities Independent Health Facilities Assessment Protocol for Pulmonary Function Studies - Technologist Assessor INSTRUCTIONS: Please complete ( ) the attached protocol during the assessment. Ensure that all the

More information

Air-Mate High Efficiency Powered Air Purifying Respirator (PAPR) Staff Training Module

Air-Mate High Efficiency Powered Air Purifying Respirator (PAPR) Staff Training Module Air-Mate High Efficiency Powered Air Purifying Respirator (PAPR) Staff Training Module PRES0389 7/05-1 Powered Air Purifying Respirators (PAPR) Reduce the risk of transmission of infectious agents spread

More information

ACV-10 Automatic Control Valve

ACV-10 Automatic Control Valve ACV-10 Automatic Control Valve Installation, Operation & Maintenance General: The Archer Instruments ACV-10 is a precision automatic feed rate control valve for use in vacuum systems feeding Chlorine,

More information

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting

Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator.

More information

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual

Welker Sampler. Model GSS-1. Installation, Operation, and Maintenance Manual Installation, Operation, and Maintenance Manual Welker Sampler Model GSS-1 The information in this manual has been carefully checked for accuracy and is intended to be used as a guide to operations. Correct

More information

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120

MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 MASK INTEGRITY TEST ACCESSORY (MITA) MODEL 8120 QUICK START GUIDE P/N 6006154, REVISION C MAY 2013 Model 8120 Mask Integrity Tester is patented under U.S. Patent No. 8,312,761. Additional patents are pending.

More information

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC. MANUFACTURERS AND DISTRIBUTORS OF: O-Two eseries Electronic Transport Ventilators O-Two Single Use Open CPAP Systems O-Two CAREvent Range of Automatic Transport Ventilators O-Two CAREvent Range of Automatic

More information

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER

Operation Manual - PN A MENSOR MODEL 73 SHOP AIR BOOSTER Operation Manual - PN 0017946001 A MENSOR MODEL 73 SHOP AIR BOOSTER Mensor Model 73 Shop Air Booster System (750 psi Version) April 23, 2012 Trademarks / Copyright Mensor is a registered trademark of Mensor

More information

TR Electronic Pressure Regulator. User s Manual

TR Electronic Pressure Regulator. User s Manual TR Electronic Pressure Regulator Page 2 of 13 Table of Contents Warnings, Cautions & Notices... 3 Factory Default Setting... 4 Quick Start Procedure... 5 Configuration Tab... 8 Setup Tab... 9 Internal

More information

5 Gallon Pressure Pot with HVLP Spray Gun and Hose

5 Gallon Pressure Pot with HVLP Spray Gun and Hose California Air Tools 5 Gallon Pressure Pot with HVLP Spray Gun and Hose Model No. 365 Technical Data Type of feed.pressure Maximum pressure in the tank... 0,413Mpa (60PSI) Working pressure in the tank.0,

More information

Mass Spec will not Autotune

Mass Spec will not Autotune Mass Spec will not Autotune Applies to 5973A/N MSD What could be the problem? There could be several things that would cause your Mass Spec not to Autotune. The most common, easily corrected Autotune problems

More information

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS

6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS 6000 SERIES OF PNEUMATIC DEADWEIGHT TESTERS USER MANUAL -1- SI-BARNET DEADWEIGHT TESTER 6000 SERIES Release: S31-0250 Revision: A Date: 05/08/03 MANUFACTURED BY: GE RUSKA RUSKA INSTRUMENT CORPORATION 10311

More information

Operation and Maintenance of the EPV200 Portable Ventilator

Operation and Maintenance of the EPV200 Portable Ventilator Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide

More information

Pressure Automated Calibration Equipment

Pressure Automated Calibration Equipment GE Measurement & control Pressure Automated Calibration Equipment Safety Instructions and User Guide - K0447 PACE5000 PACE6000 K0447 Issue No. 9 1 10 1 PACE5000 1 2 3 4 5 PACE6000 2 6 7 8 3 4 5 6 7 8 9

