Analgesic and Oxygen Demand Valve. Service Manual
|
|
- Philippa Spencer
- 6 years ago
- Views:
Transcription
1 Analgesic and Oxygen Demand Valve Service Manual May 2014
2
3 Contents 1. Symbols Warnings and Cautions Warnings! Cautions! Notes Functional Description Intended Use Technical Description Maintenance User Maintenance and Inspection When to Repair When to Disinfect Specification Servicing and Repair Procedures Demand Valve Disassembly Cleaning Cleaning and Inspection Procedure Reassembly Disinfection Procedure Demand Valve Assembly External Disinfection Demand Valve Component Disinfection Testing Troubleshooting Parts and Spares Demand Valve Main Components Distributor Details... 19
4 1. Symbols Warning! Caution! Note Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user or others Indicates a potentially hazardous situation which, if not avoided, could result in damage to the equipment or property Highlights points that might allow more convenient or efficient operation of the equipment Attention, consult accompanying documents Use no oil 0120 CE Marked to Medical Device Directive 93/42/EEC as amended by directive 2007/47/CE No smoking Single use or Single patient use Service due date 1304X12345 YYMM Date of manufacture identification Use by 2. Warnings and Cautions 2.1. Warnings! Read through this entire instruction manual before using or showing others how to use this demand valve. As with all medical equipment, attempting to use this device without a thorough understanding of its operation may result in patient or user injury. In the United States of America, Federal Law restricts this device to sale by or on the order of a physician. Only appropriately trained personnel working in controlled conditions may disassemble or assemble this demand valve. Do not smoke around analgesic gas equipment. Gas specific connectors are fitted to the demand valve. Do not attempt to modify the fittings to suit other gases or fitting systems. 2
5 N 2 O/O 2 50%/50% V/V, hereinafter referred to as analgesic gas, is or should be considered a drug and should only be used for medical purposes as prescribed by a physician or authorised clinician and in accordance with the medicinal product labelling. Oxygen and nitrous oxide analgesic gas mixture is not flammable; however the presence of it will drastically increase the rate and severity of combustion. Oil and/or grease in the presence of an oxygen enriched atmosphere will become highly combustible. Analgesic gas must never be allowed to come into contact with oil, grease or other petroleum-based substances. Do not use oil or grease on this demand valve. Continuous exposure to elevated levels of nitrous oxide can be harmful. National standards for exposure levels (Time Weighted Average (TWA)), if available, should be applied. Risk controls measures such as adequate room ventilation, gas scavenging and/or environmental monitoring may be necessary. Oxygen therapy may be a critical treatment. Additional risk control measures such as SpO 2 monitoring with an alarm and a backup oxygen supply with delivery device should be provided for patients who might suffer distress or injury if their oxygen supply is not available or is stopped unexpectedly. Ensure that the medical gas supply is sufficient for the proposed therapy and is supplied within the pressure range given in the Device Specification. If the supply is a gas cylinder, check the cylinder contents gauge regularly. Do not use 100% oxygen or analgesic gas as a pressure medium to purge obstructed pipelines or equipment, to operate pneumatic tools, or to build up any pressure in tanks. Do not use or store medical gas near excessive heat (>50 C/125 F) or below 10 C (50 F). Always refer to the medical gas suppliers recommendations. Disassemble the device before cleaning. Do not submerge an assembled demand valve in any fluid. Never permit compressed medical gases to enter a pressure regulator suddenly. Always open the cylinder valve slowly. Secure cylinders to a wall, stand or cart in accordance with local safety regulations. Arrange the gas hose carefully to avoid damage to the hose and the potential for causing a trip hazard. Never pull or apply excessive force to the gas hose. A leaking hose may result in high local oxygen and nitrous oxide concentrations and an increased risk of fire. When therapy is complete, disconnect the demand valve from the gas supply. When the source of analgesic gas is from a gas cylinder, always close the gas cylinder valve when the demand valve is not in use and disconnect the demand valve from the pressure regulator (for reference only). The demand valve should not be used as a resuscitator (for reference only). If using a cylinder and regulator, ensure that the device is connected to the regulator and the cylinder valve is properly opened before beginning therapy (for reference only). Use a new exhalation valve for each new patient or after 30 days of use for the same patient (for reference only) Cautions! The performance of the demand valve may be affected if it is stored or transported in temperatures outside of the range -20 C to 60 C (-4 F to 140 F). The demand valve is not suitable for autoclaving. The handset is protected from contamination in normal use by a single patient use exhalation valve. 3
6 The single patient use exhalation valve is not suitable for cleaning. If the exhalation valve becomes visually soiled or discoloured it should be replaced for reference only) Notes Peak flow through the demand valve might be restricted resulting in increased work of breathing for the patient in the following circumstances: o o o o regulator or gas supply source does not meet the specification inlet filter is blocked an extension hose other than those listed in this manual is used hose is kinked or occluded 3. Functional Description 3.1. Intended Use The Ultraflow Demand Valve is intended to be used for self-administration of 100% oxygen or N 2 O/O 2 50%/50% V/V, hereinafter referred to as analgesic gas, in response to the patient s inspiratory effort. The demand valve is designed for use in all types of clinical environment. Always refer to the medical gas product labelling Technical Description The demand valve comprises of two main components; a demand valve handset and a single patient use exhalation valve (supplied separately). The exhalation valve is single patient use to prevent cross contamination between patients. It may be used by a single patient for up to 30 days. The demand valve handset is designed to be reused and can be cleaned and disinfected, although routine disinfection is not necessary as it is protected from contamination by the single patient use exhalation valve. The demand valve handset contains a specially designed tilt valve mechanism that opens when the diaphragm in the back of the demand valve handset moves forward as a result of the patient inhaling. When the tilt valve opens, the gas, which is under pressure behind the tilt valve, passes through the handset and is inhaled by the patient. The