Oil Content Analyzer OCMA-500 Instruction manual CODE:GZ

Size: px
Start display at page:

Download "Oil Content Analyzer OCMA-500 Instruction manual CODE:GZ"

Transcription

1 Oil Content Analyzer OCMA-500 Instruction manual CODE:GZ

2 Preface This manual describes the operation of the Oil Content Analyzer, OCMA-500. Be sure to read this manual before using the product to ensure proper and safe operation of the instrument. Also safely store the manual so it is readily available whenever necessary. Product specifications and appearance, as well as the contents of this manual are subject to change without notice. Warranty and responsibility Trademarks HORIBA, Ltd. warrants that the Product shall be free from defects in material and workmanship and agrees to repair or replace free of charge, at option of HORIBA, Ltd., any malfunctioned or damaged Product attributable to responsibility of HORIBA, Ltd. for a period of one (1) year from the delivery unless otherwise agreed with a written agreement. In any one of the following cases, none of the warranties set forth herein shall be extended; Any malfunction or damage attributable to improper operation Any malfunction attributable to repair or modification by any person not authorized by HORIBA, Ltd. Any malfunction or damage attributable to the use in an environment not specified in this manual Any malfunction or damage attributable to violation of the instructions in this manual or operations in the manner not specified in this manual Any malfunction or damage attributable to any cause or causes beyond the reasonable control of HORIBA, Ltd. such as natural disasters Any deterioration in appearance attributable to corrosion, rust, and so on Replacement of consumables HORIBA, LTD. SHALL NOT BE LIABLE FOR ANY DAMAGES RESULTING FROM ANY MALFUNCTIONS OF THE PRODUCT, ANY ERASURE OF DATA, OR ANY OTHER USES OF THE PRODUCT. Company names and brand names are either registered trademarks or trademarks of the respective companies. (R), (TM) symbols may be omitted in this manual. Proprietary rights The font used for the displays on the LCD screen of the product is a Gothic bitmap font, which is designed and created by NEC Corporation. Consequently, any rights relating to this font shall belong to NEC Corporation. February, HORIBA, Ltd.

3 Regulations Conformable Directive This equipment conforms to the following directives and standards: Directives: Standards: The EMC Directive 2004/108/EC The Low Voltage Directive 2006/95/EC [the EMC Directive] EN : 2006 Class B, Basic requirements [the Low Voltage Directive] EN : 2010(Ed.3.0) Installation environment This product is designed for the following environment. Overvoltage Category II Pollution degree 2 Information on disposal of electrical and electronic equipment and disposal of batteries and accumulators The crossed out wheeled bin symbol with underbar shown on the product or accompanying documents indicates the product requires appropriate treatment, collection and recycle for waste electrical and electronic equipment (WEEE) under the Directive 2012/19/EU, and/or waste batteries and accumulators under the Directive 2006/66/EC in the European Union. The symbol might be put with one of the chemical symbols below. In this case, it satisfies the requirements of the Directive 2006/66/EC for the object chemical. This product should not be disposed of as unsorted household waste. Your correct disposal of WEEE, waste batteries and accumulators will contribute to reducing wasteful consumption of natural resources, and protecting human health and the environment from potential negative effects caused by hazardous substance in products. Contact your supplier for information on applicable disposal methods.

4 FCC Rules Any changes or modifications not expressly approved by the party responsible for compliance shall void the user's authority to operate the equipment. Note This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provided reasonable protection against harmful interference in a residential installation. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: Reorient or relocate the receiving antenna. Increase the separation between the equipment and receiver. Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. Consult the dealer or an experienced radio/tv technician for help. Korea certification

5 For Your Safety Hazard classification and warning symbols Warning messages are described in the following manner. Read the messages and follow the instructions carefully. Hazard classification This indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. This is to be limited to the most extreme situations. This indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. This indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury. It may also be used to alert against unsafe practices. Without safety alert indication of hazardous situation which, if not avoided, could result in property damage. Warning symbols Description of what should be done, or what should be followed Description of what should never be done, or what is prohibited

6 Safety precautions This section provides precautions for using the product safely and correctly and to prevent injury and damage. The terms of DANGER, WARNING, and CAUTION indicate the degree of imminency and hazardous situation. Read the precautions carefully as it contains important safety messages. WARNING Electric shock To prevent electric shock, ground the product. Do not ground the product to dangerous places such as a gas pipe. Samples may be dangerous substances. Fully understand the properties of the samples to be measured, and handle them appropriately. Fire For your safety, make sure to unplug the power plug from the electrical outlet when not in use. Clear dust on the power plug periodically (a few times a year). If the power supply cord is left plugged in the electrical outlet for a long period of time, electrical tracking may occur due to dust and moisture and it may ignite and result in fire. Fire or electric shock Do not bundle the power supply cord during use. Do not damage, bend, or stretch the power supply cord forcibly. If it can not be plugged into an electrical outlet firmly, stop use of the power supply cord. If may result in overheating, fire, or an electrical shock. Be sure not to disassemble or modify the product, except as instructed in this manual. It may cause electric shock or product failure. CAUTION Chemical hazard (solvent S-316) Inhalation or accidental ingestion of a large amount of solvent S-316 may be harmful. Observe the following rules when handling: Ventilate the work area sufficiently. Wear a protective mask and protective gloves. Wash hands well after handling the solvent. Chemical hazard (hydrochloric acid) Hydrochloric acid is toxic by skin or eye exposure. If it touches the skin, immediately rinse with water. If it reaches the eyes, rinse immediately under a large amount of running water and get medical attention. When dispensing liquids into the extraction tank, limit the solvent amount to 10 ml or less and the amount of the measurement liquid to 20 ml or less, so that the total volume of liquid does not exceed the upper limit line (30 ml). If the liquid exceeds the upper limit line, the liquid may leak and cause short-circuit in the internal wiring of the product. If liquid containing hydrochloric acid leaks and comes in contact with the skin, irritation and burning may occur. Discard the liquid collected in the beakers before overflow. If liquid containing hydrochloric acid overflows and comes in contact with the skin, irritation and burning may occur. Take care not to pinch your fingers when opening or closing the right cover. During closing the right cover, do not release your hand until you hear a click sound.

7 CAUTION Avoid any impact on the product. If the product is damaged and liquid leaks, the internal wiring may short-circuit. If liquid containing hydrochloric acid leaks and comes in contact with the skin, irritation and burning may occur.

8 Product Handling Information Operational precautions Using this product in a non-specified manner may impair performance and prevent the protective functions from operating. Observe the cautions below. This product is specified for use with S-316. Do not use any other solvent than S-316 to perform extraction and measurement. It may cause product failure. Samples containing emulsifying substances (surface-active substances) cannot be measured. Samples containing acetone or toluene cannot be measured. These samples may damage the product. Samples containing impurities and samples with high viscosity should be filtered, diluted, or otherwise preprocessed appropriately before measurement. If measured without preprocessing, the tubing and valves may become clogged and product may get damaged. Take care not to spill samples or solvents on the main unit. It may cause product failure. Avoid operating and storing the product under the following locations and conditions: - Humidity above 80%. - Temperature less than 0 C or over 40 C - Locations subject to sudden temperature changes - Direct exposure to sunlight - Presence of corrosive gases - Dusty locations - Poor ventilation - Locations subject to vibration - Close proximity to large electric motors or voltage transformers When handling liquids during measurement, calibration, or otherwise, remove the USB memory stick from the USB memory port and cap the port. If a liquid spills on a USB memory stick or the USB memory port, the liquid may enter the interior of the product from the USB memory port and cause product damage. Do not overturn the main unit. It may cause liquid to leak from the unit inside. Do not press the keys or the screen with a sharp or hard object. Do not block the fan vent on the back of the main unit. Before performing maintenance or inspection, read and understand the chapter " Maintenance " (page 85 ) in this manual. Wipe with water when cleaning the exterior of the product, never use the organic solvent. Make sure that the power supply voltage is correct for the product before switching the power ON. When the product will not be used for an extended period of time, remove the plug from the power outlet. Do not use the provided power cable for other than this product.

9 Solvent handling precautions It is recommended that new solvent from the same production lot is used for calibration liquid preparation, zero calibration, span calibration, and measurement. Solvents from different production lots may have different mix ratios. If it is necessary to use solvents from different production lot or reprocessed solvent, mix all the volumes to be used in a glass container, in order to equalize all mix ratios. Disposal of the product When disposing of the product, follow the related laws and/or regulations of your country.

10 Manual Information Description in this manual This interprets the necessary points for correct operation and notifies the important points for handling the product. This indicates the part where to refer for information. This indicates reference information.

11 Contents Product Outline... 1 Overview... 1 Accessories... 2 Part names... 3 Exterior... 3 Extraction tank window... 4 Right inside... 4 Operation buttons... 5 LCD... 6 Main Unit Setup 9 Putting the valve stickers... 9 Placing the absorbent sheet... 9 Basic Operation Power ON Power OFF Drainage mode operation Operations while the sequence is in progress Pause When an error occurs Connecting a USB memory stick Screen operations in the manual measurement/calibration mode Using pop-up screens Selection list display Numeric keys Character keys Preparation Measurement preparation cautions Glassware Items required Placing the beakers for drainage Cleaning the measuring cylinders or syringes... 19

12 Zero liquid for calibration Span liquid for calibration Using B-heavy oil Hydrochloric acid Hydrochloric acid preparation method Condition settings Calibration condition settings Measurement condition settings Calibration Calibration cautions Points to check prior to calibration Points to check prior to zero calibration Points to check prior to span calibration Items required Zero calibration Span calibration Automatic zero calibration Automatic span calibration Manual zero calibration Manual span calibration Measurement Measurement cautions Preprocessing Removing fine particles Performing preliminary extraction outside the product Points to check prior to measurement Items required Automatic measurement Deciding the extraction time Flow scheme of operation Procedure Semi-automatic measurement... 40

13 Flow scheme of operation Procedure Manual measurement Scheme of operation Procedure Examples of measurement by extraction outside the product Oil content in water Oil content in or on solids Data Management Data Top screen Current Alarm screen Measurement History screen Calibration History screen USB Memory screen Execution confirmation for [Save Measurement History] Execution confirmation for [Save Calibration History] Execution confirmation for [Save Settings] Memory Clear screen Execution confirmation for [Clear Measurement History] Execution confirmation for [Clear Calibration History] Execution confirmation for [Initialize Setting] Setting Setting Top screen Measurement Setting screen Extraction Time Separation Time Fill Cell Time Meas. Limit Drainage Time Number of Purge Meas. Mode Measurement Unit Solvent Vol Sample Vol Zero Shift Value Use Light Confirm Save... 76

14 Save Memo Display Negative Display Raw Data Calibration Setting screen Span Point Extraction Time Separation Time Number of Purge Calib. Mode System Setting screen Language B-Light Off Time Date Time Maintenance Maintenance item list Rinsing the flow paths Maintenance interval guideline Items required Rinsing procedure Inspecting the absorbent sheet Maintenance interval guideline Items required Work procedure Cleaning the fan filter Maintenance interval guideline Items required Work procedure Washing the fan filter Maintenance interval guideline Items required Work procedure Washing the extraction tank Maintenance interval guideline Items required Work procedure Removing liquid from the air hole of the extractor Maintenance interval guideline Items required Work procedure... 92

15 Replacing the water filter Maintenance interval guideline Items required Replacement procedure Drying the measurement cell Maintenance interval guideline Items required Drying procedure Troubleshooting Alarm displays and actions List of alarms Problems not indicated by an alarm Problems related to product operation Problems related to measured values Reference About this product Measurement principle Measurement time Conversion of measurement units mg/kg mg/g mg/pc Solvent S Characteristics Properties of S Reclamation of solvent Storing solvent Disposing of solvent Frequently asked questions Solvents Measurement Solvent reclamation unit SR Product Information 115 Specifications List of optional parts

16 Image Contents Fig. 1 Exterior... 3 Fig. 2 Extraction tank window... 4 Fig. 3 Right inside... 4 Fig. 4 Operation buttons... 5 Fig. 5 Measurement/calibration screen example... 6 Fig. 6 Example of item selection screen... 7 Fig. 7 Pop-up screen example (selection list)... 7 Fig. 8 Pop-up screen example (numeric keys)... 8 Fig. 9 Input screen example (character keys)... 8 Fig. 10 Positions of the valve stickers... 9 Fig. 11 Tray in the main unit... 9 Fig. 12 Power cable connection Fig. 13 Initial screen Fig. 14 Measurement Top screen Fig. 15 Drainage mode - Drainage Fig. 16 Inserting a USB memory stick Fig. 17 USB icon Fig. 18 Example of instantaneous value measurement screen (manual measurement) Fig. 19 Example of selection list pop-up screen Fig. 20 Example of numeric key pop-up screen Fig. 21 Example of character key pop-up screen Fig. 22 Placing beakers for drainage Fig. 23 Start of the automatic zero calibration mode Fig. 24 Display of automatic zero calibration result Fig. 25 Start of the automatic span calibration mode Fig. 26 Display of automatic span calibration result Fig. 27 Start of the manual zero calibration mode Fig. 28 Manual zero calibration - Instantaneous value measurement Fig. 29 Manual zero calibration - Measurement Fig. 30 Manual zero calibration - Result Fig. 31 Start of the manual span calibration mode Fig. 32 Manual span calibration - Instantaneous value measurement Fig. 33 Manual span calibration - Measurement Fig. 34 Manual span calibration - Result Fig. 35 Removal of fine particles Fig. 36 Operation flow of automatic measurement Fig. 37 Start of the automatic measurement mode Fig. 38 Auto Measurement - Measurement Fig. 39 Operation flow of semi-automatic measurement Fig. 40 Start of the semi-automatic measurement mode Fig. 41 Semi-automatic measurement - Extraction Fig. 42 Semi-automatic measurement - Layer separation Fig. 43 Semi-automatic measurement - Fill Cell Fig. 44 Semi-automatic measurement - Measurement Fig. 45 Operation flow of manual measurement Fig. 46 Start of the manual measurement mode Fig. 47 Manual measurement - Instantaneous value measurement Fig. 48 Manual measurement - Measurement Fig. 49 Manual measurement - Result Fig. 50 Mixing solvent and sample water by shaking Fig. 51 Checking layer separation Fig. 52 Extraction... 50

17 Fig. 53 Separating the solvent layer Fig. 54 Visually checking the solvent layer Fig. 55 Removing water from the solvent layer Fig. 56 Oil content extraction from part (immersion) Fig. 57 Data Top screen Fig. 58 Current Alarm screen Fig. 59 Measurement History screen Fig. 60 Calibration History screen Fig. 61 USB Memory screen Fig. 62 Execution confirmation for [Save Measurement History] Fig. 63 Execution confirmation for [Save Calibration History] Fig. 64 Execution confirmation for [Save Settings] Fig. 65 Setting file example Fig. 66 Memory Clear screen Fig. 67 Execution confirmation for [Clear Measurement History] Fig. 68 Execution confirmation for [Clear Calibration History] Fig. 69 Execution confirmation for [Initialize Setting] Fig. 70 Setting Top screen Fig. 71 Measurement Setting screen Fig. 72 Extraction Time screen Fig. 73 Separation Time screen Fig. 74 Fill Cell Time screen Fig. 75 Meas. Limit screen Fig. 76 Drainage Time screen Fig. 77 Number of Purge screen Fig. 78 Meas. Mode screen Fig. 79 Measurement Unit screen Fig. 80 Solvent Vol. screen Fig. 81 Sample Vol. screen Fig. 82 Zero Shift Value screen Fig. 83 Use Light screen Fig. 84 Confirm Save screen Fig. 85 Save Memo screen Fig. 86 Display Negative screen Fig. 87 Display Raw Data screen Fig. 88 Calibration Setting screen Fig. 89 Span Point screen Fig. 90 Extraction Time screen Fig. 91 Separation Time screen Fig. 92 Number of Purge screen Fig. 93 Calib. Mode screen Fig. 94 System Setting screen Fig. 95 Language screen Fig. 96 B-Light Off Time screen Fig. 97 Date screen Fig. 98 Time screen Fig. 99 Tray Fig. 100 Removing the fan filter Fig. 101 Cleaning the fan filter Fig. 102 Removing the fan filter Fig. 103 Removing the sample inlet Fig. 104 Latch knobs Fig. 105 Extractor... 91

18 Fig. 106 Removing the sample inlet Fig. 107 Latch knobs Fig. 108 Removing liquid from the air hole Fig. 109 Opening the right cover Fig. 110 Removing the joints Fig. 111 Replacing the water filter element Fig. 112 Opening the right cover Fig. 113 Removing the joints Fig. 114 Replacing the water filter element Fig. 115 Lower hole on the filter block Fig. 116 Infrared absorption spectrums of solvent S-316 and oil