More information

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities

Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities I. Introduction Pulmonary Function I (modified by C. S. Tritt, April 10, 2006) Volumes and Capacities The volume of air a person inhales (inspires) and exhales (expires) can be measured with a spirometer

More information

OPERATION MANUAL NTF-60 Plus

OPERATION MANUAL NTF-60 Plus OPERATION MANUAL NTF-60 Plus Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81264-00 (Rev A)

More information

PAPR Flow Tester. instructions. part no WARNING

PAPR Flow Tester. instructions. part no WARNING PAPR Flow Tester part no. 488903 instructions WARNING This manual, including warnings and cautions inside, must be carefully read and followed by all persons who use or maintain the product, including

More information

EasyWarePro Message Numbers

EasyWarePro Message Numbers Application Note CB 06/2016 EasyWarePro Message Numbers Important note: this application note is only valid beginning from EOPC V1.9.0.18 and EOP / Lab V1.9.0.19. Compared to the previous released version

More information

Met One E-BAM Particulate Monitor

Met One E-BAM Particulate Monitor STANDARD OPERATING PROCEDURES Met One E-BAM Particulate Monitor AMBIENT AIR MONITORING PROGRAM for the 130 LIBERTY STREET DECONSTRUCTION PROJECT LOWER MANHATTAN DEVELOPMENT CORPORATION 1 Liberty Plaza

More information

EasyWarePro Message Numbers

EasyWarePro Message Numbers Application Note CB 10/2018 EasyWarePro Message Numbers Important note: this application note is only valid beginning from EOPC V1.9.0.18 and EOP / Lab V1.9.0.19. Compared to the previous released version

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C

Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung. Susan Blonshine RRT, RPFT, FAARC, AE-C Diffusing Capacity: 2017 ATS/ERS Standards for single-breath carbon uptake in the lung Susan Blonshine RRT, RPFT, FAARC, AE-C Joint ATS/ERS Taskforce Recent literature reviewed Surveyed current technical

More information

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.

P5513. Users Manual. Pneumatic Comparison Test Pump. Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot. Test Equipment Depot - 800.517.8431-99 Washington Street Melrose, MA 02176 TestEquipmentDepot.com P5513 Pneumatic Comparison Test Pump Users Manual PN 3963372 November 2010 2010 Fluke Corporation. All

More information

OPERATION MANUAL NTF-15

OPERATION MANUAL NTF-15 OPERATION MANUAL NTF-15 Nitrogen Tire Filling Valve Stem Caps (Qty=200) Order P/N 436075 RTI Technologies, Inc 10 Innovation Drive York, PA 17402 800-468-2321 www.rtitech.com 035-81235-00 (Rev B) TABLE

More information

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators

Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators Genstar Technologies Co., Inc. 4525 Edison Avenue Chino, CA 91710 USA TEL 909-0-272 FAX 909-0-485

More information

OPERATION MANUAL ID900

OPERATION MANUAL ID900 OPERATION MANUAL ID900 Portable Refrigerant Identifier WARNING Liquid Samples will Damage Unit Do not operate if filter has red flecks. Replace filter every 3 months or after every 25 uses. Always have

More information

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator Operating Instructions Table of Contents Resource v.2 Conserving Regulator Safety Information Device Precautions Introduction Product Features Product Specifications Feature Illustrations Set Up Usage

More information

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR

CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR FR-300 RESUSCITATOR Symbols used in the Operating Manual for the Oxylator FR-300 and/or on the Oxylator FR-300 unit: WARNING! Do not use grease or

More information

Dialflow Regulator. Instructions for Use

Dialflow Regulator. Instructions for Use Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user

More information

Oxygen Dialflow Meter. Instructions for Use

Oxygen Dialflow Meter. Instructions for Use Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,