deeper the patient breathes, the greater the volume of gas delivered. When the patient exhales the diaphragm moves back, the tilt valve closes and no further gas is delivered. The demand valve offers a very low resistance to flow during both patient inhalation and exhalation, which means less effort for the patient. This is achieved by the unique, patented exhalation valve design that diverts exhaled gas out through a special valve, thus eliminating the need for the patient to exhale through a highly resistant patient filter as is the case with other demand valve systems. The demand valve can be driven directly from the terminal unit of a medical gas pipeline system or from a medical gas cylinder via a suitable pressure regulator. A gas hose complying with BS EN ISO 5359 carries the gas from the gas supply source to the demand valve. The demand valve should be used with either a mouthpiece or a facemask. The exhalation valve has a viral filter that allows the flow of inhaled gas to the patient and prevents contamination of the demand valve handset from the patient s expired breath. 4
7 4. Maintenance 4.1. User Maintenance and Inspection All demand valves should be inspected and tested on an annual basis to ensure correct performance. The demand valve handset must be serviced every 4 years to ensure that it continues to perform in accordance with its specification. Full details of this is covered in sections 6 through 8. Servicing must be carried out by a suitably qualified person working in a controlled environment When to Repair If form or function of the demand valve is affected. See sections 6 through When to Disinfect If contamination is suspected. See section 9. 5
8 5. Specification Specification Inspiratory Resistance Supply Pressure 1 Supply Flow Capacity 1 Demand Valve Peak Flow Environmental Regulatory Value <1.5 kpa (0.22 psi) at 200 l/min <0.25 kpa (0.036 psi) at 10 l/min Maximum 600 kpa (87 psi) Minimum 310 kpa (45 psi) >120 l/min >200 l/min Transport and Storage Temperature: -20 C to 60 C (-4 F to 140 F) Operating Temperature: 5 C to 40 C (41 F to 104 F) Humidity: 0-100% RH non-condensing CE: Medical Device Directive 93/42/EEC Active Medical Device Class IIa Applied Standards BS BS 5682 BS EN ISO BS EN ISO 5359 BS EN ISO BS EN ISO BS EN ISO NF S DIN SS CGA V-5 SANS 1409 AS 2896 Anaesthetic and analgesic machines. Specification for intermittent (demand) flow analgesic machines for use with 50/50% (V/V) nitrous oxide and oxygen Probes (quick connectors) for use with medical gas pipeline systems Anaesthetic and respiratory equipment. Conical connectors. Cones and sockets Low pressure hose assemblies for use with medical gases Medical devices. Application of risk management to medical devices Anaesthetic and respiratory equipment. Compatibility with oxygen Medical devices. Symbols to be used with medical device labels, labelling and information to be supplied. General requirements Medico-surgical equipment. Terminal units and related probes for medical fluids Supply systems for medical gases. Part 2: Dimensions and allocation of probes and gas specific connection points for terminal units for compressed medical gases and vacuum Medical gas pipeline systems. Connectors for medical gases DISS Diameter Index Safety System Outlet sockets and probes for medical (gas and vacuum) service used in hospitals Medical gas systems installation and testing of non-flammable medical gas pipeline systems 1 Indicates minium supply pressure at stated value of gas flow. Based upon an adult breathing at 30 bpm with a tidal volume of 1 litre and I:E ratio of 1:2 6
9 6. Servicing and Repair Procedures 6.1. Demand Valve Disassembly Note: A Demand Valve Service Kit contains replacement parts with the exception of the tilt valve and tilt valve seat. 1 Dispose of parts only if damaged or carrying out 4 year service. For repair, retain parts. Illustrations show blue analgesic cover, oxygen demand valves have a grey cover. Older devices have a black diaphragm. Equipment and Tools Required 2mm Allen key T-6 TORX driver 2mm steel punch Step Action Illustration 1. Ensure that the gas supply is turned off at the source (e.g. gas cylinder) and purge gas using test button. 2. Disconnect the device from the gas supply. 3. Remove lanyard retainer. 4. Remove and discard lanyard Prise the demand valve cover from the demand valve body and remove it by sliding it along the hose. 6. Remove the main label body. 7
10 7. Using a 2mm Allen key, remove and discard the hose retaining screws (2 off) from the main body Remove hose assembly whilst holding main body vertically in order to retain spring and tilt valve inside the main body. Discard the hose assembly, lanyard retaining collar and blue cover Remove the spring and tilt valve, taking care to avoid damage to the ruby seat on the tilt valve. Dispose of the spring 1. Do not dispose of the tilt valve. Caution! The tilt valve and seat are a matched pair unique to each device; they must not be mixed with similar items from other Ultraflow Demand Valves 10. Remove the tilt valve insert, taking care not to damage the seat area. Remove the O-ring from the insert and discard it 1. Do not dispose the tilt valve insert. Note If the tilt valve insert is not easily removable, push out from inside the main body neck using 2 mm punch. To prevent damage to the tilt valve seat, do not drive out from the centre of the tilt valve insert. 11. Using a T-6 TORX driver, remove and discard the finger guard TORX screws (2 off) To remove the finger guard: Feed a steel punch through the neck of the main body. Push on the locating tab located inside the neck. When the tab unseats from the locating hole in the main body, twist the finger guard and pull it out of the main body. Discard the finger guard 1. 8
11 13. Pull the centre of the diaphragm away from the body and work your fingers around the circumference until the diaphragm pulls away from the seat. Discard the diaphragm and main body Cleaning Warning! PPE should be used in line with local policy or in line with disinfectant instructions Cleaning and Inspection Procedure Warning! The cleaning procedure is not a disinfection procedure. Equipment and Tools Required Lint free swabs Isopropyl alcohol Carry out the following process to clean demand valve components in order to remove substances that may affect device performance. Step Action Illustration 1. Clean the following components using a lint free swab and isopropyl alcohol: a Tilt valve b Tilt valve ruby Note the following steps (1c thru 3) are not required for service as new parts are provided in Service Kit c Diaphragm seat d Insert area 2. Inspect all components for damage, including scratches, cracks and deformation 9