19 Table Contents Table 1 Operable buttons with the instantaneous value measurement screen.. 14 Table 2 Operable buttons with a selection list pop-up screen Table 3 Operable buttons with a numeric key pop-up screen Table 4 Operable buttons with a character key pop-up screen Table 5 Menu on the Data Top screen Table 6 Button functions with the Data Top screen Table 7 Button functions with the Current Alarm screen Table 8 Button functions with the Measurement History screen Table 9 Button functions with the Calibration History screen Table 10 Menu on the USB Memory screen Table 11 Button functions with the USB Memory screen Table 12 Button functions with execution confirmation for [Save Measurement History] Table 13 Button functions with execution confirmation for [Save Calibration History] Table 14 Button functions with execution confirmation for [Save Settings] Table 15 Menu on the Memory Clear screen Table 16 Button functions with the Memory Clear screen Table 17 Button functions with execution confirmation for [Clear Measurement History] Table 18 Button functions with execution confirmation for [Clear Calibration History] Table 19 Button functions with execution confirmation for [Initialize Setting] Table 20 Menu on the Setting Top screen Table 21 Button functions with the Setting Top screen Table 22 Items on the Measurement Setting screen Table 23 Button functions with the Measurement Setting screen Table 24 Items on the Calibration Setting screen Table 25 Button functions with the Calibration Setting screen Table 26 Items on the System Setting screen Table 27 Button functions with the System Setting screen Table 28 Maintenance items... 85

20

21 Product Outline Product Outline Overview OCMA-500 is a compact automated oil content analyzer using the solvent S-316. The solvent extracts the oil content from a sample, binds it and is being measured by the infrared detector. Simple button operations execute extraction, measurement, and drainage automatically. OCMA-500 allows a measurement using 8 ml of solvent and 16 ml of measurement liquid. It also allows a measurement using 10 ml of solvent and 20 ml of measurement liquid like the OCMA-300 series. 1

22 Product Outline Accessories The package contains the main unit and accessories indicated below. Make sure that none of the items are missing or damaged. Name Remarks Quantity Image Main unit OCMA Dropper Polyethylene, 2.5 ml 1 B-heavy oil 10 ml 1 Filter element For water filter 5 Absorbent sheet For the internal tray (refer to page 9 ) 1 Power cable - 1 Valve stickers For operation buttons (refer to page 9 ) 2 Manual This manual 1 2

23 Product Outline Part names Exterior Fig. 1 Exterior No. Name Description 1 LCD This displays measurement results and items necessary for various operations. 2 Operation buttons Buttons for performing a variety of operations 3 Extraction tank window Allows you to check conditions inside the extraction tank. 4 5 Drainage outlet from measurement cell Drainage outlet from extraction tank Liquid is drained from the measurement cell through this outlet. Liquid is drained from the extraction tank through this outlet. 6 Air vent pipe This vent allows keeping the flow path at the atmospheric pressure. 7 Fan vent A fan for internal temperature adjustment is located inside this vent. 8 Power switch Switches the power of this product ON and OFF. 9 Power cable connector Connects the provided power cable. 10 USB memory port A USB memory stick can be inserted into this port. 11 Sample inlet Dispense the measurement liquid into this inlet. Use a FAT/FAT32 formatted USB stick. Other formats are not available for this product. HORIBA-recommended USB memory sticks are available. Consult your dealer. 3

24 Product Outline Extraction tank window Fig. 2 Extraction tank window No. Name Description 1 Upper limit line 30 ml line. Indicates the maximum amount of liquid that can be dispensed. Right inside Fig. 3 Right inside No. Name Description 1 Right cover Open to perform water filter maintenance. 2 Tray Put the absorbent sheet on this tray to catch overflows from the extraction tank. 3 Water filter Separates water from the solvent. Always keep the right cover closed during measurement. Stable measurement cannot be performed when the right cover is open. 4

25 Product Outline Operation buttons Fig. 4 Operation buttons No. Name Image Description 1 DATA button 2 CAL button 3 MEAS button 4 SET button 5 Up button 6 Right button 7 Left button 8 Down button 9 ENT button 10 ESC button Press this button to open the Data Top screen (refer to " Data Top screen " (page 56 )). Press this button with the Measurement Top screen appearing to move to the automatic zero calibration mode or manual zero calibration mode, depending on the calibration mode setting (refer to the chapter " Calibration " (page 23 )). Press this button to move to the currently set measurement mode (refer to the chapter " Measurement " (page 33 )). Press this button to move to the Setting Top screen (refer to " Setting Top screen " (page 68 )). Press this button to change selections. The item above the currently selected item will be selected. In the manual mode, press this button to open and close the fill cell valve (refer to " Screen operations in the manual measurement/calibration mode " (page 14 )). Press this button to change selections. The item to the right of the currently selected item will be selected. If there is the next page, the next page will be displayed. In the manual mode, press this button to switch the stirring motor ON and OFF (refer to " Screen operations in the manual measurement/calibration mode " (page 14 )). Press this button to change selections. The item to the left of the currently selected item will be selected. If there is the previous page, the previous page will be displayed. Press this button to change selections. The item below the currently selected item will be selected. In the manual mode, press this button to open and close the drainage valve (refer to " Screen operations in the manual measurement/calibration mode " (page 14 )). Press this to enter the current selection or value, or move to the next action. In the manual mode, press this button to start measurement (refer to " Screen operations in the manual measurement/calibration mode " (page 14 )). Press to undo the last action and return to the previous process. When pressed during measurement, measurement stops or is paused. In the manual mode, press this button to return the mode start state (refer to " Screen operations in the manual measurement/calibration mode " (page 14 )). 5

26 Product Outline LCD The LCD backlight will be turned OFF automatically when the period of the set [B-Light Off Time] has passed after the last button operation (refer to " B-Light Off Time " (page 83 )). ). Any button operations turn ON the light again. Measurement/calibration screen example This screen appears when measurement or calibration is performed. " Calibration " (page ( 23 ) " Measurement " (page ( 33 ) Fig. 5 Measurement/calibration screen example No. Name Description 1 Alarm icon Blinks when an abnormal condition occurs during measurement (refer to " Alarm displays and actions " (page 102)). Yellow: Caution alert Red: Warning alarm 2 Warm-up icon Blinks for 30 minutes after the power is turned ON. 3 USB icon Lights up while a USB memory stick is inserted. 4 Screen title Indicates the name of the screen. 5 Process display Shows the measurement or calibration process. 6 Measured value display Shows the measured value. 7 Operation guide display Shows the button operations that are used to move to the next action. 6

27 Product Outline Example of item selection screen This screen appears for the operations of data management or setting. " Data Top screen " (page 56 ) " Setting Top screen " (page 68 ) Fig. 6 Example of item selection screen Example of pop-up screen The pop-up screen below appears when needed for selection/entry in setting operations. Selection list display This appears when a setting item of selection type is selected. " Selection list display " (page 15 ) " Setting " (page ( 68 ) Fig. 7 Pop-up screen example (selection list) 7

28 Product Outline Numeric keys These appear when the settings function is used to select a setting item that requires entry of a numerical value. " Numeric keys " (page 16 ) " Setting " (page ( 68 ) Fig. 8 Pop-up screen example (numeric keys) Character keys If [Save Memo] is set to "ON" in the measurement settings, these appear when measured values are saved. " Character keys " (page 17 ) " Save Memo " (page 77 ) Fig. 9 Input screen example (character keys) 8

29 Main Unit Setup Main Unit Setup Putting the valve stickers If you use the manual measurement/calibration mode, it may be useful to put the valve stickers on the operation button panel as shown below. Fig. 10 Positions of the valve stickers Placing the absorbent sheet Follow the steps below to place the absorbent sheet on the tray in the main unit. 1. Turn OFF the power. 2. Open the right cover. 3. Place the absorbent sheet on the tray using tweezers. Fig. 11 Tray in the main unit 4. Close the right cover. 9

30 Basic Operation Basic Operation Power ON WARNING Electric shock To prevent electric shock, ground the product. Do not ground the product to dangerous places such as a gas pipe. 1. Insert the provided power cable into the power cable connector on the back of the main unit. Fig. 12 Power cable connection 2. Turn the power ON, using the switch on the back of the main unit. The power turns ON and the initial screen appears on the display followed by the Measurement Top screen. The warm-up icon blinks for 30 minutes. Fig. 13 Initial screen The model, program number, and version that are shown on the initial screen vary by product. 10

31 Basic Operation Fig. 14 Measurement Top screen Power OFF The main unit is not stable while the warm-up icon blinks. Although measurement is possible while the warm-up icon blinks, the alarm icon will blink after measurement is finished and an invalid data error will occur (refer to " Current Alarm screen " (page 57 ) and " Alarm displays and actions " (page 102)). ). For correct measurement, be sure to wait until the warm-up icon turns off before starting calibration or measurement. For high-precision measurement, warm up the analyzer at least an hour before calibration or measurement. 3. Refer to " System Setting screen " (page 82 ) to set the date and time. 1. Clean the flow paths inside the product. " Rinsing the flow paths " (page 86 ) 2. Turn OFF the power switch. 3. Place the cap on the sample inlet. 4. Remove the power cable plug from the power outlet. 5. Dispose of the drainage liquid. " Solvent S-316 " (page 109) 11

32 Basic Operation Drainage mode operation You can start liquid drainage manually by changing to the drainage mode. The procedure for draining liquid in the drainage mode is explained below. 1. Press the up or down button with the Measurement Top screen appearing until the process display shows [Drainage]. The drainage mode is entered. 2. Press the ENT button. Liquid drainage starts. Fig. 15 Drainage mode - Drainage When the set [Drainage Time] elapses, drainage ends and the Measurement Top screen returns. If an abnormal condition occurs during measurement, calibration, or liquid drainage, operation may stop. The alarm icon will blink. Check the alarm information on the Current Alarm screen (refer to " Current Alarm screen " (page 57 ) and " Alarm displays and actions " (page 102)). ). Operations while the sequence is in progress Pause During measurement, the sequence can be paused in the automatic measurement mode, automatic zero calibration mode, and automatic span calibration mode. 1. Press the ESC button during measurement. The sequence enters the paused state. 2. To resume operation from the paused state, press the ENT button. Measurement will resume from the point where it was stopped. If the ESC button is pressed in the paused state, the measurement is aborted and the measurement liquid is drained. When an error occurs When a light source error or unstable data alarm occurs, measurement stops, the Measurement Top screen returns, and the alarm icon blinks. " Alarm displays and actions " (page 102) 12

33 Basic Operation Connecting a USB memory stick This section explains how to connect a USB memory stick with the product. When a USB memory stick is connected, the following operations can be performed. Saving the measurement history to an USB memory stick Saving the calibration history to an USB memory stick Saving the settings of the main unit to an USB memory stick Use a FAT/FAT32 formatted USB stick. Other formats are not available for this product. HORIBA-recommended USB memory sticks are available. Consult your dealer. Do not lose the cap for the USB memory port. When handling liquids during measurement, calibration, or otherwise, remove the USB memory stick from the USB memory port and cap the port. If a liquid spills on a USB memory stick or the USB memory port, the liquid may enter the interior of the product from the USB memory port and cause product damage. If any of the operations are attempted without inserting a USB memory stick into the USB memory port, a message of "Process has failed" will appear. For details on each of the operations, refer to " USB Memory screen " (page 60 ). 1. Remove the cap from the USB memory port on the left side of the product. 2. Insert the USB memory stick into the USB memory port. Fig. 16 Inserting a USB memory stick The USB icon blinks on the screen. Fig. 17 USB icon 13

34 Basic Operation Screen operations in the manual measurement/calibration mode On the instantaneous value measurement screen, which appears when the manual measurement mode, manual zero calibration mode, or manual span calibration mode starts, you can switch to any operations below to control extraction, cell filling, measurement, and drain manually. Stirring motor: Rotate/stop the stirring blades inside the extraction tank. Fill cell valve: Start/stop flow from the extraction tank to the measurement cell. Drainage valve: Start/stop drainage from the measurement cell and extraction tank. " Meas. Mode " (page 73 ) " Calib. Mode " (page 81 ) " Manual measurement " (page 44 ) " Manual zero calibration " (page 29 ) " Manual span calibration " (page 30 ) The procedure for performing operations with the instantaneous value measurement screen appearing is explained below. Fig. 18 Example of instantaneous value measurement screen (manual measurement) The buttons and button functions, which can be used with the instantaneous value measurement screen appearing, are described in the table below. Table 1 Operable buttons with the instantaneous value measurement screen Button Function ENT button Starts measurement. Up button Opens and closes the fill cell valve (UPPER V). Down button Opens and closes the drainage valve (LOWER V). Right button Switches the stirring motor ON and OFF (STIR). ESC button Returns to the mode start state. 14

35 Basic Operation The relations between part control and product operation are shown in the table below. Product operation Stirring motor Fill cell valve Drainage valve Extraction ON Closed Closed Cell filling OFF Open Closed Measurement OFF Open Closed Drainage (all) OFF Open Open Drainage (solvent) OFF Closed Open Using pop-up screens A pop-up screen for selection or entry will appear when it is necessary for you to select an item or enter a number or characters when configuring settings or performing other operations. If [Save Memo] is set to "ON" in the measurement settings, a pop-up screen will appear for entering the data name, before measured values being saved. The procedures for using the pop-up screens are explained below. Selection list display This screen is used to configure settings. Values that can be selected are shown in a list. " Setting " (page ( 68 ) Fig. 19 Example of selection list pop-up screen The buttons and button functions, which can be used with a selection list pop-up screen appearing, are described in the table below. Table 2 Operable buttons with a selection list pop-up screen Button Function ENT button Applies the currently selected setting and closes the screen. Up button Selects the next item up. Down button Selects the next item down. ESC button Cancels changes and closes the screen. 15

36 Basic Operation Follow the steps below to change a setting. 1. Press the up or down button to select the desired value. 2. Press the ENT button. The selected value is applied. Numeric keys This screen is used to configure settings. Numeric keys and an input box appear. " Setting " (page ( 68 ) Fig. 20 Example of numeric key pop-up screen The buttons and button functions, which can be used with a numeric key pop-up screen appearing, are described in the table below. Table 3 Operable buttons with a numeric key pop-up screen Button Function ENT button Applies the currently entered value and closes the screen. Up button Selects the next key up. Down button Selects the next key down. Left button Selects the next key to the left. Right button Selects the next key to the right. ESC button Cancels changes and closes the screen. Follow the steps below to change a setting. 1. Press the up/down/left/right button to select a numeric key. The selected key is shown in red. 2. Press the ENT button. The selected value appears in the input box. 3. Repeat steps 1. to 2. to enter the desired numeric value in the input box. 4. Press the up/down/left/right button to select the [ENT] key, and press the ENT button. The entered value, which appears in the input box, is applied. 16

37 Basic Operation Character keys This screen is used to enter names of memo data such as measurement conditions. If [Save Memo] is set to "ON" in the measurement settings, this screen appears immediately before measured values are saved. " Save Memo " (page 77 ) Fig. 21 Example of character key pop-up screen The buttons and button functions, which can be used with a character key pop-up screen appearing, are described in the table below. Table 4 Operable buttons with a character key pop-up screen Button Function ENT button Applies the currently entered value and closes the screen. Up button Selects the next key up. Down button Selects the next key down. Left button Selects the next key to the left. Right button Selects the next key to the right. ESC button Cancels changes and closes the screen. Follow the steps below to change a setting. 1. Press the up/down/left/right button to select a character key. The selected key is shown in red. 2. Press the ENT button. The selected character appears in the input box. 3. Repeat steps 1. to 2. to enter the desired characters in the input box. 4. Press the up/down/left/right button to select the [ENT] key, and press the ENT button. The characters that appear in the input box are applied. 17

38 Preparation Preparation Measurement preparation cautions CAUTION Chemical hazard (solvent S-316) Inhalation or accidental ingestion of a large amount of solvent S-316 may be harmful. Observe the following rules when handling: Ventilate the work area sufficiently. Wear a protective mask and protective gloves. Wash hands well after handling the solvent. Chemical hazard (hydrochloric acid) Hydrochloric acid is toxic by skin or eye exposure. If it touches the skin, immediately rinse with water. If it reaches the eyes, rinse immediately under a large amount of running water and get medical attention. To obtain correct measurement results, it is important to eliminate factors affecting measured values and to keep calibration and measurement conditions as uniform as possible. Observe the precautions below. Be sure to wash your hands before starting work. If oil from your fingers gets on the sample inlet on the product, the measuring utensils, or other parts, the measured value will be affected and correct measurement will be impossible. When measuring sample or reagent quantities, use suitable measuring utensils and measure accurately. When using a microsyringe or measuring syringe, take care that foam and bubbles are not drawn into the syringe. Foam and bubbles will increase measurement error. In addition, read "Operational precautions" and "Solvent handling precautions" in the front matter of this manual. 18

