Laser DopplerVelocimetryvs Heater Poweras Indicatorsof Skin Perfusion duringtranscutaneous02 Monitoring

Size: px
Start display at page:

Download "Laser DopplerVelocimetryvs Heater Poweras Indicatorsof Skin Perfusion duringtranscutaneous02 Monitoring"

Transcription

1 CLN. CHEM. 27/3, (1981) Laser DopplerVelocimetryvs Heater Poweras ndicatorsof Skin Perfusion duringtranscutaneous02 Monitoring Louis Enkema, Jr.,1 G. llen Holloway, Jr.,2 Daniel W. Piraino,3 David Harry,1 Gregory L. Zick,3 and Margaret. Kenny 4 Laser Doppler velocimetry (LDV) and heater power (HP) were compared as indicators of skin perfusion during transcutaneous oxygen tension (p2) monitoring in adults. The WV probe was fitted to a Radiometer Ptco2 attachment ring, which was fixed to the forearm. ptco2, electrode temperature, HP, and LDV measurements were recorded continuously and simultaneously. LDV was more specific and sensitive to changes in cutaneous perfusion. Moreover, changes in perfusion were signalled four to 24 times faster by WV than by HP. Changes in p2 corresponded to changes in LDV more closely than to changes in HP in all experiments. HP was more specific and sensitive to changes in electrode and air temperatures. Once hyperemia was established at 44 #{176}C and the electrode heater was turned off (HP = 0), LDV measurements frequently showed persistence of hyperemia (up to 25 mm). Under these circumstances ptco2 decreased linearly with decreasing electrode temperature in the range of #{176}C. LDV performance was referenced by monitoring velocity measurements taken on a cylinder revolving at known velocities. We conclude that LDV is a useful adjunct to Ptc.02monitoring and better than HP in assessing skin blood f low. t should enhance quality control of transcutaneous 02 monitoring. ddtonal Keyphrases: blood gases. peripheral perfusion skin blood flow Transcutaneous oxygen-tension measurements (po2) are now a widely accepted alternative to arterial whole-blood oxygen tension determinations (Pa02) in the management of acutely ill neonates because they are continuous and noninvasive (1-3). However, the correlation of ptco2 with Pa02 depends in part upon adequate peripheral cutaneous perfusion at the measurement site (2). For this reason, transcutaneous 02 electrodes incorporate heating elements, which can be set to #{176}C and which thus thermally induce a localized hyperemia when the electrode is fixed on the skin. The power required to maintain the heating element at the preset temperature, heater power (HP), has frequently been used as an indicator of changes in perfusion. Peabody et al. (4) have challenged the validity of HP as a measure of cutaneous blood flow. However, until the recent development by Stern et al. (5) and Watkins and Holloway (6) of a device that monitors superficial capillary blood flow by laser Doppler velocimetry (LDV), there was no practical alternative. Clinical Chemistry Division, Department of Laboratory Medicine, University of Washington School of Medicine, Seattle, W Center for Bioengineering, University of Washington. Department of Electrical Engineering, University of Washington. To whom communications should be addressed. Nonstandard abbreviations used: p,o2, transcutaneous oxygen tension; Pe02, oxygen tension in arterial blood; LDV, laser Doppler velocimetry (or laser Doppler velocimeter); HP, heater power. Received Nov. 6, 1980; accepted Dec. 16, The LDV instrument operates on the same Doppler principle as that used for diagnostic ultrasound devices, except that light is substituted for sound energy. Light from a 5-mW He-Ne laser (frequency 4.75 X 1014 Hz) is transmitted to the skin through a quartz optical fiber. The light is then backscattered from both stationary skin components and moving objects, primarily erythrocytes in the capillaries, which are encountered as the laser light penetrates to an estimated maximum depth of mm. The backscattered light, which is transmitted to a photodiode through another optical fiber, consists of a mixture of frequency-shifted and -nonshifted waves. These different frequencies are optically mixed on the surface of the photodiode and produce a beat frequency which reflects the Doppler shift. The photodiode output signal is composed of a spectrum of Doppler-shifted frequencies, generally in the range of 1-8 khz, caused by the different velocities of erythrocytes encountered. From this, a single flow parameter is obtained by taking the normalized root mean square bandwidth of the signal, which has been shown (5) to be proportional (r = 0.89) to peripheral cutaneous perfusion as measured by radioisotope (xenon) clearance studies. n this paper, using the LDV integrated with a commercially available transcutaneous oxygen monitor (Radiometer /S, Copenhagen), we report the comparison of HP and LDV as indicators of cutaneous perfusion during transcutaneous oxygen monitoring in healthy adult subjects, and we demonstrate circumstances in which LDV is a better measure of peripheral perfusion. Methods pparatus and Materials Our transcutaneous oxygen-measurement system is the Radiometer TCM-1 monitor (The London Co., Cleveland, OH 44145), which was calibrated and operated as described by Graham and Kenny (1) except that a 25-tm Teflon (fluorinated ethylene-propylene, FEP type) electrode membrane replaced the polypropylene membrane and that no spacer was used. ll measurements were made with 44 #{176}C electrode temperature except where noted. They are reported in mmhg (1 mmhg = 133 Pa = 133 N/m2). four-channel recorder was used to display the PtcO2, electrode temperature, heater power, and laser Doppler velocimetry signals simultaneously at a chart speed of either 25 mm/mm or 5 mm/mm. The LDV apparatus was that described by Piraino et al. (7), with some modifications: Our laser source (Spectra-Physics nc., Santa Clara, C 95050; Model 120), a 5-mW helium neon laser, is more powerful and our optical fibers have a smaller core diameter (63 nm). n addition, the LDV output voltage is proportional to root mean square bandwidth, our flow parameter, which was electronically calculated according to the algorithm of Stern et al. (5,8). The LDV unit is a preproduction prototype (Nuclear Pacific, nc., Seattle, W 98108). The optical fibers for the LDV are permanently mounted in the Radiometer transcutaneous oxygen electrode retaining ring as illustrated in Figure 1. The two glass optical fibers, located 1 mm apart and 2 mm from the interior edge of the CLNCL CHEMSTRY, Vol. 27, No. 3,