12 3. Replace components with new items where necessary. 8. Reassembly Note A Demand Valve Service Kit contains replacement parts with the exception of the tilt valve and tilt valve seat. Equipment Required IKV Fluor 70 lubricant 2mm steel punch Spare parts or service kit as required Step Action Illustration 1. Put a thin coating of IKV Fluor 70 lubricant around the circumference of the diaphragm. Note Older diaphragms are black in colour with a single rib on the top face and should always be replaced with the current blue diaphragm. 2. Insert the diaphragm into the outer groove in the main body; lining up the tab on the diaphragm with the cut-out in the main body 3. Ensure the diaphragm is fully seated by gently pressing down around the circumference of the diaphragm with the steel punch, ensuring there are no ripples. Take care not to perforate the diaphragm. 4. Push the finger guard into the main body, lining up the tab with the half cut-out. 5. Feed the steel punch through the neck of the main body and press down on tab located on finger guard. Whilst pressing down on tab, push finger guard fully home, ensuring the tab locks fully into place in the cut-out. 10
13 6. Fit replacement finger guard retaining TORX screws (2 off) (Part No ) 7. Fit a new O-ring (Part No ) to tilt valve insert. Apply a thin coating of IKV 70 lubricant to the O-ring. 8. Insert the tilt valve insert into the main body, ensuring that the holes align. 9. Fit rating label to main body. If repaired, re-use insert. If serviced, create new insert with service date 4 years from date of service. Annotate Year-Month (YYYY-MM) e.g Serial number to be copied from removed label. New label insert template available by visiting then clicking on Tools and Utilities and downloading the file Barcode label template.xlsm or you may contact cs@bprmedical.com to request a copy. Note Handset is tapered; ensure label is applied without wrinkles. 10. Feed the lanyard retainer and then the cover over the hose assembly, ensuring correct orientation of the cover. 11. Whilst holding the main body horizontally, slide the tilt valve into the tilt valve seat. Take extra care when inserting the tilt valve that it does not damage the insert seat. 11
14 12. Hold the main body vertical and locate the spring into the tilt valve head. 13. Apply IKV-Flour 70 (or equivalent) to both O- rings on hose tail. Align the screw holes on the thrust washer with the holes on the tilt valve insert. Insert the hose assembly into the tilt valve insert and fit replacement retaining screws (2 off). 14. Fit cover over the handset body. 15. Fit the lanyard over the handset body and into the groove in the cover, sliding the lanyard retainer over the lanyard. 9. Disinfection Procedure Warning! PPE should be used in line with local policy or in line with disinfectant instructions. The demand valve is not suitable for autoclaving. Do not immerse the demand valve handset in any fluid or attempt to disinfect internal parts whilst the demand valve handset is assembled. The Ultraflow Demand Valve has been tested and proven suitable for disinfection with a chlorine dioxide sporicidal fluid with ClO 2 concentrations up to 200 ppm 9.1. Demand Valve Assembly External Disinfection Step Action 1. Use a sporicidal wipe to thoroughly clean all external surfaces of the demand valve taking care to ensure all crevices are thoroughly covered. 2. Wipe all components thoroughly using a sterile cloth and water to ensure that all traces of the cleaning solution are removed. 3. Allow all components to air dry. 12
15 9.2. Demand Valve Component Disinfection Step Action 1. Dismantle the demand valve into its component parts accordance with section 6.1 (Demand Valve Disassembly). 2. With the exception of the hose assembly, carry out a cold disinfection process of the handset by submerging each of the components listed below in one of the following suitable disinfecting solutions in accordance with the solution suppliers recommendations. Pay particular attention not to exceed any recommended submersion time given: Suitable Disinfecting Solutions: Active Ingredient Sodium dichloroisocyanurate (NaDCC; 1000 ppm Cl) Sodium hypochlorite 1% (1000 ppm Cl) Chlorine dioxide (280 ppm available ClO2) Common Brands/Solutions Chlor-Clean H8950, Actichlor, Haz-Tab, Presept Bleach 0.55% ortho-phthaldehyde solution Cidex OPA Tristel Fuse for Instruments Component List: Various Demand Valve Cover Main Body Diaphragm Finger Guard Lanyard Retainer O-ring Lanyard 3. Remove the components from the solution and drain the excess back into the container. 4. Submerge the components in 8 litres of potable water and agitate each component. Leave the components submerged for 1 min. 5. Repeat section 6.2 step 4 two more times with a fresh batch of 8 litres of potable water. 6. Dry the components until no visible surface water is present. 7. Clean the tilt valve and tilt valve insert using a lint free swab and isopropyl alcohol. 8. Using an anti-bacterial wipe, clean the following components: Long socket cap head screw (2 off) Counter sunk head screw (2 off) Tilt Valve Spring Various Tilt Valve Tilt Valve Insert PVC Hose Assembly 13
16 10. Testing The testing procedure recommended for an Ultraflow Demand Valve is outlined in document Demand Valve Test Instrument, Instructions for Use. The Demand Valve Test Instrument is intended to verify that an Ultraflow Demand Valve is performing to the specifications outlined in BPR Medical s manufacturing procedures following service or repair. For further information on the Demand Valve Test Instrument or for a copy of document , please contact your local BPR distributor or visit Warnings! The demand valve is supplied with a gas specific connector that is designed to connect to a mating gas outlet. Gas outlets may be part of a terminal unit in a medical gas pipeline system or part of a pressure regulator outlet on a gas cylinder. Where the gas specific connector is dependent on a threaded fastener (e.g. DISS CGA V5 1240), make sure that the connection is tight before turning on the supply pressure. For quick connector probes (e.g. BS5682, SIS, AFNOR), ensure that the connection is secure by gently pulling the gas hose before turning on the supply pressure. If the demand valve has a regulator, this should be serviced in accordance with the Regulators for Medical Gases Service Manual ( ). Testing is required to ensure performance of device after servicing or repair. 11. Troubleshooting Fault Possible Cause Solution No gas flow Audible gas leak Demand valve handset is not connected properly. Gas cylinder empty. Medical gas terminal unit on a pipeline system is isolated. Demand valve probe blocked. Hole in diaphragm. The demand valve will work with the Test Button but not when used conventionally by inhaling through the mouthpiece or face mask. Valve or diaphragm has become dislodged. Tilt valve worn, bent or broken. Check gas supply. Check that the gas specific probe is correctly connected. Replace gas cylinder. Seek advice from someone authorised to operate the medical gas pipeline system isolation valves. Repair or service required. Repair or service required. Remove the exhalation valve and check that the diaphragm is laying flat and that it moves forwards and backwards when the Test Button is pressed Repair or service required. Constant gas flow Tilt valve damaged or blocked. Repair or service required. Insufficient gas flow Tilt valve damaged. Repair or service required. Diaphragm perforated. Supply pressure too low and/or gas cylinder nearly empty. Repair or service required. Check gas supply and/or replace gas cylinder. 14