More information

ETT Olfactometer: Getting started guide 2013a. ETT Olfactometer quick start

ETT Olfactometer: Getting started guide 2013a. ETT Olfactometer quick start ETT Olfactometer quick start Table of Contents Safety Information... 3 Contact to ETT... 3 Un-boxing / Installation... 3 Resources overview... 7 Working with the ETT Olfactometer... 8 Creating the first

More information

Installing ASI Umbilical Delivery Systems

Installing ASI Umbilical Delivery Systems Rev A 05 Nov 2015 Page 1 of 10 PURPOSE This guide provides an overview of tubing connections into an ASI wall-mounted Junction Box that has already been plumbed with a compressed air connection including

More information

Cover Page for Lab Report Group Portion. Head Losses in Pipes

Cover Page for Lab Report Group Portion. Head Losses in Pipes Cover Page for Lab Report Group Portion Head Losses in Pipes Prepared by Professor J. M. Cimbala, Penn State University Latest revision: 02 February 2012 Name 1: Name 2: Name 3: [Name 4: ] Date: Section

More information

DPC-30 DPC-100. Reference Manual

DPC-30 DPC-100. Reference Manual DPC-30 DPC-100 Reference Manual 1. Introduction 1.1 Description The Martel DPC Digital Pneumatic Calibrator improves upon traditional dial gauge pneumatic calibrators. The Martel DPC improves accuracy,

More information

AUTOVENT 4000 VENTILATOR

AUTOVENT 4000 VENTILATOR OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on

More information

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings

SomnoSuite FAQ. Setup. Calibration 4. What are the calibration requirements for the SomnoSuite? Settings SomnoSuite FAQ V1.3 January 2015 Setup 1. How do I connect the SomnoSuite to my oxygen source? 2. Is there a way to speed up the downward movement of the pusher block when setting the empty position? 3.

More information

Chapter 53: Clinical Center Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) MOP 53.1 Introduction

Chapter 53: Clinical Center Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) MOP 53.1 Introduction Chapter 53: Clinical Center Single-Breath Carbon Monoxide Diffusing Capacity (DLCO) MOP 53.1 Introduction The DLCO will determine the degree of pulmonary impairment in participants enrolled in the PVDOMICS

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

Pressure Regulator. Instructions for Use

Pressure Regulator. Instructions for Use Pressure Regulator Instructions for Use 702-0083.9 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient, the

More information

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance

bespoke In general health and rehabilitation Breath-by-breath multi-functional respiratory gas analyser In human performance Introduction Expired Gas Analysis or indirect calorimetry, can be used to measure ventilation and the fractions of oxygen and carbon dioxide in expired air. From these measurements, the body's oxygen consumption

More information

1020 Industrial Drive, Orlinda, TN fax

1020 Industrial Drive, Orlinda, TN fax Operation Manual Ultrafiltration for High Purity Distribution K-A-HPTUF Series 615-654-4441 sales@specialtyh2o.com 615-654-4449 fax TABLE OF CONTENTS Section 1 GENERAL 1.1 Warnings and Cautions... 1 1.2

More information

IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES. Tank valves must be installed or removed by qualified personnel.

IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES. Tank valves must be installed or removed by qualified personnel. !WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINS!WARNING! IMPORTANT CO2/ HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES GETTING STARTED This Paintball Marker is NOT A TOY. Misuse can cause serious

More information

My CoughAssist. A patient guide to CoughAssist T70. Please visit

My CoughAssist. A patient guide to CoughAssist T70. Please visit Philips Healthcare is part of Royal Philips Electronics Europe, Middle East, Africa +49 7031 463 2254 How to reach us www.philips.com/healthcare healthcare@philips.com Latin America +55 11 2125 0744 Asia

More information

CPR MEDICAL DEVICES INC. OPERATING MANUAL

CPR MEDICAL DEVICES INC. OPERATING MANUAL CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR EMX RESUSCITATION AND INHALATION SYSTEM Symbols used in the Operating Manual for the OXYLATOR EMX and/or on the OXYLATOR EMX unit: WARNING! Do not