2 TranscutonsouS 02 Electrode Lf Fiber LOV Optical Fibers 6 B C Electrode Rstaining Ring Fg. 1. Diagram llustrating the coupling of the Radiometer transcutaneous 02 electrode assembly with the laser Doppler velocimeter (LDV) probe () Transcutaneous electrode; (6) electrode attachment ring wth optical fibers ncorporated;(c) ranscutaneous electrode nstalled n the modif led attachment rng. The afferent (1) and efferent (2) optcal fbers of the wv probe are imbeddednthe electrode retaningring (4) perpendcularto and flush with the skin contact surface. The electrode hester (5) and the temperature probe (3) are located in the anode, which surrounds the cathode (6) ring, are secured with epoxy resin, then polished until they are flush with the ring s bottom surface. To secure the ring to the skin without obstructing the light beam, a small notch is cut in the double-sided adhesive rings used for attachment (item no. 2181; 3-M Co., St. Paul, MN 55101). The entire length of the glass fibers is encased in polyethylene tubing to protect them. The LDV apparatus was turned on for at least 20 mm before all experiments and was left in the operating mode between experiments, to maximize electrical stability. velocity calibration unit was fabricated by the Medical nstruments Facility of this institution. t consists of a synchronous clock motor (120 V, 7 W, 60 Hz), which drives a precision gear train with ratios of 1:10, 1:1,4.5:1, and This four-speed transmission is linked to a solid acrylic cylinder 19.1 mm in diameter. Five millimeters above the cylinder s curved surface the laser probe holder is fixed so that when in place, the laser s beam strikes the surface at an 80_boo angle. The cylinder surface rotates at 0.4,4.0, 18.0, and 40 mm/mm, 4 respectively, with the above gears. These speeds were selected to bracket the estimated range of erythrocyte velocities in the capillary (9). Experimental Procedure nstrument calibration. We made 20 determinations of LDV at each of the four speeds. The gear ratios were changed between each LDV measurement. Neither the motor nor the laser system was turned off at any time during the experiment. LDV drift was assessed with the probe on a stationary surface for 24 h. n vivo experiments. Before each experiment the output of the laser Doppler instrument was set to zero while holding it stationary in the air 90 to 120cm from a fixed surface. This procedure was used to subtract the electronic noise and thus permit a zero reading for flow on a stationary target. fter this adjustment, the ring containing the flow sensor was attached to the flexor surface of the forearm within 5 cm of the antecubital fossa. hairless site was selected, to facilitate attachment and to eliminate potential optical interference from hair trapped near the laser beam. The initial LDV mean backscattered light level was observed as an index of adequate probe function. Values greater than 500 mv were acceptable. fter confirming the LDV performance subsequent to positioning, a calibrated oxygen electrode was screwed into the attachment ring with the heater off. The experiments were performed as soon as the po2 measurements stabilized. Subjects. Seventeen ostensibly healthy adult volunteers, years of age and without excessive skin tanning at the test site, participated in 23 experiments. Each monitoring experiment lasted from h, during which the subject was supine in a hospital bed with head and test arm in the horizontal position (except as noted) before the electrode was applied and during experimental maneuvers. The maneuvers were sequenced so that each individual would serve as his or her own control. Results n Vitro Experiments The precision data for the LDV calibration are based on 20 sets of measurements. Each set consisted of one measurement at each speed (Table 1). The data in Table 1 show that LDV is directly proportional to velocity to 18 mm/mm (r = 0.994). t 40 mm/mm the output fails to represent velocity because of the 17-kHz limit in frequency response. The average drift in the zero point is ±50 mv/h. During normal heated cutaneous flow this drift affects output by less than 1%, even though this prototype was not designed for optimal electrical stability. ( new system designed to meet more stringent drift criteria is already operational.) n Vivo Experiments The first set of 15 experiments on 11 subjects involved heating the electrode from the normal resting skin-surface temperature of #{176}C directly to 44 #{176}C. Heater power was maximum until the 44 #{176}C temperature was reached. LDV increased gradually from 195 ± 90 mv, requiring 12 ± 6 mm Table 1. Precision Data for LDV Calibration Velocty, mm/mln LDV. my, mean (SD) 0-5(20) (13) (47) (40) 392 CLNCLCHEMSTRY,Vol. 27, No. 3, 1981

3 pto2, electrode temperature, HP, and LDV measured on one patient with a cuff rapidly inflated to 200 mmhg at point. The LDV decreased in 20s from an equilibrium value of 1400 mv to a value of 250 mv (in this instance 250 = zero flow) - during total occlusion. The HP current changed from 330 mw 75. at equilibrium before cuffing to 275 at equilibrium during #{231} cuffing. The pcj2 decreased from 92 too mmhg after 3 mm, of cuffing. When the cuff pressure was released suddenly and completely, the LDV increased rapidly (point B). n 50% of the experiments, the equilibrium value was restored within 3 a. n all other experiments, the equilibrium value was mi- N tially exceeded but was restored within 3 min. The over- shoot ranged up to + 125% of the initial reading. The 0-50% maximum response time of LDV is s. The response times of HP and po2 calculated in the same fashion were and s, respectively. n seven subjects, progressive partial blood flow occlusions with cuffs were performed, starting at 20 mmhg applied #{163} Temperature (#{176}C) Fig. 2. Effect of increasing electrode temperature on p2 and on WV Data from one experment on a single volunteer to reach a steady-state value of 400 ± 160 (± values are ranges), an increase of 105% over control values. Ptho2 also increased from 20±20 to 82± 15 mmhg but at a much slower rate, requiring 17 ± 5 mm. second set of experiments, closely related to the first, involved increasing the electrode temperature in three subjects from 30 to 45 #{176}C in stepwise increments to demonstrate the effect of increasing electrode temperature on po2, HP, and LDV. ll three variables were allowed to equilibrate before each successive increase in electrode temperature. Figure 2 illustrates the fmdings in one subject. puo2 remained zero below 35 #{176}C. t increased to 35 mmhg at 40 #{176}C and 52 mmhg at 42#{176}C, where it stabilized. t 45#{176}C the (cj.2 reached 95 mmhg and stabilized within 3 mm. LDV changed little between 30 and 42 #{176}C. t 45 #{176}C it increased from 150 mv to a new equilibrium of 1050 mv in less than 1 min. s shown in the comparative data in Table 2, HP went to 800 mw with each successive increase in electrode temperature. Subsequently, it established a new equilibrium that had no predictable relation to the previous or successive equilibria. Once the 44 #{176}C temperature had been reached, three different sets of experiments were performed: total arterial occlusion, partial occlusion, and modification of the external environment of the probe. n six subjects, total arterial occlusion was accomplished by using a blood-pressure cuff inflated to give a pressure mmhg greater than systolic pressure. typical recorder tracing (Figure 3) illustrates the pressure, and increasing in 20-mmHg increments every 20s up to 120 mmhg. Figure 4 shows results for two subjects, and B. LDV decreased about 20% in and 30% in B for each 20 mml-g increase in pressure up to about 80 mmhg, when zero LDV was attained in and 150 mv was attained in B. The small changes in heater power averaged <1% for each 20-mmHg increment in pressure. po2 diminished but only by about 5 mmhg with each 20 mmhg cuff pressure increase initially. t declined more rapidly to about minhg at the higher pressures of 60 and 80 mmhg. The third set of experiments, at 44 #{176}C on five subjects, involved modulation of the environment surrounding the probe (Figure 5). n the first subset, hot air from a hair dryer was blown for mm over the probe and skin from points to B, to generate an increase in ambient temperature. The probe temperature increased from 45 to 47#{176}C in <30 s. Under these conditions, HP decreased to zero and remained at or near that value until the hot air was discontinued. LDV in- 200 :3 t Table 2. Effects of ncreasing Electrode Tempeiature on Ptco2, HP, and LDV p2, nital Equlibrum mmhg HP, mw HP, mw LDV, mv Temp. oc urns (mm) Fig. 3. Simultaneous recordings of transcutaneous 02 monitor, electrodetemperature, HP, and LDV measurements during a blood flowocclusionexperiment on one volunteer s arm Cuff pressure rapdly applied at. Cuff released at B. Zero flow equals 250 mv CLNCL CHEMSTRY, Vol. 27, No. 3,