17 Exhalation valve will not fit to or be retained on the demand valve body Single patient use exhalation valve has damaged location lugs. Check the underside of the single use exhalation valve to see if the locating lugs have been bent over or otherwise damaged. It is important to rotate the single use exhalation valve until it locates in the demand valve housing before pressing it down and engaging the locating ears. Trying to force the single patient use exhalation valve down before correctly locating it will damage the exhalation valve beyond repair. Fit a new single patient use exhalation valve. 15
18 12. Parts and Spares Demand Valve Main Components Item Part Number Quantity Description Finger Guard Diaphragm Finger Guard Torx Screw Main Body Cover Blue Analgesic Cover Grey Oxygen Tilt Valve O-ring (not shown) Tilt Valve assembly (some parts not shown) Lanyard (not shown) Lanyard Retainer Tilt Valve Spring Allen Screw 13 Various see service kit 1 Rating Label Analgesic (not shown) Various see service kit 1 Rating Label Oxygen (not shown) 14 Various see service kit 1 Retainer Label Analgesic (not shown) Various see service kit 1 Retainer Label Oxygen (not shown) 15 Various see service kit 1 Hose Assembly Various 16
19 Spare Parts and Servicing Demand Valve Factory Service (Universal) Oxygen Demand Valve Service Kit - 3m Hose - AFNOR Analgesic Demand Valve Service Kit - 2m Hose - AFNOR Analgesic Demand Valve Service Kit - 3m Hose - AFNOR Analgesic Demand Valve Service Kit - 4m Hose - AFNOR Analgesic Demand Valve Service Kit - 5m Hose - AFNOR Analgesic Demand Valve Service Kit - 6m Hose - AFNOR Oxygen Demand Valve Service Kit - 3m Hose - DIN Oxygen Demand Valve Service Kit - 3m Hose - Nordica Analgesic Demand Valve Service Kit - 2m Hose - Nordica AGA Analgesic Demand Valve Service Kit - 3m Hose - Nordica AGA Analgesic Demand Valve Service Kit - 4m Hose - Nordica AGA Analgesic Demand Valve Service Kit - 5m Hose - Nordica AGA Analgesic Demand Valve Service Kit - 6m Hose - Nordica AGA Oxygen Demand Valve Service Kit - 3m Hose - BS Analgesic Demand Valve Service Kit - 2m Hose - BS Analgesic Demand Valve Service Kit - 3m Hose - BS Analgesic Demand Valve Service Kit - 4m Hose - BS Analgesic Demand Valve Service Kit - 5m Hose - BS Analgesic Demand Valve Service Kit - 6m Hose - BS Oxygen Demand Valve Service Kit - 3m Hose - DISS Analgesic Demand Valve Service Kit - 2m Hose - DISS Analgesic Demand Valve Service Kit - 3m Hose - DISS Analgesic Demand Valve Service Kit - 4m Hose - DISS Analgesic Demand Valve Service Kit - 2m Hose - SANS Analgesic Demand Valve Service Kit - 3m Hose - SANS Analgesic Demand Valve Service Kit - 4m Hose - SANS Analgesic Demand Valve Service Kit - 5m Hose - SANS Analgesic Demand Valve Service Kit - 6m Hose - SANS Analgesic Demand Valve Service Kit - 2m Hose - SIS Analgesic Demand Valve Service Kit - 3m Hose - SIS Analgesic Demand Valve Service Kit - 4m Hose - SIS Analgesic Demand Valve Service Kit - 5m Hose - SIS Analgesic Demand Valve Service Kit - 6m Hose - SIS 17
20 Analgesic Demand Valve Service Kit - 2m Hose - BS 5682 (New Zealand) Analgesic Demand Valve Service Kit - 3m Hose - BS 5682 (New Zealand) Analgesic Demand Valve Service Kit - 4m Hose - BS 5682 (New Zealand) Analgesic Demand Valve Service Kit - 5m Hose - BS 5682 (New Zealand) Analgesic Demand Valve Service Kit - 6m Hose - BS 5682 (New Zealand) Analgesic Demand Valve Service Kit - 2m Hose - ¼ BSP Analgesic Demand Valve Service Kit - 3m Hose - ¼ BSP Analgesic Demand Valve Service Kit - 4m Hose - ¼ BSP Analgesic Demand Valve Service Kit - 5m Hose - ¼ BSP Analgesic Demand Valve Service Kit - 6m Hose - ¼ BSP Demand Valve Test Instrument Demand Valve Test Instrument Demand Valve Test Instrument supply hose - MA4 CPC (BS 5682) Demand Valve Test Instrument adaptor hose - MA4 N 2 O/O 2 (BS 5682) Demand Valve Test Instrument Adapter Hose MA4 O 2 (BS 5682) 18
21 13. Distributor Details United Kingdom BOC Healthcare Customer Service Centre Priestly Road Worsley Manchester M28 2UT T: +44 (0) F: +44 (0) Australia BOC Healthcare Customer Service Centre 10 Julius Avenue North Ryde Sydney NSW, 2113 T: +61 (0) F: +61 (0) New Zealand BOC Healthcare Customer Service Centre Great South Road Penrose 1061 Auckland New Zealand T: +64 (0) Canada BOMImed Inc. 96 Terracon Place Winnipeg MB R2J 4G& Canada T: F: BPR Medical Ltd License No: South Africa AFROX Hospital Care PO Box 5404 Johannesburg 2000 T: +27 (0) F: +27 (0) ENDV25/11/09/V2 19
22 20
23 21
24 BPR Medical Limited T: +44 (0) Hamilton Way, Mansfield F: +44 (0) Nottinghamshire NG18 5BU, United Kingdom
Analgesic Demand Valve. Instructions for Use
Analgesic Demand Valve Instructions for Use 702-0038.16 May 2014 Contents 1. Symbols... 2 2. Warnings and Cautions... 2 2.1. Warnings!... 2 2.2. Cautions!... 3 2.3. Notes... 3 3. Functional Description...
More informationOxygen Demand Valve. Instructions for Use
Oxygen Demand Valve Instructions for Use 702-0059.11 May 2017 Contents 1. Symbols... 2 2. Warnings and Cautions... 2 2.1. Warnings!... 2 2.2. Cautions!... 3 2.3. Notes... 4 3. Functional Description...
More informationOxygen Dialflow Meter. Instructions for Use
Oxygen Dialflow Meter Instructions for Use 702-0031.9 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationPressure Regulator. Instructions for Use
Pressure Regulator Instructions for Use 702-0083.9 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient, the
More informationMicro Dial-Flowmeter. Instructions for Use
Micro Dial-Flowmeter Instructions for Use 702-0082.8 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationDialflow Regulator. Instructions for Use
Dialflow Regulator Instructions for Use 702-0030.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationOxygen Dialflow Meter. Instructions for Use
Oxygen Dialflow Meter Instructions for Use 702-0031.12 December 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in injury to the patient,
More informationPressure Regulators. Instructions for Use
Pressure Regulators Instructions for Use 702-0032.11 May 2014 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the user
More informationFiresafe Flowmeter. Instructions for Use
Firesafe Flowmeter Instructions for Use 702-0049.16 May 2014 Contents 1. Symbols... 2 2. Warnings and Cautions... 2 2.1. Warnings!... 2 2.2. Cautions!... 3 3. Functional Description... 4 3.1. Intended
More informationDomiciliary Oxygen Switch. Instructions for Use
Domiciliary Oxygen Switch Instructions for Use 702-0041.11 July 2017 1. Symbols Warning! Caution! Indicates a potentially hazardous situation which, if not avoided, could result in personal injury to the
More informationEquinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM (01EQ7000E) DEMAND VALVE (01EQ7001E) O-TWO MEDICAL TECHNOLOGIES INC.