More information

Helium-Oxygen Blender

Helium-Oxygen Blender Helium-Oxygen Blender Service Manual Model No. PM5400 Series PM5500 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com

More information

Gas Pressure Probe. Information Sheet. Total Dissolved Gas

Gas Pressure Probe. Information Sheet. Total Dissolved Gas Total Dissolved Gas Information Sheet T900 Total Dissolved Gas Pressure Probe Description The In-Situ Inc. T900 probe is a specially adapted submersible pressure transducer designed to monitor Total Dissolved

More information

INSTALLATION AND OPERATING INSTRUCTIONS

INSTALLATION AND OPERATING INSTRUCTIONS INSTALLATION AND OPERATING INSTRUCTIONS For 3D Instruments Direct Drive Pressure Gauges 3D Instruments Inc., 15542 Chemical Lane, Huntington Beach, CA, 92649, U.S.A. Tel. (714) 894-5351, Fax. (714) 895-4309

More information

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA

STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA STANDARD OPERATING PROCEDURES DIVISION OF COMPARATIVE MEDICINE UNIVERSITY OF SOUTH FLORIDA SOP#: 1157.1 Date Issued: 05/14 Date Revised: 5/15 Page 1 of 6 TITLE: SCOPE: RESPONSIBILITY: PURPOSE: SurgiVet

More information

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL

SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL Product Manual MST Portable Air Filtration System SERVICE MANUAL MODEL BA100BMST 100 SCFM PORTABLE AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the

More information

RhinoPro Cartridge Gun Manual

RhinoPro Cartridge Gun Manual RhinoPro Cartridge Gun Manual RhinoPro Cartridge Gun 1500 with 750ml cartridges Table of Contents Application Instructions......................................... 1 Operating Instructions..........................................

More information

Using the UltraRAE. Firmware 2.35

Using the UltraRAE. Firmware 2.35 Using the UltraRAE Firmware 2.35 Training Agenda UltraRAE features Setting up the UltraRAE Turning on the UltraRAE Idle Operation RAE-Sep Tubes Prepping for a measurement Taking a measurement Alarm modes

More information

User Manual for the Mars Calibration Bench

User Manual for the Mars Calibration Bench User Manual for the Mars Calibration Bench Fall 2013 Table of Contents Table of Contents Table of Contents... iii Introduction... v Chapter 1: The Mars Calibration Bench... 1 What Is the Mars Calibration

More information

INSTALLATION & MAINTENANCE INSTRUCTIONS

INSTALLATION & MAINTENANCE INSTRUCTIONS DESCRIPTION / IDENTIFICATION The QBX series valve uses Proportion-Air closed loop technology for Pressure control. It gives an output pressure proportional to an electrical command signal input. The QB1X

More information

Installation and Instructions 2. Product Features 3-6. Key Pad Functions 7. Distributor Information Programming Guide 8. Master Programming Guide 9-14

Installation and Instructions 2. Product Features 3-6. Key Pad Functions 7. Distributor Information Programming Guide 8. Master Programming Guide 9-14 Installation and Instructions 2 Product Features 3-6 Key Pad Functions 7 Distributor Information Programming Guide 8 Master Programming Guide 9-14 Dimensional Drawing 15 D-STC & D-SMM Valve Assembly 16-17

More information

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1.