39 Preparation Glassware Items required Beakers (200 ml): 1 or 2 (for drainage) Measuring cylinders (20 ml) or syringes (20 ml, optional): 2 (for solvent and sample) Solvent S-316 (optional) Zero liquid for calibration (refer to " Zero liquid for calibration " (page 20 )) Span liquid for calibration (refer to " Span liquid for calibration " (page 20 )) Hydrochloric acid (refer to " Hydrochloric acid " (page 21 )) Placing the beakers for drainage Follow the steps below to place the drainage beakers. 1. Put tap water into the beakers so that the water level is approx. 1 cm from the bottom. 2. Place the beaker or beakers as shown below. Fig. 22 Placing beakers for drainage Cleaning the measuring cylinders or syringes Clean the measuring cylinders or syringes (for solvent and for the sample) with pure solvent S

40 Preparation Zero liquid for calibration Prepare pure solvent S-316, the same as is used for measurement. Span liquid for calibration Use pure solvent S-316, the same as is used for measurement, to prepare the span liquid for calibration. Using B-heavy oil Use B-heavy oil (specific gravity at 20 C) to prepare span liquid for OCMA-500. When you know the oil type to be measured, you can also use that oil type for the calibration oil. Items required Scale Glass container with a lid (screw-top bottle is recommended) Measuring flask Solvent S-316 (optional) B-heavy oil (specific gravity at 20 C) or calibration oil (when the oil type to be measured is known) Clean the glass utensils to be used with pure solvent S-316, and let them air dry completely. If the utensils cannot be dried completely, purge them 3 times or 4 times using solvent S-316. It is difficult to measure B-heavy oil with a microsyringe because it has a high viscosity. Use a suitably sized glass container with a lid (a screw-top bottle is recommended) for measurement. Preparation method Prepare the liquid so that it has a concentration of approximately twice the measured value of the sample. You can change the calibration value to be input as appropriate for the volume that is actually measured. 1. Use a scale to accurately measure the B-heavy oil in a glass container (with lid) of suitable size. 2. Transfer the B-heavy oil from the glass container to the measuring flask, while cleaning with S Fill the measuring flask to the graduation with solvent S Insert the stopper into the measuring flask and mix the contents well. 20

41 Preparation Hydrochloric acid When an acid is added to a water sample containing organic matter, the solvent and the sample separate easily (salting-out effect). When performing oil content extraction, the salting-out effect can be produced by adding approx. 6 mol/l hydrochloric acid. The procedure for preparing 6 mol/l hydrochloric acid using commercially available concentrated hydrochloric acid (36%) is explained below. Hydrochloric acid preparation method Items required Glass beaker Glass measuring utensil (measuring flask, measuring cylinder, etc.) Glass rod Pure water Commercially available concentrated hydrochloric acid (36%) Clean the glass utensils to be used with pure water, and let them air dry completely. Preparation method 1. Use a measuring cup to determine a specific volume of pure water and transfer it to the glass beaker. 2. Add the same volume (as the pure water) of commercially available hydrochloric acid, adding gradually by running the hydrochloric acid down the glass rod. Always add the hydrochloric acid to the pure water. Do not pour in the hydrochloric acid all at once. Using an incorrect preparation method may cause heat generation and explosive boiling. 21

42 Preparation Condition settings Standard calibration conditions and measurement conditions are set by default in the product. Once the warm-up icon has turned off, calibration and measurement can be started immediately. Calibration condition settings The default settings for the calibration conditions and the pages to refer to for the setting procedures are shown below. Measurement condition Default setting Setting procedure page Span Point mg/l " Span Point " (page 79 ) Extraction Time 40 sec " Extraction Time " (page 80 ) Separation Time 30 sec " Separation Time " (page 80 ) Number of Purge 2 time " Number of Purge " (page 81 ) Calib. Mode AUTO " Calib. Mode " (page 81 ) Measurement condition settings The default settings for the measurement conditions and the pages to refer to for the setting procedures are shown below. Measurement condition Default setting Setting procedure page Extraction Time 40 sec " Extraction Time " (page 70 ) Separation Time 30 sec " Separation Time " (page 71 ) Fill Cell Time 60 sec " Fill Cell Time " (page 71 ) Meas. Limit 300 sec " Meas. Limit " (page 72 ) Drainage Time 30 sec " Drainage Time " (page 72 ) Number of Purge 2 time " Number of Purge " (page 73 ) Meas. Mode AUTO " Meas. Mode " (page 73 ) Measurement Unit mg/l " Measurement Unit " (page 74 ) Solvent Vol. 8.0 ml " Solvent Vol. " (page 74 ) Sample Vol ml " Sample Vol. " (page 75 ) Zero Shift Value 0.0 mg/l " Zero Shift Value " (page 75 ) Use Light ON " Use Light " (page 76 ) Confirm Save AUTO " Confirm Save " (page 76 ) Save Memo OFF " Save Memo " (page 77 ) Display Negative OFF " Display Negative " (page 77 ) Display Raw Data OFF " Display Raw Data " (page 78 ) 22

43 Calibration Calibration Calibration cautions CAUTION Chemical hazard (solvent S-316) Inhalation or accidental ingestion of a large amount of solvent S-316 may be harmful. Observe the following rules when handling: Ventilate the work area sufficiently. Wear a protective mask and protective gloves. Wash hands well after handling the solvent. Chemical hazard (hydrochloric acid) Hydrochloric acid is toxic by skin or eye exposure. If it touches the skin, immediately rinse with water. If it reaches the eyes, rinse immediately under a large amount of running water and get medical attention. When dispensing liquids into the extraction tank, limit the solvent amount to 10 ml or less and the amount of the measurement liquid to 20 ml or less, so that the total volume of liquid does not exceed the upper limit line (30 ml). If the liquid exceeds the upper limit line, the liquid may leak and cause short-circuit in the internal wiring of the product. If liquid containing hydrochloric acid leaks and comes in contact with the skin, irritation and burning may occur. Discard the liquid collected in the beakers before overflow. If liquid containing hydrochloric acid overflows and comes in contact with the skin, irritation and burning may occur. To obtain correct measurement results, it is important to eliminate factors affecting measured values and to keep calibration and measurement conditions as uniform as possible. Observe the precautions below. If the temperature of the measurement liquid or solvent is lower than the inside of the main unit, condensation may form inside the main unit and the indicated value may be unsteady or shift. If the room temperature is less than 5 C, the viscosity of solvent S-316 will increase and the indicated value may be low. In this case, raise the room temperature above 5 C, or set [Fill Cell Time] to a longer value. Avoid performing measurement at high temperature and high humidity. Moisture may cause the indicated value to shift. When the air has a high concentration of hydrocarbons, for instance in an oil refinery, hydrocarbons may be adsorbed inside the product and affect measurement. In this type of location, increase the [Number of Purge] value. Be sure to wash your hands before starting work. If oil from your fingers gets on the sample inlet on the product, the measuring utensils, or other parts, the measured value will be affected and correct measurement will be impossible. Perform measurement with the right cover closed. Measurement cannot be stable with the right cover open. Be sure to perform zero calibration before every measurement. Perform span calibration from once a day to once a week. When you change to a new solvent lot, perform zero calibration and span calibration before measurement using the calibration liquids prepared from the new lot of solvent. 23

44 Calibration Be sure to perform zero calibration before performing span calibration. If zero calibration is performed after span calibration, correct measurement will be impossible. Purging is necessary to prevent effects from the previous measurement liquid. Set [Number of Purge] from 2 to 6. If the concentration changes by 100 mg/l or more from the previous measurement, set [Number of Purge] to 5 or more. When using the built-in extractor, maintain an extraction ratio of solvent:water = 1:2. The instructions in this manual indicate 8 ml of solvent and 16 ml of clean water or sample water to provide an extraction ratio of solvent:water = 1:2. When the built-in extractor is not used, the extraction ratio can be changed. However, be sure to keep the solvent to sample water extraction ratio the same between measurement and calibration. If the extraction ratios are not the same, measurement error will occur due to the differences in the water content of the solvent. In addition, this product is designed for measurement at an extraction ratio of solvent:sample water = 1:2. Other extraction ratios may cause greater measurement error and the specified product performance cannot be guaranteed. When measuring sample or reagent quantities, use suitable measuring utensils and measure accurately. When using a microsyringe or measuring syringe, take care that foam and bubbles are not drawn into the syringe. Foam and bubbles will increase measurement error. Dispense liquid slowly into the extraction tank. A vigorous dispense of liquid may cause overflow. If an abnormality occurs during measurement, calibration, or liquid drainage, operation may stop. The alarm icon will blink. Check the alarm information on the Current Alarm screen (refer to " Current Alarm screen " (page 57 ) and " Alarm displays and actions " (page 102)). In addition, read "Operational precautions" and "Solvent handling precautions" in the front matter of this manual. 24

45 Calibration Points to check prior to calibration Points to check prior to zero calibration Is the warm-up icon off? Is the alarm icon off? Are the beakers for drainage in place? If the warm-up icon is blinking, wait until it turns off. If the alarm icon is blinking, check the error information and remove the cause (refer to " List of alarms " (page 102)). If not, set the beakers for drainage in place (refer to " Placing the beakers for drainage " (page 19 )). Points to check prior to span calibration Did you perform zero calibration? Is the warm-up icon off? Is the alarm icon off? Are the beakers for drainage in place? If not, first perform zero calibration. If the warm-up icon is blinking, wait until it turns off. If the alarm icon is blinking, check the error information and remove the cause (refer to " List of alarms " (page 102)). If not, set the beakers for drainage in place (refer to " Placing the beakers for drainage " (page 19 )). Items required Zero calibration OCMA-500 (this product) Measuring cylinders (20 ml) or syringes (20 ml, optional): 2 (for solvent and sample) Dropper (provided): 1 (for hydrochloric acid) Zero liquid for calibration (refer to " Zero liquid for calibration " (page 20 )) Hydrochloric acid (refer to " Hydrochloric acid " (page 21 )) Clean water * Span calibration OCMA-500 (this product) Measuring cylinders (20 ml) or syringes (20 ml, optional): 2 (for solvent and sample) Dropper (provided): 1 (for hydrochloric acid) Span liquid for calibration (refer to " Span liquid for calibration " (page 20 )) Hydrochloric acid (refer to " Hydrochloric acid " (page 21 )) Clean water * *: Use the same clean water for the sequence of tasks from zero calibration to measurement. It is recommended that pure water is used. When the sample to be measured contains large amounts of water-soluble substances, such as urea or NaCl, and the concentration is known, perform zero calibration using that oil-free aqueous solution. When water is not used for measurement, such as when measuring the oil content of metal parts, do not use clean water for zero calibration (do not perform step 5. (page 26 ) of the zero calibration procedure). 25

46 Calibration Automatic zero calibration 1. Set the calibration mode to "AUTO" on the Calibration Setting screen. Change other settings as needed. " Calibration condition settings " (page 22 ) " Calibration Setting screen " (page 78 ) 2. Press the CAL button or up/down button with the Measurement Top screen appearing until the process display shows [Auto Zero Calibration]. The automatic zero calibration mode is entered. Fig. 23 Start of the automatic zero calibration mode 3. Use the measuring cylinder or measuring syringe (for solvent) to dispense 8 ml of zero liquid for calibration through the sample inlet. 4. Use the dropper to add one drop of hydrochloric acid through the sample inlet. 5. Use the measuring cylinder or measuring syringe (for sample) to dispense 16 ml of clean water through the sample inlet. 6. Press the ENT button. If the set [Number of Purge] has not been completed: The purge sequence starts. The screen title changes to [Auto Zero Calibration (Purge: 1)], and the process display changes as follows as the sequence progresses: [Extraction] [Layer Separation] [Fill Cell] [Drainage] The remaining time of the current process is shown (counting down). When the purge sequence ends, the start state of the automatic zero calibration mode returns. When the remaining [Number of Purge] is 0: The measurement sequence starts. The screen title changes to [Auto Zero Calibration: Measurement], and the process display changes as follows as the sequence progresses: [Extraction] [Layer Separation] [Fill Cell] [Calibration] The remaining time of the current process is shown (counting down) during extraction, layer separation, and cell filling. The elapsed time is shown (counting up) during calibration. When calibration ends, the calibration result appears. After the result appears, the measurement liquid is drained. During drainage, the process display on the screen shows [Drainage]. 26

47 Calibration Fig. 24 Display of automatic zero calibration result When drainage is completed, the start state of the automatic zero calibration mode returns. 7. Repeat steps 3. to 6. until the set [Number of Purge] is completed and measurement starts. Automatic span calibration 1. Set the calibration mode to "AUTO" on the Calibration Setting screen. Change other settings as needed. The value set in [Span Point] in the calibration settings must be adjusted based on the extraction ratio. For example, when the extraction ratio is solvent:water = 1:2, set 1/2 the value of the actual concentration of the span liquid for calibration in [Span Point] in the calibration settings. " Calibration condition settings " (page 22 ) " Calibration Setting screen " (page 78 ) 2. Press the CAL button or up/down button with the Measurement Top screen appearing until the process display shows [Auto Span Calibration]. The automatic span calibration mode is entered. Fig. 25 Start of the automatic span calibration mode 3. Use the measuring cylinder or measuring syringe (for solvent) to dispense 8 ml of span liquid for calibration through the sample inlet. 4. Use the dropper to add one drop of hydrochloric acid through the sample inlet. 5. Use the measuring cylinder or measuring syringe (for sample) to dispense 16 ml of clean water through the sample inlet. 27

48 Calibration 6. Press the ENT button. If the set [Number of Purge] has not been completed: The purge sequence starts. The screen title changes to [Auto Span Calibration (Purge: 1)], and the process display changes as follows as the sequence progresses: [Extraction] [Layer Separation] [Fill Cell] [Drainage] The remaining time of the current process is shown (counting down). When the purge sequence ends, the start state of the automatic span calibration mode returns. When the remaining [Number of Purge] is 0: The measurement sequence starts. The screen title changes to [Auto Span Calibration: Measurement] and the process display changes as follows as the sequence progresses: [Extraction] [Layer Separation] [Fill Cell] [Calibration] The remaining time of the current process is shown (counting down) during extraction, layer separation, and cell filling. The elapsed time is shown (counting up) during calibration. When calibration ends, the calibration result appears. After the result appears, the measurement liquid is drained. During drainage, the process display on the screen shows [Drainage]. Fig. 26 Display of automatic span calibration result When drainage is completed, the start state of the automatic span calibration mode returns. 7. Repeat steps 3. to 6. until the set [Number of Purge] is completed and measurement starts. 28

49 Calibration Manual zero calibration 1. Set the calibration mode to "MANUAL" on the Calibration Setting screen. Change other settings as needed. " Calibration condition settings " (page 22 ) " Calibration Setting screen " (page 78 ) 2. Press the CAL button or up/down button with the Measurement Top screen appearing until the process display shows [Manual Zero Calibration]. The manual zero calibration mode is entered. Fig. 27 Start of the manual zero calibration mode 3. Use the measuring cylinder or measuring syringe (for solvent) to dispense 8 ml of zero liquid for calibration through the sample inlet. 4. Use the dropper to add one drop of hydrochloric acid through the sample inlet. 5. Use the measuring cylinder or measuring syringe (for sample) to dispense 16 ml of clean water through the sample inlet. 6. Press the ENT button. The instantaneous value measurement screen appears. The screen title shows [Manual Zero Calibration], and the process display shows the current status of the valves. The operation guide display shows the stirring motor and valve operation buttons and [ENT: Calib. ESC: Return]. Fig. 28 Manual zero calibration - Instantaneous value measurement 7. If needed, execute purging (extraction, cell filling, and drainage) referring to " Screen operations in the manual measurement/calibration mode " (page 14 ). 8. Execute extraction and fill the cell at any time referring to steps 3. to 6. and " Screen operations in the manual measurement/calibration mode " (page 14 ). 29

50 Calibration 9. Press the ENT button. Measurement starts. The operation guide display shows [ESC: Cancel]. The elapsed time is shown (counting up). When measurement ends, the calibration result appears. Fig. 29 Manual zero calibration - Measurement When measurement ends normally, the measured value is held. The operation guide display shows [ESC: Return]. Fig. 30 Manual zero calibration - Result 10. Press the ESC button. The instantaneous value measurement screen returns. 11. Execute drainage referring to " Screen operations in the manual measurement/calibration mode " (page 14 ) at any time. 12. Press the ESC button. The start state of the manual zero calibration mode returns. Manual span calibration 1. Set the calibration mode to "MANUAL" on the Calibration Setting screen. Change other settings as needed. The value set in [Span Point] in the calibration settings must be adjusted based on the extraction ratio. For example, when the extraction ratio is solvent:water = 1:2, set 1/2 the value of the actual concentration of the span liquid for calibration in [Span Point] in the calibration settings. " Calibration condition settings " (page 22 ) " Calibration Setting screen " (page 78 ) 30