MANUFACTURERS AND DISTRIBUTORS OF: O-Two eseries Electronic Transport Ventilators O-Two Single Use Open CPAP Systems O-Two CAREvent Range of Automatic Transport Ventilators O-Two CAREvent Range of Automatic
More informationService Manual. Regulators for Medical Gases
Service Manual Regulators for Medical Gases Copyright No part of this manual may be copied for any reason without the written permission of BPR Medical Ltd. This product is manufactured by BPR Medical
More informationO-Two Self-Study Guide. Equinox Relieve
O-Two Self-Study Guide Equinox Relieve PRODUCT DESCRIPTION/PRINCIPLES OF OPERATION The Equinox Relieve provides spontaneously breathing patients with a preset mixture of oxygen and nitrous oxide for analgesia
More informationEquinox II USER MANUAL NITROUS OXIDE/ OXYGEN 50%/50% USER ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E TABLE OF CONTENTS
TM controlled ventilation Equinox II NITROUS OXIDE/ OXYGEN 50%/50% ADMINISTRATION SYSTEM 01EQ7000E DEMAND VALVE 01EQ7001E USER TABLE OF CONTENTS CHAPTER TITLE PAGE 1 INTRODUCTION 1.1 Equinox II Administration
More informationUser manual. ZX trolley.
User manual. ZX trolley. Date July 2011 02 User manual. ZX trolley. ZX trolley. Safe transportation for medical gas cylinders. Important Please read this manual carefully before attempting to use the BOC
More informationAUTOVENT 4000 VENTILATOR
OVERVIEW AUTOVENT 4000 Only properly trained and approved Escambia County Bureau of Public Safety Paramedics are to use the AutoVent 4000 ventilator manufactured by LSP to transport patients already on
More informationService Manual. Dial Flow Regulators
Service Manual Dial Flow Regulators Copyright No part of this manual may be copied for any reason without the written permission of BPR Medical Ltd. This product is manufactured by BPR Medical Ltd on behalf
More information45A, Armthorpe Road, Brampton, Ontario, Canada, L6T 5M4. Made in Canada by O-Two Medical Technologies Inc.
O-Two eseries Electronic Transport Ventilators O-Two Single Use Open CPAP Systems O-Two CAREvent Range of Automatic Transport Ventilators O-Two CAREvent Range of Automatic Resuscitators O-Two SMART BAG
More informationBOC: Living healthcare. Manual. LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare
BOC: Living healthcare Manual LIV IQ BOC Integrated Valve with digital display portable delivery system for Medical Oxygen. BOC: Living healthcare 02 Manual LIV IQ Oxygen Manual LIV IQ Oxygen 03 Contents.
More informationCLEMENTS. ANA Oxygen Regulator with Yoke Connector. User Manual. Manual No. ANA Issue 1
ANA 74300 Oxygen Regulator with Yoke Connector User Manual Manual No. ANA 74300 001 Issue 1 Safety Thank you for purchasing this Clements Oxygen Regulator. For your safety it is imperative that this unit
More informationRev. B. Operating Instructions
780745-00 Rev. B Operating Instructions 780745-00 Rev. B TABLE OF CONTENTS PAGE Warnings, Cautions, and Notes..............................4-7 Definition of Symbols......................................
More informationMirage Vista Nasal Mask
English 60 50 40 30 20 10 0 4 6 10 12 14 16 1 20 ENGLISH Mirage Vista Nasal Mask Rx Only The Mirage Vista Nasal Mask is an accessory to a non-continuous ventilator (respirator) intended for multipatient
More informationHydraulic Punch Drivers
SERVICE MANUAL 7804SB / 7806SB Quick Draw 7704SB / 7706SB Quick Draw Flex Quick Draw Hydraulic Punch Drivers Serial Codes AHJ and YZ Read and understand all of the instructions and safety information in
More informationMedical oxygen. LIV IQ Digital therapy device.
Medical oxygen. LIV IQ Digital therapy device. BOC: Living healthcare 02 Cylinder instructions Cylinder instructions 03 Therapy device features. LIV IQ giving you safe guidance during patient treatment.
More informationCPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL
CPX EMERGENCY STANDBY MANIFOLD INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855
More informationSERIES-O REGULATOR. Instructions for use
SERIES-O REGULATOR Instructions for use 522460-05-708 Contents Page 1. INTRODUCTION 2 2. INTENDED USES 3 3. GENERAL DESCRIPTION 4 4. PARTS IDENTIFICATION 5 5. SAFETY PRECAUTIONS 6 6. EQUIPMENT USE 8 7.
More informationU S E R M A N U A L CAUTION. SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician.
U S E R M A N U A L 1600 SERIES OXYGEN REGULATOR 168715G (Shown) SAVE THESE INSTRUCTIONS Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive Tel: (+001) 610-262-6090
More informationAssembling & Installing your Pool Cleaner
Assembling & Installing your Pool Cleaner STEP 1: Check Contents of the Box Remove the Badu Clean Supreme body and all parts from the box and check that the following components are included. Refer to
More informationMMR Air Mask With. with Quick-Connect Hose
MMR Air Mask With with Quick-Connect Hose Upgrade Kits P/N 10025120 Slide to Connect P/N 10050038 Slide to Connect w/ Solid Cover P/N 10038666 Push To Connect P/N 10050037 Push To Connect w/ Solid Cover
More informationInstallation Operating Instructions for Simple Duplex Manual Manifolds PX-TSD Series
Introduction Powerex manifolds are cleaned, tested and prepared for the indicated gas service and are built in accordance with the Compressed Gas Association guidelines. The manifold consists of a regulator
More informationTechnical Specification
Medical Gas Terminal Units (Gas Outlet Points) Technical Specification EN 9170-1 Terminal Unit Product Description CPX Terminal Units comprise a first fix, second fix and check valve. The first fix assembly
More informationOperation and Maintenance of the EPV200 Portable Ventilator
Operation and Maintenance of the EPV200 Portable Ventilator 1 Applications of the EPV200 The EPV200 Portable Ventilator is a gas powered electronically controlled mechanical ventilator, designed to provide
More informationTHE HF-300 SERIES. Operating and Service Manual. Series includes all variants of HF-300/301
THE HF-300 SERIES Operating and Service Manual Series includes all variants of HF-300/301 Issue A July 2015 1 TABLE OF CONTENTS 1. Description... 3 2. Installation... 3 3. Operation... 4 3.1. Spring Loaded...
More informationAREA VALVE SERVICE UNITS
AREA VALVE SERVICE UNITS. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com U.K., SK3 0SB www.p3-phoenix.com
More informationBRITISH COMPRESSED GASES ASSOCIATION Registered office: 4a Mallard Way, Pride Park, Derby, UK. DE24 8GX Company Number: 71798, England
BRITISH COMPRESSED GASES ASSOCIATION Registered office: 4a Mallard Way, Pride Park, Derby, UK. DE24 8GX Company Number: 71798, England www.bcga.co.uk TECHNICAL INFORMATION SHEET 36-2017 MEDICAL GASES.