SPECIFICATIONS Type: Twin stack, single phase Tank: 4 gallon Air Output: PSI; PSI Max PSI: 125 PSI HP: 1. 2 GALLON TWIN STACK AIR COMPRESSOR Model: 9526 DO NOT RETURN TO STORE. Please CALL 800-348-5004 for parts and service. CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or

More information

Overview. Front Panel: Keypad and Display

Overview. Front Panel: Keypad and Display Overview The GA-200B is an analyzer that integrates a gas sampling system with sensors to measure and display the concentrations of oxygen and carbon dioxide in a sample as the percentage of a gas in the

More information

NTF-515. Operation Manual

NTF-515. Operation Manual NTF-515 Operation Manual RTI Technologies, Inc. 10 Innovation Drive York, Pennsylvania 17402 USA Phone: 717-840-0678 Toll Free: 800-468-2321 Web-site: www.rtitech.com Manual P/N: 035-81474-00 (RevA) TABLE

More information

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight

AC1810 / AC1810-A TECHNICAL SPECIFICATIONS. Operating Pressure psi ( kgs/cm²) [AC1810] Displacement. Net Weight Technical Specifications Operating Instructions Maintenance Information Troubleshooting Guide Parts Diagrams AC1810 / AC1810-A THE EVOLUTION OF PERFECTION CAUTION: Before attempting to use or service this

More information

CART FOR CENTURION 1500+

CART FOR CENTURION 1500+ CART FOR CENTURION 1500+ Operation & Maintenance Manual www.ameriwater.com 800-535-5585 AmeriWater 3345 Stop 8 Rd. Dayton, OH 45414 98-0161 Rev A Contents 1. GENERAL INFORMATION... 1 1.1. Features...

More information

AIR-OPERATED DOUBLE DIAPHRAGM PUMP USER S MANUAL

AIR-OPERATED DOUBLE DIAPHRAGM PUMP USER S MANUAL 00, 0, 000 00, 000, 00 A. TECHNICAL INFORMATION Model 00 Inlet/Outlet " Air Inlet /" 0 / 000 /" /" 00 /" /" 000 /" /" 00 /" /" Flow Rate GPM/ 0LPM GPM/ 0LPM GPM/ LPM GPM/ LPM GPM/ 0LPM GPM/ 0LPM Maximum

More information

Interface Devices, Inc. Hydraulic Mini Mule

Interface Devices, Inc. Hydraulic Mini Mule Interface Devices, Inc. Hydraulic Mini Mule INSTALLATION, OPERATION & MAINTENANCE MANUAL IMPORTANT! FILE THIS MANUAL IN A SAFE PLACE FOR FUTURE SERVICE & PARTS NEEDS ALWAYS REFERENCE THE SERIAL NUMBER

More information

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products

SATURN EVO COLOR A N A E S T H E S I A S Y S T E M. Manufacturer of high quality anaesthesia products Saturn Evo SATURN EVO COLOR A N A E S T H E S I A S Y S T E M i n t e l l i g e n t s i m p l i c i t y Manufacturer of high quality anaesthesia products SATURN EVO COLOR PATIENT SAFETY AND USER PATIENT

More information

HumiSys HF High Flow RH Generator

HumiSys HF High Flow RH Generator HumiSys HF High Flow RH Generator Designed, built, and supported by InstruQuest Inc. Versatile Relative Humidity Generation and Multi-Sensor System The HumiSys HF is a high flow version of the previously

More information

! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES

! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES ! WARNING! IMPORTANT SAFETY INSTRUCTION AND GUIDELINES! WARNING! IMPORTANT HPA AIR TANK SAFETY INSTRUCTION AND GUIDELINES This Paintball Marker is NOT A TOY. Misuse can cause serious injury or death. It

More information

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS)

Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Superconducting Susceptometer (MPMS-5S) Quantum Design Room 296 (MPMS) Sensitivity: 1x10 11 A m 2 Applied DC fields: 0 T to 5 T Applied AC fields: 0 G to 3 G (zero-to-peak), 0.01 Hz to 1000 Hz Temperatures

More information

Glove Box Installation Manual

Glove Box Installation Manual Glove Box Installation Manual 1998 by M. Braun Company File: GB-UNI-INS.DOC! Edition 08-00 by M. Boutin! Subject to be changed without notice Glovebox Installation Your Glove box has been fully assembled,