51 Calibration 2. Press the CAL button or up/down button with the Measurement Top screen appearing until the process display shows [Manual Span Calibration]. The manual span calibration mode is entered. Fig. 31 Start of the manual span calibration mode 3. Use the measuring cylinder or measuring syringe (for solvent) to dispense 8 ml of span liquid for calibration through the sample inlet. 4. Use the dropper to add one drop of hydrochloric acid through the sample port. 5. Use the measuring cylinder or measuring syringe (for sample) to dispense 16 ml of clean water through the sample inlet. 6. Press the ENT button. The instantaneous value measurement screen appears. The screen title shows [Manual Span Calibration] and the process display shows the current status of the valves. The operation guide display shows the stirring motor and valve operation buttons and [ENT: Calib. ESC: Return]. Fig. 32 Manual span calibration - Instantaneous value measurement 7. If needed, execute purging (extraction, cell filling, and drainage) referring to " Screen operations in the manual measurement/calibration mode " (page 14 ). 8. Execute extraction and fill the cell at any time referring to steps 3. to 6. and " Screen operations in the manual measurement/calibration mode " (page 14 ). 9. Press the ENT button. Measurement starts. The operation guide display shows [ESC: Cancel]. The elapsed time is shown (counting up). When measurement ends, the calibration result appears. Fig. 33 Manual span calibration - Measurement When measurement ends normally, the measured value is held. 31

52 Calibration The operation guide display shows [ESC: Return]. Fig. 34 Manual span calibration - Result 10. Press the ESC button. The instantaneous value measurement screen returns. 11. Execute drainage referring to " Screen operations in the manual measurement/calibration mode " (page 14 ) at any time. 12. Press the ESC button. The start state of the manual span calibration mode returns. 32

53 Measurement Measurement Measurement cautions CAUTION Chemical hazard (solvent S-316) Inhalation or accidental ingestion of a large amount of solvent S-316 may be harmful. Observe the following rules when handling: Ventilate the work area sufficiently. Wear a protective mask and protective gloves. Wash hands well after handling the solvent. Chemical hazard (hydrochloric acid) Hydrochloric acid is toxic by skin or eye exposure. If it touches the skin, immediately rinse with water. If it reaches the eyes, rinse immediately under a large amount of running water and get medical attention. When dispensing liquids into the extraction tank, limit the solvent amount to 10 ml or less and the amount of the measurement liquid to 20 ml or less, so that the total volume of liquid does not exceed the upper limit line (30 ml). If the liquid exceeds the upper limit line, the liquid may leak and cause short-circuit in the internal wiring of the product. If liquid containing hydrochloric acid leaks and comes in contact with the skin, irritation and burning may occur. Discard the liquid collected in the beakers before overflow. If liquid containing hydrochloric acid overflows and comes in contact with the skin, irritation and burning may occur. To obtain correct measurement results, it is important to eliminate factors affecting measured values and to keep calibration and measurement conditions as uniform as possible. Observe the precautions below. If the temperature of the measurement liquid or solvent is lower than the inside of the main unit, condensation may form inside the main unit and the indicated value may be unsteady or shift. If the room temperature is less than 5 C, the viscosity of solvent S-316 will increase and the indicated value may be low. In this case, raise the room temperature above 5 C, or set [Fill Cell Time] to a longer value. Avoid performing measurement at high temperature and high humidity. Moisture may cause the indicated value to shift. When the air has a high concentration of hydrocarbons, for instance in an oil refinery, hydrocarbons may be adsorbed inside the product and affect measurement. In this type of location, increase the [Number of Purge] value. Be sure to wash your hands before starting work. If oil from your fingers gets on the sample inlet on the product, the measuring utensils, or other parts, the measured value will be affected and correct measurement will be impossible. Perform measurement with the right cover closed. Measurement cannot be stable with the right cover open. Be sure to perform zero calibration before every measurement. Perform span calibration from once a day to once a week. When you change to a new solvent lot, perform zero calibration and span calibration before measurement using the calibration liquids prepared from the new lot of solvent. 33

54 Measurement Be sure to perform zero calibration before performing span calibration. If zero calibration is performed after span calibration, correct measurement will be impossible. Purging is necessary to prevent effects from the previous measurement liquid. Set [Number of Purge] from 2 to 6. If the concentration changes by 100 mg/l or more from the previous measurement, set [Number of Purge] to 5 or more. When using the built-in extractor, maintain an extraction ratio of solvent:water = 1:2. The instructions in this manual indicate 8 ml of solvent and 16 ml of clean water or sample water to provide an extraction ratio of solvent:water = 1:2. When the built-in extractor is not used, the extraction ratio can be changed. However, be sure to keep the solvent to sample water extraction ratio the same between measurement and calibration. If the extraction ratios are not the same, measurement error will occur due to the differences in the water content of the solvent. In addition, this product is designed for measurement at an extraction ratio of solvent:sample water = 1:2. Other extraction ratios may cause greater measurement error and the specified product performance cannot be guaranteed. When measuring sample or reagent quantities, use suitable measuring utensils and measure accurately. When using a microsyringe or measuring syringe, take care that foam and bubbles are not drawn into the syringe. Foam and bubbles will increase measurement error. Dispense liquid slowly into the extraction tank. A vigorous dispense of liquid may cause overflow. When measurement is completed for the day, clean the flow paths inside the product (refer to " Rinsing the flow paths " (page 86 )). If an abnormality occurs during measurement, calibration, or liquid drainage, operation may stop. The alarm icon will blink. Check the alarm information on the Current Alarm screen (refer to " Current Alarm screen " (page 57 ) and " Alarm displays and actions " (page 102)). In addition, read "Operational precautions" and "Solvent handling precautions" in the front matter of this manual. 34

55 Measurement Preprocessing Removing fine particles If the sample water contains fine particles or other impurities, use an ashless cellulose quantitative filter paper (particle retention: 8 µm) to remove the impurities before dispensing the sample into the product. Fig. 35 Removal of fine particles Performing preliminary extraction outside the product For the following types of sample water, perform extraction outside the product prior to dispensing the sample into the product. Sample water that has an oil membrane or oil drops floating on the surface It cannot be measured in a measuring utensil. Extraction must be performed using the entire sample. Sample water containing large amounts of suspended material (sand, organic materials, etc.) It will clog the measuring utensil or internal filter, and may damage the measuring utensil or the product. Sample water that is difficult to separate from the solvent after extraction (sample water containing emulsifying substances) If any incompletely separated solvent is sent to the water filter, emulsifying substances or water content in the solvent may cause clogging in the water filter and an accurate measurement will not be possible. " Examples of measurement by extraction outside the product " (page 47 ) Points to check prior to measurement Did you perform calibration? Is the warm-up icon off? Is the alarm icon off? Are the beakers for drainage in place? If not, first perform zero calibration, then perform span calibration. If the warm-up icon is blinking, wait until it turns off. If the alarm icon is blinking, check the error information and remove the cause (refer to " List of alarms " (page 102)). If not, set the beakers for drainage in place (refer to " Placing the beakers for drainage " (page 19 )). 35

56 Measurement Items required OCMA-500 (this product) Measuring cylinders (20 ml) or syringes (20 ml, optional): 2 (for solvent and sample) Dropper (provided): 1 (for hydrochloric acid) Clean solvent S-316 (optional) Hydrochloric acid (refer to " Hydrochloric acid " (page 21 )) Sample water Automatic measurement Deciding the extraction time 40-second extraction results in a uniform extraction state for normal sample water, however, the optimum extraction time depends on the properties of the sample water. If you do not know the optimum extraction time of the sample water to be measured, use semi-automatic measurement or manual measurement to vary the extraction time prior to automatic measurement, and compare with extraction outside the product to determine the optimum extraction time. " Examples of measurement by extraction outside the product " (page 47 ) " Semi-automatic measurement " (page 40 ) " Manual measurement " (page 44 ) " Measurement condition settings " (page 22 ) 36

57 Measurement Flow scheme of operation Fig. 36 Operation flow of automatic measurement 37

58 Measurement Procedure 1. Set the measurement mode to "AUTO" on the Measurement Setting screen. Change other settings as needed. " Measurement condition settings " (page 22 ) " Measurement Setting screen " (page 69 ) 2. Press the MEAS button or up/down button until the process display on the screen shows [Auto Measurement]. The automatic measurement mode is entered. Fig. 37 Start of the automatic measurement mode 3. Use the measuring cylinder or measuring syringe (for solvent) to dispense 8 ml of clean solvent through the sample inlet. 4. Use the dropper to add one drop of hydrochloric acid through the sample port. 5. Use the measuring cylinder or measuring syringe (for sample) to dispense 16 ml of sample water through the sample inlet. 6. Press the ENT button. If the set [Number of Purge] has not been completed: The purge sequence starts. The screen title changes to [Auto Measurement (Purge: 1)] and the process display changes as follows as the sequence progresses: [Extraction] [Layer Separation] [Fill Cell] [Drainage] The remaining time of the current process is shown (counting down). When the purge sequence ends, the start state of the automatic measurement mode returns. When the remaining [Number of Purge] is 0: The measurement sequence starts. The screen title changes to [Auto Measurement: Measurement] and the process display changes as follows as the sequence progresses: [Extraction] [Layer Separation] [Fill Cell] [Measure] The remaining time of the current process is shown (counting down) during extraction, layer separation, and cell filling. The elapsed time is shown (counting up) during measurement. When measurement ends, the measured value appears. 38

59 Measurement Fig. 38 Auto Measurement - Measurement During layer separation, make sure that the solvent and sample water separate sufficiently. If the solvent and sample water are insufficiently separated, press the ESC button to cancel measurement, and lengthen [Separation Time] in the measurement settings. If the nature of the sample is such that layer separation is difficult, it is recommended that you use semi-automatic measurement or manual measurement in advance to determine the optimum [Separation Time]. If the elapsed time exceeds 60 seconds and the measured value does not appear, replace the water filter (refer to " Replacing the water filter " (page 94 )). ). If [Confirm Save] is set to "MANUAL", the operation guide display will show [ENT: Save ESC: Skip]. To save the measured value, press the ENT button. To skip, press the ESC button. For details on the [Confirm Save] setting, refer to " Confirm Save " (page 76 ). If [Save Memo] is set to "ON" in the measurement settings, a pop-up screen to input the data name will appear immediately before the measured value is saved. To save the measurement conditions or other memo with the measured value, enter the data name and press the ENT button. To save only the measured value and not an additional memo, press the ESC button. For details on the [Save Memo] setting, refer to " Save Memo " (page 77 ).. For details on the pop-up screen to input the data name, refer to " Character keys " (page 17 ). After the result appears, the measurement liquid is drained. During drainage, the process display on the screen shows [Drainage]. When drainage ends, the start state of the automatic measurement mode returns. 7. Repeat steps 3. to 6. until the set [Number of Purge] is completed and measurement starts. 39

60 Measurement Semi-automatic measurement Flow scheme of operation Fig. 39 Operation flow of semi-automatic measurement 40

61 Measurement Procedure 1. Set the measurement mode to "SEMI-AUTO" on the Measurement Setting screen. Change other settings as needed. " Measurement condition settings " (page 22 ) " Measurement Setting screen " (page 69 ) 2. Press the MEAS button or up/down button until the process display on the screen shows [Semi-Auto Measurement]. The semi-automatic measurement mode is entered. Fig. 40 Start of the semi-automatic measurement mode 3. Use the measuring cylinder or measuring syringe (for solvent) to dispense 8 ml of clean solvent through the sample inlet. 4. Use the dropper to add one drop of hydrochloric acid through the sample inlet. 5. Use the measuring cylinder or measuring syringe (for sample) to dispense 16 ml of sample water through the sample inlet. 6. Press the ENT button. Extraction starts. The screen title changes to [Semi-Auto Measurement] and the process display shows [Extraction]. The operation guide display shows [ENT: Next ESC: Previous] The elapsed time appears (counting up). Fig. 41 Semi-automatic measurement - Extraction 7. Press the ENT button. Layer separation starts. The process display on the screen shows [Layer Separation]. The elapsed time appears (counting up). 41

62 Measurement Fig. 42 Semi-automatic measurement - Layer separation 8. Make sure that the solvent and sample water have sufficiently separated, and press the ENT button. Cell filling starts. The process display on the screen shows [Fill Cell]. The operation guide display shows [ENT: Next ESC: Drain] The elapsed time appears (counting up). Fig. 43 Semi-automatic measurement - Fill Cell 9. Press the ENT button. Measurement starts. The process display on the screen shows [Measure]. The operation guide display shows [ESC: Drain]. The elapsed time is shown (counting up). When measurement ends, the measured value appears. Fig. 44 Semi-automatic measurement - Measurement If [Confirm Save] is set to "MANUAL", the operation guide display will show [ENT: Save ESC: Skip]. To save the measured value, press the ENT button. To skip, press the ESC button. For details on the [Confirm Save] setting, refer to " Confirm Save " (page 76 ). If [Save Memo] is set to "ON" in the measurement settings, a pop-up screen to input the data name will appear immediately before the measured value is saved. To save the measurement conditions or other memo with the measured value, enter the data name and press the ENT button. To save only the measured value and not an additional memo, press the ESC button. For details on the [Save Memo] setting, refer to " Save Memo " (page 77 ).. For details on the pop-up screen to input the data name, refer to " Character keys " (page 17 ). After the result appears, the measurement liquid is drained. The process display on the screen shows [Drainage]. The operation guide display shows [ESC: Cancel]. The elapsed time appears (counting 42

63 Measurement up). 10. Press the ESC button. The start state of the semi-automatic measurement mode returns. If the measured value is not saved in step 9.,, purging is executed. After purging, you can return to step 3. to continue purging or execute measurement. 43

64 Measurement Manual measurement Scheme of operation Fig. 45 Operation flow of manual measurement 44

65 Measurement Procedure 1. Set the measurement mode to "MANUAL" on the Measurement Setting screen. Change other settings as needed. " Measurement condition settings " (page 22 ) " Measurement Setting screen " (page 69 ) 2. Press the MEAS button or up/down button until the process display on the screen shows [Manual Measurement]. The manual measurement mode is entered. Fig. 46 Start of the manual measurement mode 3. Use the measuring cylinder or measuring syringe (for solvent) to dispense 8 ml of clean solvent through the sample inlet. 4. Use the dropper to add one drop of hydrochloric acid through the sample inlet. 5. Use the measuring cylinder or measuring syringe (for sample) to dispense 16 ml of sample water through the sample inlet. 6. Press the ENT button. The instantaneous value measurement screen appears. The screen title shows [Manual Measurement] and the process display shows the current status of the valves. The operation guide display shows the stirring motor and valve operation buttons and [ENT: Save ESC: Skip]. Fig. 47 Manual measurement - Instantaneous value measurement 7. If needed, execute purging referring to " Screen operations in the manual measurement/calibration mode " (page 14 ). 8. Execute extraction and fill the cell at any time referring to steps 3. to 6. and " Screen operations in the manual measurement/calibration mode " (page 14 ). 9. Press the ENT button. Measurement starts. The elapsed time is shown (counting up). When measurement ends, the measured value appears. 45

66 Measurement Fig. 48 Manual measurement - Measurement If [Confirm Save] is set to "MANUAL", the operation guide display will show [ENT: Save ESC: Skip]. To save the measured value, press the ENT button. To skip, press the ESC button. For details on the [Confirm Save] setting, refer to " Confirm Save " (page 76 ). If [Save Memo] is set to "ON" in the measurement settings, a pop-up screen to input the data name will appear immediately before the measured value is saved. To save the measurement conditions or other memo with the measured value, enter the data name and press the ENT button. To save only the measured value and not an additional memo, press the ESC button. For details on the [Save Memo] setting, refer to " Save Memo " (page 77 ).. For details on the pop-up screen to input the data name, refer to " Character keys " (page 17 ). If measurement ends normally, the measured value is held. The operation guide display shows [ESC: Return]. Fig. 49 Manual measurement - Result 10. Press the ESC button. The instantaneous value measurement screen returns. 11. Execute drainage referring to " Screen operations in the manual measurement/calibration mode " (page 14 ) at any time. 12. Press the ESC button. The start state of the manual measurement mode returns. 46