More informationCPR MEDICAL DEVICES INC. OPERATING MANUAL FOR RESUSCITATOR
CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR FR-300 RESUSCITATOR Symbols used in the Operating Manual for the Oxylator FR-300 and/or on the Oxylator FR-300 unit: WARNING! Do not use grease or
More informationTactAir PCP Fill Kit/Adapter. Operation Manual
TactAir PCP Fill Kit/Adapter Operation Manual INTRODUCTION Thank you for choosing a TactAir PCP fill accessory! Always exercise caution when operating this item. It will allow you to fill your PCP airguns
More informationAUTOMATIC GAS MANIFOLDS
AUTOMATIC GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com
More informationMEDICAL REGULATOR RANGE (MEDIREG RANGE)
MEDICAL REGULATOR RANGE (MEDIREG RANGE) Medireg Dialreg Flowreg INSPECTION Remove the Medireg from the packaging and inspect for damage. If there is any damage, DO NOT USE, and contact Therapy Equipment
More informationREL-510H WARNING NOTICE 12 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD
OPERATORS ORS GUIDE REL-510H 12 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD Compatible with U style and RELIABLE R12 shell type 12 ton compression dies. RELIABLE EQUIPMENT & SERVICE CO., INC. 92 Steamwhistle
More informationMANUAL GAS MANIFOLDS
MANUAL GAS MANIFOLDS Phoenix Pipeline Products Limited. Unit 8, McKenzie Industrial Park, Tel No.: 44 (0) 161 428 7200 Bird Hall Lane, Fax No.: 44 (0) 161 428 7010 Stockport, Email: info@p3-phoenix.com
More informationInstallation and Maintenance Manual. Compact Medical Gas Outlets
Installation and Maintenance Manual Compact Medical Gas Outlets Contents Product Description 3 Cleaning and Lubricating 4 Inspection and Testing 4 Installation and Dimensions 5-6 Compact Outlet Indexing
More informationREL-46 WARNING NOTICE 15 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD. Compatible with RELIABLE R15 and P Style dies. REL-46 Manual
OPERATORS ORS GUIDE REL-46 15 TON SINGLE ACTING REMOTE HYDRAULIC CRIMPING HEAD Compatible with RELIABLE R15 and P Style dies. RELIABLE EQUIPMENT & SERVICE CO., INC. 301 Ivyland Road Warminster, PA 18974
More informationEquinox II NITROUS OXIDE / OXYGEN ADMINISTRATION EQUIPMENT 01EQ7000E
MANUFACTURERS AND DISTRIBUTORS OF: CPR Bio-Barrier Face Shields Universal Pocket Ventilators Rescue Breather CPR Protective Barriers Demand Valve Resuscitators CAREvent Range of Automatic Transport Ventilators
More informationMST21 Stainless Steel Balanced Pressure Thermostatic Steam Trap
1250650/6 IM-P125-07 ST Issue 6 MST21 Stainless Steel Balanced Pressure Thermostatic Steam Trap Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation
More informationCOMPACT TRIPLE WALL OUTLET STATION MODEL INSTALLATION AND OPERATING INSTRUCTIONS
COMPACT TRIPLE WALL OUTLET STATION MODEL 6255-1 INSTALLATION AND OPERATING INSTRUCTIONS To assure safe operation and conformation to local fire codes, all Porter Outlet Stations are designed to be used
More informationTable of Contents. Operating Instructions. Resource v.2 Conserving Regulator
Operating Instructions Table of Contents Resource v.2 Conserving Regulator Safety Information Device Precautions Introduction Product Features Product Specifications Feature Illustrations Set Up Usage
More informationRejuvenation Instructions
Rejuvenation Instructions #401 Air Systems UPR This NRI covers the following: Understanding the applications and operation of flow meters. Understand the application and operation of test pressure gauges.
More information5 Gallon Pressure Pot with HVLP Spray Gun and Hose
California Air Tools 5 Gallon Pressure Pot with HVLP Spray Gun and Hose Model No. 365 Technical Data Type of feed.pressure Maximum pressure in the tank... 0,413Mpa (60PSI) Working pressure in the tank.0,
More informationLIQUIP DRYBREAK COUPLER. API800 Series MAINTENANCE INSTRUCTIONS
LIQUIP DRYBREAK COUPLER API800 Series MAINTENANCE INSTRUCTIONS API LOADING COUPLER TO API RP1004 June 2015 Issue: F M:\Product-Info\API8xx\6-Service-Maintenance\API800 MAINTENANCE INSTRUCTIONS 40183.doc
More informationMaintenance and Repair. AirHawk II Air Mask. Second Stage Regulator. Order No.: /02 Prnt. Spec (I) MSAsafety.