4 SO 80 t t B C F. 0-4C j\ B B Time (min 50 S. Fig. 5. Comparison of changes in LDV flow and heater power with changes in environmental temperature Hotair applied at and discontinuedat B. Unheated air applied at C anddiscontinued at Cuff Pressure (mmhg) Fig. 4. Comparison of heater power and laser Doppler velocimetry (WV) measurements made on two patients ( and B), while sequentially increasing the arm cuff pressure to occlude blood flow The sensor assenl,ly was located on the flexor surface of the forearm, the cuff on the upper arm. Equlbrium was established at each pressure. Electrode temperature, 44 #{176}C creased by 150% and remained increased for 3-25 mm after heating was stopped. PtcO2 increased slightly during the experiment and fell to previous values once the hot air was discontinued. n the second subset, a stream of unheated air was passed over the probe from points C to D in the same manner as the heated air in the previous experimental subset. Under these conditions HP increased and a new equilibrium, 60% higher than baseline, was established within 15 a. LDV and PtcO2 remained essentially unchanged. n a final set of experiments, we turned the electrode heater off to test the effect of decreasing electrode temperature on PtcO2 and blood-flow indicators. PtcO2 and electrode temperature were linearly related over the range of 44-32#{176}Cin six of seven subjects. Data from a typical subject are presented in Figure 6. LDV remained high and steady until the electrode temperature fell below 33 #{176}C. s the temperature continued to decline, the LDV decreased dramatically, from 0 mv at 33 #{176}C to 250 mv at 31 #{176}C. However, this two-degree decrease in electrode temperature required 26 mm, so the decrease in LDV was, in fact, rather gradual. The interval required for this temperature decrease varied for the six subjects from 5 to 30 mm. Linear regression analysis of PtcO2 vs temperature (from 44 to 32 #{176}C) for the six showed correlation coefficients ranging from to The slopes of the regression lines varied from 1.77 to 5.85 and averaged n these subjects there was no change in LDV for about 5-10 mm after the electrode temperature was allowed to decrease. LDV very gradually decreased during the subsequent 0-20 mm. n the seventh subject, although the electrode temperature decreased at the same rate as in the other six, the LDV decreased immediately, unlike the other six. The decrease in po in this subject was nonlinear in relation to the decrease in electrode temperature. Discussion There has long been a need for an instrument to measure cutaneous blood flow continuously and nonmnvasively. This is particularly true when measuring po2 transcutaneously, where the accuracy of the po2 as compared with Pa02 is determined by the adequacy of cutaneous perfusion. To achieve the high perfusion necessary to minimize the effect of cutaneous respiration and oxygen consumption, we incorporated a heater into the commercially available oxygen electrodes. The power delivered to this heater is commonly thought to vary directly with the extent of cutaneous perfusion. Because blood flowing through the dermis removes heat (thermal sink mechanism) the heater power has been viewed as a satisfactory indicator of cutaneous perfusion in the region of the transcutaneous oxygen electrode. Because heater power has many limitations and the laser Doppler velocimeter has numerous advantages, we have adapted the laser Doppler velocimeter to a Radiometer transcutaneous oxygen monitor to demonstrate the effective use of this new device in monitoring cutaneous perfusion while simultaneously monitoring PtcO2. The most serious objections to the LDV measurement of cutaneous blood flow relate to problems in calibration and drift. HP cannot be absolutely calibrated either, and its dependence on uncontrollable variables such as varying thermal conductivity of skin and changing environmental conditions is greater than is the case for LDV. The simple precision-gear device for calibrating the LDV signal against known velocities shows promise. The linearity shown over the range of signals 394 CLNCL CHEMSTRY, Vol. 27, No. 3, 1981

5 75, E c Temperature (#{176}C) Fig. 6. Effect of decreasing electrode temperature on p2 on WV Data from one experiment on a single volunteer and received under in vitro experimental conditions suggests that the calibration instrument is effective. One question remains: What does the signal from this calibration device represent? The cylinder does not compare to the complex geometry of the dermal capillary bed. t best, the calibration device, like the LDV device, is a method of making a relative measurement until a better standard of reference is developed. The results we have obtained in these experiments support our hypothesis that the LDV is more specific, more sensitive, and has a faster response time than does heater power as a measure of cutaneous perfusion. n three circumstances HP is unusable. When the heater is first turned on, the power output is maximum and it cannot be used as a monitor until thermal equilibrium with the skin is reached-a situation not always easily identified. Similarly, when the heater power is turned off to allow the electrode to cool, HP obviously cannot be used as an indicator of skin blood flow. n addition, HP cannot be used to compare blood flows at different sensor temperatures. LDV is interpretable under all these conditions. The cuffing experiments were undertaken to evaluate the relative rate and magnitude of response of the two blood-flow indicators to changes in cutaneous perfusion. Response times were calculated from the point of cuff release rather than the point of initial pressure application because inflating the cuff takes more time. s illustrated in Figure 3, the blood-flow indicators respond more rapidly than the po2. We defined the LDV response time as the time interval needed to achieve 50% of the equilibrium LDV after cuff release. This time varied among individuals but was consistent with an individual within an experiment. The response time of LDV following release of total occlusion is 1.5 s, whereas HP is 6-24 s and PtcO2 is s. The LDV is at least four to 16 times faster than HP. The actual LDV response time may be faster, because cuff release time and vessel compliance are known limiting factors in this experiment. Severinghaus et al. (10) used Po2 values obtained during forearm cuffing experiments to estimate skin blood flow and metabolic rate. They used conditions similar to ours, except their subjects breathed % 02 rather than room air; they used a polypropylene membrane rather than Teflon; the oc- clusions lasted min rather than 2 mix,; and the electrode temperature was 45 #{176}C rather than 44 #{176}C. Their mean 63% response time for prj2 after cuff release was 68.6 s. Our mean 50% response time was 45 s. The relative responses of LDV and HP to complete occlu sion of arterial inflow by blood-pressure-cuff inflation consistently show LDV decreasing to zero while HP decreases as #{231} little as 7% in some cases and up to 75% in others, presumably depending on environmental conditions. Heat losses through 50 the skin and electrode have a large direct and immediate effect on HP readings, but little if any direct effect on LDV. n our experience, changes in the immediate environment of the Li.. probe generated the largest discrepancies between LDV and. HP. Blowing hot air over the electrode caused a slight increase 25 C) in LDV and po2-presumably owing to an increase in blood a flow-but a marked decrease in HP. Conversely, blowing air at room temperature over the probe, a situation in which no increase in flow would be expected, resulted in no change in LDV or po2. However, HP increased markedly. Severinghaus et a). (10) estimate that under ideal conditions 70% of HP is lost to the tissues, and conclude that HP is a poor indicator of blood flow. When HP was turned off after hyperemia had been established, LDV remained stable for at least 5-10 mm in six of seven subjects, even though all heating of the electrode had stopped. LDV gradually declined, often rather slowly, thereafter. The pattern of LDV and PtcO2 after heat is turned off suggests the possibility that under carefully controlled conditions transcutaneous monitoring of po2 might be extended beyond the 3-h limit now imposed to avoid burning the patient. This would involve developing a new protocol for heating the transcutaneous electrode. Continuous transcutaneous monitoring would be contingent on an ability to make appropriate corrections for oxygen tension determinations at diminished temperatures. We are currently investigating the feasibility of such a Po2 monitoring protocol as one application of the po2-ldv coupled system. lthough our comparative analyses of LDV and HP as blood flow indicators deal with greater challenges than one would generally encounter in a clinical setting, we have touched on characteristics of each flow monitor that are relevant to clinical applications. recent article by Whitehead et al. (11) has aroused concern over the smaller po2 values seen when the electrode is placed over ribs or sternum. The LDV device would have signalled the lowered flow, alerting us to this problem with po2, whereas HP oould never provide this information. Decreases in po2 readings caused by pressure when patients have been rolled onto the electrode during surgical procedures were not attended by changes in HP but would be signalled by decreases in LDV. n neonatal intensive-care units, changes in HP must be interpreted cautiously in view of the measurement s extreme susceptibility to change with changes in incubator air temperature, drafts, and the patient s movements. HP susceptibility to noise (i.e., environmental factors that do not influence blood flow but do affect HP) make it much less valuable than the LDV in measuring blood flow. The relative sensitivity and specificity of LDV make it an excellent adjunct to quality control in po2 monitoring. HP will be of even more limited value than it is now if proposed extended monitoring permits HP to be turned off for intervals during continuous pcj2 monitoring. n summary, we conclude that the laser Doppler velocimeter can be successfully coupled to the Radiometer transcutaneous po2 monitor. When used in this manner the LDV is more specific, more sensitive, and has a more rapid response time than HP as a measure of cutaneous blood flow. We believe that simultaneous assessment of LDV and po2 will provide a more CLNCLCHEMSTRY,Vol. 27, No. 3,