67 Measurement Examples of measurement by extraction outside the product This section describes examples of extraction performed outside the product without using the built-in extractor. The methods described below are only examples for reference. These methods may not be the optimum methods for some sample types. Use an extraction and measurement method that is appropriate for the actual sample. Oil content in water For the following types of sample water, extraction must be performed outside the product prior to dispensing the sample into the product. Sample water that has an oil membrane or oil drops floating on the surface It cannot be measured in a measuring utensil. Extraction must be performed using the entire sample. Sample water containing large amounts of suspended material (sand, organic materials, etc.) It will clog the measuring utensil or internal filter, and may damage the measuring utensil or the product. Sample water that is difficult to separate from the solvent after extraction (sample water containing emulsifying substances) If any incompletely separated solvent is sent to the water filter, emulsifying substances or water content in the solvent may cause clogging in the water filter and an accurate measurement will not be possible. The water filter in the product serves to removes water particles. However if the sample water contains emulsifying substances, fine water particles may pass through the water filter. When this type of sample is measured repeatedly, the water filter tends to clog and accurate measurement is no longer possible. If the measured value suddenly becomes a negative value or extremely low value, it is likely that water has entered the measurement cell due to water filter clogging or damage. (A negative value indicated during zero calibration is not abnormal.) Replace the water filter at regular intervals, even if it does not appear to be dirty (refer to " Replacing the water filter " (page 94 )). ). Checking layer separation If a sample potentially contains emulsifying substances, check if the solvent layer and water layer are separated in advance. The following are typical examples of water that is likely to contain emulsifying substances. Miscellaneous domestic gray water Industrial wastewater Activated sludge water (when the killed bacteria in the activated sludge is introduced, the content of the bacteria dissolves and acts in the same way as emulsifying substances.) Items required Glass container with a lid (screw-top bottle is recommended) (50 ml) Sample water Clean solvent S-316 (optional) 47

68 Measurement Checking procedure 1. Dispense 8 ml of solvent and 16 ml of sample water into the glass container with a lid. 2. Screw the lid closed and shake for 1 minute by hand. Fig. 50 Mixing solvent and sample water by shaking 3. Place on a flat surface, and check if the mixture completely separates into two layers after 20 seconds to 30 seconds. Fig. 51 Checking layer separation If a boundary surface has formed between the solvent layer and water layer, it can be concluded that the layers "have been separated". The following cases show that the layers have not been separated. Separation into two layers takes more than 40 seconds A whitish cloudy layer forms between the two layers Bubbles with a diameter of greater than 10 mm get formed in multiple places throughout the liquid and a boundary surface cannot be identified 48

69 Measurement Extraction Items required Separating funnel (300 ml) *1 : 2 (for sample extraction and for zero point checking) Measuring cylinder (200 ml) *1 : 3 (for sample, for solvent, and for clean water) Dropper (provided): 1 (for hydrochloric acid) ph meter Filter paper (if suspended material will be removed and as needed) *2 Glass container (if suspended material will be removed or water will be removed from solvent layer) *1 Teflon membrane filter (mesh diameter 20 µm to 40 µm) (if water will be removed from solvent layer) Glass funnel *1 (if water will be removed from solvent layer) Measuring utensil (if water will be removed from solvent layer) Sample water Clean water *3 Hydrochloric acid (refer to " Hydrochloric acid " (page 21 )) Clean solvent S-316 (optional) Anhydrous sodium sulfate (Na2SO4) (if water will be removed from solvent layer) *1: Clean the glass utensils with solvent in advance and let air dry. *2: Use solvent to elute and clean organic material from the filter paper in advance, and let air dry. *3: Use the same clean water for the sequence of tasks from zero calibration to measurement. Normally pure water should be used. If the sample contains large amounts of water-soluble substances, such as urea or NaCl, and the concentration is known, perform zero calibration using this oil-free aqueous solution. Extraction procedure 1. Measure 200 ml of clean water into the measuring cylinder for clean water, measure 100 ml of clean solvent into the measuring cylinder for solvent, and dispense these into the separating funnel for zero point checking. 2. Measure 160 ml to 200 ml of sample water into the measuring cylinder for sample, and dispense into the separating funnel for sample extraction. If the sample contains emulsifying substances, it is recommended that you use a smaller amount of sample water (approx. 100 ml). 3. Measure an amount of clean solvent equal to 1/2 the amount of sample water in the measuring cylinder for solvent, and transfer this to the measuring cylinder for sample. 4. Rinse the inside of the measuring cylinder for sample with the solvent, and add this solvent to the separating funnel for sample extraction. 5. Use the dropper to add 0.2 ml to 0.5 ml (5 drops to 10 drops) of hydrochloric acid to each separating funnel (for sample extraction and for zero point checking). If the sample contains emulsifying substances, there is no need to add hydrochloric acid. 6. Use the ph meter to check the ph value (ph 2 to ph 3) of each separating funnel (for sample extraction and for zero point checking). 49

70 Measurement 7. Insert the stopper in each separating funnel (for sample extraction and for zero point checking) and shake for approx. 5 minutes. If the sample contains emulsifying substances, it is recommended that you shake for a shorter time (approx. 1 minute). Fig. 52 Extraction The internal pressure will increase as you shake. Open the cock from time to time to reduce the internal pressure. 8. After shaking, wait until the water and solvent get separated, and check the status of the sample water. If almost no suspended material is visible and the water and solvent have separated, collect the solvent layer (lower layer) and use that as the measurement liquid. If there is suspended material or the water and solvent have not sufficiently separated and the solvent layer is cloudy, proceed to the "Removing suspended material" and "Removing water from the solvent layer" procedures. Removal of suspended material 1. Filter the solvent layer through the filter paper into the glass container and use this as the measurement liquid. Removing water from the solvent layer 1. Collect only parts that can be recognized as solvent in the glass container, and check the status of the liquid. If no water particles are visible and the liquid is clear, use it as measurement liquid. If you can see water particles with a diameter of 0.1 mm or larger or the liquid appears whitish overall, follow the steps below to remove water from the solvent layer. 50

71 Measurement Fig. 53 Separating the solvent layer Fig. 54 Visually checking the solvent layer 2. Filter the liquid through the Teflon membrane filter. 3. Gradually add the anhydrous sodium sulfate and shake well to mix. If the sample contains emulsifying substances, place filter paper in the funnel and filter the solvent layer into a container containing 10 g of anhydrous sodium sulfate. Fig. 55 Removing water from the solvent layer 4. Continue performing step 3. until the solvent layer is clear. 5. If anhydrous sodium sulfate crystals remain, filter with filter paper. 6. After waiting, collect the supernatant liquid using a measuring utensil and use this as measurement liquid. 51

72 Measurement Zero calibration Perform zero calibration using the conditions in the table below. Parameter Extraction Time Separation Time Number of Purge Zero liquid for calibration 0 sec 0 sec 3 or more Condition Take 24 ml from the solvent layer in the separating funnel for zero point checking. " Calibration Setting screen " (page 78 ) " Calibration " (page ( 23 ) Span calibration Perform span calibration using the conditions in the table below. Parameter Span Point Extraction Time Separation Time Number of Purge Span liquid for calibration Condition Concentration of span liquid 0 sec 0 sec 3 or more Use 24 ml of span liquid for calibration. " Calibration Setting screen " (page 78 ) " Span liquid for calibration " (page 20 ) " Calibration " (page ( 23 ) Measurement Perform measurement using the conditions in the table below. Parameter Extraction Time Separation Time Number of Purge Measurement liquid Condition 0 sec 0 sec 3 or more Take 24 ml from the measurement liquid. " Measurement Setting screen " (page 69 ) " Measurement " (page ( 33 ) 52

73 Measurement Oil content in or on solids Directly immerse the part in the solvent to extract the oil content. Although an ultrasonic cleaner can be used for extraction. The solvent volatilization volume and changes in water content will be larger and may have a greater effect on the measured value. In addition, elution and peeling of the part may occur. Extraction by the immersion method is recommended. To obtain a correct measurement result, it is important that the solvent conditions is the same for calibration and measurement. Observe the precautions below. Water is not used for extraction of oil content on metal parts. Use the same lot of solvent for zero calibration and span calibration, which is used for extraction. Do not use water. Solvent volatilization during extraction will affect measurement results. The effect of solvent volatilization can be reduced by performing, in parallel with extraction from the metal part, the same processing using solvent only, and using that solvent to perform zero calibration and span calibration. In particular, if an ultrasonic cleaner is used for extraction, always process the solvent to be used for zero calibration and span calibration with the ultrasonic cleaner as well. Extraction Items required Measuring cylinder (select a size appropriate for the amount of solvent) *1 Wide-mouthed glass container with lid (select a size appropriate for the size of the sample and the amount of solvent) *1 Ashless cellulose quantitative filter paper (particle retention: 8 µm) *2 Glass funnel *1 Glass container *1 Ultrasonic cleaner (if used) Part to be measured (referred to as sample in the rest of this section) Clean solvent S-316 (optional) *1: Clean the glass utensils with solvent in advance and let air dry. *2: Use solvent to elute and to clean organic material from the filter paper in advance, and let air dry. Extraction procedure 1. Place the sample in the wide-mouthed glass container with a lid. 2. Fill an amount of solvent into the measuring cylinder sufficient to immerse the entire sample. 3. Pour the measured solvent into the wide-mouthed container and immediately close the lid. 4. Wait with the lid closed, shaking from time to time, for 1 minute. If needed, the time can be extended. It is possible to apply the sonic cleaner with the lid closed (no more than 3 minutes). 53

74 Measurement Fig. 56 Oil content extraction from part (immersion) Do not shake the container with the solvent and with closed lid hard. If it is shaken hard, the internal pressure will increase due to vaporization of the solvent and the solvent may spray out from gaps under the lid. After shaking, open the lid briefly to lower the internal pressure and then promptly close the lid. Do not apply the ultrasonic cleaner longer than 3 minutes. 5. Filter the liquid through the filter paper to remove foreign matter (particles, peeled fragments, etc.), and use it as the measurement liquid. Zero calibration Perform zero calibration using the conditions in the table below. Parameter Extraction Time Separation Time Number of Purge Zero liquid for calibration Condition 0 sec 0 sec 3 or more Use 24 ml of zero liquid for calibration. Do not introduce water during zero calibration. " Calibration Setting screen " (page 78 ) " Zero liquid for calibration " (page 20 ) " Calibration " (page ( 23 ) Span calibration Perform span calibration using the conditions in the table below. Parameter Span Point Extraction Time Separation Time Number of Purge Span liquid for calibration Condition Concentration of span liquid 0 sec 0 sec 3 or more Use 24 ml of span liquid for calibration. 54

75 Measurement Do not introduce water during span calibration Also set [Span Point] in the calibration settings to the same value as the net concentration of the span liquid, because water is not used. " Calibration Setting screen " (page 78 ) " Span liquid for calibration " (page 20 ) " Calibration " (page ( 23 ) Measurement Perform measurement using the conditions in the table below. Parameter Extraction Time Separation Time Number of Purge Measurement liquid Condition 0 sec 0 sec 3 or more Take 24 ml from the measurement liquid. " Measurement Setting screen " (page 69 ) " Measurement " (page ( 33 ) 55

76 Data Management Data Management Data Top screen The Data Top screen appears when the DATA button is pressed on the Measurement Top screen or Setting Top screen. The Data Top screen shows a menu for data management. Select an item with the up/down button and press the ENT button to move to the screen for the selected function. Fig. 57 Data Top screen Table 5 Menu on the Data Top screen Item Description Page Current Alarm Opens the Current Alarm screen. page 57 Measurement History Opens the Measurement History screen. page 58 Calibration History Opens the Calibration History screen. page 59 USB Memory Opens the USB Memory screen. page 60 Memory Clear Opens the Memory Clear screen. page 65 The buttons and button functions, which can be used with the Data Top screen appearing, are described in the table below. Table 6 Button functions with the Data Top screen Button Function Page CAL button Switches between the zero calibration mode and span calibration mode. page 23 MEAS button Opens the Measurement Top screen. page 11 SET button Opens the Setting Top screen. page 68 ENT button Enters the selected item. - Up button Selects the next item up. - 56

77 Data Management Button Function Page Down button Selects the next item down. - ESC button Returns to the Measurement Top screen. page 11 Current Alarm screen If [Current Alarm] is selected on the Data Top screen, the Current Alarm screen opens. The Current Alarm screen shows a list of the current alarms. Fig. 58 Current Alarm screen For the alarms and their descriptions, refer to " Alarm displays and actions " (page 102). The buttons and button functions, which can be used with the Current Alarm screen appearing, are described in the table below. Table 7 Button functions with the Current Alarm screen Button Function Page Up button Selects the next item up. - Down button Selects the next item down. - Left button Shows the previous page. - Right button Shows the next page. - ESC button Returns to the Data Top screen. page 56 57

78 Data Management Measurement History screen If [Measurement History] is selected on the Data Top screen, the Measurement History screen opens. The Measurement History screen shows a list of the measurement history. Fig. 59 Measurement History screen The buttons and button functions, which can be used with the Measurement History screen appearing, are described in the table below. Table 8 Button functions with the Measurement History screen Button Function Page Up button Selects the next item up. - Down button Selects the next item down. - Left button Shows the previous page. - Right button Shows the next page. - ESC button Returns to the Data Top screen page 56 58

79 Data Management Calibration History screen If [Calibration History] is selected on the Data Top screen, the Calibration History screen opens. The Calibration History screen shows a list of the calibration history. Fig. 60 Calibration History screen The buttons and button functions, which can be used with the Calibration History screen appearing, are described in the table below. Table 9 Button functions with the Calibration History screen Button Function Page Up button Selects the next item up. - Down button Selects the next item down. - Left button Shows the previous page. - Right button Shows the next page. - ESC button Returns to the Data Top screen page 56 59

80 Data Management USB Memory screen If [USB Memory] is selected on the Data Top screen, the USB Memory screen opens. The USB Memory screen shows a menu for USB memory operations. If an operation is selected with the up/down button and the ENT button is pressed, an execution confirmation message for the selected operation appears. To execute the selected operation, press the ENT button while the execution confirmation appears. Use a FAT/FAT32 formatted USB stick. Other formats are not available for this product. HORIBA-recommended USB memory sticks are available. Consult your dealer. To execute the operations with the USB Memory screen appearing, a USB memory stick must be connected to this product. If any of the operations are attempted without inserting a USB memory stick into the USB memory port, a message of "Process has failed" will appear. To connect a USB memory stick, refer to " Connecting a USB memory stick " (page 13 ). Fig. 61 USB Memory screen Table 10 Menu on the USB Memory screen Item Description Page Save Measurement History Shows an execution confirmation for [Save Measurement History]. page 61 Save Calibration History Shows an execution confirmation for [Save Calibration History]. page 62 Save Settings Shows an execution confirmation for [Save Settings]. page 63 The buttons and button functions, which can be used with the USB Memory screen appearing, are described in the table below. Table 11 Button functions with the USB Memory screen Button Function Page ENT button Shows an execution confirmation for the selected item. - Up button Selects the next item up. - Down button Selects the next item down. - ESC button Returns to the Data Top screen page 56 60

81 Data Management Execution confirmation for [Save Measurement History] This is a confirmation message for saving the measurement history to an USB memory stick. Fig. 62 Execution confirmation for [Save Measurement History] The buttons and button functions, which can be used with the execution confirmation for [Save Measurement History] appearing, are described in the table below. Table 12 Button functions with execution confirmation for [Save Measurement History] Button Function Page ENT button Saves the measurement history to an USB memory stick. - ESC button Closes the message. - If the ENT button is pressed, the measurement history is saved to an USB memory stick as a file in CSV format. Each line describes one set of data. The values are separated by commas (","). Saved items and formats are as shown in the table below. Item (1st line is the title) Format Remarks Date yyyy/mm/dd hh:mm:ss Date and time of measurement (year, month, day, hour, minute, second) Value ddddd Concentration after display conversion Unit uuuuu Value (Raw) ddddd Raw concentration value Concentration units after display conversion (mg/l, mg/kg, mg/g, mg/pc) Units (Raw) uuuuu Units of raw concentration value (fixed at mg/l) Status sssss Error flag 0: No error 2: Warming up Memo mmmmmmmmmmmmmmm Entered memo Output example: Date,Value,Unit,Value(Raw),Unit(Raw),Status,Memo 2001/01/01 12:34:56,123,mg/g,123,mg/L,0,sample /01/02 12:34:56,123,mg/g,123,mg/L,2,sample02 61

82 Data Management Execution confirmation for [Save Calibration History] This is a confirmation message for saving the calibration history to an USB memory stick. Fig. 63 Execution confirmation for [Save Calibration History] The buttons and button functions, which can be used with the execution confirmation for [Save Calibration History] appearing, are described in the table below. Table 13 Button functions with execution confirmation for [Save Calibration History] Button Function Page ENT button Saves the calibration history to an USB memory stick. - ESC button Closes the message. - If the ENT button is pressed, the calibration history is saved to an USB memory stick as a file in CSV format. Each line describes one set of data. The values are separated by commas (","). Saved items and formats are as shown in the table below. Item (1st line is the title) Format Remarks Date Standard yyyy/mm/dd hh:mm:ss ddddd Date and time of calibration (year, month, day, hour, minute, second) Calibration concentration value (0 for zero calibration, set span value for span calibration) Unit uuuuu Units of calibration concentration (fixed at mg/l) Status sssss Error flag 0: No error 2: Warming up Output example: Date,Standard,Unit,Status 2001/01/01 12:34:56,0.0,mg/L,0 2001/01/02 12:34:56,200,mg/L,2 62