Maintenance and Repair Second Stage Regulator Order No.: 10104241/02 Prnt. Spec. 10000005389(I) MSAsafety.com WARNING! Read this manual carefully before servicing the device. The device will perform as
More informationOperation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators
Operation and Maintenance Manual for GENTEC Model 881VR Continuous/Intermittent Digital Suction Regulators Genstar Technologies Co., Inc. 4525 Edison Avenue Chino, CA 91710 USA TEL 909-0-272 FAX 909-0-485
More informationService and Repair Manual
II stage R2 Ice/ Special, II stage R 1 Pro DOWNSTREAM 2 nd STAGE REGULATOR Service and Repair Manual Introduction Safety Precautions...4 General Procedures, Maintenance Schedules...5 Initial Inspection
More informationHelium-Oxygen Blender
Helium-Oxygen Blender Service Manual Model No. PM5400 Series PM5500 Series (shown) SAVE THESE INSTRUCTIONS 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA 18067 USA Fax: (+001) 610-262-6080 www.precisionmedical.com
More informationIX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions
0575050/2 IM-P057-04 CH Issue 2 IX1 Spirax-Monnier International Compressed Air Filter Installation and Maintenance Instructions 1 Safety information 2 General product information 3 Installation and commissioning
More informationSOTR Special Operations Tactical Respirator
SOTR Special Operations Tactical Respirator OPERATOR S MANUAL FOR INDIVIDUAL RESPIRATORY PROTECTION OPS-CORE 2018 OMM G055-1000 REV. B INTRODUCTION ABOUT YOUR SOTR The Ops-Core Special Operations Tactical
More informationMSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection
1170850/1 IM-P117-36 ST Issue 1 MSC-P and MSC-N Manifolds for Steam Distribution and Condensate Collection Installation and Maintenance Instructions 1. Safety information 2. General product information
More informationCPR MEDICAL DEVICES INC. OPERATING MANUAL
CPR MEDICAL DEVICES INC. OPERATING MANUAL FOR OXYLATOR EMX RESUSCITATION AND INHALATION SYSTEM Symbols used in the Operating Manual for the OXYLATOR EMX and/or on the OXYLATOR EMX unit: WARNING! Do not
More informationRD(H)20/25 Pressure-Reducing Regulator User Manual
RD(H)20/25 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be
More informationLUBRICATOR ASSEMBLY AND OPERATING INSTRUCTIONS
AIR FILTER, REGULATOR AND LUBRICATOR 4035 ASSEMBLY AND OPERATING INSTRUCTIONS 349 Mission Oaks Blvd., Camarillo, CA 930 Visit our Web site at http://www.harborfreight.com Copyright 004 by Harbor Freight
More informationACRYLIC FLOW METERS 0-30, 0-70 & LPM
ACRYLIC FLOW METERS 0-30, 0-70 & 0-120 LPM ASSEMBLY INSTRUCTIONS & INSTRUCTIONS FOR USE R219P86 (0-120LPM) R219P87 (0-70LPM) R219P88 (0-30 LPM) R138P11 Rev. D TABLE OF CONTENTS: 1.0 PRODUCT OVERVIEW...1
More informationPeninsula Dental Social Enterprise (PDSE)
Peninsula Dental Social Enterprise (PDSE) Medical Gas Policy Version 3.0 Date approved: August 2018 Approved by: The Board Review due: August 2021 Policy will be updated as required in response to a change
More informationBefore using or working with the Grab n Go Heliox system, please read and understand the usage and safety instructions provided in this booklet.
Operating Instructions for the Before using or working with the Grab n Go Heliox system, please read and understand the usage and safety instructions provided in this booklet. Introduction The Praxair
More informationEquinox Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System USER MANUAL
Equinox Advantage Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System USER MANUAL Made in Canada by O-Two Medical Technologies Inc. Part Number: 15PL2183 Rev.7 Jan 18. 2017 TABLE OF CONTENTS
More informationAIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064
AIR HAMMERS MODEL NO: CAT138/CAT139 PART NO: 3120152 /3120153 OPERATING & MAINTENANCE INSTRUCTIONS GC064 INTRODUCTION Thank you for purchasing this CLARKE Air Hammer. Before attempting to use this product,
More informationInstructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max
Instructions for use Evacuation system compressed air injector SL 20 / SL 20 Max SL 30 / SL 30 Max Table of Contents Introduction 1. Description of Functions 2. Putting into Operation 3. Instructions for
More informationATD LB PRESSURE BLASTER INSTRUCTION MANUAL
ATD-8402 90LB PRESSURE BLASTER INSTRUCTION MANUAL SAVE THESE INSTRUCTIONS SAFETY INSTRUCTIONS FOR SANDBLASTER 1. Before opening the tank release the air pressure on the sand tank. To do this, turn off
More informationO-Two Equinox Relieve Nitrous Oxide/Oxygen Analgesic Gas Mixing and Delivery System
MANUFACTURERS AND DISTRIBUTORS OF: O-Two eseries Electronic Transport Ventilators O-Two Single Use Open CPAP Systems O-Two CAREvent Range of Automatic Transport Ventilators O-Two CAREvent Range of Automatic
More informationInstallation, Operation, and Maintenance Manual. Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model CP-30SI
Installation, Operation, and Maintenance Manual Welker Constant Pressure Cylinder With Solid Indicator (Non-mixer) Model The information in this manual has been carefully checked for accuracy and is intended
More informationTraining. tm PA90 Series - Lung Demand Valve
D Training Dräger tm12842 PA90 Series - Lung Demand Valve Index Section 1 Instructions for Use Section 2 Introduction and Safety Warnings Section 3 Product Description Section 4 Operating Principle Section
More informationManual Actuated Boiler Blowdown Valves
Manual Actuated Boiler Blowdown Valves Installation and Maintenance Instructions 1. Safety information 2. General product information 3. Installation 4. Operation 5. Maintenance 6. Spare parts p.1 1. Safety
More informationC-1 C-7. C-6 Short air tubing Upper Velcro straps. C-a C-5 C-2 C-4 C-3 B-1 B-9 B-10 B-2 B-7 B-8 B-3 B-6 D-1 D-2 D-13 D-3 D-12 D-10 D-11 D-4 D-5 D-8
A Forehead support Forehead pads C Mask Frame Mask Elbow Cushion clip Mask Cushion C-7 C-1 Vent cover Ports cap Cuff ActiveCell C-6 Short air tubing Upper Velcro straps Swivel Lower Velcro straps Headgear
More information2 PRE USE FUNCTION CHECK 3 OPERATING PROCEDURE 4 SERVICING. 5 CAREvent ALS ACCESSORIES 11
THE RESUSCITATOR MUST BE THOROUGHLY CLEANED AFTER EACH PATIENT USE. 1. Operate CAREvent ALS Handheld Resuscitator to blow out any contaminant from the patient valve. 2. Ensure CAREvent ALS Handheld Resuscitator
More informationCompact Triple Cabinet Outlet Station Model B Installation and Operating Instructions
PORTER Parker Hannifin Corporation Porter Instrument Division 245 Township Line Rd. P.O. Box 907 Hatfield, PA 19440-0907 USA (215) 723-4000 / fax (215) 723-5106 Compact Triple Cabinet Outlet Station Model
More informationSAFETY MEASURES. These important guidelines should always be followed to work safely with the PRIMATECH pneumatic stapler model Q180:
PRIMATECH PNEUMATIC 18GA STAPLER Q180 The pneumatic flooring stapler Q180 is a professional precision tool specially developed for the installation of most of 3/8" to 5/8" solid and engineered hardwood
More informationTraining. Testor Training Manual
Training Testor Training Manual Index Section 1 Introduction and Safety Warnings Section 2 Test Procedures Section 3 Test Hoses Section 4 Fault Location 1:1 1.1 Introduction The Dräger Testor test equipment
More informationCompact Triple Cabinet Outlet Station Model Installation and Operating Instructions
Compact Triple Cabinet Outlet Station Model 6258-1 Installation and Operating Instructions The Porter Compact Triple Outlet Station (6258-1) provides a quick, safe, and reliable method of connection to
More informationAuto-Rewind Hose Reels INSTRUCTION MANUAL FOR OXY-LPG MODEL
Auto-Rewind Hose Reels INSTRUCTION MANUAL FOR OXY-LPG MODEL Introduction Thank you for purchasing a Retracta Auto Rewind Hose Reel. The Retracta range of hose reels are a breakthrough in industrial quality
More informationFILTER REGULATORS MODEL NO: CAT155 & CAT156 FITTING & MAINTENANCE INSTRUCTIONS PART NO: & ORIGINAL INSTRUCTIONS
FILTER REGULATORS MODEL NO: CAT155 & CAT156 PART NO: 3120169 & 3120170 FITTING & MAINTENANCE INSTRUCTIONS ORIGINAL INSTRUCTIONS GC0117 INTRODUCTION Thank you for purchasing this CLARKE Filter/Regulator.