6 accurate and physiological understanding of changes in pj2 in the critically ill patient. The LDV data provide a valuable new dimension to Puo2 quality control under circumstances where HP is frankly erroneous or not available. dditionally, when cutaneous blood flow remains increased for considerable periods of time after the heater power has been turned off, cycling of the heater may be possible while po2 monitoring continues without adverse effect. These experiments show how LDV can be applied to the development of transcutaneous monitoring that is more convenient, safe, accurate, and reliable. This work was supported in part by NH grants HL23711, C06331, and HLO The transcutaneous oxygen monitor was provided by Radiometer /S. Copenhagen, Denmark. References 1. Graham, G., and Kenny, M.., Performance of a Radiometer transcutaneous oxygen monitor in a neonatal-intensive-care unit. Clin. Chem. 26, (1980). 2. Eberhard, P., Continuous Monitoring of Newborns by Skin Sensors. Offset Press, Basel, Switzerland, Lofgrenn, 0., On transcutaneous P02 measurements in humans. University of Lund, Malmo, Sweden, Thesis. 4. Peabody, J. L., Willis, M. M., Gregory, G.., and Severinghaus, J. W., Reliability of skin (te) P02 electrode heating power as a continuous noninvasive monitor of mean arterial pressure in sick newborns. Birth Defects 15, (1979). 5. Stern, M. D., Lappe, D. L., Chimosky, J. E., et a)., Continuous measurement of tissue blood flow by laser-doppler spectroscopy. m. J. Physiol. 232, H441-H448 (1977). 6. Watkins, D., and Holloway, G.. Jr., n instrument to measure cutaneous blood flow using the Doppler shift of laser light. EEE Trans. Biomed. Eng. BME-25, (1978). 7. Piraino, D. W., Zick, G. L., and Holloway, G.., Jr., n instrumentation system for the simultaneous measurement of transcutaneous oxygen and skin blood flow. n Frontiers of Engineering in Health Care, EEE Press, New York, NY, cat. 79CH1440-7, 1979, pp Stern, M. D., n vivo evaluation of microcirculation by coherent light scattering. Nature 254, (1975). 9. Butti, P., ntaglietta, M., Reimann, H. eta!., Capillary red cell blood velocity measurements in the human nailfold by video densitometric methods. Microvasc. Res. 10, 220 (1975). 10. Severinghaus, J. W., Stafford, M., and Thunstrom,. M., Estimation of skin metabolism and blood flow with tcpo2 and tcpco2 electrodes by cuff occlusion of the circulation. cta naesthesiol. Scand., (Suppl.) 68,9-15 (1978). 11. Whitehead, M. D., Pollitzer, M. J., and Reynolds, E. 0., rtifactual hypoxemia during estimation of Pa02 by skin electrode. Lancet ii, 157, 21 (1979). Letter. 396 CLNCLCHEMSTRY,Vol. 27, No. 3, 1981

BIOPAC Blood Flow Monitor

BIOPAC Blood Flow Monitor BIOPAC Blood Flow Monitor Technique - Laser Doppler Flowmetry (LDF) The LDF100C is a laser Doppler blood perfusion monitor used to measure real-time micro-vascular red blood cell perfusion in tissue. LDF100C

More information

TRANSCUTANEOUS CO2 MONITORING (TCpCO2) PURPOSE

TRANSCUTANEOUS CO2 MONITORING (TCpCO2) PURPOSE PURPOSE tcpco2 monitoring provides continuous non invasive monitoring of an infant s ventilatory status. The sensor consists of a Stow-Severinghaus-type electrode. The use of tcpco2 monitoring may facilitate

More information

Solutions for transcutaneous monitoring

Solutions for transcutaneous monitoring Solutions for transcutaneous monitoring About Radiometer Radiometer is a leading provider of technologically advanced acute care solutions that simplify and automate all phases of acute care testing. By

More information

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy

Real-time Analysis of Industrial Gases with Online Near- Infrared Spectroscopy Real-time Analysis of Industrial Gases with Online Near- Infrared Branch 5 Petrochemical, Biofuels Keywords Process, NIR, Gases, Gas Cell, Ethylene, Carbon Dioxide, Propane, Butane, Acetylene Summary Near-infrared

More information

Estimation of arterial oxygen tension in adult

Estimation of arterial oxygen tension in adult Estimation of arterial oxygen tension in adult subjects using a transcutaneous electrode D C S HUTCHISON, G ROCCA, AND D HONEYBOURNE From the Chest Unit, King's College Hospital, London Tharax, 1981, 36,

More information

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111

DEVICES FOR FIELD DETERMINATION OF WATER VAPOR IN NATURAL GAS Betsy Murphy MNM Enterprises 801 N. Riverside Drive Fort Worth, Texas 76111 INTRODUCTION Water vapor in natural gas has more than a substantial effect on the quality of the gas stream. Without quality measurement of water vapor the gas is basically not saleable. Contracts are

More information

transients' of large amplitude can be imposed on the arterial, cardiac and Since both coughing and the Valsalva manoeuvre raise intrathoracic pressure

transients' of large amplitude can be imposed on the arterial, cardiac and Since both coughing and the Valsalva manoeuvre raise intrathoracic pressure 351 J. Physiol. (I953) I22, 35I-357 EFFECTS OF COUGHING ON INTRATHORACIC PRESSURE, ARTERIAL PRESSURE AND PERIPHERAL BLOOD FLOW BY E. P. SHARPEY-SCHAFER From the Department of Medicine, St Thomas's Hospital

More information

SenTec OxiVenT Illuminate Ventilation and Oxygenation PCO2 PO2. Digital Transcutaneous Blood Gas Monitoring

SenTec OxiVenT Illuminate Ventilation and Oxygenation PCO2 PO2. Digital Transcutaneous Blood Gas Monitoring Digital Transcutaneous Blood Gas Monitoring SenTec OxiVenT Illuminate Ventilation and Oxygenation MONTHS COMPLETE WARRANTY PCO2 PO2 Continuous Noninvasive Safe Easy to Use Accurate OxiVenT overcoming limitations

More information

plethysmographic methods that when the subject was pinched on the upper

plethysmographic methods that when the subject was pinched on the upper 24 J. Physiol. (I95I) II2, 24-2I 6I2.I5.6II.976 THE DECREASE IN HAND BLOOD FLOW FOLLOWING INFLATION OF AN ARTERIAL OCCLUSION CUFF ON THE OPPOSITE ARM BY IAN C. RODDIE From the Department of Physiology,