83 Data Management Execution confirmation for [Save Settings] This is a confirmation message for saving the settings of the main unit to an USB memory stick. Fig. 64 Execution confirmation for [Save Settings] The buttons and button functions, which can be used with the execution confirmation for [Save Settings] appearing, are described in the table below. Table 14 Button functions with execution confirmation for [Save Settings] Button Function Page ENT button Saves the settings of the main unit to an USB memory stick. - ESC button Closes the message. - If the ENT button is pressed, the settings of the main unit are saved to an USB memory stick as a.cfg file. The data ID and data value are indicated as text for each setting item category in the file. Saved items are shown in the table below. Setting item category Item (data ID) Description Page MeasureSetting ExtractTime Extraction time page 70 LayerSeparationTime Separation time page 71 FillSampleTime Cell filling time page 71 MeasureLimitTime Measurement limit time page 72 DrainTime Drainage time page 72 PurgeNum Numbers of purges page 73 StabWaitTime Not used on this product. - MeasMode Measurement mode page 73 ConvertionType Measurement unit page 74 SolventVolume Solvent volume page 74 SampleVolume Sample volume page 75 ZeroShift Zero shift value page 75 fl_extractlight Use of extraction layer light source page 76 MeasTrig Not used on this product. - DataLog Save confirmation setting page 76 fl_memo Memo Saving setting page 77 fl_minus Display setting for negative value page 77 fl_rawdata Display setting for raw data page 78 63

84 Data Management Setting item category Item (data ID) Description Page SpanValue Set span value page 79 ExtractTime Extraction time page 80 CalibrationSetting LayerSeparationTime Separation time page 80 PurgeNum Numbers of purges page 81 CalMode Calibration mode page 81 SystemSetting Language Language setting page 83 An example of a setting file is shown in Fig. 65. Fig. 65 Setting file example 64

85 Data Management Memory Clear screen If [Memory Clear] is selected on the Data Top screen, the Memory Clear screen opens. The Memory Clear screen shows a menu for selection of data to be deleted from internal memory. If an item is selected with the up/down button and the ENT button is pressed, an execution confirmation message for deletion of the selected data appears. To delete the selected data, press the ENT button while the execution confirmation appears. Fig. 66 Memory Clear screen Table 15 Menu on the Memory Clear screen Item Description Page Clear Measurement History Shows an execution confirmation for [Clear Measurement History]. page 66 Clear Calibration History Shows an execution confirmation for [Clear Calibration History]. page 66 Initialize Setting Shows an execution confirmation for [Initialize Setting]. page 67 The buttons and button functions, which can be used with the Memory Clear screen appearing, are described in the table below. Table 16 Button functions with the Memory Clear screen Button Function Page ENT button Shows an execution confirmation for the selected item. - Up button Selects the next item up. - Down button Selects the next item down. - ESC button Returns to the Data Top screen. page 56 65

86 Data Management Execution confirmation for [Clear Measurement History] This is a confirmation message for clearing the measurement history. Fig. 67 Execution confirmation for [Clear Measurement History] The buttons and button functions, which can be used with the execution confirmation for [Clear Measurement History] appearing, are described in the table below. Table 17 Button functions with execution confirmation for [Clear Measurement History] Button Function Page ENT button Clears the measurement history. - ESC button Closes the message. - Execution confirmation for [Clear Calibration History] This is a confirmation message for clearing the calibration history. Fig. 68 Execution confirmation for [Clear Calibration History] The buttons and button functions, which can be used with the execution confirmation for [Clear Calibration History] appearing, are described in the table below. Table 18 Button functions with execution confirmation for [Clear Calibration History] Button Function Page ENT button Clears the calibration history. - ESC button Closes the message. - 66

87 Data Management Execution confirmation for [Initialize Setting] This is a confirmation message for initializing the settings. Fig. 69 Execution confirmation for [Initialize Setting] The buttons and button functions, which can be used with the execution confirmation for [Initialize Setting] appearing, are described in the table below. Table 19 Button functions with execution confirmation for [Initialize Setting] Button Function Page ENT button Initializes the settings. - ESC button Closes the message. - 67

88 Setting Setting Setting Top screen The Setting Top screen appears when the SET button is pressed on the Measurement Top screen or the Data Top screen. The Setting Top screen shows a menu for setting. Select an item with the up/down button and press the ENT button to move to the screen for the selected function. Fig. 70 Setting Top screen Table 20 Menu on the Setting Top screen Item Description Page Measurement Setting Opens the Measurement Setting screen. page 69 Calibration Setting Opens the Calibration Setting screen. page 78 System Setting Opens the System Setting screen. page 82 The buttons and button functions, which can be used with the Setting Top screen appearing, are described in the table below. Table 21 Button functions with the Setting Top screen Button Function Page DATA button Opens the Data Top screen. page 56 CAL button Opens the Measurement Top screen. page 11 MEAS button Opens the Measurement Top screen. page 11 ENT button Enters the selected item. - Up button Selects the next item up. - Down button Selects the next item down. - 68

89 Setting Button Function Page ESC button Returns to the Measurement Top screen. page 11 Measurement Setting screen If [Measurement Setting] is selected on the Setting Top screen, the Measurement Setting screen opens. The Measurement Setting screen shows the current measurement settings. If an item is selected with the up/down button and the ENT button is pressed, a pop-up screen appears to let you change the setting of the selected item. Fig. 71 Measurement Setting screen The setting items on the Measurement Setting screen are shown in the table below. Table 22 Items on the Measurement Setting screen Item Description Setting range (unit) or selections Page Extraction Time Sets the extraction time in the automatic measurement mode. 0 or 10 to 600 (sec) page 70 Separation Time Fill Cell Time Meas. Limit Drainage Time Number of Purge Meas. Mode Measurement Unit Solvent Vol. Sample Vol. Sets the layer separation time in the automatic measurement mode. Sets the cell filling time in the automatic measurement and automatic calibration modes. Sets the maximum measurement time in measurement and calibration. If the measured value does not stabilize, measurement ends when the set time (from the start of measurement or calibration) elapses and the measured value at that time is displayed. Sets the drainage time in the automatic measurement and automatic calibration modes. Sets the numbers of purges in the automatic measurement mode. Sets the measurement mode. Sets the units of measured values. Sets the solvent volume used as a coefficient when the measured value is converted to mg/kg, mg/g, or mg/pc. Sets the sample volume used as a coefficient when the measured value is converted to mg/kg, mg/g, or mg/pc. 0 or 10 to 600 (sec) page to 3600 (sec) page to 3600 (sec) page to 3600 (sec) page 72 0 to 9 (time) page 73 AUTO, SEMI-AUTO, MANUAL mg/l, mg/kg, mg/g, mg/pc page 73 page to (ml) page to (The units depend on the set [Measurement Unit].) page 75 69

90 Setting Item Description Setting range (unit) or selections Page Zero Shift Value Use Light Confirm Save Save Memo Display Negative Display Raw Data Sets the shift correction value for zero liquid. Sets whether the LED that illuminates the inside of the extraction tank is ON or OFF. Sets whether measurement data is automatically saved to internal memory after a measured value is settled or a message appears for execution of the data save. Sets whether a memo indicating the measurement conditions or other information is saved together with the measured values, if measurement data is saved. Sets whether a negative value is displayed or zero is displayed, if the measured value is a negative value. Sets whether or not the concentration prior to conversion is displayed together with the converted value, if converted values are displayed to (mg/l) page 75 OFF, ON page 76 AUTO, MANUAL page 76 OFF, ON page 77 OFF, ON page 77 OFF, ON page 78 The buttons and button functions, which can be used with the Measurement Setting screen appearing, are described in the table below. Table 23 Button functions with the Measurement Setting screen Button Function Page ENT button Shows the screen for setting the selected item. - Up button Selects the next item up. - Down button Selects the next item down. - ESC button Returns to the Setting Top screen. page 68 Extraction Time Use this screen to set the extraction time in the automatic measurement mode. Fig. 72 Extraction Time screen The following settings are available. The default setting is 40 (sec). Setting range Units 0 or 10 to 600 sec 70

91 Setting For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). Separation Time Use this screen to set the layer separation time in the automatic measurement mode. Fig. 73 Separation Time screen The following settings are available. The default setting is 30 (sec). Setting range Units 0 or 10 to 600 sec For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). Fill Cell Time Use this screen to set the cell filling time in the automatic measurement and automatic calibration modes. Fig. 74 Fill Cell Time screen The following settings are available. The default setting is 60 (sec). Setting range Units 30 to 3600 sec For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). 71

92 Setting Meas. Limit Use this screen to set the maximum measurement time in measurement and calibration. If the measured value does not stabilize, measurement ends when the set time (from the start of measurement or calibration) elapses, and the measured value at that point is displayed. Fig. 75 Meas. Limit screen The following settings are available. The default setting is 300 (sec). Setting range Units 60 to 3600 sec For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). Drainage Time Use this screen to set the drainage time in the automatic measurement and automatic calibration modes. Fig. 76 Drainage Time screen The following settings are available. The default setting is 30 (sec). Setting range Units 30 to 3600 sec For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). 72

93 Setting Number of Purge Use this screen to set the numbers of purges in the automatic measurement mode. Fig. 77 Number of Purge screen The following settings are available. The default setting is 2 (time). Setting range Units 0 to 9 time For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). Meas. Mode Use this screen to set the measurement mode. Fig. 78 Meas. Mode screen The following settings are available. The default setting is "AUTO". Selection AUTO SEMI-AUTO MANUAL Description If the MEAS button is pressed, the automatic measurement mode is entered. If the MEAS button is pressed, the semi-automatic measurement mode is entered. If the MEAS button is pressed, the manual measurement mode is entered. For the buttons and functions, which can be used with this screen appearing, refer to Table 2 (page 15 ). 73

94 Setting Measurement Unit Use this screen to set the units of measured values. " Conversion of measurement units " (page 108) Fig. 79 Measurement Unit screen The following settings are available. The default setting is "mg/l". Selection mg/l mg/kg mg/g mg/pc Description Shows measured concentration values in units of mg/l. Shows measured concentration values in units of mg/kg. Shows measured concentration values in units of mg/g. Shows measured concentration values in units of mg/pc. For the buttons and functions, which can be used with this screen appearing, refer to Table 2 (page 15 ). Solvent Vol. Use this screen to enter the solvent volume to be used as a coefficient for conversion of the measurement unit into mg/kg, mg/g, or mg/pc. " Conversion of measurement units " (page 108) Fig. 80 Solvent Vol. screen The following settings are available. The default setting is 8.0 (ml). Setting range Units 1.0 to ml For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). 74

95 Setting Sample Vol. Use this screen to enter the sample volume to be used as a coefficient for conversion of the measurement unit into mg/kg, mg/g, or mg/pc. " Conversion of measurement units " (page 108) Fig. 81 Sample Vol. screen The following settings are available. The default setting is 16.0 (the units depend on the set [Measurement Unit]). Setting range 1.0 to Units The units depend on the set [Measurement Unit] (refer to " Measurement Unit " (page 74 )). kg if [Measurement Unit] is set to mg/kg g if [Measurement Unit] is set to mg/g PC if [Measurement Unit] is set to mg/pc For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). Zero Shift Value Use this screen to set the shift correction value for zero liquid. The sum of this set value and the raw measured value is displayed as the measured value. " Conversion of measurement units " (page 108) Fig. 82 Zero Shift Value screen The following settings are available. The default setting is 0.0 (mg/l). Setting range Units to mg/l For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). 75

96 Setting Use Light Use this screen to set ON or OFF for the LED that illuminates the inside of the extraction tank. Turning ON the LED makes it easier to check liquid amounts and the boundary surface through the extraction tank window. Fig. 83 Use Light screen The following settings are available. The default setting is "ON". OFF ON Selection Description Turns OFF the LED that illuminates the inside of the extraction tank. Turns ON the LED that illuminates the inside of the extraction tank. Even if [Use Light] is set to "ON", the LED will be turned OFF automatically when the period of the set [B-Light Off Time] has passed after the last button operation (refer to " B-Light Off Time " (page 83 )). ). Any button operations turn ON the LED again. For the buttons and functions, which can be used with this screen appearing, refer to Table 2 (page 15 ). Confirm Save Use this screen to set whether saving of settled measured values to internal memory takes place automatically or by manual selection. Fig. 84 Confirm Save screen The following settings are available. The default setting is "AUTO". AUTO Selection MANUAL Description Saving of settled measured values to internal memory takes place automatically. After a measured value is settled, a message appears to let you select whether the values are saved to internal memory. For the buttons and functions, which can be used with this screen appearing, refer to Table 2 (page 15 ). 76

97 Setting Save Memo Use this screen to set whether a memo indicating the measurement conditions or other information is saved together with the measured values, if measurement data is saved. Fig. 85 Save Memo screen The following settings are available. The default setting is "OFF". OFF ON Selection Description If measurement data is saved, only measured values are saved. If measurement data is saved, a memo indicating measurement conditions or other information is saved together with the measured values. For the buttons and functions, which can be used with this screen appearing, refer to Table 2 (page 15 ). Display Negative Use this screen to set whether a negative value is displayed or zero is displayed when the measured value is a negative value. Fig. 86 Display Negative screen The following settings are available. The default setting is "OFF". OFF ON Selection Description If the measured value is a negative value, zero is displayed. If the measured value is a negative value, the negative value is displayed. For the buttons and functions, which can be used with this screen appearing, refer to Table 2 (page 15 ). 77

98 Setting Display Raw Data Use this screen to set whether the concentration prior to conversion is displayed, if converted values are displayed or not. Fig. 87 Display Raw Data screen The following settings are available. The default setting is "OFF". OFF ON Selection Description Raw data is not displayed, if converted values are displayed. Raw data is displayed, if converted values are displayed. For the buttons and functions, which can be used with this screen appearing, refer to Table 2 (page 15 ). Calibration Setting screen If [Calibration Setting] is selected on the Setting Top screen, the Calibration Setting screen opens. The Calibration Setting screen shows the current calibration settings. If an item is selected with the up/down button and the ENT button is pressed, a pop-up screen appears to let you change the setting of the selected item. Fig. 88 Calibration Setting screen The setting items on the Calibration Setting screen are as shown in the table below. Table 24 Items on the Calibration Setting screen Item Description Setting range (units) Page Span Point Sets the concentration of the span liquid for calibration. 1.0 to (mg/l) page 79 Extraction Time Sets the extraction time in the automatic calibration mode. 0 or 10 to 600 (sec) page 80 Separation Time Number of Purge Sets the layer separation time in the automatic calibration mode. Sets the numbers of purges in the automatic calibration mode. 0 or 10 to 600 (sec) page 80 0 to 9 (time) page 81 Calib. Mode Sets the calibration mode. AUTO, MANUAL page 81 78

99 Setting The buttons and button functions, which can be used with the Calibration Setting screen appearing, are described in the table below. Table 25 Button functions with the Calibration Setting screen Button Function Page ENT button Shows the screen for setting the selected item. - Up button Selects the next item up. - Down button Selects the next item down. - ESC button Returns to the Setting Top screen. page 68 Span Point Use this screen to set the concentration of the span liquid for calibration. Fig. 89 Span Point screen The following settings are available. The default setting is 200 (mg/l). Setting range Units 1.0 to mg/l For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). 79

100 Setting Extraction Time Use this screen to set the extraction time in the automatic calibration mode. Fig. 90 Extraction Time screen The following settings are available. The default setting is 40 (sec). Setting range Units 0 or 10 to 600 sec For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). Separation Time Use this screen to set the layer separation time in the automatic calibration mode. Fig. 91 Separation Time screen The following settings are available. The default setting is 30 (sec). Setting range Units 0 or 10 to 600 sec For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). 80

101 Setting Number of Purge Use this screen to set the numbers of purges in the automatic calibration mode. Fig. 92 Number of Purge screen The following settings are available. The default setting is 2 (time). Setting range Units 0 to 9 time For the buttons and functions, which can be used with this screen appearing, refer to Table 3 (page 16 ). Calib. Mode Use this screen to set the calibration mode. Fig. 93 Calib. Mode screen The following settings are available. The default setting is "AUTO". AUTO Selection MANUAL Description If the CAL button is pressed, the automatic calibration mode is entered. If the CAL button is pressed, the manual calibration mode is entered. For the buttons and functions, which can be used with this screen appearing, refer to Table 2 (page 15 ). 81

102 Setting System Setting screen If [System Setting] is selected on the Setting Top screen, the System Setting screen opens. The System Setting screen shows the current system settings. If an item is selected with the up/down button and the ENT button is pressed, a pop-up screen appears to let you change the setting of the selected item. Fig. 94 System Setting screen The setting items on the System Setting screen are as shown in the table below. Table 26 Items on the System Setting screen Item Description Setting range or selections Page Language Sets the language. ENGLISH, JAPANESE, RUSSIAN page 83 B-Light Off Time Sets the time of turning OFF the light of the LCD and extraction tank automatically. OFF, 10 min, 30 min, 60 min page 83 Date Sets the date. 2000/01/01 to 2099/12/31 page 84 Time Sets the current time. 00:00 to 23:59 page 84 Ver. Shows the software version (fixed value) in the right column. - - The buttons and button functions, which can be used with the System Setting screen appearing, are described in the table below. Table 27 Button functions with the System Setting screen Button Function Page ENT button Shows the screen for setting the selected item. - Up button Selects the next item up. - Down button Selects the next item down. - ESC button Returns to the Setting Top screen page 68 The screen does not change, if [Ver.] is selected and the ENT button is pressed. 82

Oil Content Analyzer OCMA-550 Instruction manual CODE:GZ

Oil Content Analyzer OCMA-550 Instruction manual CODE:GZ Oil Content Analyzer OCMA-550 Instruction manual CODE:GZ0000331789 Preface This manual describes the operation of the Oil Content Analyzer, OCMA-550. Be sure to read this manual before using the product

More information

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation

ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS Dionex Corporation ASE SOLVENT CONTROLLER INSTALLATION INSTRUCTIONS 2000 Dionex Corporation Document No. 031277 Revision 03 April 2000 2000 Dionex Corporation All rights reserved worldwide. Printed in the United States of

More information

SkillGuide. User Guide. English

SkillGuide. User Guide. English SkillGuide User Guide English SkillGuide SkillGuide is a feedback device designed to provide real-time and summative feedback on CPR performance. www.laerdal.com Items included SkillGuide and User Guide.