More information

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008

UNITY 2 TM. Air Server Series 2 Operators Manual. Version 1.0. February 2008 UNITY 2 TM Air Server Series 2 Operators Manual Version 1.0 February 2008 1. Introduction to the Air Server Accessory for UNITY 2...2 1.1. Summary of Operation...2 2. Developing a UNITY 2-Air Server method

More information

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution:

EasyFit Lite Nasal Mask, Silicone. Nasal Mask. EasyFit Lite Nasal Mask. Device Description and Instructions for Use. Caution: EasyFit Lite Nasal Mask EasyFit Lite Nasal Mask, Silicone Nasal Mask Device Description and Instructions for Use Caution: Federal Law (U.S.A.) restricts this device to sale by or on the order of a physician.

More information

MACS Mask CPAP System

MACS Mask CPAP System The MACS CPAP delivery system is under US patent protection as part of the pneuton Ventilator (Patent # 6,591,835) MACS Mask CPAP System Manufactured by: Airon Corporation 129 West Hibiscus Boulevard Suite

More information

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation

Appendix D: SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor. Description and Principle of Operation Page 1 of 19 : SOP of INNOVA 1412 Photoacoustic Multi-Gas Monitor Description and Principle of Operation The photoacoustic multi-gas monitor (INNOVA 1412, Innova AirTech Instruments, Denmark) is a highly

More information

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL

SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL www.modsafe.com SERVICE MANUAL MODEL BA050BMST BREATHING AIR PANEL WARNING: Do not attempt to operate this equipment without first reading and understanding the service manual enclosed with this device.

More information

STATEMENTS OF FACT BENCH BG-4 CONTEST. 1. A positive pressure leak could be caused by a leakage in or at device components.

STATEMENTS OF FACT BENCH BG-4 CONTEST. 1. A positive pressure leak could be caused by a leakage in or at device components. STATEMENTS OF FACT BENCH BG-4 CONTEST 1. A positive pressure leak could be caused by a leakage in or at device components. 2. The battery in the Sentinel should be replaced every 6 months. 3. Dow Corning

More information

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS

Equinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS TM controlled ventilation Equinox II NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E USER TABLE OF CONTENTS CHAPTER TITLE PAGE 1 INTRODUCTION 1.1 Equinox II Administration

More information

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10

Installation and operating manual. Pneumatic control station LK product no: PCS 1-10 LK product no: PCS 1-10 Article no: 74503 Revision:8 Article no: 74503 Revision: 8 2 (23) Contents 1. General information... 5 2. Safety precautions... 5 2.1 Significance of symbols... 5 2.2 Explanatory

More information

Operating Instructions Part No

Operating Instructions Part No DIGITAL AUTOMATIC TYRE INFLATOR Operating Instructions Part No. 11.0545 Thank you for selecting this Jamec Pem Automatic Tyre Inflator. Please read this manual before carrying out any installation or service

More information

Freedom8 ShoeBox Compressor Manual

Freedom8 ShoeBox Compressor Manual Freedom8 ShoeBox Compressor Manual Warning!! This product is not a toy! Use or misuse can cause severe injury or death! Use only with adult supervision. This unit is only to be used with tanks, hoses and

More information

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments

Pegas 4000 MF Gas Mixer InstructionManual Columbus Instruments Pegas 4000 MF Gas Mixer InstructionManual Contents I Table of Contents Foreword Part I Introduction 1 2 1 System overview... 2 2 Specifications... 3 Part II Installation 4 1 Rear panel connections...

More information

T EK-SUB 4800C 19 mm Submersible Level Transmitter

T EK-SUB 4800C 19 mm Submersible Level Transmitter Technology Solutions T EK-SUB 4800C 19 mm Submersible Level Transmitter Instruction Manual Document Number: IM-4800C www.tek-trol.com Table of Contents 1 Safety Instructions... 2 1.1 Intended Use... 2

More information

STATEMENTS OF FACT BENCH BG-4 CONTEST. 1. A positive pressure leak could be caused by a leakage in or at device components.