More information

LABORATORY EXPERIMENTS ON WAVE OVERTOPPING OVER SMOOTH AND STEPPED GENTLE SLOPE SEAWALLS

LABORATORY EXPERIMENTS ON WAVE OVERTOPPING OVER SMOOTH AND STEPPED GENTLE SLOPE SEAWALLS Asian and Pacific Coasts 23 LABORATORY EXPERIMENTS ON WAVE OVERTOPPING OVER SMOOTH AND STEPPED GENTLE SLOPE SEAWALLS Takayuki Suzuki 1, Masashi Tanaka 2 and Akio Okayasu 3 Wave overtopping on gentle slope

More information

Medical Instruments in the Developing World

Medical Instruments in the Developing World 2.2 Ventilators 2.2.1 Clinical Use and Principles of Operation Many patients in an intensive care and the operating room require the mechanical ventilation of their lungs. All thoracic surgery patients,

More information

SenTec Digital Monitoring System Illuminate Ventilation and Oxygenation in Neonatology PCO2 PO2

SenTec Digital Monitoring System Illuminate Ventilation and Oxygenation in Neonatology PCO2 PO2 Digital Transcutaneous Blood Gas Monitoring SenTec Digital Monitoring System Illuminate Ventilation and Oxygenation in Neonatology PCO2 PO2 Continuous Noninvasive Accurate Safe Easy to Use Transcutaneous,

More information

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator

Title: Standard Operating Procedure for R&R Environmental Devices Model MFC201 Gas Dilution Calibrator Procedure No: SOP-029 Revision No: 1.1 (December 29, 2010) Page No.: 1 of 7 1. INTRODUCTION AND SCOPE To obtain timely data for the purpose of air quality assessment, air quality trend reporting, air quality

More information

(Received 9 September 1940)

(Received 9 September 1940) 257 J. Physiol. (I 94I) 99, 257-264 6I2.2II A METHOD OF RECORDING THE RESPIRATION BY J. H. GADDUM From the College of the Pharmaceutical Society, 17 Bloomsbury Square, London, W.C. 2 (Received 9 September

More information

Lung Volumes and Capacities

Lung Volumes and Capacities Lung Volumes and Capacities Normally the volume of air entering the lungs during a single inspiration is approximately equal to the volume leaving on the subsequent expiration and is called the tidal volume.

More information

Characterizers for control loops

Characterizers for control loops Characterizers for control loops By: F. G. Shinskey (May 1999) Introduction Commercial controllers such as the PID series (proportional, integral, derivative, and their combinations) are linear devices

More information

APPENDIX. working blood volume was also rather large; Evans, Grande, and. equilibrated to the new mixture is partially dependent upon the rate

APPENDIX. working blood volume was also rather large; Evans, Grande, and. equilibrated to the new mixture is partially dependent upon the rate 612.172-5 APPENDIX A SIMPLIFIED HEART OXYGENATOR CIRCUIT FOR BLOOD- FED HEARTS. By J. YULE BOG-UE and R. A. GREGORY.' SINCE 1934 studies on the carbohydrate metabolism of the blood-fed heart without lungs

More information

Comparative temperature measurements in an experimental borehole heat exchanger. Vincent Badoux 1, Rita Kobler 2

Comparative temperature measurements in an experimental borehole heat exchanger. Vincent Badoux 1, Rita Kobler 2 European Geothermal Congress 2016 Strasbourg, France, 19-24 Sept 2016 Comparative temperature measurements in an experimental borehole heat exchanger Vincent Badoux 1, Rita Kobler 2 1 GEOTEST AG, Bernstrasse

More information

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE

STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE STEAM STERILIZABLE POLAROGRAPHIC OXYGEN ELECTRODE GENERAL INFORMATION This family of dissolved oxygen electrodes is of the steam sterilizable polarographic type, designed to be interchangeable with NBS

More information

General Accreditation Guidance. User checks and maintenance of laboratory balances

General Accreditation Guidance. User checks and maintenance of laboratory balances General Accreditation Guidance User checks and maintenance of laboratory balances January 2018 Copyright National Association of Testing Authorities, Australia 2010 All intellectual property rights in

More information

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies

29th Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies IMPROVED CALIBRATION AND OPERATION OF A STABLE-GAS QUANTIFICATION MANIFOLD AS PART OF A RADIOXENON COLLECTION SYSTEM ABSTRACT Richard M. Williams, James C. Hayes, and Randy R. Kirkham Pacific Northwest

More information

FLUORESCENCE DETERMINATION OF OXYGEN

FLUORESCENCE DETERMINATION OF OXYGEN FLUORESCENCE DETERMINATION OF OXYGEN Objectives This experiment will familiarize the students with the principles of fluorescence quenching as well as the Ocean Optics FOXY Fiber Optic Sensor System. Spectra

More information

OptonPro. High-power laser up to 7000 mw. Laser therapy natural healing with the power of light

OptonPro. High-power laser up to 7000 mw. Laser therapy natural healing with the power of light GB High-power laser up to 7000 mw Laser therapy natural healing with the power of light Application of high-power light for targeted penetration of high amounts of energy deep into the tissue. Initiates

More information

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017

Dissolved Oxygen and measurement possibilities. Berno Lüpkes, 15 th March 2017 Dissolved Oxygen and measurement possibilities Berno Lüpkes, 15 th March 2017 Content 1. Introduction to Dissolved Oxygen 2. Amperometric measurement principle 3. Optical measurement principle 4. Optical

More information

Chapter 7. SCIIB Pressure Sensor Performance Evaluations: Experiments, Results and Discussions

Chapter 7. SCIIB Pressure Sensor Performance Evaluations: Experiments, Results and Discussions : Experiments, Results and Discussions This chapter will summarize the experiments and results associated with the development of the single-mode fiber-based SCIIB pressure sensor system. We will start

More information

Generating Calibration Gas Standards

Generating Calibration Gas Standards Technical Note 1001 Metronics Inc. Generating Calibration Gas Standards with Dynacal Permeation Devices Permeation devices provide an excellent method of producing known gas concentrations in the PPM and

More information

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V

O2100C Oxygen Measurement Module Technical Use Notes do not use other wall adapters with the O2100Cmodule. 10% / V 5% / V 2% / V 1% / V 10% / V O2100C Oxygen Measurement Module The O2100C module measures the partial pressure of O2 and thus the module output is proportional to the pressure in the sample cell. Gas sampled must be free of liquids

More information

Linear Infl ation Technology Non-invasive Blood Pressure Measurement. inibp Case Report Experiences in OR

Linear Infl ation Technology Non-invasive Blood Pressure Measurement. inibp Case Report Experiences in OR Linear Infl ation Technology Non-invasive Blood Pressure inibp Case Report Experiences in OR C ONTENTS Introduction 2 Technological Description of inibp inibp advantages 2 inibp technology adapts to each

More information

Measuring range Δp (span = 100%) Pa

Measuring range Δp (span = 100%) Pa 4.4/ RLE 5: Volume-flow controller, continuous How energy efficiency is improved Enables demand-led volume flow control for the optimisation of energy consumption in ventilation systems. Areas of application

More information

IT is frequently desirable to be able to determine the

IT is frequently desirable to be able to determine the NEW INVENTIONS A PORTABLE OXYGEN ANALYSER By A. BRACKEN Research and Development Department, The British Oxygen Company, Limited IT is frequently desirable to be able to determine the oxygen concentration

More information

This test shall be carried out on all vehicles equipped with open type traction batteries.