More information

Resusci Anne Simulator

Resusci Anne Simulator EN Resusci Anne Simulator Important Product Information www.laerdal.com Cautions and Warnings A Caution identifies conditions, hazards, or unsafe practices that can result in minor personal injury or damage

More information

User Manual. Bluetooth Pedometer. AmericanPumpkins. version:1.0

User Manual. Bluetooth Pedometer. AmericanPumpkins. version:1.0 version:1.0 AmericanPumpkins User Manual Bluetooth Pedometer LS405-B AMERICAN PUMPKINS, INC 6724 Perimeter Loop Rd # 175 Dublin, Ohio 43017 www.americanpumpkins.com FCC ID: OU9LS405-B01 Thank you very

More information

Portable Gas Monitor GX User Maintenance Manual (H4-0050)

Portable Gas Monitor GX User Maintenance Manual (H4-0050) H4E-0050 Portable Gas Monitor GX-8000 User Maintenance Manual (H4-0050) Need of Maintenance and Servicing This gas monitor must be maintained in a normal state at all times to prevent accidents due to

More information

MAGICFX CONFETTIPISTOL USER AND INSTALLATION MANUAL

MAGICFX CONFETTIPISTOL USER AND INSTALLATION MANUAL English MAGICFX CONFETTIPISTOL USER AND INSTALLATION MANUAL CONFETTI FX / STREAMER FX Manual trigger PART01129 rev 01-00 DISCLAIMER WARNING Read this manual carefully before installing and/or using this

More information

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required)

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required) Quick Setup Guide Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required) Harvest Model number: 100690-BLK / GRY / WHT / RED Harvest 360 Model

More information

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required)

Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required) Quick Setup Guide Welcome to the World of In-Home Gardening! Simple Steps to Get Your AeroGarden Up and Growing (No Tools Required) Harvest Elite Model number: 100691-PPL / BSS / PCP / PSG Harvest Elite

More information

Our technical information is based on Tip: calls attention to special information.

Our technical information is based on Tip: calls attention to special information. SUPPLY TANKS (CBP) 2 General information This manual contains technical information regarding Bayer SeedGrowth Equipment. Please read and understand these instructions completely before proceeding to install

More information

RH800 & 2000 BASIC OPERATIONS

RH800 & 2000 BASIC OPERATIONS RH800 & 2000 BASIC OPERATIONS 2 General information This manual contains technical information regarding Bayer SeedGrowth Equipment. Please read and understand these instructions completely before proceeding

More information

BGA244 Binary Gas Analyzer

BGA244 Binary Gas Analyzer Quick Start Guide Revision 1.0 Certification Warranty Service certifies that this product met its published specification at the time of shipment. This product is warranted against defects in materials

More information

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1)

JOLLY2. Installation user s manual. 6 different operating modes selectable. version 3.3. DATA TO BE FILLED OUT BY THE INSTALLER (Page 1) ENGLISH ENGLISH ENGLISH ENGLISH Installation user s manual Warning! electrical scheme modified JANUARY 2005 version 3.3 JOLLY2 DATA TO BE FILLED OUT BY THE INSTALLER (Page 1) 6 different operating modes

More information

All Purpose AP-Series Operation Manual

All Purpose AP-Series Operation Manual All Purpose AP-Series Operation Manual CARDINAL SCALE MFG. CO. 8525-M176-O1 Rev J 203 E. Daugherty, Webb City, MO 64870 USA Printed in USA 10/00 Ph: 417-673-4631 Fax: 417-673-2153 www.detectoscale.com

More information

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer

OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer July 2011 OPERATOR S MANUAL Ar-Gone Weld Gas Analyzer WARNING! Before operating this product, read and understand this Operator s Manual. Become familiar with the potential hazards of this unit. Contact

More information

6800 Maintenance Instruction System Flush Procedure

6800 Maintenance Instruction System Flush Procedure Equipment Required FA74005 FA65318 FA900005 FA900003 Damper Drain Tube 6800 Cover Removal Tool Beaker 0.25 Litre Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves

More information

PitchTracker User Guide. Model: DKPT01 User Guide

PitchTracker User Guide. Model: DKPT01 User Guide PitchTracker User Guide Model: DKPT01 User Guide PACKAGE CONTENTS What Comes in the Box USING YOUR SMART BALL Turn On Your Smart Ball Pair your Smart Ball Turn Off Your Smart Ball Charge your Smart Ball

More information

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website:

USER MANUAL SAVE THESE INSTRUCTIONS. For the most current manual revision, please visit our Website: USER MANUAL Model: PM4300 Series SAVE THESE INSTRUCTIONS For the most current manual revision, please visit our Website: www.precisionmedical.com 300 Held Drive Tel: (+001) 610-262-6090 Northampton, PA

More information

6900 Maintenance Instruction System Flush

6900 Maintenance Instruction System Flush Equipment Required FA74005 Damper Drain Tube FA16005 Cover Removal Tool FA900005 Beaker 0.25 Litre FA900003 Solvent Cleaning Bottle FA940021 Syringe Polypropylene 50 ml as required FA999045 Gloves Latex

More information

PitchTracker Softball User Guide

PitchTracker Softball User Guide PitchTracker Softball User Guide Model: DKPTS01 User Guide PACKAGE CONTENTS What Comes in the Box USING YOUR SMART SOFTBALL Turn On Your Smart Softball Pair your Smart Softball Turn Off Your Smart Softball

More information

UsER manual for Watersens ph -REDOX

UsER manual for Watersens ph -REDOX UsER manual for Watersens -REDOX Cl 8 1 2 6 3 3 7 7 4 4 4 4 Parts List 1 Redox Probe 1 x 2 PH Probe 1 x 5 Tube Weight 2 x 6 Connection Valve 1 x chlorine 3 Chlorine and Pumps 2 x 7 Dosing Valve 2 x 5 5

More information

MP15 Jockey Pump Controller

MP15 Jockey Pump Controller Setup and Operating Instructions MP15 Jockey Pump Controller This manual provides general information, installation, operation, maintenance, and system setup information for Metron Model MP15 Jockey Pump

More information

Digital Melting Point Apparatus

Digital Melting Point Apparatus Digital Melting Point Apparatus Heating Plateau Ramping Start/Stop Plateau set Ramp stop Hold User Guide Version 1.1 Heating Viewing tube Sample Chamber IEC power inlet socket Power on/off Temperature

More information

The Univentor 1250 Anaesthesia Unit

The Univentor 1250 Anaesthesia Unit THE UNIVENTOR 1200/1250 ANAESTHESIA UNIT The Univentor 1250 Anaesthesia Unit TABLE OF CONTENTS EDITION 1 Section 1 - WARRANTY & SERVICE 1.1. WARRANTY 2 1.2. DAMAGED SHIPMENTS 2 1.3. SERVICE 2 Section 2

More information

Regalia Oxygen Concentrator

Regalia Oxygen Concentrator Regalia Oxygen Concentrator REGALIA INSTRUCTION MANUAL P/N 5167 Rev B January 2014 Table of Contents Warnings and Cautions... 3 Indications for Use... 4 Introduction... 4 Important Safety Instructions...

More information

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289

3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS. Item #31289 3 GALLON, OILLESS PANCAKE COMPRESSOR INSTRUCTIONS Item #31289 The EASTWOOD 3 GALLON, OILLESS PANCAKE COMPRESSOR, with an Integral Air Regulator, efficiently supplies all compressed air requirements for

More information

USER MANUAL. JET TST and JET TST S.

USER MANUAL. JET TST and JET TST S. USER MANUAL JET TST and JET TST S www.sisteven.com Types of Jet fans: JET TST UNI: Impulse axial fan single direction JET TST S UNI: Impulse axial fan single direction (short case) JET TST REV: Impulse

More information

SMART Oxygen Analyzer. User Manual

SMART Oxygen Analyzer. User Manual SMART Oxygen Analyzer User Manual TABLE OF CONTENTS 1 WELCOME... 3 2 NITROXBUDDY2 OVERVIEW... 3 3 WARNINGS... 3 4 BEFORE FIRST USE... 3 5 QUICK GUIDE... 3 6 SETTINGS... 4 6.1 BUTTON... 4 6.2 DISPLAY...

More information

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems

Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems Corporation Issued: 5/09 P/N: 600104 Rev. 2 Instructions for Assembly, Installation, and Operation of the Gas Addition Kit Accessory with the CEM Discover Systems The Gas Addition Accessory permits the

More information

USER S INFORMATION MANUAL

USER S INFORMATION MANUAL USER S INFORMATION MANUAL UPFLOW, DOWNFLOW, UPFLOW/HORIZONTAL & HORIZONTAL ONLY INDUCED DRAFT GAS FURNACES Recognize this symbol as an indication of Important Safety Information If the information in this

More information

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM 4021-PR. Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021-PR Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-PR For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4

More information

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual

Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual Bante810 Benchtop Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante810 Benchtop Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante810 benchtop dissolved oxygen

More information

MODEL 1329 Tank Gauge

MODEL 1329 Tank Gauge SERVICE INSTRUCTIONS FOR PETRO-METER 1329 GAUGE SERVICE INSTRUCTIONS MODEL 1329 Tank Gauge www.petro-meter.com Petro-Meter 1329 series Service Instructions HOW TO DETERMINE IF A PETRO-METER IS IN GOOD

More information

Bante820 Portable Dissolved Oxygen Meter Instruction Manual

Bante820 Portable Dissolved Oxygen Meter Instruction Manual Bante820 Portable Dissolved Oxygen Meter Instruction Manual BANTE INSTRUMENTS CO., LTD Bante820 Portable Dissolved Oxygen Meter 1 Introduction Thank you for selecting the Bante820 portable dissolved oxygen

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP

GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP GETZ EQUIPMENT INNOVATORS PART NO.: 9G59554 MODEL: MS 36 SC-R HYDROSTATIC TEST PUMP LIMITED WARRANTY Getz Equipment Innovators warrants its products, and component parts of any product manufactured by

More information

Owner s Manual. Model H4685

Owner s Manual. Model H4685 Owner s Manual Model H4685 TABLE OF CONTENTS Important Information.................................... 3 Parts.................................................. 4 Battery Installation......................................

More information

Elcometer 138. Bresle Kit and Patches. Operating Instructions

Elcometer 138. Bresle Kit and Patches. Operating Instructions Elcometer 138 English Bresle Kit and Patches Operating Instructions R English The Conductivity Meter model B-173 has been tested in accordance with EU regulations governing Electro-magnetic compliance

More information

Stirring Water Bath. Instruction Manual

Stirring Water Bath. Instruction Manual Stirring Water Bath Instruction Manual Catalog Nos. SWB-10L-1-110 / 220 SWB-10L-2-110 / 220 SWB-20L-1-110 / 220 SWB-20L-3-110 / 220 www.majorsci.com service@majorsci.com Version 01A October 20, 2009 Un

More information

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions

RAM Operation Manual. Worldwide Manufacturer of Gas Detection Solutions RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For Your Safety... 2 Description.... 2 Setup Mode.... 2 Lights/Alarms.... 3 Operation.... 4 Calibration....

More information

English. Introduction. Safety Instructions. All Products. Inspection and Maintenance Schedules. Parts Ordering. Specifications WARNING WARNING

English. Introduction. Safety Instructions. All Products. Inspection and Maintenance Schedules. Parts Ordering. Specifications WARNING WARNING Contents All Products... Gb-1 Control Valves... Gb-2 Control Valve Actuators... Gb-3 Regulators... Gb-3 Relief Valves... Gb-4 Instruments, Switches, and Accessories... Gb-4 Products Covered by Battery

More information

CCT-7320/ROC-2313 Reverse Osmosis Controller

CCT-7320/ROC-2313 Reverse Osmosis Controller CCT-7320/ROC-2313 Reverse Osmosis Controller 1 General The instrument is a combined control instrument of a reverse osmosis controller and an on-line conductivity instrument. It can perform the operation

More information

D10S/D20S Wall/Post Mount Inflator Quick Start Manual

D10S/D20S Wall/Post Mount Inflator Quick Start Manual PART NUMBER SERIAL NUMBER D10S/D20S Wall/Post Mount Inflator Quick Start Manual Please read and save these instructions. Read carefully before attempting to assemble, install, operate or maintain the product

More information

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury.

AIR COMPRESSOR. Failure to follow all instructions as listed below may result in electrical shock, fire, and/or serious personal injury. 2 GALLON AIR COMPRESSOR Model: 7517 DO NOT RETURN TO STORE. Please CALL 800-348-5004 for parts and service. CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials

More information

XpressFill XF2500 Counter Pressure Filler. Operating Instructions

XpressFill XF2500 Counter Pressure Filler. Operating Instructions XpressFill XF2500 Counter Pressure Filler Operating Instructions ii XPRESSFILL Systems LLC lling machine. ank you for choosing our handcrafted bottle filler as the technology to bottle your passion. We

More information

EC214 EC215 - EC215R Bench Conductivity Meters

EC214 EC215 - EC215R Bench Conductivity Meters Instruction Manual EC214 EC215 - EC215R Bench Conductivity Meters http://www.hannainst.com These Instruments are in Compliance with the CE Directives Dear Customer, Thank you for choosing a Hanna Instruments

More information

Budget Range Operators Handbook

Budget Range Operators Handbook Budget Range Operators Handbook BAMBI AIR COMPRESSORS LTD 152 Thimble Mill Lane Heartlands Birmingham B7 5HT United Kingdom Tel: 0121 322 2299 Fax: 0121 322 2297 Email: sales@bambi-air.co.uk www.bambi-air.co.uk

More information

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator

Table of Contents. Operating Instructions. Resource v.2 Conserving Regulator Operating Instructions Table of Contents Resource v.2 Conserving Regulator Safety Information Device Precautions Introduction Product Features Product Specifications Feature Illustrations Set Up Usage

More information

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL

2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL 2 GALLON TWIN STACK AIR COMPRESSOR W/ HOSE REEL Model: 52024 CALIFORNIA PROPOSITION 65 WARNING: You can create dust when you cut, sand, drill or grind materials such as wood, paint, metal, concrete, cement,

More information

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17

Operation Manual. O2 Quickstick. Oxygen Analyzer 08.17 Operation Manual O2 Quickstick Oxygen Analyzer 08.17 If you have any questions on this equipment please contact Technical Support at: Nuvair 1600 Beacon Place Oxnard, CA 93033 Phone: 805-815-4044 FAX:

More information

Read This First. SurfLink Media Controls Overview. Back View

Read This First. SurfLink Media Controls Overview. Back View MEDIA Read This First SurfLink Media can be connected to most media sources, including televisions, radios, and MP3 players. This guide provides easy step-by-step instructions for connecting and using

More information

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank

MODEL NUMBER: PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock or Electrocution: MODEL NUMBER: 20008 200 PSI AIR SOURCE KIT 200 PSI Compressor on 2.0 Gallon 200 PSI Air Tank IMPORTANT: It is