STATEMENTS OF FACT BENCH BG-4 CONTEST. 1. A positive pressure leak could be caused by a leakage in or at device components. STATEMENTS OF FACT BENCH BG-4 CONTEST 1. A positive pressure leak could be caused by a leakage in or at device components. 2. The battery in the Sentinel should be replaced every 6 months. 3. Dow Corning

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.

U S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. U S E R M A N U A L 1600 SERIES OXYGEN REGULATOR 168715G (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090

More information

accidents which arise due to non-observance of these instructions and the safety information herein.

accidents which arise due to non-observance of these instructions and the safety information herein. 3 GALLON PANCAKE COMPRESSOR Model: 50959 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement, or other masonry.

More information

O-Two Self-Study Guide. Equinox Relieve

O-Two Self-Study Guide. Equinox Relieve O-Two Self-Study Guide Equinox Relieve PRODUCT DESCRIPTION/PRINCIPLES OF OPERATION The Equinox Relieve provides spontaneously breathing patients with a preset mixture of oxygen and nitrous oxide for analgesia

More information

FX134, FX1234, FX3030 Problem and Solutions

FX134, FX1234, FX3030 Problem and Solutions FX134, FX1234, FX3030 Problem and Solutions Unit Will Not Power Up Turn the power switch on. The unit LCD will display revision program and filter life within 3 to 5 seconds after turning on unit. If this

More information

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #:

ID-Cube. Installation and Operations Manual For ID-Cube Source and Power Supply. Revision 2 Document #: ID-Cube Installation and Operations Manual For ID-Cube Source and Power Supply Revision 2 Document #: 7.1.002 IonSense Inc. 999 Broadway Suite 404 Saugus, MA 01906 Copyright 2014 by IonSense Inc. All rights

More information

Discontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves.

Discontinued. Powers Controls. Technical Instructions Document No P25 RV Rev. 1, May, RV 201 Pressure Reducing Valves. Powers Controls RV 201 Pressure Reducing Valves Description Features Product Numbers Dual Pressure PRV Technical Instructions Document No. 155-049P25 RV 201-1 Single Pressure PRV The RV 201 Pressure Reducing

More information

6900 Maintenance Instruction System Flush

6900 Maintenance Instruction System Flush Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex

More information

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual

WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, PHONE #: (970) FAX #: (970) TO-Clean Manual WSN: XXXXXX 101 ROME COURT FORT COLLINS, CO, 80524 PHONE #: (970) 221-9179 FAX #: (970) 221-9364 TO-Clean Manual Copyright 2008 by Wasson ECE Instrumentation, Inc. (SJC 02/05/2008) All rights reserved

More information

Oxygen Contamination

Oxygen Contamination MPMS Application Note 1014-210 Oxygen Contamination This application note describes potential sources for oxygen contamination in the sample chamber and discusses its possible effects. Molecular oxygen,

More information

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure

In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure PPMS Service Note 1099-412 In Response to a Planned Power Outage: PPMS EverCool II Shut Down and Re-start Procedure Introduction: Loss of electricity to the PPMS EverCool II should not cause damage to

More information

SOP for use of the SX19 Stopped-Flow Instrument

SOP for use of the SX19 Stopped-Flow Instrument SOP for use of the SX19 Stopped-Flow Instrument The main components of the stopped flow system are: Always book the system before use. Starting up and preparing the SX19 for use: 1. Turn on the compressed

More information

Technical Data Sheet MF010-O-LC

Technical Data Sheet MF010-O-LC Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the

More information

Exercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE

Exercise 8. Closed-Loop Pressure Control, Proportional-Plus-Integral Mode EXERCISE OBJECTIVE Exercise 8 Closed-Loop Pressure Control, EXERCISE OBJECTIVE To understand open and closed-loop pressure control; To learn how to sense the pressure in a pneumatic circuit; To control the pressure in a

More information