This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4. Determination of hydrogen emissions page 1 RESS-6-15 5.4.1. This test shall be carried out on all vehicles equipped with open type traction batteries. 5.4.2. The test shall be conducted following

More information

Electrochemical Sensor

Electrochemical Sensor ME3-O3 Electrochemical Sensor Manual (Model:ME3-O3) Version V1.2 Valid from May 1 st,2014 Zhengzhou Winsen Electronics Technology Co., Ltd ME3-O3 gas sensor ME3-O3 electrochemical sensor detect gas concentration

More information

Analysis of Pressure Rise During Internal Arc Faults in Switchgear

Analysis of Pressure Rise During Internal Arc Faults in Switchgear Analysis of Pressure Rise During Internal Arc Faults in Switchgear ASANUMA, Gaku ONCHI, Toshiyuki TOYAMA, Kentaro ABSTRACT Switchgear include devices that play an important role in operations such as electric

More information

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers

A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers 2011 ROCKY MOUNTAIN NASA SPACE GRANT CONSORTIUM 1 A CO 2 Waveform Simulator For Evaluation and Testing of Respiratory Gas Analyzers Christina Long, and Joseph Orr, Ph.D. Department of Bioengineering, University

More information

EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY

EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY 18 th World Conference on Non destructive Testing, 16-20 April 2012, Durban, South Africa EXPERIMENTAL RESULTS OF GUIDED WAVE TRAVEL TIME TOMOGRAPHY Arno VOLKER 1 and Hendrik VOS 1 TNO, Stieltjesweg 1,

More information

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER

METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1250 METHOD 25A - DETERMINATION OF TOTAL GASEOUS ORGANIC CONCENTRATION USING A FLAME IONIZATION ANALYZER 1.0 Scope and Application. 1.1 Analytes. Analyte CAS No. Sensitivity Total Organic Compounds N/A

More information

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES

METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES METHOD 3C - DETERMINATION OF CARBON DIOXIDE, METHANE, NITROGEN, AND OXYGEN FROM STATIONARY SOURCES 1. Applicability and Principle 1.1 Applicability. This method applies to the analysis of carbon dioxide

More information

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters.

New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters. New Generation System M, leading the World in the Non-Invasive Measurement of Critical Real-Time Parameters. System M Spectrum Medicals total commitment to continuous product improvement is demonstrated

More information

Application Note 268 Metabolic Analysis Systems and Propane-based Verification

Application Note 268 Metabolic Analysis Systems and Propane-based Verification Application Note 268 Metabolic Analysis Systems and Propane-based Verification 8.10.11 This application note examines a conventional equipment setup for metabolic measurements. The setup is exposed to

More information

Visual Observation of Nucleate Boiling and Sliding Phenomena of Boiling Bubbles on a Horizontal Tube Heater

Visual Observation of Nucleate Boiling and Sliding Phenomena of Boiling Bubbles on a Horizontal Tube Heater Proceedings of the 2 nd World Congress on Mechanical, Chemical, and Material Engineering (MCM'16) Budapest, Hungary August 22 23, 216 Paper No. HTFF 146 DOI:.11159/htff16.146 Visual Observation of Nucleate

More information

1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics

1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics 1 Design of the Portable mass spectrometric sensor for vascular and endocrine disease diagnostics Andrei Antonov, Viktor Kogan, Aleksandr Krutikov A.F. Ioffe Physical Technical Institute, Academy of Science,

More information

5.1 Introduction. Learning Objectives

5.1 Introduction. Learning Objectives Learning Objectives 5.1 Introduction Statistical Process Control (SPC): SPC is a powerful collection of problem-solving tools useful in achieving process stability and improving capability through the

More information

Level MEASUREMENT 1/2016

Level MEASUREMENT 1/2016 Level MEASUREMENT 1/2016 AGENDA 2 A. Introduction B. Float method C. Displacer method D. Hydrostatic pressure method E. Capacitance method G. Ultrasonic method H. Radar method I. Laser method J. Level

More information

Introductory Lab: Vacuum Methods

Introductory Lab: Vacuum Methods Introductory Lab: Vacuum Methods Experiments in Modern Physics (P451) In this lab you will become familiar with the various components of the lab vacuum system. There are many books on this topic one of

More information

Electrochemical Sensor

Electrochemical Sensor ME3-H2 Electrochemical Sensor Manual (Model:ME3-H2) Zhengzhou Winsen Electronics Technology Co., Ltd ME3-H2 gas sensor ME3-H2 electrochemical sensor detect gas concentration by measuring current based

More information

485 Annubar Primary Flow Element Installation Effects

485 Annubar Primary Flow Element Installation Effects ROSEMOUNT 485 ANNUBAR 485 Annubar Primary Flow Element Installation Effects CONTENTS Mounting hole diameter Alignment error Piping Geometry Induced Flow Disturbances Pipe reducers and expansions Control

More information

Homeostasis and Negative Feedback Concepts and Breathing Experiments 1

Homeostasis and Negative Feedback Concepts and Breathing Experiments 1 Homeostasis and Negative Feedback Concepts and Breathing Experiments 1 I. Homeostasis and Negative Feedback Homeostasis refers to the maintenance of relatively constant internal conditions. For example,

More information

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience -

Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Certification of AMS acc. EN 15267, Part 3 - Overview and First Experience - Dr. Wolfgang Jockel, Martin Schneider, TÜV Rheinland Group, D-51105 Cologne / Germany 1. Introduction A new basis for the certification

More information

Radnoti Liver Perfusion System

Radnoti Liver Perfusion System Radnoti Liver Perfusion System 130003 Offset Offset Gain Gain Radnoti 2006 Description Qty Part # A Base only, for 4-bar stand 1 159950-B4 B Stabilizer Bar only, for 4-Bar stand 2 159950-C4 C Rod 24 Long

More information

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage

ISE-730 Oxygen Electrode and DO2-100 Currentto-Voltage Technical Note ISE-730 and DO2-100 Overview In 1954, Dr. Leland Clark invented the first membrane-covered electrode designed to measure the concentration of oxygen in blood, or solution. This electrode

More information

(fig. 3) must be at the same temperature as the water in this chamber CALORIMETRIC STUDIES OF THE EXTREMITIES

(fig. 3) must be at the same temperature as the water in this chamber CALORIMETRIC STUDIES OF THE EXTREMITIES CALORIMETRIC STUDIES OF THE EXTREMITIES II. EXPERIMENTAL APPARATUS AND PROCEDURES' By ROY KEGERREIS (Received for publication July 1, 1926) The calorimeter used in these experiments is a modification of

More information

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist

CDI Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist Original Article Blood Parameter Monitoring System 500 A New Tool for the Clinical Perfusionist David W. Fried, MS Ed, CCP; Joseph J. Leo, BS, CCP; Gabriel J. Mattioni, BS, CCP; Hasratt Mohamed, CCP; Raymond

More information

The Variation of Muscle Oxygen Consumption With Velocity of Shortening

The Variation of Muscle Oxygen Consumption With Velocity of Shortening The Variation of Muscle Oxygen Consumption With Velocity of Shortening R.J. BASKIN From the Department of Zoology, University of California, Davis ABSTRACT Total oxygen consumption following contraction