More informationHANDBOOK. Squeeze Off Unit SOU 400. Please refer any queries to:
HANDBOOK Squeeze Off Unit SOU 400 Please refer any queries to: Hy Ram Engineering Co Ltd Pelham Street Mansfield Nottinghamshire NG18 2EY Telephone No: (01623) 422982 Please note all queries should state
More informationstormrider Instructions for Use Version 08/2017
stormrider EN Instructions for Use Version 08/2017 INTRODUCTION DESCRIPTION SAFETY INSTRUCTIONS TECHNICAL DATA HANDLING & OPERATION CARE & MAINTENANCE TROUBLESHOOTING DESCRIPTION Front Sight Barrel Rear
More informationPRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL
PRESSURE REDUCING STATION INSTALLATION, OPERATIONS & MAINTENANCE MANUAL Company Registered Office: Crown House, Stockport, Cheshire, SK13RB No. 850 0700 67 Registered in England and Wales No. 05058855
More informationPROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road
PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves
More informationInstallation, Operation and Maintenance Manual for Back Pressure Regulator
Installation, Operation and Maintenance Manual for Back Pressure Regulator Model 8860 2009 Groth Corporation IOM-8860 Rev. B 12541 Ref. ID: 95565 Page 2 of 13 Table of Contents I. INTRODUCTION 3 II. DESIGN
More informationFOR SAFE AND CORRECT USE
FOREWARD Thank you for purchasing the Model AP-20 sampling pump. The Model AP-20 pump is designed specifically for use with Kitagawa detector tubes. This system with the available Kitagawa detector tubes
More informationInstructions for use Compressed air injector with water trap
Instructions for use Compressed air injector with water trap Table of Contents Introduction 1. Description of Functions 2. Putting into Operation 3. Instructions for Cleaning and Care 3.1 Disassembling
More informationRARS5000 AIR BODY SAW OWNER S OPERATING MANUAL
RARS5000 AIR BODY SAW OWNER S OPERATING MANUAL DESCRIPTION 1. No mar 2. No mar tip 3. Housing grip 4. Trigger 5. Air inlet 6. Air inlet plug 7. Plastic board Important! It is essential that you read the
More informationISO 5359 INTERNATIONAL STANDARD. Low-pressure hose assemblies for use with medical gases
INTERNATIONAL STANDARD ISO 5359 Third edition 2008-06-15 Low-pressure hose assemblies for use with medical gases Flexibles de raccordement à basse pression pour utilisation avec les gaz médicaux Reference
More informationHallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting
Hallowell EMC 2000 and 2002 Veterinary Ventilator Set-up, Use, and Troubleshooting This is a quick reference for setting up, using, and troubleshooting the Hallowell EMC 2000 and 2002 Small Animal Ventilator.
More informationRG1200 Service and Repair Manual
Dive Rite RG 1200 Regulator Service and Repair Manual Page 1 Text and Photography by Pete Nawrocky Copyright ( ) 1999-2000, Lamartek, Inc., dba Dive Rite RG1200 Service and Repair Manual First Stage.........................................
More informationHANDBOOK. Squeeze Off Unit SOU 250. Please refer any queries to:
HANDBOOK Squeeze Off Unit SOU 250 Please refer any queries to: Hy Ram Engineering Co Ltd Pelham Street Mansfield Nottinghamshire NG18 2EY Telephone No: (01623) 422982 Please note all queries should state
More informationEASTERN ENERGY SERVICES PTE LTD. 60 Kaki Bukit Place #02-19 Eunos Tech Park Singapore, SG Singapore Telephone: Fax:
2 Table Of Contents 1. Introduction 3 2. About this Manual 3 3. Contacting YZ Systems 3 4. Vessel Components 4 5. Specifications 5 6. Application 6 7. Theory of Operation 7 8. DuraSite Installation & Use
More information500SE SECOND STAGE SERVICE PROCEDURE
500SE SECOND STAGE SERVICE PROCEDURE This 500SE Service Procedure conveys a list of components and service procedures that reflect the 500SE as it was configured at the time of this writing. CONTENTS TROUBLESHOOTING...
More informationNitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions
FM-171 Rev. B 8/00 Nitrous Oxide / Oxygen FLOOR MOUNTED OUTLET STATION Installation and Instructions MODEL 6300-1 PORTER INSTRUMENT COMPANY, INC. 245 TOWNSHIP LINE RD. P.O. BOX 907 HATFIELD, PA 19440-0907
More information1200B2 Series Service Regulators. Instruction Manual
00B Series Service Regulators Instruction Manual 00B Series Service Regulators 0 Elster American Meter 00B Series Service Regulators General Information The 00B Series Service Regulators are available
More informationPRS(TC)4,8 USER MANUAL. Read the complete manual before installing and using the regulator.
PRS(TC)4,8 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More information35 TON HYDRAULIC PUNCH WARNING
OPERATORS GUIDE REL-35T-PNC 35 TON HYDRAULIC PUNCH NOTICE Sizes, weights and tool specifications listed in this manual are subject to change without notice. Please consult factory for information and updates.
More informationFlow meter. bellow vaporizer. APL valve. Scavenging system
Introductory Lecture Series: The Anesthesia Machine PORNSIRI WANNADILOK Objectives Anesthesia Machine Ventilators Scavenging Systems System Checkout 1 Flow meter ventilator bellow vaporizer Corrugated
More informationAir Mask Low/High Pressure
Air Mask Low/High Pressure USERS MAINTENANCE INSTRUCTIONS THIS MANUAL MUST BE CAREFULLY READ AND FOLLOWED BY ALL PERSONS WHO HAVE OR WILL HAVE THE RESPONSIBILITY FOR USING OR SERVICING THIS AIR MASK. This
More informationRS(H)10,15 USER MANUAL. Read the complete manual before installing and using the regulator.
RS(H)10,15 USER MANUAL Read the complete manual before installing and using the regulator. WARNING INCORRECT OR IMPROPER USE OF THIS PRODUCT CAN CAUSE SERIOUS PERSONAL INJURY AND PROPERTY DAMAGE. Due to
More information