More information

PTG100 Precision Test Gauge

PTG100 Precision Test Gauge PTG100 Precision Test Gauge User Manual PD1007 Rev B 03/28/2014 Palmer Instruments Inc. 234 Old Weaverville Road Asheville, NC 28804 Toll Free: 800-421-2853 Phone: 828-658-3131 Fax: 828-658-0728 Email:

More information

QUICK START GUIDE TO DRINKING WATER MONITORING

QUICK START GUIDE TO DRINKING WATER MONITORING APPENDIX F QUICK START GUIDE TO DRINKING WATER MONITORING Introduction: This Quick Start guide is designed to allow users who are already familiar with the basic use of YSI 6-series sondes for surface

More information

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE

MODEL NUMBER: M20005 AIR SOURCE KIT. 30% Duty Compressor on. 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE MODEL NUMBER: M20005 AIR SOURCE KIT 30% Duty Compressor on 2.0 Gallon Air Tank SAVE THIS MANUAL FOR FUTURE REFERENCE USER MANUAL IMPORTANT SAFETY INSTRUCTIONS CAUTION - To reduce risk of electrical shock

More information

HI 2314 HI 2315 HI 23151

HI 2314 HI 2315 HI 23151 Instruction Manual HI 2314 HI 2315 HI 23151 Multi-Range Conductivity Meters for Laboratories www.hannainst.com Dear Customer, Thank you for choosing a Hanna Instruments product. Please read this instruction

More information

TABLE OF CONTENTS. INTRODUCTION to the AirVAC... 2 SAFETY INFORMATION... 3 DIMENSIONS & SPECIFICATIONS... 5 UNIT OVERVIEW... 6

TABLE OF CONTENTS. INTRODUCTION to the AirVAC... 2 SAFETY INFORMATION... 3 DIMENSIONS & SPECIFICATIONS... 5 UNIT OVERVIEW... 6 TABLE OF CONTENTS INTRODUCTION to the AirVAC... 2 SAFETY INFORMATION... 3 DIMENSIONS & SPECIFICATIONS... 5 UNIT OVERVIEW... 6 OPERATING PROCEDURE... 7 REPLACEMENT PARTS... 8 MAINTENANCE PROCEDURES... 11

More information

OPERATING AND MAINTENANCE MANUAL GOLD LINE EXTRACTION CLEANERS M14-M26

OPERATING AND MAINTENANCE MANUAL GOLD LINE EXTRACTION CLEANERS M14-M26 OPERATING AND MAINTENANCE MANUAL GOLD LINE EXTRACTION CLEANERS M14-M26 Made For: Spitwater Australia Pty Ltd 953 Metry St North Albury, NSW, Australia WARNING: FAILURE TO FOLLOW OPERATING, SAFETY AND MAINTENANCE

More information

User Manual Pocket Pro + Multi 1

User Manual Pocket Pro + Multi 1 User Manual Pocket Pro + Multi 1 DOC022.53.80398 1 Specifications Specification Dimensions (W x D x H) Enclosure rating Weight Specifications are subject to change without notice. Details 37 x 30 x 170

More information

ELECTRONIC PRINTING CALCULATOR OPERATION MANUAL CONTROLS AND OPERATIONS 5 CALCULATION EXAMPLES P a g e

ELECTRONIC PRINTING CALCULATOR OPERATION MANUAL CONTROLS AND OPERATIONS 5 CALCULATION EXAMPLES P a g e ELECTRONIC PRINTING CALCULATOR OPERATION MANUAL CONTROLS AND OPERATIONS 5 CALCULATION EXAMPLES..24 1 P a g e WARNING FCC Regulations state that any unauthorized changes or modifications to this equipment

More information

OWNER S MANUAL. Page: 1 of 8

OWNER S MANUAL. Page: 1 of 8 Air Needle Scaler OWNER S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may result in serious personal

More information

Armfield Distillation Column Operation Guidelines

Armfield Distillation Column Operation Guidelines Armfield Distillation Column Operation Guidelines 11-2016 R.Cox Safety SAFETY GLASSES ARE REQUIRED WHEN OPERATING THE DISTILLATION COLUMN Wear gloves when mixing alcohol feedstock The column will become

More information

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR

INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR INSTRUCTION MANUAL MANUAL INFLATION BLOOD PRESSURE MONITOR Model HEM-412C TABLE OF CONTENTS Introduction 3 Know Your Unit 4 Quick Reference Guide 5 Battery Installation/Replacement 6 How To Apply The Arm

More information

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax

Endo-Flush Order # ZUTR30004 OPERATION MANUAL. Zutron Medical, LLC W 98 th St #40-27 Lenexa, KS Phone Fax OPERATION MANUAL Zutron Medical, LLC 17501 W 98 th St #40-27 Lenexa, KS 66219 Phone 877-343-5873 Fax 913-967-5944 ZUT-Lab-004-30004 REV. 03312017 Table of Contents 2 Introduction 1. Intended Use 2. Labels,

More information

DC5A. Cyclone Separator Trap for Air. Copyright 2015 by TLV Co., Ltd. All rights reserved ISO 9001/ ISO MA-03 (DC5A) 19 June 2015

DC5A. Cyclone Separator Trap for Air. Copyright 2015 by TLV Co., Ltd. All rights reserved ISO 9001/ ISO MA-03 (DC5A) 19 June 2015 172-65205MA-03 (DC5A) 19 June 2015 ISO 9001/ ISO 14001 Manufacturer Kakogawa, Japan is approved by LRQA LTD. to ISO 9001/14001 Cyclone Separator Trap for Air DC5A Copyright 2015 by TLV Co., Ltd. All rights

More information

User Manual MFCS TM -EZ

User Manual MFCS TM -EZ User Manual MFCS TM -EZ Version 6A_December 2016 1. INTRODUCTION 5 2. GENERAL INFORMATION 6 2.1 General functioning 6 2.2 Different MFCS TM 6 3. PACKAGE CONTENT 8 4. MFCS TM -EZ DESCRIPTION 9 5. CONNECTIONS

More information

Operator s Manual. The Bullet Blender Gold BB24-AU, BB5E-AU

Operator s Manual. The Bullet Blender Gold BB24-AU, BB5E-AU Operator s Manual The Bullet Blender Gold BB24-AU, BB5E-AU Congratulations! Congratulations on your purchase of a Bullet Blender Gold by Next Advance, Inc., for lysing, disrupting, and homogenizing your

More information

Safety Powder Spray Systems

Safety Powder Spray Systems Instruction Sheet P/N 107 952C Safety Powder Spray Systems 1. Introduction This section contains general safety instructions for using your Nordson equipment. Task- and equipment-specific warnings are

More information

Oxygen Concentrator Instruction

Oxygen Concentrator Instruction WARNING-Read instruction before operating this equipment Oxygen Concentrator Instruction K5BW WARNING Oxygen therapy can be hazardous in certain conditions. Seeking medical advice before using an oxygen

More information

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064

AIR HAMMERS MODEL NO: CAT138/CAT139 OPERATING & MAINTENANCE INSTRUCTIONS PART NO: / GC064 AIR HAMMERS MODEL NO: CAT138/CAT139 PART NO: 3120152 /3120153 OPERATING & MAINTENANCE INSTRUCTIONS GC064 INTRODUCTION Thank you for purchasing this CLARKE Air Hammer. Before attempting to use this product,

More information

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual

ORB-400 BUBBLE/HAZE MACHINE. Item ref: UK User Manual ORB-400 BUBBLE/HAZE MACHINE Item ref: 160.462UK User Manual Caution: Please read this manual carefully before operating Damage caused by misuse is not covered by the warranty Introduction Thank you for

More information

Composite Pistol-Type Air Needle Scaler OWNER S MANUAL

Composite Pistol-Type Air Needle Scaler OWNER S MANUAL Composite Pistol-Type Air Needle Scaler OWNER S MANUAL WARNING: Read carefully and understand all INSTRUCTIONS before operating. Failure to follow the safety rules and other basic safety precautions may

More information

VS18/26 with PROFINET and EtherNet/IP Interface. ATEX Installation Instructions

VS18/26 with PROFINET and EtherNet/IP Interface. ATEX Installation Instructions VS18/26 with PROFINET and EtherNet/IP Interface ATEX Installation Instructions INDEX 1. INTENDED USAGE 3 2. OPERATING MANUAL ATEX 4 2.1 General conditions 4 2.2 Installation 5 2.3 Operating 5 2.4 Failures

More information

Model : LMW-9620 Please read this manual carefully before using the instrument Labnics Equipment

Model : LMW-9620 Please read this manual carefully before using the instrument Labnics Equipment MICROPLATE WASHER User Manual Model : LMW-9620 Please read this manual carefully before using the instrument Labnics Equipment Table of Content Sr.No. Contents Pg. No. 1 INTRODUCTION 1 1.1 Working Conditions

More information

Operation Manual. O2 Quickstick. Oxygen Analyzer

Operation Manual. O2 Quickstick. Oxygen Analyzer Operation Manual O2 Quickstick Oxygen Analyzer 01.11 www.nuvair.com If you have any questions on this equipment please contact Technical Support at: Nuvair 2949 West 5 th St. Oxnard, CA 93030 Phone: 805-815-4044

More information

RAM 4021 Operation Manual

RAM 4021 Operation Manual RAM 4021 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021 For your safety...3 Description...3 Set-up mode...4 Annunciator lights/alarms...4 Operation...5 Calibration...6

More information

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST

AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST AIR COMPRESSOR OPERATING INSTRUCTION AND PARTS LIST OIL-LESS TYPE IMPORTANT: PLEASE READ CAREFULLY BEFORE STARTING OPERATIONS. THE CONTENTS ARE FOR GENERAL INFORMATION OF ALL THE SIMILAR MODELS. Record

More information

! Warning, refer to accompanying documents.

! Warning, refer to accompanying documents. About this Manual To the best of our knowledge and at the time written, the information contained in this document is technically correct and the procedures accurate and adequate to operate this instrument

More information

SMART Carbon Monoxide Analyzer. User Manual

SMART Carbon Monoxide Analyzer. User Manual SMART Carbon Monoxide Analyzer User Manual TABLE OF CONTENTS 1 WELCOME... 3 2 MONOX OVERVIEW... 3 3 WARNINGS... 3 4 BEFORE FIRST USE... 3 5 QUICK GUIDE... 3 6 SETTINGS... 4 6.1 BUTTON... 4 6.2 DISPLAY...

More information

micro CA-100 Inspection Camera

micro CA-100 Inspection Camera micro CA-100 Operator s Manual micro CA-100 Inspection Camera! Read this Operator s Man ual carefully before using this tool. Failure to understand and follow the contents of this manual may result in

More information

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual

DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual BANTE INSTRUMENTS CO., LTD DOscan10 Pocket Dissolved Oxygen Tester 1 Thank you for selecting the DOscan10 pocket dissolved oxygen tester. This

More information

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc.

Calibration Gas Instrument INSTRUCTION MANUAL. Release I. Advanced Calibration Designs, Inc. Advanced Calibration Designs, Inc. Calibration Gas Instrument INSTRUCTION MANUAL Release I www.goacd.com Instruction Manual Gas Generator Release I TABLE OF CONTENTS I. General Description Page 2 II. Start-Up

More information

STAGE, STADIUM & X-TREME SHOT.

STAGE, STADIUM & X-TREME SHOT. USER AND INSTALLATION MANUAL ENGLISH STAGE, STADIUM & X-TREME SHOT. CONFETTI & STREAMER FX PART01339 rev 01-00 DISCLAIMER WARNING Read this manual carefully before installing and/or using this product.

More information

SPECIFICATIONS PARTICLE SENSOR KS-19F Higashimotomachi, Kokubunji, Tokyo , Japan

SPECIFICATIONS PARTICLE SENSOR KS-19F Higashimotomachi, Kokubunji, Tokyo , Japan SPECIFICATIONS PARTICLE SENSOR KS-19F 3-20-41 Higashimotomachi, Kokubunji, Tokyo 185-8533, Japan No. 13012-3E 17-06 Printed in Japan Outline The KS-19F is a sensor which uses the light scattering method

More information

RAM 4021-DPX Operation Manual

RAM 4021-DPX Operation Manual RAM 4021-DPX Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS ABL 4021-DPX / RAM 4021-DPX For Your Safety... 3 Description... 3 Setup Mode... 4 Lights/Alarms... 4 Operation...

More information

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures

TESCOM 50-4X Series Safety, Installation & Start-Up Procedures Operations & Service Manual TESCOM 50-4X Series Safety, Installation & Start-Up Procedures Do not attempt to select, install, use or maintain this product until you have read and fully understood this

More information

RD(H)20/25 Pressure-Reducing Regulator User Manual

RD(H)20/25 Pressure-Reducing Regulator User Manual RD(H)20/25 Pressure-Reducing Regulator User Manual Read the complete manual before installing and using the regulator. 2 Safe Product Selection When selecting a product, the total system design must be

More information

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road

PROPORTIONING VALVE. Model 150 INSTRUCTION MANUAL. March 2017 IMS Company Stafford Road PROPORTIONING VALVE Model 150 INSTRUCTION MANUAL March 2017 IMS Company 10373 Stafford Road Telephone: (440) 543-1615 Fax: (440) 543-1069 Email: sales@imscompany.com 1 Introduction IMS Company reserves

More information

In Vivo Scientific, LLC INSTRUCTION MANUAL

In Vivo Scientific, LLC INSTRUCTION MANUAL CO 2 Controller In Vivo Scientific, LLC INSTRUCTION MANUAL CONTENTS CONTENTS...1 ABOUT THIS MANUAL...2 INTRODUCTION...2 Cautions and Warnings...2 Parts List...2 Unpacking...2 INSTRUMENT DESCRIPTION...3

More information

Scoreboard Operator s Instructions MPCW6 Control

Scoreboard Operator s Instructions MPCW6 Control Scoreboard Operator s Instructions MPCW6 Control Horn Misc. Shot Time Sub. Horn Set Model Code 134 Basketball Time Out Timer Start Clear Options Yes Confirm Home T.O. Left Next Poss. Bonus Time 7 8 9 No

More information

Instruction Manual for Fluorocarbon Analyzer (Identifier) TA400A

Instruction Manual for Fluorocarbon Analyzer (Identifier) TA400A Instruction Manual for Fluorocarbon Analyzer (Identifier) TA400A TASCO JAPAN CO., LTD. Preface This product is designed to measure component ratio of refrigerants (fluorocarbon) for freezing and airconditioning

More information

xx TA-CMI. User Manual

xx TA-CMI. User Manual 307 161-50 2004.xx User Manual Contents Contents Guarantee 2 FCC Notification and ETL markings 2 General 3 Unpacking 4 Important information 5 Radio frequency communication 5 Storage recommendations 5

More information

Norrsken Family Booklet

Norrsken Family Booklet Section 1: Introduction Low Energy Designs produce efficient and effective LED based lighting products for commercial, retail and industry purposes. Each product may contain specific details on its operation

More information

General Safety Rules

General Safety Rules General Safety Rules 1. The goggles provided by the university are to be properly worn at all times in the lab. 2. You are to be properly dressed in the lab. a. Close-toed shoes. b. At least a short sleeve

More information

Oxygen Sensor with Electrode

Oxygen Sensor with Electrode imagine explore learn Oxygen Sensor with Electrode Product Number: ENOXY-A222 Overview The Oxygen sensor is a maintenance-free galvanic oxygen electrode, capable of measuring Oxygen both in the air and

More information

200 PSI HIGH-FLOW AIR SOURCE KIT

200 PSI HIGH-FLOW AIR SOURCE KIT 200 PSI HIGH-FLOW AIR SOURCE KIT 50% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20008 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of

More information

RAM Operation Manual

RAM Operation Manual RAM 4021-1 Operation Manual Worldwide Manufacturer of Gas Detection Solutions TABLE OF CONTENTS RAM 4021-1 For Your Safety... 2 Description... 2 Setup Mode... 3 Lights/Alarms... 3 Operation... 4 Calibration...

More information

200 PSI FAST-FILL AIR SOURCE KIT

200 PSI FAST-FILL AIR SOURCE KIT 200 PSI FAST-FILL AIR SOURCE KIT 55% Duty Compressor on 2.0 Gallon Air Tank PART NO. 20007 IMPORTANT: It is essential that you and any other operator of this product read and understand the contents of

More information

User Manual. Quantos Automated Dosing Liquid Module

User Manual. Quantos Automated Dosing Liquid Module User Manual Quantos Automated Dosing Liquid Module Overview Liquid module 4 2 1 3 5 6 7 Legend Liquid Module Pump module with bottle Liquid kit with liquid dosing head 1 Pump module (QL2) 4 Top glass

More information

Standard Operating Procedure (SOP)

Standard Operating Procedure (SOP) Microwave Digester #1 General Process Description Standard Operating Procedure (SOP) This instrument is used to digest small amounts of material for chemical analysis (ICP), by placing the sample in a

More information