More information

ME3-HCL Electrochemical sensor

ME3-HCL Electrochemical sensor ME3-HCL Electrochemical sensor Manual (Model:ME3-HCL) Zhengzhou Winsen Electronics Technology Co., Ltd ME3-HCL gas sensor ME3-HCL electrochemical sensor detect gas concentration by measuring current based

More information

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS

COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS COMPARISON OF DIFFERENTIAL PRESSURE SENSING TECHNOLOGIES IN HOSPITAL ISOLATION ROOMS AND OTHER CRITICAL ENVIRONMENT APPLICATIONS APPLICATION NOTE LC-136 Introduction Specialized spaces often times must

More information

Application Note AN-107

Application Note AN-107 SPEC Sensor TM Characterization & Calibration Considerations Scope This document is provided to describe the considerations needed to characterize, calibrate, verify and validate the measurement performance

More information

Improving distillation tower operation

Improving distillation tower operation Improving distillation tower operation Measuring differential pressure across long sections of distillation columns has always been challenging, but purpose-built sensor systems provide a solution Fast

More information

Lucid System In-Service Training

Lucid System In-Service Training Lucid System In-Service Training DC-00172 Indications for Use The Lucid M1 System is a medical ultrasound system intended for use as an adjunct to the standard clinical practices for measuring and displaying

More information

MULTICHANNEL. Labnet International, Inc.

MULTICHANNEL. Labnet International, Inc. Labnet International, Inc. A B A1 A2 C F D H E 4A 4B 4C 4D G CONTENTS 1 - INTRODUCTION 2 - VOLUME SETTING 3 - METHOD OF PIPETTING 4 - RECOMMENDATIONS - RECALIBRATION 6 - CLEANING AND STERYLIZATION 7 -

More information

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip.

Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip. Lab 1: Pressure and surface tension. Bubblers, gravity and the mighty paper clip. CEE 3310 - Summer 2012 SAFETY The major safety hazard in this laboratory is a shock hazard. Given that you will be working

More information

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives

The physiological functions of respiration and circulation. Mechanics. exercise 7. Respiratory Volumes. Objectives exercise 7 Respiratory System Mechanics Objectives 1. To explain how the respiratory and circulatory systems work together to enable gas exchange among the lungs, blood, and body tissues 2. To define respiration,

More information

What is ultrasonic plethysmography and how can we use it in airway disease?

What is ultrasonic plethysmography and how can we use it in airway disease? What is ultrasonic plethysmography and how can we use it in airway disease? Christian Buess ndd Medizintechnik AG, Zürich, Switzerland ERS 2007-1/19 Background Until approx. the year 2000 mainly the following

More information

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER

LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER LINEAR TRANSFORMATION APPLIED TO THE CALIBRATION OF ANALYTES IN VARIOUS MATRICES USING A TOTAL HYDROCARBON (THC) ANALYZER Michael T Tang, Ph.D. Grace Feng Greg Merideth Rui Huang Matheson Gas Applied Lab

More information

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator

Title: Standard Operating Procedure for Dasibi Model 5008 Gas Dilution Calibrator Procedure No: SOP-034 Revision No: 1.0 Revised Dec. 29, 2010 Page No.: 1 of 10 1. INTRODUCTION AND SCOPE This procedure is intended to describe the operations of the Dasibi model 5008 calibrator. The Dasibi

More information

Flow in a shock tube

Flow in a shock tube Flow in a shock tube April 30, 05 Summary In the lab the shock Mach number as well as the Mach number downstream the moving shock are determined for different pressure ratios between the high and low pressure

More information

CO2 incubator with additional process controls

CO2 incubator with additional process controls CB series 150 C2 incubators C2 incubator with additional process controls The BINDER CB series C2 incubator is the premium class among the C2 incubators. It is suitable for all sensitive incubation applications

More information

Electrochemical Carbon Monoxide Gas Sensor (Model:ME3-CO) Manual. Version 1.2 Valid from May 1 st 2014

Electrochemical Carbon Monoxide Gas Sensor (Model:ME3-CO) Manual. Version 1.2 Valid from May 1 st 2014 Electrochemical Carbon Monoxide Gas Sensor (Model:ME3-CO) Manual Version 1.2 Valid from May 1 st 2014 Zhengzhou Winsen Electronics Technology Co., Ltd ME3-CO Carbon Monoxide Gas Sensor ME3-CO electrochemical

More information

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT

OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT OXYGEN CONSUMPTION AND TEMPERATURE IN THE AQUATIC ENVIRONMENT BACKGROUND READING Animal Physiology by Hill, Wyse & Anderson, 2004: pp. 130 139 & 198 201. PRE-LAB (Due at the start of the lab) ** In your

More information

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator)

VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF performance verification of Bunnell Life-Pulse HFJV (High Frequency Jet Ventilator) VT PLUS HF provides a special mode for evaluating the performance of high frequency ventilators while connected

More information

Model 130M Pneumatic Controller

Model 130M Pneumatic Controller Instruction MI 017-450 May 1978 Model 130M Pneumatic Controller Installation and Operation Manual Control Unit Controller Model 130M Controller is a pneumatic, shelf-mounted instrument with a separate

More information

Columbus Instruments

Columbus Instruments 0215-003M Portable O 2 /CO 2 /CH 4 Meter User s Manual Columbus Instruments 950 NORTH HAGUE AVENUE TEL:(614) 276-0861 COLUMBUS, OHIO 43204, USA FAX:(614) 276-0529 1 www.colinst.com TOLL FREE 1-800-669-5011

More information

Laser-Induced Bubbles in Glycerol-Water Mixtures

Laser-Induced Bubbles in Glycerol-Water Mixtures Laser-Induced Bubbles in Glycerol-Water Mixtures Allison R. McCarn, Erin M. Englert, and Gary A. Williams Department of Physics and Astronomy, University of California, Los Angeles, California 90095, USA

More information

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION

CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE 3-4 BATTERY INSTALLATION IFU SBPMON107 CONTENTS SPECIFICATIONS GENERAL INFORMATION RECOMMENDED USE OPERATING PRINCIPLE TIPS ON TAKING YOUR BLOOD PRESSURE BATTERY INSTALLATION CORRECT POSITION FOR MEASUREMENT POSITIONING THE CUFF

More information

Precision level sensing with low-pressure module MS

Precision level sensing with low-pressure module MS The task on hand Level sensing as it is understood in modern process automation is much more than simply "tank half full" or "tank a quarter full". Using suitable sensors, levels, inlets and outlets can

More information

Improved Algorithm for Measurement of Blood Pressure based on a Laser Doppler Flowmetry Signal

Improved Algorithm for Measurement of Blood Pressure based on a Laser Doppler Flowmetry Signal Master of Science Thesis in Biomedical Engineering Department of Biomedical Engineering, Linköping University, 2016 Improved Algorithm for Measurement of Blood Pressure based on a Laser Doppler Flowmetry

More information

Oxygen Flue Gas Analyzer BA 2000

Oxygen Flue Gas Analyzer BA 2000 Gas Analysis Oxygen Flue Gas Analyzer BA 2000 Some combustion processes, e.g. process heaters, steam boilers or heating furnaces the air required to achieve optimal system efficiency can easily fluctuate.

More information

Technical Data Sheet MF010-O-LC

Technical Data Sheet MF010-O-LC Technical Data Sheet MF010-O-LC - 1 - 1. Properties The oxygen measuring system MF010-O-LC determines the oxygen content in gas mixtures up to a temperature of 250 C. It is particularly suitable for the

More information

Name. Lab Partners. The Dive Response. In this lab exercise, you will be designing and conducting your own experiments!

Name. Lab Partners. The Dive Response. In this lab exercise, you will be designing and conducting your own experiments! Name Lab Partners The Dive Response In this lab exercise, you will be designing and conducting your own experiments! Note: A formal report is due for this lab (see handout). Objectives 1. Characterize

More information

Ultrasound Dose Calculations

Ultrasound Dose Calculations Ultrasound Dose Calculations One of the advantages of ultrasound therapy remains the reasonably broad range of trials from which effective treatment doses can be established. In principle, there is no

More information

Dissolved Oxygen Guide

Dissolved Oxygen Guide Educat i onser i es Di ssol vedoxygengui de Dissolved Oxygen Guide Introduction Dissolved oxygen probes provide a convenient approach to essentially direct measurement of molecular oxygen. The membrane

More information

Module No GETTING ACQUAINTED GENERAL GUIDE TIMEKEEPING

Module No GETTING ACQUAINTED GENERAL GUIDE TIMEKEEPING Module No. 2196 2196-1 GETTING ACQUAINTED Congratulations upon your selection of this CASIO Pressure Monitor Watch (BP-1B, Module No. 2196). To get the most out of your purchase, be sure to carefully read

More information

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars.

better measurement Simply a question of SCHMIDT Flow Sensor SS The cost-effective alternative in pressurised systems up to 10 bars. Simply a question of better measurement SCHMIDT Flow Sensor SS 20.261 The cost-effective alternative in pressurised systems up to 10 bars. Compressed air technology Industrial processes A cost analysis

More information

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA

11/22/ (4) Harmonization: <846> SPECIFIC SURFACE AREA BRIEFING 846 Specific Surface Area, USP 27 page 2385. The European Pharmacopoeia is the coordinating pharmacopeia for the international harmonization of the Specific Surface Area General Chapter, as part

More information

Magnetic Field Therapy-Does Affect Soft Tissue? *

Magnetic Field Therapy-Does Affect Soft Tissue? * 0196-601 1 /83/0404-0241$02.00/0 THE JOURNAL OF ORTHOPAEDIC AND SPORTS PHYSICAL THERAPY Copyright O 1983 by The Orthopaedic and Sports Physical Therapy Sections of the American Physical Therapy Association

More information

Gas Pressure. Pressure is the force exerted per unit area by gas molecules as they strike the surfaces around them.

Gas Pressure. Pressure is the force exerted per unit area by gas molecules as they strike the surfaces around them. Chapter 5 Gases Gas Gases are composed of particles that are moving around very fast in their container(s). These particles moves in straight lines until they collides with either the container wall or

More information

Application Block Library Fan Control Optimization

Application Block Library Fan Control Optimization Application Block Library Fan Control Optimization About This Document This document gives general description and guidelines for wide range fan operation optimisation. Optimisation of the fan operation

More information

SEMEM3-78 Testing and calibrating instrumentation and control equipment and circuits

SEMEM3-78 Testing and calibrating instrumentation and control equipment and circuits Testing and calibrating instrumentation and control equipment and Overview This unit identifies the competences you need to carry out tests and calibration of instrumentation and control equipment and,

More information

Introduction to AAE520 Forebody/LDV Lab

Introduction to AAE520 Forebody/LDV Lab Introduction to AAE520 Forebody/LDV Lab Learn to use a water tunnel Learn about the laser doppler velocimeter (LDV), use it for some measurements Learn about flow visualization Learn about subsonic forebodies

More information

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE

Chapter 4: Ventilation Test Bank MULTIPLE CHOICE Instant download and all chapters Test Bank Respiratory Care Anatomy and Physiology Foundations for Clinical Practice 3rd Edition Will Beachey https://testbanklab.com/download/test-bank-respiratory-care-anatomy-physiologyfoundations-clinical-practice-3rd-edition-will-beachey/

More information

LEAKHUNTER Plus Model 8066

LEAKHUNTER Plus Model 8066 LEAKHUNTER Plus LEAKHUNTER Plus Model 8066 Description The Matheson LEAKHUNTER Plus is a truly universal leak detector. Extensive R&D has produced a multi-functional leak detector engineered to perform

More information

Table of Contents. By Adam Hollingworth

Table of Contents. By Adam Hollingworth By Adam Hollingworth Table of Contents Oxygen Cascade... 2 Diffusion... 2 Laws of Diffusion... 2 Diffusion & Perfusion Limitations... 3 Oxygen Uptake Along Pulmon Capillary... 4 Measurement of Diffusing

More information

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson

A Journal of Practical and Useful Vacuum Technology. By Phil Danielson A Journal of Practical and Useful Vacuum Technology From By Phil Danielson Thermal Conductivity Gauges Thermal conductivity pressure gauges are extremely common in vacuum technology, but an understanding

More information

BLOOD PRESSURE SENSOR 0377i

BLOOD PRESSURE SENSOR 0377i BLOOD PRESSURE SENSOR 0377i USER S GUIDE CENTRE FOR MICROCOMPUTER APPLICATIONS http://www.cma-science.nl Short description The Blood Pressure sensor 0377i is used to measure arterial blood pressure in

More information

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION

TG GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION GUIDELINES CONCERNING CALIBRATION INTERVALS AND RECALIBRATION Approved By: Senior Manager: Mpho Phaloane Revised By: Field Manager: Neville Tayler Date of Approval: 2015-08-26 Date of Implementation: 2015-08-26

More information

MIL-STD-883G METHOD

MIL-STD-883G METHOD STEADY-STATE LIFE 1. PURPOSE. The steady-state life test is performed for the purpose of demonstrating the quality or reliability of devices subjected to the specified conditions over an extended time

More information

Instrumentation & Data Acquisition Systems

Instrumentation & Data Acquisition Systems Instrumentation & Data Acquisition Systems Section 3 -Level Robert W. Harrison, PE Bob@TheHarrisonHouse.com Made in USA 1 Level Section Question Which level measuring technology is the best solution when

More information

AN31E Application Note

AN31E Application Note Balancing Theory Aim of balancing How an unbalance evolves An unbalance exists when the principle mass axis of a rotating body, the so-called axis of inertia, does not coincide with the rotational axis.

More information

High Frequency Acoustical Propagation and Scattering in Coastal Waters

High Frequency Acoustical Propagation and Scattering in Coastal Waters High Frequency Acoustical Propagation and Scattering in Coastal Waters David M. Farmer Graduate School of Oceanography (educational) University of Rhode Island Narragansett, RI 02882 phone: (401) 874-6222

More information

Autodesk Moldflow Communicator Process settings

Autodesk Moldflow Communicator Process settings Autodesk Moldflow Communicator 212 Process settings Revision 1, 3 March 211. Contents Chapter 1 Process settings....................................... 1 Profiles.................................................

More information

Vibration Analysis and Test of Backup Roll in Temper Mill

Vibration Analysis and Test of Backup Roll in Temper Mill Sensors & Transducers 2013 by IFSA http://www.sensorsportal.com Vibration Analysis and Test of Backup Roll in Temper Mill Yuanmin Xie College of Machinery and Automation, Wuhan University of Science and

More information